UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
f. RESEARCH TRIANGLE PARK. NC 27711

OFFICE OF AIR QUALITY PLANNING AND STANDARDS

Technical Note- Clarifications and Guidance on Gaseous Pollutant Methods

01/30 /2018

During recent revisions to the QA Regulations (March 2016) and the QA Handbook (January 2017), EPA
has received questions from the monitoring community on a few topics that will be addressed in this
technical memo. OAQPS and ORD both agree with the clarifications and guidance below and will
eventually reflect this information in future method revisions. These clarifications and guidance are
effective immediately.

Concentration Requirement for NO Cylinder Standards

The N02 method, 40 CFR Part 50 Appendix F Section 1.3.1 requires a "gas cylinder standard containing
50 to 100 ppm NO in N2 with less than 1 ppm N02." This method has not been revised in some time and
there are NIST traceable standards lower than 50 ppm NO in N2 on the market that are now suitable for
use in this method. Therefore, lower standards that are NIST traceable may be used for calibrations.

Concentration Requirement for S02 Cylinder Standards

The S02 method, 40 CFR Part 50 Appendix A-l Section 4.1.6.1 requires a "S02gas concentration transfer
standard having a certified S02 concentration of not less than 10 ppm, in N2, traceable to a NIST
Standard Reference Material (SRM)". This method has not been revised in some time and there are NIST
traceable standards lower than 10 ppm in N2 on the market that are now suitable for use in this method.
Therefore, lower standards that are considered EPA protocol gas and NIST traceable may be used for
calibrations. The transfer standard requirements for calibrations do not apply to the standards used for
the 1-point QC checks or annual performance evaluations and therefore, lower concentration standards
can be used for these checks as long as the gasses used are EPA protocol gas and NIST traceable.

Calibration Acceptance Criteria Language

For the 03 (40 CFR part 50 App D Sec 4.5.5.6), S02 (40 CFR Part 50 App C Sec. 4.2.9) and CO (40 CFR Part
50 App C Sec 4.4.7) methods, the following sentence can be found:

"Compute (or calculate) the linear regression slope and intercept and plot the regression line to verify
that no point deviates from this line by more than 2 percent of the maximum concentration tested"

The CFR criteria could be interpreted that you can take the highest test concentration (for example 400
ppb for ozone with full scale at 500 ppb) and take 2 percent of that value (8 ppb) and allow every
calibration point to diverge from the calibration line by 8 ppb.

Since 2008, the QA Handbook has guidance that suggests:


-------
"For the gaseous pollutants the verification/calibration is considered acceptable if all calibration points
fall within 2% of the full scale, best fit straight line.

In this case, all calibration points are used to develop the calibration regression line. Once the line is
developed no individual point may deviate from the line by two percent. In addition, because the
Handbook allows for calibrations at lower concentrations than full scale1, EPA will be adding a 1.5 ppb
difference for 03, S02 and N02 and 0.03 ppm CO difference acceptance criteria for those monitoring
organization that plan to calibrate at lower concentrations. If either acceptance criteria pass, the
calibration will be acceptable. EPA suggests that monitoring organizations use the Handbook language.

N02 Dynamic Parameter Specifications (40 CFR Part 50 Appendix F Sec 1.4)

Monitoring organization have found it difficult to meet both the residence times and dynamic
parameter requirements when calibrating or auditing at lower levels.

1.4 Dynamic parameter specification.

1.4.1 The O3 generator air flowrate (Fo) and NO flowrate (Fno) must be adjusted such that the
following relationship holds:

Pr = [NO] rc x tR > 2.75 ppm-minutes

t„ =	—— < 2 minutes

where:

Pr = dynamic parameter specification, determined empirically, to insure complete reaction of
the available O3, ppm-minute

[NO] rc - NO concentration in the reaction chamber, ppm

tR = residence time of the reactant gases in the reaction chamber, minute

[NO]std = concentration of the undiluted NO standard, ppm

Fno = NO flowrate, scm3/min

Fo = O3 generator air flowrate, scm3/min

Vrc = volume of the reaction chamber, scm3

With the allowance of a lower the gas standard (20 ppm NO), ORD demonstrated their
calibration system in the following configurations (see Figure 1.) can meet the residence times
and the dynamic parameter specifications. Therefore, with the ability to use lower
concentration gas standards these parameters can be met. However, as the method gets
reviewed, ORD will review the requirements for the dynamic parameter specification.

1 The QA Handbook is using the term "calibration scale" to denote a range for calibration based on routine
concentrations measured at monitoring organization sites.


-------
Figure 1. Dynamic parameter and residence time information

NO cone

PPB

200

200

400

400

500

500

03 cone

PPB

180

10

300

10

400

10

F NO

SCCM

84.5

84.8

169.1

169.2

210.5

210.5

Fo

SCCM

129

27.8

129.1

27.8

129

27.8

FZA-Act

LPM

8.31

8.42

8.24

8.34

8.2

8.3



Minutes

0.627635

1.190053

0.449363

0.680203

0.394698

0.562316

[ N 0 ] RC



7.915691

15.06217

11.34138

17.17766

12.40059

17.66681

Pr

ppm/Min

4.968162

17.92478

5.096395

11.6843

4.894489

9.934335

PR must be >2.75

ok

ok

ok

ok

ok

ok

^rc (cm 3)	[ N 0] STD

134	20 ppm


-------