Lemieux et al., Submission to Applied and Environmental Microbiology

Title: Destruction of Spores on Building Decontamination Residue in a Commercial Autoclave
Running Title: Destruction of Spores on BDR in a Commercial Autoclave

P. Lemieux*

U.S. Environmental Protection Agency
National Homeland Security Research Center
109 TW Alexander Dr. E343-06
Research Triangle Park, NC 27711
Phone: 919-541-0962
Fax: 919-541-0496
Email: lemieux.paul@epa.gov

R. Sieber, A. Osborne
Eastern Research Group, Inc.

14555 Avion Parkway, Suite 200
Chantilly, VA 20151-1102

A. Woodard

NY State Department of Environmental Conservation
625 Broadway
Albany, NY 12233-7258

* - Corresponding Author

1


-------
Lemieux et al., Submission to Applied and Environmental Microbiology

ABSTRACT

The U.S. Environmental Protection Agency (EPA) conducted an experiment to evaluate the effectiveness
of a commercial autoclave at treating simulated building decontamination residue (BDR). The BDR was
intended to simulate porous materials removed from a building deliberately contaminated with biological
agents such as Bacillus Anthracis (anthrax) in a terrorist attack. The purpose of these tests was to assess
whether the standard operating procedure in a commercial autoclave provides sufficiently robust
conditions to adequately destroy bacteria spores bound on the BDR. This study investigated the effect of
several variables related to autoclaving BDR, including time, temperature, pressure, item type, moisture
content, packing density, packing orientation, autoclave bag integrity, and autoclave process sequence.
The test team created simulated BDR from wallboard, ceiling tiles, carpet, and upholstered furniture,
embedded with 106 population Geobacillus stearothermophilus biological indicator (BI) strips and
thermocouples to obtain time/temperature profile data associated with each BI strip.

Results indicate that a single standard autoclave cycle did not effectively decontaminate the BDR.
Autoclave cycles of 120 minutes at 31.5 psig/275 °F and 75 minutes at 45 psig/292 °F effectively
decontaminate the BDR material. Two standard autoclave cycles of 40 minutes and 31.5 psig/275 °F run
in sequence proved to be particularly effective, probably because the second cycle's evacuation step
pulled the condensed water out of the pores of the materials, allowing better steam penetration. Results
also indicate that packing density and material type of the BDR in the autoclave can significantly impact
on the effectiveness of the decontamination process.

INTRODUCTION

In the event of a terrorist attack on a building
where biological weapons (BW) such as
Bacillus Anthracis (anthrax) might be used,
much of the porous material in the building may
be shipped for disposal after decontamination
activities. These materials are collectively
termed "building decontamination residue"
(BDR). Although the BDR may be disinfected
or decontaminated prior to shipment, it may
need additional decontamination to ensure the
contaminating agent has been destroyed, or
because of heightened political sensitivities
(e.g., a stigma attached to the waste), may need
to be handled as if it were still contaminated.
There are no mandated action levels for residual
spores in such materials, and the emergency
response personnel or on-scene coordinators
typically work with relevant state regulators to
make determinations of what constitutes proper
BDR disposal. Much of this BDR might be
tightly packed and possibly wet. EPA has
initiated a research program to investigate issues
related to the proper disposal of BDR (Lemieux,
2005).

Autoclaves are commonly used to effectively
treat regulated medical waste by exposing the
waste to steam at elevated pressures and
temperatures for extended periods of time (e.g.,
31.5 psig, 275 °F, and 40 minutes) (Fleming et
al., 1995). However, it is unknown whether the
standard operating procedure in a commercial
autoclave will provide sufficient
time/temperature/pressure to adequately destroy
residual bacteria spores bound on BDR.

The primary objective of this study (Sieber and
Osborne, 2005) was to establish whether
standard operating conditions at a commercial
medical waste autoclave are sufficient to destroy
bacteria spores potentially found on BDR, and if
not, what modifications to the standard operating
procedure could be recommended to assure
complete spore destruction. The secondary
objective of this study was to investigate the
time/temperature dependence of Geobacillus
stearothermophilus spore destruction as a
function of autoclave operating conditions and
BDR composition. Geobacillus
stearothermophilus was chosen because it is
widely available and commonly recommended
(Technical Assistance Manual, 1998) for moist
heat sterilization validation.

2


-------
Lemieux et al, Submission to Applied and Environmental Microbiology

MATERIALS AND METHODS

Autoclave Description

EPA conducted these tests on March 4-6, 2005
at the Healthcare Environmental, Inc. facility
located in Oneonta, New York, approximately
90 miles from Albany. The facility can treat up
to 84 tons of medical waste per day using two
identical autoclaves 8 feet in diameter and 32
feet long, which accept large metal bins (80
inches by 54 inches by 69 inches) on rollers.
Each autoclave (Bontech Model 886) can
process six bins, with a total mass of
approximately 3,000-4,000 pounds per cycle.

The State and Territorial Association on
Alternate Treatment Technologies developed a
document (Technical Assistance Manual, 1998)
to establish a framework or guideline that
defines medical waste treatment technology
efficacy criteria and delineates the components
required to establish an effective state medical
waste treatment technology approval process.
The report recommends that all medical waste
treatment technologies achieve six logs or
greater microbial inactivation of mycobacteria
and four logs or greater reduction of spores.
Approximately 32 states incorporate the
STAATT criteria for the treatment of regulated
medical waste or biohazardous waste. Effective
December 2005, the STAATT criteria now also
apply to autoclave technologies. While BDR
may not be classified at regulated medical waste,
a commercial autoclave rather than a bench-
scale autoclave, was investigated because of the
quantities of BDR that may be generated in the
event of a biological contamination.

The nominal autoclave operating cycle time is
40 minutes plus cool-down time to prepare for
subsequent loads. At the start of each cycle, the
autoclave is sealed and air is evacuated for 3
minutes using a vacuum pump to approximately
-10 psig. Steam is then injected to reach and
maintain the desired operating pressure and
temperature, typically within approximately 5
minutes. The nominal operating conditions
during the cycles are 31.5 psig and 275 °F.

Steam is injected through three ports at the top
of the autoclave, located at the front, center, and
rear. The steam is injected over distributor plates

to cause turbulent, disbursed steam flow
throughout the autoclave. At the end of each
cycle, the steam is evacuated by again pulling
vacuum.

Testing Approach

Autoclave performance was judged based on
two parameters: real-time measurements from
thermocouples and viability of BI test strips
containing 106 spores of Geobacillus
stearothermophilus embedded within each load
of simulated BDR material tested. The testing
comprised a series of test runs at different
conditions in one of the facility autoclaves (Unit
Al).

For each test run, 24 thermocouples were
embedded in the BDR material to record the
time/temperature profile at different locations
within the load. Additional control
thermocouples not embedded in BDR recorded
the temperature inside and outside the autoclave
(for data completeness and as an additional
diagnostic for temperature measurement
instrumentation operation). The thermocouple
wires passed into the autoclave through a custom
flange plate with a Swagelok bulkhead fitting
packed with high temperature RTV-silicone
sealant (see Figure 1). Real-time temperature
measurements were monitored and recorded at
each sampling point using a GEC Instruments
Model S27TC temperature measurement system
and Type "T" thermocouples. Temperatures
were recorded to the hard disk at approximately
10-second intervals.

Figure 1. Bulkhead Flange for Temperature
Measurements

3


-------
Lemieux et glm Submission to Applied and Environmental Microbiology

A BI pouch was paired with a thermocouple at
each test location (see Figure 2). Each BI pouch
contained two Geobacillus stearothermophilus
(American Type Culture Collection #7953, Lot
#3167091, expiration January 2007, Dtai Value
of 1.5 minutes, D1.32.2 Value of 0.14 minutes)
indicator strips, labeled 'A' and 'B', encased in a
GS Medical Packaging self-seal pouch
(#222100). Each BI strip contained a 10
population of spores on Schleicher & Schuell
filter paper (#470) encased in a glassine peel-
open envelope. Raven Biological Laboratories,
Inc. (Raven) manufactured the BI strips and
assembled the BI pouches. After the test, the A
strips were analyzed for a growth/no-growth
indication using the United States Pharmacopeia
(USP) Viable Spore Count procedure (United
States Pharmacopeia, XXV). The strips were
removed from their respective pouches,
transferred into Tryptic Soy Broth with
bromocresol purple indicator, and incubated at
55-60°C for 7 days. If the A strips showed
viable spores, a population assay was performed
on the corresponding B strips using the USP
Biological Indicator (spore strip) Population
Determination (United States Pharmacopeia,
XXV). The population was determined after 24-
hour incubation at 55-60°C in Tryptic Soy Agar.
Three types of control BI test pouches were also
used in the test: BI test pouches fully exposed to
the autoclave conditions (but not embedded
within BDR), BI test pouches packaged and
handled similarly to other BDR but not
autoclaved, and duplicate BI test pouches (i.e.,
two pouches placed next to each other within the
BDR). Figure 3 illustrates placement of the
fully exposed controls.

Figure 3. Control Placement (carpet and ceiling
tile shown)

The following variables were identified as
having a potential impact on penetration of hot
steam into the BDR and therefore impact spore
destruction capability:

•	Item type (wallboard, ceiling tile, carpeting,
upholstered furniture);

•	Moisture content of autoclaved materials

(wet, dry);

•	Autoclave packing density (loose, dense);

•	Packing orientation (horizontal, vertical);

•	Opening autoclave bags prior to cycle;

•	Autoclave temperature/pressure (31.5
psig/275 °F, 45 psig/292 °F);

•	Time in autoclave (up to 2 hours); and

•	Impact of multiple sequential autoclave
cycles.

The test matrix presented in Table 1 was
designed to investigate the effects of each of
these variables.

Item Types

Unpainted wallboard (LaFarge regular grade Vi"
thick drywall) was cut into approximately 2-
feet-by-2-feet sections. Sample BDR bags were
formed by placing five 2-feet-by-2-feet sections
face to face in autoclave bags. Wallboard was

Figure 2. BI Pouch

4


-------
Lemieux et al., Submission to Applied and Environmental Microbiology

Table 1. Test Matrix

Run

Item Type

Packing Arrangement

Pressure/ Temperature

Time (min)

1

Mixed

Loose, Horizontal

31.5 psig/275 °F

120

2

Wallboard

Dense, Horizontal

45 psig/292 °F

120

3

Carpet

Loose, Dense, Large Roll

31.5 psig/275 °F

120

4

Mixed (incl. sofa)

Loose, Horizontal

45 psig/292 °F

75

5

Mixed

Loose, Vertical

31.5 psig/275 °F

40 (x 2

sequential

runs)

6

Mixed

Loose, Vertical, Open Bags

31.5 psig/275 °F

40 (x 2

sequential

runs)

Mixed: wallboard, ceiling ti

e, and carpet

tested both wet and dry. In this context, dry
refers to as-is condition at ambient humidity
with no additional moisture added. Wetted
samples were submerged in a tank of water for
30 seconds and placed on a drain rack for 5
minutes prior to being placed in the bag. Dry
test bags weighed approximately 34 pounds, and
wet bags weighed approximately 37 pounds.
Samples were double bagged in 1.8 mil
polypropylene autoclave bags (to represent
likely practices that would be found at an
emergency response), and the bags were
individually goose-necked and taped shut using
duct tape. A section of nylon rope was attached
to the gooseneck to allow test personnel to easily
and safely load and unload the bags from the
autoclave bins. Three types of wallboard bags
were created. Some test bags (called "1-
sample" bags) were assembled with one
thermocouple and one test strip pouch placed
together, between the second and third
wallboard section. Other test bags (called "3-
sample" bags) were assembled with three
thermocouples paired with three test strip
pouches placed between the first and second,
second and third, and fourth and fifth wallboard
sections. Additional bags were made without
thermocouples and Bis, to be used as fillers.

Ceiling tiles (Armstrong Contractor Series 5/8"
thick Ceiling Panels Model #942) were cut into
approximately 2-feet-by-2-feet sections.

Samples were prepared similarly to wallboard;
however, bags contained nine 2-feet-by-2-feet
sections placed face to face. Dry test bags
weighed approximately 23 pounds, and wet bags

weighed approximately 31 pounds. "1 -sample"
bags contained one thermocouple and one test
strip pouch placed together, between the fourth
and fifth ceiling tile section. "3-sample" bags
contained three thermocouples paired with three
test strip pouches placed between the second and
third, fourth and fifth, and seventh and eighth
ceiling tile sections. Additional bags were made
without thermocouples and BI test pouches, to
be used as fillers.

Carpet (Mannington Nepenthe II Blue
commercial grade carpeting with Nylon 6,6
fibers) was tested in two configurations, small
and large rolls. For small rolls, the carpet was
cut into strips 26 inches wide by 20 feet long,
representing how carpet would most likely be
removed from a building. Some samples were
soaked with a hose-end sprayer. After soaking,
samples were rolled and placed on end to allow
free-flowing water to drain. Small rolls were
bagged in a similar manner to wallboard and
ceiling tiles. Dry test bags weighed
approximately 26 pounds, and wet bags weighed
approximately 40 pounds. As a worst case,
larger sections of carpet 6 feet wide and 24 feet
long were also tested. Large rolls were only
prepared wet, and weighed approximately 200
pounds, the maximum size that could be
reasonably handled by two workers. Large rolls
were wrapped in polypropylene and all seams
sealed with duct tape. For the small carpet rolls,
1- and 3-sample bags were prepared. "1-
sample" bags contained one thermocouple and
one test strip pouch placed together, at the
approximate mid-point of the radius of the

5


-------
Lemieux et glm Submission to Applied and Environmental Microbiology

carpet roll. "3-samplc " bags contained three
thermocouples paired with three test strip
pouches placed two laps in from the top. at the
mid-point of the radius, and two laps from the
center of the carpet roll. Additional bags were
made without thermocouples and BI test
pouches, to be used as fillers. The autoclave
bins at either end of the group of bins placed in
for each run were filled with uninstrumented
BDR materials to provide thermal mass and
minimize the impact of any cold spots within the
autoclave.

To represent upholstered furniture, a dry. used
queen-sized sleeper sofa was autoclaved in Run
4. Four thermocouple and test strip pouches
were paired and embedded at the following
locations in the sofa: one sample each was
inserted in holes cut approximately 6 inches
deep in a back cushion and a seat cushion (the
holes were then covered with duct tape); one
sample was placed within the folded sleeper
mattress; and one was placed between the seat
cushions. Although surface contamination of
upholstered furniture is the most likely scenario,
the BI strips were embedded in the upholstered
furniture to simulate a worst-case scenario. The
sofa was wrapped in polypropylene with all
seams sealed with duct tape, and then placed in
the autoclave on a sheet of plywood.

Packing Density

Wallboard was tested at two different packing
densities. In low-density packing, six bags were
placed in a bin, forming a single layer at the base
of the bin; some surfaces of all bags were readily
exposed to autoclave temperatures. In high-
density packing, 23 bags were placed in each
bin, forming layers approximately three to four
levels deep. In this arrangement, some bags
were exposed directly to autoclave conditions
while others were buried within the load in the
bin. Ceiling tiles were tested only in a low-
density arrangement, as described above for
wallboard. Runs of densely packed ceiling tiles
were deleted from the test matrix because
densely packed BDR material could not be
brought up to autoclave temperatures within the
120-minute duration specified in the test plan.
Carpet was tested in three configurations. Small
rolls, approximately 1 foot in diameter and 26

inches long, were placed in bags. Six bags were
placed in a bin for low-density packing, and 25
bags were placed in a bm for high-density
packing. In addition, a large, intact roll of carpet
6 feet long and approximately 1.5 feet in
diameter was tested in one run. Figure 4
illustrates the dense and loose packing
arrangements.

Figure 4. Dense and Loose Packing
Arrangements (wallboard shown)

Packing Orientation

Material was tested both lying horizontally (see
Figure 4) in the autoclave bins and positioned
vertically, with all sides exposed. Material was
positioned vertically by tying two ropes to the
top of the bag and attaching the ends of the
ropes to opposite sides of the autoclave bin wall,
as shown in Figure 5. The bags were hung
vertically to simulate a rack system to position
bags so that all sides were exposed to steam. The
test team theorized that hanging the BDR
upright would keep it from compressing from its

6


-------
Lemieux et al, Submission to Applied and Environmental Microbiology

own weight, and allow steam condensate to
drain more easily as it formed. If these
hypotheses were correct, they would facilitate
steam penetration and more effective heating of
the material in the autoclave.

Figure 5. Photograph of Vertical BDR
Positioning (wallboard, ceiling tile shown)

Open Bags

All BDR was double-bagged in 1.8 mil
polypropylene autoclave bags. The bags were
individually goose-necked and sealed with duct
tape. This procedure was adopted based on
packaging information from the State
Department Sterling, Virginia mail facility
anthrax cleanup (Army Corps of Engineers,
2002). After autoclaving, some of the bags had
clearly ruptured due to temperature and pressure
changes. However, in many cases, bag surfaces
bubbled and became deformed in the autoclave,
but it was not clear if they had fully opened. To
test if the bags opening had an effect on
decontamination, two bags in Run 5 and all of
the bags in Run 6 were opened prior to
autoclaving, by slicing open two sides of each
bag with a utility knife.

Autoclave Conditions

The test plan initially established a minimum run
time of 40 minutes at elevated temperature (275
°F), which is the standard operation of the
Healthcare Environmental autoclave. Literature
data indicate that holding the material for 15
minutes at 250 °F is required to ensure moist
heat sterilization (Boca et al., 2002; Barkley and
Richardson, 1994; Gardner and Peel, 1986;

Block, 2000). Therefore, the test plan called for
extending the run time beyond 40 minutes to
achieve a 250 °F temperature target at all, or at
least most, embedded thermocouples. Even if
the 250 °F target had not been achieved, the test
plan established a maximum run time of 120
minutes to enable the autoclave to process
multiple test runs each day. Runs 1, 2, and 3
were terminated at 120 minutes, before all
thermocouples reached the target temperature.
Run 4 was stopped at 75 minutes because the
sofa temperature was rising above the
temperature of the autoclave, indicating a
potential exothermic reaction in the sofa.
Because such a reaction and possible associated
hazards were not well understood, the run was
terminated. The BI strips in the sofa all showed
no growth, indicating that the Run 4 conditions
were sufficient to decontaminate upholstered
furniture. In Runs 5 and 6, two 40-minute runs
were conducted in sequence.

Multiple Short Cycles

As steam in the autoclave was evacuated at the
end of Runs 2, 3, and 4, the test team observed
that, as the vacuum was drawn, most
thermocouple readings converged toward a
single temperature. It was not known if this
resulted from increased turbulence during the
post-vacuum cycle, condensed water being
drawn out of BDR under vacuum, or some
combination of these and other factors. To
further investigate this phenomenon, in Runs 5
and 6, two complete normal autoclave-operating
cycles were run in succession. Each cycle
consisted of a pre-vacuum, steam pressurization,
and post-vacuum phase. The cycles were
conducted in immediate succession and the
autoclave remained sealed throughout both
cycles.

RESULTS

Figures 6 through 12 present plots of the
time/temperature data recorded during each of
the six runs. Some of the figures (Figures 6 and
10) also include readings from the control
thermocouple inside the autoclave, the reference
thermocouple outside the autoclave, and the
autoclave set point pressure/temperature.

7


-------
Lemieux et al, Submission to Applied and Environmental Microbiology

Figure 6 shows the time/temperature data from
Run 1. Note that in Run 1, there was a
significant amount of noise on several of the
thermocouple channels, believed to result from
condensation accumulating in the thermocouple
connection fittings (see Figure 1). After Run 1,
the bundle of thermocouple wire outside the
flange was positioned so that gravity would
prevent condensate from collecting - subsequent
runs only showed a minimal amount of noise. It
should be noted that although the control
thermocouple rapidly approached the autoclave
operating temperature, many of the
thermocouples never reached the targeted 250 °F
temperature. The BI viability measurements on
Run 1 were consistent with the temperature
measurements (i.e., the BI strips that were at
locations where 250 °F was maintained for 15
minutes showed no growth).

Figure 6. Time/Temperature Data (Loose
Packing)

Run 2 (see Figure 7) consisted of subjecting
only densely packed wallboard to the highest
autoclave pressure/temperature conditions.
Again, even at this higher temperature, many of
the thermocouples never reached 250 °F.
Wallboard is mostly composed of
CaSO-^tTO), and loses moisture between 212
and 302 °F (Budavari, 1996). This dehydration
step could possibly contribute to the slow heatup
times for wallboard, although the bulk density or
packing density of the wallboard could also be a
factor. The control temperature dropped in Run
2, which was explained later by the fact that the
bag containing the control BI strip and
thermocouple came loose and fell into the bin,
reducing its exposure to the steam. This run was

not repeated due to time constraints, but the
autoclave facility process monitors exhibited no
change at that point in time, which convinced
the investigators that the problem was with that
one thermocouple. In addition, the temperature
signals converged at the end of the run. This
observation led to the hypothesis that a second
autoclave cycle in sequence might be effective.

Figure 7. Temperature and Wallboard Spore
Viability for Run 2

The data sets depicted in Figure 7 are color
coded to indicate if the Bis associated with each
thermocouple were viable at the conclusion of
Run 2. A viable spore designation was used if
growth was found in both the growth/no-growth
test and the assay analysis. Decontamination or a
no viable spore designation was used if no
growth was found in the test with an initial 106
spore population. In a limited number of cases,
the growth/no-growth test indicated a positive
result; however, the subsequent assay analysis
measured no quantifiable population (reported
result of <100 CFU). These data series are
labeled as indeterminate. Note that the sample
locations that were maintained at 250 °F for 15
minutes consistently showed no growth on their
corresponding BI strips, while most of the
sample locations that did not meet that
time/temperature still showed growth.

Figure 8 shows the effect of the second
autoclave cycle (Run 5), using average
temperature data for each individual bag.

During this run, bags of various materials were
placed upright to maximize exposure during the
autoclave cycle, and then a second cycle was
run, complete with evacuation and
repressurization. At the beginning of the second

8


-------
Lemieux et al, Submission to Applied and Environmental Microbiology

autoclave cycle, almost all of the temperatures
converged to the operating temperature of the
autoclave. It is believed that when the cold,
porous BDR material is exposed to the steam
during the first cycle, condensate forms in the
pores, limiting steam penetration and subsequent
heat transfer. With the pores of the material full
of water, heat can transfer to the interior of the
material mostly through conduction, which is
slow, and the steam cannot penetrate very well
into the material. At the initiation of the second
autoclave cycle, the evacuation step pulls the
condensate out of the pores, so that when steam
is reapplied, it effectively penetrates the
preheated material and reaches the operating
temperature of the autoclave. The only
thermocouples that did not achieve the necessary
temperatures were in the wet carpeting. It was
unclear whether the bag with the wet carpet
burst open during Run 5. This led to Run 6
being performed with all bags cut open prior to
loading. Figure 9 shows the spore viability for
Run 5. As before, the samples that did not
achieve the necessary time/temperature
exhibited residual spore viability.

Figure 8. Effect of Second Autoclave Cycle:
Impact on Achieving Target Temperature
(Averaged Measurements)

—

	 No Growth

	 Growth









"\



I Run5
Multiple BDR Types
Two Autoclave Cycles
31.5 psig/275 °F

Figure 9. Effect of Second Autoclave Cycle:
Impact on Spore Survivability

Figure 10 shows the time/temperature data for
Run 6, where the bags were cut open prior to
autoclave loading, and two sequential autoclave
cycles were performed. In this case, all
thermocouples reached the necessary
time/temperature to achieve spore destruction,
supported by the fact that none of the Run 6 BI
strips showed any growth. It was not always
obvious whether any given bag ruptured during
the cycle, so no definitive conclusions can be
made about the effect of changing bag materials
as a means to promote bag burst during the
autoclave cycle. However, these observations
do suggest that packing BDR using bags made
from a material that will melt or open during
autoclaving might ensure good steam
penetration.

	 Reference and Control TC

	 Ceiling Tile



-j





J \

Run 6

Multiple BDR Types
Vertical Orientation
Two Autoclave Cycles
31 psig/275 T

Figure 10. Effect of Second Autoclave Cycle:
Impact of Cut Bags

Figure 11 shows the effect of packing density
when processing wallboard. Clearly, high-
density packing reduces the effectiveness of the

9


-------
Lemieux et al, Submission to Applied and Environmental Microbiology

autoclaving. It appears that an autoclave facility
processing BDR should minimize packing
density so that steam can readily penetrate to
each bag in the load.

Figure 11. Effect of Packing Density:
Wallboard

Figure 12 shows the effect of initial moisture on
heating BDR (except for wet carpet, which was
not present in Run 1). The wet ceiling tile
heated significantly slower than the other BDR
item types probably because the micropore
structure of the ceiling tiles completely filled
with water. The other item types showed similar
heating profiles. This supports the hypothesis
that initial condensation of steam in the pores of
the ambient-temperature BDR limits heat
transfer.

Figure 12. Effect of Moisture Content

DISCUSSION

This paper presents an empirical study to
evaluate whether or not moist heat/steam can
successfully access all surfaces of porous
building materials and furnishings with

sufficient potency such that deeply absorbed
bacterial spores may be inactivated, and to
determine the operational parameters needed to
achieve this. While spore strips can present
"easily-handled" challenges to this process, it's
important to acknowledge that weaponized
spores, or even live non-weaponized spores, will
likely behave differently. Given the dangerous
nature of biological weapon agents, and the
severely restricted access and stringent safety
protocols necessary to handle live agents, these
tests had to be performed on a simulant such as
Geobacillus stearothermophilus. However,
Geobacillus stearothermophilus is commonly
used as a simulant for agents such as Bacillus
Anthracis, particularly for studies on
technologies utilizing thermal treatment methods
to kill the spores (Lemieux et al., 2005). It must
be remembered that, in all likelihood, any BDR
brought to a disposal facility would have been
previously decontaminated and would probably
contain very small numbers of viable spores, so
testing with Bis that have lxlO6 spores
represents a worst-case scenario.

Based on the results of these tests, heating the
BDR to 250 °F for 15 minutes at the sampling
locations resulted in no viable spores. The most
effective spore destruction was obtained from:

•	Loose packing arrangement;

•	Dry BDR material;

•	Higher autoclave operating
pressure/temperature;

•	Multiple autoclave cycles in sequence;

•	Bags cut open prior to loading.

The optimal practices for processing BDR in a
commercial autoclave are:

•	Place BDR so all surfaces are exposed to
autoclave conditions;

•	Maintain a loose packing arrangement for
the materials; and

•	Use plastic film bags that allow steam
penetration.

The material that was successfully
decontaminated included:

•	Wet wallboard;

10


-------
Lemieux et al., Submission to Applied and Environmental Microbiology

•	Dry wallboard;

•	Wet ceiling tiles;

•	Dry ceiling tiles;

•	Dry carpet; and

•	Dry upholstered furniture.

Wet carpeting was successfully decontaminated
only using cut bags and two sequential autoclave
cycles.

Below are conclusions regarding autoclave-
operating conditions:

•	120 min @ 31.5 psig/275 °F decontaminated
wallboard, ceiling tiles, and dry carpet when
loaded as recommended;

•	75 min @ 45 psig/292 °F was sufficient to
decontaminate dry upholstered furniture,
although there were not sufficient runs with
upholstered furniture to determine whether
less rigorous conditions would also achieve
spore destruction;

•	75 min @ 45 psig/292 °F decontaminated
wallboard and ceiling tiles when loaded as
recommended; and

•	Two standard autoclave cycles of 40 min @
31.5 psig/275 °F in sequence
decontaminated wallboard, ceiling tiles, and
dry carpet when loaded as recommended.

The second cycle time may be able to be
shortened and still enable destruction of the
spores in the BDR. A third cycle may be
necessary for wet carpet.

The most important recommendation based on
these tests is to run at least two sequential
autoclave cycles. The second cycle exhibited a
profound effect on the time/temperature profile
of the BDR materials being processed. The
steam evacuation step between cycles appears to
be the critical step for assuring effective
decontamination of porous materials in an
autoclave.

ACKNOWLEDGMENTS

The authors would like to acknowledge Scott
Sholar, Steve Strackbein, and Dave Dayton of
ERG, Richard Geisser and Russ Hilton of

Healthcare Environmental, Inc., and Russ
Nyberg of Raven Labs for their help in making
these tests successful. This paper has been
reviewed by the U.S. Environmental Protection
Agency, and approved for publication.

Approval does not signify that the contents
necessarily reflect the views and policies of the
Agency, nor does mention of trade names or
commercial products constitute endorsement or
recommendation for use.

REFERENCES

1.	Army Corps of Engineers. 2002. Work
Plan, Waste Removal and Facility Cleaning,
Department of State SA-32 Diplomatic Pouch
and Mail Facility Sterling, VA., Submitted by
U.S. Army Corps of Engineers, Rapid Response
Program and SE&I, Rev 3, November 12, 2002.

2.	Barkley, W. and J. Richardson. 1994.
Laboratory Safety, p. 715-734, in Gerhardt, P.,
Murray, R., Wood, W., and Krieg, N. (ed.)
Methods for General and Molecular
Bacteriology. ASM. Washington, DC.

3.	Block, S.S.. 2000. Physical Sterilization and
Disinfection, p. 695-729, in Disinfection,
Sterilization, and Preservation, Lippincott,
Williams, & Wilkins.

4.	Boca, B., Pretorius, E., Gochin, R.,
Chapoullie, R., and Z. Apostolides. 2002. An
Overview of the Validation Approach for Moist
Heat Sterilization, Part I. Pharmaceutical
Technology. September, 2002, p. 62-70.

5.	Budavari, S., (ed.) 1996. The Merck Index,
12th Ed., Merck and Co., Inc., White Station,
New Jersey, 1996.

6.	Fleming, D.O., Richardson, J.H., Tulis, J.

J., and D. Vesley. (ed.). 1995. Decontamination,
Sterilization, Disinfection, and Antisepsis, p.
219-237. Laboratory Safety: Principles and
Practices, 2nd ed. American Society for
Microbiology, Washington, D.C.

7.	Gardner J.F. and M.M. Peel, (ed.). 1986.
Introduction to sterilization and disinfection,
New York, Churchill Livingstone.

8.	Lemieux, P. 2005. U.S. EPA R&D Program
for Disposal of Building Decontamination
Residue, Proceedings of the Department of

11


-------
Lemieux et al., Submission to Applied and Environmental Microbiology

Homeland Security R&D Partnership
Conference, Boston, MA.

9.	Lemieux, P., Wood, J.P., Lee, C.W.,

Serre, S.D., Denison, M., Bockelie, M.,

Sarofim, A., and J. Wendt. 2005. Thermal
Destruction of CB Contaminants Bound on
Building Materials: Experiments And Modeling.

Proceedings of the 2005 Scientific Conference
on Chemical and Biological Defense Research,

Timonium, MD.

10.	Sieber, R. and A. Osborne. 2005.

Destruction of Spores on Building
Decontamination Residue in a Commercial
Autoclave. EPA/600/R-05/081.

11.	Technical Assistance Manual, 1998.

Alternative Medical Waste Technology Efficacy
Assessment Criteria, p. 2-2, in State Regulatory
Oversight of Medical Waste Treatment
Technologies. State and Territorial Association
on Alternate Treatment Technologies, Electric
Power Research Institute, Inc., Palo Alto, CA.

12.	United States Pharmacopeia XXV,

Validation of Microbial Recovery from
Pharmacopeial Articles (U.S. Pharmaceutical
Convention), p. 2259-2261.

12


-------