Hazardous Waste Support Section
SOP No. HW-34A, Revision 1
SQM02.2
Trace Volatile Data Validation
SOP HW-34A
Revision 1
September, 2016
Approvals:

(LA*£ \—a
I	i.
Narendra Kumar
Chemist, Hazardous Waste Support Section
Li
Date


P^t'ip (iocuzzs
Chief, Hazardous Waste Support Section
Jon Gabry /'
Chief, Hazardous WTaste Support Branch
/z /W/C_
Date
/Msm
Date

-------
SOP HW-34A
Revision 1
September, 2016
NOTICE
The policies and procedures set forth here are intended as guidance to the United States
Environmental Protection Agency (hereafter referred to as USEPA) and other governmental
employees. They do not constitute rule making by USEPA, and may not be relied upon to create
a substantive or procedural right enforceable by any other person. The Government may take
action that is at variance with the policies and procedures in this manual.
The guidance for data validation set forth in the quality assurance project plan (QAPP) for the
project associated with the data in question will always take precedence over the data validation
guidance listed herein.
Validators should note that their professional judgment supersedes any guidance listed in this
document.
This document can be obtained from the USEPA's Region 2 Quality Assurance website at:
http://www.epa.gov/region2/qa/documents.htm
1

-------
SOP HW-34A
Revision 1
September, 2016
TABLE OF CONTENTS
NOTICE 	1
TABLE OF CONTENTS	2
LIST OF TABLES	3
ACRONYMS	4
DATA QUALIFIER DEFINITIONS	7
DATA PACKAGE INSPECTION	7
HWSS DATA VALIDATION PROCESS	8
PRELIMINARY REVIEW	9
Preservation	10
Gas Chromatograph/Mass Spectrometer (GC/MS) Instrument Performance Check	11
Initial Calibration	12
Continuing Calibration Verification (CCV)	16
Blanks	18
Deuterated Monitoring Compounds (DMCs)	21
Matrix Spike/Matrix Spike Duplicates (MS/MSDs)	24
Internal Standards	25
Standards Data	27
Target Compound Identification	28
Tentatively Identified Compounds (TICs)	29
Compounds Quantitation and Reported Contract Required Quantitation Limits
(CRQLs)	30
Field Duplicates	31
System Performance	32
Regional Quality Assurance (QA) and Quality Control (QC)	33
Overall Assessment of Data	34
APPENDIX A: GLOSSARY	35
APPENDIX B: ORGANIC DATA EXECUTIVE NARRATIVE TEMPLATE	38
APPENDIX C: SAMPLE ORGANIC DATA SAMPLE SUMMARY	39
APPENDIX D: ELECTRONIC DATA DELIVERABLE TEMPLATE	40
2

-------
SOP HW-34A
Revision 1
September, 2016
LIST OF TABLES
Table 1. Holding Time Actions for Trace Volatile Analyses	10
Table 2. RRF, %RSD, and %D Acceptance Criteria in Initial Calibration and CCV for Trace
Volatile Analysis	12
Table 3. Initial Calibration Actions for Trace Volatiles Analyses	15
Table 4. Continuing Calibration Verification (CCV) Actions for Trace Volatiles Analyses	17
Table 5. Blank Actions for Trace Volatiles Analyses	20
Table 6. Volatile Deuterated Monitoring Compounds (DMCs) and Recovery Limits	21
Table 7. Deuterated Monitoring Compound (DMC) Recovery Actions for Trace Volatiles
Analyses	22
Table 8. Volatile Deuterated Monitoring Compounds (DMCs) and the Associated Target
Compounds	22
Table 9. Internal Standard Actions for Trace Volatiles Analyses	26
3

-------
SOP HW-34A
Revision 1
September, 2016
ACRONYMS
%D
Percent Difference
%RSD
Percent Relative Standard Deviation
ARO
Aroclor
ASB
Analytical Services Branch
BFB
Bromofluorob enzene
CCS
Contract Compliance Screening
CCV
Continuing Calibration Verification
CF
Calibration Factor
CLP
Contract Laboratory Program
CLP PO
Contract Laboratory Program Project Officer
COR
Contracting Officer Representative
CRQL
Contract Required Quantitation Limit
CSF
Complete SDGFile
DART
Data Assessment Rapid Transmittal
DAT
Data Assessment Tool
DCB
Decachlorobiphenyl
DFTPP
Decafluorotriphenylphosphine
DMC
Deuterated Monitoring Compound
DQA
Data Quality Assessment
DQO
Data Quality Objective
EDD
Electronic Data Deliverable
EDM
EXES Data Manager
ESAT
Environmental Services Assistance Team
EXES
Electronic Data Exchange and Evaluation System
GC
Gas Chromatograph
GC/ECD
Gas Chromatograph/Electron Capture Detector
GC/MS
Gas Chromatograph/Mass Spectrometer
GPC
Gel Permeation Chromatography
HWSS
Hazardous Waste Support Section
INDA
Individual Standard Mixture A
INDB
Individual Standard Mixture B
INDC
Individual Standard Mixture C
LCS
Laboratory Control Sample
MS
Matrix Spike
MSD
Matrix Spike Duplicate
OSRTI
Office of Superfund Remediation and Technology Innovation
PCBs
Polychlorinated Biphenyls
PE
Performance Evaluation
PEM
Performance Evaluation Mixture
QA
Quality Assurance
QAC
Quality Assurance Coordinator
QAPP
Quality Assurance Project Plan
QC
Quality Control
4

-------
SOP HW-34A
Revision 1
September, 2016
RAS
Routine Analytical Services
RIC
Reconstructed Ion Chromatogram
RPD
Relative Percent Difference
RRF
Relative Response Factor
RRF
Mean Relative Response Factor
RRT
Relative Retention Time
RSCC
Regional Sample Control Center Coordinator
RSD
Relative Standard Deviation
RT
Retention Time
SAP
Sampling and Analysis Plan
SCP
Single Component Pesticide
SDG
Sample Delivery Group
SIM
Selected Ion Monitoring
SMO
Sample Management Office
SOP
Standard Operating Procedure
SOW
Statement of Work
TCL
Target Compound List
TCLP
Toxicity Characteristics Leachate Procedure
TCX
Tetrachloro-m-xylene
TIC
Tentatively Identified Compound
TOPO
Task Order Project Officer
TR/COC
Traffic Report/Chain of Custody Record
USEPA
United States Environmental Protection Agency
UV
Ultraviolet
VTSR
Validated Time of Sample Receipt
5

-------
SOP HW-34A
Revision 1
September, 2016
INTRODUCTION
This document is designed to offer the data reviewer guidance in determining the validity of
analytical data generated through the USEPA Contract Laboratory Program (CLP) Statement of
Work (SOW) for Multi-Media, Multi-Concentration Organics Analysis (SOM02.2), and any
future editorial revisions of SOM02.2, hereinafter referred to as the SOM02.2 SOW. This
guidance is somewhat limited in scope and is intended to be used as an aid in the formal
technical review process.
The guidelines presented in the document will aid the data reviewer in establishing (a) if data
meets the specific technical and QC criteria established in the SOW, and (b) the validity and
extent of bias of any data not meeting the specific technical and QC criteria established in the
SOW. It must be understood by the reviewer that acceptance of data not meeting technical
requirements is based upon many factors, including, but not limited to site-specific technical
requirements, the need to facilitate the progress of specific projects, and availability for re-
sampling.
The reviewer should note that while this document is to be used as an aid in the formal data
review process, other sources of guidance and information, as well as professional judgment,
should also be used to determine the ultimate validity of data, especially in those cases where all
data does not meet specific technical criteria.
6

-------
SOP HW-34A
Revision 1
September, 2016
DATA QUALIFIER DEFINITIONS
The following definitions provide brief explanations of the national qualifiers assigned to results
in the data review process.
U
The analyte was analyzed for, but was not detected above the level of the reported
sample quantitation limit.
J
The result is an estimated quantity. The associated numerical value is the
approximate concentration of the analyte in the sample.
J+
The result is an estimated quantity, but the result may be biased high.
J-
The result is an estimated quantity, but the result may be biased low.
NJ
The analysis indicates the presence of an analyte that has been "tentatively
identified" and the associated numerical value represents its approximate
concentration.
UJ
The analyte was analyzed for, but was not detected. The reported quantitation
limit is approximate and may be inaccurate or imprecise.
R
The data are unusable. The sample results are rejected due to serious deficiencies
in meeting Quality Control (QC) criteria. The analyte may or may not be present
in the sample.
DATA PACKAGE INSPECTION
For data obtained through the Contract Laboratory Program (CLP), the EXES Data Manager
(EDM) is a useful tool in the data review process. For more information about EDM, please refer
to the following Sample Management Office (SMO) website:
https://epasmoweb.fedcsc.com/help/guides/Submit%20and%20Inspect%20Data%20Quick%20G
uide%20%28EXES%29.pdf
EDM will identify any missing and/or incorrect information in the data package. The CLP
laboratory may submit a reconciliation package for any missing items or to correct data.
If there are any concerns regarding the data package, contact the laboratory COR (CLP COR)
from the Region where the samples were taken. For personnel contact information, please refer
to the following CLP website:
http://www.epa.gov/superfund/programs/clp/contacts.htm
7

-------
SOP HW-34A
Revision 1
September, 2016
HWSS DATA VALIDATION PROCESS
After downloading the data package from EDM, the data validator will use the recommendations
in this SOP as well as their own professional judgment to validate the data.
The data will be saved in the following location, under the appropriate case number folder:
G:\DESADIV\HWSS\DATA VALIDATION
The file naming conventions will consist of
A.	case number	i.e., 12345
B.	SDG name	i.e., BXY12
C.	level of validation performed	i.e., S3VE
Examples: 12345_BXY12_S3VE.xls
12345 BXY12 S3VEM.xls
When data validation is completed, the data package is uploaded for the client to download from
the HWSS data delivery website.
The completed data package includes the Executive Narrative (see Appendix B for template), the
Sample Summary Report (see Appendix C for example), and the Electronic Data Deliverable
(EDD) (see Appendix D for an example list of the column headers included in this document).
Additional deliverables per modified analysis request and QAPP are also included.
All data is initially marked as "reportable" (Y) in the EDM before validation is begun.
Sometimes, due to dilutions, re-analysis, or SIM/scan runs all being performed, there will be
multiple results for a single sample. The following criteria and professional judgment are used to
determine which results should be reported:
Analysis with a lower CRQL
The analysis with a better QC results
The analysis with a higher result
The analysis values and their respective CRQLs are then transferred to a single sample run.
Other runs which are not being used are updated as "Not Reportable" or (N) in the EDM.
8

-------
SOP HW-34A
Revision 1
September, 2016
PRELIMINARY REVIEW
This document is for the review of analytical data generated through the SOM02.2 SOW and any
future editorial revisions of SOM02.2 for USEPA Region 2. To use this document effectively,
the reviewer should have an understanding of the analytical method and a general overview of
the Sample Delivery Group (SDG) or sample Case at hand. The exact number of samples, their
assigned numbers, their matrix, and the number of laboratories involved in the analysis are
essential information.
It is suggested that an initial review of the data package be performed, taking into consideration
all information specific to the sample data package [e.g., Modified Analysis Requests, Traffic
Report/Chain of Custody (TR/COC) documentation, SDG Narratives, etc.].
The reviewer should also have a copy of the Quality Assurance Project Plan (QAPP) or similar
document for the project for which the samples were analyzed. The criteria for data validation
outlined in the QAPP supersede this Standard Operating Procedure. The reviewer should contact
the appropriate Laboratory COR to obtain copies of the QAPP and relevant site information.
This information is necessary in determining the final usability of the analytical data.
The SDGs or Cases routinely have unique samples that require special attention from the
reviewer. These include field blanks and trip blanks, field duplicates, and Performance
Evaluation (PE) samples which must be identified in the sampling records. The sampling records
(e.g., TR/COC records, field logs, and/or contractor tables) should identify:
1.	The Region where the samples were taken,
2.	The Case number,
3.	The complete list of samples with information on:
a.	Sample matrix;
b.	Field blanks (i.e., equipment blanks or rinsate blanks) and trip blanks;
c.	Field duplicates;
d.	Field spikes;
e.	QC audit samples;
f.	Shipping dates;
g.	Preservatives; and
h.	Laboratories involved.
The TR/COC documentation includes sample descriptions and date(s) of sampling. The reviewer
must consider lag times between sampling and start of analysis when assessing technical sample
holding times.
The laboratory's SDG Narrative is another source of general information. Notable problems with
matrices, insufficient sample volume for analysis or reanalysis, samples received in broken
containers, preservation, and unusual events should be documented in the SDG Narrative. The
reviewer should also inspect any email or telephone/communication logs detailing any
discussion of sample or analysis issues between the laboratory, the CLP Sample Management
Office (SMO), and USEPA Region 2.
9

-------
SOP HW-34A
Revision 1
September, 2016
Preservation
Action:
1.	Qualify sample results using preservation and technical holding time information as
follows (see Table 1):
a.	If there is no evidence that the samples were properly preserved (pH < 2, T =
> 6.0°C), but the samples were analyzed within the technical holding time [7
days from sample collection], qualify detects as estimated (J) and non-detects
as estimated (UJ).
b.	If there is no evidence that the samples were properly preserved, and the
samples were analyzed outside of the technical holding time [7 days from
sample collection], qualify detects for all volatile compounds as estimated (J)
and non-detects as unusable (R).
c.	If the samples were properly preserved, and the samples were analyzed
within the technical holding time [14 days from sample collection], no
qualification of the data is necessary.
d.	If the samples were properly preserved, but were analyzed outside of the
technical holding time [14 days from sample collection], qualify detects as
estimated (J) and non-detects as unusable (R).
2.	Whenever possible, the reviewer should comment on the effect of the holding time
exceedance on the resulting data in the Data Review Narrative.
3.	Use professional judgment to qualify samples whose temperature upon receipt at the
laboratory is either below 2° C or above 6.0° C.
4.	If air bubbles were present in the sample vial used for analysis, qualify detected
compounds as estimated (J) and non-detected compounds as estimated (UJ).
5.	Note, for Contract Laboratory COR action, when technical holding times are exceeded.
Table 1. Holding Time Actions for Trace Volatile Analyses
Matrix
Preserved
Criteria
Action
Detected
Associated
Compounds
Non-Detected
Associated Compounds
Aqueous
No
< 7 days
J
UJ
Aqueous
No
> 7 days
J
R
Aqueous
Yes
<14 days
No qualification
Aqueous
Yes
>14 days
J R
Aqueous
Samples^ 6.0°C or <
2°C upon arrival in the
laboratory
Professional Judgment
10

-------
SOP HW-34A
Revision 1
September, 2016
Gas Chromatograph/Mass Spectrometer (GC/MS) Instrument Performance Check
Action:
NOTES: This requirement does not apply when samples are analyzed by the Selected Ion
Monitoring (SIM) technique.
All mass spectrometer instrument conditions must be identical to those used
during the sample analysis. Background subtraction actions resulting in spectral
distortions for the sole purpose of meeting the method specifications are contrary
to the Quality Assurance (QA) objectives, and are therefore unacceptable.
No data should be qualified based on BFB or DFTTP failure. Instances of this
should be noted in the narrative and professional judgement should be used.
All ion abundance ratios must be normalized to m/z 95, the nominal base peak,
even though the ion abundance of m/z 174 may be up to 120% that of m/z 95.
1.	If samples are analyzed without a preceding instrument performance check, qualify all
data in those samples as unusable (R).
2.	If the laboratory has made minor transcription errors which do not significantly affect the
data, the data reviewer should make the necessary corrections on a copy of the form.
3.	If the laboratory has failed to provide the correct forms or has made significant
transcription or calculation errors, the Region's designated representative should contact
the laboratory and request corrected data. If the information is not available, the reviewer
must use professional judgment to assess the data. Notify the laboratory's Contract
Laboratory Program Project Officer (CLP PO).
4.	If ion abundance criteria are not met, professional judgment may be applied to determine
to what extent the data may be utilized. When applying professional judgment to this
topic, the most important factors to consider are the empirical results that are relatively
insensitive to location on the chromatographic profile and the type of instrumentation.
Therefore, the critical ion abundance criteria for BFB are the m/z 95/96, 174/175,
174/176, and 176/177 ratios. The relative abundances of m/z 50 and 75 are of lower
importance. This issue is more critical for Tentatively Identified Compounds (TICs).
5.	Note, in the Data Review Narrative, decisions to use analytical data associated with BFB
instrument performance checks not meeting contract requirements.
6.	If the reviewer has reason to believe that instrument performance check criteria were
achieved using techniques other than those described in Trace Volatiles Organic
Analysis, Section II.D.5 of the NFG SOM02.2, obtain additional information on the
instrument performance checks. If the techniques employed are found to be at variance
with the contract requirements, the performance and procedures of the laboratory may
merit evaluation. Note, for laboratory COR action, concerns or questions regarding
laboratory performance. For example, if the reviewer has reason to believe that an
inappropriate technique was used to obtain background subtraction (such as background
subtracting from the solvent front or from another region of the chromatogram rather than
from the BFB peak), note this for laboratory COR action.
7.	Use professional judgment to determine whether associated data should be qualified
based on the spectrum of the mass calibration compound.
11

-------
SOP HW-34A
Revision 1
September, 2016
Initial Calibration
1.	ICAL should be performed at the specified frequency and sequence. Each GC/MS system
must be calibrated with a minimum of five concentrations to determine instrument
sensitivity and the linearity of GC/MS response for the purgeable target analytes and
Deuterated Monitoring Compounds (DMCs).
2.	ICAL standards must be analyzed prior to any analysis of samples and required blanks
and within 12 hours of the associated instrument performance check at the beginning of
each analytical sequence, or as necessary if the CCV acceptance criteria are not met.
3.	ICAL standards must contain all required target analytes and DMCs at concentrations of
0.50, 1.0, 5.0, 10, and 20 [j,g/L for non-ketones, and 5.0, 10, 50, 100, and 200 [j,g/L for
ketones.
4.	All three xylene isomers (o-, m-, and p-xylene) must be present in calibration standards.
Concentrations for o-xylene must be at 0.50, 1.0, 5.0, 10, and 20 (J,g/L, while the total
concentrations of the m- plus the p-xylene isomers must be at 0.50, 1.0, 5.0, 10, and 20
Hg/L
5.	The Relative Response Factor (RRF), mean RRF, and Percent Relative Standard
Deviation (%RSD) must be calculated for each target analyte and DMC according to the
SOW.
6.	The RRF for each target analyte and DMC in each ICAL standard must be > Minimum
RRF value in Table 2.
7.	The %RSD of the ICAL RRF for each target analyte and DMC must be < Maximum
%RSD value in Table 2.
NOTE: The technical acceptance criteria specified in a "Request for Quote (RFQ) for Modified
Analysis" may impact some of the preceding evaluation criteria. A copy of this document should
be present in the SDG, when applicable. Refer to the CLP home page at
http://www.epa.gov/oerrpage/superfund/programs/clp/modifiedanalyses.htm for the specific
method flexibility requirements.
Table 2. RRF, %RSD, and %D Acceptance Criteria in Initial Calibration and CCV for
Trace Volatile Analysis
Analyte
Minimum
RRF
Maximum
%RSD
Opening
Maximum
%D'
Closing
Maximum
%D
Dichlorodifluoromethane
0.010
30.0
±40.0
±50.0
Chloromethane
0.010
30.0
±30.0
±50.0
Vinyl chloride
0.010
30.0
±30.0
±50.0
Bromomethane
0.010
40.0
±30.0
±50.0
Chloroethane
0.010
30.0
±30.0
±50.0
Trichlorofluoromethane
0.010
30.0
±30.0
±50.0
1,1 -Dichloroethene
0.020
30.0
±20.0
±25.0
1,1,2-Trichloro-1,2,2-trifluoroethane
0.010
30.0
±30.0
±50.0
12

-------
SOP HW-34A
Revision 1
September, 2016

Minimum
Maximum
Opening
Closing
Analyte
RRF
%RSD
Maximum
%Di
Maximum
%D
Acetone
0.010
40.0
±40.0
±50.0
Carbon disulfide
0.010
20.0
±25.0
±25.0
Methyl acetate
0.010
40.0
±40.0
±50.0
Methylene chloride
0.010
40.0
±30.0
±50.0
trans-1,2-Dichloroethene
0.070
20.0
±20.0
±25.0
Methyl tert-butyl ether
0.010
30.0
±30.0
±50.0
1,1 -Dichloroethane
0.100
20.0
±20.0
±25.0
cis-l,2-Dichloroethene
0.100
20.0
±20.0
±25.0
2-Butanone
0.010
40.0
±40.0
±50.0
Bromochloromethane
0.020
20.0
±20.0
±25.0
Chloroform
0.040
20.0
±20.0
±25.0
1,1,1 -Trichloroethane
0.050
30.0
±20.0
±25.0
Cyclohexane
0.100
30.0
±25.0
±50.0
Carbon tetrachloride
0.020
20.0
±25.0
±50.0
Benzene
0.300
20.0
±20.0
±25.0
1,2-Dichloroethane
0.010
20.0
±25.0
±50.0
Trichloroethene
0.100
20.0
±20.0
±25.0
Methylcyclohexane
0.200
30.0
±25.0
±50.0
1,2-Dichloropropane
0.100
20.0
±20.0
±25.0
Bromodichloromethane
0.090
20.0
±20.0
±25.0
cis-1,3 -Dichloropropene
0.100
20.0
±20.0
±25.0
4-Methyl-2-pentanone
0.010
30.0
±30.0
±50.0
Toluene
0.400
20.0
±20.0
±25.0
trans -1,3 -Dichloropropene
0.010
30.0
±20.0
±25.0
1,1,2-Trichloroethane
0.040
20.0
±20.0
±25.0
T etrachloroethene
0.100
20.0
±20.0
±25.0
2-Hexanone
0.010
40.0
±40.0
±50.0
Dibromochloromethane
0.050
20.0
±20.0
±25.0
1,2-Dibromoethane
0.010
20.0
±20.0
±25.0
Chlorobenzene
0.400
20.0
±20.0
±25.0
Ethylbenzene
0.500
20.0
±20.0
±25.0
m,p-Xylene
0.200
20.0
±20.0
±25.0
o-Xylene
0.300
30.0
±20.0
±25.0
Styrene
0.200
30.0
±20.0
±25.0
Bromoform
0.010
30.0
±30.0
±50.0
Isopropylbenzene
0.700
30.0
±25.0
±25.0
1,1,2,2-T etrachloroethane
0.050
20.0
±25.0
±25.0
1,3 -Dichlorobenzene
0.500
20.0
±20.0
±25.0
1,4-Dichlorobenzene
0.700
20.0
±20.0
±25.0
1,2-Dichlorobenzene
0.400
20.0
±20.0
±25.0
1,2-Dibromo-3 -chloropropane
0.010
40.0
±40.0
±50.0
1,2,4-Trichlorobenzene
0.300
30.0
±30.0
±50.0
1,2,3 -Trichlorobenzene
0.200
30.0
±40.0
±50.0
13

-------
SOP HW-34A
Revision 1
September, 2016
Analyte
Minimum
RRF
Maximum
%RSD
Opening
Maximum
%Di
Closing
Maximum
%D
Deuterated Monitoring Compound
Vinyl chloride-d3
0.010
30.0
±30.0
±50.0
Chloroethane-ds
0.010
30.0
±30.0
±50.0
1,1 -Dichloroethene-d2
0.010
30.0
±25.0
±25.0
2-Butanone-d5
0.010
40.0
±40.0
±50.0
Chloroform-d
0.010
20.0
±20.0
±25.0
1,2-Dichloroethane-d4
0.010
20.0
±25.0
±25.0
Benzene-d6
0.030
20.0
±20.0
±25.0
1,2-Dichloropropane-d6
0.100
20.0
±20.0
±25.0
Toluene-ds
0.200
20.0
±20.0
±25.0
trans -1,3 -Dichloropropene-d4
0.010
30.0
±25.0
±25.0
1,1,2,2- Tetrachloroethane-d2
0.010
20.0
±25.0
±25.0
l,2-Dichlorobenzene-d4
0.060
20.0
±20.0
±25.0
2-Hexanone-d5
0.01
40.0
±40.0
±50.0
1 If a closing CCV is acting as an opening CCV, all target analytes must meet the requirements for an
opening CCV.
Action:
Qualify all volatile target compounds using the following criteria:
a.	If any volatile target compound has an RRF value less than the minimum
criterion listed in Table 2, use professional judgment for detects, based on
mass spectral identification to qualify the data as estimated (J+).
b.	If any volatile target compound has an RRF value less than the minimum
criterion listed in Table 2 qualify non-detected compounds as unusable (R).
c.	If any of the volatile target compounds has %RSD greater than the maximum
in table 2, qualify detects as estimated (J), and non-detected compounds using
professional judgment (see Action 2).
d.	If the volatile target compounds meet the acceptance criteria for RRF and the
%RSD, no qualification of the data is necessary.
e.	No qualification of the data is necessary on the DMC RRF and %RSD data
alone. Use professional judgment and follow the guidelines in Action 2, to
evaluate the DMC RRF and %RSD data in conjunction with the DMC
recoveries to determine the need for qualification of data.
2. At the reviewer's discretion, and based on the project-specific Data Quality
Objectives (DQOs), a more in-depth review may be considered using the following
guidelines:
a. If any volatile target compound has a %RSD greater than the maximum
criterion in Table 2 and if eliminating either the high or the low-point of the
curve does not restore the %RSD to less than or equal to the required
maximum:
i.	Qualify detects for that compound(s) as estimated (J).
ii.	Qualify non-detected volatile target compounds using professional
judgment.
14

-------
SOP HW-34A
Revision 1
September, 2016
b.	If the high-point of the curve is outside of the linearity criteria (e.g., due to
saturation):
i.	Qualify detects outside of the linear portion of the curve as estimated
(J).
ii.	No qualifiers are required for detects in the linear portion of the curve.
iii.	No qualifiers are required for volatile target compounds that were not
detected.
c.	If the low-point of the curve is outside of the linearity criteria:
i.	Qualify low-level detects in the area of non-linearity as estimated (J).
ii.	No qualifiers are required for detects in the linear portion of the curve.
iii.	For non-detected volatile compounds, use the lowest point of the
linear portion of the curve to determine the new quantitation limit.
3.	If the laboratory has failed to provide adequate calibration information, the Region's
designated representative should contact the laboratory and request the necessary
information. If the information is not available, the reviewer must use professional
judgment to assess the data.
4.	Note in the Data Review Narrative, whenever possible, the potential effects on the
data due to calibration criteria exceedance.
5.	Note, for Laboratory COR action, if calibration criteria are grossly exceeded.
Table 3. Initial Calibration Actions for Trace Volatiles Analyses
Criteria
Action
Detect
Non-detect
Initial Calibration not
performed at specified
frequency and sequence
Use professional judgment
R
Use professional judgment
R
Initial Calibration not
J
UJ
performed at the specified


concentrations


RRF < Minimum RRF in
Table 2 for target analyte
Use professional judgment
J+ orR
R
RRF > Minimum RRF in
Table 2 for target analyte
No qualification
No qualification
%RSD > Maximum %RSD in
Table 2 for target analyte
J
Use professional judgment
%RSD < Maximum %RSD in
Table 2 for target analyte
No qualification
No qualification
15

-------
SOP HW-34A
Revision 1
September, 2016
Continuing Calibration Verification (CCV)
Action:
1.	If a CCV (opening and closing) was not run at the appropriate frequency, qualify data
using professional judgment.
2.	Qualify all volatile target compounds listed in table 2 using the following criteria:
a.	For an opening CCV, if any volatile target compound has an RRF value less than
the minimum stated in table 2 above use professional judgment for detects, based
on mass spectral identification, to qualify the data as estimated (J).
b.	For a closing CCV, if any volatile target compound has an RRF value less than
stated in the table 2 above use professional judgment for detects based on mass
spectral identification to qualify the data as estimated (J).
c.	For an opening CCV, if any volatile target compound has an RRF value less than
the minimum stated in table 2 above, qualify non-detected compounds as
unusable (R).
d.	For a closing CCV, if any volatile target compound has an RRF value less than
the limit stated in table 2 above, qualify non-detected compounds as unusable (R).
e.	For an opening CCV, if the Percent Difference value for any of the volatile target
compounds is outside the limits stated in table 2 above, qualify detects as
estimated (J) and non-detected compounds as estimated (UJ).
f.	For a closing CCV, if the Percent Difference value for any of the volatile target
compounds is outside the limit listed in Table 2, qualify detects as estimated (J)
and non-detected compounds as estimated (UJ).
g.	If the volatile target compounds meet the acceptable criteria for RRF and the
Percent Difference, no qualification of the data is necessary.
h.	No qualification of the data is necessary on the DMC RRF and the Percent
Difference data alone. Use professional judgment to evaluate the DMC RRF and
Percent Difference data in conjunction with the DMC recoveries to determine the
need for qualification of data.
3.	If the laboratory has failed to provide adequate calibration information, the Region's
designated representative should contact the laboratory and request the necessary
information. If the information is not available, the reviewer must use professional
judgment to assess the data.
4.	Note in the Data Review Narrative, whenever possible, the potential effects on the data
due to calibration criteria exceedance.
5.	Note, for Laboratory COR action, if calibration criteria are grossly exceeded.
16

-------
SOP HW-34A
Revision 1
September, 2016
Table 4. Continuing Calibration Verification (CCV) Actions for Trace Volatiles Analyses
Criteria for Opening CCV
Criteria for Closing CCV
Action
Detect
Non-detect
CCV not performed at
required frequency
CCV not performed at
required frequency
Use professional
judgment
R
Use professional
judgment
R
CCV not performed at
specified concentration
CCV not performed at
specified concentration
Use professional
judgment
Use professional
judgment
RRF < the Minimum RRF
in Table 2 for target
analytes
RRF < Minimum RRF in
Table 2 for target analytes
Use professional
judgment
J or R
R
RRF > the Minimum RRF
in Table 2 for target
analytes
RRF > Minimum RRF in
Table 2 for target analytes
No qualification
No qualification
%D outside the Opening
Maximum %D limits in
Table 2 for target analytes
%D outside the Closing
Maximum %D limits in
Table 2 for target analytes
J
UJ
%D within the inclusive
Opening Maximum %D
limits in Table 2 for target
analytes
%D within the inclusive
Opening Maximum %D
limits in Table 2 for target
analytes
No qualification
No qualification
17

-------
SOP HW-34A
Revision 1
September, 2016
Blanks
Action:
NOTES: The concentration of each target compound found in the storage, method, field, or
trip blanks must be less than its CRQL listed in the method, except for methylene
chloride, acetone, and 2-butanone, which must be less than 2x their respective
CRQLs. The concentration of non-target compounds in all blanks must be less
than 0.5 (J,g/L.
Data concerning the field or trip blanks are not evaluated as part of the CCS
process. If field or trip blanks are present, the data reviewer should evaluate this
data in a similar fashion as the method blanks.
"Water blanks, "drill blanks", and "distilled water blanks" are validated like any
other sample and are not used to qualify data. Do not confuse them with the other
QC blanks discussed below.
Action regarding unsuitable blank results depends on the circumstances and origin of the blank.
The method blank, like any other sample in the SDG, must meet the technical acceptance criteria
for sample analysis. In instances where more than one of the same type of blank is associated
with a given sample, qualification should be based upon a comparison with the associated blank
having the highest concentration of a contaminant. Do not correct the results by subtracting any
blank value.
1.	If a volatile compound is found in a method blank, but not found in the sample, no
qualification of the data is necessary (Table 5).
2.	If the method , storage, field, or trip blanks contain a listed volatile Target compound(s)
(TCL) at a concentration less than the CRQL (less than 2x the CRQL for methylene
chloride, 2-butanone, and acetone) and:
a.	The sample concentration is less than the CRQL (less than 2x the CRQL for
methylene chloride, 2-butanone, and acetone), report the CRQL value with a "U".
b.	The sample concentration is greater than or equal to the CRQL (greater than or
equal to 2x the CRQL for methylene chloride, 2-butanone, and acetone), and less
than 2x the CRQL (less than 4x the CRQL for methylene chloride, 2-butanone,
and acetone), report the concentration of the compound in the sample and qualify
with a "U".
c.	The sample concentration is greater than or equal to 2x the CRQL (greater than or
equal to 4x the CRQL for methylene chloride, 2-butanone, and acetone), no
qualification of the data is necessary.
3.	If the method, storage, field, or trip blanks contain a volatile TCL compound(s) at a
concentration greater than the CRQL (greater than 2x the CRQL for methylene chloride,
2-butanone, and acetone) and:
a.	The sample concentration is less than the CRQL (less than 2x the CRQL for
methylene chloride, 2-butanone, and acetone), report the CRQL value with a "U".
b.	The sample concentration is greater than or equal to the CRQL (greater than or
equal to 2x the CRQL for methylene chloride, 2-butanone, and acetone), and less
18

-------
SOP HW-34A
Revision 1
September, 2016
than the blank concentration, report the concentration of the compound in the
sample at the same concentration found in the blank and qualify with a "U".
c. The sample concentration is greater than or equal to the CRQL (greater than or
equal to 2x the CRQL for methylene chloride, 2-butanone, and acetone) and
greater than the blank concentration, no qualification is required.
4.	If the method , storage, field, or trip blanks contain a volatile TCL compound(s) at a
concentration equal to the CRQL (equal to 2x the CRQL for methylene chloride, 2-
butanone, and acetone) and:
a.	The sample concentration is less than or equal to the CRQL (less than or equal to
2x the CRQL for methylene chloride, 2-butanone, and acetone), report the CRQL
value with a "U".
b.	The sample concentration is greater than the CRQL (greater than 2x the CRQL
for methylene chloride, 2-butanone, and acetone), no qualification is required.
5.	If gross contamination exists (i.e., blank contamination > 2x the CRQL) in the method,
storage, field, or trip blanks, raise the CRQL to the level of the blank contamination and
report the associated sample data below this level as CRQL-U.
6.	If contaminants are found in the storage, field, or trip blanks, the following is
recommended:
a. Review the associated method blank data to determine if the contaminant(s) was
also present in the method blank.
i.	If the analyte was present at a comparable level in the method blank, the
source of the contamination may be in the analytical system and the action
recommended for the method blank would apply.
ii.	If the analyte was not present in the method blank, the source of
contamination may be in the storage area, in the field, or during sample
transport. Consider all associated samples for possible cross-
contamination.
7.	Tentatively Identified Compounds (TICs) should only be considered if requested.
a. For TICs, if the concentration in the sample is less than five times the
concentration in the most contaminated associated blank (TIC concentration <
5xblank concentration), qualify the sample data as unusable (R).
8.	If an instrument blank was not analyzed following a sample analysis which contained an
analyte(s) at high concentration(s) (i.e., exceeding the calibration range), evaluate the
sample analysis results immediately after the high concentration sample for carryover.
The system is considered uncontaminated if the target analyte is below the CRQL. Use
professional judgment to determine if instrument cross-contamination has affected any
positive compound identification(s). Note, for laboratory COR action, if instrument
cross-contamination is suggested and suspected of having an effect on the sample results.
NOTE: There may be instances where little or no contamination was present in the
associated blanks, but qualification of the sample is deemed necessary. If the
reviewer determines that the contamination is from a source other than the
sample, they should qualify the data. Contamination introduced through dilution
water is one example. Although it is not always possible to determine, instances
of this occurring can be detected when contaminants are found in the diluted
sample result, but are absent in the undiluted sample result.
19

-------
SOP HW-34A
Revision 1
September, 2016
Table 5. Blank Actions for Trace Volatiles Analyses
Blank Type
Blank Result
Sample Result
Action for Samples
Method,
Storage, Field,
Trip,
Instrument***
Detects
Not detected
No qualification required
< CRQL *
< CRQL*
Report CRQL value with a U
> CRQL* and < 2x
the CRQL**
Report concentration of sample
with a U
> 2x the CRQL**
No qualification required
> CRQL *
< CRQL*
Report CRQL value with a U
> CRQL* and <
blank concentration
Report blank value for sample
concentration with a U
> CRQL* and >
blank concentration
No qualification required
= CRQL*
< CRQL*
Report CRQL value with a U
> CRQL*
No qualification required
Gross
contamination **
Detects
Report blank value for sample
concentration with a U
2x the CRQL for methylene chloride, 2-butanone and acetone.
4x the CRQL for methylene chloride, 2-butanone, and acetone.
Qualifications based on instrument blank results affect only the sample analyzed
immediately after the sample that has target compounds that exceed the calibration range
or non-target compounds that exceed 100 (J,g/L.
20

-------
SOP HW-34A
Revision 1
September, 2016
Deuterated Monitoring Compounds (DMCs)
Table 6. Volatile Deuterated Monitoring Compounds (DMCs) and Recovery Limits
DMC
Recovery Limits (%)
Vinyl chloride-d3
40-130
Chloroethane-d5
65-130
1,1 -Dichloroethene-d2
60-125
2-Butanone-d5
40-130
Chloroform-d
70-125
1,2-Dichloroethane-d4
70-130
Benzene-d6
70-125
1,2-Dichloropropane-d6
60-140
Toluene-d8
70-130
trans-l,3-Dichloropropene-d4
55-130
2-Hexanone-d5
45-130
1,1,2,2-Tetrachloroethane-d2
65-120
1,2-Dichlorobenzene-d4
80-120
Action:
NOTES: Recoveries for DMCs in volatile samples and blanks must be within the limits
specified in Table 6.
The recovery limits for any of the compounds listed in Table 6 may be expanded
at any time during the period of performance if USEPA determines that the limits
are too restrictive.
NOTE: Up to three (3) DMCs per sample, excluding SIM analysis, may fail to meet the
recovery limits. As per SOM02.2, any sample which has more than 3 DMCs
outside the limits must be reanalyzed.
Table 8 lists the volatile DMCs and their associated target compounds. If any DMC recovery
in the volatiles fraction is out of specification, qualify the data considering the existence of
interference in the raw data (see Table 7). Considerations include, but are not limited to:
1.	For any recovery greater than the upper acceptance limit:
a.	Qualify detected associated volatile target compounds as estimated high (J+).
b.	Do not qualify non-detected associated volatile target compounds.
2.	For any recovery greater than or equal to 10%, and less than the lower acceptance limit:
a.	Qualify detected associated volatile target compounds as estimated low (J-).
b.	Qualify non-detected associated volatile target compounds as estimated (UJ).
3.	For any recovery less than 10%:
a.	Qualify detected associated volatile target compounds as estimated low (J-).
b.	Qualify non-detected associated volatile target compounds as unusable (R).
4.	For any recovery within acceptance limits, no qualification of the data is necessary.
5.	In the special case of a blank analysis having DMCs out of specification, the reviewer
must give special consideration to the validity of associated sample data. The basic
21

-------
SOP HW-34A
Revision 1
September, 2016
concern is whether the blank problems represent an isolated problem with the blank
alone, or whether there is a fundamental problem with the analytical process. For
example, if one or more samples in the batch show acceptable DMC recoveries, the
reviewer may choose to consider the blank problem to be an isolated occurrence.
However, even if this judgment allows some use of the affected data, note analytical
problems for Laboratory COR action.
6. If more than three DMCs are outside of the recovery limits for trace volatiles analysis and
the sample was not reanalyzed, note under Contract Problems/Non-Compliance.
Table 7. Deuterated Monitoring Compound (DMC) Recovery Actions for Trace Volatiles
Analyses
Criteria
Action

Detect
Non-detect
%R < 10%
J-
R
10% < %R < Lower Acceptance Limit
J-
UJ
Lower Acceptance Limit < %R < Upper
Acceptance Limit
No qualification
No qualification
%R > Upper Acceptance Limit
J+
No qualification
Table 8. Volatile Deuterated Monitoring Compounds (DMCs) and the Associated Target
Compounds
Vinyl chloride-d3(DMC-l)
Chloroethane-ds (DMC-2)
l,l-Dichloroethene-d2 (DMC-3)
Vinyl chloride
Dichlorodifluoromethane
trans-1,2-Dichloroethene

Chloromethane
cis-l,2-Dichloroethene

Bromomethane
1,1 -Dichloroethene

Chloroethane


Carbon disulfide

2-Butanone-ds (DMC-4)
Chloroform-d (DMC-5)
l,2-Dichloroethane-d4 (DMC-6)
Acetone
1,1 -Dichloroethane
Trichlorofluoromethane
2-Butanone
Bromochloromethane
1,1,2-Trichloro-1,2,2-trifluoroethane

Chloroform
Methyl acetate

Dibromochloromethane
Methylene chloride

Bromoform
Methyl-tert-butyl ether
1,1,1 -Trichloroethane
Carbon tetrachloride
1,2-Dibromoethane
1,2-Dichloroethane
Benzene-d6 (DMC-7)
l,2-Dichloropropane-d6 (DMC-8)
Toluene-ds (DMC-9)
Benzene
Cyclohexane
Methylcyclohexane
1,2-Dichloropropane
Bromodichloromethane
Trichloroethene
Toluene
T etrachloroethene
Ethylbenzene
o-Xylene
m,p-Xylene
Styrene
Isopropylbenzene
22

-------
SOP HW-34A
Revision 1
September, 2016
trans-l,3-Dichloropropene-
d4 (DMC-10)
2-Hexanone-ds (DMC-11)
l,l,2,2-Tetrachloroethane-d2
(DMC-12)
cis-1,3 -Dichloropropene
trans -1,3 -Dichloropropene
1,1,2-Trichloroethane
4-Methyl-2-pentanone
2-Hexanone
1,1,2,2,-Tetrachloroethane
1,2-Dibromo-3 -chloropropane
l,2-Dichlorobenzene-d4 (DMC-13)
Chlorobenzene
1,3 -Dichlorobenzene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
1,2,4-Trichlorobenzene
1,2,3 -Trichlorobenzene
23

-------
SOP HW-34A
Revision 1
September, 2016
Matrix Spike/Matrix Spike Duplicates (MS/MSDs)
Action:
NOTES:
NOTE:
Data for MS and MSDs will not be present unless requested by the Region.
Notify the Laboratory COR if a field or trip blank was used for the MS and MSD.
For a Matrix Spike that does not meet criteria, apply the action to only the field
sample used to prepare the Matrix Spike sample. If it is clearly stated in the data
validation materials that the samples were taken through incremental sampling or
some other method guaranteeing the homogeneity of the sample group, then the
entire sample group may be qualified.
1. No qualification of the data is necessary on MS and MSD data alone. However, using
professional judgment, the validator may use the MS and MSD results in conjunction
with other QC criteria and determine the need for some qualification of the data.
24

-------
SOP HW-34A
Revision 1
September, 2016
Internal Standards
Action:
1.	If an internal standard area count for a sample or blank is greater than 140%of the
area for the associated standard (opening CCV or mid-point standard from initial
calibration) (see Table 9):
a.	Qualify detects for compounds quantitated using that internal standard as
estimated low (J-).
b.	Do not qualify non-detected associated compounds.
2.	If an internal standard area count for a sample or blank is less than 60% of the area
for the associated standard (opening CCV or mid-point standard from initial
calibration):
a.	Qualify detects for compounds quantitated using that internal standard as
estimated high (J+).
b.	Qualify non-detected associated compounds as unusable (R).
3.	If an internal standard area count for a sample or blank is greater than or equal to
60%, and less than or equal to 140% of the area for the associates standard opening
CCV or mid-point standard from initial calibration, no qualification of the data is
necessary.
4.	If an internal standard RT varies by more than 10.0 seconds: Examine the
chromatographic profile for that sample to determine if any false positives or
negatives exist. For shifts of a large magnitude, the reviewer may consider partial or
total rejection of the data for that sample fraction. Detects should not need to be
qualified as unusable (R) if the mass spectral criteria are met.
5.	If an internal standard RT varies by less than or equal to 10.0 seconds, no
qualification of the data is necessary.
6.	Note, for Laboratory COR action, if the internal standard performance criteria are
grossly exceeded. Note in the Data Review Narrative potential effects on the data
resulting from unacceptable internal standard performance.
25

-------
SOP HW-34A
Revision 1
September, 2016
Table 9. Internal Standard Actions for Trace Volatiles Analyses
Criteria
Action
Detect
Non-detect
Area response < 20% of the
opening CCV or mid-point
standard CS3 from initial
calibration
J+
R
20% < Area response <
50% of the opening CCV or
mid-point standard CS3
from initial calibration
J+
UJ
50%) < Area response <
200%) of the opening CCV
or mid-point standard CS3
from initial calibration
No qualification
No qualification
Area response > 200%> of
the opening CCV or mid-
point standard CS3 from
initial calibration
J-
No qualification
RT shift between
sample/blank and opening
CCV or mid-point standard
CS3 from initial calibration
>10.0 seconds
R
R
* For volatile compounds associated to each internal standard, see Table 3 - Trace Volatile
Target Compounds and Deuterated Monitoring Compounds with Corresponding Internal
Standards for Quantitation in SOMQ2.2. Exhibit D. available at:
http ://www. epa. gov/ superfund/programs/clp/som2.htm
** Examine the chromatographic profile for that sample to determine if any false positives or
negatives exist. For shifts of a large magnitude, the reviewer may consider partial or total
rejection of the data for that sample fraction. Detects should not need to be qualified as
unusable (R) if the mass spectral criteria are met.
26

-------
SOP HW-34A
Revision 1
September, 2016
Standards Data
Action:
If any calibration standards data are missing, contact the laboratory COR to obtain an
explanation/resubmittal from the lab. If missing deliverables are unavailable, document the effect
in the Data Assessment.
27

-------
SOP HW-34A
Revision 1
September, 2016
Target Compound Identification
Action:
1.	The application of qualitative criteria for GC/MS analysis of target compounds requires
professional judgment. It is up to the reviewer's discretion to obtain additional
information from the laboratory. If it is determined that incorrect identifications were
made, qualify all such data as unusable (R).
2.	Use professional judgment to qualify the data if it is determined that cross-contamination
has occurred.
3.	Note in the Data Review Narrative any changes made to the reported compounds or
concerns regarding target compound identifications. Note, for Laboratory COR action,
the necessity for numerous or significant changes.
28

-------
SOP HW-34A
Revision 1
September, 2016
Tentatively Identified Compounds (TICs)
Action:
NOTE: Tentatively identified compounds should only be evaluated when requested by a
party from outside of the Hazardous Waste Support Section (HWSS).
1.	Qualify all TIC results for which there is presumptive evidence of a match (e.g. greater
than or equal to 85% match) as tentatively identified (NJ), with approximated
concentrations.
2.	General actions related to the review of TIC results are as follows:
a.	If it is determined that a tentative identification of a non-target compound is
unacceptable, change the tentative identification to "unknown" or another
appropriate identification, and qualify the result as estimated (J).
b.	If all contractually-required peaks were not library searched and quantitated, the
Region's designated representative may request these data from the laboratory.
3.	In deciding whether a library search result for a TIC represents a reasonable
identification, use professional judgment. If there is more than one possible match, report
the result as "either compound X or compound Y". If there is a lack of isomer specificity,
change the TIC result to a nonspecific isomer result (e.g., 1,3,5-trimethyl benzene to
trimethyl benzene isomer) or to a compound class (e.g., 2-methyl, 3-ethyl benzene to a
substituted aromatic compound).
4.	The reviewer may elect to report all similar compounds as a total (e.g., all alkanes may be
summarized and reported as total hydrocarbons).
5.	Target compounds from other fractions and suspected laboratory contaminants should be
marked as "non-reportable".
6.	Other Case factors may influence TIC judgments. If a sample TIC match is poor, but
other samples have a TIC with a valid library match, similar RRT, and the same ions,
infer identification information from the other sample TIC results.
7.	Note in the Data Review Narrative any changes made to the reported data or any
concerns regarding TIC identifications.
8.	Note, for the Laboratory COR action any failure to properly evaluate and report TICs.
29

-------
SOP HW-34A
Revision 1
September, 2016
Compounds Quantitation and Reported Contract Required Quantitation Limits (CRQLs)
Action:
1.	When a sample is analyzed at more than one dilution, the lowest CRQLs are used unless
a QC exceedance dictates the use of the higher CRQLs from the diluted sample. Replace
concentrations that exceed the calibration range in the original analysis by replacing the
"E" and its corresponding value on the original Form I and substituting the data from the
diluted sample.
2.	If any discrepancies are found, the Region's designated representative may contact the
laboratory to obtain additional information that could resolve any differences. If a
discrepancy remains unresolved, the reviewer must use professional judgment to decide
which value is the most accurate. Under these circumstances, the reviewer may determine
that qualification of data is warranted. Note in the Data Review Narrative a description of
the reasons for data qualification and the qualification that is applied to the data.
3.	Note, for laboratory COR action, numerous or significant failures to accurately quantify
the target compounds or to properly evaluate and adjust CRQLs.
4.	Results between MDL and CRQL should be qualified as estimated "J".
5.	Results < MDL should be reported at the CRQL and qualified "U". MDLs themselves are
not reported.
30

-------
SOP HW-34A
Revision 1
September, 2016
Field Duplicates
Action:
NOTE: In the absence of QAPP guidance for validating data from field duplicates, the
following action will be taken.
Identify which samples within the data package are field duplicates. Estimate the relative percent
difference (RPD) between the values for each compound. Use professional judgment to note
large RPDs (> 50%) in the narrative.
31

-------
SOP HW-34A
Revision 1
September, 2016
System Performance
Action:
Use professional judgment to qualify the data if it is determined that system performance has
degraded during sample analyses. Note, for Contract Laboratory Program Project Officer (CLP
PO) action, any degradation of system performance which significantly affected the data.
32

-------
SOP HW-34A
Revision 1
September, 2016
Regional Quality Assurance (OA) and Quality Control (OC)
Action:
Any action must be in accordance with Regional specifications and the criteria for acceptable PE
sample results. Note, for the Laboratory COR action any unacceptable results for PE samples.
33

-------
SOP HW-34A
Revision 1
September, 2016
Overall Assessment of Data
Action:
1.	Use professional judgment to determine if there is any need to qualify data which were
not qualified based on the Quality Control (QC) criteria previously discussed.
2.	Write a brief narrative to give the user an indication of the analytical limitations of the
data. Note, for the Laboratory COR action, any inconsistency of the data with the Sample
Delivery Group (SDG) Narrative. If sufficient information on the intended use and
required quality of the data is available, the reviewer should include their assessment of
the usability of the data within the given context. This may be used as part of a formal
Data Quality Assessment (DQA).
34

-------
SOP HW-34A
Revision 1
September, 2016
APPENDIX A: GLOSSARY
Analyte — The element of interest, ion, or parameter an analysis seeks to determine.
Analytical Services Branch (ASB) — Directs the Contract Laboratory Program (CLP) from
within the Office of Superfund Remediation and Technical Innovation (OSRTI) in the Office of
Solid Waste and Emergency Response (OSWER).
Analytical Sample — Any solution or media introduced into an instrument on which an analysis
is performed excluding instrument calibration, Initial Calibration Verification (ICV), Initial
Calibration Blank (ICB), Continuing Calibration Verification (CCV), and Continuing Calibration
Blank (CCB). Note that the following are all defined as analytical samples: undiluted and diluted
samples (USEPA and non-USEPA); Matrix Spike samples; duplicate samples; serial dilution
samples, analytical (post-digestion/post-distillation) spike samples; Interference Check Samples
(ICSs); Laboratory Control Samples (LCSs); and Preparation Blanks.
Associated Samples — Any sample related to a particular Quality Control (QC) analysis. For
example, for Initial Calibration Verification (ICV), all samples run under the same calibration
curve. For duplicates, all Sample Delivery Group (SDG) samples digested/distilled of the same
matrix.
Blank — A sample designed to assess specific sources of contamination. See individual
definitions for types of blanks.
Calibration — The establishment of an analytical curve based on the absorbance, emission
intensity, or other measured characteristic of known standards. The calibration standards are to
be prepared using the same type of reagents or concentration of acids as used in the sample
preparation.
Calibration Blank — A blank solution containing all of the reagents in the same concentration
as those used in the analytical sample preparation. This blank is not subject to the preparation
method.
Calibration Curve — A plot of instrument response versus concentration of standards.
Calibration Standards — A series of known standard solutions used by the analyst for
calibration of the instrument (i.e., preparation of the analytical curve). The solutions may or may
not be subjected to the preparation method, but contain the same matrix (i.e., the same amount of
reagents and/or preservatives) as the sample preparations to be analyzed.
Case — A finite, usually predetermined number of samples collected over a given time period
from a particular site. Case numbers are assigned by the Sample Management Office (SMO). A
Case consists of one or more Sample Delivery Groups (SDGs).
Contract Compliance Screening (CCS) — A screening of electronic and hardcopy data
deliverables for completeness and compliance with the contract. This screening is performed
under USEPA direction by the Contract Laboratory Program (CLP) Sample Management Office
(SMO) contractor.
Continuing Calibration Verification (CCV) — A single parameter or multi-parameter standard
solution prepared by the analyst and used to verify the stability of the instrument calibration with
time, and the instrument performance during the analysis of samples. The CCV can be one of the
calibration standards. However, all parameters being measured by the particular system must be
represented in this standard and the standard must have the same matrix (i.e., the same amount of
reagents and/or preservatives) as the samples.
35

-------
SOP HW-34A
Revision 1
September, 2016
Contract Laboratory Program (CLP) — Supports the USEPA's Superfund effort by providing
a range of state-of-the-art chemical analytical services of known quality. This program is
directed by the Analytical Services Branch (ASB) of the Office of Superfund Remediation and
Technical Innovation (OSRTI) of USEPA.
Laboratory COR — The Regional USEPA official responsible for monitoring laboratory
performance and/or requesting analytical data or services from a CLP laboratory.
Contract Required Quantitation Limit (CRQL) — Minimum level of quantitation acceptable
under the contract Statement of Work (SOW).
Duplicate — A second aliquot of a sample that is treated the same as the original sample in order
to determine the precision of the method.
Field Blank — Any sample that is submitted from the field and identified as a blank. A field
blank is used to check for cross-contamination during sample collection, sample shipment, and in
the laboratory. A field blank includes trip blanks, rinsate blanks, bottle blanks, equipment blanks,
preservative blanks, decontamination blanks, etc.
Field Duplicate — A duplicate sample generated in the field, not in the laboratory.
Holding Time — The maximum amount of time samples may be held before they are processed.
Contractual — The maximum amount of time that the Contract Laboratory Program (CLP)
laboratory may hold the samples from the sample receipt date until analysis and still be in
compliance with the terms of the contract, as specified in the CLP Analytical Services Statement
of Work (SOW). These times are the same or less than technical holding times to allow for
sample packaging and shipping.
Technical — The maximum amount of time that samples may be held from the collection date
until analysis.
Initial Calibration — Analysis of analytical standards for a series of different specified
concentrations to define the quantitative response, linearity, and dynamic range of the instrument
to target analytes.
Initial Calibration Verification (ICV) — Solution(s) prepared from stock standard solutions,
metals, or salts obtained from a source separate from that utilized to prepare the calibration
standards. The ICV is used to verify the concentration of the calibration standards and the
adequacy of the instrument calibration. The ICV should be traceable to National Institute of
Standards and Technology (NIST) or other certified standard sources when USEPA ICV
solutions are not available.
Internal Standard — A non-target element added to a sample at a known concentration after
preparation but prior to analysis. Instrument responses to internal standards are monitored as a
means of assessing overall instrument performance.
Matrix — The predominant material of which the sample to be analyzed is composed. For the
purposes of this document, the matrices are aqueous/water, soil/sediment, wipe, and filter.
Matrix Spike — Introduction of a known concentration of analyte into a sample to provide
information about the effect of the sample matrix on the digestion and measurement
methodology (also identified as a pre-distillation/digestion spike).
Method Detection Limit (MDL) — The concentration of a target parameter that, when a sample
is processed through the complete method, produces a signal with 99 percent probability that it is
different from the blank. For 7 replicates of the sample, the mean value must be 3.14s above the
blank, where "s" is the standard deviation of the 7 replicates.
Narrative (SDG Narrative) — Portion of the data package which includes laboratory, contract,
Case, Sample Number identification, and descriptive documentation of any problems
36

-------
SOP HW-34A
Revision 1
September, 2016
encountered in processing the samples, along with corrective action taken and problem
resolution.
Office of Solid Waste and Emergency Response (OSWER) - The USEPA office that provides
policy, guidance, and direction for the USEPA's solid waste and emergency response programs,
including Superfund.
Percent Difference (%D) — As used in this document and the Statement of Work (SOW), is
used to compare two values. The difference between the two values divided by one of the values.
Performance Evaluation (PE) Sample — A sample of known composition provided by USEPA
for contractor analysis. Used by USEPA to evaluate Contractor performance.
Preparation Blank — An analytical control that contains reagent water and reagents, which is
carried through the entire preparation and analytical procedure.
Relative Percent Difference (RPD) — As used in this document and the Statement of Work
(SOW) to compare two values, the RPD is based on the mean of the two values, and is reported
as an absolute value (i.e., always expressed as a positive number or zero).
Regional Sample Control Center Coordinator (RSCC) — In USEPA Regions, coordinates
sampling efforts and serves as the central point-of-contact for sampling questions and problems.
Also assists in coordinating the level of Regional sampling activities to correspond with the
monthly projected demand for analytical services.
Relative Standard Deviation (RSD) — As used in this document and the Statement of Work
(SOW), the mean divided by the standard deviation, expressed as a percentage.
Sample — A single, discrete portion of material to be analyzed, which is contained in single or
multiple containers and identified by a unique Sample Number.
Sample Delivery Group (SDG) — A unit within a sample Case that is used to identify a group
of samples for delivery. An SDG is defined by the following, whichever is most frequent:
a.	Each 20 field samples [excluding Performance Evaluation (PE) samples] within a
Case; or
b.	Each 7 calendar day period (3 calendar day period for 7-day turnaround) during
which field samples in a Case are received (said period beginning with the receipt
of the first sample in the SDG).
c.	Scheduled at the same level of deliverable.
In addition, all samples and/or sample fractions assigned to an SDG must be scheduled under the
same contractual turnaround time. Preliminary Results have no impact on defining the SDG.
Samples may be assigned to SDGs by matrix (i.e., all soil/sediment samples in one SDG, all
aqueous/water samples in another) at the discretion of the laboratory.
Sample Management Office (SMO) — A contractor-operated facility operated under the SMO
contract, awarded and administered by the USEPA. Provides necessary management, operations,
and administrative support to the Contract Laboratory Program (CLP).
Statement of Work (SOW) — A document which specifies how laboratories analyze samples
under a particular Contract Laboratory Program (CLP) analytical program.
37

-------
SOP HW-34A
Revision 1
September, 2016
APPENDIX B: ORGANIC DATA EXECUTIVE NARRATIVE TEMPLATE
UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
REGION 2
DESAi'HWSB.'HWSS
2&1MK Wood bridge Avenue, Edison, NJ 08837
EXECUTIVE NARRATIVE
Case No, :
Site:|
Number of Samples:
Analysis:
SDG No.:
Laboratory:
Sampling dates:
QAPP
HWSS #:
Contractor Document#:
SUMMARY:
Critical: Results have an unacceptable level of uncertainty and should not be used for making decisions.
Data have been qualified "R" rejected.
Major: A level of uncertainty exists that may not meet the data quality objectives for the project. A bias
is likely to be present in the results. Data has been qualified "J" estimated.
Minor: The level of uncertainty is acceptable. No significant bias in the data was observed.
Critical Findings:
Major Findings:
Minor Findings:
COMMENT:
Reviewer Name(s|:
Approver's Signature:
Name:
Affiliation:
USEPA;R2/HWSBi'HWSS
Page 1 of 3
38

-------
SOP HW-34A
Revision 1
September, 2016
APPENDIX C: SAMPLE ORGANIC DATA SAMPLE SUMMARY
Case No: OCXJOl Coasici. XYZ1234 SDGHo:. XY123 Ljb Ctxl«: 00001
Snap!. Urate: XW23 MbiIkkS: VOAJTrace SSbSk- Water IttHnAr. DIFACT.T
$ara.pjs Locatoa. SCWE'.VHERE CUTTHERE pH: 16 inapb 3ra 12322J5S Sample Tinas: 01.1* 00
°d Monturs "t >oadc

Analyte flame
Result
Units
Dilution Factor
Lab Flag:
Validation
Reportable
Validation Level
DtcKovodiEl'Uoro
methane
C.JO
uf L
1.0
U
U
Yes
S3YEM
C fcilcrcjEseiane
C.50
of L
1.0
u
u
Yes
S5VEM
Vmyl elibode
0.50
uf 1
L0
u
u
Yes
S5VEM
B[omon.«d]>4UM
C.50
Uf L
1.0
u
u
Yes
S5YEM
CMoroetkane
C.50
up L
1.0
II
u
Yes
S5YEM
Tnchk>ro£iioTacn
ethaoa
OJO
uf L
L0
u
u
Yes
S3VEM
1.1-
0,50
up'l
1.0
u
u
Yes
S3VEM
l.l.Mnshlmo-
1.2.2-
tiitluoic ethane
0.50
u| L
1.0
u
u
Yes
S3VEM
Aceton*
5.0
Uf 1
1.0

u
Yes
55YcM
CjubonDssuHil*
C 46
•og.'X.
1.0
I
I
Yes
S5VEM
Metcyl acetate
0 50
me/L
1.0
u
u
Yes
53YEM
Methylene
chlcnde
0 50
ng.'X.
1.0
JB
u
Yes
55YEM
tr3ns-l,2-
Di:hloroetuere
0 50
uf L
1.0
U
u
Yes
S3VEM
Methyl «it-batyl
ethei
0 50
uf L
1.0
u
u
Yes
S3VEM
1.1-
Dizhlorcetiiaiie
0.50
uf L
1.0
u
u
Yes
53VEM
Dickiaroetiiene
1.7
uf L
1.0


Yes
S3VEM
2-Butei;oBe
5.0
uf L
1.0
u
u
'Yes
S3VEM
B romocKloi omet
hane
0 5i">
ttf L
1.0
u
u
Yes
S3VEM
CMomfbrm
T. 6
Uf 1
1.0


Yes
SiVI.il
1,1,1-
IrkcbJorocdazM
0.50
uf L
1.0
u
u
Yes
S3VEM
Cyc'iohmx*n*
0.50
Uf 1
1.0
II
0
Yes
5 3 VIM
Carbon
tetrachloride
0.50
uf L
1.0
u
0
Yes
S3VEM
BeEszeise
0.20
uf L
1.0
J
I
Yes
S3VEM
1,2-
DkMfflroefeane
0.50
uf L
1.0
u
u
Yes
S3VEM
TricM^HsethgEe
0.50
uf L
1.0
u
u
Yes
S3VEM
M»(faTlcTcloh«j
M
0.50
«if L
1.0
u
u
Yes
S3YEM
39

-------
SOP HW-34A
Revision 1
September, 2016
APPENDIX D: ELECTRONIC DATA DELIVERABLE TEMPLATE
DATA PROVIDER
LAB MATRIX CODE
RESULT UNIT
SYS SAMPLE CODE
ANAL LOCATION
DETECTION LIMIT UNIT
SAMPLE NAME
BASIS
TIC RETENTION TIME
SAMPLE MATRIX CODE
CONTAINER ID
RESULT COMMENT
SAMPLE TYPE CODE
DILUTION FACTOR
QC ORIGINAL CONC
SAMPLE SOURCE
PREP METHOD
QC SPIKE ADDED
PARENT SAMPLE CODE
PREP DATE
QC SPIKE MEASURED
SAMPLE DEL GROUP
LEACHATE METHOD
QC SPIKE RECOVERY
SAMPLE DATE
LEACHATE DATE
QC DUP ORIGINAL CONC
SYS LOC CODE
LAB NAME CODE
QC DUP SPIKE ADDED
START DEPTH
QC LEVEL
QC DUP SPIKE MEASURED
END DEPTH
LAB SAMPLE ID
QC DUP SPIKE RECOVERY
DEPTH UNIT
PERCENT MOISTURE
QC RPD
CHAIN OF CUSTODY
SUB SAMPLE AMOUNT
QC SPIKE LCL
SENT TO LAB DATE
SUB SAMPLE AMOUNT UNIT
QC SPIKE UCL
SAMPLE RECEIPT DATE
ANALYST NAME
QC RPD CL
SAMPLER
INSTRUMENT ID
QC SPIKE STATUS
SAMPLING COMPANY CODE
COMMENT
QC DUP SPIKE STATUS
SAMPLING REASON
PRESERVATIVE
QC RPD STATUS
SAMPLING TECHNIQUE
FINAL VOLUME
BREAK 2
TASK CODE
FINAL VOLUME UNIT
SYS SAMPLE CODE
COLLECTION QUARTER
CAS RN
LAB ANL METHOD NAME
COMPOSITE YN
CHEMICAL NAME
ANALYSIS DATE
COMPOSITE DESC
RESULT VALUE
TOTAL OR DISSOLVED
SAMPLE CLASS
RESULT ERROR DELTA
COLUMN NUMBER
CUSTOM FIELD 1
RESULT TYPE CODE
TEST TYPE
CUSTOM FIELD 2
REPORTABLE RESULT
TEST BATCH TYPE
CUSTOM FIELD 3
DETECT FLAG
TEST BATCH ID
COMMENT
LAB QUALIFIERS
CASE
BREAK 1
VALIDATOR QUALIFIERS
CONTRACT NUM
SYS SAMPLE CODE
INTERPRETED QUALIFIERS
SCRIBE SAMPLE ID
LAB ANL METHOD NAME
ORGANIC YN
SAMPLE TIME
ANALYSIS DATE
METHOD DETECTION LIMIT
FRACTION
TOTAL OR DISSOLVED
REPORTING DETECTION LIMIT
PH
COLUMN NUMBER
QUANTITATION LIMIT
DATA VAL LABEL
TEST TYPE


40

-------
REQUEST FOR SOP CHANGE
Initiator
Archana Mirle/Dorina
Date of
8/18/17
Name:
Christina Aliiu
Initiation:


Dept:
ESAT DV
SOP #: HW-34A

Revision #:
1
SOP Title:
SOM02.2 Trace Volatile Data
Pi case Check One
MINOR REVISION
MAJOR REVISION
CHANG E(S)
attachment if necessary);
CHANGE FROM:
f. Preservation 1 a: If there is no evidence that the samples were properly preserved (pi I < 2, T
>6 °C), but the samples were analyzed within the technical holding time [7 days from sample
collection]. Detects J, Non-detects UJ.
TO:
2.	Preservation 1 a: Add- If there is an evidence that the samples were not properly preserved (T
>6 °C). but the samples were extracted or analyzed within the technical holding time [7 days
from sample collection], Detects J, Non-detects UJ.
3.	1 a'. Add : If there is an evidence that the samples were not properly preserved (pH >2), but the
samples were extracted or analyzed within the technical holding time [7 days from sample
collection], no qualification of the data is necessary.
REASON(S) FOR CHANGE(S):
1. To cover all possible conditions.
APPROVAL
NAME:
	j ' 	
Signature/Date
EPA Branch Chief /
Section Chief/Team
Leader
jp/n / t. t-

ESAT Sr. Designee
QA Auditor
Dorina Christina
Alliu
/ /' y ^ -nt yrjr
V ( In-' \ i :.
EPA
TOCOR
Narendra Kumar
M- k - h /
Effective Date

EPArQW*
9-


-------