THE ENVIRONMENTAL TECHNOLOGY VERIFICATION

s>EPA

ll ** » 111 It "Muni 1.11 I'r-«I

PROGRAM £

ETV

onvl

Oak Ridge National Laboratory

Verification Statement

TECHNOLOGY TYPE:

X-RAY FLUORESCENCE

APPLICATION:

MEASUREMENT OF LEAD IN DUST WIPES

TECHNOLOGY NAME:

XL-300 Series XRF Instrument

COMPANY:

NITON Corporation

ADDRESS:

900 Middlesex Turnpike, Bldg. 8 PHONE: (978) 670-7460
Billerica, MA 01821 FAX: (978) 670-7430

WEB SITE:
E-MAIL:

www.niton.com
jjshein@niton.com

The U.S. Environmental Protection Agency (EPA) has created the Environmental Technology
Verification Program (ETV) to facilitate the deployment of innovative or improved environmental
technologies through performance verification and dissemination of information. The goal of the ETV
Program is to further environmental protection by substantially accelerating the acceptance and use of
improved and cost-effective technologies. ETV seeks to achieve this goal by providing high-quality,
peer-reviewed data on technology performance to those involved in the design, distribution, financing,
permitting, purchase, and use of environmental technologies.

ETV works in partnership with recognized standards and testing organizations and stakeholder groups
consisting of regulators, buyers, and vendor organizations, with the full participation of individual
technology developers. The program evaluates the performance of innovative technologies by developing
test plans that are responsive to the needs of stakeholders, conducting field or laboratory tests (as
appropriate), collecting and analyzing data, and preparing peer-reviewed reports. All evaluations are
conducted in accordance with rigorous quality assurance protocols to ensure that data of known and
adequate quality are generated and that the results are defensible.

Oak Ridge National Laboratory (ORNL) is one of the verification organizations operating under the
Advanced Monitoring Technology (AMT) Center. AMT, which is administered by EPA's National
Exposure Research Laboratory (NERL), is one of six technology areas under ETV. In this verification
test, ORNL evaluated the performance of lead in dust wipe measurement technologies. This verification
statement provides a summary of the test results for NITON's XL-300 Series x-ray fluorescence (XRF)
instrument.

EPA-VS-SCM-51

The accompanying notice is an integral part of this verification statement.

August 2002


-------
VERIFICATION TEST DESCRIPTION

This verification test was designed to evaluate technologies that detect and measure lead in dust wipes.
The test was conducted at the Capitol Community Technical College in Hartford, CT, from November 5
through November 9, 2001. The vendors of commercially-available, field portable technologies blindly
analyzed 160 dust wipe samples containing known amounts of lead, ranging in concentration from <2 to
1,500 |ig/wipe. The experimental design was particularly focused on important clearance levels, such as
those identified in 40 CFR 745.65 of 40, 250, and 400 |ig/ft2. The samples included wipes newly-
prepared and archived from the Environmental Lead Proficiency Analytical Testing Program (ELPAT).
These samples were prepared from dust collected in households in North Carolina and Wisconsin. Also,
newly-prepared samples were acquired from the University of Cincinnati (UC). The UC dust wipe
samples were prepared from National Institute of Standards and Technology (NIST) Standard Reference
Materials (SRMs). The results of the lead analyses generated by the technology were compared with
results from analyses of similar samples by conventional laboratory methodology in a laboratory that
was recognized as proficient by the National Lead Laboratory Accreditation Program (NLLAP) for
clearance testing. Details of the test, including a data summary and discussion of results, may be found in
the report entitled Environmental Technology Verification Report: Lead in Dust Wipe Detection
Technology— NITON Corporation, XL-300 Series X-Ray Fluorescence Instrument, EPA/600/R-02/059.
NITON's XL-700 Series XRF was also evaluated in the test and a separate report has been prepared
(Environmental Technology Verification Report: Lead in Dust Wipe Detection Technology— NITON
Corporation, XL-700 Series X-Ray Fluorescence Instrument, EPA/600/R-02/055).

TECHNOLOGY DESCRIPTION

The XL-300 Series instrument, which is designed to quantify only lead, is an energy dispersive x-ray
fluorescence (EDXRF) spectrometer that uses a sealed, 10 mCi cadmium-109 radioisotope source to
excite characteristic x-rays of a test sample's constituent elements. These characteristic x-rays are
continuously detected, identified, and quantified by the spectrometer during sample analysis. Stated
simply, the energy of each x-ray detected identifies a particular element present in the sample, and the
rate at which x-rays of a given energy are counted provides a determination of the quantity of that
element that is present in the sample. Detection of the characteristic lead x-rays is achieved using a
highly efficient, thermo-electrically cooled, solid-state, silicon PIN-diode detector, a part of the Dual
Detector system. Signals from the Dual Detector are amplified, digitized, and then quantified via integral
multichannel analysis and data processing units. Sample test results are displayed in total micrograms of
lead per dust wipe. During this verification test, the instrument configuration was XL-309 and reporting
limits were approximately 15 |ig/wipe.

VERIFICATION OF PERFORMANCE

The following performance characteristics of the XL-300 Series XRF were observed:

Precision: Precision—based on the average percent relative standard deviation—was 8%.

Accuracy: Accuracy was assessed using the estimated concentrations of the ELPAT and UC samples.
The average percent recovery value for all samples reported above 15 |ig/wipe was 94%. The slight
negative bias was statistically significant, but well within the acceptable range of bias (100% ±25%).

Comparability: A comparison of the XL-300 results and the NLLAP-recognized laboratory
results was performed for all samples (ELPAT and UC) that were reported above 20 |ig/wipe. The
correlation coefficient (r) for the comparison of the entire dust wipe data set was 0.999 [slope (m) = 0.849,
intercept = 13.247], indicating good agreement with the NLLAP-laboratory data.

EPA-VS-SCM-51

The accompanying notice is an integral part of this verification statement.

August 2002


-------
Detectable blanks: All twenty samples, prepared at concentrations around 1 |ig/wipe, were reported
correctly as < 15 |ig/wipe by the XL-300.

False positive results: A false positive result is one in which the technology reports a result that is above
the clearance level when the true (or estimated) concentration is actually below. Neither the XL-300 nor
the NLLAP laboratory reported false positives, out of a possible 30 fp results.

False negative results: A false negative result is one in which the technology reports a result that is below
the clearance level when the true (or estimated) concentration is actually above. Both the XL-300 and the
NLLAP laboratory reported 21 of 30 possible fn results. The XL-300 had a lower percentage of fn results
at the 40 |ig/wipe level, a comparable amount at the 250 |ig/wipe level, and a higher percentage at the 400
|ig/wipe level than the NLLAP laboratory, but overall fn performance for the two methods was
comparable.

Completeness: The XL-300 Series instrument generated results for all 160 dust wipes samples, for a
completeness of 100%.

Sample Throughput: A single analyst achieved a sample throughput rate of 40 samples per day, with
eight measurements taken for each dust wipe, for a total of 320 measurements per day.

Overall Evaluation: The overall performance was characterized as having a slight negative bias (but
one within an acceptable range of bias), precise, and comparable to the NLLAP laboratory results. The
verification team found that the XL-300 XRF instrument was relatively simple for the trained analyst
to operate in the field, requiring less than an hour for initial setup. As with any technology selection,
the user must determine if this technology is appropriate for the application and the project data quality
objectives. Additionally, ORNL and ETV remind the reader that, while the ETV test provides valuable
information in the form of a snapshot of performance, state, tribal, or federal requirements regarding
the use of the technologies (such as NLLAP recognition where required) need to be followed. For more
information on this and other verified technologies, visit the ETV web site at http ://www .epa.gov/etv.

Gary J. Foley, Ph.D.

Director

National Exposure Research Laboratory
Office of Research and Development

W. Frank Harris, Ph.D.

Associate Laboratory Director
Biological and Environmental Sciences
Oak Ridge National Laboratory

NOTICE: EPA verifications are based on evaluations of technology performance under specific, predetermined criteria
and appropriate quality assurance procedures. EPA and ORNL make no expressed or implied warranties as to the
performance of the technology and do not certify that a technology will always operate as verified. The end user is solely
responsible for complying with any and all applicable federal, state, and local requirements. Mention of commercial
product names does not imply endorsement or recommendation.

EPA-VS-SCM-51

The accompanying notice is an integral part of this verification statement.

August 2002


-------