I lll ENVIRONMENTAL TECHNOLOGY VERIFICATION

PROGRAM J

ETV

wEPA

ETV Joint Verification Statement

Bafteiie

U.S. Environmental Protection Agency	putting Technotogy To Work

TECHNOLOGY TYPE: PORTABLE EMISSION ANALYZER
APPLICATION:	DETERMINING NITROGEN OXIDES EMISSIONS

TECHNOLOGY NAME: LANCOM Series II Portable Emissions Analyzer
COMPANY:	Land Combustion

ADDRESS:	10 Friends Lane	PHONE: 215-504-8000

Newtown, PA 18940	FAX: 215-504-1346

WEB SITE:	http://www.landinst.com

E-MAIL:	sales@land-combustion.com

The U.S. Environmental Protection Agency (EPA) has created the Environmental Technology Verification (ETV)
Program to facilitate the deployment of innovative or improved environmental technologies through performance
verification and dissemination of information. The goal of the ETV Program is to further environmental protec-
tion by substantially accelerating the acceptance and use of improved and cost-effective technologies. ETV seeks
to achieve this goal by providing high quality, peer reviewed data on technology performance to those involved in
the design, distribution, financing, permitting, purchase, and use of environmental technologies.

ETV works in partnership with recognized standards and testing organizations; stakeholder groups which consist
of buyers, vendor organizations, and permitters; and with the full participation of individual technology devel-
opers. The program evaluates the performance of innovative technologies by developing test plans that are
responsive to the needs of stakeholders, conducting field or laboratory tests (as appropriate), collecting and
analyzing data, and preparing peer reviewed reports. All evaluations are conducted in accordance with rigorous
quality assurance protocols to ensure that data of known and adequate quality are generated and that the results
are defensible.

The Advanced Monitoring Systems (AMS) Center, one of 12 technology areas under ETV, is operated by Battelle
in cooperation with EPA's National Exposure Research Laboratory. AMS has recently evaluated the performance
of portable nitrogen oxides monitors used to determine emissions from combustion sources. This verification
statement provides a summary of the test results for the Land LANCOM Series II Portable Emission Analyzer.

VERIFICATION TEST DESCRIPTION

The verification test described in this report was one of a series of tests conducted in April and May 2000 on
commercial portable nitrogen oxides analyzers at Battelie's facilities in Columbus, Ohio. Verification testing of
the analyzers involved (1) a series of laboratory tests in which certified NO and N02 standards were used to


-------
challenge the analyzers over a wide concentration range and (2) tests using realistic combustion sources, in which
data from the portable analyzers undergoing testing were compared to simultaneous measurements of NO and
NOx obtained with two chemiluminescent analyzers.

Verification testing lasted three to four days, of which two days were required for laboratory testing and the
remainder for source emissions testing. To assess inter-unit variability, two identical analyzers were tested
simultaneously in all tests, and results from the two analyzers were kept separate. The analyzers were operated at
all times by a representative of Land and supervised at all times by Battelle staff.

Verification testing focused on measurement of NO and N02, the sum of which is denoted as NOx. Laboratory
testing included a linearity test over the entire nominal ranges of the analyzers for both NO and N02; estimation
of detection limits and response times; interference testing; assessment of sample pressure and ambient tempera-
ture effects on analyzer response; and evaluation of zero and span drift during the various laboratory tests. Tests
with combustion sources assessed the accuracy of NO, N02, and NOx measurements, relative to the chemi-
luminescent NO/NOx approach that is the basis of EPA Method 7E. Sources used in the testing were a gas-fired
rangetop burner, a gas-fired water heater, and a diesel-powered electrical generator operated at both idle and at
high RPM. These sources produced NOx emissions ranging from less than 10 to about 350 ppm. Zero and span
drift resulting from exposure to source emissions were assessed, and analyzer stability was monitored during one
hour of uninterrupted sampling of diesel emissions.

Quality assurance (QA) oversight of verification testing was provided by Battelle. Battelle independent QA staff
conducted a technical systems audit, a performance evaluation audit, and a data quality audit of 10% of the test
data. Battelle testing staff conducted a performance evaluation audit, which was reviewed by independent QA
staff.

TECHNOLOGY DESCRIPTION

The LANCOM Series II weighs 13.2 pounds, has the dimensions of a standard laptop computer, and measures up
to seven flue gases (02, NO, N02, CO (low), CO (high), S02, and hydrocarbons). Analyzer options include
semi-continuous monitoring (pre-determined timed sampling intervals), printing, data logging (1,000 records),
and serial communications, plus various probe lengths. All gas measurements can be stored, downloaded, or
printed. The LANCOM Series II offers on-board diagnostics, accessible filters and water catchpot, and a "semi-
continuous" operating mode. It provides ppm conversions (mg/m3, lb/mBTU. etc.), oxygen normalization, and
total NOx, on a wet or dry basis.

The LANCOM Series II systems components are mounted on molded PVC and sheathed in corrosion-resistant
plastic. The analyzer can be operated when worn on a shoulder strap or free-standing on the ground. All controls
are on the top of the instrument. The batteries are mounted at the bottom of the case, which provides enhanced
stability when the instrument is on the floor. The large capacity water catchpot is mounted on the side of the
instrument on a hinged assembly. The particulate and chemical filters are also mounted on the side of the instru-
ment. All measured parameters and operator interface are displayed on a full function alphanumeric/graphic
liquid crystal display. The LANCOM Series II contains two 6V batteries capable of powering the instrument for
eight hours in the field.

VERIFICATION OF PERFORMANCE

Linearity: The Land Combustion LANCOM Series II analyzers provided linear response for NO and N02 over
the tested ranges of 0 to 2,000 ppm and 0 to 512 ppm, respectively. One of the LANCOM units did exhibit a
slightly low response to N02 above about 250 ppm, perhaps as a result of an older sensor used in that unit.

Detection Limit: Detection limits estimated from these wide-range linearity tests were about 2.5 ppm for NO and
1.5 to 2.3 ppm for N02.


-------
Response Time: Response times of the two LANCOM Series II analyzers were 35 and 39 seconds for NO and 77
and 90 seconds for N02, respectively.

Zero/Span Drift: Drift in LANCOM Series II NO and N02 zero readings before and after laboratory and
combustion source tests was within ± 2 ppm in nearly all circumstances. In laboratory tests, span drift for NO was
within about 2%, using 1,400 to 2,000 ppm NO span levels. For NOz, span drift was within 1%, using 350 to 512
ppm NOz span levels. In sampling gas combustion and diesel sources, absolute NO and N02 span drift was
usually within 1 ppm, and span drift exceeded 2% of the span gas value only at span gas values of 10 to 20 ppm,
reflecting the ± 1 ppm resolution of the analyzers.

Interferences: No interference was found from any of the following: 496 ppm CO; 5.03% C02; 494 ppm NH3;
605 ppm of total hydrocarbons; 501 ppm of S02; or 451 ppm S02 in the presence of 393 ppm NO.

Pressure Sensitivity: Over the tested range of + 10 to - 10 in. H20, sample gas pressure had no significant effect
on LANCOM Series II zero or span readings.

Ambient Temperature: Variations in ambient temperature over the range of 7 to 39° C (45 to 103 °F) had no
effect on the LANCOM Series II zero readings for N02, but a small effect was seen for NO, with higher tempera-
ture increasing zero readings by a few ppm. Over that entire temperature range, span response increased with
increasing temperature by 7 to 10% for NO, and by about 4% for N02.

Relative Accuracy: The relative accuracy of the LANCOM Series II analyzers for NOx ranged from 1.8 to 17.5%
over both analyzers with natural gas and diesel combustion sources. At NO and N02 levels of 6 ppm or less, the
LANCOM Series II analyzers were accurate to within about their 1 ppm measurement resolution.

Inter-Unit Repeatability: Comparison of verification results from the two LANCOM Series II analyzers showed
some slight unit-to-unit differences, primarily in relative accuracy; but overall the performance of the two
analyzers was essentially the same. Unit-to-unit agreement for NOx in source sampling ranged from 0.5 to 7.7%,
and was comparable to that of the reference analyzers.

Gabor J. Kovacs	Date	Gary J. Foley	Date

Vice President	Director

Environmental Sector	National Exposure Research Laboratory

Battelle	Office of Research and Development

U.S. Environmental Protection Agency

NOTICE: EPA verifications are based on an evaluation of technology performance under specific,
predetermined criteria and the appropriate quality assurance procedures. EPA and Battelle make no expressed or
implied warranties as to the performance of the technology and do not certify that a technology will always
operate as verified. The end user is solely responsible for complying with any and all applicable federal, state,
and local requirements. Mention of commercial product names does not imply endorsement.


-------