Generic Verification Protocol for Induction Mixers
Used for High Rate Disinfection of Wet Weather Flows

Draft 3.4

For

NSF International
P.O. Box 130140
Ann Arbor, Ml 48113-0140

By

Moffa & Associates

PO Box 26
5710 Commons Park
Syracuse, NY 13214

June 2000


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

1 Introduction

1.1	Purpose

The purpose of this document is to establish a protocol for the verification of induction
mixers intended for use in the chemical disinfection of wet weather flows, such as
combined sewer overflows and sanitary sewer overflows. The protocol established here
will serve as the basis for verification of commercially-available induction mixers under
the Environmental Technology Verification (ETV) Program of the United States
Environmental Protection Agency (EPA). This protocol describes the steps to be
followed to ensure that verification is carried out in a consistent and objective manner
that assesses the relevant performance characteristics of an induction mixer. It
describes, in general terms, the process of selecting and documenting the verification
tests to be conducted, and outlines the methodology to be employed. The protocol
provides guidelines for the preparation of a Verification Test Plan that is specific to a
given mixer and the organization conducting the test. Guidance is also provided on the
execution of testing and data reduction, analysis and reporting.

1.2	Scope

This Protocol is intended to apply to induction mixers with a motor speed greater than
3,000 rpm and which are designed for submerged service in wet weather flows such as
combined sewer overflows (CSO) and sanitary sewer overflows (SSO).

1.3	The ETV Program and the Wet Weather Flow Technologies Pilot

The Environmental Technology Verification (ETV) program was established to promote
the marketplace acceptance of commercial-ready environmental technologies. The
purpose is to provide credible third-party performance assessments of environmental
technologies so that users, developers, regulators, and consultants can make informed
decisions about such technologies. ETV is not an approval process, but rather provides
a quantitative assessment of technology performance as dtermined in accordance with
an established test protocol . Twelve ETV "Pilots" were established to verify innovative
technologies covering the range of environmental media. Each Pilot is administered

1


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

through a cooperative agreement between EPA and a Verification Partner
Organization.

The Wet Weather Flow Technologies Pilot was established to verify commercially
available technologies used in the control and abatement of urban storm water runoff,
combined sewer overflows and sanitary sewer overflows. NSF International (NSF) is
the verification partner organization on the Wet Weather Flow Technologies Pilot and
administers the WWF Pilot in cooperation with the Urban Watershed Management
Branch of the US EPA's National Risk Management Research Laboratory.

A Stakeholder Advisory Group (SAG) was formed to assist NSF and EPA in
establishing priorities for the verification of wet weather flow technologies. The SAG
consists of technology vendors, state and federal regulatory and permitting officials,
technology users (POTWs and other municipal government staff), and technology
enablers (e.g., consulting firms and universities) with an interest in the assessment and
abatement of the impacts of wet weather flows. The SAG identified high-rate
disinfection technologies as a priority for verification given their potential role in abating
the frequency and extent of pathogen discharges to surface waters associated with
combined sewer overflows.

The Technology Panel on High Rate Disinfection was established to guide the
development of a protocol for the verification of high rate disinfection technologies,
including induction mixers. NSF contracted with Moffa & Associates to develop a draft
verification protocol.

1.4 Technology Applications and Description

Experience has shown that the long disinfection contact time required for conventional
wastewater treatment is not appropriate for the disinfection of CSO due to the
infrequent peak flow rates that would require large tankage. However, disinfection of
CSO can be achieved with less contact time by providing an increased disinfection

2


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

dosage and intense mixing. Mechanical mixers at one or more disinfectant locations
can accomplish such mixing.

The induction type mixer produces relatively intense mixing due to the impeller speed.
The induction mixer was originally designed to inject chemicals into process water.
The fundamental principal is to rotate an impeller at a sufficient rpm as to cause a
vacuum behind the impeller. This vacuum pressure is then used to draw chemicals to
the impeller. Induction mixing is unique in that it blasts the wastewater with a fine spray
of chemical disinfectant. This provides an almost instantaneous dispersion due to the
high rotation of the impeller. Such dispersion allows the most effective forms of the
chemical disinfectant to react with bacteria, which greatly improves the disinfection
process. However, there are still some questions regarding the appropriate size and
effectiveness of such mixers. Therefore the ETV program will focus on induction
mixers and verifying volume of water affected by the mixer.

Using an induction mixer for CSO disinfection has proven to be economical (Moffa &
Associates, 1997). Because CSO facilities operate intermittently, reducing capital
costs by possibly incurring higher operation and maintenance is economically viable.
Using high-rate disinfection with induction mixing reduces contact times from 15-30
minutes to 5 minutes (Moffa & Associates, 1997). The reduced construction cost for
the smaller basins generally offset any higher operation and maintenance costs.

High rate disinfection is governed by the relationship:

Kill = Fn (cxGxt)

Where:	c = concentration of disinfectant

G = mixing intensity

t = time of contact (within a contained volume)

3


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

The mixing intensity is a function of the power imparted into a volume of water. Mean
velocity gradient (G) is a measure of mixing intensity and has become an industry
standard for representing the fluid mechanics of mixing. It is directly related to the
total shear per unit volume per unit of time. The G number gives an indication of
turbulence as it relates to head loss, which in turn relates to mixing (White, 1992). The
velocity gradient is therefore a parameter of disinfection efficiency. G can be
expressed by the following equation:

G = (P/ u*V)1/2

Where P is the power requirement, V is volume of affected process water, and u is the
absolute fluid viscosity. The mean velocity gradient for a typical well designed diffuser
grid system is on the order of 200-500/sec. G.C. White used to believe that the G
number should be approximately 1,000/sec for superior mixing. Some of White's more
recent research indicates that a G number between 700 and 1.000/sec may be
appropriate regardless of disinfection requirements. (White, 1992).

Collins and Kruse (EPA-670/2-73-077) have demonstrated the influence of mixing
intensity on bacterial kills and formation of chloramines with Cl2. When chlorine or
hypochlorite is added to wastewater containing ammonia, the free chlorine will react to
form chloramines. The rate of bactericidal efficacy of chloramines is significantly less
than that of free chlorine. It is theorized that by instantaneously dispersing hypochlorite
in the wastewater stream using high-rate mixing, more of the organisms in the
wastewater can come into contact with chlorine in its free form prior to the formation of
chloramines and, therefore, result in greater kills.

In the past, researchers have related bacterial reductions to Gt. This relationship holds
true when the mixing devices are operated in a mixing chamber of fixed size. This
relationship does not hold true when the mixing devices are operated in an open
channel, allowing the mixing zone volume to change as a result of horsepower (Hp)

4


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

and channel geometry. Alternatively, the extent of mixing can be directly measured by
chemical concentrations at defined locations downstream of the mixer.

1.5 Experimental Objective

A manufacturer of an induction- type mixer may make claims about the mixing
capabilities of its product and its ability to provide rapid, uniform chemical transfer
resulting in reduction or elimination of chemical breakout and stratification. Since these
claims are subjective, often the manufacturers will provide a G factor for a specific
induction mixer application. However there is not a standard method or approach used
for calculating this G factor. As presented in the previous section, G is a function of the
mixer power and the volume of the affected process water. The power and viscosity
variables are standard and therefore the manufacturers use consistent values, but each
manufacturer defines the volume variable differently. As a result each manufacturer
may claim a different G, based on their definition of volume. Currently, there is no
standard for representing volume.

In response to the manufacturer's claims regarding "rapid uniform chemical transfer"
and the inconsistencies in calculating G, this ETV protocol establishes a method for the
determining the volume of process water affected by the induction mixer. The testing
will be performed in a hydraulic laboratory setting. During the verification testing, the
high-rate induction mixers will be operated as though they were installed in a
disinfection facility. However, instead of mixing a chemical disinfectant into the process
water, a dye or conservative tracer chemical will be used in a clean water matrix. The
dye or conservative tracer chemical will allow the researcher to observe the extent of
mixing provided by the high-rate induction mixers.

The objective of this ETV protocol is to evaluate the effectiveness of induction mixers
based on their ability to transfer chemicals into the process water. The volume of water
affected by the mixer (i.e. mixing zone) will define the effectiveness of that particular
mixer. The velocity of the process water has the greatest influence on the induction
mixer's ability to transfer chemicals into the water. This ETV protocol establishes a

5


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

method for verifying the performance of induction mixers over a range of flow velocities.
This range of velocities will represent the typical range of flow conditions at wet-
weather treatment facilities.

1.6	Technical Approach

The general approach will be to define the mixing zone volume by introducing a dye or
conservative tracer chemical at the point of the impeller and measuring the dye or
tracer concentration around and just downstream the impeller. The measured dye or
tracer concentrations will then be used to define the volume of water affected by the
mixer. This approach will be performed for a combination of different flow velocities
and mixing times for each induction mixer tested.

The transfer of chemicals into the process water is a function of mechanical dispersion
and molecular diffusion. Mechanical dispersion is a function of the induction mixer and
the velocity of the process water. Molecular diffusion is a function of chemicals moving
from high concentration to low concentration. In the case of induction mixers, the
mechanical dispersion is several orders of magnitude greater than molecular diffusion,
and therefore molecular diffusion can be disregarded. Some inference between the
mechanical dispersion from the induction mixer and the process water velocity will be
made based on the tests at different velocities.

1.7	Verification Process

The process of verification of induction mixers under the ETV Program consists of three
primary phases as described below:

Planning - The planning phase involves establishing and documenting the procedures
to be followed during the verification of a specific induction mixer, including identifying
the testing laboratory and personnel responsible for performance and oversight of the
testing. The planning phase culminates in the preparation of a product-specific
Verification Test Plan by a field testing organization and its approval by NSF and EPA.
Guidelines for this phase are described in Section 2 of this Protocol.

6


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

Verification Testing - This phase involves establishing the required test conditions,
conducting the required tests, and the collection of the relevant data. Guidelines for this
phase are described in Section 3 of this Protocol.

Data Assessment and Reporting - This last phase includes all data analysis and the
preparation and dissemination of a Verification Report and Verification Statement.
Guidelines for this phase are described in Section 4 of this Protocol.

2 Development of a Verification Test Plan
2.1 Purpose of a Test Plan

Prior to the start of verification testing of an induction mixer under the ETV Program,
the field testing organization (FTO) shall prepare a Verification Test Plan that clearly
describes how and by whom testing is to be conducted. An adequate Test Plan will
help to ensure that testing is conducted and that the results are reported in a manner
consistent the requirements specified in this Protocol. A good Test Plan also ensures
that information about a vendor's mixer or series of mixers is available for incorporation
into a Verification Report upon the completion of testing. An individual Test Plan should
be developed for each mixer, or series of mixers, undergoing verification testing.

At a minimum a Test Plan for the verification of an induction mixer shall include:

•	An introduction that briefly describes the objectives of verification testing and an
overview of approach taken in this study;

•	Roles and responsibilities of participants in the verification testing of the mixer;

•	A complete description of the mixer(s) and its (their) intended functions and
capabilities;

•	A description of the site(s) where verification testing is to take place (i.e., the
hydraulic laboratory)

7


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

•	A description of the experimental design that includes the specific test procedures
to be followed and identifies any necessary deviations from the requirements
established in this Protocol;

•	A description of the Quality Assurance/Quality Control procedures to be employed
to ensure data quality objectives are met;

•	A description of how data is to be analyzed, managed, and reported

•	Health and safety procedures

Subsections 2.2 through 2.8 of this protocol establishes guidelines and requirements
for the content and scope of each section required in a test plan.

2.2 Roles and Responsibilities of Involved Organizations

A Test Plan shall specify the names and addresses of each organization having a role
in the verification of a mixer. Where possible, the Test Plan should include the names,
titles, and contact information, for specific individuals with designated roles in the
verification of the mixer. General guidelines on the roles and responsibilities for the
primary participants are listed below.

2.2.1 NSF International

NSF is the US EPA's verification partner on the Wet Weather Flow Technologies Pilot.
In the context of this Verification Protocol, NSF will select a qualified Testing
Organization to develop and implement a Test Plan. In addition, NSF International has
the following responsibilities:

•	Review and approval of the Test Plan;

•	Oversight of Quality Assurance, including the performance of technical system and
data quality audits, as prescribed in the Quality Management Plan for the Wet
Weather Flow Technologies ETV Pilot;

•	Coordination of Verification Report peer reviews, including review by the
Stakeholder Advisory Group and Technology Panel;

•	Approval of Verification Report; and

8


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

•	Preparation and dissemination of Verification Statement.

2.2.2	US Environmental Protection Agency (EPA)

The US EPA's National Risk Management Research Laboratory provides
administrative, technical and quality assurance guidance and oversight on all WWF
pilot activities. EPA personnel are responsible for the following:

•	Review and approval of Test Plan;

•	Review and approval of Verification Report;

•	Review and approval of Verification Statement; and

•	Posting of Verification Report and Statement on EPA Website.

2.2.3	Field Testing Organization

The Field Testing Organization (FTO) shall have experience with the use of high-rate
induction mixers and also have experience with pilot testing and experimental design.
The FTO will be the main consultant in charge of developing and implementing the Test
Plan. The responsibilities of the FTO may include but are not limited to the following:

•	Preparation of the site-specific Verification Test Plan, including its revision in
response to comments made during the review period;

•	Coordinating with the Manufacturer of the mixer;

•	Contracting with the hydraulic laboratory, analytical laboratory, general
contractor, and any other sub-consultants necessary for implementation of the
approved Test Plan;

•	Providing needed logistical support to the sub-consultants, as well as
establishing a communication network, and scheduling and coordinating the
activities for the verification testing;

•	Overseeing or conducting the verification testing as per the approved Test Plan;

•	Managing, evaluating, interpreting and reporting on data generated during the
verification testing;

•	Preparation and review of a Draft Verification Report.

9


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

2.2.4	Vendor

The vendor shall be a manufacturer of high-rate induction mixers. The vendor's
responsibilities may include but are not limited to the following:

•	Providing mixers and ancillary equipment required for the verification testing,

•	Providing technical support for the installation and operation of the mixer;
including the designation of at least one staff person as the point of contact;

•	Providing descriptive details about the capabilities and intended function of the
mixer;

•	Review and approval of the Verification Test Plan prior to the start of testing;

•	Review and comment on the Draft Verification Report and Verification
Statement.

2.2.5	Hydraulic Laboratory

The hydraulic laboratory shall have experience with adjusting flume dimensions to
provide flow rates required for the verification testing. Requirements for the facilities at
the hydraulic laboratory are described in Section 2.4 of this Protocol. The hydraulic
laboratory may or may not be owned and operated by the FTO. The responsibilities of
the hydraulic laboratory may include but are not limited to the following:

•	Providing flume(s) with the required dimensions.

•	Providing process water to achieve the required flow rates.

•	Measuring, evaluating and reporting on flow rates during the verification testing.

•	Providing an electrical supply sufficient enough to supply the high-rate induction
mixers and sampling equipment.

•	Providing sampling rigs.

•	Coordinating and taking the required samples. This also shall include
transporting the samples to the analytical laboratory.

•	Providing QA/QC documentation for flow rate and sampling.

10


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

2.2.6	Analytical Laboratory

The analytical laboratory shall have experience measuring the dye or tracer chemical.
The responsibilities of the analytical laboratory may include but are not limited to the
following:

•	Providing the instrumentation to measure the approved dye or tracer chemical.

•	Coordinating and measuring the approved dye or tracer chemical.

•	Measuring, evaluating and reporting on the dye or tracer chemical analyses.

•	Providing QA/QC documentation for sample analysis.

2.2.7	General Contractors

One or more general contractors may be needed for the installation of the induction
mixer and ancillary equipment. The general contractor shall have experience with
installing mixing devices in wastewater applications.

2.2.8	Technology Panel on High-Rate Disinfection

The ETV Technology Panel on High-Rate Disinfection will serve as a technical and
professional resource during all phases of the verification of a mixer, including the
review of Test Plans and Verification Reports, as requested by NSF and EPA.

2.3 Capabilities and Description of the Equipment to be Tested

2.3.1 Mixer Capabilities

The Test Plan shall identify the capabilities of the equipment to be evaluated in the
verification testing. Statements should also be made regarding the application of the
equipment, the known limitations of the equipment and what advantages it provides
over alternative equipment.

Statements of capabilities that are too easily met may not be of interest to the potential
user, while statements of capabilities that are overstated may not be achievable. The
statement of capabilities forms the basis of the entire equipment verification testing and
must be chosen carefully. Therefore the Test Plan should include a range of mixer
sizes each tested in a range of flow velocities.

11


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

One obvious limitation of measuring dye or tracer chemical dispersion to verify mixer
capabilities is that the energy of moving water is also responsible for some amount of
dye or tracer chemical dispersion. This issue shall also be addressed in the statement
of capability.

2.3.2	Mixer Description

The Test Plan shall also include a description of the mixer(s), impeller configuration
and ancillary equipment to be tested. Engineering drawings and photographs of the
mixer(s), impeller configuration and ancillary equipment shall be included in the Test
Plan. For each mixer, the impeller shall be commercially available and fully described
in the Test Plan.

2.3.3	Mixer Requirements

Data plates shall be permanently secured to each mixer. The data plates shall be
easily read and contain a least the following information:

•	Equipment name

•	Model #

•	Manufacturer's name and address

•	Electrical requirements-volts, amps, and hertz

•	Serial number

•	Warning and caution statements

2.4 Description and Requirements for Laboratory /Test Facility

2.4.1 Test Plan Content

The Test Plan shall include a description of the testing facilities to be used in the
Verification Tests, including the hydraulic and analytical laboratories. The Test Plan
shall identify the channel design, flow control apparatus and all instrumentation to be
used in creating, controlling and measuring flow and in conducting the dye/tracer

12


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

dilution studies as required by Section 3 of this Protocol. Equipment descriptions shall
include:

•	Equipment name

•	Model #

•	Manufacturer's name

The Test Plan shall include diagrams showing the test facility, including location of the
test mixer, hydraulic controls, dye/tracer injection points, sampling equipment, and
other relevant equipment. Figure 1 provides a example of a verification test set-up.

2.4.2 Facility Requirements

Verification testing shall be conducted at a hydraulic laboratory that conforms to the
criteria outlined below.

•	The hydraulic laboratory shall have a defined channel, rectangular (preferred) or
circular with a minimum width of 6 feet and water depth of 6 feet.

•	The channel shall accommodate the installation of mixers and sampling
equipment;

•	The hydraulic laboratory shall be capable of creating and sustaining hydraulic
conditions as specified in Table 1. The hydraulic laboratory shall be able to
sustain flows ranging from 10 cfs to 120 cfs within the specified channel
geometry. A calibrated flow control device such as a weir should control specific
hydraulic conditions. This hydraulic control shall be far enough away from the
mixer so that it does influence flow patterns in the vicinity of the mixer.

•	The hydraulic laboratory shall be capable of supplying the specified electrical
utilities for the mixers, dye or tracer feed pumps and sampling equipment. The
hydraulic laboratory will also be required to measure amperage draw for each
mixer during each test.

•	The hydraulic laboratory shall be capable of creating constant hydraulic
conditions during each mixer test using a water supply with characteristics that
remain consistent throughout the tests and ensure conservation of the dye/tracer

13


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

chemical to be used. Flow shall be verified by use of the calibrated flow control
device and independent flow / velocity measurements.

•	The hydraulic laboratory shall be equipped with the necessary dye or
conservative tracer chemical and associated equipment, and sampling
equipment.

2.5 Experimental Design

The Test Plan shall completely describe the test procedures to be followed during the
course of verification testing. The methods and procedures described in the test plan
shall be consistent with the requirements and guidelines established in Section 3 of this
Protocol. The Test Plan shall describe:

•	how the required test conditions are to be achieved and maintained;

•	the procedures for tracer/dye dilution study, injection, sampling and analysis;
and

•	procedures for data collection, storage, and compilation.

14


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

2.6 Quality Assurance Project Plan (QAPP)

The Test Plan shall include a QAPP that specifies procedures to be used to ensure
data quality and integrity. Careful adherence to these procedures will ensure that data
generated from the verification testing will provide sound analytical results that can
serve as the basis for performance verification. The purpose of the QAPP is to ensure
that data resulting from this verification testing is of known quality and that a sufficient
number of critical measurements are taken.

2.6.1 Required Elements

The Quality Assurance Project Plan shall include the following elements:

•	Description of methodology for measurement of accuracy for field and laboratory
measurements.

•	Description of methodology for measurement of precision for field and laboratory
measurements.

•	Outline of the procedure for determining samples to be analyzed in duplicate,
the frequency and approximate number.

•	Description of the procedures used to assure that the data are correct.

•	Development of a corrective action plan.

•	Provision of all QC information such as calibrations, blanks and reference
samples in an appendix. All raw analytical data shall also be reported in an
appendix.

•	Provision of all data in hardcopy and electronic form in a common spreadsheet
or database format.

The QAPP shall addess the following measurements/operations:

•	Calibrating hydraulic control and flow monitoring

•	Velocity distribution and flow monitoring

•	Mixer Hp monitoring

15


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

•	Induction unit flow rate monitoring

•	Metering pump verification monitoring

•	Preparation of dye or tracer chemical stock solution

•	Preparation of dye or tracer chemical standards

•	Fluorometer or other analyzer calibration

•	Fluorometer or other analyzer operation

•	Sampling rig operation

2.6.2	Quality Assurance Responsibilities

A number of individuals may be responsible for QA/QC throughout the verification
testing. Primary responsibility for ensuring that both mixer and sampling and analysis
activities comply with the QA/QC requirements of the Test Plan shall rest with the
Testing Organization.

QA/QC activities for the analytical laboratory shall be the responsibility of that
analytical laboratory's supervisor. If problems arise or any data appear unusual, they
shall be thoroughly documented and corrective actions shall be implemented as
specified in this section. The QA/QC measurements made by the analytical laboratory
are dependent on the analytical methods being used.

2.6.3	Data Quality Indicators

The data obtained during the verification testing must be of sound quality for
conclusions to be drawn on the equipment. For all measurement and monitoring
activities conducted for equipment verification, the NSF and EPA require that data
quality parameters be established based on the proposed end uses of the data. Data
quality parameters include four indicators of data quality: representativeness, accuracy,
precision, and statistical uncertainty. The Test Plan shall include a plan for identifying
such indictors.

16


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

2.6.4	Operational Control Checks

The Test Plan shall describe the QC requirements that apply to the operation of the
mixer equipment. This section will explain the methods to be used to check on the
accuracy of equipment operating parameters and the frequency with which these
quality control checks will be made.

2.6.5	Data Reduction, Validation, and Reporting

The Test Plan shall include procedures to maintain good data quality. Specific
procedures shall be followed during data reduction, validation, and reporting.

2.6.6	System Inspections

On-site system inspections for sampling activities, field operations, and laboratories
may be conducted as specified in the Test Plan. These inspections will be performed
by NSF International or its designee to determine if the Verification Test Plan is being
implemented as intended. At a minimum, NSF shall conduct one audit of the sampling
activities, field operations program and laboratories during the Verification Test.

2.6.7	Corrective Action

The Test Plan shall incorporate a corrective action plan. This plan shall establish
acceptance limits and the corrective action to be initiated whenever such acceptance
criteria are not met. The Test Plan shall identify the individuals responsible for
implementation.

Routine corrective action may result from common monitoring activities, such as:

•	Performance evaluation audits

•	Technical systems audits

2.7 Data Management and Analysis

A variety of data will be generated during a verification testing. Each piece of data or
information identified for collection in the Test Plan shall be included in the Verification

17


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

Report. The data handling section of the Test Plan shall describe types of data to be
collected and managed and how the data will be reported the Verification Report.

All raw data and validated data shall be reported. These data shall be provided in hard
copy and in electronic format. As with the data generated by the innovative equipment,
the electronic copy of the laboratory data shall be provided in a spreadsheet. In
addition to the sample results, all QA/QC summary forms must be provided.

Other items that must be provided include:

•	field notebooks;

•	photographs, slides and videotapes (copies);

•	results from the use of other field analytical methods.

2.8 Safety Measures

The Test Plan shall address safety considerations that are appropriate for the testing
laboratory and the equipment being tested. The safety procedures shall address safety
considerations, including the following as applicable:

•	storage, handling, and disposal of dye or tracer chemical

•	conformance with electrical code

•	confined space entry

•	working in and around a hydraulic conveyance system

18


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

Figure 1- Example of Verification Test Set-up

Tracer Dye
Stock Solution

Channel Bottom

Metering Pump

ixer

Sampling Pump (each sampling port may
require a separate pump)

Samnlina Bottles

Sampl'

Hydraulic control
(beyond area
which may
influence mixer
voume)

19


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

3 Verification Test Procedures- Requirements and Guidelines

This section contains requirements and guidelines to be followed by a Field Testing
Organization in establishing site-specific test procedures for the verification testing of
high-rate induction mixers. Test procedures developed by the Field Testing
Organization in conformance with these requirements and guidelines should be clearly
documented in a Verification Test Plan.

3.1	Test Objective

The objective of this test is to characterize the performance of a high-rate induction
mixer with respect to its ability to rapidly transfer chemicals into a flowing body of
process water. Mixer performance is characterized by the volume of water affected by
a mixer (i.e., the size of its mixing zone) over a range of flow conditions that are
representative of those observed in wet weather flow collection and treatment facilities.

3.2	Test Conditions

Each verification test shall evaluate a single induction mixer under the three sets of
hydraulic conditions defined in Table 1. These hydraulic conditions are intended to
represent the range of conditions typically encountered in wet-weather treatment
facilities. The induction mixer manufacturer shall provide one mixer for each
verification test. A verification test shall include the three hydraulic conditions defined
in Table 1. It is recommended verification tests be performed on a series of induction
mixers. For example a series of mixers may include, 2 Hp, 5 Hp, 10 Hp and 20 Hp
mixers.

Table 1 - Verification Test Hydraulic Test Conditions

Hydraulic
Conditions

Velocity
(fps)

Cross Section
Flow Area
(ft2)

Sampling
Distance1

(ft)

Corresponding
Sampling Time
(sec)

Hydraulic Condition 1

0.5

36

5, 10, and 15

10, 20,and 30

20


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

Hydraulic
Conditions

Velocity
(fps)

Cross Section
Flow Area
(ft2)

Sampling
Distance1

(ft)

Corresponding
Sampling Time
(sec)

Hydraulic Condition 2

2

36

5, 10, and 15

2.5, 5.0,and 7.5

Hydraulic Condition 3

3

36

5,10, and 15

1.7, 3.3, and 5.0

Downstream distance from point of dye or tracer injection to point of dye or tracer sampling.

3.3 Methods and Materials

A single mixer shall be installed in the channel with one sampling rig set at the
appropriate sampling location. The entire system, including channel flow, induction
mixer, dye or chemical tracer metering pump, and sampling pump, shall be allowed to
reach a steady state before sampling begins. These system components shall be
monitored as specified in the QAPP, to assure and document that a steady state
condition has been met. Once a steady state condition has been achieved, the
sampling can begin.

Variability in the operation of the system should be minimized through the QA/QC
procedure identified in the QAPP. However, the natural movement of water as it flows
through the channel will introduce variability in data collected at the sampling rig.
Therefore, it is recommended that an "event mean" type sample (i.e. flow weighted
sample) be taken over a 30 minute sampling duration. Sampling pumps with constant
flow rates should be used to take continuous samples for the duration of 30 minute.
The total sample volume should not be less than two liters.

3.3.1 Mixer Operation

The induction mixer will be operated as though it was being used for disinfection of wet-
weather flows. However, instead of the induction unit mixing a disinfectant, such as
sodium hypochlorite, it will mix water that is amended with a dye or conservative tracer
chemical. The mixer will run continuously during the test. Amperage shall be
measured for each mixer during each test.

21


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

3.3.2	Dye or Conservative Tracer Chemical

The dye or tracer chemical shall be a conservative material whose concentrations may
accurately measured downstream of the mixer. The selection of the dye or tracer
chemical shall be made with consideration of the water quality parameters of the
process water. The use of Rhodamine WT is recommended be used for the testing
because of the established standards associated with this chemical. The use of other
conservative tracer chemicals such as LiCI may be proposed in the Test Plan.

The concentration of the dye or tracer at the sampling locations shall correspond with
the detection limits of the dye or tracer analyzer. The concentration of the stock dye or
tracer solution must be calculated based on mass balance and the appropriate
dilutions.

3.3.3	Metering Pump

A chemical metering pump shall be used to inject the dye or tracer into the water that is
drawn by the induction unit. The metering pump shall be sized to deliver the specified
quantity of dye or tracer, based on the concentration of dye or tracer required in the
process water, the process water flow, and the flow drawn by the induction unit.

The metering pump shall be operated for each of the hydraulic conditions during each
verification testing. The metering pump shall draw from a stock solution of dye or
tracer and discharge into the hose that conveys water to the induction unit.

3.3.4	Sampling Rig

A sampling apparatus or "rig" shall be constructed such that simultaneous water
samples may be drawn at defined points across the entire cross section of the channel.
The sampling rig will consist of a number of sampling ports spaced equally throughout
the channel so to capture the complete cross sectional area. The sampling ports shall
be spaced at a minimum of one port per square foot. The sampling rig shall be
constructed so that a sample from each port can be drawn simultaneously. Each
sample will be collected, bottled and analyzed individually. The sampling rig will

22


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

require pumps to draw the process water from the sampling ports into sample
containers.

The sampling rigs shall be located at five, ten and fifteen feet downstream of the mixer
during each test run as defined in Tablel This corresponds to mixing times ranging
from 2 seconds to 30 seconds.

Only one sampling rig shall be installed in the channel during each test run. Inactive
sampling rigs left upstream of an active sampling rig may affect the mixing zone
pattern.

3.3.5	Fluorometer or other tracer analyzer

A fluorometer or tracer analyzer shall be used to measure the concentration of tracer in
the process water. As a Rhodamine WT dye was previously recommended; it is
recommended that a fluorometer be used to measure the concentration of the
Rhodamine WT. In this case, an established company shall manufacture the
fluorometer, and the procedures to be implemented for preparing and measuring the
dye shall be established by the fluorometer manufacturer.

The fluorometer or other tracer analyzer will most likely be operated after the samples
have been taken. It is recommended that the sample be analyzed after each hydraulic
condition sampling period, so that corrections can be made to the sampling protocols.

3.3.6	Velocity Meter

A velocity meter shall be used to verify the total flow and to calibrate to the hydraulic
control. The velocity meter shall also be used to verify the velocity distribution within
the channel.

3.4 General Test Procedures

23


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

3.4.1	Test Preparation

3.4.1.1	Standards

Tracer chemical standards shall be used to calibrate the fluorometer or other tracer
analyzer. These standards shall be made according to a method developed by the
manufacturer of the fluorometer.

3.4.1.2	Fluorometer or other analyzer calibration

The analyzer shall be calibrated according to a method developed by the manufacturer.
An approved QA/QC plan shall be used.

3.4.1.3	Metering pump calibration

The metering pump that injects dye or tracer chemical into the induction flow shall be
calibrated prior to testing. The delivery rate shall be verified volumetrically.

3.4.2	Test Conditions

3.4.2.1	Flow

During testing the flow in the channel shall be measured. Once the flow has been
verified, flow can be estimated using the hydraulic control. Flow shall be controlled
to create the specified velocities as defined in Table 1, within the given channel
geometry.

3.4.2.2	Channel geometry

The hydraulic laboratory shall establish the channel geometry prior to testing. The
channel shall have a minimum cross sectional width of 6 feet and a minimum water
depth of 6 feet.

3.4.2.3	Mixer size

The vendor shall establish the mixer size (Hp) prior to the test. Hp shall be verified
and documented in the field by measuring the amperage, and calculating wattage
by the following equation;

24


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

Watts = amperes x voltage
Note: 746 watts is the electrical equivalent to one (1) horsepower.

3.4.3	Mixer Testing

3.4.3.1	Sampling Locations

Each mixer shall be tested under three hydraulic conditions as defined in Table 1.
During each hydraulic condition, dilution samples shall be collected separately at
locations 5, 10 and 15 feet downstream of the mixer (or at the equivalent contact time
based on velocities presented in Table 1). This will provide information with respect to
the shape of the mixing zone. Only one sampling rig shall be installed in the channel
during each test run.

3.4.3.2	Sampling procedures

The sampling rig shall have sampling ports evenly spaced throughout the channel.
Samples shall be simultaneously drawn from each sampling port into individual sample
bottles. Drawing the samples simultaneously will likely require having a dedicated
pump for each sampling port.

Sampling at each location shall occur separately. The sampling rig should be installed
at one sampling location, the samples taken, and the rig removed. The sampling rig
should then be installed at the next sampling location and the same procedure
followed.

3.4.4	Tracer Analysis

3.4.4.1	Sample handling

Samples shall be collected in approved bottles, preserved by an approved method and
analyzed before the approved maximum holding time.

3.4.4.2	Analyzer procedures

25


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

The fluorometer or other tracer analyzer shall be used to the specification of the
manufacturer. An approved QA/QC program shall be implemented.

3.5 Data Analysis

The Test Plan shall describe the procedures used to define the mixing zone volume
based on the concentration of dye or tracer chemical in the process water throughout
the mixing channel. The measured dye or tracer concentration shall be used to
generate an isopleth diagram as depicted in Figure 2. An isopleth diagram shall be
generated for each hydraulic condition and downstream sampling distance defined in
Table 1 for a total of nine isopleth drawings for each mixer tested.

The isopleth diagrams shall be drawn to depict the uniform dye or tracer concentration.
The uniform theoretical dye or tracer concentration shall be defined by the
concentration if the dye or tracer were instantaneously dispersed over the entire cross
section of the channel. It shall be calculated by the following equation:

Uniform Tracer Concentration = (Tracer Stock Concentration * Metering Pump Feed Rate)

Process Water Flow Rate

26


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

The uniform theoretical dye or tracer concentration shall be represented by the value of
1.0. Tracer concentrations greater than the uniform dye or tracer concentration shall
represented by values greater than 1.0, as depicted in Figure 2.

Figure 2- Tracer Concentration Isopleth Diagram

A percent mix factor shall be calculated once the uniform theoretical dye or tracer
concentration isopleth has been established. The percent mix factor represents the
area of the channel that has experienced a complete mix. The percent mix factor shall
be calculated by the following equation:

Water
Depth

Channel Width

Percent Mix Factor = Channel Area with Tracer Concentration Greater than Uniform Concentration

Total Channel Cross Section Area

Other data that shall be collected are process water flow velocity, mixer amperage
draw, and dye or tracer feed pump rate. Results shall be presented in tables and
figures.

27


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

4 Reporting

4.1	Data Management and Analysis

A variety of data will be generated during a verification testing. Each piece of data or
information identified for collection in the Test Plan will need to be provided in the Final
Verification Report. The data handling section of the Test Plan shall describe what
types of data and information needs to be collected and managed, and shall also
describe how the data will be reported to the NSF for evaluation.

4.2	Verification Report

The FTO shall prepare a draft Verification Report describing the verification testing that
was carried out and the results of that testing. The Verification report shall undergo a
complete review by NSF International and the EPA, as well as a peer review as
recommended by the Technology Panel on High Rate Disinfection. The mixer vendor
shall review and be provided the opportunity for input on its content. This report should
fully describe the mixer and the verification of its performance characteristics. At a
minimum, shall include the following items:

•	Introduction

•	Executive Summary

•	Description and Identification of Product Tested

•	Procedures and Methods Used in Testing

•	Results and Discussion

•	Conclusions and Recommendations

•	References

•	Appendices, which may include test data.

4.3	Verification Statement

NSF and EPA shall prepare a Verification Statement that briefly summarizes the
Verification Report for issuance to the mixer vendor. The Verification Statement shall
provide a brief description of the testing conducted and a synopsis of the performance

28


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

results. The Statement is intended to provide verified vendors a tool by which to
promote the strengths and benefits of their product.

29


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

5 References

EPA. 1973. Collins HF and Kruse (EPA-670/2-73-077). United States Environmental
Protection Agency.

Moffa & Associates. 1997. CSO Final Disinfection Pilot Study Final Report. Spring
Creek AWPCP Upgrade Capital Project No. WP-225. Camp, Dresser & McKee.
November 1997.

White, Clifford Geo. 1992. Handbook of Chlorination and Alternative Disinfectants, 3rd
edition. Van Nostrand Reinhold, New York. 1308 pp.

30


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

6 Glossary

Terms and acronyms used in this Protocol that have special meaning are defined here.

Complete Mix - The dispersion of dye or tracer chemical resulting in an equal
concentration distribution throughout the channel. This is represented by the unit
isopleth (i.e. isopleth equal to 1.0).

Dye or Tracer Chemical - A conservative chemical material that can be measured in
water. A conservative chemical does not react or change form when diluted in water.

EPA - The United States Environmental Protection Agency, its staff or authorized
representatives.

Event Mean Sample - A flow weighted composite sample taken throughout an event.
This sample represents the average concentration throughout the sampling duration.

Field Testing Organization - An organization qualified to conduct studies and testing
of induction mixers in accordance with the Verification Protocol.

Induction Mixer - A mechanical mixer that was designed with a vacuum port at the
impellor. The mixer impellor rotates at 35,000 rpms or greater and causes a vacuum at
this port. This vacuum draws disinfectant chemical to the impellor where it is mixed
with the process water.

Isopleth - A line connecting points of equal dye or conservative tracer concentrations.

Mixing - The act of creating turbulence within a flow path for the purpose of dispersing
chemical.

NSF - NSF International, its staff, or other authorized representatives.

Percent Mixing Factor - the area of the channel that has experienced a complete mix
divided by the total area of the channel.

Quality Assurance Project Plan (QAPP) - a written document that describes the
implementation of quality assurance and quality control activities during the life cycle of
the project. The QAPP is a required component of a verification Test Plan.

31


-------
Draft 3.4 - Generic Verification Protocol for Induction Mixers Used for High Rate
Disinfection of Wet Weather Flows - June 2000

Representativeness - a measure of the degree to which data accurately and precisely
represent a characteristic of a population parameter at a sampling point or for a
process conditions or environmental condition.

Sampling Rig - A rigid device used to keep the sampling ports at specified locations
within the channel.

Test Condition - A specific mixer horsepower, sampling rig location, and channel flow
velocity.

Verification - To establish the evidence on the range of performance of equipment
and/or device under specific conditions following an established protocol(s) and test
plan(s).

Verification Protocol - a written document that clearly states the objectives, goals,
and scope of the testing under the ETV Program and that establishes the minimum
requirements for verification testing and for the development of a verification test plan.
A protocol shall be used for reference during Manufacturer participation in the
verification testing program. Often simply referred to as a Protocol.

Verification Test Plan - A written document that establishes the detailed test
procedures for verifying the performance of a specific technology. It also defines the
roles of the specific parties involved in the testing and contains instructions for sample
and data collection, sample handling and preservation, and quality assurance and
quality control requirements relevant to a given test site.

Verification Report - a written document prepared by the FTO containing all raw and
analyzed data, all QA/QC data sheets, descriptions of all collected data, a detailed
description of all procedures and methods used in the verification testing, and all
QA/QC results.

Verification Statement - A written document that summarizes a final report reviewed
and approved by NSF on behalf of EPA or directly by EPA.

32


-------
Table Of Contents

1	Introduction	1

1.1	Purpose	1

1.2	Scope	1

1.3	The ETV Program and the Wet Weather Flow Technologies Pilot	1

1.4	Technology Applications and Description	2

1.5	Experimental Objective	5

1.6	Technical Approach	6

1.7	Verification Process	6

2	Development of a Verification Test Plan	7

2.1	Purpose of a Test Plan	7

2.2	Roles and Responsibilities of Involved Organizations	8

2.2.1	NSF International	8

2.2.2	US Environmental Protection Agency (EPA)	9

2.2.3	Field Testing Organization	9

2.2.4	Vendor	10

2.2.5	Hydraulic Laboratory	10

2.2.6	Analytical Laboratory	11

2.2.7	General Contractors	11

2.2.8	Technology Panel on High-Rate Disinfection	11

2.3	Capabilities and Description of the Equipment to be Tested	11

2.3.1	Mixer Capabilities	11

2.3.2	Mixer Description	12

2.3.3	Mixer Requirements	12

2.4	Description and Requirements for Laboratory/Test Facility	12

2.4.1	Test Plan Content	12

2.4.2	Facility Requirements	13

2.5	Experimental Design	14

2.6	Quality Assurance Project Plan (QAPP)	15

2.6.1	Required Elements	15

2.6.2	Quality Assurance Responsibilities	16

2.6.3	Data Quality Indicators	16

2.6.4	Operational Control Checks	17

2.6.5	Data Reduction, Validation, and Reporting	17

2.6.6	System Inspections	17

2.6.7	Corrective Action	17

2.7	Data Management and Analysis	17

2.8	Safety Measures	18

3	Verification Test Procedures- Requirements and Guidelines	20

3.1	Test Objective	20

3.2	Test Conditions	20

3.3	Methods and Materials	21

3.3.1	Mixer Operation	21

3.3.2	Dye or Conservative Tracer Chemical	22

3.3.3	Metering Pump	22

3.3.4	Sampling Rig	22

3.3.5	Fluorometer or other tracer analyzer	23

3.3.6	Velocity Meter	23

3.4	General Test Procedures	23

3.4.1	Test Preparation	24

3.4.2	Test Conditions	24

3.4.3	Mixer Testing	25

3.4.4	Tracer Analysis	25

3.5	Data Analysis	26

4	Reporting	28

4.1	Data Management and Analysis	28

4.2	Verification Report	28

4.3	Verification Statement	28

5	References	30

6	Glossary	31


-------

-------