Method 1663

Differentiation of Diesel and
Crude Oil by GC/FID


-------
Method 1663

Differentiation of Diesel and Crude Oil by GC/FID

1.	SCOPE AND APPLICA TION

1.1	This method differentiates between diesel oil and crude oil in drilling muds and
other sources by comparing the ratio of n-alkanes in the C9-C30 range as
determined by gas chromatography (GC) with a flame-ionization detector (FID).

1.2	This method is for use in the Environmental Protection Agency's survey programs
and may be used for compliance monitoring as part of the "Effluent Limitations
Guidelines and New Source Performance Standards for the Offshore Subcategory
of the Oil and Gas Extraction Point Source Category" [50 FR 34592],

1.3	For oil in drilling muds, this method is designed to be used in conjunction with the
extraction procedure in EPA Method 1662.

1.4	This method cannot differentiate between mineral oil and diesel crude oil. EPA
Method 1654A can be used to determine that the oil is not mineral oil by
measurement of the polynuclear aromatic (PAH) content.

1.5	When used in conjunction with EPA Method 1662, the estimated detection limit
for diesel or crude oil in drilling mud is 100 mg/kg, excluding interferences caused
by other materials in the mud.

1.6	Any modification of this method beyond those expressly permitted shall be
considered as a major modification subject to application and approval of
alternative test procedures under 40 CFR 136.4 and 136.5.

1.7	The gas chromatography portions of this method are restricted to use by or under
the supervision of analysts experienced in the use of gas chromatography and in
the interpretation of gas chromatograms. Each laboratory that uses this method
must generate acceptable results using the procedures described in Sections 7.1,
8.2, and 12 of this method.

2.	SUMMAR Y OF METHOD

2.1	An oil sample is diluted in methylene chloride. An internal standard is added and an
aliquot is injected into a gas chromatograph (GC). The components of the oil are
separated by the GC and detected by a flame-ionization detector (FID).

2.2	Identification of diesel oil or crude oil (qualitative analysis) is performed by
comparing ratios of groups of n-alkanes.

2.3	Quantitative analysis is performed by calibrating the GC/FID with hexadecane
using an internal standard technique. The calibration factor is then used to
determine the amounts of the groups of n-alkanes. A quotient of these amounts
establishes that the oil is diesel or crude.

2.4	Quality is assured through reproducible calibration and testing of the extraction
and GC systems.


-------
3.	CONTAMINATION AND INTERFERENCES

3.1	Solvents, reagents, glassware, and other sample processing hardware may yield
artifacts and/or elevated baselines causing misinterpretation of chromatograms.

3.1.1	All materials used in the analysis shall be demonstrated to be free from
interferences by running method blanks initially and with a sample batch
(samples started through the extraction process at the same time, to a
maximum of ten). Specific selection of reagents and purification of solvents
by distillation in all-glass systems may be required.

3.1.2	Glassware and, where possible, reagents are cleaned by rinsing with solvent
or baking at 450°C for a minimum of 1 hour.

3.2	Interferences co-extracted from samples may vary from source to source,
depending on the diversity of the site being sampled.

4.	SAFETY

4.1	The toxicity or carcinogenicity of each reagent used in this method has not been
defined. Therefore, each chemical compound should be treated as a potential
health hazard. From this viewpoint, exposure to these chemicals must be reduced
to the lowest possible level by whatever means available.

4.2	The laboratory is responsible for maintaining a current awareness file of OSHA
regulations regarding the safe handling of the chemicals specified in this method. A
reference file of material safety data sheets (MSDSs) should also be made available
to all personnel involved in the chemical analysis. Additional references to
laboratory safety can be found in References 1 through 3.

5.	APPARA TUS AND MA TERIALS

NOTE: Brand names, suppliers, and part numbers are for illustrative purposes only. No
endorsement is implied. Equivalent performance maybe achieved using apparatus
and materials other than those specified here, but demonstration of equivalent
performance meeting the requirements of this method is the responsibility of the
	laboratory.	

5.1	Equipment for glassware cleaning.

5.1.1	Laboratory sink with overhead fume hood.

5.1.2	Kiln: Capable of reaching 450°C within 2 hours and holding 450°C within
±10°C, with temperature controller and safety switch (Cress
Manufacturing Co., Santa Fe Springs, CA. B31H or X31TS, or
equivalent).

5.2	Equipment for sample preparation.

5.2.1	Laboratory fume hood.

5.2.2	Analytical balance: Capable of weighing 0.1 mg.

5.2.3	Glassware.


-------
5.2.3.1	Disposable pipettes: Pasteur, 150 mm long by 5 mm i.d. (Fisher
Scientific 13-678-6A, or equivalent).

5.2.3.2	Glass pipettes: 0.1-, 1.0-, and 10-mL, accurate to 1% or better.

5.2.3.3	Volumetric flasks: Glass, 10- and 100-mL.

5.2.3.4	Sample vials: Amber glass, 1- to 3-mL with PTFE-lined screw- or
crimp-cap, to fit GC autosampler.

5.3	Gas Chromatograph (GC): Analytical system with split injection, capillary column,
temperature program with initial and final isothermal holds, and all required
accessories, including syringes, analytical columns, gases, detector, and recorder.
The analytical system shall meet the performance specifications in Section 12.

5.3.1	Column: 30 m long (±5 m) by 0.25 mm i.d. (±0.02 mm), 99% methyl, 1%
vinyl, 1.0 |im film thickness, bonded-phase fused-silica capillary (Supelco
SPB-1, or equivalent).

5.3.2	Detector: Flame ionization. Capable of detecting 10 ng of hexadecane.

5.4	GC data system: Shall collect and record GC data, store GC runs in magnetic
memory or on magnetic disk or tape, process GC data, compute peak areas, store
calibration data including retention times and the response factor, identify GC
peaks through retention times, and compute concentrations.

5.4.1	Data acquisition: GC data shall be collected continuously throughout the
analysis and stored on a magnetic storage device.

5.4.2	Response factor: The data system shall be used to record and maintain the
response factor (Section 7). Computations of relative standard deviation
(coefficient of variation; CV) are used for testing calibration linearity.
Statistics on initial (Section 8.2) and ongoing (Section 12.5) performance
shall be computed and maintained.

5.4.3	Data processing: The data system shall search, locate, identify, and quantify
the compounds of interest in each GC analysis. Software routines shall be
employed to compute and record retention times and peak areas. Displays
of chromatograms and library comparisons are required to verify results.

6. REAGENTS

6.1	Methylene chloride: ACS grade or equivalent.

6.2	Standards: Purchased as solutions or mixtures with certification as to their purity,
concentration, and authenticity, or prepared from materials of known purity and
composition. If compound purity is 96% or greater, the weight may be used
without correction to compute the concentration of the standard. If diesel oil in
drilling mud is to be tested, the diesel oil standard used in this method should be
from the diesel oil added to the mud on the drilling rig from which the mud sample
is taken. If this oil is not available, No. 2 diesel oil from a local source may be
substituted.

6.2.1 Stock solutions: Prepare in methylene chloride for injection into the GC.

Observe the safety precautions in Section 4.

6.2.1.1 Diesel oil (62.5 mg/mL): If QC extracts from Method 1662 are to
be tested, use the oil that was spiked to produce these extracts.


-------
Weigh 6.25 g of diesel oil to three significant figures in a 100-mL
ground-glass stoppered volumetric flask and fill to the mark with
methylene chloride. After the oil is completely dissolved, transfer
the solution to a 150-mL bottle with PTFE-lined cap.

6.2.1.2	Normal hydrocarbons - decane (C12) hexadecane (C16), and
tetracosane (C24): Dissolve an appropriate amount of reference
material in a suitable solvent. For example, weigh 10.0 mg of
decane in a 10-mL volumetric flask and fill to the mark with
methylene chloride. After the decane is completely dissolved,
transfer the solution to a 15-mL vial with PTFE-lined cap.

6.2.1.3	Internal standard: Dissolve 1.0 g of 1,3,5-trichlorobenzene (TCB,
Kodak No. 1801 or equivalent) in 100 mL methylene chloride.

After the TCB is completely dissolved, transfer the solution to a
150-mL bottle with PTFE-lined cap. Label with the concentration
and date. Mark the level of the meniscus on the bottle to detect
solvent loss.

6.2.1.4	Stock solutions should be checked for signs of degradation prior to
the preparation of calibration or performance test standards.

6.2.2	Normal hydrocarbon calibration standards (GAL): Dilute and mix the stock
solutions (Section 6.2.1.2) in methylene chloride to produce the calibration
standards shown in Table 1. The three solutions permit the response of
hexadecane to be measured as a function of concentration, and decane and
tetracosane permit the retention-time window of diesel oil to be defined.
The medium-level solution is used for calibration verification.

6.2.3	Precision and recovery standard: Dilute the stock solution of diesel oil
(Section 6.2.1.1) to produce a concentration of 1.25 mg/mL in methylene
chloride. This standard is used for initial precision and recovery (IPR,
Section 8.2) and ongoing precision and recovery (OPR, Section 12.5).

6.2.4	Addition of internal standard: Using a micropipette or microsyringe,
transfer 100 |iL of each standard solution (Section 6.2.2 or 6.2.3) to a GC
injection vial. Add 100 |iL of the TCB internal standard (Section 6.2.1.3)
to each vial and mix thoroughly. Calibration and precision and recovery
standards are made fresh daily to avoid solvent loss by evaporation.

6.2.5	Stability of standards.

6.2.5.1	When not being used, standards are stored in the dark at -20 to
-10°C in screw-capped vials with PTFE-lined lids. A mark is placed
on the vial at the level of the solution so that solvent loss by
evaporation can be detected. The vial is brought to room
temperature prior to use. Any precipitate is redissolved and solvent
is added if solvent loss has occurred.

6.2.5.2	Standard solutions used for quantitative purposes (Sections 6.2.1
through 6.2.3) shall be analyzed within 48 hours of preparation and
on a monthly basis thereafter for signs of degradation. Standards
will remain acceptable if the peak area remains within ±15% of the
area obtained in the initial analysis of the standard.


-------
7.

CALIBRATION

7.1 Establish gas chromatographic operating conditions given in Table 2. Verify that
the GC meets the minimum level in Table 4. The gas chromatograph is calibrated
	using the internal standard technique.	

NOTE:Because each GC is slightly different, it may be necessary to adjust the operating

conditions (carrier gas flow rate and column temperature and temperature
	program) slightly until the retention times in Table 3 are met.	

7.2 Internal standard calibration procedure: 1 ,3,5-Trichlorobenzene (TCB) has been
shown to be free of interferences from the diesel and crude oils tested in the
development of this method. Check for acceptability by injecting 0.5 |iL of the
internal standard solution (Section 6.2.1.3) into the GC\FID. If a major peak other
than the TCB peak appears in the chromatogram, interference with the peaks used
for determination of diesel/crude oil may occur. In this case, the analyst must
choose an alternative internal standard that is free from interferences.

7.2.1	Inject 1 |iL of each calibration standard containing the internal standard
(Table 1 and Section 6.2.2) into the GC\FID. The TCB will elute approx
8.5 minutes after injection. For the GC\FID used in the development of this
method, the TCB internal standard peak was 30 to 50% of full scale at an
attenuator setting of 8 x 10"11 amp.

7.2.2	Response factor of hexadecane (C16).

7.2.2.1 Tabulate the peak areas against concentration for the TCB and C16
peaks. Calculate response factors (RF) at each concentration for
C16 using the following equation:

Equation 1

RF U)fe.)

IMH

where:

As = Area of the peak to be measured

Cis = Concentration of the internal standard, in ng/kg

Ais = Area of the internal standard peak

Cx = Concentration of the peak to be measured, in ug kg

7.2.2.2	Calculate the average, standard deviation, and relative standard
deviation (RSD) of the response factors. If the RF is constant
(<15% RSD) over the calibration range, linearity through the origin
can be assumed and system performance is acceptable; if not, the
system must be recalibrated.

7.2.2.3	The average response factor is verified on each working 8-hour
shift by measurement of the medium-level calibration standard
(Section 12.4).

7.2.3 Single-point calibration for diesel oil: Inject the precision and recovery standard
(Section 6.2.3) to which the internal standard has been added (Section 6.2.4) to
produce a single calibration point for diesel oil.

7.2.3.1	Integrate the area of the C16 peak.

7.2.3.2	Determine the response factor for diesel oil using Equation 1.


-------
8. QUALITY ASSURANCE/QUALITY CONTROL

8.1	Each laboratory that uses this method is required to operate a formal quality
assurance program (Reference 4). The minimum requirements of this program
consist of an initial demonstration of laboratory capability, an ongoing analysis of
standards and blanks as a test of continued performance, analyses of spiked
samples to assess accuracy, and analysis of duplicates to assess precision.
Laboratory performance is compared to established performance criteria to
determine if the results of analyses meet the performance characteristics of the
method. If the determination of diesel/crude oil is to be made on extracts from
Method 1662, the quality control samples for initial precision and recovery (IPR),
spiked samples, duplicates, and ongoing precision and recovery (OPR) from
Method 1662 shall be substituted for those in the QC tests below, and the
specifications in Table 4 of this method for extracts from Method 1662 shall be
met.

8.1.1	The analyst shall make an initial demonstration of the ability to generate
acceptable accuracy and precision with this method. This ability is
established as described in Section 8.2.

8.1.2	The analyst is permitted to modify this method to improve separations or
lower the costs of measurements, provided all performance requirements
are met. Each time a modification is made to the method, the analyst is
required to achieve the minimum level (Section 7.1) and to repeat the
procedure in Section 8.2 to demonstrate method performance.

8.1.3	Analyses of spiked samples are required to demonstrate method accuracy
when extracts from Method 1662 are analyzed. The procedure and QC
criteria for spiking are described in Section 8.3.

8.1.4	Analyses of duplicate samples are required to demonstrate method
precision when extracts from Method 1662 are analyzed. The procedure
and QC criteria for duplicates are described in Section 8.4.

8.1.5	Analyses of blanks are required to demonstrate freedom from
contamination. The procedures and criteria for analysis of a blank are
described in Section 8.5.

8.1.6	The laboratory shall, on an ongoing basis, demonstrate through calibration
verification and the analysis of the precision and recovery standard (Section
6.2.3) that the analysis system is in control. These procedures are described
in Section 12.

8.1.7	The laboratory shall maintain records to define the quality of data that is
generated. Development of accuracy statements is described in Sections
8.3.2 and 12.5.4.

8.2	Initial precision and accuracy: The initial precision and recovery test is performed
using the precision and recovery standard. If extracts from Method 1662 are to be
analyzed, the extracts from the initial precision and recovery test in that method
shall be used; otherwise, the laboratory shall generate acceptable precision and
recovery by performing the following operations.

8.2.1 Using diesel oil, prepare four separate aliquots of the precision and

recovery standard (Section 6.2.3) using the procedure in Section 10. Add
the internal standard to each aliquot (Section 6.2.4). Analyze these
aliquots using the procedure in Section 11.


-------
8.2.2	Using results of the set of four analyses, compute the average recovery (X)
in mg/mL and the standard deviation of the recovery (s) in mg/mL for each
sample by the internal standard method (Sections 7.2 and 14.2).

8.2.3	For each compound, compare s and X with the corresponding limits for
initial precision and accuracy in Table 4. If s and X meet the acceptance
criteria, system performance is acceptable and analysis of samples may
begin. If, however, s exceeds the precision limit or X falls outside the range
for accuracy, system performance is unacceptable. In this event, review this
method, correct the problem, and repeat the test.

8.3	Method accuracy: If extracts from Method 1662 are to be analyzed, the extract
from the accuracy test in that method shall be used; otherwise, an accuracy test is
unnecessary. The procedure for determining method accuracy is given in Section
8.3 of Method 1662, and the specification for accuracy is given in Table 4 of this
method.

8.3.1	Compare the percent recovery for diesel oil with the corresponding QC
acceptance criteria in Table 4. If the results of the spike fail the acceptance
criteria, and the recovery of the QC standard in the ongoing precision and
recovery test (Section 12.6.3) is within the acceptance criteria in Table 4,
an interference may be present. In this case, the result may not be reported
for regulatory compliance purposes. If, however, the results of both the
spike and the ongoing precision and recovery test fail the acceptance
criteria, the analytical system is judged to be out of control, and the
problem must be immediately identified and corrected and the sample batch
reanalyzed.

8.3.2	As part of the QA program for the laboratory, method accuracy for
samples shall be assessed and records shall be maintained. After the
analysis of five spiked samples in which the recovery passes the test in
Section 8.3.1, compute the average percent recovery (P) and the standard
deviation of the percent recovery (sp). Express the accuracy assessment as
a percent recovery interval from P - 2sp to P + 2sp. For example, if P =
90% and sp = 10% for five analyses of diesel oil, the accuracy interval is
expressed as 70 to 110%. Update the accuracy assessment on a regular
basis (e.g., after each five to ten new accuracy measurements).

8.4	Duplicates: If extracts from Method 1662 are to be analyzed, the extracts from the
duplicates test in that method shall be used. The procedure for preparing duplicates
is given in Section 8.4 of Method 1662, and the specification for RPD is given in
Table 4 of this method. If extracts from Method 1662 are not to be analyzed,
duplicates of the precision and recovery standard (Section 6.2.3) are analyzed, and
the specification for RPD is given for diesel oil in Table 4 of this method.

8.4.1	Analyze each of the duplicates per the procedure in Section 11 and
compute the results per Section 14.

8.4.2	Calculate the relative percent difference (RPD) between the two results per
the following equation:


-------
Equation 2

RPD =

A -A

1 x 100

(A + D2)/2

where:

I), = Concentration of diesel oil in the sample

	P7 = Concentration of diesel oil in the second (duplicate) sample

8.4.3 The relative percent difference for duplicates shall meet the acceptance

criteria in Table 4. If the criteria are not met, the analytical system shall be
judged to be out of control, and the problem must be immediately identified
and corrected and the sample batch reanalyzed.

8.5	Blanks: If extracts from Method 1662 are to be analyzed, the extracts from blanks
in that method shall be analyzed in addition to the blanks in this method.

8.5.1	Rinse the glassware used in preparation of the extracts in this method with
hexane and analyze a l-|oL aliquot of the rinsate using the procedure in
Section 11. Compute the results per Section 14.

8.5.2	If any peak is detected in a blank at greater than the minimum level in Table
1, analysis of samples is halted until the source of contamination is
eliminated and a blank shows no evidence of contamination.

8.6	The specifications contained in this method can be met if the apparatus used is
calibrated properly, then maintained in a calibrated state. The standards used for
initial precision and recovery (IPR, Section 8.2) and ongoing precision acid
recovery (OPR, Section 12.5) precision and recovery should be identical, so that
the most precise results will be obtained. The GC instrument will provide the most
reproducible results if dedicated to the settings and conditions required for the
analyses given in this method.

8.7	Depending on specific program requirements, field replicates and field spikes of
diesel oil into samples may be required when Method 1662 and this method are
used to assess the precision and accuracy of the sampling and sample transporting
techniques:

9. SAMPLE COLLECTION, PRESER VA TION, AND HANDLING

9.1	Oil samples are collected in 20- to 40-mL vials with PTFE- or aluminum-foil-lined
caps and stored in the dark at -20 to -10°C.

9.2	If extracts from Method 1662 are to be analyzed, the laboratory should be aware
that sample and extract holding times for this method have not yet been
established. However, based on tests of wastewater for the analytes determined in
this method, samples shall be extracted within seven days of collection and extracts
shall be analyzed within 40 days of extraction.

9.3	As a precaution against analyte and solvent loss or degradation, sample extracts
are stored in glass bottles with PTFE-lined caps, in the dark, at -20 to -10°C.


-------
10.	DILUTION OF OIL AND EXTRACTS

10.1	Neat oil samples: If oil is received in neat form, it should be diluted to bring the
concentration within the range of the instrument. If the oil is No. 2 diesel oil, the
appropriate concentration will be approximately 1.25 mg/mL.

10.2	Extracts from Method 1662: If extracts of samples from Method 1662 are to be
analyzed, these extracts (from Section 10.4.2 of that method) are analyzed
undiluted unless diesel oil is known or suspected to be present.

10.3	Neat oil expected to be diesel oil.

10.3.1	Weigh 100 mg into a 1.0-mL volumetric flask and dilute to the mark with
methylene chloride to produce a concentration of 10 mg/mL. Stopper and
mix thoroughly.

10.3.2	Using a calibrated 1.0-mL volumetric pipette, withdraw 1.0-mL of the
solution created in Section 10.3.1.1 and place in a 10-mL volumetric flask.
Then withdraw an additional 0.25 mL of the solution and add it to the 10-
mL volumetric flask (for a total of 1.25 mL). Fill to the mark with
methylene chloride to produce a concentration of 1.25 mg/mL (1250
|ig/mL), This solution will be near, but not above, the limit of the
calibration range and will match the concentration of the QC samples from
Method 1662 (assuming 100% recovery).

11.	GAS CHROMATOGRAPHY

11.1	Table 2 summarizes the recommended operating conditions for the GC. Retention
times for the n-alkanes obtained under these conditions are given in Table 3. An
example of the separation achieved for diesel oil is shown in Figure 1. Other
columns, chromatographic conditions, or detectors may be used if the minimum
level (Section 7.1) and the initial precision and accuracy requirements (Section 8.2)
are met.

11.2	Using a micropipette or microsyringe, transfer equal 100-|iL volumes of the
sample, sample extract, or QC standard extract (Section 10.2) and the TCB
internal standard solution (Section 6.2.1.3) into a GC injection vial. Cap tightly
and mix thoroughly.

11.3	Inject 1 //L of the sample extract or standard into the GC, using the conditions in
Table 2.

11.4	Begin data collection and the temperature program at the time of injection.

11.5	If the area of any peak exceeds the calibration range of the system, dilute a fresh
aliquot of the extract by a factor of 10, mix 100 |iL of internal standard with a

100-|iL aliquot of the extract, and reanalyze.

11.6	Compute the concentrations of the individual n-alkane peaks using the response
factor for hexadecane from the calibration data (Section 7.2.2.2).


-------
12. SYSTEM AND LABORATORY PERFORMANCE

12.1	At the beginning of each 8-hour shift during which analyses are performed, GC
calibration and system performance are verified. For these tests, analysis of the
medium-level calibration standard (Table 1) and of the precision and recovery
standard (Section 6.2.3) shall be used to verify all performance criteria.

Adjustment and/or re-calibration (per Section 7) shall be performed until all
performance criteria are met. Only after all performance criteria are met may
samples and blanks be analyzed.

12.2	Inject 1 |iL of the medium-level calibration standard (Table 1) into the GC
instrument according to the procedure in Section 11.

12.3	Retention times.

12.3.1	Retention time of the internal standard: The absolute retention time of the
TCB internal standard shall be within the range of 7.96 to 8.08 minutes.

12.3.2	Relative retention times of the n-alkanes: The retention times of the n-
alkanes relative to the TCB internal standard shall be within the limits given
in Table 3.

12.4	Calibration verification: Compute the concentration of hexadecane based on the
average calibration factor (Section 7.2.2.2). The concentration shall be within the
limits in Table 4. If calibration is verified, system performance is acceptable and
analysis of blanks and QC samples may begin. If, however, the concentration falls
outside of the calibration verification range, system performance is unacceptable.
In this case, correct the problem and repeat the test, or recalibrate (Section 7).

12.5	Ongoing precision and recovery (OPR): If the extract is from Method 1662, the
OPR standard from that method shall be used and the specification for the OPR
from Method 1662 in Table 4 shall be met; if not, a sample of diesel oil shall be
diluted per the procedure in Section 10 and shall be used for the OPR test.

12.5.1	Analyze the appropriate OPR standard.

12.5.2	Compute the concentration of diesel oil in this standard per Section 14.2.

12.5.3	Compare the concentration with the limits for ongoing precision and
recovery in Table 4. If the concentration is in the range specified, the
analytical processes are in control and analysis of blanks and samples may
proceed. If, however, the concentration is not in the specified range, these
processes are not in control. In this event, correct the problem, re-extract
the sample batch if the OPR is from Method 1662, or redilute the oil
sample (per Section 10.3) and repeat the ongoing precision and recovery
test.

12.5.4	Add results that pass the specifications in Section 12.5.3 to initial and
previous ongoing data. Update QC charts to form a graphic representation
of continued laboratory performance. Develop statements of laboratory
data quality for diesel oil by calculating the average percent recovery (R)
and the standard deviation of percent recovery (sr). Express the accuracy
statement as a recovery interval from R - 2sr to R + 2sr. For example, if R
= 95% and sr = 5%, the accuracy is 85 to 105%.


-------
13.	QUALITATIVE IDENTIFICATION

13.1 Qualitative identification is accomplished by comparison of data from analysis of a
sample or blank with data from analysis of the calibration verification standard
(Section 12.4). Diesel and crude oil are differentiated by the presence and
concentration of the C9-C30 n-alkane peaks in the chromatogram of extracts of the
sample.

13.1	Using the calibration data, establish the identity of the C9-C30 n-alkane peaks in the
chromatogram of the sample.

13.2	Diesel oil is not present in the sample if there are less than 10 n-alkane peaks
present in the C9-C24 range at a signal-to-noise ratio equal to or greater than 3 for
each peak, and if the QC tests (Sections 8 and 12) for the sample set are
acceptable. The experience of the analyst shall weigh heavily in the determination
of the presence of peaks at a signal-to-noise ratio of 3 or greater.

13.3	If ten or more n-alkane peaks are present in the analysis of the sample, diesel oil,
mineral oil, or crude oil may be present. Mineral oil can be distinguished by its
lower polynuclear aromatic hydrocarbon content using Method 1654A. Some
crude oils may be distinguished by the presence and concentration of n-alkanes in
the C25-C30 range. If peaks are present in the C25-C30 range, the quantitative
measurements in Section 14 are used as a final determination that the oil is crude
oil.

14.	QUANTITA TIVE DETERMINA TION

14.1 Differentiation between diesel and crude oil.

14.1.1	Using the concentrations of the individual n-alkane peaks determined in
Section 11.6, sum the concentrations of the n-alkanes from C9-C30
inclusive. Similarly, sum the concentrations of the n-alkanes from C25-C30
inclusive.

14.1.2	Calculate the percentage of C25-C30 n-alkanes as follows:

Equation 3

Sum of C25 - C30 n - alkanes

Percent (C„-CJ = S(jm of ^	x luu

14.1.3If the percent of C25-C30 n-alkanes is greater than 1.2, the oil is crude oil.
14.2 Determination of diesel oil: Compute the concentration of diesel oil in the standard
or QC extract using the hexadecane peak only, and the response factor given in
Section 7.2.2.2, using the following equation:


-------
Equation 4

„ . , n (QX4)

CJms/mL> " wim

where:

Cex = Concentration of oil in the sample

Cis = Concentration of the internal standard, in mglmL.

As = Area of the peak to be measured

Ais = Area of the internal standard peak

(For RF, see Equation I)	

15. COMPLEX SAMPLES

15.1	The most common interferences in the determination of diesel oil are from mineral
oil and proprietary additives in drilling fluids, and from naturally occurring
hydrocarbons from crude oil-bearing formations.

15.2	Mineral oil can be identified by its lower polynuclear aromatic hydrocarbon
content using Method 1654A.

15.3	Crude oils can usually be distinguished by the percentage of C25-C30 n-alkanes per
Section 14 of this method. However, some crude oils may not produce peaks in
the C25-C30 range.

15.3.1	Oil condensates from gas wells are low in molecular weight and will
normally produce chromatographic: peaks in the C8-C16 range. If a sample
of the gas condensate crude oil from the formation is available, the oil can
be distinguished from diesel oil using the extract from this method and the
n-alkane ratio determinations in the section on qualitative determination in
Method 1651.

15.3.2	Asphaltene crude oils with API gravities <20 may not produce
chromatographic peaks in the C25-C30 range. In this instance, the lack of
peaks in the C25-C30 range cannot be used to prove that the oil is crude oil
and not diesel oil. However, the absence of ten peaks in the C9-C24 range
can be used to demonstrate that diesel oil is not present, per Section 13 of
this method.

16. METHOD PERFORMANCE

Specifications in this method are adopted from EPA Method 1651 (Reference 5).
Example chromatograms of diesel oil and crude oil are shown in Figure 1.


-------
References

1.	"Carcinogens-Working With Carcinogens." Department of Health, Education, and
Welfare, Public Health Service, Centers for Disease Control [available through National
Technical Information System, 5285 Port Royal Road, Springfield, VA 22161, document
no. PB277256]: August 1977.

2.	"OSHA Safety and Health Standards, General industry [29 CFR 1910], Revised."
Occupational Safety and Health Administration, OSHA 2206. Washington, DC: January
1976.

3.	"Safety in Academic Chemistry Laboratories (3rd Edition)." American Chemical
Society Publication, Committee on Chemical Safety. Washington, DC: 1979.

4.	"Handbook of Analytical Quality Control in Water and Wastewater Laboratories."
USEPA, EMSL-Ci, EPA-600/4-79-019. Cincinnati, OH: March 1979.

5.	"Method 1651, Total Oil and Diesel Oil in Drilling Muds and Drill Cuttings by
Retort, Gravimetry, and GC/FID." Available from the EPA Sample Control Center, 300
N. Lee St., Alexandria, VA 22314.


-------
Method 1663

Table 1. Concentration of Calibration Standards

Calibration Solution Concentration tyg/mLt

Analyte

Low	Medium	High

n-decane
n-hexadecane
n-tetracosane
Diesel Oil

10

40
40
100
1250

200

Table 2. Gas Chromatographic Operating Conditions

Injection port, transfer line, and detector temperatures: 275°C
Column temperature program:

Initial temperature: 90 °C
Initial time: 0 minutes
Ramp: 90°C-250°C @ 5°C per minute
Final temperature: 250°C

Final hold: 10 minutes or until all peaks have eluted
Carrier gas and flow rates:

Carrier: Nitrogen or helium
Velocity: 20-40 cm/sec @ 90°C
Split ratio: 0-120:1**

Makeup gas: As required by manufacturer
Hydrogen and air flow rates: As specified by manufacturer

Detector amplifier settings: 10 " amp full scale. Attenuation is adjusted so that the highest

peaks are on scale in the most concentrated standard.

Recorder: Chart speed of 1 -2 cm/min (fixed)

* Conditions are approximate and can be adjusted to meet the performance criteria in Section 12

(see the note in Section 7.1).

**Lower split ratios may give more reproducible results.

70


-------
Method 1663

Table 3. Retention Times and Relative Retention Time Limits for n-Alkane Peaks

Retention Time



Mean



Compound

1minutes}

Relative

TCB

8.0

1.00-1.00

n"Ci2

9.9

1,22-1.24

n-C13

12.6

1,55-1,57

n-C,4

15.3

1.89-1.92

n-C,6

17.9

2,21-2.25

n-C„

20.4

2.52-2.56

n-C„

22.9

2.82-2.88

n-C„

25.2

3.12-3.15

n-C18

27.3

3.39-3.43

n-C20

29.4

3.66-3.71

n-C21

31.5

3.90-3.97

n-C22

33.4

4.14-4.21

n-C23

35.3

4.37-4.45

n"Cj4

37.1

4.58-4,69

Table 4. QC Acceptance Criteria

Criterion

Units

Minimum Level2

MJ/mL

Method Detection Limit3

mg/kg

Initial Precision and Recovery



Precision (RSD)4

mg/mL

Recovery

mg/mL

Calibration Verification6

//g/mL

Ongoing Precision and Recovery4

mg/mL

Matrix Spike Recovery"

pet

Duplicates

RPD

Oil from Method

1662	n-hexadecane

10

0.23
1.00-1.35

34-46

0.98-1.37
0.73-1,14

Oiesei
Oil'

100
100

1	CAS Registry number 68534-30-5; #2 diesei oil used for these tests.

2	This is a minimum level at which the analytical system shall give recognizable signals and
acceptable calibration points.

3	Estimated; 40 CFR Part 136, Appendix B; MDl is diesei oil in mud.

4	Test concentration in diluted extract - 1.25 mg/mL

5	Test concentration = 40 //g/ml

71


-------
Method 1663

TJ	ID

Crude Oil	o"

Diesel Oil

A52-010-003

Figure 1. Chromatograms of Crude Oil and Diesel Oil, Showing Differences in the C25-C10
Range

72


-------