oEPA 4

United States
Environmental Protection
Aaencv



Investigation of Gaseous
Criteria Pollutant Transport
Efficiency as a Function of
Tubing Material

Office of Research and Development

Center for Environmental Measurements and Modeling


-------
EP A/600/R-22/166
August 2022

Investigation of Gaseous Criteria Pollutant Transport
Efficiency as a Function of Tubing Material

by

Kyle Digby1, Carlton Witherspoon2, Robert Yaga2, and Diya Yang2

^SAI

2Jacobs Technology, Inc
US EPA Contract: 68HERC20D0018; EP-C-15-008
Task Order: R-20F0311

Cortina Johnson, Andrew Whitehill, Russell Long, and Robert Vanderpool
U.S. Environmental Protection Agency
Office of Research and Development
Center for Environmental Measurements and Modeling
Research Triangle Park, North Carolina 27711


-------
Disclaimer

The U.S. Environmental Protection Agency (EPA), through its Office of Research and Development
(ORD) funded the research described herein under Contract: 68HERC20D0018; EP-C-15-008, Task
Order: R-20F0311. The views expressed in this report are those of the authors and do not necessarily
reflect the views or policies of EPA. This document has been subjected to Agency review and approved
for publication. Mention of trade names or commercial products do not constitute an endorsement or
recommendation for use.

The contractor role did not include establishing Agency policy.

Questions concerning this document, or its application, should be addressed to:

Robert Vanderpool
Ambient Air Branch

Air Methods and Characterization Division
Office of Research and Development
U.S. Environmental Protection Agency

Mail Code D475

109 T.W. Alexander Drive

Research Triangle Park, NC 27711

Telephone No.: (919) 541-7877

E-mail Address: Vanderpool.Robert@epa.gov

i


-------
Table of Contents

Disclaimer	i

Table of Contents	ii

List of Tables	iii

List of Figures	iii

List of Equations	iv

Acronyms and Abbreviations 	 v

Acknowledgments	vii

Executive Summary	viii

1.0 Introduction	 1

1.1	Background	 1

1.2	Obj ective	2

2.0 Methods	 3

2.1	Experimental Approach	 3

2.1.1	Pilot Testing	 3

2.1.2	Target Analytes and Test Materials	 8

2.2	Experimental Setup	 10

2.2.1	Experiment Equipment	 10

2.2.2	Initial Sample Gas Flow Path	 13

2.2.3	Sample Gas Flow Along the Rest of the Flow Path	 14

2.2.4	Data Flow Path	 15

2.3	Final Experimental Procedure	 18

3.0 Quality Assurance (QA) and Quality Control (QC)	 21

3.1	Equipment Calibration and Verification	 21

3.2	QA/QC Checks	 21

3.3	Data Quality Objectives	22

4.0 Results and Discussion	 23

4.1	O3 Data Summary	 24

4.2	SO2 Data Summary 	 26

4.3	NO2 Data Summary	 28

4.4	CO Data Summary	 30

4.5	Stati sti cal Analy si s	 31

5.0 Summary and Conclusions	 36

6.0 References	 40

Appendix A	 42

ii


-------
List of Tables

Table 1. Pilot Tests Experiment Objectives, Descriptions, Results, and Determinations	4

Table 2. Test Materials List	9

Table 3. Equipment List	11

Table 4. Gas Cylinder List	12

Table 5. Setpoint Concentrations	18

Table 6. DQOs for Critical Measurements	22

Table 7. O3 Transport Efficiency Results	24

Table 8. SO2 Transport Efficiency Results	26

Table 9. NO2 Transport Efficiency Results	28

Table 10. CO Transport Efficiency Results	30

Table 11. Summary of Statistical Testing	31

Table 12. Data reorganization Transport Efficiency vs NAAQS Level	32

Table 13. Results Transportation Efficiencies of Tubing vs NAAQS Level testing	32

Table 14. Data reorganization tubing type vs pollutant	33

Table 15. Results from ANOVA single factor test of tubing type vs pollutant	33

Table 16. Data reorganized for ANOVA two-way testing	34

Table 17. P-values of the ANOVA two-way testing (unconditioned transport efficiencies)	34

Table 18. Results from the ANOVA two-way testing with conditioned O3 transport efficiencies	35

Table 19. Summary of Transport Efficiencies for the Four Fluoropolymer Tubing Materials	37

Table 20. Properties of Fluoropolymer Tubing Materials	38

Appendix A Table 1. Ozone (Unconditioned)	42

Appendix A Table 2. Ozone (Conditioned)	43

Appendix A Table 3. Sulfur Dioxide (Unconditioned)	44

Appendix A Table 4. Sulfur Dioxide (Conditioned)	45

Appendix A Table 5. Nitrogen Dioxide (Unconditioned)	46

Appendix A Table 6. Nitrogen Dioxide (Conditioned)	47

Appendix A Table 7. Carbon Monoxide (Unconditioned)	48

List of Figures

Figure 1 Example Graph of FEP vs Grade 316 Stainless Steel Pilot Test	6

Figure 2. Example Graph of Stainless Steel with SilcoNert® Coating vs PTFE Pilot Test	7

Figure 3. Side by Side Comparison of O3 Measurements through SS316 and SNSS Before and After

Solvent Cleaning	8

Figure 4. Diagram of Flow Path Experimental Setup	13

Figure 5. Flow Chart of the Valve Switch Timing Process During Each Experiment	14

Figure 6. Diagram of the Data Flow of the System	15

Figure 7. Example Graph of an Experiment from the Envidas Reporter Application	16

Figure 8. Photograph of Flow Path Experimental Setup	17

iii


-------
List of Equations

Equation 1. Transportation Efficiency for a Target Analyte through Test Material
Equation 2. Tubing Length Based on Residence Time of Gas through Test Material

iv


-------
Acronyms and Abbreviations

CAA

Clean Air Act

CAPS

Cavity Attenuated Phase Shift

CASTNET

Clean Air Status and Trends Network

CFR

Code of Federal Regulations

CO

Carbon Monoxide

DAS

Data Acquisition System

EPA

Environmental Protection Agency

FEM

Federal Equivalent Method

FEP

Fluorinated Ethylene Propylene

FRM

Federal Reference Method

GFC

Gas Filter Correlation

GPT

Gas Phase Titration

hv

Photon

ID

Inner Diameter

IR

Infrared

LED

Light Emitting Diode

MFC

Mass Flow Controller

NAAQS

National Ambient Air Quality Standards

NO

Nitric Oxide

no2

Nitrogen Dioxide

03

Ozone

OD

Outer Diameter

ORD

Office of Research and Development

ppb

Parts Per Billion (by volume)

ppm

Parts Per Million (by volume)

PFA

Perfluoroalkoxy Alkane

PM

Particulate Matter

PMT

Photomultiplier Tube

PTFE

Poly tetrafluoroethyl ene

PVC

Polyvinyl Chloride

PVDF

Polyvinylidene Fluoride


-------
QA

Quality Assurance

RS

Recommended Standard

S02

Sulfur Dioxide

SNSS

Grade 316 Stainless Steel tubing with Silconert® 2000 internally coated

SS316

Grade 316 Stainless Steel

SQL

Structured Query Language

TAPI

Teledyne Air Pollution Instruments

TCP/IP

Transmission Control Protocol/Internet Protocol

UV

Ultraviolet

vi


-------
Acknowledgments

The study team is grateful for the administrative and quality assurance (QA) support of Libby Nessley,
QA Manager, Air Methods and Characterization Division, U.S. EPA.

vii


-------
Executive Summary

Accurate determination of National Ambient Air Quality Standards (NAAQS) pollutant concentrations
requires efficient sampling and transport of the pollutants to their respective samplers for quantitative
analysis. For reactive gas sampling, Section 9 of Appendix E of 40 CFR Part 58 specifies that only
borosilicate glass (Pyrex®) or fluorinated ethylene propylene (FEP Teflon®) "or their equivalent" are
suitable materials for sampling and transporting to Federal Reference Method (FRM) and Federal
Equivalent Method (FEM) instruments. It also specifies that the sample residence time between the inlet
probe and the analyzer cannot exceed 20 seconds. These regulatory specifications have not been updated
since 1979. The purpose of this research effort was to identify alternative transport materials for
sampling gaseous criteria pollutants. This laboratory study determined the transport efficiency of Ozone
(O3), Nitrogen Dioxide (NO2), Sulfur Dioxide (SO2), and Carbon Monoxide (CO) through six candidate
materials at multiple pollutant concentration levels. The need to condition or passivate the tubing with
the pollutant of interest was also tested.

Based on a review of the literature, six tubing materials were selected for evaluation. These included
three fluorinated polymer materials (PTFE, PVDF, and PFA), a non-fluorinated polymer (PVC), and
two metals (316 stainless steel (SS316) and Silconert-2000® coated 316 stainless steel (SNSS)). All
tubing had a 1/4" OD and nominal 3/16" ID, and lengths of tubing were selected to provide a 20 second
residence time at the tested FRM's or FEM's design flow rate. For each pollutant and tubing type, three
replicate tests were conducted at concentrations of 20%, 50%, and 120% of the NAAQS value of
greatest regulatory relevance. The predetermined acceptance criteria were transport efficiencies of
97.5%) or greater (i.e., maximum loss of 2.5%), which was selected to identify potentially alternative
materials to borosilicate glass and FEP Teflon®. This acceptance criterion was adopted from the
through-the-probe audit program (US EPA, 2011) used for quantifying O3 line losses during field
monitoring.

Testing determined that O3 was the most reactive of the four gaseous NAAQS pollutants. O3 transport
efficiencies for the four fluoropolymer tubing types (FEP, PTFE, PVDF, and PFA) ranged from 95.6%
to 98.6%) without preconditioning. SS316, SNSS, and PVC had transport efficiencies of 0%, 9.2%, and
11.1% respectively, suggesting significant losses in the tubing without preconditioning. Following
conditioning of each tubing type with 450 ppb of O3 for 1 hour, the transport efficiency of the
fluoropolymer tubing types increased to an average of 99.4%. The high transport efficiency of O3
through fluoropolymer tubing was repeatable and independent of O3 concentration. Conditioning of
SS316 tubing and Tygon® tubing resulted in 0% transport efficiency (i.e., 100% loss) of O3 for both
tubing types. The transport efficiency of O3 through conditioned SNSS tubing improved to 87.7% with
conditioning but remained below the study's 97.5% acceptance criteria. During both the unconditioned
and conditioned tests, SS316, SNSS, and Tygon® test results were highly variable and dependent on the
O3 concentration being tested.

Unlike the tests conducted with O3, the transport efficiency of SO2, NO2, and CO through the four
fluoropolymers exceeded the 97.5% acceptance criteria for transport efficiency without conditioning.
The average transport efficiency of these three gases through the fluoropolymer tubing types was 99.8%
and independent of test concentrations. For SS316, SNSS, and Tygon® tubing, test results with these
three pollutants had higher transport efficiencies than obtained with conditioned O3 tests; although only
tests conducted with CO resulted in transport efficiencies exceeding the goal of 97.5%.

Based on this study's results, EPA will propose the revision of Section 9, Appendix E of 40 CFR Part 58

viii


-------
to include PVDF, PTFE, and PFA as approved tubing types for regulatory sampling of all four gaseous
criteria NAAQS pollutants.

ix


-------
1.0 Introduction

1.1 Background

Under the Clean Air Act (CAA), the National Ambient Air Quality Standards (NAAQS) were
established by the Environmental Protection Agency (EPA) in 1971 for six criteria air pollutants, which
are carbon monoxide (CO), nitrogen dioxide (NO2), ozone (O3), sulfur dioxide (SO2), particulate matter
(PM), and lead. To ensure accurate and consistent measurements of these pollutants, specific
measurements methods and techniques for ambient air instrumentation were defined and referred to as
Federal Reference Methods (FRM) and Federal Equivalent Methods (FEM). The Code of Federal
Regulations (CFR) Title 40 Parts 50 and 53 contain all published regulations for the implementation and
verification of these methods. In order for state and local agencies to properly monitor and report these
criteria pollutants (40 CFR §50.1), additional regulations were added to the CFR for the CAA in 1979.
The 40 CFR Part 58 currently covers all aspects of establishing, maintaining, and reporting data from an
ambient air monitoring site.

In Section 9, Appendix E of Title 40 CFR Part 58, it is specified that the appropriate materials to be used
as probes and intake sampling lines for measuring reactive gases from ambient air samples are
Fluorinated Ethylene Propylene (FEP) Teflon®, Pyrex® borosilicate glass, or "other equivalent
materials" which are not specified. Sample gases must also have a residence time within the material of
less than 20 seconds (40 CFR 58, Appendix E, Section 9, QA Handbook 2017). Several inquiries have
been received by the EPA Office of Research and Development's (ORD) Reference and Equivalent
Methods Program regarding what commercially available materials might qualify as suitable
"equivalent" materials.

Research regarding other materials which could be considered equivalent and suitable for the transport
of gaseous criteria pollutants has been limited during the past 40 years. The justification for limiting
probe materials and sampling lines needs reinvestigation. The results in the research referenced in the
CFR were not inclusive of all four gaseous criteria pollutants with each study focusing only on a subset
of those pollutants. In addition, new materials have since been identified as potential alternatives, and
materials previously tested may undergo new manufacturing processes which might affect the suitability
as a material for NAAQS pollutant sampling.

Although portions of Appendix E have been amended as recently as 2013, Section 9 has remained
largely unchanged since its publication in 1979. The citations in the Section 9 publication regarding the
use of Pyrex borosilicate glass or FEP Teflon® were from four publications: 1. Altshuller et al. (1961), 2.
Hughes (1975), 3. Wechter (1976), and 4. U.S. EPA (1971) Field Operations Guide for Automatic Air
Monitoring Equipment, No. APTD-0736, Research Triangle Park (RTP).

1.	Experiments from Altshuller et al. (1961) indicated that FEP Teflon® and glass tubing were the
best materials for O3 sampling. Stainless steel and polyethylene tubing were less suitable for O3
sampling, and polyvinyl chloride (PVC) tubing was unsuitable.

2.	Hughes (1975) provided the standard preparations and overview of the storage pollutant
reference standard gases and the proper types of cylinders to use for the storage of these gases.
Hughes also investigated Teflon® permeation tubes and other reference gases including NO2 and
SO2. The materials described in the Hughes article provided specific storage applications of the
criteria pollutants.

1


-------
3.	Wechter (1976) reviewed the storage of gases in cylinders. The article evaluated a treated
aluminum cylinder to guide the specialty gas industry to provide more stable and accurate
calibration standards. Three of the four gaseous criteria pollutants, CO, SO2 and NO2 were used
in Wechter's research. The treatments on the aluminum cylinder walls aided in maintaining the
stability of the reactive calibration gases during the long-term performance testing of the
cylinder. The treatment was described as a two-part process; the first part was to enhance the
aluminum oxide layer, and the second part was listed as "proprietary in nature".

4.	The U.S. EPA Field Operations Guide for Automatic Air Monitoring Equipment (1971) stated
specifically that only borosilicate glass and FEP Teflon® can be used for inlet materials, citing
Wohler's research from 1967. The article referenced the adsorption and desorption of glass,
plastic, and metal tubing regarding CO and SO2 (Byers and Davis, 1970) and provided a
schematic of the testing apparatus and test results.

To update this research, a laboratory investigation was conducted to evaluate whether materials other
than FEP Teflon® and Pyrex® borosilicate glass are acceptable for sampling gaseous criteria pollutants
at ambient air monitoring sites. The results will inform and guide federal, state, and local agencies on
decisions concerning acceptable materials for sampling criteria pollutants. Our study's results show that
ambient air monitoring site managers and technicians can use a wider variety of materials when
sampling gaseous criteria pollutants in order to maintain the required level of accuracy for NAAQS
pollutant compliance determination.

1.2 Objective

The purpose of this study was to determine what inlet materials could be considered to be functionally
equivalent to FEP Teflon® and Pyrex® borosilicate glass for transporting and measuring gaseous criteria
pollutants at ambient air monitoring sites.

The primary objective of this work was to compare several candidate materials to FEP Teflon® to
determine which materials can produce similar transport efficiencies when measuring gaseous criteria
pollutants through those materials. Results were obtained for the four gaseous criteria pollutants, O3,
N02, S02, and CO.

Because the reactivity of these gases differ, results were different for each of the gaseous criteria
pollutants. An overall recommendation of the best performing materials based on the research results
has been presented at the conclusion of this report.

2


-------
2.0 Methods

2.1 Experimental Approach

To determine which materials would be suitable equivalents for FEP Teflon® and Pyrex® borosilicate
glass, known concentrations of each gaseous analyte (O3, SO2, CO, and NO2) were generated by a
dilution calibrator (Teledyne Air Pollution Instruments (TAPI) T700U, San Diego, CA) and delivered to
an FRM or FEM analyzer that measures each specific analyte.

The T700U dilution calibrator contains three mass flow controllers (MFCs) (one for diluent gas and two
for target analyte gas), an O3 generator, and a photometer. The calibrator was connected to a cylinder
which contained known concentrations of SO2, CO, and NO2 gases to produce accurate concentrations.

The gas analyzers selected to measure the target analytes were FRM and FEM instruments that are
utilized at ambient air monitoring sites. These analyzers communicated with a data acquisition system
(DAS) using a data collection application Envidas (DR DAS LTD Envidas, Granville, OH). Once the
analyzer was connected to Envidas, it continuously streamed data into the DAS database. The specific
details of the T700U, the gas analyzers, DAS and its software, and other related equipment will be
described in a later section.

During each experiment, the target analyte gas was delivered through a 5-foot length of 1/4 inch OD,
3/16-inch ID FEP Teflon tubing as a control. While the flow was directed through the control material,
measurements were taken by the target analyte instrument and recorded in the DAS. This control
measurement represented the concentration of gas at the inlet of the test material.

After obtaining the control value, the flow was then delivered through the test material and
measurements were collected by the analyzers at the outlet. The results from the control material and the
test material determine the transport efficiency of the test materials.

_	, rrr- ¦	Concentration o f Tar qet Analyte At The Outlet of the Test Material . __

Transport Efficiency (%) = 			-					X 100

Concentration of Target Analyte At The Inlet of the Test Material

Equation 1. Transportation Efficiency for a Target Analyte through Test Material

Once stable, the average concentration measurements were read by the analyzer during the final 1-
minute average of data from each test and recorded into the DAS. The data from the inlet and outlet of
each test material were compared, and the transport efficiency of each test material was calculated using
Equation 1.

Transport efficiencies close to 100% represent an experiment in which the loss of concentration of the
target analyte was minimal. Appreciably lower transport efficiencies demonstrated that there was a
significant loss of concentration of the gas during its contact time with the tubing material.

2.1.1 Pilot Testing

In preparation for this study, several pilot tests were run to determine how to represent the transport
efficiency. These pilot tests were performed using O3 as the target analyte. Previous publications
indicated that O3 is the most reactive criteria pollutant of the four gases being tested (Altshuller et al.,
1961). These pilot tests determined methodology used for measuring the transport efficiency through
each candidate test material.

3


-------
A list of pilot test objectives, descriptions of each test, and determinations made from their results are
presented in Table 1.

Table 1. Pilot Tests Experiment Objectives, Descriptions, Results, and Determinations

Test

Pilot Test Objective

Description of Test

Results

Determination

1

Demonstration that the flow path setup
would operate correctly

Delivered O3 to a
T265 ozone analyzer
through control
material (FEP
Teflon®) and
subsequently through
test material
(SS316).

Expected O3
concentration
through each
material matched
results,

demonstrating that
the system was
operating
appropriately.

This flow path setup was
used for all experiments.

2

Determination of the amount of time
required to stabilize the target analyte
gas

Delivered O3 to a
T265 ozone analyzer
through PTFE and
SNSS material with
a residence time of
20 seconds for both
materials.

O3 concentrations
through PTFE
tubing took
approximately 5
minutes to stabilize,
while SNSS tubing
took 20 minutes or
more to stabilize.

Upon review of data, the
minimum test
stabilization time was
determined to be 40
minutes in length, with
the control material and
test material being
exposed to the target
analyte for 20 minutes
each.

3

Determination of whether cleaning
metal tubing with solvent affected
results

Delivered O3 to a
T265 ozone analyzer
through SS316 and
SNSS material and
measured O3
concentrations before
and after cleaning
each piece of tubing
with Methylene
Chloride followed by
an Acetone Rinse.

No change in
concentration was
observed after either
metal tubing
material was cleaned
with solvent.

Metal tubing materials
were not cleaned with
solvent prior to
experiments.

Test 1. During pilot testing, initial tests were conducted to determine the most practical experimental
setups for the research project purposes. The two solenoid valves that were connected to the ADAM
5000-series have three valve ports. The first port was a common port, which was always open. The
second port was the normally open port, which was open only when the solenoids did not have a voltage
delivered to them. The third port was the normally closed port, which was only open when the solenoids
had a voltage delivered to them to engage the diaphragm within the valve. Voltage was delivered to the
solenoids through the ADAM 5000-series system, which was controlled by the DAS.

When the normally open side of both solenoid valves were open, gas flowed through the control
material to the analyzer. When the normally closed side of both solenoid valves were open, gas flowed
through the test material to the analyzer. The normally closed side of the solenoid valves were opened
by delivering a voltage to the solenoid valves to engage the diaphragm and switch the valve flow path.
For each experiment, the two solenoid valves were set to alternatively switch between the control
material and the test material at five-minute intervals. These switches were automated by the ADAM-
5000 series modular I/O system (Advantech ADAM-5000-series Modular I/O, Irvine, CA) which
delivered a voltage to the solenoid to activate the valve in order to open the normally closed side of the

4


-------
valve and divert flow through the test material. The valves were automated to switch at regular 5-minute
intervals for the duration of the experiment, diverting flow through the control material to determine the
inlet concentration and test material to determine the outlet concentration of a material for an equal time
of 20-minutes while the target analyte was generated for a total of 40 minutes. By using a known control
material (FEP Teflon®) and confirming that the inlet and outlet concentrations were identical, the flow
path would be determined to be acceptable for use in testing.

During pilot testing, manual 3-way stainless steel ball valves (Swagelok SS-43GXS4, Solon, OH) were
turned by hand to change the flow path. These were replaced during the actual tests by automated PFA
solenoid valves (3-Way Valve 203-3414-215, SemiTorr Group, Inc, Hillsboro, OR 97124) controlled by
the ADAM-5000.

After performing an initial linearity check using gas concentrations generated by the T700U to adjust the
slope and intercept of the T265 O3 analyzer (TAPI T265, San Diego, CA), initial tests compared the O3
concentration values generated and measured by the T700U to those measured by the T265 when
connected to the test setup. A 5-foot piece of FEP Teflon® served as the control material, while a 22-foot
piece of 316 stainless steel (SS316) served as the test material. The residence time of the stainless-steel
tubing was approximately 14-15 seconds for the pilot test. Because the residence time had not been
determined when this test was conducted, aligning it within the CFR specifications was a key factor for
the test material in this experiment.

FEP Teflon® was already considered to be acceptable by the CFR, and previous research indicated that
SS316 had poor transmission of O3, especially at sub-100 ppbv concentrations (Scholz, 2015). The pilot
experiment tested concentrations of 400 ppb (parts per billion) and 200 ppb O3, which were chosen
because they fell into the range of the T265 analyzer (max 500 ppb). The T265 O3 readings through the
FEP Teflon® compared favorably to the T700U photometer measurements, while T265 O3 readings
through the SS316 showed little to no O3 being delivered to the analyzer, both of which were the
expected result based on this literature review. This performance can be seen in Figure 1.

5


-------
Site Report - Lab D288-A
Date&Time: 24/032021 09:00 09:00:00 - 13:00:00



































1

Vintrnl

Matprisl Tp«;t





est Material Test









Linearity Test













































































































































































































































































































1





























































































































































































































































































































































































































































































































































































































































































































































































J























































































































































































































































































Troubleshoot













































r 1





1







































































09:00 09:10 09:20 09:30 09:40 09:50 10:00 10:10 10:20 10:30

10:40 10:50 11:00 11:10
Date & Time

11:20 11:30 11:40 11:50 12:00 12:10 12:20 12:30 12:40 12:50 13:00

Figure 1 Example Graph of FEP vs Grade 316 Stainless Steel Pilot Test.

The graph in Figure 1 includes a linearity check, pilot test measuring O3 concentrations at the inlet and
at the outlet of the test material (SS316), and confirmation check (troubleshoot) to confirm the dilution
calibrator was still generating O3 gas as it was unknown at the time that SS316 would be rather low in
transport efficiency. Time is displayed on the x-axis, while the O3 concentration is displayed in "ppb" on
the y-axis. As mentioned previously, these results confirmed that the flow path system was functioning
as expected, and this test setup was planned for use going forward for all test material experiments.

Test 2. Tests were conducted to determine the time required for the O3 concentration to stabilize. The
control material was always planned to have a very short residence time to represent the inlet
concentration of the test material. However, the residence times for the test material, the target gas
concentrations for each experiment, and the amount of time each experiment would run had not been
determined. Therefore, tests to determine the length of time to allow O3 to stabilize during each
experiment were conducted by using a flow of 450 ppb O3 through PTFE (fluoropolymer tubing) and a
metal test material (grade-316 stainless steel with SilcoNert® 2000 (SilcoTek® Corporation, Bellefonte,
PA) non-reactive coating [SNSS]), with an initial residence time of 20 seconds for both materials. This
time was chosen as it is the maximum allowable residence time in the Section 9, Appendix E of 40 CFR
part 58. During these tests, the O3 concentrations at the analyzer stabilized after approximately
5 minutes for the PTFE tubing. The SNSS required 20 minutes or more to stabilize. An example of this
test can be seen in Figure 2.

6


-------
Site Report - Lab D2B8-A
Date&Time : 15/04/2021 07:30 07:30:00- 13:30:00

cm _

















550-
500-
450-
400-
350-

® 300 ~

ZD

>250-
200-
150-
100-
50-

























































































































































































































,	





















































































































































































0 -





























ir







U -

_*\fl



«





















30

CnJ -

07

1 1 1

30 08:00 08:30 09:00

1 1

09:30 10:00

1 1

10:30 11:00
Date &Trme

l

11:30

12

1 1

00 12:30 13:00 13

T2fi5_D3[ppb]

Figure 2. Example Graph of Stainless Steel with SilcoNert® Coating vs PTFE Pilot Test

Figure 2 displays the results from the SNSS and PTFE residence time pilot tests. Time is displayed on
the x-axis while O3 concentration is displayed in "ppb" on the y-axis. The portion of the graph from 8:25
to 10:10 was when O3 gas was being measured at the outlet of the SNSS material, while the portion of
the graph from 10:40 to 11:40 was when O3 gas was being measured at the outlet of the PTFE material.
Each material received the same two concentration steps of O3 gas during those periods of time, which
can be seen during their respective rises in concentration.

Based on these stabilization pilot tests, it was decided that the total amount of time that each experiment
would run would be 40 minutes, with each material being exposed to the target analyte for 20 minutes.
By observing the stabilization of O3 concentrations in each material, the longest time (in SS316) was
used to determine the length of the run time of the experiment.

Test 3. Lastly, the original conditions prior to testing of the test material candidates were investigated,
such as whether the stainless-steel tubing (SS316 or SNSS) should be pre-cleaned with solvent as there
could be leftover oils or other foreign material remaining from the manufacturing process. Tests were
run with stainless steel materials before and after solvent cleaning to see if any changes in O3
responsivity occurred.

Two tests were run in which each of these materials was connected as the test material both before and
after the solvent cleaning procedure. Figure 3 shows a side-by-side comparison of the graphs displaying
the results, with the materials before solvent cleaning on the left and the materials after solvent cleaning
on the right. Time is on the x-axis, while O3 concentration in "ppb" is on the y-axis. The graphs show

7


-------
that the solvent cleaning resulted in no significant change in measured O3 concentration through either
material, demonstrating that addition of solvent cleaning did not significantly affect the results. For
example, the percent change for SS316 at the highest concentration of O3, 400 ppb, was approximately
0.8%, while the percent change for SNSS at 400 ppb was -0.5%. Therefore, it was decided that no
solvent cleaning would be performed for the experimental runs with stainless steel materials.

~~1——1——1——1——1——1——1——1——1——1——1——1——1——1——1

10:00 10:15 10:30 10:45 11.-00 11:15 11:30 11:45 12:00 12:15 12:30 12:45 13:00 13:15 13:30 13:45
Date & Time













400ppbSS316 1







400 ppb SN







































jK

/



























J



























/





























/















200 ppb

SNSS

































j





























































200ppbSS316

1,1,1

1 1 1











1





1

1

1

1 1 1

1,1,1

1 1 1 1 1 1

Stainless Steel and Silconert Stainless Steel Before Cleaning

Stainless Steel and Silconert Stainless Steel After Cleaning

Figure 3. Side by Side Comparison of Os Measurements through SS316 and SNSS
Before and After Solvent Cleaning

2.1.2 Target Analytes and Test Materials

The target analytes for these experiments were four gaseous criteria pollutants from the NAAQS, O3,
S02, N02, and CO.

The test materials that were studied and compared to FEP Teflon* during these experiments were
polytetrafluoroethylene (PTFE), perfluoroalkoxy (PFA), polyvinylidene fluoride (PVDF, often referred
to by its trade name Kynar®), a flexible thermoplastic tubing polyvinyl chloride (PVC, often referred to
by its trade name Tygon® S-3, Formulation B-44-3 | Saint-Gobain Performance Plastic, Courbevoie
France]), grade 316 stainless steel (SS316), and grade 316 stainless steel with a non-reactive amorphous
silicon coating (Pac Stainless Ltd, Ftouston, TX [SNSS]). These materials, as well as the part numbers
and vendors are listed in Table 2.

8


-------
Table 2. Test Materials List

Material

Manufacturer/Supplier

OD (in)

ID (in)

Part Number

Description

FEP Teflon®
Tubing

St. Gobain/McMaster-
Carr (Atlanta, GA)

0.250

0.188

52355K57

Fluorinated ethylene
propylene tubing (control
material)

PFA Tubing

McMaster-Carr

0.250

0.188

5773K14

Perfluoroalkoxy tubing

PTFE Tubing

Zeus/McMaster-Carr

0.250

0.188

5239K12

Polytetrafluoroethylene
tubing

PVDF Tubing

McMaster-Carr

0.250

0.172

51105K22

Polyvinylidene fluoride

SS316 Tubing

Webco

Ind./McMaster-Carr

0.250

0.194

89995K83

Stainless steel tubing of
alloy 316

SNSS Tubing

Pac Stainless
(Houston, TX)

0.250

0.180

76C0250035BAO

Stainless steel tubing of
alloy 316 treated with the
Silconert® 2000 non-
reactive coating

PVC Tygon® S3
Tubing

St. Gobain/McMaster-
Carr

0.250

0.188

6516T17

Thermoplastic Polyvinyl
Chloride, Polymer Tubing

To determine the inlet concentration, FEP Teflon® was chosen for use in all experiments as the control
material tubing as seen in Figure 4. It is the most readily available CFR approved tubing, as the other
CFR approved tubing, glass, must be custom made to suit the purposes of a specific project or
monitoring site. The use of FEP Teflon® also provided flexibility in the set-up of the testing apparatus.

For polymer tubing materials, PFA and PTFE were chosen as readily available alternatives to FEP
Teflon®. While some research articles have referenced using these or similar materials for various air
sampling methods (Campos et al., 2013), no specific testing of these materials for ambient air sampling
has been conducted. Tygon® was also chosen although it had been tested in the past and was considered
unsuitable for sampling O3. PVDF was chosen as another fluoropolymer alternative used in applications,
such as filter elements for previously approved FEM instruments and tubing connecting fittings inside
many types of devices. No additional research had been performed specifically with PVDF as a material
in a sampling system to measure ambient air quality.

For metal tubing materials, 316 stainless steel with SilcoNert® 2000 non-reactive coating (SNSS) was
chosen for testing, as it is commonly used in other pollutant measurement applications in industry. In
particular, the company that produces the SilcoNert® coating, SilcoTek®, markets its products for use in
gas chromatographs, coatings for stainless steel-based gas cylinders, and other apparatuses for analytical
equipment. Additionally, the Marshik (2015) presentation at EPA's Annual Quality Assurance
Conference provided a summary of their research with guidance of materials for continuous emissions
monitoring, sampling line testing, and best practices which demonstrated that the SilcoNert® coated
stainless steel sampling probe and sampling line outperformed stainless steel, PTFE, and PFA for NO2,
hydrogen chloride, ammonia, and formaldehyde regarding storage and retention effects. SilcoNert® was
demonstrated to prevent the adsorption of gases measured from refineries in 40 CFR 60.100 Subpart Ja.
The gases of interest in that research were SO2, hydrogen sulfide, nitric oxide (NO), CO, and a general
array of volatile organic compounds. The coating was incorporated in the sampling probes, tubing,
valves, fittings, and analyzer components. Results from testing by the Shell Corporation indicated that
SilcoNert® improved sulfur detection and response compared to other stainless-steel tubing.

Because SNSS was chosen for testing, grade 316 stainless steel without the SilcoNert® coating (SS316)
was also chosen for comparison purposes. Additionally, testing has been performed using stainless steel
in the past for other purposes. For example, the results of transport efficiency tests through 316 stainless

9


-------
steel tubing conducted by Obermiller and Charlier (1968) showed no significant difference in efficiency
using "hot" and "cold" chromatography columns. Palmes et al. (2010) used stainless steel to make
screens for a NO2 personal sampler, the screens were made to be efficient, usable in the field,
lightweight, and unbreakable. Research performed Scholz (2015) also demonstrated that SS316 does not
transport O3 efficiently at low concentrations.

For each test material, the length of tubing necessary for testing was calculated by setting a target
residence time of 20 seconds. In Section 9, Appendix E of 40 CFR Part 58, this is specified as the
longest residence time acceptable for a sampling manifold. Using the inner diameter of each test
material tubing, the target residence time of 20 seconds, and the flow rate of the analyzer used to
measure gas concentrations, the length needed for each test material tubing could be calculated. The
equation for tubing length based on residence time can be seen in Equation 2.

Instrument Flow Rate * Residence Time

Tubing Length = 	^ , .					;	—	

Tubing Cross Sectional Area

Equation 2. Tubing Length Based on Residence Time of Gas through Test Material

While the inner diameter (ID) of the tubing could vary somewhat depending on the manufacturer or
material used, the tubing outer diameter (OD) for each material was Vi inch. All outlets from the T700U
dilution calibrator and all inlets for the target analyte instruments were designed for Vi inch OD tubing.
Other equipment for the flow apparatus was also designed around this Vi inch OD, including the ports
for the solenoids, T-fittings used, and other hardware used.

Additionally, in most cases, the test material was not pre-treated in any way, nor was it exposed to the
target analytes prior to testing. However, in cases where the test material in its unconditioned state had a
transport efficiency observed to be below 97.5% during the initial experiments, an additional test was
performed following conditioning of the material. This conditioning and additional testing also applied
to all test materials tested with O3.

2.2 Experimental Setup

2.2.1 Experiment Equipment

The experimental equipment used to generate, measure, and collect data for the target analytes are
shown in Table 3.

10


-------
Table 3. Equipment List

Manufacturer

Model Number

Description

Advantech

ADAM-5000 Series
Modular I/O

Modular I/O Controller for Solenoid Valves

DR DAS LTD

Envidas System

Data Acquisition System

TAPI

T700U

Dilution Calibrator/03 generator/Photometer

TAPI

T701H

Zero Air Generator

TAPI

T265

O3 Analyzer

TAPI

T500U CAPS N02

NO2 Analyzer

Thermo-Fisher
Scientific

43c

SO2 Analyzer

Thermo-Fisher
Scientific

48c

CO analyzer

SemiTorr Group,
Inc.

Valve-GalTek-Solenoid
1/4 3-WA

3-Way Solenoids

Swagelok®

Various

Stainless Steel and Teflon nuts, ferrules, and
caps

All tubing used for the flow path was secured to the instrumentation using Swagelok® compression
fittings.

The T700U was equipped with three Mass Flow Controllers (MFCs), one for diluent gas and two for
target analyte gas depending on the required flow rate, an O3 generator, and a photometer to generate
and quantify each of the target analyte gases at a known concentration.

The T700U used an O3 generator and a calibrated and certified photometer to generate and measure
accurate concentrations of O3.

The T700U used a zero-air generator (TAPI T701H, San Diego, CA) to generate "zero-air" (filtered air
with no contaminants) as its diluent gas and a gas cylinder with known concentrations of SO2 and CO
was used to generate accurate concentrations of those gases at a particular flow rate for each experiment.
For these experiments, the cylinder used contained a tri-blend of gases consisting of approximately
300 ppm CO, 10 ppm SO2, and 10 ppm NO (purchased and received from AirGas in Morrisville, NC,
see Table 4).

The T700U used the O3 generator and a gas cylinder (see Table 4) with known concentrations of NO in
a process called Gas Phase Titration (GPT) to create the NO2 gas samples. The T700U generated user
set concentrations of NO and O3 gases and mixed them in a container with a known volume and
residence time within the instrument. The O3 gas was set as the limiting reactant, and the concentration
of NO2 gas generated was directly related to the concentration of O3 gas generated and mixed during this
process.

The analyzers used to assess the target analyte concentrations were the FRM and FEM instruments used
previously at ambient air monitoring sites to measure the appropriate gas concentrations.

O3 was measured by a TAPI T265 (TAPI, San Diego, CA) analyzer. This analyzer operates through a
chemiluminescent reaction of O3 and NO, and the analyzer requires a 1% NO cylinder (see Table 4) to
be connected, as that gas reacts with the O3 in the sample gas to produce the chemiluminescent reaction.
The reaction of O3 with NO results in the production of electronically excited NO2 molecules. Each of
these excited NO2 molecules release its excess energy by emitting a photon (hv) and dropping to a lower

11


-------
energy level. The number of emitted photons is directly proportional to the O3 concentration in the
sample stream. A photomultiplier tube (PMT) detects the hv emissions from the excited NO2 during this
process.

To measure SO2 concentrations, a Thermo-Fisher Scientific 43c (Thermo-Fisher Scientific, Waltham,
MA) analyzer was used. This instrument operates on the principle that SO2 molecules absorb ultraviolet
(UV) light and become excited at one wavelength, then decay to a lower energy state emitting UV light
at a different wavelength. A PMT detects the UV light emissions from the decaying SO2 molecules
during this process.

For N02, a T500U (TAPI T500U Cavity Attenuated Phase Shift (CAPS), TAPI, San Diego, CA)
analyzer was used. This analyzer uses light from a blue UV light emitting diode (LED) centered at a
wavelength of 450 nm, a measurement cell with high reflectivity mirrors at either end, and a vacuum
photodiode detector to measure the phase shift in a regular fluctuating light signal, which is proportional
to the absorption cross sectional area.

Lastly, a Thermo-Fisher Scientific 48c (Thermo-Fisher Scientific, Waltham, MA) analyzer was used to
measure CO values. This instrument measures CO using Gas Filter Correlation (GFC). This is based on
the principle that CO absorbs infrared (IR) radiation at a wavelength of 4.6 microns. The radiation from
an IR source within the instrument is chopped by a chopper wheel and passed through a gas filter
alternating between CO and nitrogen gas. The CO gas filter produces a reference beam, while the
nitrogen gas filter produces a measurement beam, which can be absorbed by the sample CO in the cell.
In this way, the GFC system responds specifically to CO.

The gas cylinders used in operation are listed in Table 4. The gases in these cylinders were certified by
AirGas using EPA protocol to obtain accurate concentrations (US EPA, 2012).

Table 4. Gas Cylinder List

Supplier

Location

Cylinder Type

Description

AirGas

Morrisville, NC

Tri-blend gas cylinder

EPA protocol certified gases for NO,
CO, and SO2 (NO and SO2 at
approximately 10 ppm, CO at
approximately 300 ppm)

AirGas

Morrisville, NC

1% NO gas cylinder

EPA protocol certified 1% NO gas for
T265 analyzer operation

12


-------
2.2.2 Initial Sample Gas Flow Path

=>

Figure 4. Diagram of Flow Path Experimental Setup

Figure 4 displays a flow diagram of the final set-up of the sampling apparatus used for all experiments.
During each experiment, the zero-air generator T701H, was used to generate zero air diluent gas, and the
gas cylinder of the target analyte, were continuously connected to the MFCs and O3 generation, which
were contained within the T700U dilution calibrator.

The "exhaust point t-fitting" was installed along the flow path prior to the valves to ensure that the target
analyte FRM or FEM monitor did not get over-pressurized, and to ensure that the residence time was
based on the target analyte FRM or FEM monitor internal pump flow rate rather than the output from the
dilution calibrator. Flow rates were adjusted to ensure that the gas analyzer was exposed only to the
target analyte rather than to room air.

The O3 generator within the T700U, directed O3 gas through the instrument's internal photometer, which
provided a measurement by which to adjust the generator output accordingly to achieve the target O3
concentration. The O3 gas was then delivered throughout the rest of the flow path in Figure 4, where the
T265 analyzer would draw in the sample gas.

For NO2, the T700U dilution calibrator generated NO2 through GPT of NO and O3 gases. The two gases
mixed inside a chamber of a known volume and retention time in the T700U, allowing the two gases to
react to create NO2 gas with O3 being the limiting reactant. This resulted in the output generation of the
target NO2 gas concentration along with excess NO gas. The NO2 gas was then delivered throughout the
rest of the flow path in Figure 4 where the T500U analyzer would draw in the sample gas.

For SO2 and CO, a cylinder containing known concentrations of both SO2 and CO was connected to the
dilution calibrator. The MFCs within the dilution calibrator then routed appropriate flows of the diluent

13


-------
and target analyte gases to achieve the target concentration and target flow rate. These sample gases
were then delivered throughout the rest of the flow path in Figure 4, where an analyzer would draw in
the sample gas. For SO2, the analyzer was a Thermo 43c while for CO, the analyzer was a Thermo 48c.

2.2.3 Sample Gas Flow Along the Rest of the Flow Path

The T700U dilution calibrator routed the target analyte gases into the common inlet of the first 3-way
solenoid valve (SemiTorr Group, Inc., Valve-Galtek-Solenoid 1/4 3-WA, Tualatin, OR) in the setup
through FEP Teflon® tubing, or control material (valve 1 in Figure 4). The first two solenoid valves have
three valve ports. The first port was a common port, which was always open. The second port was the
normally open port, which was open only when the solenoids did not have a voltage delivered to them.
The third port was the normally closed port, which was only open when the solenoids had a voltage
delivered to them to engage the diaphragm within the valve. Voltage was delivered to the solenoids
through the ADAM 5000 series modular I/O system, which was connected to the DAS. The solenoid
valves were automatically switched every 5-minutes by sequences programmed into the DAS during
each experiment.

The normally open outlet of the 3-way solenoid valve was connected to the control material, while the
normally closed outlet was connected to the test material. Both materials were connected to a second 3-
way solenoid valve (valve 2 in Figure 4), where the control material was connected to the normally open
inlet and the test material was connected to the normally closed inlet. The common outlet of this 3-way
solenoid valve was then connected to the appropriate analyzer chosen per test material.

From this setup, the target analyte instruments pulled in the sample gas using their internal pumps, while
any excess sample gas flow was directed into the exhaust for the laboratory.

The timing of switching the valves was based on the results of the gas stabilization pilot test performed
before these experiments were conducted.

Based on the previously described stabilization pilot tests, it was decided that the total experiment run
length was 40 minutes, with each material being exposed to the target analyte for 20 minutes. A visual
representation of the timing process used for each transport efficiency experiment is presented in
Figure 5.

Data used in calculations (1 minute)

Beginning of test Initial entry of target analyte

5-Minutes through
Test Material



5-Minutes through
Test Material

/j\



\

i

)

20 Minutes
Zero Air to Purge Both
Materials

5- Minutes through
Control Material

Sequence Re[
Total 30 minutes, 15

seated 3 times
minutes per material

5- Minutes through
Control Material

Last 10 minute

\

Data used in calculat
s of experiment

n/

ons (1 minute)

Figure 5. Flow Chart of the Valve Switch Timing Process During Each Experiment

The flow chart in Figure 5 displays the full sequence of events as they occurred during each transport
efficiency experiment. For the initial 20 minutes, the control material and test material were purged with
zero air for 10 minutes each. Following that, while the dilution calibrator generated the target analyte at
the specified concentration for 40 minutes, the solenoid valves switched between the control material
and test material every 5 minutes. The data used in the calculations for transport efficiency were the
final 1-minute averages taken when the target analyte was flowing through the control material and the

14


-------
test material, shown in the colored portions the last 10 minutes of the experiment.
2.2.4 Data Flow Path

Dilution
Calibrator

(T700U)

ADAM 5000-Series

Data
Flow

Data
Flow



Data
Flow

Data Acquisition
System
(Envidas)

Target Analyte
Instrument

Figure 6. Diagram of the Data Flow of the System

Figure 6 shows the flow of data for the Envidas system and the associated equipment. The Envidas DAS
was used to record the data from the target analyte analyzers (different for each analyte) and to control
the solenoid valves. Each of the analyzers were connected to the DAS through either Transmission
Control Protocol/Internet Protocol (TCP/IP) or Recommended Standard 232 (RS-232) serial cable
connection. The analyzers continuously streamed their data to the DAS, which recorded their data into a
Structured Query Language (SQL) database. The DAS also had an application that allowed users to
view the data as it was streamed into the system in real time via the Envidas Viewer application. The
Envidas Reporter application allowed users to generate graphs and tables of data that had already been
recorded into the database.

An example of a graph recorded by the DAS and displayed by the Envidas Reporter application is
shown in Figure 7.

15


-------
Site Report - Lab D2SS-A
Date&Time: 16/09/2021 13:4913:49:00- 14:49:00

90 -i

85-
30-
75-
70-
65-
60-
55-
50-
45-
40-
35-
30-
25-
20-
15-
10-
5-
0 -































V1"-—









n

L	

























































....













































































































































































































































































































































































































































-R -

























13:50 13:55 14:00 14:05 14:10 14:15 14:20 14:25 14:30 14:35 14:40 14:45

Date & Time

T265_03[ppb] 	 Phctometer_03[ppb]

Figure 7. Example Graph of an Experiment from the Envidas Reporter Application

Figure 7 is an example of the graphical representation of the data recorded during an O3 experiment, as
both the T700U photometer readings of the generator output and the T265 O3 analyzer readings are
displayed on the graph. Time is displayed on the x-axis, while O3 concentration in "ppb" is displayed on
the y-axis. The orange trace on the graph is the data recorded by the T265 during the experiment, while
the red trace on the graph is the data readings of the photometer in the T700U dilution calibrator, which
represents the "actual measured output" of O3 from the T700U by the O3 generator.

The Envidas system recorded analyzer data at three different time resolutions. For the analyzers T265,
43c, and 48c, the time resolutions were 5 seconds (lowest resolution for these instruments) and 1 minute,
while for the T500U analyzer, the time resolutions were 15 seconds (lowest resolution for this
instrument) and 1 minute.

16


-------
Figure 8. Photograph of Flow Path Experimental Setup

An additional piece of equipment that was connected to the Envidas DAS was an ADAM 5000-series
modular I/O system. This equipment allowed the DAS to control the solenoid valves, as the DAS could
give the ADAM 5000-series system a command to deliver a voltage to the solenoids to engage the valve
diaphragm and divert the flow path. The ADAM 5000-series system was connected to the DAS through
a RS-232 cable, and the solenoids had an analog connection to the ADAM 5000-series system.

17


-------
Figure 8 is a photograph of the real-time flow path setup used for all experiments. The valve on the right
side labeled Solenoid 1 was connected to the T700U dilution calibrator through its common port. The
tubing from the common port in that solenoid valve can be seen leading off the right side of the
photograph, which is where the T700U is located. Solenoid 1 was switched off to direct flow through
the normally open port when delivering gas through the control material, while it was switched on to
deliver flow through the normally closed port when delivering gas through the test material. Regardless
of which material had gas flowing through it, the gas would arrive at the next valve on the left side of
the setup labeled Solenoid 2. The common port of this solenoid was connected to the respective
analyzers, which were located behind the sampling system apparatus and obscured by the teal bar on the
laboratory bench. For this solenoid, when gas was arriving from the control material, the solenoid was
switched off to direct flow through the normally open port, and when gas was arriving from the test
material, the solenoid was switched on to direct flow through the normally closed port.

2.3 Final Experimental Procedure

The pilot tests provided insights into the components of the testing apparatus and each materials' ability
to passivate large concentrations of O3. However, these large concentrations were based upon the range
of the O3 analyzer. Following discussions with project stakeholders, a decision was made to base the
experimental test concentrations upon the NAAQS concentration values. The three target concentrations
were 20%, 50% and 120% of the current NAAQS concentrations, as these targets would provide a range
likely to be encountered in the field.

During each experiment and following QA/QC checks, the first step was to deliver zero air to the
analyzer and through the control and test materials for 20 minutes (10 minutes in each material). The
purpose of this was two-fold. It would ensure that the analyzer was appropriately reading zero before an
experiment, and it would purge any potential outside contaminants that the flow apparatus may have
prior to an experiment. Following this, the pre-determined concentrations of each target analyte were
generated by the dilution calibrator, T700U, for a pre-set amount of time. These gas concentrations are
presented in Table 5.

Table 5. Setpoint Concentrations

Setpoint

O3 Cone.

NO2 Cone.

SO2 Cone.

CO Cone.

Zero Air

Oppb

Oppb

Oppb

Oppb

20% of the NAAQS

15 ppb

20 ppb

15 ppb

1,800 ppb

50% of the NAAQS

35 ppb

50 ppb

40 ppb

4,500 ppb

120% of the NAAQS

85 ppb

120 ppb

90 ppb

10,800 ppb

While the dilution calibrator generated the target analyte gas for each experiment, the two 3-way
solenoid valves were set to switch between the control material and test material at regular 5-minute
intervals. This process was automated by an ADAM 5000-series modular I/O system that would deliver
a voltage to the solenoid to engage the diaphragm and change the valve flow path to the test material,
when necessary, while the diaphragm would not be engaged to have the flow path go through the control
material. Through this process, the target analyte was delivered to the appropriate analyzer through the
control material and the test material for an equal amount of time, which was 20 minutes total for each
material. Once the target analyte was generated for the pre-determined amount of total time of
40 minutes and the analyzer data for the experiment were collected by the DAS, the system was set to
run zero air through both materials before the experiment was over. This process was repeated in
triplicate for all materials, for all target analyte concentrations, and for different material conditions

18


-------
when necessary.

Each new experiment in the triplicate runs required a new piece of test material tubing so that every
experiment had an unused, unconditioned piece of test material. Before the experiment was conducted,
each test material used for any experiment was labeled with the appropriate information, such as target
analyte, concentration, and other relevant information. Once the experiment was completed, the material
was uninstalled from the experiment setup, capped by placing a stainless-steel Swagelok fitting and a
stainless-steel Swagelok cap on both open ends of the tubing, and stored in the laboratory.

Test materials that had poor transport efficiency results during the initial experiments with new, unused
tubing were included in a follow-up test in which the test material tubing sections were conditioned to
the target analyte before transport efficiency experiments were conducted. For the purposes of these
experiments, a test material transport efficiency was classified as unacceptable if it was calculated to be
below a threshold of approximately 97.5%. However, O3 experiments were an exception, as all materials
had experiments in which the material was used in the conditioning process for that target analyte.

For the conditioning tests, a new section of the test material was conditioned using the target analyte and
was then installed into the testing setup. After the two solenoid valves were engaged so that the flow
was delivered to the analyzer through the test material, the T700U dilution calibrator generated the
target analyte gas at a high concentration, and the installed test material was exposed to this high
concentration for at least an hour. The conditioning concentration was set to 450 ppb. The timing and
the concentrations were chosen based on the Field Standard Operating Procedures (SOP) for the EPA,
Through-the-Probe National Performance Audit Program (NPAP), (EPA, 2011). In that document, in
Section 3.1.7.1 titled "Quarterly Ozone Line Loss Set-up", it states that the last step in the preparation
for the O3 line loss test is to "Allow the system to condition with approximately 0.450 ppm of ozone for
one hour".

Following this conditioning process, testing commenced on this section of test material tubing in a
similar manner as previously described. However, there was one primary difference between the original
material procedure and the conditioned material procedure. Instead of using a new piece of tubing for
each concentration, all three concentration levels were delivered through one piece of conditioned
tubing. Then, once all three concentrations had been completed, a new piece of tubing was conditioned,
and the process was repeated in triplicate.

For all experiments, all data from the analyzers were continuously logged into a DAS, which had
proprietary Envidas software installed that collected all gas analyzer data. Data being collected were
received into the SQL database at a 5 second time resolution for the T265, and 43c, and the 48c, while
data were received into the SQL database at a 15 second time resolution for the T500U. This difference
was because the T500U could not transmit data to the SQL database at a time resolution lower than
15 seconds.

Experimental data were visually inspected in the software's viewer application by the experimenter
present while tests were being run. As the data were streamed from the analyzers and recorded in the
DAS, the DAS stored these data in an internal SQL database. This database was accessed and visually
inspected through the software's reporter application. This application was also used to export the data
into Excel format and stored on the DAS. For the T265, and 43c, and the 48c, 5-second and 1-minute
data were referenced and exported, while for the T500U, 15-second and 1-minute data were referenced
and exported.

Once exported to Excel, the data were further reviewed by the experimenter and other members of the

19


-------
research team. Following that review, Excel was used to process the data, where transport efficiencies
were calculated based on the final minute average concentrations read by the analyzer when the gas
passed through the control material and when the gas passed through the test material.

20


-------
3.0	Quality Assurance (QA) and Quality Control

(QC)

To maintain quality assurance through quality control (QA/QC), this project was conducted under the
approved category B, Applied Research Quality Assurance Project Plan (QAPP), "Reference and
Equivalent Methods Program Support, QAPP-1J16-009.3" (QA Track J-AMCD-0031471-QP-1-2).

3.1	Equipment Calibration and Verification

The T700U dilution calibrator was purchased from TAPI in April of 2021, and at that time, the vendor
provided the certificates for the instrument's MFCs and photometer. A year after that date, the
ORD/RTP Metrology Laboratory (Met Lab) performed the calibration and recertification of the
photometer and MFCs within the T700U. The Met Lab also performed the verification and
recertification of the DryCal flow meter used to measure flow rates. The frequency of these calibrations
occurred at a minimum of annually or a year after the current certificate was performed.

The photometer that measures the O3 concentration output of the instrument must be checked against a
standard reference photometer, and the MFCs must be checked against a NIST-Traceable flow meter.
The records of these calibrations were stored with the instrument in the laboratory, and the dates of these
calibrations were recorded in the project laboratory notebook. Scanned digital copies of these records
were stored on the Microsoft Teams Site.

3.2	QA/QC Checks

Each target analyte analyzer (e.g., T265 O3 analyzer, Thermo 43c SO2 analyzer, T500U CAPS NO2
analyzer, and Thermo 48c CO analyzer) was required to have an intercept and slope adjustment QC
check prior to each one of several regular intervals, depending on the instruments' current condition.
These intervals included: an annual basis, when a new gas cylinder had been installed for the T700U
dilution calibrator, or when the MFCs and photometer within T700U dilution calibrator had been
calibrated by the Met Lab.

QC checks were also used to verify instrument operation if the experimenter noted unexpected results
during testing. Unexpected results could include the FRM instrument measuring high concentrations
when delivering zero-air to the instrument or the FRM instrument reading more than +/- 5% away from
the expected value of a reference measurement. These checks were used by the experimenter to
troubleshoot the instrument if the problems were observed during an experiment and any observations
were recorded in the project notebook.

The QC verification check followed the 40 CFR 53.21(b) guidelines. These guidelines stated that a
linear curve would be used to compare the target analyte instrument readings to the dilution calibrator
output by plotting at least seven approximately equally spaced points, which included a setpoint at zero
and a setpoint at 90 +/- 5 percent of the upper range limit of each analyzer. For the purposes of this
project, the setpoints were set at 0%, 15%, 30%, 45%, 60%, 75%, and 90% of the upper range limit of
each test analyzer. If the slope in this linear curve calculation was more than +/- 5% off a slope of 1.0,
the slope was adjusted on the instrument accordingly. If the intercept was over +/- 0.5 ppb off zero, the
intercept was adjusted on the instrument accordingly. These adjustments were made at the intervals
stated above and all QC checks and adjustments made were documented in the laboratory notebook used
for the research project, and the results were uploaded to the Teams site.

21


-------
3.3 Data Quality Objectives

The precision and accuracy goals for each critical measurement parameter were set based on the
requirements needed to achieve the data quality objectives (DQO) of the experiment, the knowledge of
and experience with sampling systems similar to the one being used in this experiment, and other similar
research study parameters.

The DQOs for each critical measurement parameter are listed in Table 6. If these were not met during
any experiment, the settings for the critical measurement (e.g., the instrument slope or the instrument
intercept) were adjusted using the QC verification check guidelines or the measurements were not
reported as final data. It should be noted that the DQOs were met for the T700U photometer and MFCs
during these experiments and no adjustments were performed until their yearly recertification by the Met
Lab.

Instances of a target analyte instrument not meeting the DQOs were observed during the QC verification
checks, and any changes made were documented in the appropriate laboratory notebook. Data regarding
these QC verification checks were uploaded to the Teams site. These data were reviewed by the project
leads for both Jacobs and EPA.

Table 6. DQOs for Critical Measurements

Critical Measurement

Measurement Device

Accuracy/Precision

Ozone Concentration

T700U Photometer

+/-1.1% from reading on Standard Reference
Photometer; background zero of 1.0 ppb or less

Other Analyte
Concentration

T700U Mass Flow
Controllers

+/- 0.25% from calibrated reference flow meter

Instrument Slope

Target analyte instruments

+/- 5% from a slope of 1.00 during the QC Check

Instrument Intercept

Target analyte instruments

+/-1.0 ppb from 0.0 ppb during the QC Check

22


-------
4.0 Results and Discussion

Each candidate test material was compared to the selected control material, CFR approved FEP Teflon®,
by comparing target analyte concentrations when routing identified target gas analytes through both the
control material and the test material.

All data summaries are separated out into sections based on the target analyte of each experiment.
Transport Efficiency was calculated using Equation 1 in Section 2.1, in which the final 1-minute average
of the 5 second or 15 second data records of each analyzer concentration measurement through the test
material (outlet) was divided by the final 1-minute average of the 5 second or 15 second data records of
each analyzer concentration measurement through the control material (inlet) for each experiment.

In the final data reported, the mean of all related experiments was used. "Adequate" is noted when
transport efficiency results were reported as a mean efficiency of 97.5% and above. This value was
based on the 2.5% threshold for O3 line loss tests as mentioned in Section 2.3 of this report (field SOP
NPAP, EPA, 2011).

The transport efficiencies and standard deviations for each gas type and replicate are listed in
Appendix A of this report.

For each of the following sections, per target analyte, each of the data tables has the same format:

•	The first column labeled "Tubing" represents the material sampled for the associated tests in the
three columns to the right.

•	The next column, labeled "Cone, (ppb)" represents the specific target analyte concentration at
which the transport efficiency tests were conducted.

•	Transport efficiency results are divided into two categories and are reported in the next two
columns. One category/column is labeled "Unconditioned", in which the test material was not
prepared in any way before testing, while the other category/column is "Conditioned", in which
the test material was exposed to a high concentration of the target analyte for an hour prior to
recording data.

•	The columns labeled "n", represent the number of replicate tests conducted at each concentration
for each category.

•	Lastly, the "Mean Transport Eff (%)" columns represent the average of the "n" number of tests
at each concentration for each category.

23


-------
4.1 O3 Data Summary
Table 7. O3 Transport Efficiency Results

Tubing

03

Cone,
(ppb)

Unconditioned

Conditioned

n

Mean Transport
Eff. (%)

n

Mean Transport
Eff. (%)



15

3

100.1

3

99.8

FEP Teflon®

35

3

98.0

3

99.8



85

3

97.6

3

99.2







98.6



99.6















15

3

94.9

3

100.5

PVDF

35

3

95.6

3

99.2



85

3

96.2

3

98.8







95.6



99.5















15

3

95.5

3

100.2

PTFE

35

3

97.3

3

98.6



85

3

98.2

3

99.7







97.0



99.5















15

3

96.5

3

98.6

PFA

35

3

97.2

3

99.0



85

3

96.8

3

99.5







96.8



99.0















15

3

-0.3

1

-0.2

SS316

35

3

-0.3

1

-0.4



85

3

-0.2

1

-0.1







-0.3



-0.2















15

3

-0.1

1

79.7

SNSS

35

3

0.2

1

89.6



85

3

27.6

1

93.8







9.2



87.7















15

3

4.1

3

0.9

Tygon®

35

3

9.0

3

0.1



85

3

20.2

3

0.1







11.1



0.4

24


-------
In their original conditions, all the polymer materials excluding FEP Teflon® did not have an adequate
mean transport efficiency of above 97.5%. They were relatively close to the 97.5% threshold, excluding
Tygon®, but none of them met the adequate threshold. Additionally, neither the SS316 nor the SNSS
compared favorably to the FEP control material, with very low transport efficiencies of -0.3% and 9.2%,
respectively. The negative transport efficiency values seen in Table 7 were due to the analyzer reading
values close to the intercept for the instrument, and in these cases, readings would be very close to zero,
with most averages being approximately -0.1 to -0.3 ppb.

When polymer materials were conditioned, all excluding Tygon® improved in their mean transport
efficiency results. Notably, this improvement meant that all the materials went from below the 97.5%
transport efficiency threshold to well above it. This demonstrated that any of these fluoropolymer
materials should be acceptable following their conditioned with O3 prior to use. However, for Tygon®,
conditioning the material to O3 worsened its transport efficiency.

For SS316, conditioning did not improve results, with its mean transport efficiency remaining around
0%. While the mean transport efficiency results for the SNSS greatly improved, the mean transport
efficiency did not reach the threshold of 97.5% to be considered acceptable.

25


-------
4.2 S02 Data Summary
Table 8. SO2 Transport Efficiency Results

Tubing

S02
Cone,
(ppb)

Unconditioned

Conditioned

n

Mean Transport
Eff. (%)

n

Mean Transport
Eff. (%)



15

3

102.0

3

99.5

FEP Teflon®

40

3

99.7

3

99.8



90

3

100.1

3

100.2







100.6



99.8















15

3

100.4

-

-

PVDF

40

3

99.6

-

-



90

3

100.1

-

-







100.0



-















15

3

100.8

-

-

PTFE

40

3

100.3

-

-



90

3

100.2

-

-







100.4



-















15

3

100.6

-

-

PFA

40

3

100.4

-

-



90

3

100.6

-

-







100.5



-















15

3

2.9

3

71.7

SS316

40

3

3.0

3

72.9



90

3

3.2

3

86.4







3.0



77.0















15

3

100.0

-

-

SNSS

40

3

100.0

-

-



90

3

100.2

-

-







100.1



-















15

3

88.4

3

87.7

Tygon®

40

3

88.6

3

87.8



90

3

89.6

3

87.6







88.9



87.7

It should be noted that for the experiments related to the target analyte SO2, conditioned tests were not

26


-------
conducted for certain test materials due to the high transport efficiencies observed in the tubing's
unconditioned state.

In their original conditions, all the polymer materials excluding Tygon® resulted in adequate mean
transport efficiencies of above 97.5%. Specifically, Tygon® had a mean transport efficiency of 88.9%,
which did not meet the 97.5% threshold. The SS316 did not compare favorably to the control material,
with very low transport efficiencies of 3.0%. However, the SNSS performed very well with a mean
transport efficiency of 100.1%.

The only materials that were tested after being conditioned with SO2 were FEP Teflon®, SS316, and
Tygon®, because the other test materials already had mean transport efficiencies above 97.5%. The FEP
Teflon® was conditioned and tested as a control condition and performed very well with a mean
transport efficiency of 99.8%. Conditioning SS316 greatly improved its mean transport efficiency,
raising it to 77.0%>, but this result still did not meet the 97.5% acceptance criterion for acceptability.
Conditioning had little to no effect on Tygon®, as the mean transport efficiency after conditioning was
87.7%), which was 1.2% less than the previous result.

27


-------
4.3 N02 Data Summary
Table 9. NO2 Transport Efficiency Results

Tubing

NOz
Cone,
(ppb)

Unconditioned

Conditioned

n

Mean Transport
Eff. (%)

n

Mean Transport
Eff. (%)



20

4

96.6

3

99.8

FEP Teflon®

50

3

99.7

-

-



120

3

99.8

-

-







98.7



99.8















20

4

99.5

-

-

PVDF

50

3

99.5

-

-



120

3

99.2

-

-







99.4



-















20

4

99.5

-

-

PTFE

50

3

99.1

-

-



120

3

99.0

-

-







99.2



-















20

4

99.6

-

-

PFA

50

3

99.0

-

-



120

3

98.9

-

-







99.1



-















20

4

-1.0

3

40.1

SS316

50

3

-0.7

-

-



120

3

-0.2

-

-







-0.7



40.1















20

3

94.0

3

97.8

SNSS

50

3

95.4

3

98.6



120

3

97.3

3

99.0







95.6



98.5















20

3

79.9

3

85.1

Tygon®

50

3

79.1

-

-



120

3

80.8

-

-







79.9



85.1

It should be noted that for the experiments related to the target analyte NO2, conditioned tests were not

28


-------
conducted for certain test materials due to the high transport efficiencies observed in the tubing's
unconditioned state.

In their original conditions, all the polymer materials excluding Tygon® resulted in adequate mean
transport efficiencies of above 97.5%. Specifically, Tygon® had a mean transport efficiency of 79.9%,
which did not meet the 97.5% threshold. The SS316 did not compare favorably to the control material,
with very low transport efficiencies of 3.0%. The SNSS had a mean transport efficiency of 95.6%, close
but still below the 97.5% threshold to be considered acceptable.

The only materials that were tested after being conditioned were FEP Teflon®, SS316, Tygon®, and
SNSS. Additionally, the tests for FEP Teflon®, SS316, and Tygon® were only conducted at 20 ppb NO2.
The SNSS was tested at all concentrations since its original condition mean transport efficiency was
close to the 97.5% threshold. The FEP Teflon® was conditioned and tested as a control condition, and it
performed very well with a mean transport efficiency of 99.8%. Conditioning SS316 noticeably
improved its mean transport efficiency to 40.1%, but this efficiency still did not meet the threshold of
97.5% to be considered acceptable. Conditioning had little to no effect on Tygon®, as the mean transport
efficiency after conditioning was 85.1%>, only 5.2% better than the previous result. Lastly, conditioning
improved the mean transport efficiency of the SNSS, which had a mean transport efficiency of 98.5%,
which surpassed the 97.5% threshold to be considered acceptable.

29


-------
4.4 CO Data Summary
Table 10. CO Transport Efficiency Results

Tubing

CO
Cone,
(ppm)

Unconditioned

Conditioned

n

Mean Transport
Eff. (%)

n

Mean Transport
Eff. (%)



1.8

3

100.0

-

-

FEP Teflon®

4.5

3

100.3

-

-



10.8

3

100.1

-

-







100.1



-















1.8

3

100.6

-

-

PVDF

4.5

3

100.5

-

-



10.8

3

100.3

-

-







100.5



-















1.8

3

99.0

-

-

PTFE

4.5

3

99.7

-

-



10.8

3

100.2

-

-







99.6



-















1.8

3

100.9

-

-

PFA

4.5

3

100.3

-

-



10.8

3

99.6

-

-







100.3



-















1.8

3

98.7

-

-

SS316

4.5

3

100.3

-

-



10.8

3

99.6

-

-







99.5



-















1.8

3

100.5

-

-

SNSS

4.5

3

99.9

-

-



10.8

3

99.8

-

-







100.1



-















1.8

3

100.9

-

-

Tygon®

4.5

3

100.3

-

-



10.8

3

100.1

-

-







100.4



-

It should be noted that for the experiments related to the target analyte CO, no conditioned tests were

30


-------
conducted, therefore no data results were collected.

In their original conditions, all the materials resulted in adequate mean transport efficiencies of above
97.5%. In fact, all materials had between 99% and 100% mean transport efficiency, confirming that CO
transport efficiency was acceptable regardless of the material that it traveled through. None of the
materials were conditioned and retested because all materials displayed acceptable transport efficiencies
during the initial experiments.

4.5 Statistical Analysis

Transport efficiencies were compared using Excel's Data Analysis tool pack for ease of communication
and transparency. Three statistical tests, Analysis of variance (ANOVA) Single Factor, ANOVA Single
Factor with Ad hoc testing and ANOVA Two Factor with replicates, were performed upon multiple
arrangements of the transport efficiencies, replicates, and gas concentrations to calculate the statistical
evidence that the results are repeatable. The single factor ANOVA was chosen to test the means between
two or more groups of values. The two factor ANOVA with replicates analysis was chosen to evaluate
the difference between the means of more than two groups and to determine the interaction between
both the replicates and groups. Table 11 provides a brief overview of the statistical questions, the tests
performed, if ad hoc analysis was performed, and the name of the respective data table.

Table 11. Summary of Statistical Testing

Question

Test performed

Ad hoc Analysis
Performed

Table Name and
Document Reference

Does concentration (20%, 50% and
120% of NAAQS value) change the
transport efficiency for a certain gas
vs the tubing type?

ANOVA-Single
Factor

Not performed

Transport Efficiency vs
NAAQS Level testing
(Table 13)

Does Tubing Material Matter vs
Particular Gas?

ANOVA-Single
Factor

Performed,
Tukey

ANOVA Single Factor,
with ad hoc Results -
Tubing vs Gas type,
unconditioned tubing
Table 15)

Are there any statistical differences
between the groups of data?

ANOVA - Two Factor
with Replicates

Not performed

ANOVA - Two Way with
Replication Results for
unconditioned tubing (p-
value)

Table 17)

Will conditioning the fluoropolymer
tubing with Ozone increase the
transport efficiency to match the
transport efficiencies of the
remaining three gases?

ANOVA - Two Factor
with Replicates

Not performed

ANOVA - Two Way with
Replication Results (p-
values) for

unconditioned tubing
(S02, N02 and CO) vs
conditioned Tubing (03)
(Table 18)

Table 11 provides a brief overview of the three statistical testing methods used to support the overall
observations provided in the aforementioned data summaries.

To review the statistical testing methods, Excel bases the null hypothesis of the single factor ANOVA

31


-------
test on whether the mean is the same for all groups or the columns of data. Calculated p-values are
compared to the requested alpha value, which is 0.05 for all ANOVA testing in this report. If a p-Value
is greater than 0.05, we cannot conclude that a significant difference exists. The Tukey statistical ad hoc
analysis uses a Critical Value of Studentized Range Distribution or Q-value combined with the number
of treatments, replicates number and calculated standard error values to determine the statistical
differences of the data means.

Table 12 provides an example of how the transport efficiencies were reorganized from Table 7 to
execute the ANOVA Single Factor analysis. The analysis was performed to determine if the
concentration value changed transport efficiencies vs tubing type.

Table 12. Data reorganization Transport Efficiency vs NAAQS Level

Data corresponding to question 1 in Table 11 for Ozone



15ppb Ozone

35ppb Ozone

85ppb Ozone

FEP

100.05%

98.01%

97.64%

PTFE

95.52%

97.31%

98.15%

PFA

96.53%

97.20%

96.80%

PVDF

94.92%

95.62%

96.15%

Tygon®

4.14%

9.01%

20.18%

SS316

-0.07%

-0.32%

-0.14%

SNSS

-0.08%

0.17%

27.56%

Table 12 displays the transport efficiency data arrangement to compare the three NAAQS concentration
values against the type of tubing. Transport efficiency data for unconditioned data can be found in
Appendix A Table 1, Appendix A Table 3, Appendix A Table 5 and Appendix A Table 7.

Table 13. Results Transportation Efficiencies of Tubing vs NAAQS Level testing

Transport Efficiency of Tubing

vs NAAQS Level testing

Gas

p-value

Ozone

0.96412

Sulfur Dioxide

0.99983

Nitrogen Dioxide

0.99951

Carbon Monoxide

0.78306

Table 13 displays the p-values for single factor ANOVA test comparing the mean transport efficiencies
of the different concentrations for the four NAAQS gases. The p-values for the four gases are greater
than the alpha level of 0.05, indicating that there aren't statistical differences. In other words, the
transport efficiencies are not dependent on the tested NAAQS concentrations.

To provide insight into the transport efficiency interaction between each particular gas and tubing
material, a single factor ANOVA test with a Tukey ad hoc analysis was performed. Table 14 is an
example of the data reorganization of the data sets for the unconditioned mean transport efficiencies.

32


-------
Table 14. Data reorganization tubing type vs pollutant

NAAQS
level

Ozone -FEP

Sulfur Dioxide -
FEP

Nitrogen
Dioxide - FEP

Carbon

Monoxide - FEP

20%

100.05%

102.00%

99.61%

100.01%

50%

98.01%

99.66%

99.72%

100.27%

120%

97.64%

100.10%

99.79%

100.05%

Table 15. Results from ANOVA single factor test of tubing type vs pollutant

ANOVA Single Factor, with ad hoc Results - Tubing vs Gas type, unconditioned tubing



P-values of

Statistical Difference of Transport Efficiencies Between Gases



Statistical











Testing











Tubing vs Gas









Tubing

Type

03

S02

N02

CO

FEP

0.112357

Not Different

Not Different

Not Different

Not Different





Different from S02,







PTFE

1.02E-06

N02 and CO

Not Different

Not Different

Not Different





Different from S02,







PFA

3.19E-03

N02 and CO

Not Different

Not Different

Not Different





Different from S02,







PVDF

1.02E-06

N02 and CO

Not Different

Different from CO

Different from N02





Different from S02,



Different from CO

Different from S02,

Tygon®

6.97E-17

N02 and CO

Different from CO

and 03

N02 and 03





Different from S02

Different from CO

Different from S02

Different from S02,

SS316

1.41E-06

and CO

and 03

and CO

N02 and 03





Different from S02,







SNSS

1.90E-08

N02 and CO

Not Different

Not Different

Not Different

Table 15 displays the p-values from the ANOVA single factor test comparing the individual tubing
versus the four target analyte gases. The p-values listed in Table 15 indicate that the transport
efficiencies are statistically different across all four gases for the tubing materials tested besides FEP.
Results of ad hoc analysis via Tukey analysis indicate that O3 is the most reactive gas and transport
efficiencies statistically differ from the majority of other gases. The remaining gases, SO2, NO2 and CO
were generally not as reactive with the fluoropolymer tubing; except for PVDF. As previously
mentioned, transport efficiencies for Tygon® and Stainless Steel only reached the 97.5% threshold for
CO. The ad hoc analysis statistically confirmed the transport efficiency observations reported.

To provide insight into whether there was an interaction between the concentration groups and the
replicates, an ANOVA two-way statistical test with replicates was performed. Briefly, the two-way
ANOVA tests in excel are based upon evaluating three null hypotheses. The three hypotheses of concern
are as follows.

Hypothesis 1: the means of observations grouped by one factor are the same (i.e., no difference between
in the 20%, 50% and 120% groups [each group is the testing replicates])

Hypothesis 2: the means of observations grouped by the other factor are the same (i.e., no difference
between the gases)

33


-------
Hypothesis 3: there is no interaction between the two factors (i.e., no difference between the 20%, 50%
and 120% vs the gases).

Table 16 displays how the data was reorganized for the ANOVA two-way testing, while Table 17
provides the p-values.

Table 16. Data reorganized for ANOVA two-way testing





Sulfur

Nitrogen

Carbon





Dioxide -

Dioxide -

Monoxide -

NAAQS

Ozone -FEP

FEP

FEP

FEP

20%

98.31%

103.97%

99.75%

99.48%

20%

100.29%

102.80%

99.48%

99.83%

20%

101.55%

99.23%

99.48%

100.72%

50%

97.24%

98.91%

99.70%

100.19%

50%

96.72%

100.02%

99.75%

100.39%

50%

100.07%

100.04%

99.70%

100.23%

120%

97.96%

100.68%

99.65%

100.11%

120%

95.90%

99.67%

99.82%

100.26%

120%

99.07%

99.96%

99.90%

99.78%

Table 17. P-values of the ANOVA two-way testing (unconditioned transport efficiencies)



ANOVA - Two Way with Replication Results for unconditioned tubing (p-value)

FEP PTFE

PFA

PVDF

Tygon®

SS316

SNSS

NAAQS value













comparison

0.06630384 0.082876

0.910589

0.974245

0.000907

0.04358

3.3E-08

Gases













Comparison

0.00727193 6.8E-09

6.21E-05

6.03E-09

1.34E-29

8.91E-49

3.43E-31

Interaction

0.2369388 0.017502

0.783657

0.837164

0.000116

0.089765

3.47E-10

Table 17 contains the p-values from the ANOVA two-way testing with unconditioned transport
efficiencies. The p-values greater than 0.05 were shaded in a light gray color while p-values less than
0.05 were shaded in light blue. (The color shading is only provided to aid in the viewing of the data.)

The fluoropolymer tubing materials have p-values greater than 0.05, indicating that there aren't
significant differences between the NAAQS concentrations groups. Tygon®, SS316 and SNSS have p-
values less than 0.05 indicating there are significant differences. All tubing materials have p-values less
than 0.05 for the gas comparison portion, indicating that there are significant differences between the
gases. This provides further statistical proof that O3 is the most reactive of the gases as the averages
were lower for each tubing material result. The interaction p-values were less than 0.05 for PTFE,
Tygon® and SNSS, and therefore indicate that the transport efficiency is dependent on NAAQS level
concentrations. Additionally, this indicates that the effect of the efficiencies at the different NAAQS
levels is dependent on gas type.

34


-------
Alternatively, the interaction between p-values greater than 0.05 for FEP, PFA, PVDF and SS316
indicates that the effect of different gas on the transport efficiency is not dependent on the concentration.

The transport efficiencies for O3 were statistically lower than the other gases as seen in Tables 16 and
17. Only FEP tubing achieved the 97.5% transport efficiency threshold for O3, and experiments were
performed with conditioned tubing to investigate if conditioning tubing would increase the transport
efficiencies over the previously mentioned threshold. In an effort to compare if conditioning would
increase the transportation efficiencies to the bulk average of unconditioned transport efficiencies, the
ANOVA two-way test was reperformed with the conditioned O3 values.

Table 18. Results from the ANOVA two-way testing with conditioned O3 transport
efficiencies.



ANOVA - Two Way with Replication Results (p-
values) for unconditioned tubing (S02, N02 and
CO) vs conditioned Tubing (03)

FEP PTFE PFA PVDF

NAAQS value

comparison

Gases

Comparison
Interaction

0.30793611 0.264976 0.975098 0.188877

0.13553282 0.004551 0.022389 0.085892
0.23736644 0.096694 0.33202 0.730326

Table 18 contains the p-values from the ANOVA two-way testing with both unconditioned transport
efficiencies and conditioned transport efficiencies.

Results indicate that conditioned fluoropolymer tubing with O3 increased the transport efficiencies to be
statistically similar to the remaining unconditioned tubing. Although the p-values remained under 0.05
for PTFE and PFA, the p-values greatly increased from the previous test results (in Table 17) providing
further support to confirm the positive aspect of the conditioning process.

35


-------
5.0 Summary and Conclusions

Ambient air monitoring sites operating under the guidelines given in Section 9 of Appendix E of
40 CFR 58 are currently required to use FEP Teflon® or borosilicate glass for their sampling probes and
sample lines. The requirement to use these materials was made to ensure that during testing, the
concentration of the criteria pollutants at the inlet of each analyzer was as close as possible to the
ambient air's concentration. While research has been conducted in the past regarding other materials that
might be considered equivalent, recent advances in the production process of tubing materials and the
development of the Silconert® non-reactive coating for stainless steel material provided an opportunity
to further investigate potentially acceptable alternative materials.

A review of pertinent literature was conducted to determine what tubing materials would be the best
potential candidates for the equivalence tests. Through this review, several commercially available
tubing materials were identified, which included several polymer materials, stainless steel with the
Silconert® non-reactive coating (SNSS), and untreated grade 316 stainless steel (SS316). The pollutant
types to be tested were the four gaseous criteria pollutants: O3, NO2, SO2, and CO.

Following the literature review, an experimental test setup was designed to determine the fractional
transport efficiency as a function of material type. During each experiment, known concentrations of
criteria pollutant were generated by a dilution calibrator and delivered to the FRM or FEM instrument
through the control material and through the test material for an equal amount of time. The control
material was a 5-foot section of FEP and the length of the candidate tubing material was adjusted to
provide a 20 second residence time to correspond with Section 9 of Appendix E. The lengths of
candidate tubing varied as a function of actual tubing ID and the operating flow rate of the FRM or FEM
analyzers used during the tests. While the pollutant generator delivered each criteria pollutant for
40 minutes, valves automatically switched between the control and candidate materials every 5 minutes.
The fractional transport efficiency was calculated by dividing the final minute average of the output
concentration by the final minute average of the input concentration. For the laboratory tests, all
candidate tubing types had a Vi" OD and a nominal 3/16 ID.

To determine the influence of concentration on transport efficiency, experiments were conducted for
each material at 20%, 50%, and 120% of the NAAQS value for each criteria pollutant being tested. The
repeatability of the test results was determined by conducting three replicate tests under identical test
conditions. This procedure thus resulted in nine separate experiments for each candidate test material for
each of the three criteria pollutant concentrations. Each of these replicate tests was conducted using a
new, unconditioned section of candidate tubing.

Research regarding increase of transport efficiency through sections of repeated use tubing until that
material's surface became physically and chemically conditioned was conducted in the past (Altshuller
et al. (1961), Scholz and Wallner (2015)). This need for conditioning of the tubing's surface was
demonstrated for very reactive pollutants such as O3. As a result, during this study, additional tests were
conducted in several cases where previously unused tubing was conditioned to a concentration of
450 ppb of the criteria pollutant for 1 hour before being tested at the three concentrations. This
procedure was adopted from the O3 line loss test procedure in the field SOP Through-the-Probe NPAP
(EPA, 2011). Section 9 of Appendix E of 40 CFR 58 indicates that alternate tubing materials can be
used if they are deemed equivalent by EPA. However, the exact definition of what constitutes
"equivalent" was never specified in the CFR. Therefore, ORD adopted NPAP's O3 line loss acceptance
criteria of 97.5%, or a pollutant loss of 2.5%, as the acceptance criteria for an equivalent candidate
material in the CFR.

36


-------
As summarized in Section 4.0, this study confirmed the FEP Teflon® was a highly efficient material for
transporting all four criteria pollutants, regardless of pre-conditioning. Specifically, for the most reactive
gas, O3, its mean transport efficiency was 98.4% in its unconditioned state, which represents a 0.9%
efficiency above the 97.5% acceptance threshold.

Comparatively, the transport efficiency of O3 through the other fluoropolymers (PVDF, PTFE, PFA)
was between 95.6% and 97.0%, meaning they were all slightly below the 97.5% acceptance threshold.
However, when any of the fluoropolymers was conditioned to O3, each material's individual transport
efficiency of O3 increased, with all results observed between 99.0% and 99.6%. The repeatability of the
three replicates tests for each of these materials was very favorable and the transport efficiency was
found to be independent of the concentration being tested. Additionally, the three fluoropolymer test
materials and FEP Teflon® all had excellent transport efficiency for SO2, NO2, and CO in their
unconditioned states. The repeatability for these pollutant tests was very favorable, and the transport
efficiency was once again independent of the concentration being tested, similar to the O3 test results.

For O3, unconditioned SNSS displayed a mean transport efficiency of 9.2%. While its transport
efficiency improved to 87.7% when conditioned to O3, the results were still below the 97.5% acceptance
criteria. For NO2, SNSS had a mean transport efficiency of 95.6% in its unconditioned state. This
improved to 98.5% when conditioned with 450 ppb NO2 for one hour, raising its results to within the
acceptance criterion. The repeatability of these tests was favorable, and the results were determined to
be independent of the concentration being tested. Additionally, SNSS performed well for SO2 and CO,
with mean transport efficiencies with unconditioned tubing of 100.1% for both gases.

SS316 and Tygon® performed poorly for O3, NO2, and SO2, with all associated transport efficiencies
falling well below the 97.5% threshold, regardless of the conditioning of the material. However, both
materials resulted in a high transport efficiency of CO, independent of the concentration being tested.

Based on these test results, a revision is proposed for Section 9 of Appendix E of 40 CFR 58 to include
PVDF, PTFE, and PFA in the list of approved materials for use in sample probes and sample lines, as
they have successfully demonstrated their ability to effectively transport gaseous criteria pollutants. The
recommendation to condition all sample probe materials using O3 will remain for all materials.

However, even with conditioning, SS316SS, SNSS, and Tygon® are still considered to be unacceptable
materials for efficient sampling of the criteria gaseous pollutants.

A summary of the transport efficiency test results for the four approved fluoropolymers is provided in
Table 19. Reported values for each pollutant and tubing material represent the mean of three replicate
tests conducted using the three pollutant concentrations. All tests were conducted using 20 second
residence times. For conditioned O3 and unconditioned NO2, SO2, and CO, mean transport efficiencies
for FEP Teflon®, PVDF, PTFE, and PFA were measured to be 99.8%, 99.9%, 99.1%, and 99.1%,
respectively.

Table 19. Summary of Transport Efficiencies for the Four Fluoropolymer Tubing
Materials

Tubing Material

Mean Trans

Mil l Efficiency for 20 Second Residence Time (%)

03

(Unconditioned)

03

(Conditioned)

so2

(Unconditioned)

no2

(Unconditioned)

CO

(Unconditioned)

Mean of the
Four Gases

FEP Teflon®

98.6

99.6

100.6

98.7

100.1

99.8

PVDF (Kvnar)

95.6

99.5

100.0

99.4

100.5

99.9

PTFE

97.0

99.5

100.4

99.2

99.6

99.7

PFA

96.8

99.0

100.5

99.1

100.3

99.7

Mean of Four Fluoropolymers

97.0

99.4

100.4

99.1

100.1



CV (%)

1.27

0.27

0.26

0.30

0.39



37


-------
Selection of which fluoropolymer to incorporate into a given NAAQS monitoring station is largely a
subjective decision. As previously mentioned and documented in Table 19, the transfer efficiencies of
the gaseous pollutants through each fluoropolymer types are statistically identical independent of
pollutant type and pollutant concentration. All four fluoropolymer types display high resistance to UV
light and are thus suitable for indoor and outdoor use. FEP Teflon® and PFA have one significant
advantage over PVDF and PTFE in that they are transparent and thus provide a better visual indication
when they've become internally contaminated and require replacement.

Table 20. Properties of Fluoropolymer Tubing Materials

Tubing
Material

Trade Name & Owner

3/16"
ID

Cost/ft

UV

Resistance

Optical
Properties

FEP

Teflon® FEP (DuPont)

Neoflon (Daikin)

Hyflon (Solvay Specialty
Polymers)

$3.00

High

Transparent

PVDF

Kynar (Arkema, Inc.)

Solef (Solvay Specialty
Polymers)

Hylar (Solvay Specialty
Polymers)

$2.13

High

Opaque

PTFE

Teflon® (DuPont)

Algoflon (Solvay Specialty
Polymers)

Polyflon (Daikan)

$1.85

High

Opaque

PFA

Fluon PFA (AGC
Chemicals)

$3.17

High

Transparent

Note: Costs accessed July 2022 from https://mcmaster.com

All fluoropolymer materials are inherently flexible, although overall flexibility and strength are a
function of each tubing's wall thickness. In some cases, monitoring organizations have reported
crimping problems with 3/16" ID tubing and therefore have adopted 1/8" ID tubing to minimize
crimping. Because an equivalent length of 1/8" ID tubing has one-half the surface area and 45% of the
pollutant residence time as that of 3/16" ID tubing at the same flow rate, the loss of pollutants through
1/8" ID tubing would be predicted to be even lower than that of 3/16" ID tubing. Due to the greater
amount of material required to fabricate 1/8" ID tubing, however, its cost is approximately twice that of

38


-------
the cost listed in Table 20 for 3/16" ID tubing. It's also important to ensure that flow systems can
provide the proper flow rate given the 1/8" ID tubing's higher pressure drop.

This report has identified and documented three additional fluoropolymers (PVDF, PTFE, and PFA)
which have statistically identical pollutant transport performance as that of FEP Teflon® and thus meet
the regulatory requirements for efficient NAAQS pollutant sampling. It was also determined that
sampling lines intended for O3 sampling require conditioning prior to use. No conditioning is required
for sampling lines intended for sampling of SO2, NO2 and CO.

EPA intends to revise Section 9 of Appendix E of 40 CFR Part 58 to include these three fluoropolymer
materials. In conjunction with the previously approved borosilicate glass and FEP Teflon® materials,
including PVDF, PTFE, and PFA will provide monitoring organizations with a larger variety of efficient
sampling and transport materials with which to meet their Congressionally mandated obligations to
conduct accurate NAAQS compliance monitoring.

39


-------
6.0 References

Altshuller, A. P., Wartburg, A., Taft, R.A. (1961). The interaction of ozone with plastic and metallic

materials in a dynamic flow system. International Journal of Air and Water Pollution, 4, 70-78.
Byers, R.L., Davis, J.W. (1970) Sulfur Dioxide Adsorption and Desorption on Various Filter Media, J. Air
Pollution Control Assoc., Vol. 20, pp. 236-238.

Campos, V. P., Cruz, L. P. S., Godoi, R. H. M., Godoi, A. F. L., Tavares, T. M. (2020). Development and
validation of passive samplers for atmospheric monitoring of SO2, NO2, O3, and H2S in tropical
areas. Microchemical Journal, 96, 132-138. https://doi.Org/10.1016/i.microc.2010.02.015
Code of Federal Regulations. (2018). Probe and monitoring path siting criteria for ambient air quality

monitoring, 40 CFR 58, Appendix E, § 9.

Hughes, E.E. (1975). Development of standard reference materials for air quality measurement. ISA

Transaction, 14(4), 281-291.

Marshik, B. (2015, October 19-23). Material and process conditions for successful use of extractive
sampling techniques. EPA Region 6 - 25th Annual Quality Assurance Conference, Dallas, TX,
United States.

Obermiller, E. L., Charlier, (1968). Gas chromatographic separation

of nitrogen, oxygen, argon, carbon monoxide, carbon dioxide, hydrogen sulfide, and sulfur
dioxide. Journal of Gas Chromatography, 6, 446-447. https://doi.Org/10.1093/chromsci/6.8.446
Palmes, E. D., Gunnison, A. F., Dimattio, J., & Tomczyk, C. (1976). Personal sampler for nitrogen dioxide.
American Industrial Hygiene Association Journal, 37(10), 570-577.
https://doi.org/10.1080/00028897685Q7522
Scholz, A (2015) "The Influence of Sample Tubing Material on Accuracy of Low Ozone Concentration
and Measurement", IOA-EA3G Conference, Barcelona, Spain 28 June - 3 July, 2015

40


-------
SilcoTek. (2020). SilcoNert® 2000 Coating Data Sheet.

https://www.silcotek.com/hubfs/Literature%20Cataloe/Data%20Sheets/DATA-SILCQNERT2.pdf
US EPA (2012)

"Traceability Protocol for Assay and Certification of Gaseous Calibration Standards" (PDF) (174
pp, 1.7 M, About PDF) Publication No. EPA/600/R-12/53, available at https://www.epa.gov/air-
research/epa-traceability-protocol-assay-and-certification-gaseous-calibration-standards
US EPA (2011) "Field Standard Operating Procedures for the EPA Through-the Probe National

Performance Audit Program" available via https://www.epa.gov/sites/default/files/2020-
10/documents/20110719npapttpdraftoperatorsfieldsop O.pdf

Wechter, S. (1976). Preparation of stable pollution gas standards using treated aluminum cylinders.

STP598-EB Calibration in Air Monitoring, 40-54. https://doi.org/10.1520/STP32353S
Wohlers, H. C., Newstein, H., Daunis, D. (1967). Carbon monoxide and sulfur dioxide adsorption on —
and desorption from glass, plastic, and metal tubings. Journal of the Air Pollution Control Association,
17(11), 753-756. https://doi.org/lQ.lQ8Q/00022470.1967.10469Q68

41


-------
Appendix A

Transport efficiencies and standard deviations for each tubing type and gas concentration.

Appendix A Table 1. Ozone (Unconditioned)

Ozone

15ppb

Repl

Rep2

Rep3

Tubing

Transport

Standard

Transport

Standard

Transport

Standard

Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

98.31%



0.67%

100.29%

0.41%

101.55%

0.43%

PTFE

94.77%



0.25%

96.56%

0.25%

95.24%

0.39%

PFA

93.88%



0.00%

99.49%

0.55%

96.22%

0.39%

PVDF

93.58%



0.42%

96.06%

0.31%

95.12%

0.42%

Tygon®

0.12%



0.27%

8.18%

0.03%

4.12%

0.01%

SS316

0.58%



0.27%

-0.28%

-0.35%

-0.50%

-0.30%

SNSS

-0.23%



-0.33%

0.67%

0.00%

-0.69%

0.00%

Ozone

35ppb

Repl

Rep2

Rep3

Tubing

Transport

Standard

Transport

Standard

Transport

Standard

Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

97.24%



0.27%

96.72%

0.34%

100.07%

0.57%

PTFE

96.56%



0.41%

97.21%

0.19%

98.15%

0.27%

PFA

95.43%



0.27%

98.44%

0.17%

97.71%

0.26%

PVDF

93.09%



0.27%

95.13%

0.36%

98.65%

0.25%

Tygon®

1.10%



0.13%

14.17%

0.02%

11.77%

0.12%

SS316

-0.58%



0.00%

0.00%

0.00%

-0.37%

-0.13%

SNSS

-0.36%



-0.13%

0.58%

0.00%

0.29%

0.00%

Ozone

85ppb

Repl

Rep2

Rep3

Tubing

Transport

Standard

Transport

Standard

Transport

Standard

Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

97.96%



0.30%

95.90%

0.16%

99.07%

0.14%

PTFE

97.34%



0.19%

98.35%

0.16%

98.77%

0.11%

PFA

95.44%



0.14%

97.55%

0.14%

97.42%

0.22%

PVDF

95.76%



0.07%

96.38%

0.11%

96.30%

0.10%

Tygon®

19.44%



0.02%

21.55%

0.05%

19.56%

0.01%

SS316

-0.24%



0.00%

-0.12%

0.00%

-0.05%

-0.06%

SNSS

31.76%



1.20%

18.83%

1.04%

32.10%

1.08%

42


-------
Appendix A Table 2. Ozone (Conditioned)

Ozone

15ppb

Conditioned

Repl

Rep2

Rep3



Transport

Standard

Transport

Standard

Transport

Standard

Tubing Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

97.89%



0.38%

100.23%



0.38%

101.29%



0.54%

PTFE

99.94%



0.34%

100.00%



0.58%

100.64%



0.46%

PFA

99.22%



0.48%

98.32%



0.35%

98.24%



0.44%

PVDF

98.16%



0.43%

101.69%



0.43%

101.72%



0.44%

Tygon®

1.03%



0.36%

0.22%



0.33%

1.48%



0.27%

SS316

-0.05%



-0.06%

N/A



N/A

N/A



N/A

SNSS

79.66%



0.49%

N/A



N/A

N/A



N/A

Ozone

35ppb

Conditioned

Repl

Rep2

Rep3



Transport

Standard

Transport

Standard

Transport

Standard

Tubing Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

100.31%



0.25%

100.05%



0.26%

99.01%



0.21%

PTFE

99.19%



0.21%

98.60%



0.11%

98.13%



0.22%

PFA

98.97%



0.19%

99.23%



0.20%

98.71%



0.18%

PVDF

98.96%



0.49%

98.82%



0.19%

99.73%



0.23%

Tygon®

-0.27%



-0.09%

0.00%



0.00%

0.47%



0.15%

SS316

-0.36%



-0.13%

N/A



N/A

N/A



N/A

SNSS

89.57%



0.17%

N/A



N/A

N/A



N/A

Ozone

85ppb

Conditioned

Repl

Rep2

Rep3



Transport

Standard

Transport

Standard

Transport

Standard

Tubing Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

98.78%



0.09%

99.15%



0.17%

99.59%



0.09%

PTFE

99.59%



0.14%

100.31%



0.18%

99.04%



0.10%

PFA

99.50%



0.07%

99.51%



0.16%

99.59%



0.14%

PVDF

98.30%



0.11%

99.41%



0.11%

98.59%



0.17%

Tygon®

-0.19%



-0.06%

0.12%



0.00%

0.47%



0.06%

SS316

-0.17%



-0.31%

N/A



N/A

N/A



N/A

SNSS

93.84%



0.17%

N/A



N/A

N/A



N/A

\Tote that blank values indicate that test was not performed due to lack in of tubing availability (SNSS)
or lack of efficiency increase due to Conditioning (SS316).

43


-------
Appendix A Table 3. Sulfur Dioxide (Unconditioned)

Sulfur

15ppb

Dioxide

Repl

Rep2

Rep3

Tubing
Type

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

FEP

103.97%



1.08%

102.80%



0.51%

99.23%



0.49%

PTFE

101.12%



0.63%

101.84%



1.14%

99.37%



1.07%

PFA

101.44%



0.91%

96.01%



0.51%

100.59%



0.61%

PVDF

98.95%



0.74%

102.92%



0.49%

99.42%



0.77%

Tygon®

88.21%



0.93%

86.14%



0.93%

90.98%



0.71%

SS316

2.96%



0.01%

3.33%



0.39%

2.29%



0.28%

SNSS

98.68%



0.98%

100.24%



0.77%

101.20%



0.47%

Sulfur

40ppb

Dioxide

Repl

Rep2

Rep3

Tubing
Type

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

FEP

98.91%



0.35%

100.02%



0.29%

100.04%



0.37%

PTFE

99.51%



0.34%

100.00%



0.19%

101.48%



0.43%

PFA

100.68%



0.50%

100.83%



0.23%

99.66%



0.25%

PVDF

99.39%



0.43%

99.70%



0.29%

99.83%



0.56%

Tygon®

89.70%



0.48%

88.35%



0.78%

87.87%



0.21%

SS316

3.07%



36.10%

2.95%



0.14%

3.07%



0.12%

SNSS

99.76%



0.39%

100.36%



0.35%

99.96%



0.46%

Sulfur

90ppb

Dioxide

Repl

Rep2

Rep3

Tubing
Type

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

FEP

100.68%



0.31%

99.67%



0.17%

99.96%



0.18%

PTFE

100.10%



0.44%

100.42%



0.15%

100.15%



0.24%

PFA

101.27%



0.28%

100.86%



0.17%

99.64%



0.34%

PVDF

100.54%



0.16%

99.86%



0.22%

99.75%



0.13%

Tygon®

90.52%



0.39%

89.86%



0.27%

88.33%



0.24%

SS316

3.16%



0.01%

3.24%



0.05%

3.05%



0.00%

SNSS

100.46%



0.28%

100.37%



0.21%

99.69%



0.18%

44


-------
Appendix A Table 4. Sulfur Dioxide (Conditioned)

Sulfur

15ppb

Dioxide







(Conditioned)

Repl

Rep2

Rep3



Transport Standard

Transport Standard

Transport Standard

Tubing Type

Efficiency Deviation

Efficiency Deviation

Efficiency Deviation

FEP

99.64% 0.62%

100.82% 0.49%

98.09% 0.42%

PTFE

N/A N/A

N/A N/A

N/A N/A

PFA

N/A N/A

N/A N/A

N/A N/A

PVDF

N/A N/A

N/A N/A

N/A N/A

Tygon®

87.90% 1.13%

85.46% 0.68%

87.90% 0.61%

SS316

79.26% 0.53%

63.46% 1.98%

72.34% 0.97%

SNSS

N/A N/A

N/A N/A

N/A N/A

Sulfur

40ppb

Dioxide







(Conditioned)

Repl

Rep2

Rep3



Transport Standard

Transport Standard

Transport Standard

Tubing Type

Efficiency Deviation

Efficiency Deviation

Efficiency Deviation

FEP

100.47% 0.22%

99.68% 0.40%

99.14% 0.43%

PTFE

N/A N/A

N/A N/A

N/A N/A

PFA

N/A N/A

N/A N/A

N/A N/A

PVDF

N/A N/A

N/A N/A

N/A N/A

Tygon®

88.30% 0.44%

87.13% 0.41%

87.94% 0.36%

SS316

73.05% 1.14%

72.29% 0.59%

73.22% 0.56%

SNSS

N/A N/A

N/A N/A

N/A N/A

Sulfur

90ppb

Dioxide







(Conditioned)

Repl

Rep2

Rep3



Transport Standard

Transport Standard

Transport Standard

Tubing Type

Efficiency Deviation

Efficiency Deviation

Efficiency Deviation

FEP

99.36% 0.33%

100.27% 0.15%

100.90% 0.37%

PTFE

N/A N/A

N/A N/A

N/A N/A

PFA

N/A N/A

N/A N/A

N/A N/A

PVDF

N/A N/A

N/A N/A

N/A N/A

Tygon®

88.22% 0.18%

87.70% 0.21%

87.01% 0.23%

SS316

85.91% 0.50%

87.09% 0.34%

86.16% 0.44%

SNSS

N/A N/A

N/A N/A

N/A N/A

Note that blank values indicate that test was not performed due to original set of data reaching the 97.5%
transport efficiency threshold.

45


-------
Appendix A Table 5. Nitrogen Dioxide (Unconditioned)

Nitrogen

20ppb

Dioxide

Verification

Repl

Rep2

Rep3

Tubing
Type

Transport Standard
Efficiency Deviation

Transport Standard
Efficiency Deviation

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

FEP

99.75%

0.29%

99.48%

0.00%

99.48%



0.00%

99.74%



0.31%

PTFE

99.50%

0.00%

99.47%

0.00%

99.47%



0.00%

99.47%



0.00%

PFA

99.50%

0.00%

99.48%

0.00%

99.74%



0.31%

99.47%



0.00%

PVDF

99.38%

0.25%

99.61%

0.40%

99.48%



0.00%

99.47%



0.00%

Tygon®

78.90%

0.31%

80.80%

0.23%

80.66%



0.38%

79.35%



0.33%

SS316

-1.51%

-0.02%

-0.52%

0.00%

-1.04%



0.00%

-1.04%



0.00%

SNSS





96.77%

0.23%

92.39%



0.24%

92.85%



0.36%

Nitrogen

50ppb

Dioxide

N/A

Repl

Rep2

Rep3

Tubing
Type



Transport Standard
Efficiency Deviation

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

FEP

N/A

N/A

99.70%

0.14%

99.75%



0.15%

99.70%



0.12%

PTFE

N/A

N/A

98.90%

0.12%

99.40%



0.00%

98.85%



0.10%

PFA

N/A

N/A

99.15%

0.10%

98.99%



0.00%

98.79%



0.00%

PVDF

N/A

N/A

99.50%

0.14%

99.45%



0.15%

99.65%



0.15%

Tygon®

N/A

N/A

79.80%

0.09%

79.12%



0.08%

78.36%



0.00%

SS316

N/A

N/A

-0.61%

0.00%

-0.75%



-0.10%

-0.81%



0.00%

SNSS

N/A

N/A

94.82%

0.16%

95.92%



0.15%

95.49%



0.11%

Nitrogen



Dioxide

N/A

Repl

Rep2

Rep3

Tubing
Type



Transport Standard
Efficiency Deviation

Transport
Efficiency

Standard
Deviation

Transport
Efficiency

Standard
Deviation

FEP

N/A

N/A

99.65%

0.06%

99.82%



0.30%

99.90%



0.13%

PTFE

N/A

N/A

98.85%

0.15%

99.32%



0.10%

98.92%



0.08%

PFA

N/A

N/A

98.81%

0.09%

98.85%



0.09%

99.06%



0.09%

PVDF

N/A

N/A

99.09%

0.07%

99.73%



0.15%

98.78%



0.06%

Tygon®

N/A

N/A

80.66%

0.05%

80.30%



0.09%

81.47%



0.08%

SS316

N/A

N/A

-0.25%

0.00%

-0.25%



0.00%

-0.16%



0.00%

SNSS

N/A

N/A

96.98%

0.09%

97.51%



0.04%

97.37%



0.09%

An extra test or verification test was performed due to an error in data recording by the Envidas System.
The verification data set was used in the overall transport efficiencies table in Table 9.

46


-------
Appendix A Table 6. Nitrogen Dioxide (Conditioned)

Nitrogen Dioxide

20ppb

(Conditioned)

Repl

Rep2

Rep3



Transport Standard

Transport Standard

Transport Standard

Tubing Type

Efficiency Deviation

Efficiency Deviation

Efficiency Deviation

FEP

N/A N/A

N/A N/A

N/A N/A

PTFE

N/A N/A

N/A N/A

N/A N/A

PFA

N/A N/A

N/A N/A

N/A N/A

PVDF

N/A N/A

N/A N/A

N/A N/A

Tygon®

85.49% 0.00%

84.90% 0.00%

84.82% 0.00%

SS316

34.27% 1.61%

42.66% 0.70%

43.41% 0.27%

SNSS

98.53% 0.00%

97.28% 0.28%

97.54% 0.00%

Nitrogen Dioxide

50ppb

(Conditioned)

Repl

Rep2

Rep3



Transport Standard

Transport Standard

Transport Standard

Tubing Type

Efficiency Deviation

Efficiency Deviation

Efficiency Deviation

FEP

N/A N/A

N/A N/A

N/A N/A

PTFE

N/A N/A

N/A N/A

N/A N/A

PFA

N/A N/A

N/A N/A

N/A N/A

PVDF

N/A N/A

N/A N/A

N/A N/A

Tygon®

N/A N/A

N/A N/A

N/A N/A

SS316

N/A N/A

N/A N/A

N/A N/A

SNSS

99.10% 0.12%

98.25% 0.15%

98.35% 0.10%

Nitrogen Dioxide

120ppb

(Conditioned)

Repl

Rep2

Rep3



Transport Standard

Transport Standard

Transport Standard

Tubing Type

Efficiency Deviation

Efficiency Deviation

Efficiency Deviation

FEP

N/A N/A

N/A N/A

N/A N/A

PTFE

N/A N/A

N/A N/A

N/A N/A

PFA

N/A N/A

N/A N/A

N/A N/A

PVDF

N/A N/A

N/A N/A

N/A N/A

Tygon®

N/A N/A

N/A N/A

N/A N/A

SS316

N/A N/A

N/A N/A

N/A N/A

SNSS

99.18% 0.09%

99.00% 0.09%

98.95% 0.08%

\Tote that blank values indicate that test was not performed due to original set of data reaching the 97.5%
transport efficiency threshold. Tygon® and SS316 were not tested at higher concentrations due to lower
concentrations not increasing over the 97.5% transport efficiency threshold.

47


-------
Appendix A Table 7. Carbon Monoxide (Unconditioned)

Carbon

1.8ppb

Monoxide

Repl

Rep2

Rep3



Transport

Standard

Transport

Standard

Transport

Standard

Tubing Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

99.48%

0.79%

99.83%

0.71%

100.72%



1.47%

PTFE

100.03%

1.12%

99.06%

2.16%

97.79%



0.96%

PFA

101.81%

0.98%

100.91%

1.89%

99.82%



1.05%

PVDF

101.14%

2.50%

100.02%

2.04%

100.65%



1.17%

Tygon®

101.52%

0.93%

100.75%

1.90%

100.50%



1.01%

SS316

97.32%

1.07%

98.88%

1.41%

99.93%



1.11%

SNSS

100.40%

1.15%

98.77%

2.23%

102.27%



1.43%

Carbon

4.5ppb

Monoxide

Repl

Rep2

Rep3



Transport

Standard

Transport

Standard

Transport

Standard

Tubing Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

100.19%

0.36%

100.39%

1.00%

100.23%



0.45%

PTFE

99.12%

0.40%

99.35%

0.74%

100.51%



0.73%

PFA

99.65%

0.41%

100.64%

0.55%

100.71%



0.49%

PVDF

100.52%

0.66%

101.00%

0.24%

99.91%



0.43%

Tygon®

99.76%

0.51%

100.60%

0.42%

100.64%



0.52%

SS316

100.09%

0.58%

100.71%

0.61%

99.95%



0.59%

SNSS

100.00%

0.41%

100.55%

0.74%

99.14%



0.95%

Carbon

10.8ppb

Monoxide

Repl

Rep2

Rep3



Transport

Standard

Transport

Standard

Transport

Standard

Tubing Type

Efficiency

Deviation

Efficiency

Deviation

Efficiency

Deviation

FEP

100.11%

0.20%

100.26%

0.17%

99.78%



0.36%

PTFE

100.02%

0.46%

100.22%

0.37%

100.40%



0.34%

PFA

99.42%

0.42%

99.56%

0.41%

99.92%



0.46%

PVDF

100.04%

0.36%

100.46%

0.20%

100.38%



0.26%

Tygon®

100.08%

0.30%

100.18%

0.48%

100.00%



0.29%

SS316

99.91%

0.29%

99.42%

0.54%

99.58%



0.21%

SNSS

100.06%

0.39%

99.33%

0.56%

100.06%



0.33%

48


-------
oEPA

United States
Environmental Protection
Agency

PRESORTED
STANDARD POSTAGE
& FEES PAID EPA
PERMIT NO.G-35

Office of Research and
Development (8101R)
Washington, DC 20460

Official Business
Penalty for Private Use
$300


-------