METHOD 515.2

DETERMINATION OF CHLORINATED ACIDS IN WATER USING LIQUID-SOLID
EXTRACTION AND GAS CHROMATOGRAPHY WITH AN ELECTRON

CAPTURE DETECTOR

Revision 1.0
August 1992

R.C. Dressman and J.J. Lichtenberg — EPA 600/4-81-053, Revision 1.0 (1981)

J.W. Hodgeson — Method 515, Revision 2.0 (1986)

T. Engels (Battelle Columbus Laboratories) - National Pesticide Survey
Method 3, Revision 3.0 (1987)

R.L. Graves — Method 515.1, Revision 4.0 (1989)

J.W. Hodgeson — Method 515.2, Revision 1.0 (1992)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY
OFFICE OF RESEARCH AND DEVELOPMENT
U.S. ENVIRONMENTAL PROTECTION AGENCY
CINCINNATI, OHIO 45268

515.2-1


-------
METHOD 515.2

DETERMINATION OF CHLORINATED ACIDS IN WATER USING LIQUID-SOLID
EXTRACTION AND GAS CHROMATOGRAPHY WITH AN ELECTRON

CAPTURE DETECTOR

1.0 SCOPE AND APPLICATION

1.1 This is a gas chromatographic (GC) method applicable to the determination of
certain chlorinated acids in ground water and finished drinking water. The
following compounds can be determined by this method:

Analyte

Chemical Abstract Services
Registry Number

Acifluorfen

50594-66-6

Bentazon

25057-89-0

2,4-D

94-75-7

2,4-DB

94-82-6

Dacthal3

1861-32-1

Dicamba

1918-00-9

3,5-Dichlorobenzoic acid

51-36-5

Dichlorprop

120-36-5

Dinoseb

88-85-7

5-Hydroxydicamba

7600-50-2

Pentachlorophenol (PCP)

87-86-5

Picloram

1918-02-1

2,4,5-T

93-76-5

2,4,5-TP (Silvex)

93-72-1

'Dacthal monoacid and diacid metabolites included in method scope;
Dacthal diacid metabolite used for validation studies.

1.2	This method is applicable to the determination of salts and esters of analyte
acids. The form of each acid is not distinguished by this method. Results are
calculated and reported for each listed analyte as the total free acid.

1.3	Single laboratory accuracy and precision data and method detection limits
(MDLs) have been determined for the analytes above (Section 13.0). Observed
detection limits may vary among water matrices, depending upon the nature
of interferences in the sample matrix and the specific instrumentation used.

1.4	This method is restricted to use by or under the supervision of analysts
experienced in the use of GC and in the interpretation of gas chromatograms.
Each analyst must demonstrate the ability to generate acceptable results with
this method using the procedure described in Section 9.3.

515.2-2


-------
1.5	Analytes that are not separated chromatographically, (i.e., have very similar
retention times) cannot be individually identified and measured in the same
calibration mixture or water sample unless an alternative technique for
identification and quantitation exists (Section 11.6).

1.6	When this method is used to analyze unfamiliar samples for any or all of the
analytes above, analyte identifications should be confirmed by analysis on a
second gas chromatographic column or by gas chromatography/mass
spectrometry (GC/MS).

2.0 SUMMARY OF METHOD

2.1 A 250 mL measured volume of sample is adjusted to pH 12 with 6 N sodium
hydroxide for one hour to hydrolyze derivatives. Extraneous organic material
is removed by a solvent wash. The sample is acidified, and the chlorinated
acids are extracted with a 47 mm resin based extraction disk. The acids are
converted to their methyl esters using diazomethane. Excess derivatizing
reagent is removed, and the esters are determined by capillary column GC
using an electron capture detector (ECD).

3.0 DEFINITIONS

3.1	Internal Standard (IS) — A pure analyte(s) added to a sample, extract, or
standard solution in known amount(s), and used to measure the relative
responses of other method analytes and surrogates that are components of the
same sample or solution. The IS must be an analyte that is not a sample
component.

3.2	Surrogate Analyte (SA) — A pure analyte(s), which is extremely unlikely to be
found in any sample, and which is added to a sample aliquot in known
amount(s) before extraction or other processing, and is measured with the
same procedures used to measure other sample components. The purpose of
the SA is to monitor method performance with each sample.

3.3	Laboratory Duplicates (LD1 AND LD2) — Two aliquots of the same sample
taken in the analytical laboratory and analyzed separately with identical
procedures. Analyses of LD1 and LD2 indicate the precision associated with
laboratory procedures, but not with sample collection, preservation, or storage
procedures.

3.4	Field Duplicates (FD1 AND FD2) — Two separate samples collected at the
same time and place under identical circumstances and treated exactly the
same throughout field and laboratory procedures. Analyses of FD1 and FD2
give a measure of the precision associated with sample collection, preservation
and storage, as well as with laboratory procedures.

3.5	Laboratory Reagent Blank (LRB) — An aliquot of reagent water or other blank
matrix that is treated exactly as a sample including exposure to all glassware,

515.2-3


-------
equipment, solvents, reagents, internal standards, and surrogates that are used
with other samples. The LRB is used to determine if method analytes or other
interferences are present in the laboratory environment, the reagents, or the
apparatus.

3.6	Field Reagent Blank (FRB) — An aliquot of reagent water or other blank matrix
that is placed in a sample container in the laboratory and treated as a sample
in all respects, including shipment to the sampling site, exposure to sampling
site conditions, storage, preservation and all analytical procedures. The
purpose of the FRB is to determine if method analytes or other interferences
are present in the field environment.

3.7	Instrument Performance Check Solution (IPC) — A solution of one or more
method analytes, surrogates, internal standards, or other test substances used
to evaluate the performance of the instrument system with respect to a defined
set of criteria.

3.8	Laboratory Fortified Blank (LFB) — An aliquot of reagent water or other blank
matrix to which known quantities of the method analytes are added in the
laboratory. The LFB is analyzed exactly like a sample, and its purpose is to
determine whether the methodology is in control, and whether the laboratory
is capable of making accurate and precise measurements.

3.9	Laboratory Fortified Sample Matrix (LFM) — An aliquot of an environmental
sample to which known quantities of the method analytes are added in the
laboratory. The LFM is analyzed exactly like a sample, and its purpose is to
determine whether the sample matrix contributes bias to the analytical results.
The background concentrations of the analytes in the sample matrix must be
determined in a separate aliquot, and the measured values in the LFM
corrected for background concentrations.

3.10	Stock Standard Solution (SSS) — A concentrated solution containing one or
more method analytes prepared in the laboratory using assayed reference
materials or purchased from a reputable commercial source.

3.11	Primary Dilution Standard Solution (PDS) — A solution of several analytes
prepared in the laboratory from stock standard solutions, and diluted as
needed to prepare calibration solutions and other needed analyte solutions.

3.12	Calibration Standard (CAL) — A solution prepared from the primary dilution
standard solution or stock standard solutions and the internal standards and
surrogate analytes. The CAL solutions are used to calibrate the instrument
response with respect to analyte concentration.

3.13	Quality Control Sample (QCS) — A solution of method analytes of known
concentrations which is used to fortify an aliquot of LRB or sample matrix.
The QCS is obtained from a source external to the laboratory and different

515.2-4


-------
from the source of calibration standards. It is used to check laboratory
performance with externally prepared test materials.

INTERFERENCES

4.1	Method interferences may be caused by contaminants in solvents, reagents,
glassware and other sample processing apparatus that lead to discrete artifacts
or elevated baselines in gas chromatograms. All reagents and apparatus must
be routinely demonstrated to be free from interferences under analytical
conditions by analyzing laboratory reagent blanks as described in Section 9.2.

4.1.1	Glassware must be scrupulously cleaned.1 Clean all glassware as soon
as possible after use by thoroughly rinsing with the last solvent used in
it. Follow by washing with hot water and detergent and thorough
rinsing with dilute acid, tap and reagent water. Drain dry, and heat in
an oven or muffle furnace at 400°C for one hour. Do not heat
volumetric ware. Thermally stable materials such as PCBs might not be
eliminated by this treatment. Thorough rinsing with acetone may be
substituted for the heating. After glassware is dry and cool, store it in

a clean environment to prevent any accumulation of dust or other
contaminants. Store inverted or capped with aluminum foil.

4.1.2	The use of high purity reagents and solvents helps to minimize
interference problems. Purification of solvents by distillation in
all-glass systems may be required.

WARNING: When a solvent is purified, stabilizers and preservatives
added by the manufacturer are removed, thus potentially making the
solvent hazardous and reducing the shelf life.

4.2	The acid forms of the analytes are strong organic acids which react readily
with alkaline substances and can be lost during sample preparation.

Glassware and glass wool must be acid-rinsed with 1 N hydrochloric acid and
the sodium sulfate must be acidified with sulfuric acid prior to use to avoid
analyte losses due to adsorption.

4.3 Organic acids and phenols, especially chlorinated compounds, cause the most
direct interference with the determination. Alkaline hydrolysis and subsequent
extraction of the basic sample removes many chlorinated hydrocarbons and
phthalate esters that might otherwise interfere with the electron capture
analysis.

4.4 Interferences by phthalate esters can pose a major problem in pesticide analysis
when using the ECD. Phthalates generally appear in the chromatogram as
large peaks. Common flexible plastics contain varying amounts of phthalates,
that are easily extracted or leached during laboratory operations. Cross-
contamination of clean glassware routinely occurs when plastics are handled
during extraction steps, especially when solvent-wetted surfaces are handled.

515.2-5


-------
Interferences from phthalates can best be minimized by avoiding the use of
plastics in the laboratory. Exhaustive purification of reagents and glassware
may be required to eliminate background phthalate contamination.2,3

4.5	Interfering contamination may occur when a sample containing low
concentrations of analytes is analyzed immediately following a sample
containing relatively high concentrations of analytes. Between-sample rinsing
of the sample syringe and associated equipment with methyl-tert-butyl-ether
(MTBE) can minimize sample cross- contamination. After analysis of a sample
containing high concentrations of analytes, one or more injections of MTBE
should be made to ensure that accurate values are obtained for the next
sample.

4.6	Matrix interferences may be caused by contaminants that are coextracted from
the sample. Also, note that all analytes listed in the Scope and Application
section are not resolved from each other on any one column, i.e., one analyte
of interest may interfere with another analyte of interest. The extent of matrix
interferences will vary considerably from source to source, depending upon the
water sampled. The procedures in Section 11.0 can be used to overcome many
of these interferences. Tentative identifications should be confirmed (Section
11.6).

4.7	It is important that samples and working standards be contained in the same
solvent. The solvent for working standards must be the same as the final
solvent used in sample preparation. If this is not the case, chromatographic
comparability of standards to sample extracts may be affected.

SAFETY

5.1	The toxicity or carcinogenicity of each reagent used in this method has not
been precisely defined; however, each chemical compound must be treated as
a potential health hazard. Accordingly, exposure to these chemicals must be
reduced to the lowest possible level. The laboratory is responsible for
maintaining a current awareness file of OSHA regulations regarding the safe
handling of the chemicals specified in this method. A reference file of material
safety data sheets should also be made available to all personnel involved in
the chemical analysis. Additional references to laboratory safety are available
and have been identified5 7 for the information of the analyst.

5.2	Diazomethane — A toxic carcinogen which can explode under certain
conditions. The following precautions must be followed:

5.2.1 Use the diazomethane generator behind a safety shield in a well

ventilated fume hood. Under no circumstances can the generator be
heated above 90°C, and all grinding surfaces such as ground glass
joints, sleeve bearings, and glass stirrers must be avoided.

Diazomethane solutions must not be stored. Only generate enough for
the immediate needs. The diazomethane generator apparatus used in

515.2-6


-------
the esterification procedure (Section 11.4) produces micromolar amounts
of diazomethane in solution to minimize safety hazards. If the
procedure is followed exactly, no possibility for explosion exists.

5.3 Methyl-Tert-Butyl Ether — Nanograde, redistilled in glass, if necessary. Must
be free of peroxides as indicated by EM Quant test strips (available from
Scientific Products Co., Cat. No. PI 126-8, and other suppliers).

WARNING: When a solvent is purified, stabilizers added by the manufacturer
are removed, thus potentially making the solvent hazardous.

EQUIPMENT AND SUPPLIES (All specifications are suggested. Catalog numbers

are included for illustration only.)

6.1	Kontes Filter Funnels — Fisher Cat. No. 953755-0000 or equivalent.

6.2	Vacuum Flasks — 1000 mL with glass side arm.

6.3	Vacuum Manifold — The manifold should be capable of holding six to eight
filter flasks in series with house vacuum. Commercial manifolds are available
from a number of suppliers, e.g., Baker, Fisher, and Varian.

6.4	Culture Tubes (25 x 200 mm) with Teflon-lined Screw Caps — Fisher Cat.
No. 14-933-1C, or equivalent.

6.5	Pasteur Pipets — Glass disposable (5 mL).

6.6	Large Volume Pipets — Disposable, Fisher Cat. No. 13-678-8 or equivalent.

6.7	Balance — Analytical, capable of weighing to .0001 g.

6.8	pH Meter — Wide range capable of accurate measurements in the pH = 1-12
range.

6.9	Diazomethane Generator — See Figure 1 for a diagram of an all glass system
custom made for these validation studies. A micromolar generator is also
available from Aldrich Chemical.

6.10	Analytical Concentrator — Six or 12 positions, Organomation N-EVAP Model
No. 111-6917 or equivalent.

6.11	Gas Chromatography — Analytical system complete with gas chromatograph
equipped with ECD, split/splitless capillary injector, temperature
programming, differential flow control and all required accessories. A data
system is recommended for measuring peak areas. An autoinjector is
recommended to improve precision of analysis.

515.2-7


-------
6.12	GC Columns and Recommended Operating Conditions

6.12.1	Primary — DB-5 or equivalent, 30 m x .32 mm ID, 0.25 |im film
thickness. Injector Temp. = 200°C, Detector Temp. = 280°C, Helium
linear velocity is 30 cm/sec at 200°C and 10 psi, 2 |iL splitless injection
with purge on three minutes. Program: Hold at 60°C one minute,
increase to 260°C at 5°C/min. and hold five minutes.

6.12.2	Confirmation — DB-1701 or equivalent, 30 m x .32 mm ID, 0.25 |im film
thickness. Injector Temp. = 200°C, Detector Temp. = 280°C, Helium
linear velocity is 30 cm/sec at 200°C and 10 psi, 2 |iL splitless injection
with purge on three minutes. Program: Hold at 60°C one minute,
increase to 260°C at 5°C/min. and hold five minutes.

6.13	Glass Wool — Acid washed with IN HC1 and heated at 450°C for four hours.

6.14	Short Range pH Paper (pH=0-3).

6.15	Volumetric Flasks — 50 mL, 100 mL, and 250 mL.

6.16	Microsyringes — 25 |iL, 50 |iL, 100 |iL, 250 |iL, and 500 |iL.

6.17	Amber Bottles -15 mL, with Teflon-lined screw caps.

6.18	Graduated Cylinder — 250 mL.

6.19	Separatory Funnel — 500 mL.

6.20	Graduated Centrifuge Tubes — 15 mL or 10 mL Kuderna Danish Concentrator
tubes.

REAGENTS AND STANDARDS

7.1	Extraction Disks, 47 mm — Resin based polystyrenedivinylbenzene.

7.2	Reagent Water — Reagent water is defined as a water in which an interference
is not observed at the MDL of each analyte of interest.

7.2.1	A Millipore Super-Q water system or its equivalent may be used to
generate deionized reagent water. Distilled water that has been passed
through granular charcoal may also be suitable.

7.2.2	Test reagent water each day it is used by analyzing according to
Section 11.0.

7.3	Methanol — Pesticide quality or equivalent.

515.2-8


-------
7.4	Methyl-Tert-Butyl Ether (MTBE) — Nanograde, redistilled in glass if necessary.
Ether must be demonstrated to be free of peroxides. One test kit (EM Quant
Test Strips), is available from EM Science, Gibbstown, NJ. Procedures for
removing peroxides from the ether are provided with the test strips. Ethers
must be periodically tested (at least monthly) for peroxide formation during
use. Any reliable test kit may be used.

7.5	Sodium Sulfate — ACS, granular, anhydrous. Heat in a shallow tray at 400°C
for a minimum of four hours to remove phthalates and other interfering
organic substances. Alternatively, extract with methylene chloride in a Soxhlet
apparatus for 48 hours.

7.5.1 Sodium sulfate drying tubes — Plug the bottom of a large volume
disposable pipet with a minimum amount of acidified glass wool
(Supelco Cat. No. 20383 or equivalent). Fill the pipet halfway (3 g)
with acidified sodium sulfate (See Section 7.9).

7.6	Sulfuric Acid — Reagent grade

7.6.1 Sulfuric acid, 12 N — Slowly add 335 mL concentrated sulfuric acid to
665 mL of reagent water.

7.7	Sodium Hydroxide — ACS reagent grade or equivalent.

7.7.1	Sodium hydroxide IN — Dissolve 4.0 g reagent grade sodium
hydroxide in reagent water and dilute to 100 mL in volumetric flasks.

7.7.2	Sodium hydroxide 6N

7.8	Ethyl Ether, Unpreserved — Nanograde, redistilled in glass if necessary. Must
be free of peroxides as indicated by EM Quant test strips (available from
Scientific Products Co., Cat. No. PI 126-8, and other suppliers). Procedures
recommended for removal of peroxides are provided with the test strips.

7.9	Acidified Sodium Sulfate — Cover 500 g sodium sulfate (Section 7.5) with ethyl
ether (Section 7.8). While agitating vigorously, add dropwise approximately
0.7 mL concentrated sulfuric acid. Remove the ethyl ether overnight under
vacuum and store the sodium sulfate in a 100°C oven.

7.10	Carbitol — ACS grade. Available from Aldrich Chemical.

7.11	Diazald — ACS grade. Available from Aldrich Chemical.

7.12	Diazald Solution — Prepare a solution containing 10 g Diazald in 100 mL of a
50:50 by volume mixture of ethyl ether and carbitol. This solution is stable for
one month or longer when stored at 4°C in an amber bottle with a Teflon-lined
screw cap.

515.2-9


-------
7.13	4,4'-Dibromooctafluorobiphenyl (DBOB) — 99% purity, for use as internal
standard.

7.14	2,4-Dichlorophenylacetic Acid (DCAA) — 99% purity, for use as surrogate
standard.

7.15	Potassium Hydroxide — ACS reagent grade or equivalent.

7.15.1 Potassium hydroxide solution, 37% — Using extreme caution, dissolve
37 g reagent grade potassium hydroxide in reagent water and dilute to
100 mL.

7.16	Stock Standard Solutions (1.00-2.00 |ig/|_iL) — Stock standard solutions may be
purchased as certified solutions or prepared from pure standard materials
using the following procedure:

7.16.1	Prepare stock standard solutions by accurately weighing approximately
0.0100-0.0200 g of pure material. Dissolve the material in methanol and
dilute to volume in a 10 mL volumetric flask. Larger volumes may be
used at the convenience of the analyst. If compound purity is certified
at 96% or greater, the weight may be used without correction to
calculate the concentration of the stock standard. Commercially
prepared stock standards may be used at any concentration if they are
certified by the manufacturer or by an independent source.

7.16.2	Transfer the stock standard solutions into 15 mL
TFE-fluorocarbon-sealed screw cap amber vials. Store at 4°C or less
when not in use.

7.16.3	Stock standard solutions should be replaced after two months or sooner
if comparison with laboratory fortified blanks, or QC samples indicate a
problem.

7.16.4	Primary dilution standards — Prepare two sets of standards according
to the sets labeled A and B in Table 1. For each set, add approximately
25 mL of methanol to a 50 mL volumetric flask. Add aliquots of each
stock standard in the range of approximately 20-400 |iL and dilute to
volume with methanol. Individual analyte concentrations will then be
in the range of 0.4-8 ]ig/mL (for a 1.0 mg/mL stock). The minimum
concentration would be appropriate for an analyte with strong electron
capture detector (ECD) response, e.g. pentachlorophenol. The
maximum concentration is for an analyte with weak response, e.g., 2,4-
DB. The concentrations given in Table 2 reflect the relative volumes of
stock standards used for the primary dilution standards used in
generating the method validation data. Use these relative values to
determine the aliquot volumes of individual stock standards above.

515.2-10


-------
7.17	Internal Standard Solution — Prepare a stock internal standard solution by
accurately weighing approximately 0.050 g of pure DBOB. Dissolve the DBOB
in methanol and dilute to volume in a 10 mL volumetric flask. Transfer the
DBOB solution to a TFE-fluorocarbonsealed screw cap bottle and store at room
temperature. Prepare a primary dilution standard at approximately 1.00
|ig/mL by the addition of 20 |iL of the stock standard to 100 mL of methanol.
Addition of 100 |iL of the primary dilution standard solution to the final 5 mL
of sample extract (Section 11.0) results in a final internal standard
concentration of 0.020 |ig/mL. Solution should be replaced when ongoing QC
(Section 9.0) indicates a problem. Note that DBOB has been shown to be an
effective internal standard for the method analytes, but other compounds may
be used if the QC requirements in Section 9.0 are met.

7.18	Surrogate Analyte Solution — Prepare a surrogate analyte stock standard
solution by accurately weighing approximately 0.050 g of pure DCAA.

Dissolve the DCAA in methanol and dilute to volume in a 10 mL volumetric
flask. Transfer the surrogate analyte solution to a TFE-fluorocarbon-sealed
screw cap bottle and store at room temperature. Prepare a primary dilution
standard at approximately 2.0 |ig/mL by addition of 40 |iL at the stock
standard to 100 mL of methanol. Addition of 250 |iL of the surrogate analyte
solution to a 250 mL sample prior to extraction results in a surrogate
concentration in the sample of 2 Hg/L and, assuming quantitative recovery of
DCAA, a surrogate analyte concentration in the final 5 mL extract of

0.1 |ig/mL. The surrogate standard solution should be replaced when ongoing
QC (Section 9.0) indicates a problem. DCAA has been shown to be an
effective surrogate standard for the method analytes, but other compounds
may be used if the QC requirements in Section 10.0 are met.

7.19	Instrument Performance Check Solution — Prepare a diluted dinoseb solution
by adding 10 |iL of the 1.0 \ig/\iL dinoseb stock solution to the MTBE and
diluting to volume in a 10 mL volumetric flask. To prepare the check solution,
add 40 |iL of the diluted dinoseb solution, 16 |iL of the 4-nitrophenol stock
solution, 6 |iL of the 3,5-dichlorobenzoic acid stock solution, 50 |iL of the
surrogate standard solution, 25 pL of the internal standard solution, and 250
|iL of methanol to a 5 mL volumetric flask and dilute to volume with MTBE.
Methylate sample as described in Section 11.4. Dilute the sample to 10 mL in
MTBE. Transfer to a TFE-fluorocarbon-sealed screw cap bottle and store at
room temperature. Solution should be replaced when ongoing QC (Section
9.0) indicates a problem.

SAMPLE COLLECTION. PRESERVATION. AND STORAGE

8.1 Grab samples should be collected in 1 L amber glass containers. Conventional
sampling practices7 should be followed; however, the bottle must not be
prerinsed with sample before collection.

515.2-11


-------
8.2	Sample Preservation and Storage

8.2.1	Add hydrochloric acid (diluted 1:1 in water) to the sample at the
sampling site in amounts to produce a sample pH <2. Short range (0-3)
pH paper (Section 6.14) may be used to monitor the pH.

8.2.2	If residual chlorine is present, add 80 mg of sodium thiosulfate per liter
of sample to the sample bottle prior to collecting the sample.

8.2.3	After the sample is collected in the bottle containing preservative(s),
seal the bottle and shake vigorously for one minute.

8.2.4	The samples must be iced or refrigerated at 4°C away from light from
the time of collection until extraction. Preservation study results
indicate that the sample analytes (measured as total acid), except 5-
hydroxy-dicamba, are stable in water for 14 days when stored under
these conditions (Tables 8 and 9). The concentration of 5-
hydroxydicamba is seriously degraded over 14 days in a biologically
active matrix. However, analyte stability will very likely be affected by
the matrix; therefore, the analyst should verify that the preservation
technique is applicable to the samples under study.

8.3	Extract Storage

8.3.1 Extracts should be stored at 4°C or less away from light. Preservation
study results indicate that most analytes are stable for 14 days (Tables 8
and 9); however, the analyst should verify appropriate extract holding
times applicable to the samples under study.

QUALITY CONTROL

9.1	Minimum QC requirements are initial demonstration of laboratory capability,
determination of surrogate compound recoveries in each sample and blank,
monitoring internal standard peak area or height in each sample and blank
(when internal standard calibration procedures are being employed), analysis
of laboratory reagent blanks, laboratory fortified samples, laboratory fortified
blanks, and QC samples.

9.2	Laboratory Reagent Blanks (LRB) — Before processing any samples, the analyst
must demonstrate that all glassware and reagent interferences are under
control. Each time a set of samples is extracted or reagents are changed, a LRB
must be analyzed. If within the retention time window of any analyte the LRB
produces a peak that would prevent the determination of that analyte,
determine the source of contamination and eliminate the interference before
processing samples.

515.2-12


-------
9.3

Initial Demonstration of Capability

9.3.1	Select a representative fortified concentration (about 10 to 20 times
MDL) for each analyte. Prepare a sample concentrate (in methanol)
containing each analyte at 1000 times selected concentration. With a
syringe, add 250 |iL of the concentrate to each of at least four 250 mL
aliquots of reagent water, and analyze each aliquot according to
procedures beginning in Section 11.0.

9.3.2	For each analyte the recovery value for all four of these samples must
fall in the range of ±40% of the fortified concentration. For those
compounds that meet the acceptance criteria, performance is considered
acceptable and sample analysis may begin. For compounds failing this
criteria, this procedure must be repeated using five fresh samples until
satisfactory performance has been demonstrated for all analytes.

9.3.3	The initial demonstration of capability is used primarily to preclude a
laboratory from analyzing unknown samples via a new, unfamiliar
method prior to obtaining some experience with it. As laboratory
personnel gain experience with this method the quality of data should
improve beyond those required here.

9.4	The analyst is permitted to modify GC columns, GC conditions, detectors,
concentration techniques (i.e., evaporation techniques), internal standard or
surrogate compounds. Each time such method modifications are made, the
analyst must repeat the procedures in Section 9.3.

9.5	Assessing Surrogate Recovery

9.5.1	When surrogate recovery from a sample or a blank is <60% or >140%,
check (1) calculations to locate possible errors, (2) fortifying solutions
for degradation, (3) contamination, and (4) instrument performance. If
those steps do not reveal the cause of the problem, reanalyze the
extract.

9.5.2	If a blank extract reanalysis fails the 60-140% recovery criteria, the
problem must be identified and corrected before continuing.

9.5.3	If sample extract reanalysis meets the surrogate recovery criteria, report
only data for the reanalyzed extract. If sample extract continues to fail
the recovery criteria, report all data for that sample as suspect.

9.6	Assessing the Internal Standard

9.6.1 When using the internal standard (IS) calibration procedure, the analyst
is expected to monitor the IS response (peak area or peak height) of all
samples during each analysis day. The IS response for any sample

515.2-13


-------
chromatogram should not deviate from the daily calibration check
standard's IS response by more than 30%.

9.6.2	If >30% deviation occurs with an individual extract, optimize
instrument performance and inject a second aliquot of that extract.

9.6.2.1	If the reinjected aliquot produces an acceptable internal standard
response, report results for that aliquot.

9.6.2.2	If a deviation of greater than 30% is obtained for the reinjected
extract, analysis of the samples should be repeated beginning
with Section 11.0, provided the sample is still available.
Otherwise, report results obtained from the reinjected extract,
but annotate as suspect.

9.6.3	If consecutive samples fail the IS response acceptance criteria,
immediately analyze a medium calibration standard.

9.6.3.1	If the standard provides a response factor (RF) (Section 10.2.2)
within 20% of the predicted value, then follow procedures
itemized in Section 9.6.2 for each sample failing the IS response
criterion.

9.6.3.2	If the check standard provides a response factor which deviates
more than 20% of the predicted value, then the analyst must
recalibrate as specified in Section 10.0.

Assessing Laboratory Performance — Laboratory Fortified Blank (LFB)

9.7.1	The laboratory must analyze at least one LFB sample with every 20
samples or one per sample set (all samples extracted within a 24-hour
period) whichever is greater. The concentration of each analyte in the
LFB should be 10 times the MDL. Calculate percent recovery (Xj). If
the recovery of any analyte falls outside the control limits (See Section
9.7.2), that analyte is judged out of control, and the source of the
problem should be identified and resolved before continuing analyses.

9.7.2	Until sufficient data become available, usually a minimum of results
from 20-30 analyses, each laboratory should assess laboratory perfor-
mance against the control limits in Section 9.3.2 that are derived from
the data in Table 2. When sufficient internal performance data become
available, develop control limits from the mean percent recovery (X)
and standard deviation (S) of the percent recovery. These data are used
to establish upper and lower control limits as follows:

UPPER CONTROL LIMIT = X + 3S
LOWER CONTROL LIMIT = X - 3S

515.2-14


-------
After each 5-10 new recovery measurements, new control limits should
be calculated using only the most recent 20-30 data points. These
calculated control limits should never exceed those established in
Section 9.3.2.

9.7.3	Method detection limits (MDL) must be determined using the proce-
dure given in reference8. The MDLs must be sufficient to detect
analytes at the required levels according to SDWA regulations.

9.7.4	At least quarterly, analyze a QCS (Section 3.13) from an outside source.

9.7.5	Laboratories are encouraged to participate in external performance
evaluation studies such as the laboratory certification programs offered
by many states or the studies conducted by USEPA.

Assessing Analyte Recovery — Laboratory Fortified Sample Matrix (LFM)

9.8.1	Each laboratory must analyze a LFM for 10% of the samples or one
sample concentration per set, whichever is greater. The concentration
should not be less then the background concentration of the sample
selected for fortification. Ideally, the concentration should be the same
as that used for the laboratory fortified blank (Section 9.7). Over time,
samples from all routine sample sources should be fortified.

9.8.2	Calculate the percent recovery, P of the concentration for each analyte,
after correcting the measured concentration, X, from the fortified
sample for the background concentration, b, measured in the unfortified
sample.

P = 100 (X - b) / fortified concentration,

and compare these values to control limits appropriate for reagent
water data collected in the same fashion. If the analyzed unfortified
sample is found to contain NO background concentrations and the
added concentrations are those specified in Section 9.7, then the appro-
priate control limits would be the acceptance limits in Section 9.7. If,
on the other hand, the analyzed unfortified sample is found to contain
background concentration, b, estimate the standard deviation at the
background concentration, sb, using regressions or comparable back-
ground data and, similarly, estimate the mean, Xa and standard
deviation, sa, of analytical results at the total concentration after
fortifying. Then the appropriate percentage control limits would be P
±3sP , where:

P = 100X / (b + fortifying concentration)
and sP = 100 (s2 + s^)172/fortifying concentration

515.2-15


-------
For example, if the background concentration for Analyte A was found
to be 1 Hg/L and the added amount was also 1 Hg/L, and upon
analysis the laboratory fortified sample measured 1.6 Hg/L, then the
calculated P for this sample would be (1.6 Hg/L minus 1.0 |ig/L)/l
]ig/L or 60%. This calculated P is compared to control limits derived
from prior reagent water data. Assume that analysis of an interference
free sample at 1 ]Jg/L yields_an s of 0.12 Hg/L and similar analysis at
2.0 Hg/L yields X and s of 2.01 Hg/L and 0.20 \ig/L, respectively. The
appropriate limits to judge the reasonableness of the percent recovery,
60%, obtained on the fortified matrix sample is computed as follows:

[100 (2.01 (xg/L) / 2.0 (xg/L]

±3 (100) [(0.12 ng/L)2 + (0.20 (ig/L)2],/2/1.0 (xg/L =

100.5% ± 300 (0.233) =

100.5% ± 70% or 30% to 170 recovery of the added analyte.

9.8.3 If the recovery of any such analyte falls outside the designated range,
and the laboratory performance for that analyte is shown to be in
control (Section 9.7), the recovery problem encountered with the
fortified sample is judged to be matrix related, not system related. The
result for that analyte in the unfortified sample is labeled
suspect/matrix to inform the data user that the results are suspect due
to matrix effects.

9.9	Assessing Instrument System/Instrument Performance Check (IPC) Sample —
Instrument performance should be monitored on a daily basis by analysis of
the IPC sample. The IPC sample contains compounds designed to indicate
appropriate instrument sensitivity, column performance (primary column) and
chromatographic performance. IPC sample components and performance
criteria are listed in Table 11. Inability to demonstrate acceptable instrument
performance indicates the need for reevaluation of the instrument system. The
sensitivity requirements are set based on the MDLs published in this method.
MDLs will vary from laboratory to laboratory.

9.10	The laboratory may adopt additional QC practices for use with this method.
The specific practices that are most productive depend upon the needs of the
laboratory and the nature of the samples. For example, field or laboratory
duplicates may be analyzed to assess the precision of the environmental
measurements or field reagent blanks may be used to assess contamination of
samples under site conditions, transportation, and storage.

515.2-16


-------
10.0 CALIBRATION AND STANDARDIZATION

10.1	Establish GC operating parameters equivalent to those indicated in
Section 6.12. This calibration procedure employs procedural standards, i.e.,
fortified aqueous standards which are processed through most of the method
(Section 11.0). The GC system is calibrated by means of the internal standard
technique (Section 10.2).

NOTE: Calibration standard solutions must be prepared such that no
unresolved analytes are mixed together (See Table 1).

10.2	Internal Standard Calibration Procedure — To use this approach, the analyst
must select one or more internal standards compatible in analytical behavior to
the compounds of interest. The analyst must further demonstrate that the
measurement of the internal standard is not affected by method or matrix
interferences. DBOB (Section 7.13) has been identified as a suitable internal
standard.

10.2.1	Prepare aqueous calibration standards at a minimum of three (five are
recommended) concentration levels for each method analyte as follows:
for each concentration, fill a 250 mL volumetric flask with 240 mL of
reagent water at pH 1 and containing 20% by weight of dissolved
sodium sulfate. Add an appropriate aliquot of the primary dilution
standard (Section 7.16.4) and dilute to 250 mL with the same reagent
water. Process each aqueous calibration sample through the analytical
procedure beginning with Section 11.2, i.e., omit the hydrolysis and
cleanup step (Section 11.1). The lowest calibration standard should
represent analyte concentrations near, but above, the respective MDLs.
The remaining standards should bracket the analyte concentrations
expected in the sample extracts, or should define the working range of
the detector. The internal standard is added to the final 5 mL extract as
specified in Section 11.0.

10.2.2	Analyze each calibration standard according to the procedure beginning
in Section 11.2. Tabulate response (peak height or area) against
concentration for each compound and internal standard. Calculate the
response factor (RF) for each analyte and surrogate using Equation 1.

Equation 1

(Ais) (Cs)

where: As = Response for the analyte to be measured.
Ais = Response for the internal standard.
Cis = Concentration of the internal standard (|ig/L).
Cs = Concentration of the analyte to be measured (|ig/L).

515.2-17


-------
10.2.3	If the RF value over the working range is constant (30% RSD or less)
the average RF can be used for calculations. Alternatively, the results
can be used to plot a calibration curve of response ratios (As/Ais) vs. Cs.
A data station may be used to collect the chromatographic data,
calculate response factors and generate linear or second order
regression curves.

10.2.4	The working calibration curve or RF must be verified on each working
shift by the measurement of one or more calibration standards. A new
calibration standard need not be derivatized each day. The same
standard extract can be used up to 14 days. If the response for any
analyte varies from the predicted response by more than ±30%, the test
must be repeated using a fresh calibration standard. If the repetition
also fails, a new calibration curve must be generated for that analyte
using freshly prepared standards.

10.2.5	Verify calibration standards periodically, at least quarterly is
recommended, by analyzing a standard prepared from reference
material obtained from an independent source. Results from these
analyses must be within the limits used to routinely check calibration.

11.0 PROCEDURE

11.1 Manual Hydrolysis and Clean-up

11.1.1	Remove the sample bottles from cold storage and allow them to
equilibrate to room temperature. Acidify and add sodium thiosulfate to
blanks and QC check standards as specified in Section 8.0.

11.1.2	Measure a 250 mL aliquot of each sample with a 250 mL graduated
cylinder and pour into a 500 mL separatory funnel. Add 250 |iL of the
surrogate primary dilution standard (Section 7.18) to each 250 mL
sample. The surrogate will be at a concentration of 2 Hg/L. Dissolve
50 g sodium sulfate in the sample.

11.1.3	Add 4 mL of 6 N NaOH to each sample, seal, and shake. Check the
pH of the sample with pH paper or a pH meter; if the sample does not
have a pH greater than or equal to 12, adjust the pH by adding more 6
N NaOH. Let the sample sit at room temperature for one hour,
shaking the separatory funnel and contents periodically.

11.1.4	Add 15 mL methylene chloride to the graduated cylinder to rinse the
walls, transfer the methylene chloride to the separatory funnel and
extract the sample by vigorously shaking the funnel for two minutes
with periodic venting to release excess pressure. Allow the organic
layer to separate from the water phase for a minimum of 10 minutes. If
the emulsion interface between layers is more than one-third the
volume of the solvent layer, the analyst must employ mechanical

515.2-18


-------
techniques to complete the phase separation. The optimum technique
depends upon the sample, but may include stirring, filtration through
glass wool, centrifugation, or other physical methods. Discard the
methylene chloride phase.

11.1.5	Add a second 15 mL volume of methylene chloride to the separatory
funnel and repeat the extraction procedure a second time, discarding
the methylene chloride layer. Perform a third extraction in the same
manner.

11.1.6	Drain the contents of the separatory funnel into a 500 mL beaker.

Adjust the pH to 1.0 ±0.1 by the dropwise addition of concentrated
sulfuric acid with constant stirring. Monitor the pH with a pH meter
(Section 6.8) or short range (0-3) pH paper (Section 6.14).

11.2 Sample Extraction

11.2.1	Vacuum manifold — Assemble a manifold (Section 6.3) consisting of six
to eight vacuum flasks with filter funnels (Sections 6.1 and 6.2).
Individual vacuum control, on-off and vacuum release valves and
vacuum gauges are desirable. Place the 47 mm extraction disks
(Section 7.1) on the filter frits.

11.2.2	Add 20 mL of 10% by volume of methanol in MTBE to the top of each
disk without vacuum and allow the solvent to remain for two minutes.
Turn on full vacuum and pull the solvent through the disks, followed
by room air for five minutes.

11.2.3	Adjust the vacuum to approximately 5 in. (mercury) and add the
following in series to the filter funnel (a) 20 mL methanol (b) 20 mL
reagent water (c) sample. Do not allow the disk to dry between steps
and maintain the vacuum at 5 in.

11.2.4	After the sample is extracted completely, apply maximum vacuum and
draw room air through the disks for 20 minutes.

11.2.5	Place the culture tubes (Section 6.4) in the vacuum tubes to collect the
eluates. Elute the disks with two each 2 mL aliquots of 10% methanol
in MTBE. Allow each aliquot to remain on the disk for one min before
applying vacuum.

11.2.6	Rinse each 500 mL beaker (Section 11.1.6) with 4 mL of pure MTBE and
elute the disk with this solvent as in Section 11.2.5.

11.2.7	Remove the culture tubes and cap.

515.2-19


-------
11.3 Extract Preparation

11.3.1	Pre-rinse the drying tubes (Section 7.5.1) with 2 mL of MTBE.

11.3.2	Remove the entire extract with a 5 mL pipet and drain the lower
aqueous layer back into the culture tube. Add the organic layer to the
sodium sulfate drying tube (Section 7.5.1). Maintain liquid in the
drying tube between this and subsequent steps. Collect the dried
extract in a 15 mL graduated centrifuge tube or a 10 mL Kuderna-
Danish tube.

11.3.3	Rinse the culture tube with an additional 1 mL of MTBE and repeat
Section 11.3.2.

11.3.4	Repeat step Section 11.3.3 and finally add a 1 mL aliquot of MTBE to
the drying tube before it empties. The final volume should be 6-9 mL.
In this form the extract is esterified as described below.

11.4 Extract Esterification

11.4.1	Assemble the diazomethane generator (Figure 1) in a hood.

11.4.2	Add 5 mL of ethyl ether to Tube 1. Add 4 mL of Diazald solution
(Section 7.12) and 3 mL of 37% KOH solution (Section 7.15.1) to the
reaction Tube 2. Immediately place the exit tube into the collection
tube containing the sample extract. Apply nitrogen flow (10 mL/min)
to bubble diazo-methane through the extract. Each charge of the
generator should be sufficient to esterify four samples. The appearance
of a persistent yellow color is an indication that esterification is
complete. The first sample should require 30 seconds to one minute
and each subsequent sample somewhat longer. The final sample may
require two to three minutes.

11.4.3	Cap each collection tube and allow to remain stored at room
temperature in a hood for 30 minutes. No significant fading of the
yellow color should occur during this period. Fortify each sample with
100 |iL of the internal standard primary dilution solution (Section 7.17)
and reduce the volume to 5.0 mL with the analytical concentrator
(Section 6.10), a stream of dry nitrogen, or an equivalent concentration
technique.

NOTE: The excess diazomethane is volatilized from the extract during
the concentration procedure.

11.4.4	Cap the tubes and store in a refrigerator if further processing will not
be performed immediately. Analyze by GC-ECD.

515.2-20


-------
11.5 Gas Chromatography

11.5.1	Section 6.12 summarizes the recommended GC operating conditions.
Included in Table 1 are retention times observed using this method.
Figures 2A and 2B illustrate the chromatographic performance of the
primary column (Section 6.12.1) for Groups A and B of the method
analytes. Other GC columns, chromatographic conditions, or detectors
may be used if the requirements of Section 9.3 are met.

11.5.2	Calibrate the system daily as described in Section 10.0.

11.5.3	Inject 2 |iL of the sample extract. Record the resulting peak size in area
units.

11.5.4	If the response for any sample peak exceeds the working range of the
detector, dilute the extract and reanalyze.

11.6 Identification of Analytes

11.6.1	Identify a sample component by comparison of its retention time to the
retention time of a reference chromatogram. If the retention time of an
unknown compound corresponds, within limits, to the retention time of
a standard compound, then an analyte is considered to be identified.

11.6.2	The width of the retention time window used to make identifications
should be based upon measurements of actual retention time variations
of standards over the course of a day. Three times the standard
deviation of a retention time can be used to calculate a suggested
window size for a compound. However, the experience of the analyst
should weigh heavily in interpretation of chromatograms.

11.6.3	Identification requires expert judgment when sample components are
not resolved chromatographically. When GC peaks obviously represent
more than one sample component (i.e., broadened peak with
shoulder (s) or valley between two or more maxima, or any time doubt
exists over the identification of a peak in a chromatogram, appropriate
alternative techniques to help confirm peak identification need to be
employed. For example, more positive identification may be made by
the use of an alternative detector which operates on a
chemical/physical principle different from that originally used, e.g.,
mass spectrometry, or the use of a second chromatography column. A
suggested alternative column is described in Section 6.12.2.

515.2-21


-------
12.0 DATA ANALYSIS AND CALCULATIONS

12.1	Calculate analyte concentrations in the sample from the response for the
analyte using the calibration procedure described in Section 10.0.

12.2	Calculate the concentration (C) in the sample using the response factor (RF)
determined in Section 10.2.2 and Equation 2, or determine sample
concentration from the calibration curve (Section 10.2.3).

Equation 2

(As) ds)

C (xg/L) =

(A. ) (RF) (Vo)

where: As = Response for the parameter to be measured.
Ais = Response for the internal standard.

Is = Amount of internal standard added to each extract (|ig).
VQ = Volume of water extracted (L).

13.0 METHOD PERFORMANCE

13.1	In a single laboratory, analyte recoveries from reagent water were determined
at three concentration levels, Tables 2-4. Results were used to determine the
analyte MDLs8 listed in Table 2.

13.2	In a single laboratory, analyte recoveries from dechlorinated tap water were
determined at two concentrations, Tables 5 and 6. In addition, analyte
recoveries were determined at two concentrations from an ozonated surface
(river) water, Tables 7 and 8, and at one level from a high humectant surface
(reservoir) water, Table 10. Finally, a holding study was conducted on the
preserved, ozonated surface water and recovery data are presented for Day 1
and Day 14 of this study, Tables 8 and 9. The ozonated surface water was
chosen as the matrix in which to study analyte stability during a 14-day
holding time because it was very biologically active.

14.0 POLLUTION PREVENTION

14.1 This method utilizes the new liquid-solid extraction technology which requires
the use of very small quantities of organic solvents. This feature eliminates the
hazards involved with the use of large volumes of potentially harmful organic
solvents needed for conventional liquid-liquid extractions. Also, mercuric
chloride, a highly toxic and environmentally hazardous chemical, has been
replaced with hydrochloric acid as the sample preservative. These features
make this method much safer and a great deal less harmful to the
environment. Some of the phenolic herbicides on the analyte list are very
difficult to methylate and diazomethane is still required to derivatize these
compounds.

515.2-22


-------
14.2 For information about pollution prevention that may be applicable to
laboratory operations, consult "Less is Better: Laboratory Chemical
Management for Waste Reduction" available from the American Chemical
Society's Department of Government Relations and Science Policy, 1155 16th
Street N.W., Washington, D.C. 20036.

15.0 WASTE MANAGEMENT

15.1 Due to the nature of this method, there is little need for waste management.
No large volumes of solvents or hazardous chemicals are used. The matrices
of concern are finished drinking water or source water. However, the Agency
requires that laboratory waste management practices be conducted consistent
with all applicable rules and regulations, and that laboratories protect the air,
water, and land by minimizing and controlling all releases from fume hoods
and bench operations. Also, compliance is required with any sewage
discharge permits and regulations, particularly the hazardous waste
identification rules and land disposal restrictions. For further information on
waste management, consult "The Waste Management Manual for Laboratory
Personnel," also available from the American Chemical Society at the address
in Section 14.2.

16.0 REFERENCES

1.	ASTM Annual Book of Standards, Part 11, Volume 11.02, D3694-82, "Standard
Practice for Preparation of Sample Containers and for Preservation," American
Society for Testing and Materials, Philadelphia, PA, p. 86, 1986.

2.	Giam, C.S., Chan, H.S., and Nef, G.S. "Sensitive Method for Determination of
Phthalate Ester Plasticizers in Open-Ocean Biota Samples," Analytical
Chemistry. 47. 2225 (1975).

3.	Giam, C.S. and Chan, H.S. "Control of Blanks in the Analysis of Phthalates in
Air and Ocean Biota Samples," U.S. National Bureau of Standards, Special
Publication 442, pp. 701-708, 1976.

4.	"Carcinogens - Working with Carcinogens," Department of Health, Education,
and Welfare, Public Health Service, Center for Disease Control, National
Institute for Occupational Safety and Health, Publication No. 77-206,

August 1977.

5.	"OSHA Safety and Health Standards, General Industry," (29 CFR 1910),
Occupational Safety and Health Administration, OSHA 2206, (Revised,

January 1976).

6.	"Safety in Academic Chemistry Laboratories," American Chemical Society
Publication, Committee on Chemical Safety, 3rd Edition, 1979.

515.2-23


-------
7.	ASTM Annual Book of Standards, Part 11, Volume 11.01, D3370-82, "Standard
Practice for Sampling Water," American Society for Testing and Materials,
Philadelphia, PA, p. 130, 1986.

8.	Glaser, J.A., Foerst, D.L., McKee, G.D., Quave, S.A., and Budde, W.L. "Trace
Analyses for Wastewaters," Environ. Sci. Technol. 1981, JJ3, 1426-1435.

9.	40 CFR, Part 136, Appendix B.

515.2-24


-------
17.0 TABLES. DIAGRAMS. FLOWCHARTS AND VALIDATION DATA

TABLE 1. RETENTION DATA

Retention Time (min)°

Analyte

Group3

Primary

Confirmation

3,5-Dichlorobenzoic acid

A

16.72

18.98

2,4-Dichlorophenylacetic acid (SA)

A,B

19.78

22.83

Dicamba

B

20.18

23.42

Dichlorprop

A

22.53

25.90

2,4-D

B

23.13

27.01

4,4' -Dibromooctafluorobipheny 1 (IS)

A,B

24.26

26.57

Pentachlorophenol

A

25.03

27.23

Silvex

B

25.82

29.08

5-Hydroxydicamba

B

26.28

30.18

2,4,5-T

A

26.57

30.33

2,4-DB

B

27.95

31.47

Dinoseb

A

28.03

33.02

Bentazon

B

28.70

33.58

Picloram

B

29.93

35.90

Dacthal

A

31.02

34.32

Acifluorfen

B

35.62

40.58

aAnalytes were divided into two groups during method development to avoid
chromatographic overlap.

bColumns and chromatographic conditions are described in Section 6.12.

515.2-25


-------
TABLE 2. SINGLE LABORATORY RECOVERY, PRECISION DATA AND METHOD
DETECTION LIMIT WITH FORTIFIED REAGENT WATER - LEVEL 1

Analyte

Fortified
Cone.

(Pg/L)

Mean3
Recovery
(%)

Relative
Std. Dev.
(%)

MDL
(l-ig/L)

Acifluorfen

0.50

70

21

0.25

Bentazon

2.50

70

11

0.63

2,4-D

0.25

96

38

0.28

2,4-DB

2.50

79

12

0.72

Dacthalb

0.25

96

16

0.13

Dicamba

0.75

109

11

0.28

3,5-Dichlorobenzoic acid

1.25

126

24

1.23

Dichlorprop

0.25

106

15

0.13

Dinoseb

0.50

87

22

0.28

5-Hydroxydicamba

0.75

90

12

0.25

Pentachlorophenol

0.25

103

18

0.16

Picloram

0.75

95

15

0.35

2,4,5-T

0.25

116

18

0.16

2,4,5-TP

0.25

98

9

0.06

aBased on the analyses of seven replicates.
bMeasurement includes the mono- and diacid metabolites.

515.2-26


-------
TABLE 3. SINGLE LABORATORY RECOVERY AND PRECISION DATA
FOR FORTIFIED REAGENT WATER - LEVEL 2



Fortified

Mean3

Relative



Cone.

Recovery

Std. Dev.

Analyte

(l-ig/L)

(%)

(%)

Acifluorfen

0.80

61

27

Bentazon

4.0

81

8

2,4-D

0.40

96

38

2,4-DB

4.0

90

13

Dacthalb

0.40

96

16

Dicamba

1.20

109

11

3,5-Dichlorobenzoic acid

2.00

126

24

Dichlorprop

0.40

76

21

Dinoseb

0.80

87

22

5-Hydroxydicamba

1.20

90

12

Pentachlorophenol

0.40

66

26

Picloram

1.20

68

21

2,4,5-T

0.40

116

18

2,4,5-TP

0.40

105

7

aBased on the analyses of six to seven replicates.
bMeasurement includes the mono- and diacid metabolites.

515.2-27


-------
TABLE 4. SINGLE LABORATORY RECOVERY AND PRECISION DATA
FOR FORTIFIED REAGENT WATER - LEVEL 3



Fortified

Mean3

Relative



Cone.

Recovery

Std. Dev.

Analyte

(Pg/L)

(%)

(%)

Acifluorfen

2.0

59

13

Bentazon

10.0

68

8

2,4-D

1.0

90

20

2,4-DB

10.0

74

6

Dacthalb

1.0

60

10

Dicamba

3.0

75

9

3,5-Dichlorobenzoic acid

5.0

62

18

Dichlorprop

1.0

97

17

Dinoseb

2.0

63

10

5-Hydroxydicamba

3.0

77

8

Pentachlorophenol

1.0

69

11

Picloram

3.0

66

9

2,4,5-T

1.0

64

15

2,4,5-TP

1.0

68

8

aBased on the analyses of six to seven replicates.
bMeasurement includes the mono- and diacid metabolites.

515.2-28


-------
TABLE 5. SINGLE LABORATORY RECOVERY AND PRECISION DATA
FOR FORTIFIED, DECHLORINATED TAP WATER - LEVEL 1



Fortified

Mean3

Relative



Cone.

Recovery

Std. Dev.

Analyte

(Pg/L)

(%)

(%)

Acifluorfen

0.50

117

21

Bentazon

2.50

96

12

2,4-D

0.25

59c

55

2,4-DB

2.50

112

15

Dacthalb

0.25

101

10

Dicamba

0.75

91

14

3,5-Dichlorobenzoic acid

1.25

103

15

Dichlorprop

0.25

218d

37

Dinoseb

0.50

134

10

5-Hydroxydicamba

0.75

90

14

Pentachlorophenol

0.25

91

8

Picloram

0.75

76

28

2,4,5-T

0.25

118

16

2,4,5-TP

0.25

99

10

aBased on the analyses of six to seven replicates.
bMeasurement includes the mono- and diacid metabolites.
C2,4-D background value was 0.29 Hg/L.
dProbable interference.

515.2-29


-------
TABLE 6. SINGLE LABORATORY RECOVERY AND PRECISION DATA
FOR FORTIFIED, DECHLORINATED TAP WATER - LEVEL 2



Fortified

Mean3

Relative



Cone.

Recovery

Std. Dev.

Analyte

(Pg/L)

(%)

(%)

Acifluorfen

2.0

150

7

Bentazon

10.0

112

9

2,4-D

1.0

90

16

2,4-DB

10.0

111

10

Dacthalb

1.0

118

8

Dicamba

3.0

86

10

3,5-Dichlorobenzoic acid

5.0

111

5

Dichlorprop

1.0

88

30

Dinoseb

2.0

121

6

5-Hydroxydicamba

3.0

96

6

Pentachlorophenol

1.0

96

6

Picloram

3.0

132

12

2,4,5-T

1.0

108

10

2,4,5-TP

1.0

115

7

2,4-Dichlorophenylacetic acidc

1.0

120

19

aBased on the analyses of six to seven replicates.
bMeasurement includes the mono- and diacid metabolites.
Surrogate analyte.

515.2-30


-------
TABLE 7. SINGLE LABORATORY RECOVERY AND PRECISION DATA
FOR FORTIFIED, OZONATED SURFACE WATER - LEVEL 1



Fortified

Mean3

Relative



Cone.

Recovery

Std. Dev.

Analyte

(|Jg/L)

(%)

(%)

Acifluorfen

0.50

172

14

Bentazon

2.50

92

22

2,4-D

0.25

127

13

2,4-DB

2.50

154

19

Dacthalb

0.25

113

17

Dicamba

0.75

107

13

3,5-Dichlorobenzoic acid

1.25

100

17

Dichlorprop

0.25

115

20

Dinoseb

0.50

134

28

5-Hydroxydicamba

0.75

89

13

Pentachlorophenol

0.25

110

22

Picloram

0.75

109

27

2,4,5-T

0.25

102

19

2,4,5-TP

0.25

127

8

2,4-Dichlorophenylacetic acidc

0.25

72

31

aBased on the analyses of six to seven replicates.
bMeasurement includes the mono- and diacid metabolites.
Surrogate analyte.

515.2-31


-------
TABLE 8. SINGLE LABORATORY RECOVERY AND PRECISION DATA FOR
FORTIFIED, OZONATED SURFACE WATER - LEVEL 2, STABILITY

STUDY DAY lc



Fortified

Mean3

Relative



Cone.

Recovery

Std. Dev.

Analyte

(l-ig/L)

(%)

(%)

Acitluorfen

2.0

173

11

Bentazon

10.0

122

7

2,4-D

1.0

126

10

2,4-DB

10.0

130

7

Dacthalb

1.0

116

11

Dicamba

3.0

109

9

3,5-Dichlorobenzoic acid

5.0

115

11

Dichlorprop

1.0

116

11

Dinoseb

2.0

116

9

5-Hydroxydicamba

3.0

121

9

Pentachlorophenol

1.0

118

10

Picloram

3.0

182

14

2,4,5-T

1.0

112

9

2,4,5-TP

1.0

122

10

2,4-Dichlorophenylacetic acidd

1.0

110

26

aBased on the analyses of six to seven replicates.
bMeasurement includes the mono- and diacid metabolites.
cSamples preserved at pH = 2.0.

Surrogate analyte.

515.2-32


-------
TABLE 9. SINGLE LABORATORY RECOVERY AND PRECISION DATA FOR
FORTIFIED, OZONATED SURFACE WATER - LEVEL 2, STABILITY

STUDY DAY 14c



Fortified

Mean3

Relative



Cone.

Recovery

Std. Dev.

Analyte

(Pg/L)

(%)

(%)

Acifluorfen

2.0

151

18

Bentazon

10.0

97

9

2,4-D

1.0

84

11

2,4-DB

10.0

128

10

Dacthalb

1.0

116

7

Dicamba

3.0

103

9

3,5-Dichlorobenzoic acid

5.0

81

12

Dichlorprop

1.0

107

11

Dinoseb

2.0

118

7

5-Hydroxydicamba

3.0

20

14

Pentachlorophenol

1.0

94

7

Picloram

3.0

110

32

2,4,5-T

1.0

113

8

2,4,5-TP

1.0

113

11

2,4-Dichlorophenylacetic acidd

1.0

87

6

aBased on the analyses of six to seven replicates.
bMeasurement includes the mono- and diacid metabolites.
cSamples preserved at pH = 2.0.

Surrogate analyte.

515.2-33


-------
TABLE 10. SINGLE LABORATORY RECOVERY AND PRECISION DATA FOR
FORTIFIED, HIGH HUMIC CONTENT SURFACE WATER



Fortified

Mean3

Relative



Cone.

Recovery

Std. Dev.

Analyte

(Pg/L)

(%)

(%)

Acifluorfen

2.0

120

13

Bentazon

10.0

87

11

2,4-D

1.0

59

7

2,4-DB

10.0

80

14

Dacthalb

1.0

100

6

Dicamba

3.0

76

9

3,5-Dichlorobenzoic acid

5.0

87

4

Dichlorprop

1.0

110

22

Dinoseb

2.0

97

6

5-Hydroxydicamba

3.0

82

9

Pentachlorophenol

1.0

70

5

Picloram

3.0

124

9

2,4,5-T

1.0

101

4

2,4,5-TP

1.0

80

6

aBased on the analyses of six to seven replicates.
bMeasurement includes the mono- and diacid metabolites.

515.2-34


-------
TABLE 11. LABORATORY PERFORMANCE CHECK SOLUTION

Test

Analyte

Cone,
Hg/mL

Requirements

Sensitivity

Dinoseb

0.004

Detection of analyte; S/N >3

Chromatographic performance

4-Nitrophenol

1.6

0.70 < PGF <1.05

Column performance

3,5-Dichlorobenzoic acid

0.6

Resolution >0.40b



4-Nitrophenol

1.6



aPGF - peak Gaussian factor. Calculated using the equation:

PGF _ 1-83 x W (1/2)

W (1/10)

where: W(l/2) is the peak width at half height and W(l/10) is the peak width at tenth height.
bResolution between the two peaks as defined by the equation:

R = —

W

where: t is the difference in elution times between the two peaks and W is the average peak width, at the baseline, of
the two peaks.


-------
n, aow—*

FIGURE 1. DIAZOMETHANE GENERATOR

515.2-36


-------
DiCflUL

On
[NJ

00
-<1

4.00-

£ 3.00

o
>

IF*

t>

*4

K 2.00-

1.00-

jsocn

—I—
1.50

II

SU» "F

DIWSRI

J4ST



vv*

LvjJkJc=:

2.00

2.50
x 101

i	|

3.00
¦inutes

	,—

3.50

Figure 2A. Chromatogran of Group A Analytes Extracted From
Ozonated Surface Water (bottota chromatogran is
the laboratory reagent blank)


-------
DICAH1A

Figure 28. Chromatogram of Group B Analytes Extracted Fro«i
Ozonated Surface Water (bottom chromatogram is
the laboratory reagent blank)


-------