THE ENVIRONMENTAL TECHNOLOGY VERIFICATION

s>EPA

ll ** » 111 It "Muni 1.11 I'r-«I

PROGRAM £

ETV

onvl

Oak Ridge National Laboratory

Verification Statement

TECHNOLOGY TYPE:

ANODIC STRIPPING VOLTAMMETRY

APPLICATION:

MEASUREMENT OF LEAD IN DUST WIPES

TECHNOLOGY NAME:

PDV 5000 Trace Element Analyzer

COMPANY:

Monitoring Technologies International

ADDRESS:

78 Collingwood Street PHONE: +61 8 9204 3600
Osborne Park 6017 FAX: +61 8 9204 2602
Perth, Western Australia

WEB SITE:
E-MAIL:

www.mti.com.au
cgreen@colingreen.idps.co.uk

The U.S. Environmental Protection Agency (EPA) has created the Environmental Technology
Verification Program (ETV) to facilitate the deployment of innovative or improved environmental
technologies through performance verification and dissemination of information. The goal of the ETV
Program is to further environmental protection by substantially accelerating the acceptance and use of
improved and cost-effective technologies. ETV seeks to achieve this goal by providing high-quality,
peer-reviewed data on technology performance to those involved in the design, distribution, financing,
permitting, purchase, and use of environmental technologies.

ETV works in partnership with recognized standards and testing organizations and stakeholder groups
consisting of regulators, buyers, and vendor organizations, with the full participation of individual
technology developers. The program evaluates the performance of innovative technologies by developing
test plans that are responsive to the needs of stakeholders, conducting field or laboratory tests (as
appropriate), collecting and analyzing data, and preparing peer-reviewed reports. All evaluations are
conducted in accordance with rigorous quality assurance protocols to ensure that data of known and
adequate quality are generated and that the results are defensible.

Oak Ridge National Laboratory (ORNL) is one of the verification organizations operating under the
Advanced Monitoring Technology (AMT) Center. AMT, which is administered by EPA's National
Exposure Research Laboratory (NERL), is one of six technology areas under ETV. In this verification
test, ORNL evaluated the performance of lead in dust wipe measurement technologies. This verification
statement provides a summary of the test results for Monitoring Technologies International's (MTI)
PDV5000 trace metal analyzer.

EPA-VS-SCM-53

The accompanying notice is an integral part of this verification statement.

August 2002


-------
VERIFICATION TEST DESCRIPTION

This verification test was designed to evaluate technologies that detect and measure lead in dust wipes.
The test was conducted at the Capitol Community Technical College in Hartford, CT, from November 5
through November 9, 2001. The vendors of commercially-available, field portable technologies blindly
analyzed 160 dust wipe samples containing known amounts of lead, ranging in concentration from <2 to
1,500 |ig/wipe. The experimental design was particularly focused on important clearance standards,
such as those identified in 40 CFR Part 745.227(e)(8)(viii) of 40 |ig/ft2 for floors, 250 |ig/ft2 for window
sills, and 400 |ig/ft2 for window troughs. The samples included wipes newly-prepared and archived from
the Environmental Lead Proficiency Analytical Testing Program (ELPAT). These samples were prepared
from dust collected in households in North Carolina and Wisconsin. Also, newly-prepared samples were
acquired from the University of Cincinnati (UC). The UC dust wipe samples were prepared from
National Institute of Standards and Technology (NIST) Standard Reference Materials (SRMs). The
results of the lead analyses generated by the technology were compared with results from analyses of
similar samples by conventional laboratory methodology in a laboratory that was recognized as
proficient by the National Lead Laboratory Accreditation Program (NLLAP) for dust testing. Details of
the test, including a data summary and discussion of results, may be found in the report entitled
Environmental Technology Verification Report: Lead in Dust Wipe Detection Technology—Monitoring
Technologies International, PDV 5000 Trace Metal Analyzer, EPA/600/R-02/060.

TECHNOLOGY DESCRIPTION

MTI's PDV 5000, a field portable instrument, is a self-contained anodic stripping analyzer. Anodic
Stripping Voltammetry (ASV) works by electroplating metals in solution onto an electrode. This
concentrates the metal. The metals on the electrode are then sequentially stripped off, which generates a
current that can be measured. The current (milliamps) is proportional to the amount of metal being stripped
off. The potential (voltage in millivolts) at which the metal is stripped off is characteristic for each metal.
This means the metal can be identified as well as quantified.

VERIFICATION OF PERFORMANCE

The following performance characteristics of the PDV 5000 were observed:

Precision: Precision—based on the average percent relative standard deviation— 22% for the ELPAT
samples and 21% for the UC samples, excluding two outlier values.

Accuracy: Accuracy was assessed using the estimated concentrations of the ELPAT and UC samples.
The average percent recovery value for all samples reported above 30 |ig/wipe was 88% for the UC
samples and 93% for the ELPAT samples. The range of percent recoveries values was from 35% to
137%. This negative bias is statistically significant, but the average value is within the acceptable bias
range of 100% ± 25%. For the NLLAP laboratory results, the average percent recovery values were 98%
and 91%, respectively, for the ELPAT and UC samples. The negative bias for both the ELPAT and UC
samples was statistically significant.

Comparability: A comparison of the PDV 5000 results and the NLLAP-recognized laboratory results
was performed for all samples (ELPAT and UC) that were reported above 30 |ig/wipe. The correlation
coefficient (r) for the comparison to NLLAP lab results for the UC samples was 0.999 [slope (m) = 1.074,
intercept = -14.345], and for the ELPAT samples was 0.988 [m = 0.885, intercept = 15.633], While the
slopes for both the ELPAT and UC samples were statistically different than 1.00, the correlation
coefficients show a strong linear agreement with the NLLAP laboratory data.

Detectable blanks: All twenty samples, prepared at concentrations around 1 |ig/wipe, were reported	

EPA-VS-SCM-53

The accompanying notice is an integral part of this verification statement.

August 2002


-------
correctly as < 20 |ig/wipe by the PDV 5000. The instrument reported the eight samples near 17 |ig/wipe
as < 20 |ig/wipe, but the four samples around 30 |ig/wipe were reported as < 20, < 20, 24, and 25.

False positive results: A false positive result is one in which the technology reports a result that is
above the clearance level when the true (or estimated) concentration is actually below. For the UC
samples, the PDV 5000 reported four of a possible 29 results as false positives, while the NLLAP
laboratory did not report any false positives. For the ELPAT samples, the PDV 5000 reported three of a
possible 12 fp results and the NLLAP laboratory reported two of 12.

False negative results: A false negative result is one in which the technology reports a result that is below
the clearance level when the true (or estimated) concentration is actually above. For the UC samples, the
PDV 5000 reported 17 of 29 possible fn results, while the NLLAP laboratory reported 23 of 30 fn results.
For the ELPAT samples, the PDV 5000 reported 12 of a possible 28 fn results, while the NLLAP
laboratory reported seven of 12.

Completeness: The PDV 5000 generated results for all 160 dust wipe samples. However, two results for
UC samples were reported as non-detects for sample concentrations of 40 and 250 |ig/wipe. These were
considered outliers and excluded from the data analysis. Therefore, completeness was 99%.

Sample Throughput: With two analysts each operating their own instrument, the MTI team (one expert
and one notice analyst) accomplished a sample throughput rate of approximately eighty samples per 12-
hour day.

Overall Evaluation: The overall performance was characterized as having an acceptable amount of
negative bias, larger than acceptable precision, but in linear agreement with an NLLAP-recognized
laboratory's data. The verification team found that the PDV 5000 was relatively simple for the trained
analyst to operate in the field, requiring less than an hour for initial setup. As with any technology
selection, the user must determine if this technology is appropriate for the application and the project
data quality objectives. Additionally, ORNL and ETV remind the reader that, while the ETV test
provides valuable information in the form of a snapshot of performance, state, tribal, or federal
requirements regarding the use of the technologies (such as NLLAP recognition where required) need to
be followed. For more information on this and other verified technologies, visit the ETV web site at
http: //www. epa. gov/etv.

Gary J. Foley, Ph.D.

Director

National Exposure Research Laboratory
Office of Research and Development

W. Frank Harris, Ph.D.

Associate Laboratory Director
Biological and Environmental Sciences
Oak Ridge National Laboratory

NOTICE: EPA verifications are based on evaluations of technology performance under specific, predetermined
criteria and appropriate quality assurance procedures. EPA and ORNL make no expressed or implied warranties as
to the performance of the technology and do not certify that a technology will always operate as verified. The end
user is solely responsible for complying with any and all applicable federal, state, and local requirements. Mention
of commercial product names does not imply endorsement or recommendation.

EPA-VS-SCM-53

The accompanying notice is an integral part of this verification statement.

August 2002


-------