Method 630

The Determination
of Dithiocarbamate
Pesticides in Municipal and
Industrial Wastewater


-------
Method 630

The Determination of Dithiocarbamate Pesticides in Municipal and

Industrial Wastewater

1. Scope and Application

1.1 This method covers the determination of dithiocarbamate pesticides. The following
parameters can be determined by this method:

Parameter

CAS No.

Amoban

3566-10-7

AOP

—

Busan 40

51026-28-9

Busan 85

128-03-0

Ferbam

14484-64-1

KN Methyl

137-41-7

Mancozeb

8018-01-7

Maneb

12427-38-1

Metham

137-42-8

Nabam

142-59-6

Niacide

8011-66-3

Polyram

9006-42-2

Sodium dimethyldithiocarbamate

128-04-1

Thiram

137-26-8

ZAC

—

Zineb

12122-67-7

Ziram

137-30-4

1.2	This method fails to distinguish between the individual dithiocarbamates. The
compounds above are reduced to carbon disulfide and the total dithiocarbamate
concentration is measured. Unless the sample can be otherwise characterized, all results
are reported as Ziram. Carbon disulfide is a known interferent.

1.3	This is a colorimetric method applicable to the determination of the compounds listed
above in industrial and municipal discharges as provided under 40 CFR 136.1. Any
modification of this method beyond those expressly permitted shall be considered a
major modification subject to application and approval of alternative test procedures
under 40 CFR 136.4 and 136.5.

1.4	The method detection limit (MDL, defined in Section 12) for maneb, metham and ziram
are listed in Table 1. The MDL for a specific dithiocarbamate or wastewater may differ
from those listed, depending upon the nature of interferences in the sample matrix.

1.5	This method is restricted to use by or under the supervision of analysts experienced in
trace organic analyses. Each analyst must demonstrate the ability to generate acceptable
results with this method using the procedure described in Section 8.2.


-------
Method 630

2.	Summary of Method

2.1 A measured volume of sample, approximately 1 L, is digested with acid to yield carbon
disulfide by hydrolysis of the dithiocarbamate moiety. The evolved CS2 is purged from
the sample and absorbed by a color reagent. The absorbance of the solution is measured
at 380 and 435 nm using a UV-visible spectrophotometer.1

3.	Interferences

3.1	Method interferences may be caused by contaminants in reagents, glassware, and other
sample processing hardware that lead to high blank values and biased results. All of
these materials must be routinely demonstrated to be free from interferences under the
conditions of the analysis by running laboratory reagent blanks as described in Section
8.5.

3.1.1	Glassware must be scrupulously cleaned.2 After each use, rinse the
decomposition flask and condenser with 4N NaOH and reagent water. Overnight
soaking in 4N NaOH may be necessary. Clean the H2S scrubber between each
use with 0.1N HC1 in methanol, rinse three times with methanol, and bake at
200°C for 15 minutes. Rinse the CS2 trap with methanol three times between each
use and follow by heating for 15 minutes at 200°C. Should it become difficult to
force the color reagent through the glass frit of the CS2 trap, clean in the same
manner as the H2S scrubber. After cooling, store glassware sealed to prevent any
accumulation of dust or other contaminants.

3.1.2	The use of high-purity reagents and solvents helps to minimize interference
problems.

3.2	Carbon disulfide may be a significant direct interferent in wastewaters. Its elimination
or control is not addressed in this method. If correction for background carbon disulfide
is required, the CS2 should be measured by an independent procedure, such as direct
aqueous injection gas chromatography.

3.3	Additional matrix interferences may be caused by contaminants that are codistilled from
the sample. The extent of matrix interferences will vary considerably from source to
source, depending upon the nature and diversity of the industrial complex or
municipality being sampled. The cleanup provided by the H2S trap will eliminate or
reduce some of these interferences, but unique samples may require additional clean-up
approaches to achieve the MDL listed in Table 1.

4.	Safety

4.1 The toxicity or carcinogenicity of each reagent used in this method has not been precisely
defined; however, each chemical compound should be treated as a potential health
hazard. From this viewpoint, exposure to these chemicals must be reduced to the lowest
possible level by whatever means available. The laboratory is responsible for maintaining
a current awareness file of OSHA regulations regarding the safe handling of the
chemicals specified in this method. A reference file of material data handling sheets
should also be made available to all personnel involved in the chemical analysis.


-------
Method 630

Additional references to laboratory safety are available and have been identified3 5 for the
information of the analyst.

5. Apparatus and Materials

5.1	Sampling equipment, for discrete or composite sampling.

5.1.1	Grab-sample bottle: Amber borosilicate or flint glass, 1-L or 1-quart volume,
fitted with screw-caps lined with TFE-fluorocarbon. Aluminum foil may be
substituted for TFE if the sample is not corrosive. If amber bottles are not
available, protect samples from light. The container and cap liner must be
washed, rinsed with acetone or methylene chloride, and dried before use to
minimize contamination.

5.1.2	Automatic sampler (optional): Must incorporate glass sample containers for the
collection of a minimum of 250 mL. Sample containers must be kept refrigerated
at 4°C and protected from light during compositing. If the sampler uses a
peristaltic pump, a minimum length of compressible silicone rubber tubing may
be used. Before use, however, the compressible tubing must be thoroughly rinsed
with methanol, followed by repeated rinsings with reagent water to minimize the
potential for contamination of the sample. An integrating flow meter is required
to collect flow-proportional composites.

5.2	Dithiocarbamate hydrolysis apparatus (Figure 1): Available from Southern Scientific Inc.,
Box 83, Micanopy, Florida 32267. Apparatus includes the following or equivalent
components.

5.2.1	Hot plate with magnetic stirrer.

5.2.2	Hydrolysis flask: 2-L, flat bottom with ground-glass joints, two necks.

5.2.3	Condenser: Low internal volume, ground-glass joints, Liebig (Kontes K-447000,
100 mm or equivalent).

5.2.4	Gas-washing bottles: 125-mL, with extra-coarse porosity (Kontes K-657750 or
equivalent).

5.2.5	Addition funnel: 60-mL, ground-glass joint to fit hydrolysis flask, with long stem
to reach at least 2 cm below the liquid level in the hydrolysis flask.

5.2.6	Dust trap (adapter): To fit top of addition funnel (Kontes K-174000 or equivalent).

5.2.7	Vacuum source: Stable pressure with needle valve for control.

5.3	UV-visible spectrophotometer: Double beam with extended cell path length capability
of 1.0 and 4.0 cm cells.


-------
Method 630

5.4 Balance: Analytical, capable of accurately weighing to the nearest 0.0001 g. The
preparation of calibration standards for some dithiocarbamates (e.g., metham) requires
the use of a balance capable of weighing 10 |ig.

6. Reagents

6.1	Reagent water: Reagent water is defined as a water in which an interferent is not
observed at the method detection limit of each parameter of interest. Prepare by boiling
distilled water 15 minutes immediately before use.

6.2	Acetonitrile, diethanolamine, methanol: ACS grade.

6.3	Ethanol: 95%.

6.4	Cupric acetate: Monohydrate, ACS grade.

6.5	Hydrochloric acid: Concentrated.

6.6	Hydrochloric acid, 0.1N in methanol: Slowly add 8.3 mL concentrated HC1 to methanol
and dilute to 100 mL.

6.7	Sodium hydroxide, 4N: Dissolve 16 g ACS grade NaOH pellets in reagent water and
dilute to 100 mL.

6.8	Stannous chloride: SnCl2*2H20, ACS grade.

6.9	Zinc acetate solution, 20%: Dissolve 20 g ACS grade ZnfC^O^ 2HzO in reagent water
and dilute to 100 mL.

6.10	Color reagent: Add 0.012 g cupric acetate monohydrate to 25 g diethanolamine. Mix
thoroughly while diluting to 250 mL with ethanol. Store in amber bottle with TFE-
fluorocarbon-lined cap.

6.11	Decomposition reagent: Dissolve 9.5 g stannous chloride in 300 mL concentrated
hydrochloric acid. Prepare fresh daily.

6.12	Stock standard solutions (1.00 |ig/|_iL): Stock standard solutions may be prepared from
pure standard materials or purchased as certified solutions.

6.12.1 Prepare a stock standard solution for ziram by accurately weighing approximately
0.0100 g of pure material. Dissolve the material in acetonitrile and dilute to
volume in a 1-mL volumetric flask. Larger volumes may be used at the
convenience of the analyst. If compound purity is certified at 96% or greater, the
weight may be used without correction to calculate the concentration of the stock
standard. Commercially prepared stock standards may be used at any
concentration if they are certified by the manufacturer or by an independent
source.


-------
Method 630

6.12.2	Transfer the stock standard solution into a TFE-fluorocarbon-sealed screw-cap
vial. Store at 4°C and protect from light. Frequently check stock standard
solutions for signs of degradation or evaporation, especially just prior to
preparing calibration standards from them.

6.12.3	Stock standard solutions must be replaced after 6months, or sooner if comparison
with check standards indicates a problem.

6.12.4	When using other dithiocarbamates for calibration, such as maneb or metham, it
may be necessary to weigh microgram amounts of the pure material into small
aluminum foil boats and place them directly in the hydrolysis flask.

7.	Calibration

7.1	Use ziram as the standard for total dithiocarbamates when a mixture of dithiocarbamates
is likely to be present. Use the specific dithiocarbamate as a standard when only one
pesticide is present and its identity has been established.

7.2	With the apparatus assembled and reagents in place (Section 10), pour 1500 mL of
reagent water into each decomposition flask, add 30 mL of decomposition reagent, and
start aspiration.

7.3	Spike the water in each flask with an accurately known weight of dithiocarbamate
standard. Use a series of weights equivalent to 5 to 200 pg of CS2. Follow the procedure
outlined Section 10.

7.4	Prepare calibration curves at a minimum of three concentrations by plotting absorbance
vs. weight of dithiocarbamate. A separate curve is prepared from readings taken at 435
nm and at 380 nm for each cell path length used. Normally the 435 nm curve is used for
calibration above 30 |ig ziram (4 cm cell), and the 380 nm curve is used for calibration
below 30 |ig ziram. The choice of which curve to use is left to the discretion of the
analyst. It is recommended that the curves be transformed into mathematical equations
using linear least squares fit for the data from 435 nm and quadratic least squares fit for
data from the 380 nm.

7.5	The working calibration curve must be verified on each working shift by the
measurement of one or more calibration standards. If the response varies from the
predicted response by more than ±10%, the test must be repeated using a fresh
calibration standard. Alternatively, a new calibration curve must be prepared.

8.	Quality Control

8.1 Each laboratory using this method is required to operate a formal quality control
program. The minimum requirements of this program consist of an initial demonstration
of laboratory capability and the analysis of spiked samples as a continuing check on
performance. The laboratory is required to maintain performance records to define the
quality of data that is generated.


-------
Method 630

8.1.1	Before performing any analyses, the analyst must demonstrate the ability to
generate acceptable accuracy and precision with this method. This ability is
established as described in Section 8.2.

8.1.2	The laboratory must spike and analyze a minimum of 10% of all samples to
monitor continuing laboratory performance. This procedure is described in
Section 8.4.

8.2	To establish the ability to generate acceptable accuracy and precision, the analyst must

perform the following operations.

8.2.1	Select a representative spike concentration for each compound to be measured.

8.2.2	Add the known amount of dithiocarbamate standard to each of a minimum of
four 1000-mL aliquots of reagent water. A representative wastewater may be
used in place of the reagent water, but one or more additional aliquots must be
analyzed to determine background levels, and the spike level must exceed twice
the background level for the test to be valid. Analyze the aliquots according to
the method beginning in Section 10.

8.2.3	Calculate the average percent recovery (R), and the standard deviation of the
percent recovery (s), for the results. Wastewater background corrections must be
made before R and s calculations are performed.

8.2.4	Using the appropriate data from Table 1, determine the recovery and single-
operator precision expected for the method, and compare these results to the
values calculated in Section 8.2.3. If the data are not comparable, review potential
problem areas and repeat the test.

8.3	The analyst must calculate method performance criteria and define the performance of

the laboratory for each spike concentration and parameter being measured.

8.3.1	Calculate upper and lower control limits for method performance as follows:

Upper Control Limit (UCL) = R + 3s
Lower Control Limit (LCL) = R - 3s

where R and s are calculated as in Section 8.2.3. The UCL and LCL can be used
to construct control charts6 that are useful in observing trends in performance.

8.3.2	The laboratory must develop and maintain separate accuracy statements of
laboratory performance for wastewater samples. An accuracy statement for the
method is defined as R ± s. The accuracy statement should be developed by the
analysis of four aliquots of wastewater as described in Section 8.2.2, followed by
the calculation of R and s. Alternatively, the analyst may use four wastewater
data points gathered through the requirement for continuing quality control in
Section 8.4. The accuracy statements should be updated regularly.6

8.4	The laboratory is required to collect in duplicate a portion of their samples to monitor

spike recoveries. The frequency of spiked sample analysis must be at least 10% of all


-------
Method 630

samples or one spiked sample per month, whichever is greater. One aliquot of the
sample must be spiked and analyzed as described in Section 8.2. If the recovery for a
particular parameter does not fall within the control limits for method performance, the
results reported for that parameter in all samples processed as part of the same set must
be qualified as described in Section 11.3. The laboratory should monitor the frequency
of data so qualified to ensure that it remains at or below 5%.

8.5	Before processing any samples, the analyst must demonstrate through the analysis of a
1-L aliquot of reagent water that all glassware and reagent interferences are under
control. Each time a set of samples is extracted or there is a change in reagents, a
laboratory reagent blank must be processed as a safeguard against laboratory
contamination.

8.6	It is recommended that the laboratory adopt additional quality assurance practices for
use with this method. The specific practices that are most productive depend upon the
needs of the laboratory and the nature of the samples. Field duplicates may be analyzed
to monitor the precision of the sampling technique. Whenever possible, the laboratory
should perform analysis of quality control materials and participate in relevant
performance evaluation studies.

9.	Sample Collection, Preservation, and Handling

9.1	Grab samples must be collected in glass containers. Conventional sampling practices7
should be followed; however, the bottle must not be prerinsed with sample before
collection. Composite samples should be collected in refrigerated glass containers in
accordance with the requirements of the program. Automatic sampling equipment must
be as free as possible of plastic and other potential sources of contamination.

9.2	The samples must be iced or refrigerated at 4°C from the time of collection until
extraction.

9.3	All samples must be analyzed within 7 days of collection.

10.	Sample Analysis

10.1	Assemble the hydrolysis apparatus as follows (see Figure 1).

10.1.1	Place the hydrolysis flask on the hot plate.

10.1.2	Place the addition funnel in one of the necks of the hydrolysis flask and the dust
trap in the top of the funnel.

10.1.3	Place the condenser in the other neck and attach two gas-washing bottles in
succession to the condenser outlet.

10.1.4	Attach a vacuum line with a flow valve to the second scrubber.

10.2	Allow the sample to warm to room temperature. Mark the water meniscus on the side
of the sample bottle for later determination of sample volume. Pour the entire sample


-------
Method 630

into the 2-L hydrolysis flask. Rinse the bottle four times with 100-mL aliquots of reagent
water, adding the washes to the hydrolysis flask. Bring the volume in the hydrolysis
flask to approximately 1500 mL with reagent water.

10.3	Place 5.0 mL of color reagent into the CS2 trap (second gas-washing bottle). Place 9 mL
of zinc acetate solution into the HjS scrubber (first gas washing bottle). Add 2 mL of
ethanol to the HjS scrubber. Place a magnetic stirring bar in the hydrolysis flask and
place the flask on the hotplate/magnetic stirrer (ambient at this time). Assemble the
apparatus providing adequate support for all glassware. The addition funnel stem
opening must be below the water level. Ground-glass joints may be slightly coated with
silicone grease.

10.4	Start the stirrer, begin water flow through the condenser, and turn on hot plate and begin
heating the flask. Open the needle valve slightly and start the aspirator. By closing the
needle valve, adjust the airflow through the absorption train until the proper flow is
attained. (The column of bubbles extends to the bottom of the spherical expansion
chamber at the top of the CS2 trap.) Add 30 mL of decomposition reagent to the flask.

NOTE: The analyst must ensure that the sample pH is less than 2 during hydrolysis.

10.5	Bring the liquid in the flask to a gentle boil. Continue the boiling for 60 minutes, then
remove the heat. Continue aspiration until boiling ceases.

10.6	Transfer the contents of the CS2 trap into a 25.0-mL volumetric flask by forcing the liquid
through the glass frit and out of the inlet arm with pressure from a large pipette bulb.
Ensure quantitative transfer by rinsing the trap three times with ethanol. Bring the
colored solution to volume with ethanol. Mix thoroughly and allow the color to develop
for at least 15 minutes but not more than two hours before determining the absorbance.

10.7	Determine the absorbance of the sample at 435 nm and 380 nm using a 1-cm cell or a 4-
cm cell as necessary. Determine the weight of dithiocarbamate from the appropriate
calibration curve prepared in Section 7.4.

10.8	Determine the original sample volume by refilling the sample bottle to the mark and
transferring the liquid to a 1000-mL graduated cylinder. Record the sample volume to
the nearest 5 mL. If a smaller measured aliquot of sample was used to remain within the
range of the color reagent, this step may be omitted.

11. Calculations

11.1	Determine the concentration of total dithiocarbamates in the sample as ziram directly
from the calibration curve. When a specific dithiocarbamate is being measured,
quantitate in terms of the selected pesticide.

11.2	Report results in micrograms per liter without correction for recovery data. When
duplicate and spiked samples are analyzed, report all data obtained with the sample
results.


-------
Method 630

11.3 For samples processed as part of a set where the laboratory spiked sample recovery falls
outside of the control limits in Section 8.3, data for the affected parameters must be
labeled as suspect.

12. Method Performance

12.1	The method detection limit (MDL) is defined as the minimum concentration of a
substance that can be measured and reported with 99% confidence that the value is above
zero.8 The MDL concentrations listed in Table 1 were determined using wastewater, and
are expressed in concentration units of the spiked materials.1

12.2	In a single laboratory, Environmental Science and Engineering, using spiked wastewater
samples, the average recoveries presented in Table 1 were obtained. The percent
standard deviation of the recovery is also included in Table l.1 All recoveries are based
on calibrations using the specific dithiocarbamate instead of ziram.


-------
Method 630

References

1.	"Pesticides Methods Development," Report for EPA Contract 68-03-2897 (in preparation).

2.	ASTM Annual Book of Standards, Part 31, D3694, "Standard Practice for Preparation of
Sample Containers and for Preservation," American Society for Testing and Materials,
Philadelphia, PA, p. 679, 1980.

3.	"Carcinogens—Working with Carcinogens," Department of Health, Education, and
Welfare, Public Health Service, Center for Disease Control, National Institute for
Occupational Safety and Health, Publication No. 77-206, August 1977.

4.	"OSHA Safety and Health Standards, General Industry" (29 CFR 1910), Occupational
Safety and Health Administration, OSHA 2206 (Revised, January 1976).

5.	"Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
Committee on Chemical Safety, 3rd Edition, 1979.

6.	"Handbook for Analytical Quality Control in Water and Wastewater Laboratories,"
EPA-600/4-79-019, U. S. Environmental Protection Agency, Environmental Monitoring
and Support Laboratory - Cincinnati, Ohio 45268, March 1979.

7.	ASTM Annual Book of Standards, Part 31, D3370, "Standard Practice for Sampling
Water," American Society for Testing and Materials, Philadelphia, PA, p. 76, 1980.

8.	Glaser, J.A. et al. "Trace Analysis for Wastewaters," Environmental Science & Technology,
15, 1426 (1981).


-------
Method 630

Table 1. Method Performance
Method



Detection







Mean

Standard



Limit

Sample

Number of

Spike

Recovery

Deviation

Parameter

(Vg/L)

Type*

Replicates

(Vg/L)

(%)

(%)

Maneb

15.3

1

7

31.5

97.1

15.5

Metham

3.7

2

7

20.1

94.5

5.9





3

7

250.0

65.2

2.8

Ziram

1.9

4

8

32.2

100.0

2.0





5

8

1050.0

96.2

10.0

*Sample type:

1	= Municipal wastewater

2	= Mixture of 13% industrial (pesticide manufacturing) wastewater and 87%

municipal wastewater

3	= Industrial wastewater, pesticide manufacturing

4	= Mixture of 40% industrial and 60% municipal wastewater

5	= 7% industrial process water, 7% industrial wastewater, 86% municipal wastewater


-------
Method 630

A52-002-60A

Figure 1.

Dithiocarbamate Hydrolysis Apparatus


-------