METHOD 200.9

DETERMINATION OF TRACE ELEMENTS BY STABILIZED TEMPERATURE
GRAPHITE FURNACE ATOMIC ABSORPTION

Revision 2.2
EMMC Version

J.T. Creed, T.D. Martin, L.B. Lobring, and J.W. O'Dell - Method 200.9, Revision 1.2 (1991)
J.T. Creed, T.D. Martin, and J.W. O'Dell - Method 200.9, Revision 2.2 (1994)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY
OFFICE OF RESEARCH AND DEVELOPMENT
U.S. ENVIRONMENTAL PROTECTION AGENCY
CINCINNATI, OHIO 45268

200.9-1


-------
METHOD 200.9

DETERMINATION OF TRACE ELEMENTS BY STABILIZED TEMPERATURE
GRAPHITE FURNACE ATOMIC ABSORPTION

SCOPE AND APPLICATION

1.1 This method1 provides procedures for the determination of dissolved and total
recoverable elements by graphite furnace atomic absorption (GFAA) in ground
water, surface water, drinking water, storm runoff, industrial and domestic
wastewater. This method is also applicable to the determination of total
recoverable elements in sediment, sludges, and soil. This method is applicable
to the following analytes:

Chemical Abstract Services

Analyte



Registry Number (CASRN)

Aluminum

(Al)

7429-90-5

Antimony

(Sb)

7440-36-0

Arsenic

(As)

7440-38-2

Beryllium

(Be)

7440-41-7

Cadmium

(Cd)

7440-43-9

Chromium

(Cr)

7440-47-3

Cobalt

(Co)

7440-48-4

Copper

(Cu)

7440-50-8

Iron

(Fe)

7439-89-6

Lead

(Pb)

7439-92-1

Manganese

(Mn)

7439-96-5

Nickel

(Ni)

7440-02-0

Selenium

(Se)

7782-49-2

Silver

(Ag)

7440-22-4

Thallium

(Tl)

7440-28-0

Tin

(Sn)

7440-31-5

1.2	For reference where this method is approved for use in compliance monitoring
programs [e.g., Clean Water Act (NPDES) or Safe Drinking Water Act (SDWA)]
consult both the appropriate sections of the Code of Federal Regulation (40 CFR
Part 136 Table IB for NPDES, and Part 141 § 141.23 for drinking water), and the
latest Federal Register announcements.

1.3	Dissolved analytes can be determined in aqueous samples after suitable filtration
and acid preservation.

200.9-2


-------
1.4	With the exception of silver, where this method is approved for the determination
of certain metal and metalloid contaminants in drinking water, samples may be
analyzed by direct injection into the furnace without acid digestion if the sample
has been properly preserved with acid, has turbidity of <1 NTU at the time of
analysis, and is analyzed using the appropriate method matrix modifiers. This
total recoverable determination procedure is referred to as "direct analysis".
However, in the determination of some primary drinking water metal
contaminants, such as arsenic and thallium preconcentration of the sample may
be required prior to analysis in order to meet drinking water acceptance
performance criteria (Section 10.5).

1.5	For the determination of total recoverable analytes in aqueous and solid samples
a digestion/extraction is required prior to analysis when the elements are not in
solution (e.g., soils, sludges, sediments and aqueous samples that may contain
particulate and suspended solids). Aqueous samples containing suspended or
particulate material >1% (w/v) should be extracted as a solid type sample.

1.6	Silver is only slightly soluble is the presence of chloride unless there is a sufficient
chloride concentration to form the soluble chloride complex. Therefore, low
recoveries of silver may occur in samples, fortified sample matrices and even
fortified blanks if determined as a dissolved analyte or by "direct analysis" where
the sample has not been processed using the total recoverable digestion. For this
reason it is recommended that samples be digested prior to the determination of
silver. The total recoverable sample digestion procedure given in this method is
suitable for the determination of silver in aqueous samples containing
concentrations up to 0.1 mg/L. For the analysis of wastewater samples
containing higher concentrations of silver, succeeding smaller volume, well mixed
aliquots should be prepared until the analysis solution contains <0.1 mg/L silver.
The extraction of solid samples containing concentrations of silver >50 mg/kg
should be treated in a similar manner.

1.7	Method detection limits and instrument operating conditions for the applicable
elements are listed in Table 2. These are intended as a guide and are typical of
a system optimized for the element employing commercial instrumentation.
However, actual method detection limits and linear working ranges will be
dependent on the sample matrix, instrumentation and selected operating
conditions.

1.8	The sensitivity and limited linear dynamic range (LDR) of GFAA often implies
the need to dilute a sample prior to analysis. The actual magnitude of the
dilution as well as the cleanliness of the labware used to perform the dilution can
dramatically influence the quality of the analytical results. Therefore, samples
types requiring large dilutions (>50:1) should be analyzed by an another
approved test procedure which has a larger LDR or which is inherently less
sensitive than GFAA.

1.9	Users of the method data should state the data-quality objectives prior to analysis.
Users of the method must document and have on file the required initial

200.9-3


-------
demonstration performance data described in Section 9.2 prior to using the
method for analysis.

SUMMARY OF METHOD

2.1	An aliquot of a well mixed, homogeneous aqueous or solid sample is accurately
weighed or measured for sample processing. For total recoverable analysis of a
solid or an aqueous sample containing undissolved material, analytes are first
solubilized by gentle refluxing with nitric and hydrochloric acids. After cooling,
the sample is made up to volume, is mixed and centrifuged or allowed to settle
overnight prior to analysis. For the determination of dissolved analytes in a
filtered aqueous sample aliquot, or for the "direct analysis" total recoverable
determination of analytes where sample turbidity is < 1 NTU, the sample is made
ready for analysis by the appropriate addition of nitric acid, and then diluted to
a predetermined volume and mixed before analysis.

2.2	The analytes listed in this method are determined by stabilized temperature
platform graphite furnace atomic absorption (STPGFAA). In STPGFAA, the
sample and the matrix modifier are first pipetted onto the platform or a device
which provides delayed atomization. The furnace chamber is then purged with
a continuous flow of a premixed gas (95% argon - 5% hydrogen) and the sample
is dried at a relatively low temperature (about 120°C) to avoid spattering. Once
dried, the sample is pretreated in a char or ashing step which is designed to
minimize the interference effects caused by the concomitant sample matrix. After
the char step the furnace is allowed to cool prior to atomization. The atomization
cycle is characterized by rapid heating of the furnace to a temperature where the
metal (analyte) is atomized from the pyrolytic graphite surface into a stopped gas
flow atmosphere of argon containing 5% hydrogen. (Only selenium is determined
in an atmosphere of high purity argon.) The resulting atomic cloud absorbs the
element specific atomic emission produced by a hollow cathode lamp (HCL) or
an electrodeless discharge lamp (EDL). Following analysis the furnace is
subjected to a cleanout period of high temperature and continuous argon flow.
Because the resulting absorbance usually has a nonspecific component associated
with the actual analyte absorbance, an instrumental background correction device
is required to subtract from the total signal the component which is nonspecific
to the analyte. In the absence of interferences, the background corrected
absorbance is directly related to the concentration of the analyte. Interferences
relating to STPGFAA (Section 4.0) must be recognized and corrected.
Suppressions or enhancements of instrument response caused by the sample
matrix must be corrected by the method of standard addition (Section 11.5).

200.9-4


-------
DEFINITIONS

3.1	Calibration Blank - A volume of reagent water acidified with the same acid
matrix as in the calibration standards. The calibration blank is a zero standard
and is used to auto-zero the AA instrument (Section 7.10.1).

3.2	Calibration Standard (CAL) - A solution prepared from the dilution of stock
standard solutions. The CAL solutions are used to calibrate the instrument
response with respect to analyte concentration (Section 7.9).

3.3	Dissolved Analyte - The concentration of analyte in an aqueous sample that will
pass through a 0.45 urn membrane filter assembly prior to sample acidification
(Section 11.1).

3.4	Field Reagent Blank (FRB) - An aliquot of reagent water or other blank matrix
that is placed in a sample container in the laboratory and treated as a sample in
all respects, including shipment to the sampling site, exposure to the sampling
site conditions, storage, preservation, and all analytical procedures. The purpose
of the FRB is to determine if method analytes or other interferences are present
in the field environment (Section 8.5).

3.5	Instrument Detection Limit (IDL) - The concentration equivalent to the analyte
signal which is equal to three times the standard deviation of a series of ten
replicate measurements of the calibration blank signal at the same wavelength.

3.6	Instrument Performance Check (IPC) Solution - A solution of method analytes,
used to evaluate the performance of the instrument system with respect to a
defined set of method criteria (Sections 7.11 and 9.3.4).

3.7	Laboratory Duplicates (LD1 and LD2) - Two aliquots of the same sample taken
in the laboratory and analyzed separately with identical procedures. Analyses of
LD1 and LD2 indicates precision associated with laboratory procedures, but not
with sample collection, preservation, or storage procedures.

3.8	Laboratory Fortified Blank (LFB) - An aliquot of LRB to which known quantities
of the method analytes are added in the laboratory. The LFB is analyzed exactly
like a sample, and its purpose is to determine whether the methodology is in
control and whether the laboratory is capable of making accurate and precise
measurements (Sections 7.10.3 and 9.3.2).

3.9	Laboratory Fortified Sample Matrix (LFM) - An aliquot of an environmental
sample to which known quantities of the method analytes are added in the
laboratory. The LFM is analyzed exactly like a sample, and its purpose is to
determine whether the sample matrix contributes bias to the analytical results.
The background concentrations of the analytes in the sample matrix must be
determined in a separate aliquot and the measured values in the LFM corrected
for background concentrations (Section 9.4).

200.9-5


-------
3.10	Laboratory Reagent Blank (LRB) - An aliquot of reagent water or other blank
matrices that are treated exactly as a sample including exposure to all glassware,
equipment, solvents, reagents, and internal standards that are used with other
samples. The LRB is used to determine if method analytes or other interferences
are present in the laboratory environment, reagents, or apparatus (Sections 7.10.2
and 9.3.1).

3.11	Linear Dynamic Range (LDR) - The concentration range over which the
instrument response to an analyte is linear (Section 9.2.2).

3.12	Matrix Modifier - A substance added to the graphite furnace along with the
sample in order to minimize the interference effects by selective volatilization of
either analyte or matrix components.

3.13	Method Detection Limit (MDL) - The minimum concentration of an analyte that
can be identified, measured, and reported with 99% confidence that the analyte
concentration is greater than zero (Section 9.2.4 and Table 2).

3.14	Quality Control Sample (QCS) - A solution of method analytes of known
concentrations which is used to fortify an aliquot of LRB or sample matrix. The
QCS is obtained from a source external to the laboratory and different from the
source of calibration standards. It is used to check either laboratory or instrument
performance (Sections 7.12 and 9.2.3).

3.15	Solid Sample - For the purpose of this method, a sample taken from material
classified as either soil, sediment or sludge.

3.16	Standard Addition - The addition of a known amount of analyte to the sample
in order to determine the relative response of the detector to an analyte within the
sample matrix. The relative response is then used to assess either an operative
matrix effect or the sample analyte concentration (Sections 9.5.1 and 11.5).

3.17	Stock Standard Solution - A concentrated solution containing one or more
method analytes prepared in the laboratory using assayed reference materials or
purchased from a reputable commercial source (Section 7.8).

3.18	Total Recoverable Analyte - The concentration of analyte determined to be in
either a solid sample or an unfiltered aqueous sample following treatment by
refluxing with hot dilute mineral acid(s) as specified in the method (Sections 11.2
and 11.3).

3.19	Water Sample - For the purpose of this method, a sample taken from one of the
following sources: drinking, surface, ground, storm runoff, industrial or domestic
wastewater.

200.9-6


-------
INTERFERENCES

4.1	Several interference sources may cause inaccuracies in the determination of trace
elements by GFAA. These interferences can be classified into three major
subdivisions, namely spectral, matrix, and memory.

4.2	Spectral interferences are caused by the resulting absorbance of

light by a molecule or atom which is not the analyte of interest or emission from
black body radiation.

4.2.1	Spectral interferences caused by an element only occur if there is a spectral
overlap between the wavelength of the interfering element and the analyte
of interest. Fortunately, this type of interference is relatively uncommon
in STPGFAA because of the narrow atomic line widths associated with
STPGFAA. In addition, the use of appropriate furnace temperature
programs and high spectral purity lamps as light sources can minimize the
possibility of this type of interference. However, molecular absorbances
can span several hundred nanometers producing broadband spectral
interferences. This type of interference is far more common in STPGFAA.
The use of matrix modifiers, selective volatilization, and background
correctors are all attempts to eliminate unwanted nonspecific absorbance.
The nonspecific component of the total absorbance can vary considerably
from sample type to sample type. Therefore, the effectiveness of a
particular background correction device may vary depending on the actual
analyte wavelength used as well as the nature and magnitude of the
interference. The background correction device to be used with this
method is not specified, however, it must provide an analytical condition
that is not subject to the occurring interelement spectral interferences of
palladium on copper, iron on selenium, and aluminum on arsenic.

4.2.2	Spectral interferences are also caused by the emissions from black body
radiation produced during the atomization furnace cycle. This black body
emission reaches the photomultiplier tube, producing erroneous results.
The magnitude of this interference can be minimized by proper furnace
tube alignment and monochromator design. In addition, atomization
temperatures which adequately volatilize the analyte of interest without
producing unnecessary black body radiation can help reduce unwanted
background emission during analysis.

4.3	Matrix interferences are caused by sample components which inhibit the
formation of free atomic analyte atoms during the atomization cycle.

4.3.1 Matrix interferences can be of a chemical or physical nature. In this
method the use of a delayed atomization device which provides stabilized
temperatures is required. These devices provide an environment which
is more conducive to the formation of free analyte atoms and thereby
minimize this type of interference. This type of interference can be
detected by analyzing the sample plus a sample aliquot fortified with a

200.9-7


-------
known concentration of the analyte. If the determined concentration of
the analyte addition is outside a designated range, a possible matrix effect
should be suspected (Section 9.4.3).

4.3.2 The use of nitric acid is preferred for GFAA analyses in order to minimize
vapor state anionic chemical interferences, however, in this method
hydrochloric acid is required to maintain stability in solutions containing
antimony and silver. When hydrochloric acid is used, the chloride ion
vapor state interferences must be reduced using an appropriate matrix
modifier. In this method a combination modifier of palladium,
magnesium nitrate and a hydrogen(5%)-argon(95%) gas mixture is used
for this purpose. The effects and benefits of using this modifier are
discussed in detail in Reference 2 of Section 16.0. Listed in Section 4.4 are
some typical observed effects when using this modifier.

Specific Element Interferences

Antimony: Antimony suffers from an interference produced by K2S04.3 In the
absence of hydrogen in the char cycle (1300°C), K2S04 produces a relatively high
(1.2 abs) background absorbance which can produce a false signal, even with
Zeeman background correction. However, this background level can be
dramatically reduced (0.1 abs) by the use of a hydrogen/argon gas mixture in the
char step. This reduction in background is strongly influenced by the
temperature of the char step.

Note: The actual furnace temperature may vary from instrument to instrument.
Therefore, the actual furnace temperataure should be determined on an individual
basis.

Aluminum: The palladium matrix modifier may have elevated levels of A1 which
will cause elevated blank absorbances.

Arsenic: The HC1 present from the digestion procedure can influence the
sensitivity for As. Twenty |iL of a 1% HC1 solution with Pd used as a modifier
results in a 20% loss in sensitivity relative to the analyte in a 1% HNOs solution.
Unfortunately, the use of Pd/Mg/H2 as a modifier does not significantly reduce
this suppression, and therefore, it is imperative that each sample and calibration
standard alike contain the same HC1 concentration.2

Cadmium: The HC1 present from the digestion procedure can influence the
sensitivity for Cd. Twenty |iL of a 1% HC1 solution with Pd used as a modifier
results in a 80% loss in sensitivity relative to the analyte in a 1% HNOs solution.
The use of Pd/Mg/H2 as a matrix modifier reduces this suppression to less than
10%.2

Lead: The HC1 present from the digestion procedure can influence the sensitivity
for Pb. Twenty |iL of a 1% HC1 solution with Pd used as a modifier results in a
75% loss in sensitivity relative to the analyte response in a 1% HNOs solution.

200.9-8


-------
The use of Pd/Mg/H2 as a matrix modifier reduces this suppression to less than
10%.2

Selenium: Iron has been shown to suppress Se response with continuum
background correction.3 In addition, the use of hydrogen as a purge gas during
the dry and char steps can cause a suppression in Se response if not purged from
the furnace prior to atomization.

Silver: The palladium used in the modifier preparation may have elevated levels
of Ag which will cause elevated blank absorbances.

Thallium: The HC1 present from the digestion procedure can influence the
sensitivity for Tl. Twenty |iL of a 1% HC1 solution with Pd used as a modifier
results in a 90% loss in sensitivity relative to the analyte in a 1% HNOs solution.
The use of Pd/Mg/H2 as a matrix modifier reduces this suppression to less than
10%.2

4.5 Memory interferences result from analyzing a sample containing a high
concentration of an element (typically a high atomization temperature element)
which cannot be removed quantitatively in one complete set of furnace steps. The
analyte which remains in the furnace can produce false positive signals on
subsequent sample (s). Therefore, the analyst should establish the analyte
concentration which can be injected into the furnace and adequately removed in
one complete set of furnace cycles. If this concentration is exceeded, the sample
should be diluted and a blank analyzed to assure the memory effect has been
eliminated before reanalyzing the diluted sample.

SAFETY

5.1	The toxicity or carcinogenicity of each reagent used in this method have not been
fully established. Each chemical should be regarded as a potential health hazard
and exposure to these compounds should be as low as reasonably achievable.
Each laboratory is responsible for maintaining a current awareness file of OSHA
regulations regarding the safe handling of the chemicals specified in this method.4
7 A reference file of material data handling sheets should also be made available
to all personnel involved in the chemical analysis. Specifically, concentrated nitric
and hydrochloric acids present various hazards and are moderately toxic and
extremely irritating to skin and mucus membranes. Use these reagents in a fume
hood whenever possible and if eye or skin contact occurs, flush with large
volumes of water. Always wear safety glasses or a shield for eye protection,
protective clothing and observe proper mixing when working with these reagents.

5.2	The acidification of samples containing reactive materials may result in the release
of toxic gases, such as cyanides or sulfides. Acidification of samples should be
done in a fume hood.

200.9-9


-------
5.3	All personnel handling environmental samples known to contain or to have been
in contact with human waste should be immunized against known disease
causative agents.

5.4	The graphite tube during atomization emits intense UV radiation. Suitable
precautions should be taken to protect personnel from such a hazard.

5.5	The use of the argon/hydrogen gas mixture during the dry and char steps may
evolve a considerable amount of HC1 gas. Therefore, adequate ventilation is
required.

5.6	It is the responsibility of the user of this method to comply with relevant disposal
and waste regulations. For guidance see Sections 14.0 and 15.0.

6.0 EQUIPMENT AND SUPPLIES

6.1	Graphite Furnace Atomic Absorbance Spectrophotometer

6.1.1	The GFAA spectrometer must be capable of programmed heating of the
graphite tube and the associated delayed atomization device. The
instrument must be equipped with an adequate background correction
device capable of removing undesirable non-specific absorbance over the
spectral region of interest and provide an analytical condition not subject
to the occurrence of interelement spectral overlap interferences. The
furnace device must be capable of utilizing an alternate gas supply during
specified cycles of the analysis. The capability to record relatively fast
(< 1 s) transient signals and evaluate data on a peak area basis is preferred.
In addition, a recirculating refrigeration bath is recommended for
improved reproducibility of furnace temperatures.

6.1.2	Single element hollow cathode lamps or single element electrodeless
discharge lamps along with the associated power supplies.

6.1.3	Argon gas supply (high-purity grade, 99.99%) for use during the
atomization of selenium, for sheathing the furnace tube when in operation,
and during furnace cleanout.

6.1.4	Alternate gas mixture (hydrogen 5% - argon 95%) for use as a continuous
gas flow environment during the dry and char furnace cycles.

6.1.5	Autosampler capable of adding matrix modifier solutions to the furnace,
a single addition of analyte, and completing methods of standard
additions when required.

6.2	Analytical balance, with capability to measure to 0.1 mg, for use in weighing
solids, for preparing standards, and for determining dissolved solids in digests
or extracts.

200.9-10


-------
6.3	A temperature adjustable hot plate capable of maintaining a temperature of 95°C.

6.4	(Optional) A temperature adjustable block digester capable of maintaining a
temperature of 95°C and equipped with 250 mL constricted digestion tubes.

6.5	(Optional) A steel cabinet centrifuge with guard bowl, electric timer and brake.

6.6	A gravity convection drying oven with thermostatic control capable of
maintaining 180°C ± 5°C.

6.7	(Optional) An air displacement pipetter capable of delivering volumes ranging
from 100-2500 |iL with an assortment of high quality disposable pipet tips.

6.8	Mortar and pestle, ceramic or nonmetallic material.

6.9	Polypropylene sieve, 5-mesh (4 mm opening).

6.10	Labware - All reusable labware (glass, quartz, polyethylene, PTFE, FEP, etc.)
should be sufficiently clean for the task objectives. Several procedures found to
provide clean labware include washing with a detergent solution, rinsing with tap
water, soaking for four hours or more in 20% (v/v) nitric acid or a mixture of
dilute HNOs and HC1 (1+2+9), rinsing with reagent water and storing clean.1
Ideally, ground glass surfaces should be avoided to eliminate a potential source
of random contamination. When this is impractical, particular attention should
be given to all ground glass surfaces during cleaning. Chromic acid cleaning
solutions must be avoided because chromium is an analyte.

6.10.1	Glassware - Volumetric flasks, graduated cylinders, funnels and centrifuge
tubes (glass and/or metal-free plastic).

6.10.2	Assorted calibrated pipettes.

6.10.3	Conical Phillips beakers, 250 mL with 50 mm watch glasses.

6.10.4	Griffin beakers, 250 mL with 75 mm watch glasses and (optional) 75 mm
ribbed watch glasses.

6.10.5	(Optional) PTFE and/or quartz Griffin beakers, 250 mL with PTFE covers.

6.10.6	Evaporating dishes or high-form crucibles, porcelain, 100 mL capacity.

6.10.7	Narrow-mouth storage bottles, FEP (fluorinated ethylene propylene) with
screw closure, 125 mL to 1 L capacities.

6.10.8	One-piece stem FEP wash bottle with screw closure, 125 mL capacity.
REAGENTS AND STANDARDS

200.9-11


-------
7.1	Reagents may contain elemental impurities which might affect analytical data.
Only high-purity reagents that conform to the American Chemical Society
specifications8 should be used whenever possible. If the purity of a reagent is in
question, analyze for contamination. All acids used for this method must be of
ultra high-purity grade or equivalent. Suitable acids are available from a number
of manufacturers. Redistilled acids prepared by sub-boiling distillation are
acceptable.

7.2	Hydrochloric acid, concentrated (sp.gr. 1.19) - HC1.

7.2.1	Hydrochloric acid (1 + 1) - Add 500 mL concentrated HC1 to 400 mL
reagent water and dilute to 1 L.

7.2.2	Hydrochloric acid (1+4) - Add 200 mL concentrated HC1 to 400 mL
reagent water and dilute to 1 L.

7.3	Nitric acid, concentrated (sp.gr. 1.41) - HNOs.

7.3.1	Nitric acid (1 + 1) - Add 500 mL concentrated HNOs to 400 mL reagent
water and dilute to 1 L.

7.3.2	Nitric acid (1+5) - Add 50 mL concentrated HNOs to 250 mL reagent
water.

7.3.3	Nitric acid (1+9) - Add 10 mL concentrated HNOs to 90 mL reagent water.

7.4	Reagent water. All references to water in this method refer to ASTM Type I
grade water.9

7.5	Ammonium hydroxide, concentrated (sp. gr. 0.902).

7.6	Tartaric acid, ACS reagent grade.

7.7	Matrix Modifier, dissolve 300 mg palladium (Pd) powder in conc. HNOs (1 mL
of HN03, adding 0.1 mL of concentrated HC1 if necessary). Dissolve 200 mg of
Mg(N03)2 in ASTM Type I water. Pour the two solutions together and dilute to
100 mL with ASTM Type I water.

Note: It is recommended that the matrix modifier be analyzed separately in order
to assess the contribution of the modifier to the absorbance of calibration and
reagent blank solutions.

7.8	Standard stock solutions may be purchased or prepared from ultra-high purity
grade chemicals (99.99-99.999% pure). All compounds must be dried for one hour
at 105°C, unless otherwise specified. It is recommended that stock solutions be
stored in FEP bottles. Replace stock standards when succeeding dilutions for
preparation of calibration standards can not be verified.

200.9-12


-------
CAUTION: Many of these chemicals are extremely toxic if inhaled or
swallowed (Section 5.1). Wash hands thoroughly after handling.

Typical stock solution preparation procedures follow for 1 L quantities, but for the
purpose of pollution prevention, the analyst is encouraged to prepare smaller
quantities when possible. Concentrations are calculated based upon the weight
of the pure element or upon the weight of the compound multiplied by the
fraction of the analyte in the compound.

From pure element,

Concentration = wdght

volume (L)

From pure compound,

Pnnr.pntr.tinn = weight (mg) x gravimetric factor

volume (L)

where:	gravimetric factor = the weight fraction of the analyte in the

compound

7.8.1 Aluminum solution, stock, 1 mL = 1000 |ig Al: Dissolve 1.000 g of
aluminum metal, weighed accurately to at least four significant figures, in
an acid mixture of 4.0 mL of (1+1) HC1 and 1.0 mL of concentrated HN03
in a beaker. Warm beaker slowly to effect solution. When dissolution is
complete, transfer solution quantitatively to a 1 L flask, add an additional
10.0 mL of (1 + 1) HC1 and dilute to volume with reagent water.

7.8.2 Antimony solution, stock, 1 mL = 1000 |ig Sb: Dissolve 1.000 g of
antimony powder, weighed accurately to at least four significant figures,
in 20.0 mL (1+1) HNOs and 10.0 mL concentrated HC1. Add 100 mL
reagent water and 1.50 g tartaric acid. Warm solution slightly to effect
complete dissolution. Cool solution and add reagent water to volume in
a 1 L volumetric flask.

7.8.3 Arsenic solution, stock, 1 mL = 1000 |ig As: Dissolve 1.320 g of As203
(As fraction = 0.7574), weighed accurately to at least four significant
figures, in 100 mL of reagent water containing 10.0 mL concentrated
NH4OH. Warm the solution gently to effect dissolution. Acidify the
solution with 20.0 mL concentrated HNOs and dilute to volume in a 1 L
volumetric flask with reagent water.

200.9-13


-------
7.8.4	Beryllium solution, stock, 1 mL = 1000 |ig Be: DO NOT DRY. Dissolve
19.66 g BeS04*4H20 (Be fraction = 0.0509), weighed accurately to at least
four significant figures, in reagent water, add 10.0 mL concentrated HN03,
and dilute to volume in a 1 L volumetric flask with reagent water.

7.8.5	Cadmium solution, stock, 1 mL = 1000 |ig Cd: Dissolve 1.000 g Cd metal,
acid cleaned with (1+9) HN03, weighed accurately to at least four
significant figures, in 50 mL (1+1) HNOs with heating to effect dissolution.
Let solution cool and dilute with reagent water in a 1 L volumetric flask.

7.8.6	Chromium solution, stock, 1 mL = 1000 |ig Cr: Dissolve 1.923 g Cr03
(Cr fraction = 0.5200), weighed accurately to at least four significant
figures, in 120 mL (1+5) HNOs. When solution is complete, dilute to
volume in a 1 L volumetric flask with reagent water.

7.8.7	Cobalt solution, stock, 1 mL = 1000 |ig Co: Dissolve 1.000 g Co metal,
acid cleaned with (1+9) HNOs, weighed accurately to at least four
significant figures, in 50.0 mL (1 + 1) HNOs. Let solution cool and dilute
to volume in a 1 L volumetric flask with reagent water.

7.8.8	Copper solution, stock, 1 mL = 1000 |ig Cu: Dissolve 1.000 g Cu metal,
acid cleaned with (1+9) HNOs, weighed accurately to at least four
significant figures, in 50.0 mL (1 + 1) HNOs with heating to effect
dissolution. Let solution cool and dilute in a 1 L volumetric flask with
reagent water.

7.8.9	Iron solution, stock, 1 mL = 1000 |ig Fe: Dissolve 1.000 g Fe metal, acid
cleaned with (1 + 1) HC1, weighed accurately to four significant figures, in
100 mL (1+1) HC1 with heating to effect dissolution. Let solution cool and
dilute with reagent water in a 1 L volumetric flask.

7.8.10	Lead solution, stock, 1 mL = 1000 |ig Pb: Dissolve 1.599 g PbfNO^
(Pb fraction = 0.6256), weighed accurately to at least four significant
figures, in a minimum amount of (1 + 1) HNOs. Add 20.0 mL (1 + 1) HNQ
and dilute to volume in a 1 L volumetric flask with reagent water.

7.8.11	Manganese solution, stock, 1 mL = 1000 |ig Mn: Dissolve 1.000 g of
manganese metal, weighed accurately to at least four significant figures,
in 50 mL (1 + 1) HNOs and dilute to volume in a 1 L volumetric flask with
reagent water.

7.8.12	Nickel solution, stock, 1 mL = 1000 |ig Ni: Dissolve 1.000 g of nickel
metal, weighed accurately to at least four significant figures, in 20.0 mL
hot concentrated HNOs, cool, and dilute to volume in a 1 L volumetric
flask with reagent water.

7.8.13	Selenium solution, stock, 1 mL = 1000 |ig Se: Dissolve 1.405 g SeOz
(Se fraction = 0.7116), weighed accurately to at least four significant

200.9-14


-------
figures, in 200 mL reagent water and dilute to volume in a 1 L volumetric
flask with reagent water.

7.8.14	Silver solution, stock, 1 mL = 1000 |ig Ag: Dissolve 1.000 g Ag metal,
weighed accurately to at least four significant figures, in 80 mL (1 + 1)
HNOs with heating to effect dissolution. Let solution cool and dilute with
reagent water in a 1 L volumetric flask. Store solution in amber bottle or
wrap bottle completely with aluminum foil to protect solution from light.

7.8.15	Thallium solution, stock, 1 mL = 1000 |ig Tl: Dissolve 1.303 g TlNOs
(T1 fraction = 0.7672), weighed accurately to at least four significant
figures, in reagent water. Add 10.0 mL concentrated HNOs and dilute to
volume in a 1 L volumetric flask with reagent water.

7.8.16	Tin solution, stock, 1 mL = 1000 pg Sn: Dissolve 1.000 g Sn shot, weighed
accurately to at least four significant figures, in an acid mixture of 10.0 mL
concentrated HC1 and 2.0 mL (1 + 1) HNOs with heating to effect
dissolution. Let solution cool, add 200 mL concentrated HC1, and dilute
to volume in a 1 L volumetric flask with reagent water.

7.9	Preparation of Calibration Standards - Fresh calibration standards (CAL Solution)
should be prepared every two weeks, or as needed. Dilute each of the stock
standard solutions to levels appropriate to the operating range of the instrument
using the appropriate acid diluent (see note). The element concentrations in each
CAL solution should be sufficiently high to produce good measurement precision
and to accurately define the slope of the response curve. The instrument
calibration should be initially verified using a quality control sample (Sections
7.12 and 9.2.3).

Note: The appropriate acid diluent for the determination of dissolved elements
in water and for the "direct analysis" of drinking water with turbidity <1 NTU is
1% HN03. For total recoverable elements in waters, the appropriate acid diluent
is 2% HN03 and 1% HC1, and the appropriate acid diluent for total recoverable
elements in solid samples is 2% HNOs and 2% HC1. The reason for these
different diluents is to match the types of acids and the acid concentrations of the
samples with the acid present in the standards and blanks.

7.10	Blanks - Four types of blanks are required for this method. A calibration blank
is used to establish the analytical calibration curve, the laboratory reagent blank
(LRB) is used to assess possible contamination from the sample preparation
procedure and to assess spectral background, the laboratory fortified blank is
used to assess routine laboratory performance, and a rinse blank is used to flush
the instrument autosampler uptake system. All diluent acids should be made
from concentrated acids (Sections 7.2 and 7.3) and ASTM Type I water.

7.10.1 The calibration blank consists of the appropriate acid diluent (Section 7.9
note) (HC1/HN03) in ASTM Type I water. The calibration blank should
be stored in a FEP bottle.

200.9-15


-------
7.10.2	The laboratory reagent blank (LRB) must contain all the reagents in the
same volumes as used in processing the samples. The LRB must be
carried through the same entire preparation scheme as the samples
including sample digestion, when applicable.

7.10.3	The laboratory fortified blank (LFB) is prepared by fortifying an aliquot
of the laboratory reagent blank with all analytes to provide a final
concentration which will produce an absorbance of approximately 0.1 for
each analyte. The LFB must be carried through the same entire
preparation scheme as the samples including sample digestion, when
applicable.

7.10.4	The rinse blank is prepared as needed by adding 1.0 mL of conc. HNOs
and 1.0 mL conc. HC1 to 1 L of ASTM Type I water and stored in a
convenient manner.

7.11 Instrument Performance Check (IPC) Solution - The IPC solution is used to
periodically verify instrument performance during analysis. It should be
prepared in the same acid mixture as the calibration standards (Section 7.9 note)
by combining method analytes at appropriate concentrations to approximate the
midpoint of the calibration curve. The IPC solution should be prepared from the
same standard stock solutions used to prepare the calibration standards and
stored in a FEP bottle. Agency programs may specify or request that additional
instrument performance check solutions be prepared at specified concentrations
in order to meet particular program needs.

7.12 Quality Control Sample (QCS) - For initial and periodic verification of calibration
standards and instrument performance, analysis of a QCS is required. The QCS
must be obtained from an outside source different from the standard stock
solutions and prepared in the same acid mixture as the calibration standards
(Section 7.9 note). The concentration of the analytes in the QCS solution should
be such that the resulting solution will provide an absorbance reading of
approximately 0.1. The QCS solution should be stored in a FEP bottle and
analyzed as needed to meet data-quality needs. A fresh solution should be
prepared quarterly or more frequently as needed.

SAMPLE COLLECTION. PRESERVATION. AND STORAGE

8.1	Prior to the collection of an aqueous sample, consideration should be given to the
type of data required, (i.e., dissolved or total recoverable), so that appropriate
preservation and pretreatment steps can be taken. The pH of all aqueous samples
must be tested immediately prior to aliquoting for processing or "direct analysis"
to ensure the sample has been properly preserved. If properly acid preserved, the
sample can be held up to six months before analysis.

8.2	For the determination of the dissolved elements, the sample must be filtered
through a 0.45 |im pore diameter membrane filter at the time of collection or as
soon thereafter as practically possible. (Glass or plastic filtering apparatus are

200.9-16


-------
recommended to avoid possible contamination.) Use a portion of the filtered
sample to rinse the filter flask, discard this portion and collect the required
volume of filtrate. Acidify the filtrate with (1+1) nitric acid immediately following
filtration to pH <2.

8.3	For the determination of total recoverable elements in aqueous samples, samples
are not filtered, but acidified with (1 + 1) nitric acid to pH <2 (normally, 3 mL of
(1 + 1) acid per liter of sample is sufficient for most ambient and drinking water
samples). Preservation may be done at the time of collection, however, to avoid
the hazards of strong acids in the field, transport restrictions, and possible
contamination it is recommended that the samples be returned to the laboratory
within two weeks of collection and acid preserved upon receipt in the laboratory.
Following acidification, the sample should be mixed, held for 16 hours, and then
verified to be pH <2 just prior withdrawing an aliquot for processing or "direct
analysis". If for some reason such as high alkalinity the sample pH is verified to
be >2, more acid must be added and the sample held for 16 hours until verified
to be pH <2. See Section 8.1.

Note: When the nature of the sample is either unknown or is known to be
hazardous, acidification should be done in a fume hood. See Section 5.2.

8.4	Solid samples usually require no preservation prior to analysis other than storage
at 4°C. There is no established holding time limitation for solid samples.

8.5	For aqueous samples, a field blank should be prepared and analyzed as required
by the data user. Use the same container and acid as used in sample collection.

200.9-17


-------
QUALITY CONTROL

9.1	Each laboratory using this method is required to operate a formal quality control
(QC) program. The minimum requirements of this program consist of an initial
demonstration of laboratory capability, and the periodic analysis of laboratory
reagent blanks, fortified blanks and other laboratory solutions as a continuing
check on performance. The laboratory is required to maintain performance
records that define the quality of the data thus generated.

9.2	Initial Demonstration of Performance (mandatory)

9.2.1	The initial demonstration of performance is used to characterize
instrument performance (determination of linear dynamic ranges and
analysis of quality control samples) and laboratory performance
(determination of method detection limits) prior to samples being
analyzed by this method.

9.2.2	Linear dynamic range (LDR) - The upper limit of the LDR must be
established for the wavelength utilized for each analyte by determining
the signal responses from a minimum of six different concentration
standards across the range, two of which are close to the upper limit of
the LDR. Determined LDRs must be documented and kept on file. The
linear calibration range which may be used for the analysis of samples
should be judged by the analyst from the resulting data. The upper LDR
limit should be an observed signal no more than 10% below the level
extrapolated from the four lower standards. The LDRs should be verified
whenever, in the judgement of the analyst, a change in analytical
performance caused by either a change in instrument hardware or
operating conditions would dictate they be redetermined.

Note: Multiple cleanout furnace cycles may be necessary in order to fully
define or utilize the LDR for certain elements such as chromium. For this
reason the upper limit of the linear calibration range may not correspond
to the upper LDR limit.

Determined sample analyte concentrations that exceed the upper limit of
the linear calibration range must either be diluted and reanalyzed with
concern for memory effects (Section 4.4) or analyzed by another approved
method.

9.2.3	Quality control sample (QCS) - When beginning the use of this method,
on a quarterly basis or as required to meet data-quality needs, verify the
calibration standards and acceptable instrument performance with the
preparation and analyses of a QCS (Section 7.12). If the determined
concentrations are not within ± 10% of the stated values, performance of
the determinative step of the method is unacceptable. The source of the
problem must be identified and corrected before either proceeding on with

200.9-18


-------
the initial determination of method detection limits or continuing with on-
going analyses.

9.2.4 Method detection limit (MDL) - MDLs must be established for all analytes,
using reagent water (blank) fortified at a concentration of two to three
times the estimated instrument detection limit.10 To determine MDL
values, take seven replicate aliquots of the fortified reagent water and
process through the entire analytical method. Perform all calculations
defined in the method and report the concentration values in the
appropriate units. Calculate the MDL as follows:

MDL = (t) x (S)

where:	t = Student's t value for a 99% confidence level and a

standard

deviation estimate with n-1 degrees of freedom [t = 3.14 for
seven replicates]

S = standard deviation of the replicate analyses

Note: If additional confirmation is desired, reanalyze the seven replicate
aliquots on two more nonconsecutive days and again calculate the MDL
values for each day. An average of the three MDL values for each analyte
may provide for a more appropriate MDL estimate. If the relative
standard deviation (RSD) from the analyses of the seven aliquots is <10%,
the concentration used to determine the analyte MDL may have been
inapprop-riately high for the determination. If so, this could result in the
calculation of an unrealistically low MDL. Concurrently, determination
of MDL in reagent water represents a best case situation and does not
reflect possible matrix effects of real world samples. However, successful
analyses of LFMs (Section 9.4) and the analyte addition test described in
Section 9.5.1 can give confidence to the MDL value determined in reagent
water. Typical single laboratory MDL values using this method are given
in Table 2.

The MDLs must be sufficient to detect analytes at the required levels
according to compliance monitoring regulation (Section 1.2). MDLs
should be determined annually, when a new operator begins work or
whenever, in the judgement of the analyst, a change in analytical
performance caused by either a change in instrument hardware or
operating conditions would dictate they be redetermined.

Assessing Laboratory Performance (mandatory)

9.3.1 Laboratory reagent blank (LRB) - The laboratory must analyze at least one
LRB (Section 7.10.2) with each batch of 20 or fewer samples of the same

200.9-19


-------
matrix. LRB data are used to assess contamination from the laboratory
environment. LRB values that exceed the MDL indicate laboratory or
reagent contamination should be suspected. When LRB values constitute
10% or more of the analyte level determined for a sample or is 2.2 times
the analyte MDL whichever is greater, fresh aliquots of the samples must
be prepared and analyzed again for the affected analytes after the source
of contamination has been corrected and acceptable LRB values have been
obtained.

9.3.2 Laboratory fortified blank (LFB) - The laboratory must analyze at least one
LFB (Section 7.10.3) with each batch of samples. Calculate accuracy as
percent recovery using the following equation:

R

LFB - LRB

x 100

where: R	= percent recovery

LFB = laboratory fortified blank
LRB = laboratory reagent blank

s = concentration equivalent of analyte added to fortify the
LRB solution

If the recovery of any analyte falls outside the required control limits of
85-115%, that analyte is judged out of control, and the source of the
problem should be identified and resolved before continuing analyses.

9.3.3 The laboratory must use LFB analyses data to assess laboratory
performance against the required control limits of 85-115% (Section 9.3.2).
When sufficient internal performance data become available (usually a
minimum of 20-30 analyses), optional control limits can be developed from
the mean percent recovery (x) and the standard deviation (S) of the mean
percent recovery. These data can be used to establish the upper and lower
control limits as follows:

UPPER CONTROL LIMIT = x + 3S
LOWER CONTROL LIMIT = x - 3S

The optional control limits must be equal to or better than the required
control limits of 85-115%. After each five to ten new recovery
measurements, new control limits can be calculated using only the most
recent 20-30 data points. Also, the standard deviation (S) data should be
used to established an on-going precision statement for the level of
concentrations included in the LFB. These data must be kept on file and
be available for review.

200.9-20


-------
9.3.4 Instrument performance check (IPC) solution - For all determinations the
laboratory must analyze the IPC solution (Section 7.11) and a calibration
blank immediately following each calibration, after every 10th sample (or
more frequently, if required) and at the end of the sample run. Analysis
of the calibration blank should always be less than the IDL, but greater
than a negative signal in concentration units equal to the IDL. Analysis
of the IPC solution immediately following calibration must verify that the
instrument is within ±5% of calibration. Subsequent analyses of the IPC
solution must be within ±10 % of calibration. If the calibration cannot be
verified within the specified limits, reanalyze either or both the IPC
solution and the calibration blank. If the second analysis of the IPC
solution or the calibration blank confirm the calibration to be outside the
limits, sample analysis must be discontinued, the cause determined
and/or in the case of drift the instrument recalibrated. All samples
following the last acceptable IPC solution must be reanalyzed. The
analysis data of the calibration blank and IPC solution must be kept on
file with the sample analyses data.

Assessing Analyte Recovery and Data Quality

9.4.1	Sample homogeneity and the chemical nature of the sample matrix can
affect analyte recovery and the quality of the data. Taking separate
aliquots from the sample for replicate and fortified analyses can in some
cases assess these effects. Unless otherwise specified by the data user,
laboratory or program, the following laboratory fortified matrix (LFM)
procedure (Section 9.4.2) is required. Also, the analyte addition test
(Section 9.5.1) can indicate if matrix and other interference effects are
operative in selected samples. However, all samples must demonstrate a
background absorbance <1.0 before the test results obtained can be
considered reliable.

9.4.2	The laboratory must add a known amount of each analyte to a minimum
of 10% of the routine samples. In each case the LFM aliquot must be a
duplicate of the aliquot used for sample analysis and for total recoverable
determinations added prior to sample preparation. For water samples, the
added analyte concentration must be the same as that used in the
laboratory fortified blank (Section 9.3.2). For solid samples, however, the
concentration added should be expressed as mg/kg and is calculated for
a 1 g aliquot by multiplying the added analyte concentration (ng/L) in
solution by the conversion factor 0.1 (0.001 x Hg/L x O.lL/O.OOlkg = 0.1,
Section 12.4). Over time, samples from all routine sample sources should
be fortified.

9.4.3	Calculate the percent recovery for each analyte, corrected for
concentrations measured in the unfortified sample, and compare these
values to the designated LFM recovery range of 70-130%. Recovery
calculations are not required if the concentration added is less than 25%

200.9-21


-------
of the unfortified sample concentration. Percent recovery may be
calculated in units appropriate to the matrix, using the following equation:

C - C

R = —	 x 100

s

where: R = percent recovery

Cs = fortified sample concentration
C = sample background concentration

s = concentration equivalent of analyte added to fortify
thesample

9.4.4	If the recovery of any analyte falls outside the designated LFM recovery
range (but is still within the range of calibration) and the laboratory
performance for that analyte is shown to be in control (Section 9.3), the
recovery problem encountered with the LFM is judged to be either matrix
or solution related, not system related. If the analyte recovery in the LFM
is <70% and the background absorbance is <1.0, complete the analyte
addition test (Section 9.5.1) on an undiluted portion of the unfortified
sample aliquot. The test results should be evaluated as follows:

1.	If recovery of the analyte addition test (<85%) confirms the a low
recovery for the LFM, a suppressive matrix interference is
indicated and the unfortified sample aliquot must be analyzed by
method of standard additions (Section 11.5).

2.	If the recovery of the analyte addition test is between 85-115%, a
low recovery of the analyte in the LFM (<70%) may be related to
the heterogeneous nature of the sample, the result of precipitation
loss during sample preparation, or an incorrect addition prior to
preparation. Report analyte data determined from the analysis of
the unfortified sample aliquot.

9.4.5	If laboratory performance is shown to be in control (Section 9.3), but
analyte recovery in the LFM is either >130% or above the upper
calibration limit and the background absorbance is <1.0, complete the
analyte addition test (Section 9.5.1) on a portion of the unfortified sample
aliquot. (If the determined LFM concentration is above the upper
calibration limit, dilute a portion of the unfortified aliquot accordingly
with acidified reagent water before completing the analyte addition test.)
Evaluate the test results as follows:

200.9-22


-------
1.	If the percent recovery of the analyte addition test is >115%, an
enhancing matrix interference (albeit rare) is indicated and the
unfortified sample aliquot or its appropriate dilution must be
analyzed by method of standard additions (Section 11.5).

2.	If the percent recovery of the analyte addition test is between
85-115%, high recovery in the LFM may have been caused by
random sample contamination, an incorrect addition of the analyte
prior to sample preparation, or sample heterogeneity. Report
analyte data determined from the analysis of the unfortified
sample aliquot or its appropriate dilution.

3.	If the percent recovery of the analyte addition test is <85%, either
a case of both random contamination and an operative matrix
interference in the LFM is indicated or a more plausible answer is
a heterogenous sample with an suppressive matrix interference.
Reported data should be flagged accordingly.

9.4.6	If laboratory performance is shown to be in control (Section 9.3), but the
magnitude of the sample (LFM or unfortified aliquot) background
absorbance is >1.0, a non-specific spectral interference should be
suspected. A portion of the unfortified aliquot should be diluted (1+3)
with acidified reagent water and reanalyzed. (Dilution may dramatically
reduce a molecular background to an acceptable level. Ideally, the
background absorbance in the unfortified aliquot diluted (1+3) should be
<1.0. However, additional dilution may be necessary.) If dilution reduces
the background absorbance to acceptable level (< 1.0), complete the analyte
addition test (Section 9.5.1) on a portion of the diluted unfortified aliquot.
Evaluate the test results as follows:

1.	If the recovery of the analyte addition test is between 85-115%,
report analyte data determined on the dilution of the unfortified
aliquot.

2.	If the recovery of the analyte addition test is outside the range of
85-115%, complete the sample analysis by analyzing the dilution
of the unfortified aliquot by method of standard additions
(Section 11.5).

9.4.7	If either the analysis of a LFM sample(s) or application of the analyte
addition test routine indicate an operative interference, all other samples
in the batch which are typical and have similar matrix to the LFMs or the
samples tested must be analyzed in the same manner. Also, the data user
must be informed when a matrix interference is so severe that it prevents
the successful analysis of the analyte or when the heterogeneous nature
of the sample precludes the use of duplicate analyses.

200.9-23


-------
9.4.8 Where reference materials are available, they should be analyzed to
provide additional performance data. The analysis of reference samples
is a valuable tool for demonstrating the ability to perform the method
acceptably.

9.5 The following test can be used to assess possible matrix interference effects and
the need to complete the sample analysis by method of standard additions (MSA).
Results of this test should not be considered conclusive unless the determined
sample background absorbance is <1.0. Directions for MSA are given in Section
11.5.

9.5.1 Analyte addition test: An analyte standard added to a portion of a
prepared sample, or its dilution, should be recovered to within 85-115%
of the known value. The analyte addition may be added directly to
sample in the furnace and should produce a minimum level absorbance
of 0.1. The concentration of the analyte addition plus that in the sample
should not exceed the linear calibration range of the analyte. If the
analyte is not recovered within the specified limits, a matrix effect should
be suspected and the sample must be analyzed by MSA (Section 11.5).

10.0 CALIBRATION AND STANDARDIZATION

10.1	Specific wavelengths and instrument operating conditions are listed in Table 2.
However, because of differences among makes and models of spectrophotometers
and electrothermal furnace devices, the actual instrument conditions selected may
vary from those listed.

10.2	Prior to the use of this method the instrument operating conditions must be
optimized. The analyst should follow the instructions provided by the
manufacturer while using the conditions listed in Table 2 as a guide. Of
particular importance is the determination of the charring temperature limit for
each analyte. This limit is the furnace temperature setting where a loss in analyte
will occur prior to atomization. This limit should be determined by conducting
char temperature profiles for each analyte and when necessary, in the matrix of
question. The charring temperature selected should minimize background
absorbance while providing some furnace temperature variation without loss of
analyte. For routine analytical operation the charring temperature is usually set
at least 100°C below this limit. The optimum conditions selected should provide
the lowest reliable MDLs and be similar to those listed in Table 2. Once the
optimum operating conditions are determined, they should be recorded and
available for daily reference.

10.3	Prior to an initial calibration the linear dynamic range of the analyte must be
determined (Section 9.2.2) using the optimized instrument operating conditions
(Section 10.2). For all determinations allow an instrument and hollow cathode
lamp warm up period of not less than 15 min. If an EDL is to be used, allow
30 minutes for warm up.

200.9-24


-------
10.4	Before using the procedure (Section 11.0) to analyze samples, there must be data
available documenting initial demonstration of performance. The required data
and procedure are described in Section 9.2. This data must be generated using
the same instrument operating conditions and calibration routine (Section 11.4)
to be used for sample analysis. These documented data must be kept on file and
be available for review by the data user.

10.5	In order to meet or achieve lower MDLs than those listed in Table 2 for "direct
analysis" of drinking water with turbidity <1 NTU preconcentration of the analyte
is required. This may be accomplished prior to sample introduction into the
GFAA or with the use of multiple aliquot depositions on the GFAA platform or
associated delayed atomization device. When using multiple depositions, the
same number of equal volume aliquots alike of either the calibration standards
or acid preserved samples must be deposited prior to atomization. Following
each deposition the drying cycle is completed before the next subsequent
deposition. The matrix modifier is added along with each deposition and the
total volume of each deposition must not exceed the instrument manufactures
recommended capacity of the delayed atomization device. To reduce analysis
time the minimum number of depositions required to achieve the desired
analytical result should be used. Use of this procedural technique for the "direct
analysis" of drinking water must be completed using determined optimized
instrument operating conditions for multiple depositions (Section 10.2) and
comply with the method requirements described in Sections 10.3 and 10.4. (See
Table 3 for information and data on the determination of arsenic by this
procedure.)

11.0 PROCEDURE

11.1	Aqueous Sample Preparation - Dissolved Analytes

11.1.1 For the determination of dissolved analytes in ground and surface waters,
pipet an aliquot (>20 mL) of the filtered, acid preserved sample into a
50 mL polypropylene centrifuge tube. Add an appropriate volume of
(1 + 1) nitric acid to adjust the acid concentration of the aliquot to
approximate a 1% (v/v) nitric acid solution (e.g., add 0.4 mL (1+1) HNOs
to a 20 mL aliquot of sample). Cap the tube and mix. The sample is now
ready for analysis (Section 1.3). Allowance for sample dilution should be
made in the calculations.

Note: If a precipitate is formed during acidification, transport, or storage,
the sample aliquot must be treated using the procedure described in
Sections 11.2.2 through 11.2.7 prior to analysis.

11.2	Aqueous Sample Preparation - Total Recoverable Analytes

11.2.1 For the "direct analysis" of total recoverable analytes in drinking water
samples containing turbidity < 1 NTU, treat an unfiltered acid preserved
sample aliquot using the sample preparation procedure described in

200.9-25


-------
Section 11.1.1 while making allowance for sample dilution in the data
calculation (Sections 1.2 and 1.4). For the determination of total
recoverable analytes in all other aqueous samples follow the procedure
given in Sections 11.2.2 through 11.2.7.

11.2.2	For the determination of total recoverable analytes in aqueous samples
(other than drinking water with <1 NTU turbidity), transfer a 100 mL
(±1 mL) aliquot from a well mixed, acid preserved sample to a 250 mL
Griffin beaker (Sections 1.2 and 1.6). (When necessary, smaller sample
aliquot volumes may be used.)

Note: If the sample contains undissolved solids >1%, a well mixed, acid
preserved aliquot containing no more than 1 g particulate material should
be cautiously evaporated to near 10 mL and extracted using the acid-
mixture procedure described in Sections 11.3.3 through 11.3.6.

11.2.3	Add 2 mL (1 + 1) nitric acid and 1.0 mL of (1 + 1) hydrochloric acid to the
beaker containing the measured volume of sample. Place the beaker on
the hot plate for solution evaporation. The hot plate should be located in
a fume hood and previously adjusted to provide evaporation at a
temperature of approximately but no higher than 85°C. (See the following
note.) The beaker should be covered with an elevated watch glass or
other necessary steps should be taken to prevent sample contamination
from the fume hood environment.

Note: For proper heating adjust the temperature control of the hot plate
such that an uncovered Griffin beaker containing 50 mL of water placed
in the center of the hot plate can be maintained at a temperature
approximately but no higher than 85°C. (Once the beaker is covered with
a watch glass the temperature of the water will rise to approximately
95°C.)

11.2.4	Reduce the volume of the sample aliquot to about 20 mL by gentle heating
at 85°C. DO NOT BOIL. This step takes about two hours for a 100 mL
aliquot with the rate of evaporation rapidly increasing as the sample
volume approaches 20 mL. (A spare beaker containing 20 mL of water
can be used as a gauge.)

11.2.5	Cover the lip of the beaker with a watch glass to reduce additional
evaporation and gently reflux the sample for 30 minutes. (Slight boiling
may occur, but vigorous boiling must be avoided to prevent loss of the
HCl-HzO azeotrope.)

11.2.6	Allow the beaker to cool. Quantitatively transfer the sample solution to
a 50 mL volumetric flask, make to volume with reagent water, stopper
and mix.

200.9-26


-------
11.2.7 Allow any undissolved material to settle overnight, or centrifuge a portion
of the prepared sample until clear. (If after centrifuging or standing
overnight the sample contains suspended solids that would clog or affect
the sample introduction system, a portion of the sample may be filtered
for their removal prior to analysis. However, care should be exercised to
avoid potential contamination from filtration.) The sample is now ready
for analysis. Because the effects of various matrices on the stability of
diluted samples cannot be characterized, all analyses should be performed
as soon as possible after the completed preparation.

11.3 Solid Sample Preparation - Total Recoverable Analytes

11.3.1	For the determination of total recoverable analytes in solid samples, mix
the sample thoroughly and transfer a portion (>20 g) to tared weighing
dish, weigh the sample and record the wet weight (WW). (For samples
with <35% moisture a 20 g portion is sufficient. For samples with
moisture >35% a larger aliquot 50-100 g is required.) Dry the sample to
a constant weight at 60°C and record the dry weight (DW) for calculation
of percent solids (Section 12.6). (The sample is dried at 60°C to prevent
the possible loss of volatile metallic compounds, to facilitate sieving, and
to ready the sample for grinding.)

11.3.2	To achieve homogeneity, sieve the dried sample using a 5-mesh
polypropylene sieve and grind in a mortar and pestle. (The sieve, mortar
and pestle should be cleaned between samples.) From the dried, ground
material weigh accurately a representative 1.0 ± 0.01 g aliquot (W) of the
sample and transfer to a 250 mL Phillips beaker for acid extraction
(Section 1.6).

11.3.3	To the beaker add 4 mL of (1 + 1) HNOs and 10 mL of (1+4) HC1. Cover
the lip of the beaker with a watch glass. Place the beaker on a hot plate
for reflux extraction of the analytes. The hot plate should be located in a
fume hood and previously adjusted to provide a reflux temperature of
approximately 95°C. (See the following note.)

Note: For proper heating adjust the temperature control of the hot plate
such that an uncovered Griffin beaker containing 50 mL of water placed
in the center of the hot plate can be maintained at a temperature
approximately but no higher than 85°C. (Once the beaker is covered with
a watch glass the temperature of the water will rise to approximately
95°C.) Also, a block digester capable of maintaining a temperature of
95°C and equipped with 250 mL constricted volumetric digestion tubes
may be substituted for the hot plate and conical beakers in the extraction
step.

11.3.4	Heat the sample and gently reflux for 30 minutes. Very slight boiling may
occur, however vigorous boiling must be avoided to prevent loss of the
HCl-HzO azeotrope. Some solution evaporation will occur (3-4 mL).

200.9-27


-------
11.3.5	Allow the sample to cool and quantitatively transfer the extract to a
100 mL volumetric flask. Dilute to volume with reagent water, stopper
and mix.

11.3.6	Allow the sample extract solution to stand overnight to separate insoluble
material or centrifuge a portion of the sample solution until clear. (If after
centrifuging or standing overnight the extract solution contains suspended
solids that would clog or affect the sample introduction system, a portion
of the extract solution may be filtered for their removal prior to analysis.
However, care should be exercised to avoid potential contamination from
filtration.) The sample extract is now ready for analysis. Because the
effects of various matrices on the stability of diluted samples cannot be
characterized, all analyses should be performed as soon as possible after
the completed preparation.

11.4 Sample Analysis

11.4.1	Prior to daily calibration of the instrument inspect the graphite furnace,
the sample uptake system and autosampler injector for any change in the
system that would affect instrument performance. Clean the system and
replace the graphite tube and/or platform when needed or on a daily
basis.

11.4.2	Before beginning daily calibration the instrument system should be
reconfigured to the selected optimized operating conditions as determined
in Sections 10.1 and 10.2 or 10.5 for the "direct analysis" drinking water
with turbidity < 1 NTU. Initiate data system and allow a period of not less
than 15 minutes for instrument and hollow cathode lamp warm up. If an
EDL is to be used, allow 30 minutes for warm up.

11.4.3	After the warm up period but before calibration, instrument stability must
be demonstrated by analyzing a standard solution with a concentration
20 times the IDL a minimum of five times. The resulting relative standard
deviation (RSD) of absorbance signals must be <5%. If the RSD is >5%,
determine and correct the cause before calibrating the instrument.

11.4.4	For initial and daily operation calibrate the instrument according to the
instrument manufacturer's recommended procedures using the calibration
blank (Section 7.10.1) and calibration standards (Section 7.9) prepared at
three or more concentrations within the usable linear dynamic range of the
analyte (Sections 4.4 and 9.2.2).

11.4.5	An autosampler must be used to introduce all solutions into the graphite
furnace. Once the standard, sample or QC solution plus the matrix
modifier is injected, the furnace controller completes furnace cycles and
cleanout period as programmed. Analyte signals must be integrated and
collected as peak area measurements. Background absorbances,
background corrected analyte signals, and determined analyte

200.9-28


-------
concentrations on all solutions must be able to be displayed on a CRT for
immediate review by the analyst and be available as hard copy for
documentation to be kept on file. Flush the autosampler solution uptake
system with the rinse blank (Section 7.10.4) between each solution injected.

11.4.6	After completion of the initial requirements of this method (Section 10.4),
samples should be analyzed in the same operational manner used in the
calibration routine.

11.4.7	During the analysis of samples, the laboratory must comply with the
required quality control described in Sections 9.3 and 9.4. Only for the
determination of dissolved analytes or the "direct analysis" of drinking
water with turbidity of < 1 NTU is the sample digestion step of the LRB,
LFB, and LFM not required.

11.4.8	For every new or unusual matrix, when practical, it is highly
recommended that an inductively coupled plasma atomic emission
spectrometer be used to screen for high element concentration.
Information gained from this may be used to prevent potential damage to
the instrument and to better estimate which elements may require analysis
by graphite furnace.

11.4.9	Determined sample analyte concentrations that are 90% or more of the
upper limit of calibration must either be diluted with acidified reagent
water and reanalyzed with concern for memory effects (Section 4.4), or
determined by another approved test procedure that is less sensitive.
Samples with a background absorbance >1.0 must be appropriately diluted
with acidified reagent water and reanalyzed (Section 9.4.6). If the method
of standard additions is required, follow the instructions described in
Section 11.5.

11.4.10	When it is necessary to assess an operative matrix interference
(e.g., signal reduction due to high dissolved solids), the test
described in Section 9.5 is recommended.

11.4.11	Report data as directed in Section 12.0.

200.9-29


-------
11.5 Standard Additions - If the method of standard addition is required, the following
procedure is recommended:

11.5.1 The standard addition technique11 involves preparing new standards in
the sample matrix by adding known amounts of standard to one or more
aliquots of the processed sample solution. This technique compensates for
a sample constituent that enhances or depresses the analyte signal, thus
producing a different slope from that of the calibration standards. It will
not correct for additive interference, which causes a baseline shift. The
simplest version of this technique is the single-addition method. The
procedure is as follows: Two identical aliquots of the sample solution,
each of volume Vx, are taken. To the first (labeled A) is added a small
volume Vs of a standard analyte solution of concentration C . To the
second (labeled B) is added the same volume Vs of the solvent. The
analytical signals of A and B are measured and corrected for nonanalyte
signals. The unknown sample concentration Cx is calculated:

Q 		D J J

x = (Sa-Sb) Vx

where: S A and SB = the analytical signals (corrected for the blank) of
Solutions A and B, respectively. Vs and Cs should be chosen so that SA
is roughly twice S B on the average. It is best if Y is made much less than
Vx, and thus Q i much greater than £ , to avoid excess dilution of the
sample matrix.

If a separation or concentration step is used, the additions are best made
first and carried through the entire procedure. For the results from this
technique to be valid, the following limitations must be taken into
consideration:

1.	The analytical curve must be linear.

2.	The chemical form of the analyte added must respond in the same
manner as the analyte in the sample.

3.	The interference effect must be constant over the working range of
concern.

4.	The signal must be corrected for any additive interference.

12.0 DATA ANALYSIS AND CALCULATIONS

12.1 Sample data should be reported in units of (Jg/L for aqueous samples and mg/kg
dry weight for solid samples.

200.9-30


-------
12.2	For dissolved aqueous analytes (Section 11.1) report the data generated directly
from the instrument with allowance for sample dilution. Do not report analyte
concentrations below the IDL.

12.3	For total recoverable aqueous analytes (Section 11.2), multiply solution analyte
concentrations by the dilution factor 0.5, when 100 mL aliquot is used to produce
the 50 mL final solution, round the data to the tenths place and report the data
in Hg/L up to three significant figures. If a different aliquot volume other than
100 mL is used for sample preparation, adjust the dilution factor accordingly.
Also, account for any additional dilution of the prepared sample solution needed
to complete the determination of analytes exceeding the upper limit of the
calibration curve. Do not report data below the determined analyte MDL
concentration or below an adjusted detection limit reflecting smaller sample
aliquots used in processing or additional dilutions required to complete the
analysis.

12.4	For total recoverable analytes in solid samples (Section 11.3), round the solution
analyte concentrations (pg/L) to the tenths place. Report the data up to three
significant figures as mg/kg dry-weight basis unless specified otherwise by the
program or data user. Calculate the concentration using the equation below:

Sample Cone, (mg/kg) _ C x V x D
dry -weight basis

where:	C = Concentration in the extract (mg/L)

V = Volume of extract (L, 100 mL = 0.1L)

D = Dilution factor (undiluted = 1)

W = Weight of sample aliquot extracted (g x 0.001 = kg)

Do not report analyte data below the estimated solids MDL or an adjusted MDL
because of additional dilutions required to complete the analysis.

200.9-31


-------
12.5 To report percent solids in solid samples (Section 11.3) calculate as follows:

DW

% solids (S) = x 100
WW

where:	DW = Sample weight (g) dried at 60°C

WW = Sample weight (g) before drying

Note: If the data user, program or laboratory requires that the reported percent
solids be determined by drying at 105°C, repeat the procedure given in
Section 11.3 using a separate portion (>20 g) of the sample and dry to constant
weight at 103-105°C.

12.6 The QC data obtained during the analyses provide an indication of the quality of
the sample data and should be provided with the sample results.

13.0 METHOD PERFORMANCE

13.1	Instrument operating conditions used for single laboratory testing of the method
and MDLs are listed in Table 2.

13.2	Data obtained from single laboratory testing of the method are summarized in
Table 1A-C for three solid samples consisting of SRM 1645 River Sediment, EPA
Hazardous Soil, and EPA Electroplating Sludge. Samples were prepared using
the procedure described in Section 11.3. For each matrix, five replicates were
analyzed, and an average of the replicates was used for determining the sample
background concentration. Two other pairs of duplicates were fortified at
different concentration levels. The sample background concentration, mean spike
percent recovery, the standard deviation of the average percent recovery, and the
relative percent difference between the duplicate-fortified determinations are
listed in Table 1A-C. In addition, Table 1D-F contains single-laboratory test data
for the method in aqueous media including drinking water, pond water, and well
water. Samples were prepared using the procedure described in Section 11.2. For
each aqueous matrix five replicates were analyzed, and an average of the
replicates was used for determining the sample background concentration. Four
samples were fortified at the levels reported in Table 1D-1F. A percent relative
standard deviation is reported in Table 1D-1F for the fortified samples. An
average percent recovery is also reported in Tables 1D-F.

Note: Antimony and aluminum manifest relatively low percent recoveries (see
Table 1A, NBS River Sediment 1645).

200.9-32


-------
14.0 POLLUTION PREVENTION

14.1	Pollution prevention encompasses any technique that reduces or eliminates the
quantity or toxicity of waste at the point of generation. Numerous opportunities
for pollution prevention exist in laboratory operation. The EPA has established
a preferred hierarchy of environmental management techniques that places
pollution prevention as the management option of first choice. Whenever
feasible, laboratory personnel should use pollution prevention techniques to
address their waste generation. When wastes cannot be feasibly reduced at the
source, the Agency recommends recycling as the next best option.

14.2	For information about pollution prevention that may be applicable to laboratories
and research institutions, consult "Less is Better: Laboratory Chemical
Management for Waste Reduction, available from the American Chemical
Society's Department of Government Relations and Science Policy", 1155 16th
Street N.W., Washington D.C. 20036, (202)872-4477.

15.0 WASTE MANAGEMENT

15.1 The Environmental Protection Agency requires that laboratory waste management
practices be conducted consistent with all applicable rule and regulations. The
Agency urges laboratories to protect the air, water, and land by minimizing and
controlling all releases from hoods and bench operations, complying with the
letter and spirit of any sewer discharge permits and regulations, and by
complying with all solid and hazardous waste regulations, particularly the
hazardous waste identification rules and land disposal restrictions. For further
information on waste management consult "The Waste Management Manual for
Laboratory Personnel", available from the American Chemical Society at the
address listed in the Section 15.2.

16.0 REFERENCES

1.	U.S. Environmental Protection Agency. Method 200.9, Determination of Trace
Elements by Stabilized Temperature Graphite Furnace Atomic Absorption
Spectrometry, Revision 1.2, 1991.

2.	Creed, J.T., T.D. Martin, L.B. Lobring and J.W. O'Dell. Environ. Sci. Technol.,
26:102-106, 1992.

3.	Waltz, B., G. Schlemmar and J.R. Mudakavi. JAAS, 3, 695, 1988.

4.	Carcinogens - Working With Carcinogens, Department of Health, Education, and
Welfare, Public Health Service, Center for Disease Control, National Institute for
Occupational Safety and Health, Publication No. 77-206, Aug. 1977.

5.	OSHA Safety and Health Standards, General Industry, (29 CFR 1910),
Occupational Safety and Health Administration, OSHA 2206, (Revised,
January 1976).

200.9-33


-------
6.	Safety in Academic Chemistry Laboratories, American Chemical Society
Publication, Committee on Chemical Safety, 3rd Edition, 1979.

7.	Proposed OSHA Safety and Health Standards, Laboratories, Occupational Safety
and Health Administration, Federal Register, July 24, 1986.

8.	Rohrbough, W.G. et al. Reagent Chemicals, American Chemical Society
Specifications, 7th edition. American Chemical Society, Washington, DC, 1986.

9.	American Society for Testing and Materials. Standard Specification for Reagent
Water, D1193-77. Annual Book of ASTM Standards, Vol. 11.01. Philadelphia, PA,
1991.

10.	Code of Federal Regulation 40, Ch. 1, Pt. 136, Appendix B.

11.	Winefordner, J.D., Trace Analysis: Spectroscopic Methods for Elements,
Chemical Analysis, Vol. 46, pp. 41-42.

200.9-34


-------
TABLE 1A. PRECISION AND RECOVERY DATA FOR NBS RIVER SEDIMENT 1645

Solid Sample

Certified
Value+

Average
Sed Cone
(mg/kg)

% RSD

Average
Percent
Recovery
(20 mg/kg)x

S (r)

RPD

Average
Percent
Recovery
(100 mg/kg)x

S (r)

RPD

Aluminum

22600

6810

4.6

*

_ —

	

*

— —

	

Antimony

(51)

25.8

8.2

74.9

8.3

9.5

99.0

1.5

2.7

Arsenic

(66)

69.2

3.4

69.8

19.0

12.0

89.2

4.3

7.3

Cadmium

10.2

10.8

3.7

115.3

2.6

4.0

110.7

0.7

1.7

Chromium

29600

32800

1.6

*

—

—

*

—

—

Copper

109

132

4.8

99.1

14.2

0

111.5

3.6

2.6

Manganese

785

893

5.1

*





103.2

26.4

4.7

Selenium

1.5

0.7

20.4

96.0

15.9

45.2

105.4

4.0

10.7

Silver

—

1.7

3.1

101.8

3.8

9.7

93.5

1.9

5.6

Tin

—

439

4.4

—

—

—

—

—

—

% RSD	Percent Relative Standard Deviation (n=5)

S (r)	Standard Deviation of Average Percent Recovery

RPD	Relative Percent Difference between duplicate recovery determinations

*	Fortified concentration <10% of sample concentration

—	Not determined

+	Values in parenthesis are noncertified

x	Fortified concentration


-------
TABLE IB. PRECISION AND RECOVERY DATA FOR EPA HAZARDOUS SOIL 884

Solid Sample

Average
Sed Cone
(mg/kg)

% RSD

Average
Percent
Recovery
(20 mg/kg)x

S (r)

RPD

Average
Percent
Recovery
(100 mg/kg)x

S (r)

RPD

Aluminum

6410

3.3

*

_ —

_ —

*

	

	

Antimony

4.6

14.7

61.4

2.7

7.4

60.9

1.7

7.1

Arsenic

8.7

4.6

109.8

2.1

3.5

103.7

1.5

3.6

Cadmium

1.8

10.3

115.4

0.8

1.4

99.0

4.3

12.1

Chromium

84.0

4.2

95.5

33.8

17.9

120.8

6.6

8.9

Copper

127

4.3

108.0

15.2

2.6

117.7

5.4

5.7

Manganese

453

6.0

*





99.2

13.9

1.6

Selenium

0.6

7.5

95.0

8.4

24.1

96.9

3.3

9.7

Silver

0.9

18.5

100.1

3.8

10.2

93.5

1.3

3.8

Tin

18.4

3.7

—

—

—

—

—

—

% RSD	Percent Relative Standard Deviation (n=5)

S (r)	Standard Deviation of Average Percent Recovery

RPD	Relative Percent Difference between duplicate recovery determinations

*	Fortified concentration <10% of sample concentration

—	Not determined

x	Fortified concentration


-------
TABLE 1C. PRECISION AND RECOVERY DATA FOR EPA ELECTROPLATING SLUDGE 286

Solid Sample

Average
Sed Cone
(mg/kg)

% RSD

Average
Percent
Recovery
(20 mg/kg)x

S (r)

RPD

Average
Percent
Recovery
(100 mg/kg)x

S (r)

RPD

Aluminum

6590

2.7

*

_ —

_ —

*

	

	

Antimony

7.7

3.9

68.6

2.3

5.7

60.7

3.1

12.8

Arsenic

33.7

2.7

87.6

2.6

1.7

100.2

1.5

3.1

Cadmium

119

1.3

81.9

7.9

3.0

112.5

3.9

4.7

Chromium

8070

4.5

*

—

—

*

—

—

Copper

887

1.6

*

—

—

99.5

21.9

6.0

Manganese

320

1.6

*





101.0

6.4

4.0

Selenium

0.8

6.7

99.4

0.8

2.3

96.8

0.7

1.9

Silver

6.5

2.3

102.8

2.5

5.3

92.3

1.9

5.4

Tin

21.8

3.2

—

—

—

—

—

—

% RSD	Percent Relative Standard Deviation (n=5)

S (r)	Standard Deviation of Average Percent Recovery

RPD	Relative Percent Difference between duplicate recovery determinations

*	Fortified concentration <10% of sample concentration

—	Not determined

x	Fortified concentration


-------
TABLE ID. PRECISION AND RECOVERY DATA FOR POND WATER







Fortified

% RSD at

Average



Average



Cone. |ig/L'

Fortified

Percent

Element

Cone. ]Jg/L

% RSD



Cone.2

Recovery

Ag

<0.5

*

1.25

3.7

107.5

A1

550

1.2

—

—

—

As3

3.2

4.1

10

0.8

100.5

Be

0.05

36.4

2.5

14.0

90.0

Cd

<0.05

*

0.5

4.5

99.1

Co

<0.7

*

10

2.8

97.3

Cr

0.75

8.7

2.5

1.8

98.5

Cu

2.98

11.2

10

2.9

101.9

Fe

773

5.7

—

—

—

Mn

751

2.2

—

—

—

Ni

2.11

6.8

20

1.6

105.6

Pb

1.22

20.5

25

1.8

101.6

Sb3

4

*

25

0.4

115.2

Se3

<0.8

*

25

1.6

97.8

Sn3

<0.6

*

50

3.3

117.5

T1

<1.7

75.0

50

5.2

101.0



<0.7









<	Sample concentration less than the established method detection limit

*	Not determined on sample concentrations less than the method detection limit

1	Fortified sample concentration based on 100 mL sample volumes

2	RSD are reported on 50 mL sample volumes

3	Electrodeless discharge lamps were used


-------
TABLE IE. PRECISION AND RECOVERY DATA FOR DRINKING WATER







Fortified

% RSD at

Average



Average



Cone. |ig/L'

Fortified

Percent

Element

Cone. ]Jg/L

% RSD



Cone.2

Recovery

Ag

<0.5

*

1.25

5.6

94.6

A1

163.6

2.5

150

6.4

111.7

As3

0.5

10.5

10

0.6

88.4

Be

<0.02

*

2.5

9.4

106.0

Cd

<0.05

*

0.5

6.3

105.2

Co

<0.7

*

10

3.9

88.5

Cr

<0.1

*

2.5

3.1

105.7

Cu

2.6

7.3

10

1.2

111.5

Fe

9.1

17.6

150

5.9

107.8

Mn

0.9

1.3

2.5

0.7

96.7

Ni

0.8

32.7

20

4.3

103.8

Pb

<0.7

*

10

4.0

101.8

Sb3

<0.8

*

15

14.7

101.4

Se3

<0.6

*

25

1.5

88.9

Sn3

<1.7

*

50

0.4

100.7

T1

<0.7

*

20

2.8

95.4

<	Sample concentration less than the established method detection limit

*	Not determined on sample concentrations less than the method detection limit

1	Fortified sample concentration based on 100 mL sample volumes

2	RSD are reported on 50 mL sample volumes

3	Electrodeless discharge lamps were used


-------
TABLE IF. PRECISION AND RECOVERY DATA FOR WELL WATER







Fortified

% RSD at

Average



Average



Cone. |ig/L'

Fortified

Percent

Element

Cone. ]Jg/L

% RSD



Cone.2

Recovery

Ag

<0.5

*

1.25

3.6

108.3

A1

14.4

26.7

150

1.5

97.1

As3

0.9

14.2

10

2.1

101.6

Be

<0.02

*

2.5

3.4

103.7

Cd

1.8

11.9

0.5

4.6

109.3

Co

4.0

2.9

10

1.0

95.8

Cr

<0.1

*

2.5

4.0

102.6

Cu

35.9

1.2

10

0.6

90.2

Fe

441

6.6

—

—

—

Mn

3580

2.7

—

—

—

Ni

11.8

3.2

20

4.0

105.7

Pb

<0.7

*

25

0.7

102.2

Sb3

<0.8

*

25

1.2

114.3

Se3

<0.6

*

25

1.2

95.9

Sn3

<1.7

*

50

3.0

106.1

T1

<0.7

*

50

1.4

98.0

<	Sample concentration less than the established method detection limit

*	Not determined on sample concentrations less than the method detection limit

1	Fortified sample concentration based on 100 mL sample volumes

2	RSD are reported on 50 mL sample volumes

3	Electrodeless discharge lamps were used


-------
TABLE 2. RECOMMEND GRAPHITE FURNACE OPERATING CONDITIONS
AND RECOMMENDED MATRIX MODIFIER13

Element

Wavelength

Slit

Temperature
Char

(C)5 Atom

MDL4
(Pg/L)

Ag

328.1

0.7

1000

1800

0.59

A1

309.3

0.7

1700

2600

7.89

As7

193.7

0.7

1300

2200

0.5

Be

234.9

0.7

1200

2500

0.02

Cd

228.8

0.7

800

1600

0.05

Co

242.5

0.2

1400

2500

0.7

Cr

357.9

0.7

1650

26006

0.1

Cu

324.8

0.7

1300

26006

0.7

Fe

248.3

0.2

1400

2400

-

Mn

279.5

0.2

1400

2200

0.3

Ni

232.0

0.2

1400

2500

0.6

Pb

283.3

0.7

1250

2000

0.7

Sb7

217.6

0.7

1100

2000

0.8

Se7

196.0

2.0

1000

2000

0.6

Sn7

286.3

0.7

1400s

2300

1.7

T1

276.8

0.7

1000

1600

0.7

Matrix Modifier = 0.015 mg Pd + 0.01 mg Mg(N03)2.

A 5% H2 in Ar gas mix is used during the dry and char steps at 300 mL/min.
for all elements.

A cool down step between the char and atomization is recommended.

Obtained using a 20 |iL sample size and stop flow atomization.

Actual char and atomization temperatures may vary from instrument to
instrument and are best determined on an individual basis. The actual drying
temperature may vary depending on the temperature of the water used to cool
the furnace.

A 7-s atomization is necessary to quantitatively remove the analyte from the
graphite furnace.

An electrodeless discharge lamp was used for this element.

An additional low temperature (approximately 200°C) per char is recommended.
Pd modifier was determined to have trace level contamination of this element.

200.9-41


-------
TABLE 3. MULTIPLE DEPOSITION - ARSENIC PRECISION AND

RECOVERY DATA12

Drinking Water
Source

Average
Cone. ]Jg/L

% RSD

Fortified
Cone. ]Jg/L

% RSD

Percent
Recovery

Cinti. Ohio

0.3

41%

3.8

3.9%

88%

Home Cistern

0.2

15%

4.1

1.7%

98%

Region I

0.7

7.3%

5.0

1.9%

108%

Region VI

2.6

3.4%

6.7

4.3%

103%

Region X

1.1

4.8%

5.0

1.7%

97%

NIST 1643c*

3.9

7.1%





95%

'The recommended instrument conditions given in Table 2 were used in this
procedure except for using diluted (1+2) matrix modifier and six - 36 |iL depositions
(30 |iL sample + 1 |iL reagent water + 5 |iL matrix modifier) for each determination
(Section 10.5). The amount of matrix modifier deposited on the platform with each
determination (6x5 |iL) = 0.030 mg Pd + 0.02 mg MgfNO^. The determined
arsenic MDL using this procedure is 0.1 Hg/L.

2Sample data and fortified sample data were calculated from four and five replicate
determinations, respectively. All drinking waters were fortified with 4.0 Hg/L arsenic.
The instrument was calibrated using a blank and four standard solutions (1.0, 2.5,
5.0, and 7.5 \ig/L).

*The NIST 1643c reference material Trace Elements in Water was diluted (1 + 19) for
analysis. The calculated diluted arsenic concentration is 4.1 Hg/L. The listed
precision and recovery data are from 13 replicate determinations collected over a
period of four days.

200.9-42


-------