US Environmental Protection Agency
Office of Pesticide Programs

SOP for Maintenance and Calibration of
Thermometers and Thermometer/Hygrometers

Date Revised: 7-14-11


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 1 of 10

EPA/OPP MICROBIOLOGY LABORATORY
ESC, Ft. Meade, MD

Standard Operating Procedure
for

Maintenance and Calibration of Thermometers
and Thermometer/Hygrometers

SOP Number: EQ-02-06

Date Revised: 07-14-11

Initiated By:		Date: / /

Print Name:	

Technical Review: 	Date: / /

Print Name:

Technical Staff

QA Review:		Date: / /

Print Name:	

QA Officer

Approved By:		Date: / /

Print Name:	

Branch Chief

Effective Date:	/ /

Controlled Copy No.:

Withdrawn By:

Date: / /


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 2 of 10

TABLE OF CONTENTS

Contents
Page Number

1.0 SCOPE AND APPLICATION	 3

2.0 DEFINITIONS	3

3.0 HEALTH AND SAFETY	4

4.0 CAUTIONS	4

5.0 INTERFERENCES	4

6.0 PERSONNEL QUALIFICATIONS	 4

7.0 SPECIAL APPARATUS AND MATERIALS	 4

8.0 INSTRUMENT OR METHOD CALIBRATION	 5

9.0 SAMPLE HANDLING AND STORAGE	 5

10.0 PROCEDURE AND ANALYSIS	 5

11.0 DATA ANALYSIS/CALCULATIONS	 6

12.0 DATA MANAGEMENT/RECORDS MANAGEMENT	 7

13.0 QUALITY CONTROL	7

14.0 NONCONFORMANCE AND CORRECTIVE ACTION	 8

15.0 REFERENCES	8

16.0 FORMS AND DATA SHEETS	8


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 3 of 10

1.0 SCOPE AND APPLICATION:

1.1	This protocol describes the thermometers and thermometer/hygrometers in use in
the laboratory, calibration requirements for ISO 17025 accredited verification
thermometer and the procedure to point-check the accuracy of thermometers (i.e.,
liquid in glass and mercury in teflon-coated glass) used to measure the
temperature of autoclaves, water baths, incubators, refrigerators, and freezers.

1.2	See section 2.1 for calibration requirements for ISO 17025 accredited vendor-
calibrated liquid in glass thermometers, mercury thermometers, digital
thermometers, and thermometer/hygrometers.

1.3	The laboratory utilizes an Environmental Monitoring and Alarm System (EMAS;
see SOP QC-05, Monitoring Environmental Parameters) to collect environmental
data electronically for incubators, refrigerators, freezers, laboratories, and sample
storage rooms. In the event that EMAS is not working, the appropriate
thermometers will be point-checked against an ISO 17025 accredited verification
liquid in glass thermometer and employed to measure the temperature of
incubators, refrigerators, and freezers. ISO 17025 accredited vendor-calibrated
thermometer/hygrometers are used to monitor the temperature/humidity of
laboratories and sample storage areas.

2.0 DEFINITIONS:

2.1	ISO 17025 accredited verification thermometer - a thermometer (liquid in glass,
mercury in teflon-coated glass, digital, or digital thermometer/hygrometer) that
has been calibrated by a recognized ISO 17025 accredited vendor. The
calibration certificate must contain the stamp of the accrediting body (e.g., A2LA)
and the calibration vendor certificate number.

2.1.1	Representative liquid in glass and mercury in teflon-coated glass ISO
17025 accredited verification thermometers are used to point-check the
accuracy of other laboratory thermometers of a similar type.

2.1.2	All digital thermometers and thermometer/hygrometers used in the
laboratory are calibrated annually by an ISO 17025 accredited vendor.

2.2	Correction Factor = the difference in the vendor-certified ISO 17025 accredited
verification thermometer and the Observed Temperature measured by the
thermometer to be point-checked. This is calculated and reported during the
annual certification of the thermometers.


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 4 of 10

3.0 HEALTH AND SAFETY:

3.1 Teflon-coated mercury in glass thermometers are permissible in the laboratory.
Non-coated mercury thermometers are not permissible. The teflon coating
reduces the likelihood that mercury will be spilled in the event that the
thermometer is broken. If a teflon-coated mercury in glass thermometer (section
7.2) breaks (teflon remains intact), place the thermometer in the chemical waste
bin in the fume hood in B209 for disposal. If the integrity of the teflon coating is
questionable, call the SHEM manager (or contact security at extension 52800 if
SHEM manager cannot be reached) immediately for assistance.

4.0 CAUTIONS:

4.1	Analysts are not to attempt to clean up any spills of mercury.

4.2	Discard mercury thermometers appropriately.

5.0 INTERFERENCES:

5.1 Allow the thermometer to equilibrate with the solutions before taking the
temperature reading.

6.0 PERSONNEL QUALIFICATIONS:

6.1 Personnel are required to be knowledgeable of the procedures in this SOP.

Documentation of training and familiarization with this SOP can be found in the
training file for each employee.

7.0 SPECIAL APPARATUS AND MATERIALS:

To determine which thermometers are currently in use in the laboratory, refer to the

Point-Check Calibration of Laboratory Thermometers Record Book.

7.1 ISO 17025 accredited verification thermometer. The ISO 17025 accredited

verification thermometers are certified annually (see section 2.1) by an ISO 17025
accredited vendor. Representative liquid and mercury in teflon-coated glass ISO
17025 accredited verification thermometers are used to verify the accuracy of like
thermometers specified in sections 7.2 and 7.5.

7.2

Kessler maximum registration thermometers (ERTCO, mercury-
containing/teflon-coated) are used to verify autoclave temperatures. One
representative thermometer is certified annually by an ISO 17025 accredited


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 5 of 10

vendor (see section 2.1 for certification requirements) and is used to check the
accuracy of the remaining thermometers.

7.3	Digital thermometers for freezers. All digital thermometers are certified annually
by an ISO 17025 accredited vendor. See section 2.1 for certification requirements.

7.4	Thermometer/hygrometers are used to measure temperature and humidity of
laboratory rooms, incubators, and other spaces. All thermometer/hygrometers
thermometers are certified annually by an ISO 17025 accredited vendor. See
section 2.1 for certification requirements.

7.5	Thermometers (i.e., liquid in glass) used to measure the temperature of water
baths, incubators, refrigerators, and freezers, including FRIO-Temp Precision
Thermometers.

8.0 INSTRUMENT OR METHOD CALIBRATION:

8.1 Thermometers, liquid in glass and mercury in teflon-coated glass, are checked for
accuracy using an ISO 17025 accredited verification thermometer.

9.0 SAMPLE HANDLING AND STORAGE: Not applicable

10.0 PROCEDURE AND ANALYSIS:

10.1	Annual thermometer calibration reports (see section 2.1 for certificate
requirements) for thermometers certified by ISO 17025 accredited vendors are
stored in the Thermometer and Hygrometer Calibration Certificates notebook.

10.2	Once a year, all of the thermometers falling under sections 7.2 and 7.5 of this SOP
are checked at operating temperatures against a similar type thermometer certified
by an ISO 17025 accredited vendor.

10.3	To point-check the accuracy of FRIO-Temp Precision Thermometers, place a
flask of de-ionized water or ethanol next to the thermometer in the incubator,
refrigerator, or freezer. Place the liquid in glass ISO 17025 accredited verification
thermometer into the flask of de-ionized water or ethanol. Once the de-ionized
water or ethanol has reached the temperature of interest, conduct the accuracy
check. The ISO 17025 accredited verification thermometer must be suspended in
the liquid in the flask, not resting on the bottom of the flask.

10.4

A water bath is used, where applicable, to check the temperature of the
thermometers (excluding the FRIO-Temp Precision Thermometers) by


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 6 of 10

simultaneous immersion of the ISO 17025 accredited verification thermometer
and the thermometer to be calibrated into the water bath at the temperature of
interest. Otherwise, point-checks are taken by simultaneous immersion of the
thermometers in a flask of deionized water or ethanol and then placing the flask
containing both thermometers in the instrument being monitored until the
deionized water or ethanol reaches the temperature of interest.

10.5	To point-check the accuracy of the Kessler maximum registration thermometers,
place the thermometer to be checked and the ISO 17025 accredited verification
thermometer (mercury in teflon-coated glass; section 7.2) side-by-side in one of
the laboratory's autoclaves and run a sterilization cycle (e.g., gravity 15 minute
cycle).

10.6	Any difference in temperature readings between the ISO 17025 accredited
verification thermometer and the laboratory thermometer is recorded on the Point-
Check Calibration of Laboratory Thermometers Record Form (see 16.0).

10.7	A label displaying the date of calibration and correction factor (even if it is zero),
is placed around the top of the corresponding thermometer.

10.8	Digital thermometers and thermometer/hygrometers are indicated to be within
tolerance. No correction factor is indicated on the calibration certificates.

11.0 DATA ANALYSIS/CALCULATIONS:

11.1	The Check Point for both the ISO 17025 accredited verification thermometer and
the thermometer being calibrated is the reading on the ISO 17025 accredited
verification thermometer during the point-check process. The temperature reading
on the ISO 17025 verification thermometer must be within the range at which the
thermometer to be calibrated will be used or is being used (see SOP QC-05,
Monitoring Temperatures, for acceptable temperature ranges for equipment).

11.2	The Observed Temperature is the temperature reading of the thermometer during
the check point process.

11.3	The correction factor for the ISO 17025 accredited verification thermometer is
determined by the accredited company that performs the calibration verification
for this thermometer. The True Temperature measured by the ISO 17025
accredited verification thermometer = the Observed Temperature of the ISO
17025 accredited verification thermometer plus its correction factor.

11.4

The correction factor for the laboratory thermometer that is being calibrated


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 7 of 10

against the ISO 17025 accredited verification thermometer is calculated by
subtracting the Observed Temperature of the laboratory thermometer from the
True temperature of the ISO 17025 accredited verification thermometer.

11.5 The True Temperature measured by the laboratory thermometer = the Observed
Temperature plus the correction factor for that thermometer.

12.0 DATA MANAGEMENT/RECORDS MANAGEMENT:

12.1	Data will be recorded promptly, legibly and in indelible ink on the Point-Check
Calibration of Laboratory Thermometers Record Form. Completed forms are
archived in the Point-Check Calibration of Laboratory Thermometers Record
Book. Annual Thermometer Calibration Reports for ISO 17025 accredited
verification thermometers, including vendor-certified digital thermometers,
Kessler maximum registration thermometers, and thermometer/hygrometers, are
stored in the Thermometer and Hygrometer Calibration Certificates notebook.
The books are kept in a secured file cabinet in the file room D217. Only
authorized personnel have access to the secured files. Archived data are subject
to OPP's official retention schedule contained in SOP ADM-03, Records and
Archives.

12.2	An inventory of ISO 17025 accredited verification thermometers requiring vendor
calibration will be maintained electronically (refer to 16.2). After each addition to
or deletion from the inventory, a hard copy of the inventory will be filed in the
Thermometer and Hygrometer Calibration Certificates notebook.

13.0 QUALITY CONTROL:

13.1	The OPP Microbiology laboratory conforms to 40 CFR Part 160, Good
Laboratory Practices. Appropriate quality control measures are integrated into
each SOP.

13.2	ISO 17025 accredited verification thermometers are certified once a year by an
ISO 17025 accredited vendor.

13.3	Autoclave (section 7.2) and water bath thermometers (included in section 7.5) are
point-checked once a year in the laboratory. All other thermometers in section
7.5 are point-checked on an as-needed basis, depending upon the functionality of
the EMAS. Once performed, thermometers may be used for one year, at which
point the point-check must be repeated.

13.4

For quality control purposes, the required information is documented on the


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 8 of 10

appropriate form(s) (see 16.0).

14.0 NONCONFORMANCE AND CORRECTIVE ACTION:

14.1	When routinely recording temperatures for the laboratory, laboratory equipment,
media etc., the observed temperature reading of the thermometer read plus the
correction factor for that specific thermometer must be recorded.

14.2	On occasion, thermometers must be discarded (e.g., broken columns,
unreasonably large correction factor). Contact the facility Safety, Health, and
Environmental Management Program Manager for proper disposal procedures.

15.0 REFERENCES: None

16.0 FORMS AND DATA SHEETS:

16.1 Point-Check Calibration of Laboratory Thermometers Record Form

16.2 Sample Inventory of Thermometers Requiring Vendor Calibration


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 9 of 10

16.1 Point-Check Calibration of Laboratory Thermometers Record
OPP Microbiology Laboratory

Dale

Initials

Thermometer

Type

Serial #

Check-
point'!'

Observed
Tcmp.ff

Correction
Factor *

True

Temp. **





ISO 17025
Accredited
Verification
Thermometer























Comments





Date

Initials

Thermometer

Type

Serial #

Check-
point f

Observed
Tcmp.ff

Correction
Factor *

True

Temp. **





ISO 17025
Accredited
Verification
Thermometer























Comments





Date

Initials

Thermometer

Type

Serial #

Check-
point t

Observed
Tcmp.ff

Correction
Factor *

True

Temp. **





ISO 17025
Accredited
Verification
Thermometer























Comments





f Reading on the ISO 17025 accredited verification thermometer during the point-check process. The temperature reading on
the ISO 17025 accredited verification thermometer must be within the range at which the thermometer to be calibrated will be or
is being used (See SOP QC-05, Monitoring Temperatures, for acceptable temperature ranges for equipment).

Temperature reading of thermometer during the point-check process.

* Correction factor for ISO 17025 accredited verification Thermometer determined by the Certification company. Correction
factor for the thermometer being calibrated is calculated by subtracting its observed temperature from true temperature of ISO
17025 accredited verification Thermometer.

** True Temperature of ISO 17025 accredited verification Thermometer = Observed Temp.+ Correction factor: True
temperature of Thermometer being Calibrated = Observed Temp. + Correction Factor for that thermometer.


-------
SOP No. EQ-02-06
Date Revised 07-14-11
Page 10 of 10

16.2 Sample Inventory of Thermometers Requiring Vendor Calibration
OPP Microbiology Laboratory

Thermometer Type
(Calibrated by)

Manufacturer/Model #
(if applicable)

Serial Number

Liquid in Glass
(Innocal Solutions)

ERTCO

1294

Kessler Maximum
Registration Thermometer
(autoclaves)
(Innocal Solutions)

ERTCO

7283

Digital Thermometers
(Control Company)

Control Company,
model #4015

111320371

Control Company,
model # 06-664-23

240106521

Control Company,
model #4015

91313194

Thermometer/Hygrometers
(Control Company)

Control Company, model #
4184

102115627

Control Company, model #
11-661-12

111288367

Control Company, model #
61161-324

99372818

Control Company, model #
4184

102023201


-------