Batteiie

The Business of Innovation

Environmental Technology
Verification Program

Advanced Monitoring
Systems Center

Test/QA Plan for Verification of
Ambient Hydrogen Sulfide Analyzers
at a Swine Finishing Farm

ET^r^EW


-------
TEST/QA PLAN

for

Verification of
Ambient Hydrogen Sulfide Analyzers
at a Swine Finishing Farm

April 12, 2005

Prepared by

Battelle
505 King Avenue
Columbus, OH 43201-2693


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 2 of 46
Version 1.0
April 12, 2005

A2 TABLE OF CONTENTS

Section	Page

A	PROJECT MANAGEMENT

A1	Title Page	 1

A2	Table of Contents	 2

A3	Distribution List	 5

A4	Verification Test Organization 	 6

A5	Background 	 13

A6	Verification Test Description and Schedule 	 14

A7	Quality Objectives	 20

A8	Special Training/Certification	20

A9	Documentation and Records 	21

B	MEASUREMENT AND DATA ACQUISITION

B1	Experimental Design	22

B2	Reference Sample Collection	33

B3	Sample Handling and Custody Requirements	34

B4	Laboratory Reference Methods	34

B5	Quality Control 	36

B6	Instrument/Equipment Testing, Inspection, and Maintenance 	37

B7	Instrument Calibration and Frequency 	37

B8	Inspection/Acceptance of Supplies and Consumables	38

B9	Non-Direct Measurements	38

B10	Data Management 	38

C	ASSESSMENT AND OVERSIGHT

CI	Assessments and Response Actions 	41

C2	Reports to Management	44

D	DATA VALIDATION AND USABILITY

D1	Data Review, Validation, and Verification Requirements 	45

D2	Validation and Verification Methods 	45

D3	Reconciliation with User Requirements 	45

E	REFERENCES 	46


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 3 of 46
Version 1.0
April 12, 2005

List of Figures	Page

Figure 1 Organization Chart	 7

Figure 2 Test Site	19

List of Tables

Table 1	Planned Verification Schedule	 16

Table 2	Planned Weekly Test Activities During the Field Period 	18

Table 3	Verification Test Performance Parameters 	22

Table 4	Approximate H2S Concentrations and Order for Multi-point Challenges	25

Table 5	Interferants and Approximate Concentrations for Interference Checks 	26

Table 6	Reference Method Quality Control Requirements and Target Acceptance

Criteria	37

Table 7	Summary of Data Recording Process 	39


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 4 of 46
Version 1.0
April 12, 2005

ETV Advanced Monitoring Systems Center

Test/QA Plan for Verification of
Ambient Hydrogen Sulfide Analyzers
at a Swine Finishing Farm

Version 1.0

April 12, 2005

VENDOR ACCEPTANCE:

Name

Company
Date


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 5 of 46
Version 1.0
April 12, 2005

A3 DISTRIBUTION LIST

Hydrogen Sulfide Analyzer Vendor
Mr. Bill Taylor
Teledyne Instruments API
3318 Hwy. 5, PMB 526
Douglasville, GA 30135

Hydrogen Sulfide Analyzer Vendor
Mr. Dick Bates
Horiba Instruments, Inc.
17671 Armstrong Avenue
Irvine, CA 92614

Ms. Elizabeth A. Betz

U.S. Environmental Protection Agency -

HEASD

National Exposure Research Laboratory
E205-01 USEPA Mailroom
Research Triangle Park, NC 27711

Mr. Robert Fuerst

U.S. Environmental Protection Agency -
HEASD

National Exposure Research Laboratory
D205-05 USEPA Mailroom
Research Triangle Park, NC 27711

Dr. Gary Norris

U.S. Environmental Protection Agency -
HEASD

National Exposure Research Laboratory
D205-03 USEPA Mailroom
Research Triangle Park, NC 27711

Dr. D. Bruce Harris

U.S. Environmental Protection Agency -
NRMRL

E343-02 EPA Mailroom
109 TW Alexander Drive
Research Triangle Park, NC 27711

Dr. William M. Ollison
American Petroleum Institute
1220 L St., NW
Washington, DC 20005

Verification Test Collaborators:

Dr. Jerry Hatfield

Mr. Richard Pfeiffer

U.S. Department of Agriculture

National Soil Tilth Laboratory

2150 Pammel Drive

Ames, IA 50011-4420

Dr. Eric Winegar
Applied Measurement Science
4764 Concord Drive
Fair Oaks, CA 95628

Ms. Karen Riggs

Dr. Ann Louise Sumner

Ms. Amy Dindal

Mr. Zachary Willenberg

Battelle

505 King Ave.

Columbus, OH 43201

Peer Reviewers:

Dr. Raul Dominguez, Jr.

South Coast Air Quality Management District

21865 Copley Drive

Diamond Bar, CA 91765


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 6 of 46
Version 1.0
April 12, 2005

SECTION A
PROJECT MANAGEMENT

A4 VERIFICATION TEST ORGANIZATION

The verification test will be conducted under the auspices of the U.S. Environmental
Protection Agency (EPA) through the Environmental Technology Verification (ETV) Program.
It will be performed by Battelle, which is managing the ETV Advanced Monitoring Systems
(AMS) Center through a cooperative agreement with EPA. The scope of the AMS Center covers
verification of monitoring technologies for contaminants and natural species in air, water, and
soil.

The day to day operations of this verification test will be coordinated and supervised by
Battelle personnel, with the participation of the vendors who will be having the performance of
their hydrogen sulfide (H2S) analyzers verified. The testing will be conducted at a large swine
finishing farm near Ames, Iowa in collaboration with the U.S. Department of Agriculture
National Soil Tilth Laboratory (USDA-NSTL) and Applied Measurement Science, which is a
private consulting company located in Fair Oaks, California. Staff from the USD A will support
this test by helping to install the analyzers to be tested, providing infrastructure at the test site,
overseeing operation of the analyzers during periods of routine operation, and performing
reference method sample collection and analysis. Applied Measurement Science, who will be
operating with funding from the American Petroleum Institute, will provide reference method
sampling and analysis. ASTM method D5504-011 will be used as the reference method with the
following substitution: pulsed flame photometric detection (PFPD) will be used instead of sulfur
chemiluminescence detection (SCD). Reference H2S measurements in ambient air will be carried
out by both USDA-NSTL and Applied Measurement Science using gas chromatography (GC)
with PFPD, but will differ in the sample collection method. Each analyzer vendor will install
their respective analyzer, operate the analyzer through portions of the test (unless they give
written consent for Battelle staff to operate it), and repair or maintain their analyzer during the
test. Quality Assurance (QA) oversight will be provided by the Battelle Quality Manager and


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 7 of 46
Version 1.0
April 12, 2005

also by the EPA AMS Center Quality Manager, at her discretion. The organization chart in
Figure 1 identifies the responsibilities of the organizations and individuals associated with the
verification test. Roles and responsibilities are defined further below.

Figure 1. Organization Chart


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 8 of 46
Version 1.0
April 12, 2005

A4.1 Battelle

Dr. Ann Louise Sumner is the AMS Center's Verification Test Coordinator for this test.
In this role, Dr. Sumner will have overall responsibility for ensuring that the technical, schedule,
and cost goals established for the verification test are met. Specifically, Dr. Sumner will:

•	Prepare the draft test/QA plan, verification reports, and verification statements.

•	Revise the draft test/QA plan, verification reports, and verification statements in
response to reviewers' comments.

•	Assemble a team of qualified technical staff to conduct the verification test.

•	Establish a budget for the verification test and manage staff to ensure the budget is not
exceeded.

•	Direct the team (Battelle, USD A, and Applied Measurement Science staff) in
performing the verification test in accordance with this test/QA plan.

•	Ensure that all quality procedures specified in the test/QA plan and in the AMS Center
Quality Management Plan2 (QMP) are followed.

Serve as the primary point of contact for vendor representatives.

•	Ensure that confidentiality of sensitive vendor information is maintained.

•	Assist vendors and USD A staff as needed during the analyzer installation and
verification testing.

•	Become familiar with the operation and maintenance of the H2S analyzers through
instruction by the vendors, if needed.

•	Perform testing activities and data acquisition as specified in this test/QA plan.

•	Respond to any issues raised in assessment reports and audits, including instituting
corrective action as necessary.

Coordinate distribution of the final test/QA plan, verification reports, and verification
statements.

Ms. Amy Dindal is a Verification Testing Leader for the AMS Center. As such, Ms. Dindal
will provide technical guidance and oversee the various stages of verification testing. She will:
Support Dr. Sumner in preparing the test/QA plan and organizing the testing.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 9 of 46
Version 1.0
April 12, 2005

Review the draft and final test/QA plan.

Review the draft and final verification reports and verification statements.

Ms. Karen Riggs is Battelle's manager for the AMS Center. Ms. Riggs will

•	Review the draft and final test/QA plan.

•	Review the draft and final verification reports and verification statements.

•	Ensure that necessary Battelle resources, including staff and facilities, are committed
to the verification test.

•	Ensure that confidentiality of sensitive vendor information is maintained.

•	Support Dr. Sumner in responding to any issues raised in assessment reports and
audits.

•	Maintain communication with EPA's technical and quality managers.

•	Issue a stop work order if Battelle or EPA QA staff discovers adverse findings that
will compromise test results.

Battelle Technical Staff will support Dr. Sumner in planning and conducting the
verification test. The responsibilities of the technical staff will be to:

•	Assist in planning for the test, and making arrangements for the installation of the
analyzers.

•	Perform statistical calculations specified in this test/QA plan on the H2S analyzer data
as needed.

•	Provide results of statistical calculations and associated discussion for the verification
reports as needed.

•	Support Dr. Sumner in responding to any issues raised in assessment reports and
audits related to statistics and data reduction as needed.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 10 of 46
Version 1.0
April 12, 2005

Mr. Zacharv Willenberg is Battelle's Quality Manager for the AMS Center. Mr.

Willenberg will:

•	Review the draft and final test/QA plan.

•	Conduct a technical systems audit once during the verification test, or designate other
QA staff to conduct the audit.

•	Audit at least 10% of the verification data.

•	Prepare and distribute an assessment report for each audit.

•	Verify implementation of any necessary corrective action.

•	Request that Battelle's AMS Center Manager issue a stop work order if audits
indicate that data quality is being compromised.

•	Provide a summary of the QA/QC activities and results for the verification reports.

•	Review the draft and final verification reports and verification statements.

•	Assume overall responsibility for ensuring that the test/QA plan is followed.

A4.2 Hydrogen Sulfide Analyzer Vendors

The responsibilities of the H2S analyzer vendors are as follows:

•	Review and provide comments on the draft test/QA plan.

•	Accept (by signature of a company representative) the final test/QA plan prior to test
initiation (see page 4).

•	Provide an H2S analyzer for evaluation during the verification test.

•	Provide all other equipment/supplies/reagents/consumables needed to operate their
analyzer for the duration of the verification test.

•	Supply a representative to install and maintain their technology, and to operate it in
portions of the test specified in this test/QA plan, or provide written consent and
instructions for Battelle and USD A staff to carry out these activities.

•	Provide written instructions for routine operation of their analyzers, including a daily
checklist of diagnostic and/or maintenance activities.

•	Provide maintenance and repair support for their analyzers, on-site if necessary,
throughout the duration of the verification test.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 11 of 46
Version 1.0
April 12, 2005

•	Review and provide comments on the draft verification report and statement for their
respective analyzers.

A4.3 EPA

EPA's responsibilities in the AMS Center are based on the requirements stated in the
"Environmental Technology Verification Program Quality Management Plan" (EPA QMP).3 The
roles of specific EPA staff are as follows:

Ms. Elizabeth Betz is EPA's AMS Center Quality Manager. For the verification test, Ms.
Betz will:

•	Review the draft test/QA plan.

•	Perform at her option one external technical systems audit during the verification test.

•	Notify the EPA AMS Center Manager of the need for a stop work order if the
external audit indicates that data quality is being compromised.

•	Prepare and distribute an assessment report summarizing results of the external audit.

•	Review draft verification reports and verification statements.

Mr. Robert Fuerst is EPA's manager for the AMS Center. Mr. Fuerst will:

•	Review the draft test/QA plan.

•	Approve the final test/QA plan.

•	Review the draft verification reports and verification statements.

•	Oversee the EPA review process for the test/QA plan, verification reports, and
verification statements.

•	Coordinate the submission of verification reports and verification statements for final
EPA approval.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 12 of 46
Version 1.0
April 12, 2005

A4.4 U.S. Department of Agriculture National Soil Tilth Laboratory

This test will be conducted in collaboration with the USDA-NSTL, who will provide in-
kind support for this test. The responsibilities of personnel from the USDA-NSTL include the
following:

•	Coordinate use of the test site for the purposes of ETV testing, assuring access to the
test site for Battelle, USDA-NSTL, Applied Measurement Science, EPA, and vendor
representatives.

•	Support the test by providing space and needed utilities (e.g., instrument trailer,
electricity, compressed and cryogenic gases) for the H2S analyzers and reference
methods during testing.

•	Assist Battelle and vendor staff in the installation, operation, testing, and removal of
the H2S analyzers at the test site.

•	Plan and perform reference H2S sampling and analysis.

•	Perform testing activities and data acquisition on the H2S analyzers specified in this
test/QA plan.

•	Collect basic meteorological data (e.g., wind speed, wind direction, temperature, and
relative humidity) at the test site during the field period

•	Calculate the H2S reference results in terms of ambient H2S concentrations and provide
a data package to Battelle that includes all sampling data sheets, analysis records,
calibration data, and QA information, and that presents the H2S sample analysis results,
QA results, and calculated ambient H2S concentrations as well as the recorded
meteorological data.

•	Provide daily oversight of the H2S analyzers during periods of routine operation,
checking diagnostic indicators according to vendor directions and contacting Battelle if
faults in analyzer operation are observed.

•	Record observations about the maintenance and operation of the H2S analyzers during
the field period.

•	Review the draft verification reports, and verification statements.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 13 of 46
Version 1.0
April 12, 2005

A4.5 Applied Measurement Science

Dr. Eric Winegar of Applied Measurement Science will be providing in-kind support on
this test, with support from the American Petroleum Institute. The responsibilities of Dr.

Winegar include the following:

•	Perform on-site reference H2S measurements and conduct QA efforts as specified in
this test/QA plan.

•	Calculate the H2S reference results in terms of ambient H2S concentrations and provide
a data package to Battelle that includes all sampling data sheets, analysis records,
calibration data, and QA information, and that presents the H2S sample analysis results,
QA results, and calculated ambient H2S concentrations.

•	Review that portion of the verification reports that describes the in-situ reference
method and analysis.

A5 BACKGROUND

The ETV Program's AMS Center conducts third-party performance testing of
commercially available technologies that detect or monitor natural species or contaminants in
air, water, and soil. Stakeholder committees of buyers and users of such technologies
recommend technology categories, and technologies within those categories, as priorities for
testing. Hydrogen sulfide analyzers were identified as a priority technology category through the
AMS Center stakeholder process.

Hydrogen sulfide is formed at animal feeding operations (AFOs) during the bacterial
decomposition of sulfur-containing organic compounds present in manure produced by livestock.
Also known as sewer gas, H2S has the characteristic odor of rotten eggs and, at high levels, can
cause death from even brief exposure. Ambient H2S concentrations at swine farms, for example,
are expected to range from sub-part per billion (ppb) concentrations to 100 ppb or more.4
Ammonia and volatile organic compounds (VOCs) are also produced from bacterial processing
of livestock waste and are likely to be present in significant concentrations at AFOs.5'6


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 14 of 46
Version 1.0
April 12, 2005

The National Academy of Sciences 2003 report, "Air Emissions from Animal Feeding
Operations,"6 identified the need for need for improved methods for measuring and estimating
air emissions from animal feeding operations (AFO), including emissions of H2S. The analytical
approach of the H2S analyzers that will be evaluated in this verification test has been identified
for use in the National Air Emissions Monitoring Study Protocol that will be used to conduct
measurements of emissions from AFOs as directed by the U.S. EPA Animal Feeding Operations
Consent Agreement.7 The data collected as a result of the monitoring study will be used to
ensure compliance of AFOs with applicable provisions of the Clean Air Act, Comprehensive
Environmental Response, Compensation and Liability Act (CERCLA), and Environmental
Planning and Community Right-to-Know Act (EPCRA) and to promote a national consensus on
methods for estimating AFO emissions. In addition to this federal effort, several states, including
Iowa, California, and North Carolina, have developed or are developing standards for ambient
H2S.

A6 VERIFICATION TEST DESCRIPTION AND SCHEDULE
A6.1 Summary of Technology Category

The analyzers to be tested in this verification test are single point analyzers designed for
quantifying gas phase H2S in ambient air. The analyzers will be tested in the range of H2S
mixing ratios expected at a large AFO, in this case a swine finishing farm, and must be capable
of H2S quantification at low (5-20) ppb levels. A number of analytical techniques have been
employed by vendors of H2S analyzers, and include electrochemical detection, chromatography
with sulfur chemiluminescence, oxidation of H2S followed by ultra-violet (UV) sulfur dioxide
(S02) fluorescence, and others.

The analyzers that will be evaluated in this verification test are stand-alone, automated
instruments that continuously determine H2S concentrations in ambient air by oxidizing H2S in
the air sample to S02, which is detected using UV fluorescence. The analyzers draw ambient air
into the unit at flow rates less than one liter per minute, passing the air through a scrubber to
remove S02 from the air. The H2S remaining in the sample is oxidized to S02 in a catalytic
converter maintained at approximately 300ฐC. The S02 produced from the conversion is then
detected by UV fluorescence. Both analyzers can be configured to measure both H2S and S02,


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 15 of 46
Version 1.0
April 12, 2005

alternating between the two at user-specified intervals. Optional features include on-board H2S
or S02 permeation tubes and/or valves for zero air and span gases, which allow for automated
zero and span checks. Results from multi-point H2S calibrations are stored internally and used to
determine the H2S concentration in ambient air in units of ppb or parts per million (ppm),
depending on the selected analyzer range. Although data are collected continuously, averages at
specified intervals (e.g., 3 minutes) are calculated internally and relayed to analog and other data
outputs (e.g., RS-232 or ethernet).

In performing the verification test, Battelle will follow the technical and QA procedures
specified in this test/QA plan and will comply with the data quality requirements in the AMS
Center QMP.2

A6.2 Verification Schedule

Table 1 shows the planned schedule of activities in field testing and data
analysis/reporting in this verification test. As shown in Table 1, the field test of H2S analyzers is
planned to begin in April 2005 with installation of the analyzers at the swine finishing farm, and
to extend into May 2005. The period of operation of the analyzers at the facility will be
approximately 5 weeks, during which time the analyzers will monitor and record H2S in the
ambient air at the test site. Periodically over the duration of the field period, the analyzers will be
challenged with H2S and other compressed gas standards. Measurements will also be conducted
using two H2S reference methods during the field period. Subsequent to the field testing, a
separate verification report will be drafted for each analyzer, reviewed, revised, and submitted to
EPA for final signature.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 16 of 46
Version 1.0
April 12, 2005

Table 1. Planned Verification Schedule

Month

Test Activity

(2005)

Analyzer Field Activities

Data Analysis and Reporting

April

Set up/install H2S analyzers
Analyzer training by vendors
H2S standard gas challenges
H2S reference sampling
Routine operation

Begin preparation of report template
Analyze H2S reference samples
Compile data from H2S gas challenges

May

H2S standard gas challenges
Interferant challenges
H2S reference sampling
Routine operation
Remove analyzers from test site

Review and summarize operator observations
Compile data from H2S gas challenges
Compile data from interferant challenges
Analyze H2S reference samples
Compile data packages for reference methods

June



Complete summary of operator observations
Finalize data from all gas challenges
Finalize results for reference methods
Complete common sections of reports
Complete report sections on H2S and other
gas challenges and operator observations

July



Complete report sections on reference method
comparisons

Internal review of draft reports
Vendor review of draft reports

August



Revision of draft reports
Peer review of draft reports

September



Revision of draft reports

Submission of final reports for EPA approval

Table 2 shows the activities to be conducted in each week of the test during the field
period in April and May, 2005. The test procedures are described in Section B of this test/QA
plan. Repeated challenges with zero air and H2S gas standards will be conducted early in the
field period to obtain the analyzer "baseline" responses and the analyzer response time. The
same challenges with zero air and H2S gas standards will be conducted twice each week to
address analyzer span and zero drift. Multi-point challenges with H2S gas standards will be
conducted once early in the field period and again late in the field period to address analyzer
measurement accuracy, bias, precision, and linearity. Once during the field period, the analyzers


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 17 of 46
Version 1.0
April 12, 2005

will be challenged with gas standards of sulfur-containing species to evaluate analyzer
interference effects. Reference method sampling will be conducted throughout the field period
for comparison to the analyzer response to ambient air. In substantial portions of the field period,
the analyzers will routinely monitor ambient H2S to allow assessment of operational factors and
data completeness under continuous operation.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 18 of 46
Version 1.0
April 12, 2005

Table 2. Planned Weekly Test Activities During the Field Period

Week of
(Planned Month
and Day, 2005)

Test Activities

April 11

•	Install H2S analyzers

•	Establish inlet connections

•	Training of USDA and Battelle staff by vendor representatives

•	Conduct trial operations

April 18

•	Zero air/H2S standard challenge for analyzer response (baseline) and
analyzer response time

•	H2S standard challenges for linearity, accuracy, bias, precision

•	One zero/span check

•	Two USDA reference samples collected and analyzed

•	Routine operation

April 25

•	Two zero/span checks

•	Two USDA reference samples collected and analyzed

•	Install Applied Measurement Science reference method at test site

•	Begin Applied Measurement Science reference method measurements

•	Routine operation

May 2

•	Continue/complete Applied Measurement Science reference method
measurements

•	Two zero/span checks

•	Two USDA reference samples collected and analyzed

•	Routine operation

May 9

•	Demobilize Applied Measurement Science reference method analyzer

•	Two zero/span checks

•	Two USDA reference samples collected and analyzed

•	Gas standard challenges for interference check

•	Routine operation

May 16

•	Two zero/span checks

•	Two USDA reference samples collected and analyzed

•	H2S standard challenges for linearity, accuracy, bias, precision

•	Routine operation

May 23

• Remove analyzers from test site

A6.3 Test Site

This verification test will take place at a large swine finishing farm near Ames, Iowa. The
layout of the farm is shown in Figure 2. The farm has ten animal barns arranged in two parallel
rows of five, with each barn housing up to 2,000 swine. The overall test schedule shown in Table


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 19 of 46
Version 1.0
April 12, 2005

1 (Section A6.2) begins approximately six weeks after the farm will be populated by feeder pigs.
The urine and feces from the swine exit the barns through metal gratings in the floor and are
deposited in two nutrient lagoons located on the southern end of the farm; the primary H2S
source is expected to be the lagoons. The perimeter of the farm is lined with trees and
agricultural fields surround the perimeter. A temperature-regulated instrument trailer will be
placed on-site during the test to house the monitoring equipment and to provide a sheltered work
space. The H2S analyzers will be installed inside the instrument trailer and a Teflon inlet line or
manifold will be used to sample ambient air. Sample tubing lengths will be minimized both for
ambient air sampling and for delivery of gas standards.

$

2
$

$

2
$

2
$

g

$

8

Ng

ฃ

Entrance ^

O

2
ฃ

*
&

O

Trailer | |

Nutrient
Lagoons

\

\



0

2
$

#

$
$

2
$

2

2

2
$

$

2
$

S

2

2

2
$

$

2
$

	 		o

Figure 2. Test site


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 20 of 46
Version 1.0
April 12, 2005

A7 QUALITY OBJECTIVES

This verification test will evaluate the performance of analyzers for determining H2S in
ambient air at a swine finishing farm. This evaluation will include a comparison of the analyzer
results to the results of two reference methods for H2S. The quality of the reference
measurements will be monitored by inclusion of blank samples, performance evaluation (PE)
audit samples, and duplicate samples, when applicable. The PE audit samples will be prepared
using H2S compressed gas cylinders or permeation devices that are independent of those used for
reference method calibration. These samples are meant to independently confirm that the
reference measurements are being performed correctly and are producing accurate results.

Control limits on the duplicate and PE samples are given in Section CI. All H2S standards to be
used to challenge the H2S analyzers and to prepare PE audit samples must meet National
Institute of Standards and Technology (NIST) traceability. Commercially available compressed
gas standards or permeation devices may be used in this test.

The Battelle Quality Manager or his designate will perform a technical systems audit
(TSA) at least once during this verification test. The EPA Quality Manager also may conduct an
independent TSA, at her discretion.

A8 SPECIAL TRAINING/CERTIFICATION

Documentation of training related to technology testing, field testing, data analysis, and
reporting is maintained for all Battelle technical staff in training files. Documentation of the
expertise and experience of USDA-NSTL and Applied Measurement Science staff in H2S
reference method sampling and analysis is similarly available. The Battelle Quality Manager
may verify the presence of appropriate training records prior to the start of testing. If Battelle or
USDA staff operate and/or maintain an analyzer during the verification test, the analyzer vendor
will be required to train those staff prior to the start of testing. Battelle will document this
training with a consent form, signed by the vendor, that states which specific Battelle/USDA
staff have been trained on their analyzer. Battelle technical staff will have a minimum of a
bachelor's degree in science/engineering or have equivalent work experience.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 21 of 46
Version 1.0
April 12, 2005

A9 DOCUMENTATION AND RECORDS

The records for this verification test will include the test/QA plan, chain-of-custody
(COC) forms, laboratory record books (LRB), data collection forms, electronic files (both raw
data and spreadsheets), and the final verification report and verification statement. All of these
records will be maintained in the Verification Test Coordinator's office or the test site during the
test and will be transferred to permanent storage at Battelle's Records Management Office at the
conclusion of the verification test. The location (e.g., specific personal computer, server, or
media type and storage location) of final versions of the electronic files will be noted in the test
records. All Battelle LRBs are stored indefinitely, either by the Verification Test Coordinator or
Battelle's Records Management Office. EPA will be notified before disposal of any files. The
documentation and results of the H2S reference measurements made by USD A and Applied
Measurement Science will be submitted to Battelle immediately upon completion of all sample
analyses, review of the H2S data, and calculation of ambient H2S concentrations, preferably on a
daily basis. Section BIO further details the data recording practices and responsibilities.

All written records must be in ink. Any corrections to notebook entries, or changes in
recorded data, must be made with a single line through the original entry. The correction is then
to be entered, initialed, and dated by the person making the correction. In all cases, strict
confidentiality of data from each vendor's analyzer, and strict separation of data from different
analyzers, will be maintained. Separate files (including manual records, printouts, and/or
electronic data files) will be kept for each analyzer.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 22 of 46
Version 1.0
April 12, 2005

SECTION B
MEASUREMENT AND DATA ACQUISITION

B1 EXPERIMENTAL DESIGN

This test will specifically address verification of analyzers for ambient H2S under the
conditions of a swine finishing farm by evaluating the accuracy, bias, linearity, and selectivity of
H2S measurements made by each analyzer in ambient air and/or synthetic gas mixtures, their
response to rapid changes in H2S concentration, the extent of baseline and calibration drift, and
the ability of the analyzers to perform continuous monitoring with minimal intervention. The
precision of the analyzer measurement data while sampling synthetic H2S gas mixtures will also
be determined. Specifically, the H2S analyzers will be evaluated for the performance parameters
summarized in Table 3 and discussed in detail in the following section.

Table 3. Verification Test Performance Parameters

Performance
Parameter

Method of Evaluation

Accuracy

Analyzer response to H2S standards compared to nominal concentrations

Bias

Analyzer response to H2S standards compared to nominal concentrations

Precision

Percent relative standard deviation (%RSD) of repeated analysis H2S standards
with the same concentration

Linearity

Analyzer response to H2S standards compared to nominal concentrations

Span and Zero Drift

Stability of analyzer response to zero air and H2S gas standard over time

Response Time

Time required to reach 95% of response to H2S gas standards

Interference Effects

Analyzer response to non-H2S sulfur-containing compounds, butyric acid, and
ammonia compared to nominal concentrations

Comparability

Analyzer response to ambient air compared to results of two reference methods

Data Completeness

Percentage of maximum data return over field period

Operational Factors

Operator observations, records of needed and performed maintenance, vendor
activities, use of expendable supplies


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 23 of 46
Version 1.0
April 12, 2005

In addition to the testing activities specified in this test/QA plan, Battelle and/or USDA
staff will perform regular maintenance and other routine procedures requested by the vendor for
their analyzer. This information will be summarized by the vendor in a "daily checklist" that will
include specific instructions and frequency for each regular maintenance activity or routine
procedure, any diagnostic values that should be recorded and typical or acceptable ranges for
those values, vendor contact information, and a space for Battelle and/or USDA staff to sign and
date the form daily upon completion of the items on the checklist.

Bl.l Test Procedures

The following sections describe the test procedures that will be used to evaluate each of
the H2S analyzer performance parameters listed in Table 3. Procedures will be conducted on
each analyzer separately unless a common manifold is used to deliver challenge gases and
sample ambient air. If a manifold is used, testing activities will be performed on the H2S
analyzers simultaneously. The manifold will be constructed of a bored-out 2.5 inch (in) inner-
diameter Teflon cylinder that is 10 inches length. The manifold has 3/8 in inlet and outlet
fittings, and four 1/4 in male connectors (at the same position along the length) to which the H2S
analyzers can be connected. A pump will be installed at the outlet port and configured to draw a
flow through the manifold that is in excess of that required by the H2S analyzers [approximately
3 liters per minute (Lpm)].

The electronic analyzer responses will be recorded continuously throughout the
verification test and used for all analyses. Each gas standard will be delivered for a minimum of
20 minutes to allow for stabilization of the analyzer response. Gas standards will be prepared
from the dilution of higher concentration (i.e., 2-500 part per million) standard cylinders in zero
air using a calibrated dilution system with heated internal components. Gas standards will be
supplied in excess of the analyzer sample flow rate; the excess flow will be vented outside the
trailer to ambient pressure.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 24 of 46
Version 1.0
April 12, 2005

B 1.1.1 Accuracy, Bias, Precision, and Linearity

Twice during the verification test, the analyzers will be challenged with compressed H2S
gas standards diluted in zero air to achieve measurements over a range of concentrations from
approximately 0 to 100 ppb (or the upper range of measurement of the analyzer being tested,
whichever is lower). If the ambient H2S concentrations are found to be significantly larger than
100 ppb, additional H2S gas dilutions may be tested to reflect the actual range, if possible.

Three non-consecutive measurements will be recorded at each of five different nominal
concentration levels. Each concentration will be supplied to the analyzer(s) for at least twenty
minutes. A programmable dilution system may be used to automatically supply the diluted gas
standards to the H2S analyzer(s) at fixed time intervals. Table 4 shows the nominal H2S
concentration values to be supplied to the analyzers being tested, and the order in which the
concentrations will be supplied, for conditions where the maximum H2S level encountered at the
site is -100 ppb. If H2S concentrations greater than 200 ppb are encountered, the H2S challenge
gas concentrations may be modified upwards (e.g., to 0, 20, 60, 100, and 200 ppb) to more
accurately reflect the range of ambient H2S levels. As Table 4 indicates, the H2S concentrations
will first be supplied to the analyzers in increasing order, then in random order, and finally in
decreasing order. After the last measurement has been recorded, the analyzer will be returned to
sampling of ambient air.

The analyzer response to the series of H2S gas standards will be used to evaluate
accuracy, bias, precision, and linearity. Section B 1.2 presents the statistical procedures that will
be used. Accuracy will be calculated at each concentration and for each replicate relative to the
nominal H2S concentration. Bias will be calculated for each series of multi-point H2S challenges.
The analyzer precision will be demonstrated by the reproducibility of the analyzer response at
each nominal H2S concentration after a stable reading is achieved. Linearity will be assessed by
establishing a multi-point calibration curve from the analyzer response.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 25 of 46
Version 1.0
April 12, 2005

Table 4. Approximate H2S Concentrations and Order for Multi-point Challenges

Concentration

0 ppb

10 ppb

30 ppb

50 ppb

100 ppb



1

2

3

4

5

Measurement
Number

7

10

6

9

8

15

14

13

12

11

Bl. 1.2 Span and Zero Drift

The "baseline" response of each analyzer to zero air and a 30 ppb dilution of a
compressed H2S gas standard will be determined during the first week of testing. Each analyzer
will be alternately challenged with the diluted H2S gas standard and zero air, for a total of five
replicates of both the gas standard and zero air. Each gas will be supplied sequentially to the
analyzer for at least twenty minutes. The switch between zero air and the H2S standard will be
made as quickly as possible if conducted manually; a programmable dilution system will be used
if available. The mean and standard deviation of the analyzer response to zero air and to the 30
ppb H2S standard will be calculated from the five replicates.

Twice each week (preferably Monday and Friday) during the verification test, zero air
and a 30 ppb H2S standard will again be supplied to each of the analyzers being tested for twenty
minutes each for a total of nine zero/span checks. Each response will be compared to baseline
response to determine whether or not drift has occurred in the analyzer response to zero air or the
30 ppb H2S standard.

Bl. 1.3 Response Time

The data collected for the zero/span check (Section B 1.1.2) will also be used to determine
the analyzer response time. The 95% rise time will be calculated for changes from zero air to the
30 ppb H2S standard and the fall time will be calculated for changes from the 30 ppb standard to
zero air. A minimum of three individual measurements will be used to determine the average rise
and fall times.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 26 of 46
Version 1.0
April 12, 2005

Bl. 1.4 Interference effects

The analyzers being tested will be challenged with a series of gases (interferants, see
Table 5) that may be present at an AFO and could interfere with the analyzer response to H2S.
Each interferant will be supplied at approximately 100 ppb in the presence and absence of 100
ppb H2S. A 100 ppb H2S standard will be supplied to the analyzers for at least twenty minutes
and the responses will be recorded. The analyzer will then be flushed with zero air for at least
two minutes. The first interferant will be diluted with zero air and delivered to the analyzers for
at least twenty minutes. The analyzer responses will be recorded and each analyzer flushed for at
least two minutes with zero air. A mixture of the first interferant with 100 ppb H2S in zero air
will then be supplied to each analyzer for at least twenty minutes. The analyzer responses will be
recorded, and zero air will be supplied to each analyzer for approximately two minutes. This
process will be repeated for each of the interferants. The interferant challenges need not all be
completed in a single day.

Table 5. Interferants and Approximate Concentrations for Interference Checks

Interferant

Approximate
Concentration (ppb)

Sulfur dioxide

100

Carbonyl sulfide

100

Carbon disulfide

100

Methyl mercaptan

100

Dimethyl sulfide

100

Hydrocarbon blend

500

Ammonia

500

Bl.1.5 Comparability

The comparability of the H2S analyzer response to ambient air will be evaluated by
comparing the analyzer response to two H2S reference methods, which will be carried out by


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 27 of 46
Version 1.0
April 12, 2005

USDA and Applied Measurement Science. Both methods follow ASTM International Method
D5504-011 for the determination of sulfur compounds using GC, but will use PFPD instead of
SCD. Although the analytical approach of both methods is the same, the two methods differ in
sample handling. The two reference methods will be compared to one another by calculating the
relative percent difference (RPD) (the difference between the two reference method values
divided by the average). This comparison will assess whether the different sample collection
techniques produce differences in the measured H2S concentrations. Details of the reference
sample collection procedure and laboratory method are provided in sections B2 and B4,
respectively. The QA procedures for the reference methods are described in section B5.

B1.1.1,5a Time-integrated Comparability

The USDA reference method will utilize time-integrated air samples that will be
collected in Silonite canisters and taken to the USDA laboratory for analysis. Samples will be
collected over two eight-hour intervals on each sampling day. Sampling is expected to be
conducted on approximately the following schedule: 10:00 p.m. to 6:00 a.m. and 6:00 a.m. to
2:00 p.m. The sampling times have been selected to accommodate the schedule of the USDA
analysis laboratory. Time-integrated reference measurements will be conducted on a total of ten
days during the field period. The samples may be collected twice per week or some samples may
be rescheduled to collect more frequently during the in-situ H2S reference method measurement
period. The results of a minimum of 15 time-integrated reference method measurements will be
compared to the time-averaged analyzer responses over the same time periods to give the time-
integrated comparability of H2S analyzers.

B1.1.1,5b In-situ Comparability

The in-situ H2S reference method will be conducted by Applied Measurement Science.
The instrumentation for the in-situ method will be installed in the instrument trailer at the test
site. Discrete air samples will be drawn from the same location outside of the trailer, or
preferably from the common sampling manifold, over a relatively short time period. Volatile
compounds in the samples will be cryotrapped, thermally desorbed, and injected directly onto


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 28 of 46
Version 1.0
April 12, 2005

the GC-PFPD system. To the extent possible, the duration of sample collection will be
coordinated with the sampling frequency of the H2S analyzers being tested (approximately two
to 30 minutes). In-situ reference measurements will be conducted for approximately 8 hours per
day for ten days during the second and third weeks of the verification test. This in-situ system
decreases the chance for H2S loss due to sample handling, but does require a more complex air
handling system to automate the sample collection and injection. The results of a minimum of 60
in-situ H2S reference method measurements will be compared to the analyzer responses to
ambient air recorded at the same time. If the measurement frequency of the in-situ H2S reference
method and the H2S analyzer being tested do not match in time within ฑ two minutes, each data
set will be averaged to common time intervals (e.g., hourly averages).

Bl. 1.6 Data Completeness

No additional test procedures will be carried out specifically to address data
completeness. This parameter will be assessed based on the overall data return achieved by each
analyzer.

Bl. 1.7 Operational Factors

Operational factors such as maintenance needs, data output, consumables used, ease of
use, repair requirements, etc., will be evaluated based on observations recorded by Battelle and
USD A staff. A separate LRB will be maintained at the test site for each analyzer undergoing
testing, and will be used to enter daily observations on these factors. Examples of information to
be recorded in the record books include the daily status of diagnostic indicators for the analyzer;
use or replacement of any consumables; the effort or cost associated with maintenance or repair;
vendor effort (e.g., time on site) for repair or maintenance; the duration and causes of any
analyzer down time or data acquisition failure; and operator observations about ease of use of the
analyzer. These observations will be summarized to aid in describing analyzer performance in
the verification report on each analyzer.


-------
Ambient Hydrogen Sulfide Analyzers

Test/QA Plan
Page 29 of 46

Version 1.0
April 12, 2005

B1.2 Statistical Analysis

The statistical methods and calculations used for evaluation of the quantitative
performance parameters are described in the following sections.

B 1.2.1 Accuracy

Accuracy of the H2S analyzers with respect to the individual H2S gas standards will be
assessed as the percent recovery (%R), using Equation 1:

where Y is the average measured H2S analyzer value and X is the nominal H2S gas standard
concentration. The average, minimum, and maximum %R values will be reported for each series
of multi-level H2S challenges.

Bias of the H2S analyzers is defined as a systematic error in measurement that results in
measured error that is consistently positive or negative compared to the true value. The bias will
be calculated as the average percent difference (%D) of the H2S analyzer compared to the
nominal H2S gas standard concentration and will be calculated for each series of multi-point H2S
challenges, using Equation 2:

where k is the number of valid comparisons, and Y and X are the same as stated in B 1.2.1.

(1)

Bl. 2.2 Bias

(2)

B 1.2.3 Precision

The precision of the H2S analyzers will be evaluated from the triplicate responses to each
H2S gas standard supplied during the multi-point challenges (outlined in Table 4). The precision


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 30 of 46
Version 1.0
April 12, 2005

will be defined as the percent relative standard deviation (%RSD) of the triplicate measurements
and calculated for each H2S concentration listed in Table 4, using Equation 3:

%RSDj=J^xl00

where Y is the average analyzer response at H2S concentration i, and s the standard deviation of
the analyzer responses at that concentration. The overall average %RSD will also be calculated
for each series of multi-point H2S challenges and will include the %RSD for all H2S
concentrations tested.

B 1.2.4 Linearity

Linearity will be assessed by a linear regression analysis using the diluted H2S standard
gas concentrations as the independent variable and results from the H2S analyzers being tested as
the dependent variable. Linearity will be expressed in terms of slope, intercept, and coefficient of
determination (r2).

B 1.2.5 Span and Zero Drift

The "baseline" response of the H2S analyzers to zero air and the 30 ppb H2S standard will
be established on the first day of testing, as outlined in Section B 1.1.2. The mean CO and
standard deviation (s) of the analyzer response to zero air and 30 ppb H2S will each be calculated
from the five replicate measurements conducted on the first day of testing. From these values, a
control chart will be constructed and the Y ฑ 2s "warning limit" and the Y ฑ 3s "action limit"
calculated. Span drift will be defined as having occurred if three consecutive span checks all fall
either above or below the warning limit. Zero drift will also be defined as having occurred if
three consecutive zero checks all fall either above or below the warning limit. However, if the
mean and/or standard deviation from the baseline zero check are equal to zero, the warning
limits may not be meaningful. In this case, the absolute differences to the zero air baseline mean
will be reported for each zero check.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 31 of 46
Version 1.0
April 12, 2005

B 1.2.6 Response Time

Response time will be assessed in terms of both the rise and fall times of each H2S
analyzer when sampling the 30 ppb H2S gas standard and zero air on the first day of testing
(Section B 1.1.2). Rise time (i.e., 0% - 95% response time for the change in H2S concentration)
will be determined from the analyzer response to a rapid increase in the delivered H2S
concentration. Once a stable response is achieved with the H2S standard, the fall time (i.e., the
100% to 5% response time) will be determined in a similar way, switching from the H2S
standard back to zero air.

B1.2.7 Interference Effects

The interference effects of the H2S analyzers will be calculated in terms of the ratio of the
response of the analyzer to the interferant relative to the actual concentration of the interfering
species. For example, if 100 ppb of an interfering species results in a 1 ppb change in the
response of the analyzer, the interference effect will be reported as 1% (i.e., 1 ppb/100 ppb).
Interference effects will be reported separately for each interferant both in the absence and in the
presence of H2S in zero air.

B 1.2.8 Comparability

Comparability between the H2S analyzer results and the reference method results will be
assessed by linear regression using the reference method H2S concentrations as the independent
variable and results from the H2S analyzers being tested as the dependent variable. Linearity will
be expressed in terms of slope, intercept, and r2, and will be calculated independently for the
time-integrated H2S reference method and the in-situ H2S reference method. It is expected that
the measured concentration of H2S will vary by at least a factor of five during each phase of
testing. However, if this magnitude of variation is not achieved for one or both of the reference
methods, comparability for that method will be calculated using Equation 1 and Equation 2 and
reported as a percent recovery and bias, rather than in terms of the linear regression results.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 32 of 46
Version 1.0
April 12, 2005

B 1.2.9 Data Completeness

Data completeness will be calculated as the percentage of the total possible data return
over the entire field period that is achieved by each analyzer. This calculation will use the total
hours of data recorded from each analyzer, divided by the total hours of data in the entire field
period. The field period is defined to begin at 8:00 a.m. on the first day of testing and to end at
the completion of the last testing activity or 5:00 p.m. on the final day of the field period,
whichever is later. No distinction will be made in this calculation between data recorded during a
specific test activity (e.g., data recorded for comparison to H2S reference method data) and that
recorded during routine ambient air monitoring. The causes of any substantial incompleteness of
data return will be established from operator observations or vendor records, and noted in the
discussion of data completeness results.

B1.3 Reporting

The statistical comparisons described above will be conducted separately for each of the
analyzers being tested, and information on the operational parameters will be compiled and
reported. The data for each analyzer will be kept separate from data for all other analyzers, and
no intercomparison of the analyzer data will be performed at any time. A separate verification
report will be prepared for each analyzer tested, presenting the test procedures and test data, as
well as the results of the statistical evaluation of those data.

Operational aspects of the analyzers will be recorded by testing staff at the time of
observation during the field test, and summarized in the verification report. For example,
descriptions of the data-acquisition procedures, use of vendor-supplied proprietary software,
consumables used, repairs and maintenance needed, and the nature of any problems will be
presented in the report. Each verification report will briefly describe the ETV program, the AMS
Center, and the procedures used in verification testing. The results of the verification test will be
stated quantitatively, without comparison to any other analyzer tested, or comment on the
acceptability of the analyzer's performance. Each draft verification report will first be subjected
to review by the respective analyzer vendor, then revised and subjected to a review by EPA and
other peer reviewers. The peer review comments will be addressed in further revisions of the


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 33 of 46
Version 1.0
April 12, 2005

report, and the peer review comments and responses will be tabulated to document the peer
review process. The reporting and review process will be conducted according to the
requirements of the AMS Center QMP.2

B2 REFERENCE SAMPLE COLLECTION
B2.1 Time-integrated H2S Reference Method

The time-integrated reference method will utilize air samples that will be collected in
evacuated 1.4 L Silonite Canisters (Entech Instruments, Inc.), which will be transported from the
field site to the USDA laboratory for analysis. A restrictive sampler will be used to control the
fill rate over 8 hours and maintain sub-ambient pressure in the canister at the completion of
sampling (i.e., fill to -80% capacity). Under sub-ambient pressure, H2S scavenging by water
vapor has been shown to be negligible in canisters with Silcosteelฎ-treated surfaces at 50%
relative humidity for at least 48 hours.8 All components that contact ambient air samples will be
Teflon or passivated metal (e.g., Silcosteelฎ or Silonite) to minimize scavenging of H2S by bare
metal or other surfaces. In accordance with ASTM Method D5504-011, canister samples will be
analyzed by GC-PFPD within 24 hours.

As indicated in ASTM Method D5504-01,1 it has been demonstrated that H2S in
passivated canisters does not degrade over 24 hours; therefore, detailed holding time tests are not
necessary. However, the acceptability of this holding time will be verified by analyzing an
ambient air sample several times over the period (at least 24 hours) following sample collection.
The H2S concentration at 24 hours following sample collection should be within 15% of the
initial measured value. Holding time verifications will be performed at least once before the start
of the verification test and again mid-way through the test.

B2.2 In-situ H2S Reference Method

Reference samples need not be collected for the in-situ H2S reference method, since this
approach relies on direct cryotrapping of volatile components from a small sample air flow over


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 34 of 46
Version 1.0
April 12, 2005

a relatively short time period, followed immediately by thermal desorption and GC-PFPD
analysis.

B3 SAMPLE HANDLING AND CUSTODY REQUIREMENTS

All reference samples will be entirely in the custody of USD A or Battelle from sample
collection through sample recovery, transport, and analysis. Samples will be carried by USDA or
Battelle staff to the USDA laboratory for analysis. Sample custody will be documented
throughout installation of the canisters at the field site, ambient air collection, recovery of the
sample canisters, transport, and analysis of the reference samples, using standard forms used by
USDA for this purpose or forms provided by Battelle. Each COC form will be signed by the
person relinquishing samples once that person has verified that the COC form is accurate. Upon
receipt at the laboratory, COC forms will be signed by the person receiving the samples once
that person has verified that all samples identified on the COC forms are present. Any
discrepancies will be noted on the form and the sample receiver will immediately contact the
USDA sampling leader to report missing or compromised samples. Copies of all COC forms will
be delivered to the Verification Test Coordinator upon request, and maintained with the test
records.

B4 LABORATORY REFERENCE METHODS

Although the ASTM H2S reference method1 is primarily for determination of sulfur
compounds in natural gas and gaseous fuels, it has successfully been applied to H2S detection in
ambient air. Both of the reference methods described in this test/QA plan follow the ASTM
method1 and adhere to the general measurement principle of H2S determination by GC with
sulfur chemiluminescence or other detection technique. In both methods, a cryofocus interface
and thermal desorption will be utilized to deliver the ambient air sample to the GC for
separation. Instead of SCD, PFPD will be used to detect H2S. QA procedures for both reference
methods are described in section B5.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 35 of 46
Version 1.0
April 12, 2005

B4.1 Time-integrated H2S Reference Method

The time-integrated H2S reference method will use an Agilent 5890 GC equipped with a
PFPD from 01 Analytical. The inlet device will be a Model 7100A Preconcentrator (Entech) that
utilizes cryofocusing and thermal desorption to transfer the sample to the head of the column.
The column is a GS-Gaspro, 60 meter (m) x 0.32 millimeter (mm) inner diameter (i.d.) capillary
column (J & W).

A multi-point calibration curve for H2S will be constructed before reference analyses are
conducted using dilute standards prepared from NIST-traceable certified H2S compressed gas
standards (Scott Specialty Gases). Several gas concentrations will be prepared using a dynamic
dilution system (Environics), transferred into Silonite canisters, and analyzed in the same manner
as the reference samples. In addition to H2S, chromatograms will include retention time data for
sulfur dioxide, methyl mercaptan, dimethyl sulfide, carbonyl sulfide, and carbon disulfide, if
present at sufficient concentration.

All analyses of reference and QA/QC samples will be conducted by USD A staff. USD A
is responsible for providing the analytical instrumentation, calibrating that instrumentation,
performing method QA/QC (see Section B5), and maintaining calibration records for any
instrumentation used. USDA will be required to provide Battelle with documentation on
calibration and quality control of the reference analyses.

B4.2 In-situ H2S Reference Method

The in-situ H2S reference method will use a Varian 3800 GC with PFPD. The inlet
system contains an internal valving system that allows for automated operation. The column is a
GS-Gaspro, 30 m x 0.32 mm i.d. capillary column (J & W).

A multi-point calibration curve for H2S will be constructed at the test site before
reference method measurements will be conducted, using dilute standards prepared from NIST-
traceable certified H2S compressed gas standards or certified permeation tubes and sampled in
the same way as ambient air. In addition to H2S, several sulfur-containing species, including,


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 36 of 46
Version 1.0
April 12, 2005

sulfur dioxide, methyl mercaptan, dimethyl sulfide, carbonyl sulfide, and carbon disulfide, will
also be quantified if present at sufficient concentration.

All analyses of reference and QA/QC samples for the in-situ reference method will be
conducted by Applied Measurement Science staff. Applied Measurement Science is responsible
for providing the analytical instrumentation, calibrating that instrumentation, performing method
QA/QC (see Section B5), and maintaining calibration records for any instrumentation used.
Applied Measurement Science will be required to provide Battelle with documentation on
calibration and quality control of the reference analyses.

B5 QUALITY CONTROL

Steps will be taken to maintain the quality of the data collected during this verification
test. Table 6 summarizes the quality control requirements for the reference methods during this
test. Although these requirements differ from the suggested QA procedures outlined in ASTM
method D5504-01,1 they are more appropriate for the detection of ambient H2S at ppb levels
(versus ppm levels) in an ambient air matrix. Both reference methods will be required to analyze
continuing calibration verifications (CCV), quality control samples (QCS), and field blanks. The
time-integrated H2S reference method will also be required to repeat analyses of 10% of the
samples to verify the method precision. The reference analytical procedure will be maintained to
meet these requirements. If the analytical performance is outside of the required tolerances, the
relevant QC samples will be prepared again and reanalyzed. If performance problems persist, the
reference instruments) will be recalibrated, and/or affected samples will be reanalyzed.
Reference sample results not meeting these requirements will be excluded from comparison to
the H2S analyzer results. A minimum of 15 time-integrated reference measurements and 60 in-
situ reference measurements that meet these QC requirements must be collected for use in the
comparability comparisons, as stated in sections B 1.1.1.5a and B 1.1.1.5b, respectively.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 37 of 46
Version 1.0
April 12, 2005

Table 6. Reference Method Quality Control Requirements and Target Acceptance Criteria

QC Parameter

Addressed By

Required Performance

CCV

CCV run before analysis of reference
samples each day

%D of CCV result within 30%
compared to expected value

QCS

QCS run every 4 hours and after analysis
of reference samples each day

%D of QCS result within 30%
compared to expected value

Replicate H2S
precision

Analyze 10% of all samples twice(a)

Results within 30% of one
another

Measurement
accuracy

Analyze H2S standard from independent
sourceฎ

Results within 30% of expected
value

Field blanks

Analyze canisters filled with zero air
recovered from the field site (weekly)1-3-1
Analyze zero air passed through sample
manifold (weekly)(c)

If blank >30% of sample H2S,
data must be flagged

(a) Time-integrated H2S reference method only.

:l" This standard will be provided as part of the Performance Evaluation audit (Section C. 1.1).
(c) In-situ H2S reference method only.

B6 INSTRUMENT/EQUIPMENT TESTING, INSPECTION, AND MAINTENANCE

The equipment used for the reference sampling and analysis will be tested, inspected, and
maintained as per the standard operating procedures of USD A and Applied Measurement
Science, so as to meet the performance requirements established in this document. When Battelle
or USD A staff operate and maintain the H2S analyzers undergoing testing, those activities will
be done as directed by the vendor. Otherwise, operation and maintenance of the analyzers will be
the responsibility of the analyzer vendors.

B7 INSTRUMENT CALIBRATION AND FREQUENCY

The instrumentation used for the reference H2S analyses will be calibrated daily on days
when reference samples will be analyzed. The calibration of other instrumentation used in this
verification test, such as dilution systems and flow readers, will be verified immediately prior to
use in this verification test. A minimum of three flow rates for each flow controller or flow
reader will be verified with an independent factory-calibrated flow meter.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 38 of 46
Version 1.0
April 12, 2005

The H2S analyzers undergoing testing will be calibrated initially by the respective
analyzer vendors at the time of installation at the test site using H2S gas standards independent of
those used for testing activities. Calibration checks will be performed upon direction by the
analyzer vendor. In the event that recalibration is necessary, that recalibration will be carried out
by the analyzer vendor, or by Battelle staff under the direction of the vendor. All calibrations
performed will be documented by Battelle or USD A staff in the project record book dedicated to
the respective analyzer.

B8 INSPECTION/ACCEPTANCE OF SUPPLIES AND CONSUMABLES

All materials, supplies, and consumables will be ordered by the Verification Test
Coordinator or designee. Where possible, Battelle will rely on sources of materials and
consumables that have been used previously as part of ETV verification testing without
problems. Battelle will also rely on previous experience or recommendations from peer
reviewers, EPA advisors, USDA staff, Applied Measurement Science, or analyzer vendors.
Hydrogen sulfide gas standards will have NIST-traceable certifications.

B9 NON-DIRECT MEASUREMENTS

Data published previously in the scientific literature will not be used during this
verification test.

BIO DATA MANAGEMENT

Various types of data will be acquired and recorded electronically or manually by
Battelle, vendor, USDA, and Applied Measurement Science staff during this verification test.
Table 7 summarizes the types of data to be recorded. All maintenance activities, repairs,
calibrations, and operator observations relevant to the operation of the H2S analyzers will be
documented by Battelle or USDA staff in LRBs. A separate record book will be provided for
each participating analyzer. Results from the reference methods, including raw data, analysis,


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 39 of 46
Version 1.0
April 12, 2005

and final results, will be compiled by USDA and Applied Measurement Science staff in
electronic format, and submitted to Battelle at the conclusion of reference H2S analyses.

Table 7. Summary of Data Recording Process

Data to Be
Recorded

Where Recorded

How Often
Recorded

By Whom

Disposition of
Data

Dates, times, and
details of test
events, analyzer
maintenance,
down time, etc.

ETV LRBs or
data recording
forms

Start/end of test
procedure, and at
each change of a
test parameter or
change of
analyzer status

Battelle if on-site;
USDA if Battelle
not on-site

Used to organize
and check test
results; manually
incorporated in
data spreadsheets
as necessary

Analyzer

calibration

information

ETV LRBs or
electronically

At analyzer
calibration or
recalibration

Electronic data by
Vendor; Battelle
if on-site; USDA
if Battelle not on-
site

Incorporated in
verification report
as necessary

Analyzer H2S
readings

Recorded
electronically by
each analyzer and
then downloaded
to computer at
least weekly

Recorded
continuously

Analyzer vendor,
for transfer to
Battelle if on-site;
transfer to USDA
if Battelle not on-
site

Converted to
spreadsheet for
statistical analysis
and comparisons

Reference sample

collection

procedures,

reference method

procedures,

calibrations, QA,

etc.

LRBs, or data
recording forms

Throughout
sampling and
analysis processes

USDA and
Applied
Measurement
Science

Retained as
documentation of
reference method
performance

Reference method
H2S analysis
results

Electronically
from H2S
analytical method

Every sample
analysis

USDA and
Applied
Measurement
Science

Converted to
spreadsheets for
calculation of
ambient H2S
results, and
statistical analysis
and comparisons

Records received by or generated by any Battelle or USDA staff during the verification
test will be reviewed by a Battelle staff member within two weeks of generation or receipt,


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 40 of 46
Version 1.0
April 12, 2005

respectively, before the records are used to calculate, evaluate, or report verification results. If a
Battelle staff member generated the record, this review will be performed by a Battelle technical
staff member involved in the verification test, but not the staff member who originally generated
the record. The review will be documented by the person performing the review by adding
his/her initials and date to the hard copy of the record being reviewed. In addition, any
calculations performed by Battelle or USD A staff will be spot-checked by Battelle technical staff
to ensure that calculations are performed correctly. Calculations to be checked include any
statistical calculations described in this test/QA plan. The data obtained from this verification
test will be compiled and reported independently for each H2S analyzer. Results for analyzers
from different vendors will not be compared with each other.

Among the QA activities conducted by Battelle QA staff will be an audit of data quality.
This audit will consist of a review by the Battelle Quality Manager of at least 10% of the test
data. During the course of any such audit, the Battelle Quality Manager will inform the technical
staff of any findings and any immediate corrective action that will be taken. If serious data
quality problems exist, the Battelle Quality Manager will request that Battelle's AMS Center
Manager issue a stop work order. Once the assessment report has been prepared, the Verification
Test Coordinator will ensure that a response is provided for each adverse finding or potential
problem, and will implement any necessary follow-up corrective action. The Battelle Quality
Manager will ensure that follow-up corrective action has been taken.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 41 of 46
Version 1.0
April 12, 2005

SECTION C
ASSESSMENT AND OVERSIGHT

CI ASSESSMENTS AND RESPONSE ACTIONS

Every effort will be made in this verification test to anticipate and resolve potential
problems before the quality of performance is compromised. One of the major objectives of this
test/QA plan is to establish mechanisms necessary to ensure this. Internal quality control
measures described in this test/QA plan, which is peer reviewed by a panel of outside experts,
implemented by the technical staff and monitored by the Verification Test Coordinator, will give
information on data quality on a day-to-day basis. The responsibility for interpreting the results
of these checks and resolving any potential problems resides with the Verification Test
Coordinator. Technical staff have the responsibility to identify problems that could affect data
quality or the ability to use the data. Any problems that are identified will be reported to the
Verification Test Coordinator, who will work to resolve any issues. Action will be taken to
control the problem, identify a solution to the problem, and minimize losses and correct data,
where possible. Independent of any EPA QA activities, Battelle will be responsible for ensuring
that the following audits are conducted as part of this verification test.

Cl.l Performance Evaluation Audits

A Performance Evaluation (PE) audit will be conducted to assess the quality of the H2S
reference method measurements made in this verification test. In the PE audit, key aspects of the
reference measurement will be checked by comparison with an independent instrument, or an
independent NIST-traceable standard. The PE audit of the H2S reference methods will be
performed by supplying to each reference method a blind, independent, NIST-traceable H2S
standard provided by Battelle. The PE samples will be analyzed in the same manner as for all
other ambient air samples, and the analytical results for the PE samples will be compared to the


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 42 of 46
Version 1.0
April 12, 2005

nominal concentration. The target criterion for this PE audit is agreement of the analytical result
within 30% of the nominal H2S concentration. If the PE audit results do not meet the tolerances
required, they will be repeated. If the outlying results persist, a change in reference instrument
and a repeat of the PE audit may be considered. This audit will be performed once prior to the
start of the test and two times during the verification test, and will be the responsibility of the
Verification Test Coordinator or her designee. Since the PE audit samples will be sampled by the
reference methods in the same way as reference samples, they will assess each component of the
reference method, including inertness of metal sampling components, H2S stability in the
sampling container (if applicable), and analytical accuracy of the GC-PFPD systems.

A PE audit of the ambient air sample flow rate for the time-integrated reference method
will be performed by comparing to an independent flow measurement device. The target
criterion for this PE audit is agreement within the expected range (i.e., 2 to 3 milliliters per
minute). If used in this verification test, a PE audit of the programmable dilution system will be
performed by comparing to an independent flow measurement device. One mid-range flow rate
will be audited for each flow controller (i.e., 0.03, 0.3, and 5 Lpm) within the dilution system.
The target criterion for this PE audit is agreement within 5% of the flow readings. These audits
will be performed once during the verification test, and will be the responsibility of the
Verification Test Coordinator or her designee.

C1.2 Technical Systems Audits

The Battelle Quality Manager will perform a technical systems audit (TSA) at least once
during this verification test. The purpose of this audit is to ensure that the verification test is
being performed in accordance with the AMS Center QMP2, this test/QA plan, published
reference methods, and any Standard Operating Procedures (SOPs) used by USD A or Applied
Measurement Science. In the TSA, the Battelle Quality Manager, or a designee, may review the
reference methods used, compare actual test procedures to those specified or referenced in this
plan, and review data acquisition and handling procedures. In the TSA, the Battelle Quality


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 43 of 46
Version 1.0
April 12, 2005

manager will tour the test site, observe the H2S reference method sampling and sample recovery,
inspect documentation of H2S sample chain of custody; and review analyzer-specific record
books. He will also check gas standard certifications and analyzer data acquisition procedures,
and may confer with the analyzer vendors, USD A, and Applied Measurement Science personnel.
He may also visit the USDA laboratories where the time-integrated H2S reference method
analysis is conducted, to review procedures and adherence to this plan and applicable SOP's. A
TSA report will be prepared, including a statement of findings and the actions taken to address
any adverse findings. The EPA AMS Center Quality Manager will receive a copy of Battelle's
TSA report. At EPA's discretion, EPA QA staff may also conduct an independent on-site TSA
during the verification test. The TSA findings will be communicated to technical staff at the time
of the audit and documented in a TSA report.

C1.3 Data Quality Audits

The Battelle Quality Manager will audit at least 10% of the verification data acquired in
the verification test. The Battelle Quality Manager will trace the data from initial acquisition,
through reduction and statistical comparisons, to final reporting. All calculations performed on
the data undergoing the audit will be checked.

CI.4 QA/QC Reporting

Each assessment and audit will be documented in accordance with Section 3.3.4 of the
AMS Center QMP.2 The results of the technical systems audit will be submitted to EPA.
Assessment reports will include the following:

•	Identification of any adverse findings or potential problems

•	Response to adverse findings or potential problems

•	Recommendations for resolving problems


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 44 of 46
Version 1.0
April 12, 2005

•	Confirmation that solutions have been implemented and are effective

•	Citation of any noteworthy practices that may be of use to others.

C2 REPORTS TO MANAGEMENT

The Battelle Quality Manager, during the course of any assessment or audit, will identify
to the technical staff performing experimental activities any immediate corrective action that
should be taken. If serious quality problems exist, the Battelle Quality Manager is authorized to
request that Battelle's AMS Center Manager issue a stop work order. Once the assessment report
has been prepared, the Verification Test Coordinator will ensure that a response is provided for
each adverse finding or potential problem and will implement any necessary follow-up corrective
action. The Battelle Quality Manager will ensure that follow-up corrective action has been taken.
The test/QA plan and final report are reviewed by EPA AMS Center QA staff and EPA AMS
Center program management staff. Upon final review and approval, both documents will then be
posted on the ETV website (www.epa.gov/etv).


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 45 of 46
Version 1.0
April 12, 2005

SECTION D
DATA VALIDATION AND USABILITY

D1 DATA REVIEW, VALIDATION, AND VERIFICATION REQUIREMENTS

The key data review requirements for the verification test are the analysis of QC samples
described in Section B5, a comparison of field data sheet comments against final data to flag any
suspect data, and a review of final data to resolve any questions about apparent outliers. The QA
audits described within Section C of this document, including the audit of data quality, are
designed to assure the quality of the data.

D2 VALIDATION AND VERIFICATION METHODS

Section C of this test/QA plan provides a description of the validation safeguards
employed for this verification test. Data validation and verification efforts include the collection
of QC samples as required in this document, and the performance of TSA and PE audits as
described in Section C.

D3 RECONCILIATION WITH USER REQUIREMENTS

This test/QA plan and the resulting ETV verification report(s) will be subjected to review
by the H2S analyzer vendors, USD A, Applied Measurement Science, EPA, and external expert
peer reviewers. These reviews will assure that this test/QA plan and the resulting report(s) meet
the needs of potential users and permitters of H2S analyzers. The final report(s) will be submitted
to EPA in Word Perfect and Adobe pdf format and subsequently posted on the ETV website.


-------
Ambient Hydrogen Sulfide Analyzers
Test/QA Plan
Page 46 of 46
Version 1.0
April 12, 2005

SECTION E
REFERENCES

1.	ASTM International. Standard Test Method for Determination of Sulfur Compounds in
Natural Gas and Gaseous Fuels by Gas Chromatography and Chemiluminescence.
Designation: D5504-01, 2001.

2.	Quality Management Plan for the ETV Advanced Monitoring Systems Center, Version 5.0,
U.S. EPA Environmental Technology Verification Program, Battelle, Columbus, Ohio,

March 2004.

3.	Environmental Technology Verification Program Quality Management Plan,
EPA/600/R-03/021, U.S. Environmental Protection Agency, Cincinnati, Ohio, December
2002.

4.	Minnesota Pollution Control Agency. "Feedlot Air Quality Summary: Data Collection,
Enforcement and Program Development."
http://www.pca.state.mn.us/hot/pubs/feedlot-aq.pdf. March 1999.

5.	Zahn, J. A., et al. "Correlation of Human Olfactory Responses to Airborne Concentrations of
Malordorous Volatile Organic Compounds Emitted from Swine Effluent," J. Environ. Qual.,
30, 624-634, 2001.

6.	National Academy of Sciences. "Air Emissions from Animal Feeding Operations: Current
Knowledge, Future Needs." National Research Council, 2003.

7.	Animal Feeding Operations Consent Agreement and Final Order, Federal Register, Volume
70(19), January 31, 2005.

8.	Restek Corporation. "Stability of Low-level (lppb - 20 ppb) Reactive Surfurs in SilcoCan™
Canisters." Applications note #59347-INT, 2001.


-------