Total Coliforms and E. coli
Membrane Filtration Method

June 2003

Draft


-------
Total Coliforms and E. coli
Membrane Filtration Method

Sensitivity: 1 CFU/lOOmL
m-ColiBlue24® Broth
U.S. Environmental Protection Agency (USEPA) Approved for Drinking Water

Method No. 10029
Revision 3, 2/13/02

1.0	Scope and Application

1.1	This method determines the presence or absence of total coliforms and E. coli in finished potable
water using a selective and differential membrane filtration (MF) medium, m-ColiBlue24 Broth.

1.2	This method can detect the presence or absence of both total coliforms and E. coli simultaneously
within 24 hours and without the need for a confirmation step.

1.3	The detection limit of the method is one colony forming unit (CFU) of coliform bacteria per 100
mL of sample. See Attachment 1.3.

2.0	Summary of Method

2.1	Coliform bacteria are identified in water either as total coliforms or E. coli. Total coliforms can
be present in water without E. coli being present; E. coli cannot be present in water without total
coliforms also being present.

2.2	A 100-mL volume of sample is filtered through a 47-mm membrane filter using standard
techniques. The filter is then transferred to a 50-mm petri plate containing an absorbent pad
saturated with m-ColiBlue24 Broth. The filter is then incubated at 35°C ± 0.5°C for 24 hours. If
coliform bacteria are present in the sample, both red and blue colonies may appear; the blue
colonies are specific to the presence of E. coli.

2.3	m-ColiBlue24 Broth is a nutritive, lactose-based medium, containing inhibitors to selectively
eliminate growth of non-coliforms. It is analogous to an improved version of m-Endo. Total
coliform colonies growing on the medium are highlighted by a non-selective dye, 2,3,5-
Triphenoltetrazolium Chloride (TTC), which produces red colored colonies. Among the total
coliform colonies, which grow up on the medium, any E. coli colonies are distinguishable by a
selective blue color, resulting from the action of b-glucuronidase enzyme on 5-Bromo-4-Chloro-
3 -Indolyl-Beta-D-glucuronide (BCIG).

3.0	Definitions

3.1	Material Safety Data Sheet (MSDS) - Written information provided for each chemical reagent or
standard about a chemical's toxicity, health hazards, physical properties, flammability, and
reactivity. It also includes storage, spill, and handling precautions.

Draft

1

June 2003


-------
3.2	Total Coliform Bacteria - Bacteria belonging to the genera Klebsiella sp., Enterobacter sp.,
Citrobacter sp., or Escherichia sp.

3.3	Coliform Positive Colony - A red or blue colony.

3.4	Coliform Negative Colony - A clear or white colony.

3.5	Escherichia coli or E. coli Bacteria - A genus within the total coliform group typified by
possession of the enzyme b-Glucuronidase, ability to grow at 44.5°C, and form indole from
tryptophan.

3.6	E. coli Positive Colony - A blue colony.

3.7	E. coli Negative Colony - A non-blue colony.

4.0	Interferences

4.1	No interferences to the colony color development have been found in finished potable water
samples. Similarly, particulates in water samples do not alter the efficacy of the medium,
although excess particulates may cause colonies to grow together on crowded filters or slow the
sample filtration process.

5.0	Safety

5.1	Standard safety practices appropriate to microbiology laboratories should be followed. Follow the
test procedure carefully and observe all precautionary measures.

5.2	Solid and liquid waste materials containing or suspected to contain viable bacteria should be
decontaminated using an autoclave or by using an appropriate disinfectant before discarding.

5.3	Refer to the appropriate Material Safety Data Sheets supplied for each reagent for comprehensive
safety data essential to proper use.

6.0	Equipment and Supplies

6.1	Equipment

6.1.1	Air Incubator - Capable of operating at 35°C ± 0.5°C.

6.1.2	Vacuum pump.

6.1.3	Membrane filtration funnel unit and flask.

6.1.4	Dissecting microscope, capable of 10-15X magnification. The microscope should be
equipped with a fluorescent illuminator.

Draft

2

June 2003


-------
6.2 Supplies/Glassware - Cleanse all glassware thoroughly with a suitable detergent and hot water,

rinse with hot water to remove traces of detergent residual, and rinse again with laboratory-pure

water. Sterilize all glassware by autoclaving 15 min. at 121°C or by heating in an oven for at least

1 hour at 170°C.

6.2.1	Presterilized 50-mm MF petri plates with pads.

6.2.2	45-mm presterilized membrane filters.

6.2.3	Sterile Forceps.

6.2.4	Sterile glass or plastic sample collection containers.

6.2.5	Sterile graduated cylinders.

6.2.6	Sterile pipets.

6.2.7	Sterile MF filtration unit.

6.2.8	Side-arm flask.

6.2.9	Biohazard bag.

6.2.10150 mL of sterile rinse water.

7.0	Reagents and Standards

7.1	Growth Medium

7.1.1 The membrane filtration method for total coliforms and E. coli utilizes m-ColiBlue24

Broth, Hach product 26084-20, 26084-42, or 26084-50. The broth is provided in ready to
use 2-mL ampules or 100-mL glass bottles. Under proper storage conditions (2-8°C) the
broth has a shelf life of one year. The broth contains a nutritive medium and colorimetric
indicators.

7.2	Dechlorinating Reagent

7.2.1	Hach decholorinating reagent Powder Pillow (14363-69) contains sodium thiosulfate
(Na2S2C>3 ) and sodium sulfate (Na2SC>4).

7.2.2	To make a 3% sodium thiosulfate (Na2S2C>3 ) solution, add 48.18 g of sodium
thiosulfate pentahydrate to approximately 500 mL of deionized water, then dilute to 1 L
with deionized water.

Draft

3

June 2003


-------
8.0	Sample Collection, Dechlorination, Preservation, Shipment and
Storage

8.1	Water Sample Collection

8.1.1	Sample Collection Containers - Samples should be collected in sterile, clean glass or
heat-resistant bottles. Presterilized Whirl-Pak®1 Bags may also be used.

8.1.2	Sampling Procedure - Potable water samples are taken by first flushing the tap 2-3
minutes to clear service line. Collect samples using aseptic techniques to avoid
contamination. For other water samples, aseptically collect water representative of the
source.

8.2	Dechlorination - Water containing chlorine or other halogens must be treated with sodium
thiosulfate to allow accurate evaluation of microbial content. Add one decholorinating reagent
Powder Pillow (14363-69) by aseptically cutting off the tip of an alcohol-rinsed pillow and
pouring the contents into 100 mL of chlorinated water sample. Alternatively, pipet 0.1 mL of a
3% sodium thiosulfate (Na2S2C>3) solution into 100 mL of a chlorinated water sample.
Presterilized Whirl-Pak® Bags contain sufficient sodium thiosulfate powder to neutralize a 100-
mL chlorinated water sample.

8.3	Preservation, Shipment, Storage - Samples should be tested as soon as possible. If analysis cannot
be done within 1 hr of collection, potable water samples should be held on ice or at 2-8°C in a
refrigerator for a maximum holding time of 30 hours (16.1). Non-potable samples should be held
at <10°C for a maximum period of 8 hours from sampling to analysis.

9.0	Quality Control

9.1	m-ColiBlue24 Broth undergoes quality control (QC) testing at the time of manufacture. A
Certificate of Analysis is included with every m-ColiBlue24 product, stating that the
m-ColiBlue24 Broth, as received by the analyst, is ready for use in analyzing water samples by
the membrane filtration procedure. Each laboratory should test quality control using the following
procedure. Three ampules will be required, as well as membrane filtration equipment. The three
ampules will be used as an uninoculated blank, a positive control for both total coliforms and E.
coli, and a negative control.

9.2	The day before the QC procedure, inoculate three 10-mL Tryptic Soy Broth tubes with
Psuedomonas aeruginosa (ATCC 27853). Incubate at 35°C ± 0.5°C for 24 ± 4 hours.

9.3	Aseptically open three ampules of m-ColiBlue24 Broth and saturate the pads in three MF petri
dishes.

9.4	Filter 100 mL of sterile water and place filter on the first pad. This is the uninoculated blank.

* Whirl-Pak® is a registered trademark and patented product ofNasco; U.S. Patent 2,973,131.

Draft	4	June 2003


-------
9.5	Prepare serial dilutions as described in Standard Methods 9222B (16.2). Prepare a serial dilution
of the 24 hr. P. aeruginosa culture so that filtration of a 100-mL volume would theoretically yield
20-80 colonies. Filter the bacteria-containing diluent. After filtration, rinse the funnel with
several 20-3 0-mL volumes of sterile rinse water. Place membrane filter into the second plate of
m-ColiBlue24 Broth; this is the negative control. Prepare appropriate serial dilutions of the E.
coli culture and filter. Rinse the funnel as previously indicated. Using this same membrane filter,
serially dilute the E. cloacae culture, filter and rinse. This will introduce both E. coli cells and E.
cloacae cells onto the same filter for a dual positive control. Incubate all three MF petri plates for
24 hours at 35°C ± 0.5°C. The uninoculated control should remain blank. The P. aeruginosa
control should have no colonies. The positive control should have both blue (E. coli) and red (E.
cloacae) colonies.

9.6	Colonies may be picked from membrane filters and inoculated into Lauryl Tryptose Broth (LTB),
Brilliant Green Lactose Bile (BGLB), EC+MUG, or other media for further QC testing if desired.

10.0	Calibration and Standardization

10.1	No calibration or standardization of m-ColiBlue24 Broth is required.

10.2	All equipment used in Section 6 should be calibrated and maintained according to manufacturer's
instructions.

11.0	Procedure

11.1	Test Procedure

11.1.1	Aseptically open an ampule containing m-ColiBlue24 Broth and pour the broth onto the
pad in a 50-mm MF petri plate.

11.1.2	Place a sterile filter onto a sterile filter holder. Using sterile graduated cylinders and
pipets, measure an appropriate sample aliquot. Pour water sample into reservoir funnel
on filter holder and draw the water through the filter using a vacuum pump. Rinse the
funnel with several 20-3 0-mL volumes of sterile rinse water. With sterile forceps,
transfer the filter to a petri plate containing the pad saturated with m-ColiBlue24. Invert
plate and incubate at 35°C ± 0.5°C for 24 hours.

11.2	Interpretation

11.2.1 Examine filters for colony growth. Colonies are typically readily visible, but a dissecting
microscope may prove useful.

Draft

5

June 2003


-------
11.2.2	A red or blue colony is a Total Coliform Positive result. A clear or white colony is a

Total Coliform Negative result. A blue colony is specifically an E. coli Positive result. A
non-blue colony is an E. coli Negative result:



Positive Result

Negative Result

Total Coliform

Red or Blue Colony

Clear or White Colony

E. coli

Blue Colony only

Non-Blue Colony

11.2.3 The incubation time is 24 hours. If no colonies are visible after 24 hours, the sample is
free of total coliforms and E. coli.

12.0	Data Analysis, Calculations, Interpretation and Reporting

12.1	Presence/Absence

12.1.1 The presence of at least one red or blue colony indicates the sample is total coliform
positive. The presence of at least one blue colony also indicates the sample contains E.
coli. Absence of red or blue colonies indicates the sample contains no total coliforms or
E. coli. No further analysis or calculation is required.

12.2	Quantitation

12.2.1	This method is approved by the USEPA for the purposes of determining the presence or
absence of total coliforms and E. coli in drinking water. Since a protocol for evaluating
the use of a new method for the enumeration of total coliforms and/or E. coli in drinking
water does not exist, new methods, such as m-ColiBlue24, cannot be approved by the
USEPA as quantitative methods for USEPA reporting purposes. However, we have
included the enumeration protocol for other (non-USEPA reporting) purposes.

12.2.2	Coliform density determinations can often provide additional data for problem solving
within the water plant or for analysis of source water samples. m-Coliblue24 Broth can
be used to make density determinations.

12.2.3	If enumeration of total coliform and/or E. coli populations is desired, refer to Standard
Methods 9222B for appropriate dilutions of the sample to filter so that 20-80 coliform
colonies are present after incubation. For analysis of potable water samples, a standard
sample of 100 mL is always required. Calculate the number of microorganisms per 100-
mL sample also as described in Standard Methods 9222B.

Draft

6

June 2003


-------
13.0	Method Performance Characteristics

13.1	Specificity: The specificity of m-ColiBlue24 Broth for recovery of total coliforms and E. coli
was evaluated. These experiments were conducted according to the EPA protocol of June 30,
1992 (16.3). For E. coli, the false positive error was 2.5% and the undetected target error was 0%
(Section 17.0, Table 17.1). Overall agreement between m-ColiBlue24 and li. coli reference
methods was 98.8%. For total coliforms, the false positive error was 26.8% and the undetected
target error was 1.6% (Table 17.2). Using m-Endo as a comparison for total coliform recovery to
m-ColiBlue24, m-Endo total coliform false positive error was 29.6% and the undetected target
error was 3.4 % (Table 17.3). Overall agreement for total coliform recovery was 86.2% for m-
ColiBlue24, and 85.7% with m-Endo. Although the total coliform false positive errors seemed
large initially, review of the literature revealed that many existing media for total coliform
recovery have similar false positive error rates (16.4- 16.10).

13.2	Precision and Bias: Not applicable, as this protocol was conducted on a presence/absence basis.
If individuals wish to use this medium for quantitative determinations they may use precision and
bias calculations specified by ASTM (16.11).

14.0	Pollution Prevention

14.1	Base the quantity of chemicals purchased on expected usage during its shelf life and disposal cost
of unused material. Actual reagent preparation volume should reflect anticipated usage and
reagent stability.

14.2	For information about pollution prevention that may be applicable to laboratories and research
institutions, consult Less is better: Laboratory Chemical Management for Waste Reduction,
available from the American Chemical Society's department of Government Regulations and
Science Policy, 1155 16th Street N.W., Washington D.C., 20036. Phone: (202) 872-4477.

15.0	Waste Management

15.1	It is the laboratory's responsibility to comply with all federal, state, and local regulations
governing waste management, particularly the hazardous waste identification rules and land
disposal restrictions, and to protect the air, water, and land by minimizing and controlling all
releases from fume hoods and bench operations. Compliance with all sewage discharge permits
and regulations is also required.

15.2	See the MSDS for product composition information and further guidance on waste disposal.

15.3	For more information on laboratory waste management, consult Waste Management Manual for
Laboratory Personnel, available from the American Chemical Society's department of
Government Regulations and Science Policy, 1155 16th Street N.W., Washington D.C., 20036.
Phone: (202) 872-4477.

Draft

1

June 2003


-------
16.0 References

16.1	Federal Register, Title 40, Part 141, Section 141.21 Coliform sampling. December 5, 1994.

16.2	"American Public Health Association, American Water Works Association, Water Environment
Federation. Microbiological Examination, Part 9000" in: Standard Methods for the Examination
of Water and Wastewater, 18th ed., A.E. Greenberg, L.S. Clesceri, A.D. Eaton, eds. Washington,
D.C., American Public Health Association, 1993.

16.3	U. S. Environmental Protection Agency. Requirements for the nationwide approval of new or
optionally revised methods for total coliforms, fecal coliforms, and/or E. coli, in national drinking
water monitoring. Revision 1.2, June 30, 1992. Environmental Monitoring Systems Laboratory,
Cincinnati, OH.

16.4	Brenner, K.P., C. C. Rankin, Y.R. Roybal, G.N. Stelma, P.V. Scarpino, and A.P. Dufour. 1993.
New medium for the simultaneous detection of total coliforms and Escherichia coli in water.
Appl. Environ. Microbiol. 59: 3534-3544.

16.5	Cenci, G., A. De Bartolomeo, and G. Caldini.1993. Comparison of fluorogenic and conventional
membrane filter media for enumerating coliform bacteria. Microbios 76:47-54.

16.6	Covert T.C., L.C. Shadix, E.W. Rice, J.R. Haines, and R.W. Freyberg. 1989. Evaluation of the
autoanalysis colilert test for detection and enumeration of total coliforms. Appl. Environ.
Microbiol. 55:2443-2447.

16.7	Edberg, S.C., M.J. Allen, D.B. Smith, and the national collaborative study. 1988. National field
evaluation of a defined substrate method for the simultaneous enumeration of total coliforms and
Escherichia coli from drinking water: comparison with the standard multiple tube fermentation
method. Appl. Environ. Microbiol. 54:1595-1601.

16.8	Jacobs, N.J., W. L. Zeigler, F.C. Reed, T.A. Stukel, and E.W. Rice. 1986. Comparison of
membrane filter, multiple-fermentation-tube, and presence-absence techniques for detecting total
coliforms in small community water systems. Appl. Environ. Microbiol. 51:1007-1012.

16.9	Lupo, L., E. Stickland, A. Dufour, and V. Cabelli. 11977. The effect of oxidase positive bacteria
on total coliform density estimates. Health Lab. Sci. 14: 117-121.

16.10	Sartory, D P. and L. Howard. 11992. A medium detecting b-glucuronidase for

the simultaneous membrane filtration enumeration of Escherichia coli and coliforms from
drinking water. Lett. Appl.Microbiol. 15:273-276.

16.11	American Society for Testing and Materials. Determination of precision and bias applicable
methods of committee D-19 on water. D 2777-86. Annual Book of ASTM Standards, Vol. 11.01.
ASTM, Philadelphia, 1993.

Draft

8

June 2003


-------
17.0	Tables

17.1	E. coli recovery on m-ColiBlue24

Reference



Positive

Negative

Total

Positive

234

6

240

Negative

0

250

250

Total

234

256

490

Sensitivity	100.0%

Specificity	97.7%

False Positive Error	2.5%

Undetected Target Error	0%

Overall Agreement	98.8%

17.2 Coliform recovery on m-ColiBlue24

Reference



Positive

Negative

Total

Positive

183

67

250

Negative

2

248

250

Total

185

315

500

Sensitivity

96.8%

Specificity

80.1%

False Positive Error

29.0%

Undetected Target Error

3.2%

Overall Agreement

85.7%

Reference



Positive

Negative

Total

Positive

149

61

210

Negative

5

245

250

Total

154

306

460

Sensitivity

96.8%

Specificity

80.1%

False Positive Error

29.0%

Undetected Target Error

3.2%

Overall Agreement

85.7%

Draft

9