Method 1624

Revision C

Volatile Organic Compounds by Isotope Dilution

GCMS


-------
Method 1624

Volatile Organic Compounds by Isotope Dilution GCMS

1.	Scope and application

1.1	This method is designed to meet the survey requirements of the USEPA ITD. The method
is used to determine the volatile toxic organic pollutants associated with the Clean Water
Act (as amended 1987); the Resource Conservation and Recovery Act (as amended in
1986); the Comprehensive Environmental Response, Compensation, and Liability Act (as
amended in 1986); and other compounds amenable to purge and trap gas
chromatography/mass spectrometry (GCMS).

1.2	The chemical compounds listed in Tables 1 and 2 may be determined in waters, soils, and
municipal sludges by the method.

1.3	The detection limits of the method are usually dependent on the level of interferences
rather than instrumental limitations. The levels in Table 3 typify the minimum quantities
that can be detected with no interferences present.

1.4	The GCMS portions of the method are for use only by analysts experienced with GCMS
or under the close supervision of such qualified persons. Laboratories unfamiliar with
analysis of environmental samples by GCMS should run the performance tests in
Reference 1 before beginning.

2.	Summary of method

2.1	The percent solids content of the sample is determined. If the solids content is known or
determined to be less than 1%, stable isotopically labeled analogs of the compounds of
interest are added to a 5-mL sample and the sample is purged with an inert gas at 20 to
25°C in a chamber designed for soil or water samples. If the solids content is greater than
one, mL of reagent water and the labeled compounds are added to a 5-aliquot of sample
and the mixture is purged at 40°C. Compounds that will not purge at 20 to 25°C or at
40°C are purged at 75 to 85°C (see Table 2). In the purging process, the volatile
compounds are transferred from the aqueous phase into the gaseous phase where they
are passed into a sorbent column and trapped. After purging is completed, the trap is
backflushed and heated rapidly to desorb the compounds into a gas chromatograph (GC).
The compounds are separated by the GC and detected by a mass spectrometer (MS)
(References 2 and 3). The labeled compounds serve to correct the variability of the
analytical technique.

2.2	Identification of a pollutant (qualitative analysis) is performed in one of three ways: (1)
For compounds listed in Table 1 and other compounds for which authentic standards are
available, the GCMS system is calibrated and the mass spectrum and retention time for
each standard are stored in a user created library. A compound is identified when its
retention time and mass spectrum agree with the library retention time and spectrum. (2)
For compounds listed in Table 2 and other compounds for which standards are not
available, a compound is identified when the retention time and mass spectrum agree
with those specified in this method. (3) For chromatographic peaks which are not
identified by (1) and (2) above, the background corrected spectrum at the peak maximum

2


-------
is compared with spectra in the EPA/NIH mass spectral file (Reference 4). Tentative
identification is established when the spectrum agrees (see Section 12).

2.3	Quantitative analysis is performed in one of four ways by GCMS using extracted ion
current profile (EICP) areas: (1) For compounds listed in Table 1 and other compounds
for which standards and labeled analogs are available, the GCMS system iscalibrated and
the compound concentration is determined using an isotope dilution technique. (2) For
compounds listed in Table 1 and for other compounds for which authentic standards but
no labeled compounds are available, the GCMS system is calibrated and the compound
concentration is determined using an internal standard technique. (3) For compounds
listed in Table 2 and other compounds for which standards are not available, compound
concentrations are determined using known response factors. (4) For compounds for
which neither standards nor known response factors are available, compound
concentration is determined using the sum of the EICP areas relative to the sum of the
EICP areas of the nearest eluted internal standard.

2.4	The quality of the analysis is assured through reproducible calibration and testing of the
purge and trap and GCMS systems.

3


-------
Table 1. Volatile Organic Compounds Determined by GCMS Using Isotope Dilution and Internal
Standard Techniques

Pollutant	Labeled Compound





CAS

EPA







CAS

EPA

Compound

STORET

Registry

EGD

NPDES

Analog

Registry

EGD

Acetone

81552

67-64-1

516

V





d6

666-52-4

616

V

Acrolein

34210

107-02-8

002

V

001

V

d4

33984-05-3

202

V

Acrylonitrile

34215

107-13-1

003

V

002

V

d3

53807-26-4

203

V

Benzene

34030

71-43-2

004

V

003

V

d6

1076-43-3

204

V

Bromodichloromethane

32101

75-27-4

048

V

012

V

13C

93952-10-4

248

V

Bromoform

32104

75-25-2

047

V

005

V

13C

72802-81-4

247

V

Bromomethane

34413

74-83-9

046

V

020

V

d3

1111-88-2

246

V

Carbon tetrachloride

32102

56-23-5

006

V

006

V

13C

32488-50-9

206

V

Chlorobenzene

34301

108-90-7

007

V

007

V

d5

3114-55-4

207

V

Chloroethane

34311

75-00-3

016

V

009

V

d5

19199-91-8

216

V

2-Chloroethylvinyl ether

34576

110-75-8

019

V

010

V









Chloroform

32106

67-66-3

023

V

011

V

13C

31717-44-9

223

V

Chloromethane

34418

74-87-3

045

V

021

V

d3

1111-89-3

245

V

Dibromochloromethane

32105

124-48-1

051

V

008

V

13C

93951-99-6

251

V

1,1 -Dichloroethane

34496

75-34-3

013

V

014

V

d3

56912-77-7

213

V

1,2-Dichloroethane

32103

107-06-2

010

V

015

V

d4

17070-07-0

210

V

1,1 -Dichloroethene

34501

75-35-4

029

V

016

V

d2

22280-73-5

229

V

trans-1,2-Dichlorethene

34546

156-60-5

030

V

026

V

d3

42366-47-2

230

V

1,2-Dichloropropane

34541

78-87-5

032

V

017

V

d6

93952-08-0

232

V

trans-1,3-

34699

10061-02-6

033

V





d4

93951-86-1

233

V

Dichloropropene





















Diethyl ether

81576

60-29-7

515

V





dio

2679-89-2

615

V

p-Dioxane

81582

123-91-1

527

V





d8

17647-74-4

627

V

Ethylbenzene

34371

100-41-4

038

V

019

V

dio

25837-05-2

238

V

Methylene chloride

34423

75-09-2

044

V

022

V

d2

1665-00-5

244

V

Methyl ethyl ketone

81595

78-93-3

514

V





d3

53389-26-7

614

V

1,1,2,2-

34516

79-34-5

015

V

023

V

d2

33685-54-0

215

V

T etrachlor oethane





















T etrachloroethene

34475

127-18-4

085

V

024

V

13c2

32488-49-6

285

V

Toluene

34010

108-88-3

086

V

025

V

d8

2037-26-5

286

V

1,1,1 -T richloroethane

34506

71-55-6

011

V

027

V

d3

2747-58-2

211

V

1,1,2-T richloroethane

34511

79-00-5

014

V

028

V

13C

93952-09-1

214

V

Trichloroethene

39180

79-01-6

087

V

029

V

13c2

93952-00-2

287

V

Vinyl chloride

39175

75-01-4

088

V

031

V

d3

6745-35-3

288

V

4


-------
Table 2. Volatile Organic Compounds to be Determined by Reverse Search and Quantitation
Using Known Retention Times, Response Factors, Reference Compounds, and Mass Spectra

EGD No.	Compound	CAS Registry

532

Allyl alcohol 1

107-18-6

533

Carbon disulfide

75-15-0

534

2-Chloro-1,3-butadiene (Chloroprene)

126-99-8

535

Chloroacetonitrile1

107-14-2

536

3-Chloropropene

107-05-1

537

Crotonaldehyde1

123-73-9

538

1,2-Dibromoethane (EDB)

106-93-3

539

Dibromomethane

74-95-3

540

trans-1,4-Dichloro-2-butene

110-57-6

541

1,3-Dichloropropane

142-28-9

542

cis-1,3-Dichloropropene

10061-01-5

543

Ethyl cyanide 1

107-12-0

544

Ethyl methacrylate

97-63-2

545

2-Hexanone

591-78-6

546

Iodomethane

74-88-4

547

Isobutyl alcohol1

78-83-1

548

Methacrylonitrile

126-98-7

549

Methyl methacrylate

78-83-1

550

4-Methyl-2-pentanone

108-10-1

551

1,1,1,2-T etrachloroethane

630-20-6

552

T richlorofluoromethane

75-69-4

553

1,2,3-Trichloropropane

96-18-4

554

Vinyl acetate

108-05-4

951

m-Xylene

108-38-3

952

o- and p-Xylene



1 Determined at a purge temperature of 75-85°C.

5


-------
3.

Contamination and interferences

3.1	Impurities in the purge gas, organic compounds out-gassing from the plumbing upstream
of the trap, and solvent vapors in the laboratory account for the majority of contamination
problems. The analytical system is demonstrated to be free from interferences under
conditions of the analysis by analyzing reagent water blanks initially and with each
sample batch (samples analyzed on the same 8-hour shift), as described in Section 8.5.

3.2	Samples can be contaminated by diffusion of volatile organic compounds (particularly
methylene chloride) through the bottle seal during shipment and storage. A field blank
prepared from reagent water and carried through the sampling and handling protocol
may serve as a check on such contamination.

3.3	Contamination by carry-over can occur when high level and low level samples are
analyzed sequentially. To reduce carry-over, the purging device (Figure 1 for samples
containing less than one percent solids; Figure 2 for samples containing one percent solids
or greater) is cleaned or replaced with a clean purging device after each sample is
analyzed. When an unusually concentrated sample is encountered, it is followed by
analysis of a reagent water blank to check for carry-over. Purging devices are cleaned by
washing with soap solution, rinsing with tap and distilled water, and drying in an oven
at 100 to 125°C. The trap and other parts of the system are also subject to contamination;
therefore, frequent bakeout and purging of the entire system may be required.

3.4	Interferences resulting from samples will vary considerably from source to source,
depending on the diversity of the site being sampled.

6


-------
Table 3. Gas Chromatography of Purgeable Organic Compounds

Method Detection
Retention Time	Limit4

Minimum Low High

EGD
No.1

Compound

Mean
(sec)

EGD
Ref

Relative 2

Level3

(ub/l)

Solids
(Ug/kg)

Solids
(Ug/kg)

245

Chloromethane-d3

147

181

0.141-0.270

50





345

Chloromethane

148

245

0.922-1.210

50

2077

13

246

Bromomenthane-d3

243

181

0.233-0.423

50





346

Bromomethane

246

246

0.898-1.195

50

00

11

288

Vinyl chloride-d3

301

181

0.286-0.501

50





388

Vinyl chloride

304

288

0.946-1.023

10

1907

11

216

Chloroethane-d5

378

181

0.373-0.620

50





316

Chloroethane

386

216

0.999-1.060

50

7897

24

244

Methylene chloride-d2

512

181

0.582-0.813

10





344

Methylene chloride

517

244

0.999-1.017

10

566 7

O
00
Cvl

546

Iodomethane

498

181

0.68







616

Acetone-d6

554

181

0.628-0.889

50





716

Acetone

565

616

0.984-1.019

50

3561 7

322 7

202

Acrolein-d4

564

181

0.641-0.903

5

50



302

Acrolein

566

202

0.984—1.0185

50

3777

18

203

Acrylonitrile-dj

606

181

0.735-0.926

50





303

Acrylonitrile

612

203

0.985-1.030

50

360 7

9

533

Carbon disulfide

631

181

0.86







552

T richlorofluoromethane

663

181

0.91







543

Ethyl cyanide

672

181

0.92







229

1,1 -Dichloroethene-d2

696

181

0.903-0.976

10





329

1,1 -Dichloroethene

696

229

0.999-1.011

10

31

5

536

3-Chloropropene

696

181

0.95







532

Allyl alcohol

703

181

0.96







181

Bromochloromethane (I.S.)

730

181

1.000-1.000

10





213

1,1 -Dichloroethane-dj

778

181

1.031-1.119

10





313

1,1 -Dichloroethane

786

213

0.999-1.014

10

16

1

615

Diethyl ether-d10

804

181

1.067-1.254

50





715

Diethyl ether

820

615

1.010-1.048

50

63

12

230

trans-l,2-Dichloroethene-d2

821

181

1.056-1.228

10





330

trans-1,2-Dichloroethene

821

230

0.996-1.011

10

41

3

614

Methyl ethyl ketone-d3

840

181

0.646-1.202

50





714

Methyl ethyl ketone

848

614

0.992-1.055

50

241 7

O
00

223

Chloroform-13Cj

861

181

1.092-1.322

10





323

Chloroform

861

223

0.961-1.009

10

21

2


-------
EGD

No.1

535

210

310

539

548

547

211

311

627

727

206

306

554

248

348

534

537

232

332

542

287

387

541

204

304

251

351

214

314

233

333

019

538

182

8

Method Detection
Retention Time	Limit4

Minimum Low High



Mean

EGD



Level3

Solids

Solids

Compound

(sec)

Ref

Relative 2

(Ufi/L)

(Ufi/ks)

(Ufi/ks)

Chloroacetonitrile

884

181

1.21







l,2-Dichloroethane-d4

901

181

1.187-1.416

10





1,2-Dichloroethane

910

210

0.973-1.032

10

23

3

Dibromomethane

910

181

1.25







Methacrylonitrile

921

181

1.26







Isobutyl alcohol

962

181

1.32







l,l,l-Trichloroethane-13C2

989

181

1.293-1.598

10





1,1,1 -T richloroethane

999

211

0.989-1.044

10

16

4

p-Dioxane-d8

982

181

1.262—1.4485

50





p-Dioxane

1001

627

1.008—1.0405

50

-

1407

Carbon tetrachloride-13C2

1018

182

0.754-0.805

10





Carbon tetrachloride

1018

206

0.938-1.005

10

87

9

Vinyl acetate

1031

182

0.79







Bromodichloromethane-13Cj

1045

182

0.766-0.825

10





Bromodichloromethane

1045

248

0.978-1.013

10

28

3

2-Chloro-1,3-butadiene

1084

182

0.83







Crotonaldehyde

1098

182

0.84







1,2-Dichloropropane-d6

1123

182

0.830-0.880

10





1,2-Dichloropropane

1134

232

0.984-1.018

10

29

5

cis-1,3-Dichloropropene

1138

182

0.87







Trichloroethene-13C2

1172

182

0.897-0.917

10





Trichloroethene

1187

287

0.991-1.037

10

41

2

1,3-Dichloropropane

1196

182

0.92







Benezene-d6

1200

182

0.888-0.952

10





Benezene

1212

204

1.002-1.026

10

23

8

Chlorodibromomethane-13Cj

1222

182

0.915-0.949

10





Chlorodibromomethane

1222

231

0.989-1.030

10

15

2

l,l,2-Trichloroethane-13C2

1224

182

0.922-0.953

10





1,1,2-T richloroethane

1224

214

0.975-1.027

10

26

1

trans-1,3-Dichloropropene-
cL

1226

182

0.922-0.959

10





trans-1,3-Dichloropropene

1226

233

0.993-1.016

10

	6,7

__6,7

2-Chloroethyvinyl ether

1278

182

0.983-1.026

10

122

21

1,2-Dibromoethane

1279

182

0.98







2-bromo-1 -chloropropane

1306

182

1.000-1.000

10






-------
Method Detection
Retention Time	Limit4

Minimum Low High

EGD
No.1

Compound

Mean
(sec)

EGD
Ref

Relative 2

Level3

(Ufi/L)

Solids Solids
(Ug/kg) (ug/kg)

549

Methyl methacrylate

1379

182

1.06





247

Bromoform-13Cj

1386

182

1.048-1.087

10



347

Bromoform

1386

247

0.992-1.003

10

91 7

551

1,1,1,2-T etrachloroethane

1408

182

1.08





550

4-Methyl-2-pentanone

1435

183

0.92





553

1,2,3-Trichloropropane

1520

183

0.98





215

l,l,2,2-Tetrachloroethane-d2

1525

183

0.969-0.996

10



315

1,1,2,2-T etrachloroethane

1525

215

0.890-1.016

10

20 6

545

2-Hexanone

1525

183

0.98





285

Tetrachloroethene-13C2

1528

183

0.966-0.996

10



385

T etrachloroethene

1528

285

0.997-1.003

10

106 10

540

trans-l,4-Dichloro-2-butene

1551

183

1.00





183

1,4-Dichlorobutane (int std)

1555

183

1.000-1.000

10



544

Ethyl methacrylate

1594

183

1.03





286

Toluene-d8

1603

183

1.016-1.054

10



386

Toluene

1619

286

1.001-1.019

10

27 4

207

Chlorobenzene-d5

1679

183

1.066-1.135

10



307

Chlorobenzene

1679

207

0.914-1.019

10

21 587

238

Ethylbenzene-d10

1802

183

1.144-1.293

10



338

Ethylbenzene

1820

238

0.981-1.018

10

28 4

185

Bromofluorobenzene

1985

183

1.255-1.290

10



951

m-Xylene

2348

183

1.51

10



952

o- and p-Xylene

2446

183

1.57

10



1

2

Reference numbers beginning with 0, 1, 5, or 9 indicate a pollutant quantified by the
internal standard method; reference numbers beginning with 2 or 6 indicate a labeled
compound quantified by the internal standard method; reference numbers beginning with
3 or 7 indicate a pollutant quantified by isotope dilution.

The retention time limits in this column are based on data from four wastewater

laboratories. The single values for retention times in this column are based on data from
one wastewater laboratory.

This is a minimum level at which the analytical system shall give recognizable mass
spectra (background corrected) and acceptable calibration points when calibrated using
reagent water. The concentration in the aqueous or solid phase is determined using the
equations in Section 13.

Method detection limits determined in digested sludge (low solids) and in filter cake or
compost (high solids).

Specification derived from related compound.


-------
6	An unknown interference in the particular sludge studied precluded measurement of the
method detection limit (MDL) for this compound.

7	Background levels of these compounds were present in the sludge resulting in higher than
expected MDLs. The MDL for these compounds is expected to be approximately 20
\ig/kg (100 to 200 \ig/kg for the gases and water soluble compounds) for the low solids
method and 5 to 10 \ig/kg (25 to 50 \ig/kg for the gases and water soluble compounds)
for the high solids methods, with no interferences present.

Column: 2.4 m (8 ft) x 2 mm I.D. glass, packed with 1% SP-1000 coated on 60/80 Carbopak B.

Carrier gas: Helium at 40 mL/min.

Temperature program: 3 min at 45°C, 8°C/min to 240°C, hold at 240°C for 15 minutes.

4.	Safety

4.1	The toxicity or carcinogenicity of each compound or reagent used in this method has not
been precisely determined; however, each chemical compound should be treated as a
potential health hazard.

Exposure to these compounds should be reduced to the lowest possible level. The
laboratory is responsible for maintaining a current awareness file of OSHA regulations
regarding the safe handling of the chemicals specified in this method. A reference file of
data handling sheets should also be made available to all personnel involved in these
analyses. Additional information on laboratory safety can be found in References 5
through 7.

4.2	The following compounds covered by this method have been tentatively classified as
known or suspected human or mammalian carcinogens: benzene, carbon tetrachloride,
chloroform, and vinyl chloride. Primary standards of these toxic compounds should be
prepared in a hood, and a NIOSH/MESA approved toxic gas respirator should be worn
when high concentrations are handled.

5.	Apparatus and materials

5.1	Sample bottles for discrete sampling.

5.1.1	Bottle: 25- to 40-mL with screw—cap (Pierce 13075, or equivalent). Detergent
—wash, rinse with tap and distilled water, and dry at >105°C for a minimum of
1 hour before use.

5.1.2	Septum: Teflon-faced silicone (Pierce 12722, or equivalent), cleaned as above and
baked at 100 to 200°C for 1 hour minimum.

5.2	Purge and trap device: Consists of purging device, trap, and desorber.

5.2.1 Purging devices for water and soil samples.

5.2.1.1 Purging device for water samples Designed to accept 5-mL samples
with water column at least 3 cm deep. The volume of the gaseous
head space between the water and trap shall be less than 15 mL. The
purge gas shall be introduced less than 5 mm from the base of the
water column and shall pass through the water as bubbles with a

10


-------
diameter less than 3 mm. The purging device shown in Figure 1 meets
these criteria.

5.2.1.2 Purging device for solid samples: Designed to accept 5 g of solids plus
5 mL of water. The volume of the gaseous head space between the
water and trap shall be less than 25 mL. The purge gas shall be
introduced less than 5 mm from the base of the sample and shall pass
through the water as bubbles with a diameter less than 3 mm. The
purging device shall be capable of operating at ambient temperature
(20 to 25°C) and of being controlled at temperatures of 40°C (±2°C) and
80°C (±5°C) while the sample is being purged. The purging device
shown in Figure 2 meets these criteria.

5.2.2	Trap: 25 to 30 cm long x 2.5 mm I.D. minimum, containing the following:

5.2.2.1	Methyl silicone packing: 1cm (±0.2cm), 3% OV-1 on 60/80 mesh
Chromosorb W, or equivalent.

5.2.2.2	Porous polymer: 15cm (±1.0 cm), Tenax GC (2,6-diphenylene oxide
polymer), 60/80 mesh, chromatographic grade, or equivalent.

5.2.2.3	Silica gel: 8cm (±1.0 cm), Davison Chemical, 35/60 mesh, grade 15, or
equivalent. The trap shown in Figure 3 meets these specifications.

5.2.3	Desorber: Shall heat the trap to 175°C (±5°C) in 45 seconds or less. The polymer
section of the trap shall not exceed a temperature of 180°C and the remaining
sections shall not exceed 220°C during desorb, and no portion of the trap shall
exceed 225°C during bakeout. The desorber shown in Figure 3 meets these
specifications.

5.2.4	The purge and trap device may be a separate unit, or coupled to a GC as shown
in Figures 4 and 5.

5.3	Gas chromatograph: Shall be linearly temperature programmable with initial and final
holds, shall contain a glass jet separator as the MS interface, and shall produce results
which meet the calibration (Section 7), quality assurance (Section 8), and performance tests
(Section 11) of this method.

5.3.1 Column: 2.8 • 0.4 m x 2 • 0.5 mm I.D. glass, packed with 1% SP-1000 on Carbopak
B, 60/80 mesh, or equivalent.

5.4	Mass spectrometer: 70 eV electron impact ionization; shall repetitively scan from 20 to
250 amu every 2 to 3 seconds, and produce a unit resolution (valleys between m/z 174
to 176 less than 10% of the height of the m/z 175 peak), background corrected mass
spectrum from 50 ng 4-bromofluorobenzene (BFB) injected into the GC. The BFB
spectrum shall meet the mass-intensity criteria in Table 4. All portions of the GC column,
transfer lines, and separator which connect the GC column to the ion source shall remain
at or above the column temperature during analysis to preclude condensation of less
volatile compounds.

5.5	Data system: Shall collect and record MS data, store mass-intensity data in spectral
libraries, process GCMS data and generate reports, and shall calculate and record response
factors.

11


-------


Table 4



BFB Mass-Intensity Specifications

m/z

Intensity Required

50

15 to 40% of m/z 95

75

30 to 60% of m/z 95

95

base peak, 100%

96

5 to 9% of m/z 95

173

less than 2% of m/z 174

174

greater than 50% of m/z 95

175

5 to 9% of m/z 174

176

95 to 101% of m/z 174

177

5 to 9% of m/z 176

5.5.1	Data acquisition: Mass spectra shall be collected continuously throughout the
analysis and stored on a mass storage device.

5.5.2	Mass spectral libraries: User-created libraries containing mass spectra obtained
from analysis of authentic standards shall be employed to reverse search GCMS
runs for the compounds of interest (Section 7.2).

5.5.3	Data processing: The data system shall be used to search, locate, identify, and
quantify the compounds of interest in each GCMS analysis. Software routines
shall be employed to compute retention times and EICP areas. Displays of
spectra, mass chromatograms, and library comparisons are required to verify
results.

5.5.4	Response factors and multipoint calibrations: The data system shall be used to
record and maintain lists of response factors (response ratios for isotope dilution)
and generate multi-point calibration curves (Section 7). Computations of relative
standard deviation (coefficient of variation) are useful for testing calibration
linearity. Statistics on initial and ongoing performance shall be maintained
(Sections 8 and 11).

5.6	Syringes: 5-mL glass hypodermic, with Luer-lok tips.

5.7	Micro syringes: 10-, 25-, and 100 |iL.

5.8	Syringe valves: 2-way, with Luer ends (Teflon or Kel-F).

5.9	Syringe: 5-mL, gas-tight, with shut-off valve.

5.10	Bottles: 15-mL, screw-cap with Teflon liner.

5.11	Balances.

5.11.1	Analytical, capable of weighing 0.1 mg.

5.11.2	Top-loading, capable of weighing 10 mg.

5.12	Equipment for determining percent moisture.

12


-------
5.12.1	Oven,capable of being temperature-controlled at 110°C (±5°C).

5.12.2	Dessicator.

5.12.3	Beakers: 50 to 100-mL.

6. Reagents and standards

6.1	Reagent water: Water in which the compounds of interest and interfering compounds are
not detected by this method (Section 11.7). It may be generated by any of the following
methods:

6.1.1	Activated carbon: pass tap water through a carbon bed (Calgon Filtrasorb-300,
or equivalent).

6.1.2	Water purifier: Pass tap water through a purifier (Millipore Super Q, or
equivalent).

6.1.3	Boil and purge: Heat tap water to between 90 and 100°C and bubble contaminant
free inert gas through it for approximately 1 hour. While still hot, transfer the
water to screw-cap bottles and seal with a Teflon-lined cap.

6.2	Sodium thiosulfate: ACS granular.

6.3	Methanol: Pesticide-quality or equivalent.

6.4	Standard solutions: Purchased as solutions or mixtures with certification to their purity,
concentration, and authenticity, or prepared from materials of known purity and
composition. If compound purity is 96% or greater, the weight may be used without
correction to calculate the concentration of the standard.

6.5	Preparation of stock solutions: Prepare in methanol using liquid or gaseous standards per
the steps below. Observe the safety precautions given in Section 4.

6.5.1	Place approximately 9.8 mL of methanol in a 10-mL ground-glass-stoppered
volumetric flask. Allow the flask to stand unstoppered for approximately 10
minutes or until all methanol wetted surfaces have dried. In each case, weigh the
flask, immediately add the compound, then immediately reweigh to prevent
evaporation losses from affecting the measurement.

6.5.1.1	Liquids: Using a 100 |iL syringe, permit 2 drops of liquid to fall into
the methanol without contacting the neck of the flask. Alternatively,
inject a known volume of the compound into the methanol in the flask
using a micro-syringe.

6.5.1.2	Gases (chloromethane, bromomethane, chloroethane, vinyl chloride):
Fill a valved 5-mL gas-tight syringe with the compound. Lower the
needle to approximately 5 mm above the methanol meniscus. Slowly
introduce the compound above the surface of the meniscus. The gas
will dissolve rapidly in the methanol.

6.5.2	Fill the flask to volume, stopper, then mix by inverting several times. Calculate
the concentration in mg/mL ((Jg/ pL) from the weight gain (or density if a known
volume was injected).

13


-------
6.5.3	Transfer the stock solution to a Teflon-sealed screw-cap bottle. Store, with
minimal headspace, in the dark at -10 to -20°C.

6.5.4	Prepare fresh standards weekly for the gases and 2-chloroethylvinyl ether. All
other standards are replaced after one month, or sooner if comparison with check
standards indicate a change in concentration. Quality control check standards
that can be used to determine the accuracy of calibration standards are available
from the US Environmental Protection Agency, Environmental Monitoring and
Support Laboratory, Cincinnati, Ohio.

6.6	Labeled compound spiking solution: From stock standard solutions prepared as above,
or from mixtures, prepare the spiking solution to contain a concentration such that a 5-
to 10- |iL spike into each 5-mL sample, blank, or aqueous standard analyzed will result
in a concentration of 20 ug/L of each labeled compound. For the gases and for the water
soluble compounds (acrolein, acrylonitrile, acetone, diethyl ether, p-dioxane, and MEK),
a concentration of 100 ug/L may be used. Include the internal standards (Section 7.5) in
this solution so that a concentration of 20 ug/L in each sample, blank, or aqueous
standard will be produced.

6.7	Secondary standards: Using stock solutions, prepare a secondary standard in methanol
to contain each pollutant at a concentration of 500 |ig/mL. For the gases and water
soluble compounds (Section 6.6), a concentration of 2.5 mg/mL may be used.

6.7.1	Aqueous calibration standards: Using a 25-|iL syringe, add 20 |iL of the
secondary standard (Section 6.7) to 50, 100, 200, 500, and 1000 mL of reagent
water to produce concentrations of 200, 100, 50, 20, and 10 Ug/L, respectively.
If the higher concentration standard for the gases and water soluble compounds
was chosen (Section 6.6), these compounds will be at concentrations of 1000, 500,
250, 100, and 50 Ug/L in the aqueous calibration standards.

6.7.2	Aqueous performance standard: An aqueous standard containing all pollutants,
internal standards, labeled compounds, and BFB is prepared daily, and analyzed
each shift to demonstrate performance (Section 11). This standard shall contain
either 20 or 100 Ug/L of the labeled and pollutant gases and water soluble
compounds, 10 Ug/L BFB, and 20 Ug/L of all other pollutants, labeled
compounds, and internal standards. It may be the nominal 20 Ug/L aqueous
calibration standard (Section 6.7.1).

6.7.3	A methanolic standard containing all pollutants and internal standards is
prepared to demonstrate recovery of these compounds when syringe injection and
purge-and-trap analyses are compared. This standard shall contain either 100
|ig/mL or 500 y\g/mL of the gases and water soluble compounds, and 100 |ig/mL
of the remaining pollutants and internal standards (consistent with the amounts
in the aqueous performance standard in 6.7.2).

6.7.4	Other standards which may be needed are those for test of BFB performance
(Section 7.1) and for collection of mass spectra for storage in spectral libraries
(Section 7.2).

7. Calibration

14


-------
Calibration of the GCMS system is performed by purging the compounds of interest and their

labeled analogs from reagent water at the temperature to be used for analysis of samples.

7.1	Assemble the gas chromatographic apparatus and establish operating conditions given in
Table 3. By injecting standards into the GC, demonstrate that the analytical system meets
the minimum levels in Table 3 for the compounds for which calibration is to be
performed, and the mass-intensity criteria in Table 4 for 50 ng BFB.

7.2	Mass spectral libraries: Detection and identification of the compounds of interest are
dependent upon the spectra stored in user created libraries.

7.2.1	For the compounds in Table 1 and other compounds for which the GCMS is to
be calibrated, obtain a mass spectrum of each pollutant and labeled compound
and each internal standard by analyzing an authentic standard either singly or as
part of a mixture in which there is no interference between closely eluted
components. Examine the spectrum to determine that only a single compound
is present. Fragments not attributable to the compound under study indicate the
presence of an interfering compound. Adjust the analytical conditions and scan
rate (for this test only) to produce an undistorted spectrum at the GC peak
maximum. An undistorted spectrum will usually be obtained if five complete
spectra are collected across the upper half of the GC peak. Software algorithms
designed to "enhance" the spectrum may eliminate distortion, but may also
eliminate authentic m/z's or introduce other distortion.

7.2.2	The authentic reference spectrum is obtained under BFB tuning conditions
(Section 7.1 and Table 4) to normalize it to spectra from other instruments.

7.2.3	The spectrum is edited by saving the five most intense mass spectral peaks and
all other mass spectral peaks greater than 10% of the base peak. The spectrum
may be further edited to remove common interfering masses. If five mass
spectral peaks cannot be obtained under the scan conditions given in Section 5.4,
the mass spectrometer may be scanned to an m/z lower than 20 to gain
additional spectral information. The spectrum obtained is stored for reverse
search and for compound confirmation.

7.2.4	For the compounds in Table 2 and other compounds for which the mass spectra,
quantitation m/z's, and retention times are known but the instrument is not to
be calibrated, add the retention time and reference compound (Table 3); the
response factor and the quantitation m/z (Table 5); and spectrum (Appendix A)
to the reverse search library. Edit the spectrum per Section 7.2.3, if necessary.

7.3	Assemble the purge-and-trap device. Pack the trap as shown in Figure 3 and condition
overnight at 170 to 180°C by backflushing with an inert gas at a flow rate of 20 to 30
mL/min. Condition traps daily for a minimum of 10 minutes prior to use.

7.3.1 Analyze the aqueous performance standard (Section 6.7.2) according to the
purge-and-trap procedure in Section 10. Compute the area at the primary m/z
(Table 5) for each compound. Compare these areas to those obtained by injecting
1 |iL of the methanolic standard (Section 6.7.3) to determine compound recovery.
The recovery shall be greater than 20% for the water soluble compounds (Section
6.6), and 60 to 110% for all other compounds. This recovery is demonstrated
initially for each purge-and-trap GCMS system. The test is repeated only if the

15


-------
purge-and-trap or GCMS systems are modified in any way that might result in
a change in recovery.

7.3.2 Demonstrate that 100 ng toluene (or toluene-dg) produces an area at m^z 91 (or
99) approximately one-tenth that required to exceed the linear range of the
system. The exact value must be determined by experience for each instrument.
It is used to match the calibration range of the instrument to the analytical range
and detection limits required.

Table 5. Volatile Organic Compound Characteristic M/Z'S









Response purge



labeled

Primary

Reference

temp. Of:

Compound

Analog

m/z1

Compound2

O

0

O
00

o

0

O
CM

Acetone

d6

58/64





Acrolein

d4

56/60





Acrylonitrile

d3

53/56





Allyl alcohol



57

181

~3 0.20

Benzene

d6

78/84





2-Bromo-1 -chloropropane



77





Bromochloromethane4



128





Bromodichloromethane

13c

83/86





Bromoform

13c

173/176





Bromomethane

d3

96/99





Carbon disulfide



76

181

1.93 2.02

Carbon tetrachloride

13c

47/48





2-Chloro-1,3-butadiene



53

182

0.29 0.50

Chloroacetonitrile



75

181

-3 1.12

Chlorobenzene

d5

112/117





Chloroethane

d5

64/71





2-Chloroethylvinyl ether

d7

106/113





Chloroform

13c

85/86





Chloromethane

d3

50/53





3-Chloropropene



76

181

0.43 0.63

Crotonaldehyde



70

182

-3 0.090

Dibromochloromethane

13c

129/130





1,2-Dibromoethane



107

182

0.86 0.68

Dibromomethane



93

181

1.35 1.91

1,4-Dichlorobutane



55





trans-1,4-Dichloro-2-bu-



75

183

0.093 0.014

tene









1,1 -Dichloroethane

d3

63/66





16


-------








Response purge



labeled

Primary

Reference

temp. Of:

Compound

Analog

m/z1

Compound2

20 °C

OO

o

0

O

1,2-Dichloroethane

d4

62/67







1,1 -Dichloroethene

d2

61/65







trans-1,2-Dichlorethene

d2

61/65







1,2-Dichloropropane

d6

63/67







1,3-Dichloropropane



76

182

0.89

0.88

cis-1,3-Dichloropropene



75

182

0.29

0.41

trans-1,3-Dichloropropene

d4

75/79







Diethyl ether

dio

74/84







p-Dioxane

d8

88/96







Ethyl cyanide



54

181

(3)

1.26

Ethyl methacrylate



69

183

0.69

0.52

Ethylbenzene

dio

106/116







2-Hexanone



58

183

0.076

0.33

Iodomethane



142

181

4.55

2.55

Isobutyl alcohol



74

181

(3)

0.22

Methylene chloride

d2

84/88







Methyl ethyl ketone

d8

72/80







Methyl methacrylate



69

182

0.23

0.79

4-Methyl-2-pentanone



58

183

0.15

0.29

Methacrylonitrile



67

181

0.25

0.79

1,1,1,2-T etrachloroethane



131

182

0.20

0.25

1,1,2,2-T etrachloroethane

d2

83/84







T etrachloroethene

13c2

164/172







Toluene

d8

92/100







1,1,1 -Trichloroethane

d3

97/102







1,1,2-Trichloroethane

13c2

83/84







Trichloroethene

13c2

95/136







T richlorofluoromethane



101

181

2.31

2.19

1,2,3-Trichloropropane



75

183

0.89

0.72

Vinyl acetate



86

182

0.054

0.19

Vinyl chloride

d3

62/65







m-Xylene



106

183

1.69

-

0- and p-Xylene



106

183

3.33

-

181	= bromochloromethane

182	= 2-bromo-l-chloropropane

183	= 1,4-dichlorobutane

Not detected at a purge temperature of 20°C
Internal standard


-------
Note: Because the composition and purity of commercially-supplied isotopically labeled stan-
dard's may vary, the primary m/z of the labeled analogs given in this table should be used as
guidance. The appropriate m/z of the labeled analogs should be determined prior to use for
sample analysis. Deviations from the m/z's listed here must be documented by the laboratory
and submitted with the data.

7.4 Calibration by isotope dilution: The isotope dilution approach is used for the purgeable
organic compounds when appropriate labeled compounds are available and when
interferences do not preclude the analysis. If labeled compounds are not available, or
interferences are present, the internal standard method (Section 7.5) is used. A calibration
curve encompassing the concentration range of interest is prepared for each compound
determined. The relative response (RR) vs. concentration (l-ig/L) is plotted or computed
using a linear regression. An example of a calibration curve for toluene using toluene-d8
is given in Figure 6. Also shown are the ±10% error limits (dotted lines). Relative
response is determined according to the procedures described below. A minimum of five
data points are required for calibration (Section 7.4.4).

7.4.1	The relative response (RR) of pollutant to labeled compound is determined from
isotope ratio values calculated from acquired data. Three isotope ratios are used
in this process:

Rx = the isotope ratio measured in the pure pollutant (Figure 7A).

Ry = the isotope ratio of pure labeled compound (Figure 7B).

Rm = the isotope ratio measured in the analytical mixture of the
pollutant and labeled compounds (Figure 7 C.)

The correct way to calculate RR is:

_ (V (*,+ 1)

(*m - RX) (Ry + D

If Rm is not between 2Ry and 0.55 > the method does not apply and the sample
is analyzed by the internal standard method (Section 7.5).

7.4.2	In most cases, the retention times of the pollutant and labeled compound are the
same, and isotope ratios (R's) can be calculated from the EICP areas, where:

(area at mJz)

R = 		—

(area at m2/z)

If either of the areas is zero, it is assigned a value of one in the calculations; that
is, if: area of m/z = 50721,

area of mz/Z = 0,
then R = 50721/1 = 50720

The data from these analyses are reported to three significant figures (see Section
13.6). In order to prevent rounding errors from affecting the values to be

18


-------
reported, all calculations performed prior to the final determination of
concentrations should be carried out using at least four significant figures.
Therefore, the calculation of R above is rounded to four significant figures.
The m/z's are always selected such that Rx > Ry. When there is a difference in
retention times (RT) between the pollutant and labeled compounds, special
precautions are required to determine the isotope ratios.

Rx, Ry, and Rmare defined as follows:

[area m^z (at

Rx"



R =



1



1

[area

m2/z (at RT2)]

[area

mx!z (at

[area mJz (at RT2)]

7.4.3 An example of the above calculations can be taken from the data plotted in Figure
7 for toluene and toluene-d8. For these data:

= 168920 = 168900
1

IT = —— = 0.00001640
y 60960

R = 96868 = u?4
m 82508

The RR for the above data is then calculated using the equation given in Section
7.4.1. For the example, rounded to four significant figures, RR = 1.174. Not all
labeled compounds elute before their pollutant analogs.

7.4.4	To calibrate the analytical system by isotope dilution, analyze a 5-mL aliquot of
each of the aqueous calibration standards (Section 6.7.1) spiked with an
appropriate constant amount of the labeled compound spiking solution (Section
6.6), using the purge-and-trap procedure in Section 10. Compute the RR at each
concentration.

7.4.5	Linearity: If the ratio of relative response to concentration for any compound is
constant (less than 20% coefficient of variation) over the five point calibration
range, an averaged relative response/concentration ratio may be used for that
compound; otherwise, the complete calibration curve for that compound shall be
used over the five point calibration range.

Calibration by internal standard: Used when criteria for isotope dilution (Section 7.4)
cannot be met. The method is applied to pollutants having no labeled analog and to the
labeled compounds. The internal standards used for volatiles analyses are
bromochloromethane, 2-bromo-l-chloropropane, and 1,4-dichlorobutane. Concentrations

19


-------
of the labeled compounds and pollutants without labeled analogs are computed relative
to the nearest eluting internal standard, as shown in Tables 3 and 5.

7.5.1	Response factors: Calibration requires the determination of response factors (RF)
which are defined by the following equation:

R (As * CJ,

(a,-s * Cs)

Where:

A = is the EICP area at the characteristic m/z for the compound in the daily standard.
Ais = is the EICP area at the characteristic m/z for the internal standard.

Cjs = is the concentration ([ig/L) of the internal standard.

Cs = is the concentration of the pollutant in the daily standard.

7.5.2	The response factor is determined at 10, 20, 50, 100, and 200 ]ig/L for the
pollutants (optionally at five times these concentrations for gases and water
soluble pollutants; see Section 6.7), in a way analogous to that for calibration by
isotope dilution (Section 7.4.4). The RF is plotted against concentration for each
compound in the standard (CJ to produce a calibration curve.

7.5.3	Linearity: If the response factor (RF) for any compound is constant (less than 35%
coefficient of variation) over the five-point calibration range, an averaged response
factor may be used for that compound; otherwise, the complete calibration curve
for that compound shall be used over the five-point range.

7.6	Combined calibration: By adding the isotopically labeled compounds and internal
standards (Section 6.6) to the aqueous calibration standards (Section 6.7.1), a single set of
analyses can be used to produce calibration curves for the isotope dilution and internal
standard methods. These curves are verified each shift (Section 11.5) by purging the
aqueous performance standard (Section 6.7.2). Recalibration is required only if calibration
and ongoing performance (Section 11.5) criteria cannot be met.

7.7	Elevated purge temperature calibration: Samples containing greater than 1% solids are
analyzed at a temperature of 40°C (±2°C) (Section 10). For these samples, the analytical
system may be calibrated using a purge temperature of 40°C(±2°C) in order to more
closely approximate the behavior of the compounds of interest in high solids samples.

8. Quality assurance/quality control

8.1 Each laboratory that uses this method is required to operate a formal quality assurance
program (Reference 8). The minimum requirements of this program consist of an initial
demonstration of laboratory capability, analysis of samples spiked with labeled
compounds to evaluate and document data quality, and analysis of standards and blanks
as tests of continued performance. Laboratory performance is compared to established
performance criteria to determine if the results of analyses meet the performance
characteristics of the method.

8.1.1 The analyst shall make an initial demonstration of the ability to generate
acceptable accuracy and precision with this method. This ability is established as
described in Section 8.2.

20


-------
8.1.2	The analyst is permitted to modify this method to improve separations or lower
the costs of measurements, provided all performance specifications are met. Each
time a modification is made to the method, the analyst is required to repeat the
procedure in Section 8.2 to demonstrate method performance.

8.1.3	Analyses of blanks are required to demonstrate freedom from contamination and
that the compounds of interest and interfering compounds have not been carried
over from a previous analysis (Section 3). The procedures and criteria for analysis
of a blank are described in Section 8.5.

8.1.4	The laboratory shall spike all samples with labeled compounds to monitor
method performance. This test is described in Section 8.3. When results of these
spikes indicate atypical method performance for samples, the samples are diluted
to bring method performance within acceptable limits (Section 14.2).

8.1.5	The laboratory shall, on an ongoing basis, demonstrate through the analysis of the
aqueous performance standard (Section 6.7.2) that the analysis system is in
control. This procedure is described in Sections 11.1 and 11.5.

8.1.6	The laboratory shall maintain records to define the quality of data that is
generated. Development of accuracy statements is described in Sections 8.4 and
11.5.2.

21


-------
Table 6. Acceptance Criteria for Performance Tests

Acceptance criteria at 20 Hg/L or as noted









Labeled







Labeled and native

compound

Labeled and





compound initial

recovery

native compound

EGD



precision and ac-

(Sect. 8.3 and

ongoing accuracy





curacy (Sect

. 8.2.3)

14.2)

(Sect. 11.5)

No.1

Compound

s ( ug/L)

X (ug/L)

P (%)

R (ug/L)

516

acetone*

51.0

77 - 153

35 - 165

55 - 145

002

acrolein*

72.0

32 - 168

37 - 163

7 - 190

003

acrylonitrile*

16.0

70 - 132

ns - 204

58 - 144

004

benzene

9.0

13 - 28

ns - 196

4 - 33

048

bromodichloro-

8.2

7 - 32

ns - 199

4 - 34



methane









047

bromoform

7.0

7 - 35

ns - 214

6 - 36

046

bromomethane

25.0

d - 54

ns - 414

d - 61

006

carbon

6.9

16 - 25

42 - 165

12 - 30



tetrachloride









007

chlorobenzene

8.2

14 - 30

ns - 205

4 - 35

016

chloroethane

15.0

d - 47

ns - 308

d - 51

019

2-chloroethylvinyl

36.0

d - 70

ns - 554

d - 79



ether









023

chloroform

7.9

12 - 26

18 - 172

8 - 30

045

chloromethane

26.0

d - 56

ns - 410

d - 64

051

dibromochloro-

7.9

11 - 29

16 - 185

8 - 32



methane









013

1,1 -dichloroethane

6.7

11 - 31

23 - 191

9 - 33

010

1,2-dichloroethane

7.7

12 - 30

12 - 192

8 - 33

029

1,1 -dichloroethene

12.0

d - 50

ns - 315

d - 52

030

trans-1,2-dichloro-

7.4

11 - 32

15 - 195

8 - 34



ethene









032

1,2-dichloropropane

19.0

d - 47

ns - 343

d - 51

033

trans-1,3-dichloro-

15.0

d - 40

ns - 284

d - 44



propene









515

diethyl ether*

44.0

75 - 146

44 - 156

55 - 145


-------
Acceptance criteria at 20 Hg/L or as noted

EGD

No.

Compound

Labeled and native
compound initial
precision and ac-
curacy (Sect. 8.2.3)

s ( ug/L) X (ug/L)

Labeled
compound
recovery
(Sect. 8.3 and
14.2)

Labeled and
native compound
ongoing accuracy
(Sect. 11.5)

R (ua/L)

527

p-dioxane*

7.2

13 - 27

ns - 239

11 - 29

038

ethylbenzene

9.6

16 - 29

ns - 203

5 - 35

044

methylene chloride

9.7

d - 50

ns - 316

d - 50

514

methyl ethyl
ketone*

57.0

66 - 159

36 - 164

42 - 158

015

1,1,2,2 -tetr achlor o-
ethane

9.6

11 - 30

5 - 199

7 - 34

085

tetrachloroethane

6.6

15 - 29

31 - 181

11 - 32

086

toluene

6.3

15 - 29

4 - 193

6 - 33

011

1,1,1-

trichloroethane

5.9

11 - 33

12 - 200

8 - 35

014

1,1,2-

trichloroethane

7.1

12 - 30

21 - 184

9 - 32

087

trichloroethene

8.9

17 - 30

35 - 196

12 - 34

088

vinyl chloride

28.0

d - 59

ns - 452

d - 65

* acceptance criteria at 100 Ug/L

d = detected; result must be greater than zero.

ns = no specification; limit would be below detection limit.

1 Reference numbers beginning with 0, 1, or 5 indicate a pollutant quantified by the internal
standard method; reference numbers beginning with 2 or 6 indicate a labeled compound
quantified by the internal Standard method; reference numbers beginning with 3 or 7
indicate a pollutant quantified by isotope dilution.

23


-------
8.2 Initial precision and accuracy: To establish the ability to generate acceptable precision and
accuracy, the analyst shall perform the following operations for compounds to be
calibrated:

8.2.1	Analyze two sets of four 5-mL aliquots (8 aliquots total) of the aqueous
performance standard (Section 6.7.2) according to the method beginning in Section
10.

8.2.2	Using results of the first set of four analyses in Section 8.2.1, compute the average
recovery (X) in Hg/L and the standard deviation of the recovery (s) in Hg/L for
each compound, by isotope dilution for pollutants with a labeled analog, and by
internal standard for labeled compounds and pollutants with no labeled analog.

8.2.3	For each compound, compare s and X with the corresponding limits for initial
precision and accuracy found in Table 6. If s and X for all compounds meet the
acceptance criteria, system performance is acceptable and analysis of blanks and
samples may begin. If, however, any individual s exceeds the precision limit or
any individual X falls outside the range for accuracy, system performance is
unacceptable for that compound.

NOTE: The large number of compounds in Table 6 present a substantial probability
that one or more will fail one of the acceptance criteria when all compounds are
analyzed. To determine if the analytical system is out of control, or if the
failure can be attributed to probability, proceed as follows:

8.2.4 Using the results of the second set of four analyses, compute s and X for only
those compounds which failed the test of the first set of four analyses (Section
8.2.3). If these compounds now pass, system performance is acceptable for all
compounds and analysis of blanks and samples may begin. If, however, any of
the same compounds fail again, the analysis system is not performing properly
for the compound (s) in question. In this event, correct the problem and repeat
the entire test (Section 8.2.1).

8.3	The laboratory shall spike all samples with labeled compounds to assess method
performance on the sample matrix.

8.3.1	Spike and analyze each sample according to the method beginning in Section 10.

8.3.2	Compute the percent recovery (P) of the labeled compounds using the internal
standard method (Section 7.5).

8.3.3	Compare the percent recovery for each compound with the corresponding labeled
compound recovery limit in Table 6. If the recovery of any compound falls
outside its warning limit, method performance is unacceptable for that compound
in that sample. Therefore, the sample matrix is complex and the sample is to be
diluted and reanalyzed, per Section 14.2.

8.4	As part of the QA program for the laboratory, method accuracy for wastewater samples
shall be assessed and records shall be maintained. After the analysis of five wastewater
samples for which the labeled compounds pass the tests in Section 8.3.3, compute the

24


-------
average percent recovery (P) and the standard deviation of the percent recovery (sp) for
the labeled compounds only. Express the accuracy assessment as a percent recovery
interval from P - 2sp to P + 2sp. For example, if P = 90% and sp = 10%, the accuracy
interval is expressed as 70 to 110%. Update the accuracy assessment for each compound
on a regular basis (e.g., after each 5 to 10 new accuracy measurements).

8.5	Blanks: Reagent water blanks are analyzed to demonstrate freedom from carry-over
(Section 3) and contamination.

8.5.1	The level at which the purge and trap system will carry greater than 5 Hg/L of
a pollutant of interest (Tables 1 and 2) into a succeeding blank shall be
determined by analyzing successively larger concentrations of these compounds.
When a sample contains this concentration or more, a blank shall be analyzed
immediately following this sample to demonstrate no carry-over at the 5 Hg/L
level.

8.5.2	With each sample lot (samples analyzed on the same 8-hour shift), a blank shall
be analyzed immediately after analysis of the aqueous performance standard
(Section 11.1) to demonstrate freedom from contamination. If any of the
compounds of interest (Tables 1 and 2) or any potentially interfering compound
is found in a blank at greater than 10 Hg/L (assuming a response factor of 1
relative to the nearest eluted internal standard for compounds not listed in Tables
1 and 2), analysis of samples is halted until the source of contamination is
eliminated and a blank shows no evidence of contamination at this level.

8.6	The specifications contained in this method can be met if the apparatus used is calibrated
properly, then maintained in a calibrated state. The standards used for calibration
(Section 7), calibration verification (Section 11.5) and for initial (Section 8.2) and ongoing
(Section 11.5) precision and accuracy should be identical, so that the most precise results
will be obtained. The GCMS instrument in particular will provide the most reproducible
results if dedicated to the settings and conditions required for the analyses of volatiles by
this method.

8.7	Depending on specific program requirements, field replicates may be collected to
determine the precision of the sampling technique, and spiked samples may be required
to determine the accuracy of the analysis when the internal method is used.

9. Sample collection, preservation, and handling

9.1	Grab samples are collected in glass containers having a total volume greater than 20 mL.
For aqueous samples which pour freely, fill sample bottles so that no air bubbles pass
through the sample as the bottle is filled and seal each bottle so that no air bubbles are
entrapped. Maintain the hermetic seal on the sample bottle until time of analysis.

9.2	Samples are maintained at 0 to 4°C from the time of collection until analysis. If an
aqueous sample contains residual chlorine, add sodium thiosulfate preservative (10 mg/40
mL) to the empty sample bottles just prior to shipment to the sample site. EPA Methods
330.4 and 330.5 may be used for measurement of residual chlorine (Reference 9). If
preservative has been added, shake the bottle vigorously for one minute immediately after
filling.

9.3	For aqueous samples, experimental evidence indicates that some aromatic compounds,
notably benzene, toluene, and ethyl benzene are susceptible to rapid biological

25


-------
degradation under certain environmental conditions. Refrigeration alone may not be
adequate to preserve these compounds in wastewaters for more than seven days. For this
reason, a separate sample should be collected, acidified, and analyzed when these
aromatics are to be determined. Collect about 500 mL of sample in a clean container.
Adjust the pH of the sample to about 2 by adding HC1 (1 + 1) while stirring. Check pH
with narrow range (1.4 to 2.8) pH paper. Fill a sample container as described in Section
9.1. If residual chlorine is present, add sodium thiosulfate to a separate sample container
and fill as in Section 9.1.

9.4 All samples shall be analyzed within 14 days of collection.

10. Purge, trap, and GCMS analysis

Samples containing less than one percent solids are analyzed directly as aqueous samples
(Section 10.4). Samples containing one percent solids or greater are analyzed as solid
samples utilizing one of two methods, depending on the levels of pollutants in the
sample. Samples containing one percent solids or greater and low to moderate levels of
pollutants are analyzed by purging a known weight of sample added to 5 mL of reagent
water (Section 10.5). Samples containing 1% solids or greater and high levels of pollutants
are extracted with methanol, and an aliquot of the methanol extract is added to reagent
water and purged (Section 10.6).

10.1 Determination of percent solids.

10.1.1	Weigh 5 to 10 g of sample into a tared beaker.

10.1.2	Dry overnight (12 hours minimum) at 110°C (±5°C), and cool in a dessicator.

10.1.3	Determine percent solids as follows:

% solids = weight of sample dryx 100
weight of sample wet

10.2	Remove standards and samples from cold storage and bring to 20 to 25°C.

10.3	Adjust the purge gas flow rate to 40 (±4mL/min).

10.4	Samples containing less than 1% solids.

10.4.1	Mix the sample by shaking vigorously. Remove the plunger from a 5-mL syringe
and attach a closed syringe valve. Open the sample bottle and carefully pour the
sample into the syringe barrel until it overflows. Replace the plunger and
compress the sample. Open the syringe valve and vent any residual air while
adjusting the sample volume to 5 mL (±0.1 mL). Because this process of taking
an aliquot destroys the validity of the sample for future analysis, fill a second
syringe at this time to protect against possible loss of data.

10.4.2	Add an appropriate amount of the labeled compound spiking solution (Section
6.6) through the valve bore, then close the valve.

10.4.3	Attach the syringe valve assembly to the syringe valve on the purging device.
Open both syringe valves and inject the sample into the purging chamber. Purge
the sample per Section 10.7.

26


-------
10.5	Samples containing 1% solids or greater and low to moderate levels of pollutants.

10.5.1	Mix the sample thoroughly using a clean spatula.

10.5.2	Weigh 5 g (±1 g) of sample into a purging vessel (Figure 2). Record the weight
to three significant figures.

10.5.3	Add 5 mL (±0.1 mL) of reagent water to the vessel.

10.5.4	Using a metal spatula, break up any lumps of sample to disperse the sample in
the water.

10.5.5	Add an appropriate amount of the labeled compound spiking solution (Section
6.6) to the sample in the purge vessel. Place a cap on the purging vessel and and
shake vigorously to further disperse the sample. Attach the purge vessel to the
purging device, and purge the sample per Section 10.7.

10.6	Samples containing 1% solids or greater and high levels of pollutants, or samples
requiring dilution by a factor of more than 100 (see Section 13.4).

10.6.1	Mix the sample thoroughly using a clean spatula.

10.6.2	Weigh 5g (±1 g) of sample into a calibrated 15- to 25-mL centrifuge tube. Record
the weight of the sample to three significant figures.

10.6.3	Add 10 mL of methanol to the centrifuge tube. Cap the tube and shake it
vigorously for 15 to 20 seconds to disperse the sample in the methanol. Allow
the sample to settle in the tube. If necessary, centrifuge the sample to settle
suspended particles.

10.6.4	Remove approximately 0.1% of the volume of the supernatant methanol using a
15- to 25- |iL syringe. This volume will be in the range of 10 to 15 |iL.

10.6.5	Add this volume of the methanol extract to 5 mL reagent water in a 5 mL syringe,
and analyze per Section 10.4.1.

10.6.6	For further dilutions, dilute 1 mL of the supernatant methanol (Section 10.6.4) to
10 mL, 100 mL, 1000 mL, etc., in reagent water. Remove a volume of this
methanol extract/reagent water mixture equivalent to the volume in Section
10.6.4, add it to 5 mL reagent water in a 5 mL syringe, and analyze per Section
10.4.1.

10.7	Purge the sample for 11 minutes (±0.1 minute) at 20 to 25°C for samples containing less
than 1% solids. Purge samples containing one percent solids or greater at 40°(±2°). If the
compounds in Table 2 that do not purge at 20 to 40°C are to be determined, a purge
temperature of 80°C (±5°C) is used.

10.8	After the 11 minute purge time, attach the trap to the chromatograph and set the purge-
and- trap apparatus to the desorb mode (Figure 5). Desorb the trapped compounds into
the GC column by heating the trap to between 170 and 180°C while backflushing with
carrier gas at 20 to 60 mL/min for 4 minutes. Start MS data acquisition upon start of the
desorb cycle, and start the GC column temperature program 3 minutes later. Table 3
summarizes the recommended operating conditions for the gas chromatograph. Included
in this table are retention times and minimum levels that can be achieved under these
conditions. An example of the separations achieved by the column listed is shown in
Figure 9. Other columns may be used provided the requirements in Section 8 are met.

27


-------
If the priority pollutant gases produce GC peaks so broad that the precision and recovery
specifications (Section 8.2) cannot be met, the column may be cooled to ambient or
subambient temperatures to sharpen these peaks.

10.9	After desorbing the sample for four minutes, recondition the trap by purging with purge
gas while maintaining the trap temperature at between 170 and 180°C. After
approximately 7 minutes, turn off the trap heater to stop the gas flow through the trap.
When cool, the trap is ready for the next sample.

10.10	While analysis of the desorbed compounds proceeds, remove and clean the purge device.
Rinse with tap water, clean with detergent and water, rinse with tap and distilled water,
and dry for aminimum of 1 hour in an oven at a temperature greater than 150°C.

System performance

At the beginning of each 8 hour shift during which analyses are performed, system
calibration and performance shall be verified for the pollutants and labeled compounds
(Table 1). For these tests, analysis of the aqueous performance standard (Section 6.7.2)
shall be used to verify all performance criteria. Adjustment and/or recalibration (per
Section 7) shall be performed until all performance criteria are met. Only after all
performance criteria are met may blanks and samples be analyzed.

BFB spectrum validity: The criteria in Table 4 shall be met.

Retention times: The absolute retention times of the internal standards shall be as follows:
bromochloromethane: 653 to 782 seconds; 2-bromo-l-chloropropane: 1270 to 1369 seconds;

I,4-dichlorobutane:	1510 to 1605 seconds. The relative retention times of all pollutants
and labeled compounds shall fall within the limits given in Table 3.

GC resolution: The valley height between toluene and toluene-d8 (at m/z and 99
plotted on the same graph) shall be less than 10% of the taller of the two peaks.

Calibration verification and ongoing precision and accuracy: Compute the concentration
of each pollutant (Table 1) by isotope dilution (Section 7.4) for those compounds which
have labeled analogs. Compute the concentration of each pollutant which has no labeled
analog by the internal standard method (Section 7.5). Compute the concentrations of the
labeled compounds themselves by the internal standard method. These concentrations
are computed based on the calibration data determined in Section 7.

II.5.1	For each pollutant and labeled compound, compare the concentration with the
corresponding limit for ongoing accuracy in Table 6. If all compounds meet the
acceptance criteria, system performance is acceptable and analysis of blanks and
samples may continue. If any individual value falls outside the range given,
system performance is unacceptable for that compound.

NOTE: The large number of compounds in Table 6 present a substantial probability
that one or more will fail the acceptance criteria when all compounds are
analyzed. To determine if the analytical system is out of control, or if the
failure may be attributed to probability, proceed as follows:

11.5.1.1 Analyze a second aliquot of the aqueous performance standard (Section
6.7.2).

28

11.
11.1

11.2

11.3

11.4

11.5


-------
11.5.1.2 Compute the concentration for only those compounds which failed the
first test (Section 11.5.1). If these compounds now pass, system
performance is acceptable for all compounds, and analyses of blanks
and samples may proceed. If, however, any of the compounds fail
again, the measurement system is not performing properly for these
compounds. In this event, locate and correct the problem or recalibrate
the system (Section 7), and repeat the entire test (Section 11.1) for all
compounds.

11.5.2 Add results which pass the specification in Section 11.5.1.2 to initial (Section 8.2)
and previous on-going data. Update QC charts to form a graphic representation
of laboratory performance (Figure 8). Develop a statement of accuracy for each
pollutant and labeled compound by calculating the average percent recovery (R)
and the standard deviation of percent recovery (sr). Express the accuracy as a
recovery interval from R - 2sr to R + 2sr. For example, if R = 95% and sr = 5%,
the accuracy is 85 to 105%.

12. Qualitative determination

Identification is accomplished by comparison of data from analysis of a sample or blank
with data stored in the mass-spectral libraries. For compounds for which the relative
retention times and mass spectra are known, identification is confirmed per Sections 12.1
and 12.2. For unidentified GC peaks, the spectrum is compared to spectra in the
EPA/NIH mass spectral file per Section 12.3.

12.1	Labeled compounds and pollutants having no labeled analog (Tables 1 and 2):

12.1.1	The signals for all characteristic m/z's stored in the spectral library (Section 7.2.3)
shall be present and shall maximize within the same two consecutive scans.

12.1.2	Either (1) the background corrected EICP areas or (2) the corrected relative
intensities of the mass spectral peaks at the GC peak maximum shall agree within
a factor of 2 (0.5 to 2 times) for all masses stored in the library.

12.1.3	In order for the compounds for which the system has been calibrated (Table 1) to
be identified, their relative retention times shall be within the retention-time
windows specified in Table 3.

12.1.4	The system has not been calibrated for the compounds listed in Table 2; however,
the relative retention times and mass spectra of these compounds are known.
Therefore, for a compound in Table 2 to be identified, its relative retention time
must fall within a retention-time window of ±60 seconds or ±20 scans (whichever
is greater) of the nominal retention time of the compound specified in Table 3.

12.2	Pollutants having a labeled analog (Table 1):

12.2.1	The signals for all characteristic m/z's stored in the spectral library (Section 7.2.3)
shall be present and shall maximize within the same two consecutive scans.

12.2.2	Either (1) the background corrected EICP areas or (2) the corrected relative
intensities of the mass spectral peaks at the GC peak maximum shall agree within
a factor of two for all masses stored in the spectral library.

29


-------
12.2.3 The relative retention time between the pollutant and its labeled analog shall be
within the windows specified in Table 3.

12.3	Unidentified GC peaks.

12.3.1	The signals for m/z's specific to a GC peak shall all maximize within the same
two consecutive scans.

12.3.2	Either (1) the background corrected EICP areas or (2) the corrected relative
intensities of the mass spectral peaks at the GC peak maximum shall agree within
a factor of 2 with the masses stored in the EPA/NIH mass-spectral file.

12.4	The m/z's present in the sample mass spectrum that are not present in the reference mass
spectrum shall be accounted for by contaminant or background ions. If the sample mass
spectrum is contaminated, or if identification is ambiguous, an experienced spectrometrist
(Section 1.4) is to determine the presence or absence of the compound.

13. Quantitative determination

13.1	Isotope dilution: Because the pollutant and its labeled analog exhibit the same effects
upon purging, desorption, and gas chromatography, correction for recovery of the
pollutant can be made by adding a known amount of a labeled compound to every
sample prior to purging. Relative response (RR) values for sample mixtures are used in
conjunction with the calibration curves described in Section 7.4 to determine
concentrations directly, so long as labeled compound spiking levels are constant. For the
toluene example given in Figure 7 (Section 7.4.3), RR would be equal to 1.174. For this
RR value, the toluene calibration curve given in Figure 6 indicates a concentration of 31.8
\ig/L.

13.2	Internal standard: For the compounds for which the system was calibrated (Table 1)
according to Section 7.5, use the response factor determined during the calibration to
calculate the concentration from the following equation.

Concentration = ^

(Aix x RF)

where the terms are as defined in Section 7.5.1. For the compounds for which the system
was not calibrated (Table 2), use the response factors in Table 5 to calculate the
concentration.

13.3 The concentration of the pollutant in the solid phase of the sample is computed using the
concentration of the pollutant detected in the aqueous solution, as follows:

„ ^ ^	Ti,n\ 0.005 L x aqueous cone (iig/L)

Concentration in solid \]ig/kg) = 	3	—-

0.01 x percent solids(g)

where

"percent solids" is from Section 10.1.3

13.4 Dilution of samples: If the EICP area at the quantitation m/z exceeds the
calibration range of the system, samples are diluted by successive factors of 10
until the area is within the calibration range.

30


-------
13.4.1	For aqueous samples, bring 0.50 mL, 0.050 mL, 0.0050 mL, etc., to 5-mL volume
with reagent water and analyze per Section 10.4.

13.4.2	For samples containing high solids, substitute 0.50 or 0.050 g in Section 10.5.2 to
achieve a factor of 10 or 100 dilution, respectively.

13.4.3	If dilution of high solids samples by greater than a factor of 100 is required, then
extract the sample with methanol, as described in Section 10.6.

13.5	Dilution of samples containing high concentrations of compounds not in Table 1: When
the EICP area of the quantitation m/z of a compound to be identified per Section 12.3
exceeds the linear range of the GCMS system, or when any peak in the mass spectrum
is saturated, dilute the sample per Sections 13.4.1 through 13.4.3.

13.6	Report results for all pollutants, labeled compounds, and tentatively identified compounds
found in all standards, blanks, and samples to three significant figures. For samples
containing less than 1% solids, the units are Hg/L; and for undiluted samples containing
1% solids or greater, units are |ig/kg.

13.6.1 Results for samples which have been diluted are reported at the least dilute level
at which the area at the quantitation m/z is within the calibration range (Section
13.4), or at which no m/z in the spectrum is saturated (Section 13.5). For
compounds having a labeled analog, results are reported at the least dilute level
at which the area at the quantitation m/z is within the calibration range (Section
13.4) and the labeled compound recovery is within the normal range for the
method (Section 14.2).

14. Analysis of complex samples

14.1	Some samples may contain high levels (>1000 |ig/kg) of the compounds of interest and
of interfering compounds. Some samples will foam excessively when purged. Others will
overload the trap or the GC column.

14.2	When the recovery of any labeled compound is outside the range given in Table 6, dilute
0.5 mL of samples containing less than 1% solids, or 0.5 g of samples containing 1% solids
or greater, with 4.5 mL of reagent water and analyze this diluted sample. If the recovery
remains outside of the range for this diluted sample, the aqueous performance standard
shall be analyzed (Section 11) and calibration verified (Section 11.5). If the recovery for
the labeled compound in the aqueous performance standard is outside the range given
in Table 6, the analytical system is out of control. In this case, the instrument shall be
repaired, the performance specifications in Section 11 shall be met, and the analysis of the
undiluted sample shall be repeated. If the recovery for the aqueous performance standard
is within the range given in Table 6, then the method does not apply to the sample being
analyzed, and the result may not be reported for regulatory compliance purposes.

14.3	When a high level of the pollutant is present, reverse search computer programs may
misinterpret the spectrum of chromatographically unresolved pollutant and labeled
compound pairs with overlapping spectra. Examine each chromatogram for peaks greater
than the height of the internal standard peaks. These peaks can obscure the compounds
of interest.

31


-------
15. Method performance

15.1	The specifications for this method were taken from the interlaboratory validation of EPA
Method 624 (Reference 10). Method 1624 has been shown to yield slightly better
performance on treated effluents than method 624. Results of initial tests of this method
at a purge temperature of 80°C can be found in Reference 11 and results of initial tests
of this method on municipal sludge can be found in Reference 12.

15.2	A chromatogram of the 20 Hg/L aqueous performance standards (Sections 6.7.2 and 11.1)
is shown in Figure 9.

Reference

1.	"Performance Tests for the Evaluation of Computerized Gas Chromatography/Mass
Spectrometry Equipment and Laboratories," USEPA, EMSL Cincinnati, OH 45268,
EPA-600/4-80-025 (April 1980).

2.	Bellar, T. A. and Lichtenberg, J. J., "Journal American Water Works Association," 66, 739
(1974).

3.	Bellar, T. A. and Lichtenberg, J. J., "Semi-Automated Headspace Analysis of Drinking
Waters and Industrial Waters for Purgeable Volatile Organic Compounds," in Measurement
of Organic Pollutants in Water and Wastewater, C. E. VanHall, ed., American Society for
Testing Materials, Philadelphia, PA, Special Technical Publication 686, (1978).

4.	National Standard Reference Data System, "Mass Spectral Tape Format," U.S. National
Bureau of Standards (1979 and later attachments).

5.	"Working with Carcinogens," DHEW, PHS, NIOSH, Publication 77-206 (1977).

6.	"OSHA Safety and Health Standards, General Industry," 29 CFR 1910, OSHA 2206, (1976).

7.	"Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
Committee on Chemical Safety (1979).

8.	"Handbook of Analytical Quality Control in Water and Wastewater Laboratories," USEPA,
EMSL Cincinnati, OH 45268, EPA-4-79-019 (March 1979).

9.	"Methods 330.4 and 330.5 for Total Residual Chlorine," USEPA, EMSL Cincinnati, OH
45268, EPA-4-79-020 (March 1979).

10.	"Method 624-Purgeables", 40 CFR Part 136 (49 FR 43234), 26 October 1984.

11.	"Narrative for SAS 106: Development of an Isotope Dilution GC/MS Method for Hot
Purge and-Trap Volatiles Analysis," S-CUBED Division of Maxwell Laboratories, Inc.,
Prepared for W. A. Telliard, Industrial Technology Division (WH-552), USEPA, 401 M St.
SW, Washington DC 20460 (July 1986).

12.	Colby, Bruce N. and Ryan, Philip W., "Initial Evaluation of Methods 1634 and 1635 for the
Analysis of Municipal Wastewater Treatment Sludges by Isotope Dilution GCMS," Pacific
Analytical Inc., Prepared for W. A. Telliard, Industrial Technology Division (WH-552),
USEPA, 401 M St. SW, Washington DC 20460 (July 1986).

32


-------
Appendix A Mass Spectra in the Form of Mass/Intensity Lists

532 allyl alcohol

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

42

30

43

39

44

232

45

12

53

13

55

59

56

58

57

1000

58

300

61

15









533 carbon disulfide

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

44

282

46

10

64

14

76

1000

77

27

78

82

534 2-chloro-l,3-butadiene (chloroprene)

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

48

21

49

91

50

223

51

246

52

241

53

1000

54

41

61

30

62

54

63

11

64

16

73

21

87

12

88

452

89

22

90

137









535 chloroacetonitrile

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

47

135

48

1000

49

88

50

294

51

12

73

22

74

43

75

884

76

39

77

278









536 3-chloropropene

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

35

39

36

40

40

44

42

206

47

40

58

35

49

176

51

64

52

31

61

29

73

22

75

138

76

1000

77

74

78

324













537 crotonaldehyde

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

35

26

40

28

42

339

43

48

44

335

49

27

50

40

51

20

52

21

53

31

55

55

68

24

69

511

70

1000

71

43













33


-------
Appendix A Mass Spectra in the Form of Mass/Intensity Lists (continued)

538 1,2-dibromoethane (EDB)

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

79

50

80

13

31

51

82

15

93

54

95

42

105

32

106

29

107

1000

108

38

109

922

110

19

186

13

188

27

190

13













539 dibromomethane

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

43

99

44

101

45

30

79

184

80

35

81

175

91

142

92

61

93

1000

94

64

95

875

160

18

172

375

173

14

174

719

175

12

176

342





540 trans-l,4-dichloro-2-butene

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

49

166

50

171

51

289

52

85

53

878

54

273

62

286

64

91

75

1000

77

323

88

246

89

415

90

93

91

129

124

138

126

86

128

12





541 1,3-dichloropropane

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

40

15

42

44

47

19

48

20

49

193

51

55

61

18

62

22

63

131

65

38

75

47

76

1000

77

46

78

310

79

12













542 cis-l,3-dichloropropene

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

37

262

38

269

39

998

49

596

51

189

75

1000

77

328

110

254

112

161













543 ethyl cyanide

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

44

115

50

34

51

166

52

190

53

127

54

1000

55

193





















544 ethyl methacrylate

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

m/z

Int.

42

127

43

48

45

155

55

32

58

39

68

60

69

1000

70

83

71

25

85

14

86

169

87

21

96

17

99

93

113

11

114

119









34


-------
Appendix A Mass Spectra in the Form of Mass/Intensity Lists (continued)



545 2-hexanone (methvl butvl ketone)





m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

42

61

43

1000

44

24

55

12

57

130

58

382

59

21

71

36

85

37

100

56









546 iodomethane

m/z

int.

m/z

int.

m/z

int.

m/z.

int.

m/z

int.

m/z

int.

44

57

127

328

128

17

139

39

140

34

141

120

142

1000

143

12

















547 isobutyl alcohol

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

34

21

35

13

36

13

37

11

39

10

42

575

43

1000

44

42

45

21

55

40

56

37

57

21

59

25

73

12

74

63













548 methacrylonitrile

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

38

24

39

21

41

26

42

100

49

19

50

60

51

214

52

446

53

19

62

24

63

59

64

136

65

55

66

400

67

1000

68

51









549 meth

/I methacrylate

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

42

127

43

52

45

48

53

30

55

100

56

49

59

124

68

28

69

1000

70

51

82

26

85

45

98

20

99

89

100

442

101

22









550 4-methyl-2-pentanone (methyl isoboutyl ketone; MIBK)

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

42

69

43

1000

44

54

53

11

55

15

56

13

57

205

58

346

59

20

67

12

69

10

85

96

100

94





















551 1,1,1,2-tetrachloroethane

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

47

144

49

163

60

303

61

330

62

98

82

45

84

31

95

416

96

152

97

270

98

84

117

804

121

236

131

1000

133

955

135

301









35


-------
Appendix A Mass Spectra in the Form of Mass/Intensity Lists (continued)



552 trichlorofluoromethane





m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

44

95

47

153

49

43

51

21

52

14

66

162

68

53

82

40

84

28

101

1000

102

10

103

671

105

102

117

16

119

14













553 1,2,3-trichloropro

pane

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

49

285

51

87

61

300

62

107

63

98

75

1000

76

38

77

302

83

23

96

29

97

166

98

20

99

103

110

265

111

28

112

164

114

25





554 vinyl acetate

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

36

5

42

103

43

1000

44

70

45

8

86

57

951 m-xylene

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

65

62

77

124

91

1000

105

245

106

580





951 0- + p-xylene

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

m/z

int.

51

88

77

131

91

1000

105

229

106

515





36


-------
OPTIONAL
FOAM TRAP

EXIT 1/4 IN. O.D,

10 MM GLASS FRIT
MEDIUM POROSITY

EXIT 1/4 IN. O.D.

14 MM O.D.

NLET 1/4 IN, O.D.

SAMPLE INLET

? WAY SYRINGE VALVE

17 CM 20 GAUGE SYRINGE NEEDLE

6 MM O.D. RUBBER SEPTUM

INLET 1/4 IN. O.D.

1/16 IN, O.D.
y STAINLESS STEEL

13X

MOLECULAR SIEVE
PURGE GAS FILTER

PURGE GAS
FLOW CONTROL

FIGURE 1

Purging Device for Waters


-------
PURGE INLET FITTING

SAMPLE OUTLET FITTING

3" x 6 MM O.O. GLASS TUBING

40 ML VIAL



k;

SEPTUM

CAP

FIGURE 2 Purging Device for Soils or Waters

38


-------
PACKING DETAIL	CONSTRUCTION DETAIL

FIGURE 3 Trap Construction and Packings

39


-------
CARRIER GAS
FLOW CONTROL

PRESSURE
REGULATOR

PURGE GAS
FLOW CONTROL

1'iX MOLECULAR
SIEVE FILTER

LIQUID INJECTION PORTS
i— COLUMN OVEN

B-JUUU*-,
^-lTUIAJV

CONFIRMATORY COLUMN
TO DETECTOR

ANALYTICAL COLUMN

OPTIONAL 4 PORT COLUMN
SELECTION VALVE

TRAP INLET

TRAP
22 X

PURGING
DEVICE

note„

ALL LINES BETWEEN TRAP
AND GC SHOULD BE HEATED
TO 80"C

FIGURE 4 Schematic of Purge and Trap
Device-Purge Mode

40


-------
CARRIER GAS
FLOW CONTROL

PRESSURE

regulator

liquid injection ports

	COLUMN OVEN

CONFIRMATORY COLUMN
TO DETECTOR

PURGE GAS
FLOW CONTROL

13X MOLECULAR V/
SIEVE FILTER

ANALYTICAL COLUMN

OPTIONAL 4-PORT COLUMN
SELECTION VALVE

TRAP INLET

TRAP

200 "C

NOTE:

ALL LINES BETWEEN TRAP
AND GC SHOULD BE HEATED
TO 80°C.

FIGURE 5 Schematic of Purge and Trap

Device—Desorb Mods

41


-------
HI
tfi

z

o

CL
tfi
Lit

cc

LU
>

Lit

oc

i	1	r

10 20 50
CONCENTRATION (ug/L)

FIGURE 6 Relative Response Calibration Curve for
Toluene. The Dotted Lines Enclose a +/- 10 Percent
Error Window

42


-------
(B)

AREA=60960

I±

M/Z 100

M/Z 92

FIGURE 7 Extracted Ion Current Profiles for (A)
Toluene, (B) Toluene-ds, and (C) a Mixture of
Toluene and Toluene-ds

43


-------
N

<

<
ID

DC

<

*

<
LLl

OL

120,000

100,000

80,000

+ 3s

ANALYSIS NUMBER

LU Q
% lii

a. 3

W J

Hi o

eg

LU uJ

> z

I- LU

53

0.90

t i i i	1	1	1	1	r

6/1 6/1 6/1 6/1 6/2 6/2 6/3 6/3 6/4 6/5

DATE ANALYZED

FIGURE 8 Quality Control Charts Showing Area
(top graph) and Relative Response of Toluene to
ToIuene-d8 (lower graph) Plotted as Function of
Time or Analysis Number

44


-------
MASS CHROMATOGRAM	DATA: U0A101945 II

09/01/84 23:65:00	CALI: M0AID1945 II

SAMPLE: VO,S,OPR,00020,00,U,NA:NA.NAS
CONDS.! 16248,3.0n,2«f1,3845,45-24088,156240,20HL/MINS
RANGE: G 1,1200 LABEL: N 0, 4.0 QUANi A 0, 1.0 J

SCANS

1 TO 1208

0 BASE: U 20, 3

100.0-1

47
251

w

U

222376.

46.514
250.575

400
13:40

S00
20:30

800
27:28

1000
34:10

=t

1200 SCAN
41:00 TIME

FIGURE 9 Chromatogram of Aqueous Performance Standard

45


-------