444444444444444444444444444444444444444444444444444444444444444 EMISSION MEASUREMENT TECHNICAL INFORMATION CENTER NSPS TEST METHOD 444444444444444444444444444444444444444444444444444444444444444 (EMTIC M-10B, 8/16/94) Method lOB-Determination of Carbon Monoxide Emissions from Stationary Sources 1. Applicability and Principle 1.1 Applicability. This method applies to the measurement of carbon monoxide (CO) emissions at petroleum refineries and from other sources when specified in an applicable subpart of the regulations. 1.2 Principle. An integrated gas sample is extracted from the sampling point and analyzed for CO. The sample is passed through a conditioning system to remove interferences and collected in a Tedlar bag. The CO is separated from the sample by gas chromatography (GC) and catalytically reduced to methane (CH4) prior to analysis by flame ionization detection FID. The analytical portion of this method is identical to applicable sections in Method 25 detailing CO measurement. The oxidation catalyst required in Method 25 is not needed for sample analysis. Complete Method 25 analytical systems are acceptable alternatives when calibrated for CO and operated by the Method 25 analytical procedures. Note: Mention of trade names or commercial products in this method does not constitute the endorsement or recommendation for use by the Environmental Protection Agency. 1.3 Interferences. Carbon dioxide (C02) and organics potentially can interfere with the analysis. Carbon dioxide is primarily removed from the sample by the alkaline permanganate conditioning system; any residual C02 and organics are separated from the CO by GC. 2. Apparatus 2.1 Sampling. Same as in Method 10A, section 2.1. 2.2 Analysis. 2.2.1 Gas Chromatographic (GC) Analyzer. A semicontinuous GC/FID analyzer capable of quantifying CO in the sample and containing at least the following major components. 2.2.1.1 Separation Column. A column that separates CO from C02 and organic compounds that may be present. A \l/8\-in. OD stainless-steel column packed with 5.5 ft of 60/80 mesh Carbosieve S-II (available from Supelco) has been used successfully for this ------- ))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))) Prepared by Emission Measurement Branch EMTIC M-10B Technical Support Division, OAQPS, EPA ))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))) purpose. The column listed in Addendum 1 of Method 25 is also acceptable. 2.2.1.2 Reduction Catalyst. Same as in Method 25, section 2.3.2. 2.2.1.3 Sample Injection System. Same as in Method 25, section 2.3.4, equipped to accept a sample line from the Tedlar bag. 2.2.1.4 Flame Ionization Detector. Linearity meeting the specifications in section 2.3.5.1 of Method 25 where the linearity check is carried out using standard gases containing 20-, 200-, and 1,0 0 0-ppm CO. The minimal instrument range shall span 10 to 1,000 ppm CO. 2.2.1.5 Data Recording System. Same as in Method 25, section 2.3.6. 3. Reagents 3.1 Sampling. Same as in Method 10A, section 3.1. 3.2 Analysis. 3.2.1 Carrier, Fuel, and Combustion Gases. Same as in Method 25, sections 3.2.1, 3.2.2, and 3.2.3. 3.2.2 Linearity and Calibration Gases. Three standard gases with nominal CO concentrations of 20-, 200-, and 1,000-ppm CO in nitrogen. 3.2.3 Reduction Catalyst Efficiency Check Calibration Gas. Standard CH4 gas with a concentration of 1,000 ppm in air. 4. Procedure 4.1 Sample Bag Leak-checks, Sampling, and C02 Measurement. Same as in Method 10A, sections 4.1, 4.2, and 4.3. 4.2 Preparation for Analysis. Before putting the GC analyzer into routine operation, conduct the calibration procedures listed in section 5. Establish an appropriate carrier flow rate and detector temperature for the specific instrument used. 4.3 Sample Analysis. Purge the sample loop with sample, and then inject the sample. Analyze each sample in triplicate, and ------- calculate the average sample area (A) . Determine the bag CO concentration according to section 6.2. 5. Calibration 5.1 Carrier Gas Blank Check. Analyze each new tank of carrier gas with the GC analyzer according to section 4.3 to check for contamination. The corresponding concentration must be less than 5 ppm for the tank to be acceptable for use. 5.2 Reduction Catalyst Efficiency Check. Prior to initial use, the reduction catalyst shall be tested for reduction efficiency. With the heated reduction catalyst bypassed, make triplicate injections of the 1,0 0 0-ppm CH4 gas (section 3.2.3) to calibrate the analyzer. Repeat the procedure using 1,000-ppm CO (section 3.2.2) with the catalyst in operation. The reduction catalyst operation is acceptable if the CO response is within 5 percent of the certified gas value. 5.3 Analyzer Linearity Check and Calibration. Perform this test before the system is first placed into operation. With the reduction catalyst in operation, conduct a linearity check of the analyzer using the standards specified in section 3.2.2. Make triplicate injections of each calibration gas, and then calculate the average response factor (area/ppm) for each gas, as well as the overall mean of the response factor values. The instrument linearity is acceptable if the average response factor of each calibration gas is within 2.5 percent of the overall mean value and if the relative standard deviation (calculated in section 6.9 of Method 25) for each set of triplicate injections is less than 2 percent. Record the overall mean of the response factor values as the calibration response factor (R). 6. Calculations Carry out calculations retaining at least one extra decimal figure beyond that of the acquired data. Round off results only after the final calculation. 6.1 Nomenclature. A=Average sample area. Bw=Moisture content in the bag sample, fraction. C=CO concentration in the stack gas, dry basis, ppm. Cb=CO concentration in the bag sample, dry basis, ppm. F=Volume fraction of C02 in the stack, fraction. Pbar=Barometric pressure, mm Hg. ------- Pw=Vapor pressure H20 in the bag (from Table 10A-2, Method 1 OA), mm Hg. R=Mean calibration response factor, area/ppm. 6.2 CO Concentration in the Bag. Calculate Cb using Equations 10B-1 and 10B-2. If condensate is visible in the Tedlar bag, calculate Bw using Table 10A-2 of Method 10A and the temperature and barometric pressure in the analysis room. If condensate is not visible, calculate Bw using the temperature and barometric pressure at the sampling site. See Table Table 10A-2. Moisture Correction 444444444444444444444444444444444444444444444444444444444444444 Temp.°C Vapor Press.H20 mm Hg Temp.°C Vapor Press.H20 mm HH ))))))))0)))))))))))))))))))))))0)))))))))0)))))))))))))))))))) 4 * 6.1 * 18 * 15.5 6 * 7.0 * 20 * 17.5 8 * 8.0 * 22 * 19.8 10 * 9.2 * 24 * 22.4 12 * 10.5 * 26 * 25.2 14 * 12.0 * 28 * 28.3 16 * 13.6 * 30 * 31.8 ))))))))2)))))))))))))))))))))))2)))))))))2)))))))))))))))))))) Eq. 10B-1 B„ P„ = )))) P, • bar A Eq. 10B-2 Cb = )))) R(1-BJ 6.3 CO Concentration in the Stack, Eq. 10B-3 C = Cb (1-F) 7. Bibliography 1. Butler, F.E, J.E. Knoll, and M.R. Midgett. Development and Evaluation of Methods for Determining Carbon Monoxide Emissions. Quality Assurance Division, Environmental Monitoring Systems Laboratory, U.S. Environmental Protection Agency, Research Triangle Park, NC 27711. June 1985. 33p. 2. Salo, A.E., S. Witz, and R.D. MacPhee. Determination of ------- Solvent Vapor Concentrations by Total Combustion Analysis: A Comparison of Infrared with Flame Ionization Detectors. Paper No. 75-33.2. (Presented at the 68th Annual Meeting of the Air Pollution Control Association. Boston, MA. June 15, 1975.) 14 p. 3. Salo, A.E., W.L. Oaks, and R.D. MacPhee. Measuring the Organic Carbon Content of Source Emissions for Air Pollution Control. Paper No. 74-190. (Presented at the 67th Annual Meeting of the Air Pollution Control Association. Denver, CO. June 9, 1974.) 25 p. ------- |