Battelle

The Business of Innovation

Environmental Technology
Verification Program

Advanced Monitoring
Systems Center

Test/QA Plan for Verification of
Dioxin Emission Monitoring Systems

(EMSs)

ETvETvETv


-------
TEST/QA PLAN

for

Verification of
Dioxin Emission Monitoring Systems (EMSs)

Version 1.0

September 2,2005
Prepared by

Battelle

505 King Avenue
Columbus, OH 43201-2693


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 2 of 48
Version: 1.0
September 2, 2005

SECTION A
PROJECT MANAGEMENT

A1 VENDOR APPROVAL PAGE

ETV Advanced Monitoring Systems Center

Draft Test/QA Plan for Verification of
Dioxin Emission Monitoring Systems (EMSs)

Version 1

September 2, 2005
APPROVAL:

Name

Company
Date


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 3 of 48
Version: 1.0
September 2, 2005

A2 TABLE OF CONTENTS

Section	Page

A	PROJECT MANAGEMENT

A1 Vendor Approval Page	2

A2 Table of Contents	3

A3 Di stributi on Li st	4

A4 Verification Test Organization	5

A5 Background	13

A6 Verification Test Description and Schedule	14

A7 Quality Objectives	20

A8 Special Training/Certification	23

A9 Documentation and Records	23

B	MEASUREMENT AND DATA ACQUISITION

B1 Experimental Design	24

B2 Reference Sample Collection	30

B3 Sample Handling and Custody Requirements	31

B4 Laboratory Reference Methods	33

B5 Quality Control Audits and Requirements	34

B6 Instrument/Equipment Testing, Inspection, and Maintenance	35

B7 Instrument Calibration and Frequency	36

B8 Inspection/Acceptance of Supplies and Consumables	38

B9 Non-Direct Measurements	38

BIO Data Management	38

C	AS SES SMENT AND OVERSIGHT

C1 Assessments and Response Actions	41

C2 Reports to Management	42

D	DATA VALIDATION AND USABILITY

Dl Data Review, Validation, and Verification Requirements	45

D2 Validation and Verification Methods	45

D3 Reconciliation with User Requirements	46

E REFERENCES	48

APPENDIX A	A1


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 4 of 48
Version: 1.0
September 2, 2005

A3 DISTRIBUTION LIST

Dioxin EMS Vendors

Jiirgen Reinmann
Becker Messtechnik
Kolner Strasse 6
D-65760 Eschborn
Germany

Tsunehisa (Tom) Onishi
IDX Technologies
NIBld. 3-12-9 Kayaba-cho
Nihonbashi, Chuo
Tokyo, Japan

Battelle

Kenneth Cowen
Thomas Kelly
Karen Riggs
Zachary Willenberg
Robyn Kroeger
Battelle
505 King Ave.
Columbus, OH 43201

ARCADIS

Dahman Touati
ARCADIS

4915 Prospectus Drive
Suite F

Durham, N.C. 27713
Brian Gullett

U.S. Environmental Protection Agency
E305-01 USEPA Mailroom
Research Triangle Park, NC 27711

EPA

Thomas Sterner
Monitoring Systems
Schloss 2

A-2542 Kottingbrunn Niederosterreich
Austria

Elizabeth A. Betz

U.S. Environmental Protection Agency-
National Exposure Research Laboratory
E205-01 EPA Mailroom
Research Triangle Park, NC 27711

Robert Fuerst

U.S. Environmental Protection Agency-
National Exposure Research Laboratory
D205-05 EPA Mailroom
Research Triangle Park, NC 27711


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 5 of 48
Version: 1.0
September 2, 2005

A4 VERIFICATION TEST ORGANIZATION

The verification test will be conducted under the auspices of the U.S. Environmental Protection
Agency (EPA) through the Environmental Technology Verification (ETV) Program. It will be
performed by Battelle, which is managing the ETV Advanced Monitoring Systems (AMS)

Center through a cooperative agreement with EPA. The scope of the AMS Center covers
verification of monitoring technologies for contaminants and natural species in air, water, and
soil.

This verification test will be coordinated and supervised by Battelle, with the support of
ARCADIS Inc., and in cooperation with EPA. The testing will be conducted at the EPA's
Research Triangle Park (RTP) campus and will involve the evaluation of commercial dioxin
emission monitoring systems (EMSs), where the term "dioxin" is used to genetically represent
polychlorinated dibenzo-p-dioxins and polychlorinated dibenzofurans. Staff from ARCADIS
will support this test under subcontract from Battelle. ARCADIS will provide support in
preparing for the test, during installation of the dioxin EMSs to be tested, and by overseeing
operation of the EMSs during periods of routine operation. Additionally, ARCADIS will collect
reference dioxin samples and will arrange for the reference samples to be analyzed using a
modified version of EPA Method 23.1

Each EMS vendor will install their respective EMS, operate the EMS through portions of the test
(unless they give written consent for Battelle or ARCADIS staff to operate it), and repair or
maintain their EMS during the test.

Quality assurance (QA) oversight will be provided by the Battelle Quality Manager, and also by
the EPA AMS Center Quality Manager at her discretion. The organization chart in Figure 1
identifies the responsibilities of the organizations and individuals associated with the verification
test. Roles and responsibilities are defined further below.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 6 of 48
Version: 1.0
September 2, 2005

Figure 1. Organizational Chart


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 7 of 48
Version: 1.0
September 2, 2005

A4.1 Battelle

Dr. Kenneth Cowen is the AMS Center Verification Test Coordinator for this test. In this role,
Dr. Cowen will have overall responsibility for ensuring that the technical, schedule, and cost
goals established for the verification test are met. Specifically, he will:

>	Assemble a team of qualified technical staff to conduct the verification test.

>	Direct the team (Battelle and ARCADIS staff) in performing the verification test in
accordance with this test/QA plan.

>	Ensure that all quality procedures specified in the test/QA plan and in the AMS
Center Quality Management Plan2 (QMP) are followed.

>	Manage the subcontract(s) under which ARCADIS conducts the testing activities.

>	Prepare the draft and final test/QA plan, verification reports, and verification
statements.

>	Revise the draft test/QA plan, verification reports, and verification statements in
response to reviewers' comments.

>	Respond to any issues raised in assessment reports and audits, including instituting
corrective action as necessary.

>	Serve as the primary point of contact for vendor representatives.

>	Coordinate distribution of the final test/QA plan, verification reports, and statements.

>	Establish a budget for the verification test and manage staff to ensure the budget is
not exceeded.

>	Ensure that confidentiality of sensitive vendor information is maintained.

Dr. Thomas Kelly is Battelle's Verification Testing Leader for the AMS Center. Dr. Kelly will:

>	Support Dr. Cowen in preparing the test/QA plan and organizing the test.

>	Review the draft and final test/QA plan.

>	Review the draft verification reports and statements.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 8 of 48
Version: 1.0
September 2, 2005

>	Support Dr. Cowen in responding to any issues raised in assessment reports and
audits.

Ms. Karen Riggs is Battelle's manager for the AMS Center. Ms. Riggs will:

>	Review the draft and final test/QA plan.

>	Review the draft and final verification reports and verification statements.

>	Ensure that necessary Battelle resources, including staff and facilities, are committed
to the verification test.

>	Ensure that confidentiality of sensitive vendor information is maintained.

>	Maintain communication with EPA's technical and quality managers.

>	Facilitate a stop work order if Battelle or EPA QA staff discovers adverse findings
that will compromise data quality or test results.

Battelle Testing Staff will conduct the testing of the dioxin EMSs during the verification test.
Battelle staff (including Dr. Cowen) will be on-site at the EPA test facility during the verification
test, and will be in daily communication with facility personnel, and with EMS vendors as
needed. The responsibilities of the technical staff will be to:

>	Maintain and operate the dioxin EMSs if so instructed by the vendors, assuming
proper training in EMS operation is provided.

>	Assure that verification testing is performed as described in the test/QA plan.

>	Communicate and coordinate with ARCADIS staff and the EMS vendor
representatives on the installation, operation, testing, and removal of the EMSs.

>	Communicate with ARCADIS testing staff on the planning, performance, and
reporting of the reference dioxin sampling and analysis.

>	Record qualitative observations about the maintenance and operation of the dioxin
EMSs during testing.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 9 of 48
Version: 1.0
September 2, 2005

>	Assure that the data from each dioxin EMS are recorded and transmitted to the
Verification Test Coordinator on at least a weekly basis.

>	Provide input on test procedures, EMS operation and maintenance, and field
conditions for the draft verification reports.

Mr. Zachary Willenberg is Battelle's Quality Manager for the AMS Center. Mr. Willenberg
will:

>	Review the draft and final test/QA plan.

>	Conduct a technical systems audit at least once during the verification test, or
designate other QA staff to conduct the audit.

>	Audit at least 10% of the verification data.

>	Prepare and distribute an assessment report for each audit.

>	Verify implementation of any necessary corrective action.

>	Notify Battelle's AMS Center Manager to issue a stop work order if audits indicate
that data quality is being compromised.

>	Provide a summary of the QA/QC activities and results for the verification reports.

>	Review the draft and final verification reports and verification statements.

>	Assume overall responsibility for ensuring that the test/QA plan is followed.

A4.2 Dioxin EMS Vendors

The responsibilities of the EMS vendors are as follows:

>	Review and provide comments on the draft test/QA plan.

>	Approve the final test/QA plan prior to test initiation.

>	Provide a dioxin EMS with complete flue gas sampling inlet for evaluation during the
verification test.

>	Provide all other equipment/supplies/reagents/consumables needed to operate their
EMSs for the duration of the verification test.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 10 of 48
Version: 1.0
September 2, 2005

>	Supply a representative to install and maintain their technology, and to operate it in
portions of the test specified in this test/QA plan, or provide written consent and
instructions for Battelle staff to carry out these activities.

>	Provide written instructions for routine operation of their EMSs, including a daily
checklist of diagnostic and/or maintenance activities.

>	Review and provide comments on the draft verification report and statement for their
respective EMSs.

A4.3 EPA

EPA's responsibilities for the AMS Center are based on the requirements stated in the
"Environmental Technology Verification Program Quality Management Plan" (EPA QMP).3 The
roles of specific EPA staff are as follows:

Ms. Elizabeth Betz is EPA's AMS Center Quality Manager. Ms. Betz will:

>	Review the draft test/QA plan.

>	Perform at her option one external technical systems audit during the verification test.

>	Notify the EPA AMS Center Manager of the need for a stop work order if the
external audit indicates that data quality is being compromised.

>	Prepare and distribute an assessment report summarizing results of the external audit.

>	Review draft verification reports and statements.

Mr. Robert Fuerst is EPA's manager for the AMS Center. Mr. Fuerst will:

>	Review the draft test/QA plan.

>	Approve the final test/QA plan.

>	Review the draft verification reports and statements.

>	Oversee the EPA review process for the verification reports and statements.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 11 of 48
Version: 1.0
September 2, 2005

>	Coordinate the submission of verification reports and statements for final EPA
approval.

A4.4 ARCADIS

ARCADIS personnel are responsible for performing most of the actual testing activities,
including providing required emissions sampling and analytical services. In addition, ARCADIS
will purchase incidental materials as needed for the verification test, as well as provide repair and
preventive maintenance support to equipment at the host facility that will be used during the
verification test.

Dr. Dahman Touati is the ARCADIS Technical Lead for the dioxin EMS verification test. In this
role, Dr. Touati is responsible for ensuring that the project meets the scheduled technical
milestones agreed upon by Battelle through subcontracts with ARCADIS. Dr Touati will:

>	Be the primary ARCADIS contact for Battelle's Verification Test Coordinator.

>	Ensure that designated ARCADIS staff, and the EPA test facility are ready for the
verification test.

>	Coordinate distribution of the test/QA plan to ARCADIS staff.

>	Coordinate the operations of the boiler and testing facilities.

>	Direct the ARCADIS testing staff in performing the baseline testing in accordance
with the test/QA plan.

>	Review and approve all data and records related to facility operation

>	Review the draft test/QA plan.

>	Assist Battelle and EMS vendor staff in the installation, operation, testing, and
removal of the EMSs at the EPA boiler facility, including connection of EMS flue gas
sampling inlets to the facility stack.

>	Assist in the planning and performance of the reference dioxin sampling and analysis.

>	Ensure the availability of appropriate space and needed utilities (e.g., electricity, air,
water) for the EMSs during testing.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 12 of 48
Version: 1.0
September 2, 2005

>	Support Battelle Testing Staff in providing daily oversight of the EMSs during
periods of routine operation, checking diagnostic indicators and contacting Battelle if
faults in EMS operation are observed.

>	Support Battelle Testing Staff in recording observations about the maintenance and
operation of the dioxin EMSs during the test period.

>	At the option of EPA facility staff, review the draft verification reports and
statements.

>	Coordinate with the ARCADIS QA and safety officers to ensure appropriate
procedures are in place and followed during the verification test.

The responsibilities of ARCADIS Testing Staff in this test are as follows:

>	Provide equipment and personnel to carry out dioxin reference sampling using
Method 23 sampling trains as described in this test/QA plan.

>	Recover collected samples from the Method 23 trains, and transfer the recovered
samples for dioxin analysis.

>	Coordinate analysis on flue gas samples and QA samples per the modified Method 23
described in this test/QA plan.

>	Calculate the Method 23 sampling results in terms of flue gas dioxin concentrations
as specified in the method, and submit a report to Battelle that describes the sampling,
and presents the sample analysis results, QA results, and calculated dioxin
concentrations in the flue gas

>	Review that portion of the verification reports that describes the Method 23 sampling
and analysis.

The ARCADIS QA Officer is Laura Nesslev who is responsible for:

>	Reviewing and approving procedures and testing performed by ARCADIS and
described in this QAPP.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 13 of 48
Version: 1.0
September 2, 2005

>	Ensuring that the ARCADIS portion of the QAPP is implemented by performing
routine assessments.

>	Communicating with Battelle's Quality Manager for the AMS center to coordinate
any planned audits.

The ARCADIS Safety Officer is Jerry Revis. Mr. Revis will be responsible for:

>	Ensuring that this project is carried out in accordance with all permit and EPA safety
requirements.

>	Ensuring that anyone working on the project has fulfilled all of the safety training
requirements.

A5 BACKGROUND

The ETV Program's AMS Center conducts third-party performance testing of commercially
available technologies that detect or monitor natural species or contaminants in air, water, and
soil. Stakeholder committees of buyers and users of such technologies recommend technology
categories, and technologies within those categories, as priorities for testing. Dioxin EMSs were
identified as a priority technology category through the AMS Center stakeholder process, since
these emerging technologies have significant potential to improve upon the standard method for
the determination of dioxins in flue gas.

EPA Method 231 is the certified extractive method used for quantification of dioxin emissions
from incinerators in the United States as well as in many other countries. This method is labor
intensive, expensive, and requires an extended time for subsequent laboratory analysis of
collected samples. As a result, Method 23 measurements are made infrequently (-once each
year) for compliance purposes and not for long-term or short-term performance monitoring.
New emerging technologies are being developed to provide semi-continuous monitoring or long-
term sampling of dioxins, and may have the potential to provide more information on dioxin


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 14 of 48
Version: 1.0
September 2, 2005

source emissions than the relatively few samples required under federal or state regulations.
However, the performance of these newly introduced technologies has not been evaluated in the
United States to determine their relative operational capabilities.

The purpose of this verification test is to generate performance data on dioxin EMS technologies
so organizations and users interested in installing and operating dioxin EMSs on their municipal
waste incinerators and industrial plants can be assured of their benefit. The test will be
conducted over a period of approximately two weeks and will involve the continuous operation
of several dioxin EMSs at a well-controlled, pilot-scale boiler facility located at EPA laboratories
in Research Triangle Park, North Carolina. The accuracy and range of the EMSs will be
determined through comparisons to the standard EPA integrated sampling method for dioxin.
Other performance parameters such as data completeness, maintenance requirements, ease of
use, and operational costs will be determined from operator observations. This test is not
intended to simulate long-term performance of these technologies on a full-scale incinerator or
plant.

A6 VERIFICATION TEST DESCRIPTION AND SCHEDULE
A6.1 Verification Test Description

In general, the dioxin EMS technologies to be evaluated in this verification test can be
categorized into two groups:

(1)	technologies that collect long-term integrated samples of dioxin from flue gas onto
sorbent media, for subsequent laboratory analysis, and

(2)	technologies that allow for continuous or semi-continuous sampling and automated,
on-site analysis of dioxins or marker compounds using in-situ laser ionization/mass
spectrometric techniques.

Typically, the technologies which collect integrated samples for laboratory analysis consist of a
sampling unit which mounts to the duct and isokinetically samples the flue gas, and a control
unit that remotely controls the gas sampling based on pre-set sampling times. In addition to


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 15 of 48
Version: 1.0
September 2, 2005

sample collection, these systems can be configured to measure a variety of flue gas parameters
and can be programmed to automatically stop sampling in the event of unusual operating
conditions. After sampling, the sample media are retrieved and sent to a laboratory for analysis.

The operating principles of the laser-ionization/mass spectrometric technologies include laser-
induced ionization of a molecule of interest using a multi-step resonant process, followed by
mass spectrometric measurement of the resulting ions. Typically, this process involves
absorption of a single ultra violet (UV) photon followed by the subsequent absorption of a
second UV photon to bring the internal energy of the molecule above its ionization energy, and
resulting in the formation of a molecular ion. After ionization, the created ions are extracted into
a time-of-flight (TOF) mass spectrometer and detected. Generally, the spectral absorption
"fingerprint" of a specific molecule/ion is sufficiently unique for unambiguous identification of
target molecules and can be obtained by systematically changing the wavelength of the laser
radiation.

This verification test will involve the simultaneous evaluation of multiple dioxin EMSs under
realistic operating conditions on a pilot-scale boiler. The EMSs will be operated for
approximately two weeks, during which time a series of Method 23 reference samples will be
collected. The operational parameters of the boiler will be systematically varied during testing to
provide a range of expected dioxin concentrations in the flue gas.

In performing the verification test, Battelle will follow the technical and QA procedures
specified in this test/QA plan and will comply with the data quality requirements in the AMS
Center QMP.2

A6.2 Verification Test Schedule

Table A1 shows the planned schedule of testing and data analysis/reporting activities to be
conducted in this verification. The verification test of dioxin EMSs is planned to be conducted
in September 2005, with installation of the EMSs at the host facility in late August 2005.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 16 of 48
Version: 1.0
September 2, 2005

Table A1. Planned Verification Test Schedule

Date(s)

Testing Activities

Data Analysis and Reporting

August 29-
September 9

Set up/install EMSs
EMS shakedown

Begin preparation of report template

September 12-
September 23

Routine operation
Reference sampling periods
Remove EMSs from host facility
Begin analysis of first reference samples

Review and summarize operator

observations

Compile data from EMSs

October 30

Complete analysis of reference samples

Complete common sections of reports

November 30



Complete draft reports
Vendor review of draft reports

January 30, 2006



Revise draft reports

Peer review of draft reports

Submit final reports for EPA approval

Subsequent to the verification test, a separate verification report will be drafted for each EMS.
These reports will be reviewed by the respective vendors and by peer reviewers, and submitted to
EPA for final signature. The period of operation of the EMSs at the facility will be
approximately 2 weeks. Installation of the EMSs is expected to begin during the week of August
29, 2005 and is expected to be completed by September 9. Routine operation of the EMSs is
expected to begin on September 12 and continue until September 23, 2005, or until all testing
activities are completed. During testing, it is anticipated that at least one set of reference samples
will be collected each day. Depending on timing, more than one set of reference samples may be
collected in one day.

The test procedures are described in Section B of this test/QA plan. Changes in the operational
conditions of the boiler will be introduced to provide a range of expected dioxin concentrations
in the flue gas.

A6.3 Test Facility

A 2.94 MBtu/hr, 3-Pass Wetback Scotch Marine Packaged Boiler (SMPB) manufactured by
Superior Boiler Works, Inc., and located at the EPA RTP facility, will be used for the
verification test. This boiler is capable of firing natural gas or a variety of fuel oils. The oil
burner used is a low pressure, air atomizing nozzle that delivers a fine spray at an angle which


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 17 of 48
Version: 1.0
September 2, 2005

ensures proper mixing with the air stream. The burner can be set to fire automatically or
manually at any desired rate between the minimum and the maximum firing rates. Fuel oil
temperature can be adjusted using an electric heater to maintain proper viscosity; the fuel and
atomizing airflow rates are variable to ensure adequate oil atomization. The boiler has 33-square
meters of heating surface and generates up to 1,090 kg/hr of saturated steam at pressures up to 15
psig. Fuel flows are measured with a liquid volume totalizer and stoichiometric ratios are
verified through O2 and CO2 emission concentrations. The SMPB is shown in Figure 1.

During this verification test, the SMPB will be fully instrumented with continuous emission
monitors (CEMs) for a variety of species including 02, CO, H20, HC1, and S02. Continuous
emission monitoring of chemical species is performed with two shared CEMs for the package
boiler facility. The first CEM bench includes four gas analyzers: high range CO, low range CO,
O2, and CO2, each with multiple ranges. HC1, and SO2, on the other hand, will be measured by a
self-contained bench-scale CEM system (Bodenseewerk). The system uses an Altech Hot/Wet
(FEW) sampling system and a Perkin-Elmer (PE) MCS-100 Infrared (IR) Multi-Component
Analyzer. The MCS is capable of measuring up to eight compounds simultaneously, using gas
filter correlation and single beam dual wavelength techniques. The compounds that can be
measured include HC1, NH3, N20, H20, N02, CO, C02, and S02. The HW probe assembly
provides functions for reliable sampling of flue gases, while closely maintaining temperatures at
elevated levels. The functions include probe blowback with instrument air, calibration gas
injection, and processes to protect the system from corrosion. The Altech HW sampling system
is designed to maintain elevated temperatures, up to 250 °C, throughout the entire gas analyzer.
Monitored concentrations are continuously output as linear 4-20 mA signals.

The flue gas from the unit passes through a manifold to an air pollution control system (APCS)
consisting of a natural-gas-fired secondary combustion chamber, a fabric filter, and an acid gas
scrubber to ensure proper removal of pollutants. All emission measurements are taken prior to
the APCS. The SMPB facility is equipped with several sampling ports located at the exit of the
boiler. The vertical section of the duct (8 in. steel pipe) is sufficient in length and free of flow


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 18 of 48
Version: 1.0
" ' mber 2, 2005

Figure 1. Wetback Scotch Marine Packaged Boiler

disturbances so that particulate matter can be sampled at an axial location that meets EPA
Method IA particulate matter sampling requirements. Several sampling ports are located along
the horizontal section of duct approximately 3 meters above the facility catwalk. The horizontal
section of the duct (20 cm. steel pipe) is also sufficient in length and free of flow disturbances so
that particulate matter can be sampled at an axial location that meets EPA Method 1A particulate
matter sampling requirements. The boiler stack has been modified to accommodate sampling
stations for each of the EMSs and for Method 23 sampling trains.

A surrogate chlorinated chemical (1,2-dichlorobenzene) and a source of metal atoms (copper
naphthenate) will be added to the boiler fuel to promote dioxin formation for the EMS testing.4
A surrogate feed system was designed to safely tap the surrogate feed line to the fuel line just
before the burner nozzle. The feed system consists of a 37 liter pressurized stainless steel tank, in


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 19 of 48
Version: 1.0
September 2, 2005

which the surrogate and the copper naphthenate are mixed. A series of check valves were
incorporated in the feed system to avoid contaminating the fuel oil recirculation in case of an
accidental shutdown of the boiler, as well as a flow meter for measurement of the feed flow rate.
A solenoid valve electrically tied to the burner management/safety system is included to stop any
surrogate feed to the burner when the burner is off. The pressurized stainless steel tank is
contained in a secondary container to minimize any spill that may occur during the surrogate
addition process. The amount of copper to be injected with the surrogate is calculated to simulate
the copper content in ash of refuse-derived fuel, generating approximately 20% ash and an
estimated concentration of copper of 100 mg per kg of ash. Since the mixture is prepared for
different test conditions, the concentration of copper injected varies with the surrogate injection
rate, the chlorine content in the boiler being the main driver. This injection rate is determined by
the calculated HC1 concentration to be maintained in the flue gas. Prior to testing, the boiler
emissions were characterized under a series of test conditions similar to those that will be used in
the verification test. The results of the boiler characterization are presented below.

Table A2 summarizes the characteristics of the flue gas from the EPA boiler at the point where
the dioxin EMSs will be installed. These characteristics show the average and the range of
several key constituents.

Table A2. Stack Gas Characteristics of the EPA SMPB

Parameter

Typical Value

Range or Maximum

o2(%)

2.2

0.5-4.0

co2 (%)

13.2

12-15

CO (ppm)

230

5-650

Calculated HCI (ppm)

463

35-1150

Total TEQ(a) (TEQ/dscm)

15.2

1.6-56

w' TEQ - Toxic Equivalents


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 20 of 48
Version: 1.0
September 2, 2005

A7 QUALITY OBJECTIVES AND CRITERIA FOR MEASUREMENT DATA

The objective of this verification test is to evaluate the performance of dioxin EMSs under
realistic operating conditions. This evaluation will in part assess the capabilities of the dioxin
EMSs for determining dioxins in the flue gas of a boiler which is cofiring copper napthenate and
dichlorobenzene, and will include a comparison of the EMS results to those of reference samples
collected by Method 23*and analyzed for dioxins according to a modified version of Method 23
as described in Section B4. Additionally, this evaluation will rely upon operator observations to
assess other performance characteristics of the EMSs. Below is a discussion of the quality
objectives and the criteria for measurement data that have been established to assure that the
objectives of this test are met.

A7.1 Quality Objectives

The data quality objectives indicate the minimum quality of data required to meet the objectives
of the verification of dioxin EMSs. The data quality objectives for this verification test include
those for the collection and analysis of reference samples, those for the operation of the boiler at
the test facility and consequently of the flue gas conditions, as well as those for the
documentation of operator observations. The data quality objectives for the collection and
analysis of the reference samples are based on the requirements of Method 231 and the
modifications to Method 23 described in Section B4, and are presented in terms of data quality
indicators (DQI) criteria for the critical measurements associated with those methods. The data
quality of the reference samples relies, in part, on the proper operation of the test boiler, and
consequently on the flue gas conditions. As such, data quality objectives for the operation of the
boiler have been established for this verification test and are defined in terms of DQI goals for
the boiler controls, and for the CEMs which monitor the flue gas conditions. The data quality
objectives for the operator observations have not been defined quantitatively but have been
incorporated into documentation requirements and data review, verification, and validation
requirements for this verification test.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 21 of 48
Version: 1.0
September 2, 2005

A7.2 Criteria for Measurement Data

Table A3 presents the DQIs and criteria for the critical measurements of the reference method.
Table A4 presents the criteria for DQIs for several important but non-critical measurements
made by the CEMs at the SMPB facility.

The quality of the reference measurements will be assured by adherence to these DQI criteria
and the requirements of Method 23 with the exception of the modifications described in Section
B4, including the QA/QC requirements of those methods. The quality of the reference
measurements will be monitored by inclusion of blank samples and performance evaluation (PE)
samples (Section Cl.l) as appropriate. The quality of the data relating to the boiler operation
and the flue gas conditions will be assured through the accurate measurement of O2, CO, H20,
dopant flow injection rate, and total flue gas flow rate.

Various calibration requirements and QA/QC checks are associated with Method 23 and are
discussed in detail in Sections B2-B7 of this test/QA plan. Dioxin standards to be used in
calibration standards and QC samples must meet National Institute of Standards and Technology
(NIST) traceability, when available. Acceptance criteria for PE samples and PE audit
measurements are given in Section Cl.l.

The Battelle Quality Manager or his designee will perform a technical systems audit (TSA) at
least once during this verification test to augment these QA/QC requirements. The EPA Quality
Manager also may conduct an independent TSA, at her discretion.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 22 of 48
Version: 1.0
September 2, 2005

Table A3. DQIs and Criteria for Critical Measurements for EPA Method 23.

Measurement

DQI

Criteria

Internal standard recovery

Accuracy

Recovery:

40 to 130% fortetra- through hexachlorinated
compounds;
and

25 to 130% for hepta- and octachlorinated
compounds

Surrogate standard recovery

Accuracy

Goal of 70 to 130% recovery, otherwise
correction is necessary

Flue gas flow rate

Precision

±5%

Flue gas temperature

Precision

l+
cn

o

O

Flue gas pressure

Precision

±2%

Leak check

Bias

Leak rate of less than 0.02 ftVmin

Field blanks

Bias

if blank >30% of sample concentration, data
must be flagged

GC column performance

Bias

Retention times within 10 seconds of expected
retention times

Solution blanks

Bias

Blank <10% of sample concentration or <10 x

LOD, subtract blank from samples;
if blank >10% of sample concentration, data
must be flagged

Duplicate reference samples

Precision

Concentration differences < 30%

Table A4. DQI Goals for SMPB Continuous Emission Monitors

Measurement Parameter

Method

Precision1
(RSD) (%)

Bias2 (%)

Completeness (%)

o2

CEM

<7

<10

>90

CO

CEM

<20

<10

>90

h2o

CEM

<20

<10

>90

1Analyzer precision will be assessed from repeated measurements under nominally constant operation
(i.e., during bias testing).

2Analyzer bias is assessed daily; it measures the degree of disagreement between an averaged
measurement and an accepted reference value, expressed as a percentage of the reference value.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 23 of 48
Version: 1.0
September 2, 2005

A8 SPECIAL TRAINING/CERTIFICATION

Documentation of training related to technology testing, field testing, data analysis, and reporting
is maintained for all Battelle technical staff in training files at their respective locations.
Documentation of the expertise and experience of ARCADIS staff in Method 23 sampling and
analysis is similarly available. The Battelle Quality Manager may verify the presence of
appropriate training records prior to the start of testing. If Battelle or ARCADIS staff operate
and/or maintain an EMS during the verification test, the EMS vendor will be required to train
those staff prior to the start of testing. Battelle will document this training with a consent form,
signed by the vendor that specifies which Battelle/ARCADIS staff have been trained on their
EMS. Battelle technical staff will have a minimum of a bachelor's degree in science/engineering
or have equivalent work experience.

A9 DOCUMENTATION AND RECORDS

The records for this verification test will include the test/QA plan, chain-of-custody forms,
laboratory record books (LRB), data collection forms, electronic files (both raw data and
spreadsheets), and the final verification report. All of these records will be maintained at the
host facility or in the Verification Test Coordinator's office during the test and may be
transferred to permanent storage at Battelle's Records Management Office (RMO) at the
conclusion of the verification test. All Battelle LRBs are stored indefinitely, either by the
Verification Test Coordinator or Battelle's RMO. EPA will be notified before disposal of any
files. The documentation and results of the Method 23 measurements made by ARCADIS will
be submitted to Battelle after completion of all sample analyses, review of the data, and
calculation of dioxin concentrations in the flue gas. Section BIO further details the data
recording practices and responsibilities.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 24 of 48
Version: 1.0
September 2, 2005

SECTION B
MEASUREMENT AND DATA ACQUISITION

B1 EXPERIMENTAL DESIGN

This test will specifically address verification of EMSs for dioxins in flue gas by evaluating the
accuracy and operational range of the EMS measurements, as well as the ease of use, reliability,
and maintenance needs of each EMS. Specifically, the dioxin EMSs will be evaluated for the
following performance parameters:

Relative Accuracy
Range

Data Completeness

Operational factors such as maintenance, ease of use, reliability, and operational
costs.

Relative accuracy and range will be determined for each EMS by comparison of EMS results to
results from Method 23 reference samples collected simultaneously with the EMS
measurements. Range will be determined from measurements over a variety of defined
operating conditions expected to produce differing levels of dioxins. Data completeness will be
assessed as the percentage of maximum data return that is achieved by each EMS over the test
period. Operational factors will be evaluated by means of operator observations, and records of
needed maintenance, vendor activities, and expendables used.

Bl.l Test Procedures

The following sections describe the test procedures that will be followed during the verification
test.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 25 of 48
Version: 1.0
September 2, 2005

Bl. 1.1 Relative Accuracy

The relative accuracy of each dioxin EMS will be evaluated by comparison of EMS results to
simultaneous results obtained by sampling the flue gas with Method 23. During the verification
test a series of nine (9) Method 23 test runs will be conducted using duplicate Method 23 trains.
The Method 23 trains will sample from ports located at each end of the sampling region where
the EMSs are installed. The reference samples will be recovered and submitted for analysis by
the modified version of Method 23 described in Section B4. The dioxin concentrations
determined by the reference methods will be compared to corresponding results from each EMS,
averaged over the period of each Method 23 test run. During each of the test runs the boiler
operation will be maintained as constant as possible based on measurements of the flue gas
conditions and boiler control parameters (e.g., dopant injection rate). However, the duration of
the sampling periods and the operating conditions of the boiler will be changed from run to run
to provide a range of conditions under which the EMSs will be evaluated. Two sets of operating
conditions will be used for the test runs to generate expected high and low dioxin concentrations.
Test runs of various durations will be conducted under each set of operating conditions. The
duration of the sampling periods for the Method 23 test runs will be varied to assess the ability of
the EMSs to accurately determine dioxin concentrations over a range of sampling times.

Sampling periods of four hours will be used to assess short-term accuracy of the EMSs, whereas
long-term accuracy of the EMSs will be assessed from composite samples collected over two 8-
hour sampling periods on successive days (i.e., totaling 16 hours per sample). For samples
collected over multiple days, the Method 23 trains used for the sample collection will be
removed from the duct after each session. The samples will be recovered and analyzed for each
8-hour session. The EMSs that collect long-term samples will be shut off after each 8-hour
session but the sampling media may, at the vendor's discretion, remain in place until sampling
has been completed for the appropriate number of 8-hour sessions. The results of the long term
EMS samples will be compared with the cumulative average of the appropriate 8-hour reference
samples. Table Bl shows the sampling durations and boiler operating conditions for each of the
test runs, and Table B2 shows a tentative schedule for completion of the test runs. Note: The
test runs may not be completed in the order shown in the tables.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 26 of 48
Version: 1.0
September 2, 2005

Table B1. Test Run Summary

Test Run

Sampling Duration

Expected Dioxin Concentration'3'

1

4 hours

Low

2, 3

16 hours (2x8 hours)

High

4

4 hours

Low

5

8 hours

Low

6

4 hours

High

7, 8

16 hours (2x8 hours)

Low

9

4 hours

High

(a) - Expected concentrations based on results of baseline testing. High corresponds to expected concentrations near
the upper end of the range, and low corresponds to expected concentrations near the lower end of the range.

Table B2. Tentative Schedule for Collection of Reference Samples

Test Run

Week 1

Week 2

M

T

w

Th

F

M

T

w

Th

F

1

X



















2, 3



X

X















4







X













5









X











6











X









7, 8













X

X





9

















X



Two Method 23 trains will be used to collect each reference sample during each test run. These
trains will each sample isokinetically from a single point in the gas flow, with one of the trains
sampling at each end of the sampling region where the EMSs are sampling. To assure
comparability of the EMS and Method 23 results, each reference method sampling run will start
no sooner than a time previously agreed upon with the EMS vendors. The vendors, Battelle


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 27 of 48
Version: 1.0
September 2, 2005

staff, and ARCADIS staff will be given at least 15 minutes notice prior to the start of each
reference sampling run. However, there will be no obligation to delay the start of a reference
method run because of a lack of readiness on the part of an EMS vendor. Vendors will similarly
be notified as the end of each Method 23 run approaches, so that they can stop sampling or
define the reporting period for data, as appropriate for their EMS.

Upon completion of each test run, the Method 23 trains will be dismantled for sample recovery
in the field by ARCADIS staff, and all collected sample fractions will be logged and stored for
transfer to the analytical laboratory. All sample handling, QA/QC activities, and dioxin analyses
will be conducted by ARCADIS or EPA staff, adhering to all requirements of Method 23 and the
modifications described in Section B4 of this test/QA plan. Subsequent to analysis, ARCADIS
will review the data, and report final dioxin concentrations from all trains in units of toxic
equivalents per dry standard cubic meter (TEQ/dscm), corrected to 7% O2. The results from the
simultaneously collected Method 23 trains will be used to assess the degree of dioxin loss (if
any) in the duct between the two reference method sampling ports. Unless discrepancies of
greater than 30% are observed for total measured TEQs between the reference samples collected
simultaneously, the results from the reference method samples will be averaged together, to
produce the final reference data used for comparison to the EMS results. If discrepancies of
greater than 30% are observed, the data will be flagged and the samples will be treated as
independent samples for comparison to the EMSs.

Bl.1.2 Range

No additional test procedures will be carried out specifically to address range. Rather, this
parameter will be assessed in terms of relative accuracy over the range of measured dioxin
concentrations and sampling periods. The reference method samples will be collected over a
range of expected dioxin concentrations in order to assess the degree of agreement of each of the
EMSs with the reference method under a range of conditions and sampling times. Based on
results from baseline testing of the boiler conducted prior to the verification test, the dopant
injection rate and firing conditions will be changed for different Method 23 runs to achieve


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 28 of 48
Version: 1.0
September 2, 2005

different expected dioxin concentrations. Additionally, the duration of the test runs will be
varied to achieve a range of sampling periods. The flue gas HC1 level will be used as an indicator
of the expected dioxin concentrations in the flue gas.

B 1.1.3 Data Completeness

No additional test procedures will be carried out specifically to address data completeness. This
parameter will be assessed based on the overall data return achieved by each EMS.

Bl.1.4 Operational Factors

Operational factors such as maintenance needs, data output, consumables used, ease of use,
repair requirements, etc., will be evaluated based on observations recorded by Battelle and
facility staff, and in some cases by the EMS vendors. A laboratory record book will be
maintained at the test facility, and will be used to enter daily observations on these factors.
Examples of information to be recorded in the record books include the daily status of diagnostic
indicators for the EMS; use or replacement of any consumables; the effort or cost associated with
maintenance or repair; vendor effort (e.g., time on site) for repair or maintenance; the duration
and causes of any EMS down time or data acquisition failure; and operator observations about
ease of use of the EMS. These observations will be summarized to aid in describing EMS
performance in the verification report on each EMS.

B1.2 Statistical Analysis

The statistical methods and calculations used for evaluation of the quantitative performance
parameters are described in the following sections.

Bl.2.1 Relative Accuracy

The relative accuracy (RA) of the EMSs with respect to the reference sample results will be
assessed as a percent bias, using Equation 1:


-------
d

+ t

0.975

RA=-





Dioxin Emission Monitoring Systems
Test/QA Plan
Page 29 of 48
Version: 1.0
September 2, 2005

x 100 (1)

where:

d

= the absolute value of the mean of the differences between the EMS and

reference sample results for each test run,
t0 975 = the lvalue,

Sd = the standard deviation of the differences between the EMS and reference sample
results for each test run, and

RM = the mean of the reference method results.

Bl.2.2 Range

The measurement range of the EMSs will be reported in terms of the accuracy of the EMSs
relative to the reference method under the variety of boiler operating conditions and sampling
durations used during the test runs.

B 1.2.3 Data Completeness

Data completeness will be calculated as the percentage of the total possible data return over the
entire field period that is achieved by each EMS. For each EMS that collects real-time data this
calculation will use the total hours of data recorded, divided by the total hours of data in the
entire field period. For EMSs that collect integrated samples, data completeness will be
assessed in terms of the percentage of successfully recovered samples collected during the
corresponding test runs. The causes of any substantial incompleteness of data return will be
established from operator observations or vendor records, and noted in the discussion of data
completeness results.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 30 of 48
Version: 1.0
September 2, 2005

B1.3 Reporting

The statistical comparisons described above will be conducted separately for each of the EMSs
being tested, and information on the operational parameters will be compiled and reported. The
data for each EMS will be kept separate from data for all other EMSs, and no intercomparison of
the data from different EMSs will be performed at any time. A separate verification report will
be prepared for each EMS tested, that presents the test procedures and test data, as well as the
results of the statistical evaluation of those data.

Operational aspects of the EMSs will be recorded by testing staff at the time of observation
during the field test, and summarized in the verification report. For example, descriptions of the
data acquisition procedures, use of vendor-supplied proprietary software, consumables used,
repairs and maintenance needed, and the nature of any problems will be presented in the report.
Each verification report will briefly describe the ETV program, the AMS Center, and the
procedures used in verification testing. The results of the verification test will be stated
quantitatively, without comparison to any other EMS tested, or comment on the acceptability of
the EMS's performance. Each draft verification report will first be subjected to review by the
respective EMS vendor, then revised and subjected to a review by EPA and other peer reviewers.
The peer review comments will be addressed in further revisions of the report, and the peer
review comments and responses will be tabulated to document the peer review process. The
reporting and review process will be conducted according to the requirements of the ETV/AMS
Center QMP.2

B2 SAMPLING METHOD REQUIREMENTS

The collection of reference samples will be conducted as described above (Section B 1.1.1) by
ARCADIS, under subcontract to Battelle, according to the requirements of Method 23.1 Method
23 is the standard sampling method for dioxins from municipal waste combustors. The method
uses sampling trains that consist of a heated probe, heated box containing a cyclone and a filter,
water-cooled condenser, water-cooled XAD-2 resin cartridge, impinger train for water


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 31 of 48
Version: 1.0
September 2, 2005

determination, leak-free vacuum line, vacuum pump, and a dry gas and orifice meter with flow
control valves and vacuum gauge. During the verification test, temperatures will be measured
and recorded in the hot box (set at 125 °C), at the impinger train outlet, at the XAD-2 cartridge
outlet (maintained to be below ambient temperature) and at the inlet and outlet of the dry gas
meter. Leak checks will be conducted at the beginning and end of each sample run to ensure
integrity of the sampling train. Prior to sampling, all glassware, probe, glass wool and aluminum
foil will be cleaned following the Method 23 cleaning procedure, and the XAD traps will be
spiked with carbon-13 labeled dioxin surrogate standards according to Method 23 to assess the
efficiency of sample recovery from the sampling train. Sampling will be conducted at a
sampling rate of approximately 0.75 cubic feet per minute (CFM), based on isokineticity
between sampling nozzle flow and flow from the source. Section B5 describes the QA/QC
requirements of Method 23.

The preparation of calibration and QC samples, and the analysis of samples for dioxins will be
carried out according to Method 23 as described in Section B5. In addition, independent audits
of sampling procedures will be carried out by Battelle as part of the technical systems audit
procedure (Section CI.2) and the performance evaluation audit procedure (Section Cl.l).

B3 SAMPLE HANDLING AND CUSTODY REQUIREMENTS

Following completion of each Method 23 run, each sampling train will be recovered in a clean
area, and the cleanup procedure will begin as soon as the probe is removed from the sample
source location. During the transportation between the test facility and the designated recovery
area, both ends of the heated probe and openings of the impinger assembly will be sealed with
aluminum foil or glass caps.

The sample fractions from each train shall include the following:

> Filter


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 32 of 48
Version: 1.0
September 2, 2005

>	XAD-2 Resin Cartridge

>	Front-half acetone/dichloromethane rinse collection jar

>	Front-half toluene rinse collection jar

>	Back-half acetone/dichloromethane rinse collection jar

>	Back-half toluene rinse collection j ar.

The filter will be recovered and placed in a Petri dish that is sealed with Teflon tape. The probe
and front half of the filter housing will be rinsed with acetone followed by dichloromethane. The
solvents will be collected in a single 250 mL amber jar. This fraction will be designated as the
acetone/dichloromethane front-half rinse sample. The probe and filter housing will then be
rinsed with toluene. The toluene will be collected in a separate 250 mL amber jar. This sample
will be designated as the toluene front-half rinse sample. The collection of the rinses in separate
bottles is a modification to Method 23.

The XAD-2 cartridges will be kept refrigerated prior to use and during transport to the facility to
prevent evaporation of the pre-sampling surrogate standards. After sampling, the XAD-2 resin
cartridge from each train shall be capped at both ends and wrapped in aluminum foil during
transport. This fraction of the sampling train will be designated as the XAD-2 fraction. As with
all sample fractions, the XAD-2 fractions will remain refrigerated during storage and transport.
The back half of the filter housing, glass connection and condenser will be rinsed with acetone
followed by dichloromethane. The solvents will be collected in a single 250 mL amber jar. This
fraction will be designated as the acetone/dichloromethane back-half rinse sample. This
glassware will then be rinsed with toluene and the solvent will be collected in a separate 250 mL
amber jar. This fraction will be designated as the toluene back-half rinse sample. The solvent
rinse jars shall be capped with Teflon lined caps and sealed with Teflon tape to prevent leakage
and evaporation during transport. The recovered samples will be uniquely identified for each
test run and stored in a refrigerated space before they are sent for analysis. The samples shall be
refrigerated during transport to the analytical laboratory.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 33 of 48
Version: 1.0
September 2, 2005

All reference samples will be in the custody of ARCADIS from sample collection through
sample recovery and in the custody of EPA staff for analysis. Recovered samples will be carried
by ARCADIS staff to the EPA laboratory for analysis. Sample custody will be documented
throughout collection, recovery, and analysis of the reference samples, using standard forms used
by ARCADIS for this purpose. Each chain-of-custody form will be signed by the person
relinquishing samples once that person has verified that the chain-of-custody form is accurate.
Upon receipt at the laboratory, chain-of-custody forms will be signed by the person receiving the
samples once that person has verified that all samples identified on the chain-of-custody forms
are present in the shipping container. Any discrepancies will be noted on the form and the
sample receiver will immediately contact the ARCADIS technical lead to report missing, broken,
or compromised samples. Copies of all chain-of-custody forms will be delivered to the
Verification Testing Coordinator upon request, and maintained with the test records.

B4 ANALYTICAL METHOD REQUIREMENTS

Analysis of the reference samples for dioxins will be conducted at the EPA RTP Campus, using
a modified version of Method 23. The modifications to Method 23 that will be followed for this
verification test include:

>	Analysis will be completed by high resolution gas chromatography/low resolution
mass spectrometry (HRGC/LRMS). (The expected resolution for the LRMS is
approximately 400, rather than 10,000 as stated for high resolution mass
spectrometry.)

>	Mass locking will not be used with LRMS.

The extraction and cleanup procedures for the target compounds of interest shall follow Method
23, with the exception that:


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 34 of 48
Version: 1.0
September 2, 2005

>	The front and back halves of the reference samples will be extracted and analyzed
together rather than separately.

>	The internal, surrogate, and recovery standards that are used in the modified method
include several that are not included in the standard method. A list of these standards
is presented in Appendix A.

ARCADIS will coordinate the analysis of the reference method samples, which will be
conducted by EPA staff at the EPA RTP facility. EPA staff will be responsible for ensuring that
the calibration of the analytical instrumentation and the analysis of the samples are conducted
according to the requirements of the modified Method 23, and for ensuring that the appropriate
QA/QC activities are conducted according to the method (see Section B5). ARCADIS will
ensure that the calibration records for any instrumentation used are maintained and will be
responsible for providing Battelle with documentation on calibration and quality control of the
reference analyses, upon request.

B5 QUALITY CONTROL REQUIREMENTS

As described in Section A7, reference dioxin sampling will be carried out using Method 23,1 and
will be subject to the data quality criteria of that method. The analysis of the reference samples
will be conducted according to Method 23, including the modifications described in Section B4,
and will be subject to the data quality criteria of that method. Table B3 summarizes the quality
control requirements of those two methods. If the sampling or analytical performance strays
outside the required tolerances, the relevant QC checks will be conducted again or the relevant
QC samples will be prepared again and reanalyzed. If performance problems persist, the
reference instrument will be recalibrated, and/or affected samples will be reanalyzed. Reference
sample results not meeting the requirement will be excluded from comparison to the dioxin EMS
results.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 35 of 48
Version: 1.0
September 2, 2005

Table B3. QC Checks for Method 23

Measured Parameter

QC Check

Required Performance

Overall cleanup and analysis
method efficiency

Internal standard
recovery

Recovery:

40 to 130% fortetra- through hexachlorinated
compounds;
and

25 to 130% for hepta- and octachlorinated
compounds

Sampling train integrity

Leak check

Leak rate of less than 0.02 ffVmin

Sample train collection
efficiency

Surrogate standard
recovery

Goal of 70 to 130% recovery, otherwise
correction is necessary

GC column performance

Performance check
samples

Retention times within 10 seconds of expected
retention times

Reagent solution
contamination

Solution blanks

Blank <10% of sample or <10 x LOD, subtract

blank from samples;
if blank >10% of sample, data must be flagged

Sampling media
contamination

Field blanks

if blank >30% of sample, data must be flagged

Precision of Method 23
sampling trains

Duplicate reference
samples collected at
each end of
sampling duct

If concentrations do not agree within ±30% the
data must be flagged and samples treated
independently

B6 INSTRUMENT/EQUIPMENT TESTING, INSPECTION, AND MAINTENANCE

The equipment used for the reference sampling and analysis will be tested, inspected, and
maintained so as to meet the performance requirements established in Method 23. System
preventive maintenance will be performed prior to the start of each test as needed. All major
components will be checked to ensure operability and repaired when required. Laboratory
equipment maintenance is conducted as recommended by the manufacturer on an as-needed
basis. Daily calibrations of the CEMs will be conducted to ensure continued reliable operation
and provide the operator warnings of abnormal operation.

The EPA Metrology Laboratory, prior to the start of the sampling program, will calibrate field-
sampling equipment, such as Method 5 meter boxes for volumetric flow rates. Any leaks that


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 36 of 48
Version: 1.0
September 2, 2005

have developed will be repaired, parts will be lubricated as recommended by the manufacturer,
and manometers will be filled and checked for leaks. Replacement parts, including fuses, pumps,
spare tubing, compression fittings, etc., are maintained in the laboratory to minimize downtime.
Equipment manufacturers and overnight delivery services will be utilized for repair parts in
emergency situations.

If Battelle or ARCADIS staff operate and maintain the dioxin EMSs undergoing testing, those
activities will be done as directed by the vendor. Otherwise, operation and maintenance of the
EMSs will be the responsibility of the EMS vendors.

B7 INSTRUMENT CALIBRATION AND FREQUENCY
B7.1 Sampling Equipment Calibration

The instrumentation used for the reference sample collection and analysis will be calibrated per
the requirements stated in the EPA certified methods. EPA methods require that a laboratory
record be maintained of all calibrations. The calibration requirements of Method 23 include the
following minimal calibration activities:

>	Standard Pitot tube will be inspected and cleaned before each Method 23 test run.

>	The volume metering system will be calibrated within six months of use during the
verification test using a wet-test meter, as permitted in the Method.

>	All thermocouples and dial thermometers will be calibrated within six months of use
during the verification test. Thermometric fixed points (i.e. ice bath and boiling
water) are adequate standards for this task.

>	The portion of the volume metering system from the pump to the orifice meter will be
leak checked following each test, using the procedure described in EPA Method 5,
Section 8.4.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 37 of 48
Version: 1.0
September 2, 2005

>	Barometers will be calibrated within six months of use during the verification test by
reference to a mercury barometer or a local National Weather Service station.
Corrections will be made at a rate of -0.1 inches Hg per 100 feet of elevation above
sea level.

Additionally, to ensure the accurate measurement of the flue gas conditions,

>	The two self-contained bench-scale CEM systems at the SMPB will be calibrated at
the start of each sampling day, and verified at the end of each sampling day to correct
for possible sample drift and bias.

B7.2 Analytical Instrumentation Calibration

Prior to sample analysis, a calibration of the analytical instrumentation must be conducted
according to Section 6 of Method 23. Also, a daily performance check must be conducted daily
according to Section 6.1.2 of Method 23.

B7.3 Dioxin EMS Calibration

The dioxin EMSs undergoing testing will be calibrated initially by the respective EMS vendors
at the time of installation at the host facility. In the event that recalibration is necessary, that
recalibration will be carried out by the EMS vendor, or by Battelle staff under the direction of the
vendor. All calibrations performed will be documented by Battelle or host facility staff in the
project record book dedicated to the respective EMS.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 38 of 48
Version: 1.0
September 2, 2005

B8 INSPECTION/ACCEPTANCE OF SUPPLIES AND CONSUMABLES

All materials, supplies, and consumables will be ordered by the Verification Test Coordinator or
designee. Where possible, Battelle will rely on sources of materials and consumables that have
been used previously as part of ETV verification testing without problems. Battelle will also rely
on previous experience or recommendations from EPA advisors, ARCADIS staff, or EMS
vendors. Upon receipt of any supplies or consumables, the Verification Test Coordinator or his
designee will visually inspect and ensure that the materials received are those that were ordered
and that there are no visual signs of damage that could compromise the suitability of the
materials. If damaged or inappropriate goods are received they will be returned or disposed of
and arrangements will be made to receive replacement materials. Certificates of analysis (COA)
or other documentation of analytical purity will be checked for all gases, reagents, and standards
to ensure suitability for this verification test. Unsuitable materials will be returned or disposed of
and arrangements for the receipt of replacement materials will be made.

B9 NON-DIRECT MEASUREMENTS

No non-direct measurements will be used during this verification test.

BIO DATA MANAGEMENT

Various types of data will be acquired and recorded electronically or manually by Battelle,
vendor, and ARCADIS staff during this verification test. All data will be recorded in permanent
ink. Corrections to records will be made by drawing a single line through the entry to be
corrected and providing a simple explanation for the correction, along with a date and the initials
of the person making the correction. Table B4 summarizes the types of data to be recorded. All
maintenance activities, repairs, calibrations, and operator observations relevant to the operation
of the dioxin EMSs will be documented by Battelle or ARCADIS staff in laboratory record


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 39 of 48
Version: 1.0
September 2, 2005

Table B4. Summary of Data Recording Process

Data to Be
Recorded

Where Recorded

How Often
Recorded

By Whom

Disposition of Data

Dates, times, and
details of test
events, EMS
maintenance,
down time, etc.

ETV test notebooks

Start/end of test
procedure, and at
each change of a
test parameter or
change of EMS
status

Battelle and
ARCADIS

Used to

organize/check test
results; manually
incorporated in data
spreadsheets as
necessary

EMS calibration
information

ETV test notebooks,
or electronically

At EMS calibration
or re-calibration

Vendor,
Battelle, and
ARCADIS

Incorporated in
verification report as
necessary

EMS readings

Recorded
electronically by
each monitor and
then downloaded to
computer at the
close of each day.

Recorded
continuously, or as
determined by
each EMS.

EMS vendor,
for transfer to
Battelle

Converted to
spreadsheet for
statistical analysis
and comparisons

Integrated EMS

measurement

results

Electronically from
analytical method

Every sample
analysis

Analytical
laboratory

Converted to
spreadsheets for
calculation of flue
gas dioxin
concentrations, and
statistical analysis
and comparisons

Reference
method
procedures,
calibrations, QA,
etc.

Laboratory record
books, or data
recording forms

Throughout
sampling and
analysis processes

ARCADIS

Retained as
documentation of
reference method
performance

Reference
method analysis
results

Electronically from
analytical method

Every sample
analysis

ARCADIS

Converted to
spreadsheets for
calculation of flue
gas dioxin
concentrations, and
statistical analysis
and comparisons

books. Results from the reference method will be compiled by ARCADIS staff in electronic
format, and submitted to Battelle in the form of a sampling and analysis report at the conclusion
of reference dioxin analyses.

Records received by or generated by any Battelle or ARCADIS staff during the verification test
will be reviewed by a Battelle staff member within two weeks of receipt or generation,
respectively, before the records are used to calculate, evaluate, or report verification results. If a


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 40 of 48
Version: 1.0
September 2, 2005

Battelle staff member generated the record, this review will be performed by a Battelle technical
staff member involved in the verification test, but not the staff member who originally received
or generated the record. The review will be documented by the person performing the review by
adding his/her initials and date to the hard copy of the record being reviewed. In addition, any
calculations performed by Battelle or ARCADIS staff will be spot-checked by Battelle technical
staff to ensure that calculations are performed correctly. Calculations to be checked include any
statistical calculations described in this test/QA plan. The data obtained from this verification
test will be compiled and reported independently for each dioxin EMS. Results for EMSs from
different vendors will not be compared with each other.

Among the QA activities conducted by Battelle QA staff will be an audit of data quality. This
audit will consist of a review by the Battelle Quality Manager of at least 10% of the test data.
During the course of any such audit, the Battelle Quality Manager will inform the technical staff
of any findings and any immediate corrective action that should be taken. If serious data quality
problems exist, the Battelle Quality Manager will inform the AMS Center Manager who is
authorized to stop work. Once the assessment report has been prepared, the Verification Test
Coordinator will ensure that a response is provided for each adverse finding or potential
problem, and will implement any necessary follow-up corrective action. The Battelle Quality
Manager will ensure that follow-up corrective action has been taken.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 41 of 48
Version: 1.0
September 2, 2005

SECTION C
ASSESSMENT AND OVERSIGHT

CI ASSESSMENTS AND RESPONSE ACTIONS

Every effort will be made in this verification test to anticipate and resolve potential problems
before the quality of performance is compromised. One of the major objectives of this test/QA
plan is to establish mechanisms necessary to ensure this. Internal quality control measures
described in this test/QA plan, which is peer reviewed by a panel of outside experts,
implemented by the technical staff and monitored by the Verification Test Coordinator, will give
information on data quality on a day-to-day basis. The responsibility for interpreting the results
of these checks and resolving any potential problems resides with the Verification Test
Coordinator. Technical staff have the responsibility to identify problems that could affect data
quality or the ability to use the data. Any problems that are identified will be reported to the
Verification Test Coordinator, who will work with the Battelle Quality Manager to resolve any
issues. Action will be taken to control the problem, identify a solution to the problem, and
minimize losses and correct data, where possible. Independent of any EPA QA activities,

Battelle will be responsible for ensuring that the following audits are conducted as part of this
verification test.

Cl.l Performance Evaluation Audit

A Performance Evaluation (PE) audit will be conducted to assess the quality of the critical
measurements associated with the reference sampling and analysis methods. In the PE audit,
critical measurements associated with the reference methods will be checked by comparison with
an independent instrument, or an independent NIST-traceable standard. Table CI shows the
critical measurements to be audited, with the audit procedures and acceptance criteria for the
audit comparisons. If the PE audit results do not meet the acceptance criteria shown, they will be
repeated. If the outlying results persist, a change in reference instrument and a repeat of the PE


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 42 of 48
Version: 1.0
September 2, 2005

Table CI. Methods and Acceptance Criteria for PE Audit Measurements

Critical Measurement

PE Audit Method

Acceptance Criteria

Method 23 gas sample flow rate

Compare to independent flow
measurement device

±5%

Method 23 stack gas
temperature

Compare to independent temperature
measurement device

±2% absolute temperature

Barometric pressure

Compare to independent pressure
gauge

±1% absolute pressure

Dioxin internal standard recovery

Method spike with an independent
dioxin standard

40 to 130% for tetra-
through hexachlorinated

compounds; and
25 to 130% for hepta- and
octachlorinated
compounds

Dioxin surrogate standard
recovery

Field spike with an independent dioxin
standard

70 to 130% recovery

audit may be considered, and data will be flagged until the PE audit results are acceptable. This
audit will be performed once during the verification test, and will be the responsibility of the
Verification Test Coordinator or his designee.

The PE audit of the surrogate standard recovery will be performed by spiking one blank Method
23 train with a NIST-traceable dioxin solution, provided by Battelle. The spiked train will not be
used to collect a flue gas sample but will be recovered and analyzed in the same manner as for all
other Method 23 trains, and the analytical results will be compared to the spike amount to assess
recovery. The target criteria for this PE audit are 40-130% recovery of the surrogate standards
for the tetra-through hexachlorinated compounds, and 25-130% for the hepta- and octa-
chlorinated compounds. If these criteria are not met, the data will be flagged and noted in the
verification report.

The PE audit of the internal standard recovery will be performed by spiking one blank XAD-
cartridge with a NIST-traceable dioxin internal standard solution provided by Battelle rather than
the internal standard solution typically used by the laboratory. This spiked cartridge will be
extracted and analyzed in the same manner as for all the other cartridges. The target criterion for


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 43 of 48
Version: 1.0
September 2, 2005

this PE audit is 70-130% recovery of the internal standards. If this criterion is not met, the data
will be flagged and congener-specific correction factors will be applied to account for material
loss (or gain).

C1.2 Technical Systems Audits

The Battelle Quality Manager will perform a technical systems audit (TSA) at least once during
this verification test. The purpose of this audit is to ensure that the verification test is being
performed in accordance with the AMS Center QMP,2 this test/QA plan, published reference
methods, and any SOPs used by the test facility. In this audit, the Battelle Quality Manager, or
designee, may review the reference methods used, compare actual test procedures to those
specified or referenced in this plan, and review data acquisition and handling procedures. In the
TSA, the Battelle Quality Manager will tour the test site and EMS locations; observe the Method
23 sampling and sample recovery; inspect documentation of reference sample chain of custody;
and review laboratory record books. He will also check gas standard certifications and EMS data
acquisition procedures, and may confer with the EMS vendors and ARCADIS testing staff. A
TSA report will be prepared, including a statement of findings and the actions taken to address
any adverse findings. The EPA AMS Center Quality Manager will receive a copy of Battelle's
TSA report. At EPA's discretion, EPA QA staff may also conduct an independent on-site TSA
during the verification test. The TSA findings will be communicated to technical staff at the time
of the audit and documented in a TSA report.

CI.3 Data Quality Audits

The Battelle Quality Manager, or his designee, will audit at least 10% of the verification data
acquired in the verification test. The Battelle Quality Manager, or his designee, will trace the
data from initial acquisition, through reduction and statistical comparisons, to final reporting. All
calculations performed on the data undergoing the audit will be checked.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 44 of 48
Version: 1.0
September 2, 2005

CI.4 QA/QC Reporting

Each assessment and audit will be documented in accordance with Section 3.3.4 of the AMS
Center QMP.2 The results of the TSA will be submitted to EPA. Assessment reports will
include the following:

•	Identification of any adverse findings or potential problems

•	Response to adverse findings or potential problems

•	Recommendations for resolving problems

•	Confirmation that solutions have been implemented and are effective

•	Citation of any noteworthy practices that may be of use to others.

C2 REPORTS TO MANAGEMENT

The Battelle Quality Manager, during the course of any assessment or audit, will identify to the
technical staff performing experimental activities any immediate corrective action that should be
taken. If serious quality problems exist, the Battelle Quality Manager will notify the AMS Center
Manager, who is authorized to stop work. Once the assessment report has been prepared, the
Verification Test Coordinator will ensure that a response is provided for each adverse finding or
potential problem and will implement any necessary follow-up corrective action. The Battelle
Quality Manager will ensure that follow-up corrective action has been taken. The test/QA plan
and final report are reviewed by EPA AMS Center quality assurance staff and the EPA AMS
Center program management staff. Upon final review and approval, both documents will then be
posted on the ETV website (www.epa.gov/etv).


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 45 of 48
Version: 1.0
September 2, 2005

SECTION D
DATA VALIDATION AND USABILITY

D1 DATA REVIEW, VERIFICATION, AND VALIDATION REQUIREMENTS

The key data review and data verification requirements for this test are stated in Section BIO of
this test/QA plan. In general, the data review requirements specify that data generated during
this test will be reviewed by a Battelle technical staff member within two weeks of generation of
the data. The reviewer will be familiar with the technical aspects of the verification test but will
not be the person who generated the data. This process will serve both as the data review and the
data verification, and will ensure that the data have been recorded, transmitted and processed
properly. Furthermore, this process will ensure that the EMS data and reference method data
were collected under appropriate testing conditions and that the reference sample data meet the
specifications of Method 23.

The data validation requirements for this test involve an assessment of the quality of the data
relative to the DQIs and audit acceptance criteria specified for this test. The DQIs listed in
Section B5 will be used to validate the quality of the data. The QA audits described within
Section C of this document, including the performance evaluation audit and the audit of data
quality, are also designed to validate the quality of the data.

D2 VERIFICATION AND VALIDATION METHODS

Data verification is conducted as part of the data review as described in Section BIO of this
test/QA plan. A visual inspection of handwritten data will be conducted to ensure that all entries
were properly recorded or transcribed, and that any erroneous entries were properly noted (i.e.,
single line through the entry, with an error code and the initials of the recorder and date of entry).
Electronic data from the CEMs, EMSs, and other instruments used during the test will be


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 46 of 48
Version: 1.0
September 2, 2005

inspected to ensure proper transfer from the datalogging system. All calculations used to
transform the data will be reviewed to ensure the accuracy and the appropriateness of the
calculations. Calculations performed manually will be reviewed and repeated using a handheld
calculator or commercial software (e.g., Excel). Calculations performed using standard
commercial office software (e.g., Excel) will be reviewed by inspection of the equations used for
the calculations and verification of selected calculations by handheld calculator. Calculations
performed using specialized commercial software (i.e., for analytical instrumentation) will be
reviewed by inspection and, when feasible, verified by handheld calculator, or standard
commercial office software.

To ensure that the data generated from this test meet the goals of the test, a number of data
validation procedures will be performed. Section C of this test/QA plan provides a description of
the validation safeguards employed for this verification test. Data validation efforts include the
completion of QC activities, and the performance of TSA and PE audits as described in Section
C. The data from this test will be evaluated relative to the measurement DQIs described in
Section B5, and the PE audit acceptance criteria given in Section CI. 1 of this test/QA plan. Data
failing to meet these criteria will be flagged in the data set and not used for evaluation of the
EMS, unless these deviations are accompanied by descriptions of their potential impacts on the
data quality.

An audit of data quality will be conducted by the Battelle Quality Manager to ensure that data
review, verification, and validation procedures were completed, and to assure the overall quality
of the data.

D3 RECONCILIATION WITH USER REQUIREMENTS

This purpose of this verification test is to evaluate the performance of commercial dioxin EMSs.
In part, this evaluation will include comparisons of results from the EMSs to the results from
reference samples generated from a well-established EPA method for sample collection, and a


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 47 of 48
Version: 1.0
September 2, 2005

second well-established EPA method for sample analysis. To meet the requirements of the user
community, the reference data collected during this verification test should meet the QA
requirements of the reference methods. Additional performance data will be collected by testing
personnel regarding operational characteristics of the EMSs. To meet the requirements of the
user community, these data should include thorough documentation of the performance of the
EMSs during the verification test. The data review, verification, and validation procedures
described above will assure that data meeting these requirements is accurately presented in the
verification reports generated from this test, and will assure that data not meeting these
requirements will be appropriately flagged and discussed in the verification reports.

This test/QA plan and the resulting ETV verification report(s) will be subjected to review by the
EMS vendors, the host facility, EPA, and expert peer reviewers. The reviews of this test/QA
plan will assure that this verification test and the resulting report(s) meet the needs of potential
users and permitters of dioxin EMSs.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page 48 of 48
Version: 1.0
September 2, 2005

SECTION E
REFERENCES

1.	Method 23 - Determination of Poly chlorinated Dibenzo-p-dioxins and Poly chlorinated
Dibenzofurans from Municipal Waste Combustors. U.S. Environmental Protection Agency,
February 1991. Available at: http://www.epa.gov/ttn/emc/promgate/m-23.pdf

2.	Quality Management Plan for the ETV Advanced Monitoring Systems Center, Version 5.0,
U.S. EPA Environmental Technology Verification Program, Battelle, Columbus, Ohio,

March 2004.

3.	Environmental Technology Verification Program Quality Management Plan,
EPA/600/R-03/021, U.S. Environmental Protection Agency, Cincinnati, Ohio, December
2002

4.	George C. Clark, Michael Chu, Dahman Touati, Barry Rayfield, Jon Stone, and Marcus
Cooke, A Novel Low-Cost Air Sampling Device (Amb Stack Sampler) and Detection
System (CALUX Bioassay) for Measuring Air Emissions of Dioxin, Furan, and PCB on a
TEQ Basis Tested With a Model Industrial Boiler, Organohalogen Compounds, 40 (1999),
79-82.


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page A1
Version: 1.0
October 13, 2005

APPENDIX A
Modified Method 23 Spiking Scheme


-------
Dioxin Emission Monitoring Systems
Test/QA Plan
Page A2
Version: 1.0
October 13, 2005

Modified Method 23 Spiking Scheme
Internal Standards Pre Extraction Spike

13C12-2-MCDF

13C12-2-MCDD

13C12-2,4-DCDF

13C12-2,7-DCDD

13C12-2,4,8-TrCDF

13C12-2,3,7,8-TeCDF

13C12-2,3,7,8-TeCDD

13C12-1,2,3,7,8-PCDF

13C12-1,2,3,7,8-PCDD

13C12-1,2,3,6,7,8-HxCDF

13C12-1,2,3,6,7,8-HxCDD

13C12-l,2,3,4,6,7,8-HpCDF

13C12-l,2,3,4,6,7,8-HpCDD

13Ci2-1,2,3,4,6,7,8,9-OCDD

Surrogate Standards Pre-Sampling Spike on the XAD

13C12-2,8-DCDF

13C12-2,3-DCDD

13C12-2,3,7-TrCDD

37Cl4-2,3,7,8-TeCDD

13C12-2,3,4,7,8-PCDF

13C12-1,2,3,4,7,8-HxCDF

13C12-1,2,3,4,7,8-HxCDD

13C12-l,2,3,4,7,8,9-HpCDF

Recovery Standards Pre-Analysis Spike (Spiked after Clean-up)

13C12-l,2,3,4-TeCDD
13C12-1,2,3,7,8,9-HxCDD


-------