THE ENVIRONMENTAL TECHNOLOGY VERIFICATION

PROGRAM ^

ETV

SEPA

ETV Joint Verification Statement

Batteiie

U.S. Environmental Protection Agency	Puttjng Tcchnohgy To Work

TECHNOLOGY TYPE: RAPID TOXICITY TESTING SYSTEM
APPLICATION:	DETECTING TOXICITY IN DRINKING WATER

TECHNOLOGY NAME: POLYTOX™

COMPANY:	Interlab Supply, Ltd.

ADDRESS:	9200 New Trails Drive	PHONE: 888-876-2844

The Woodlands, TX, 77381 FAX: 281-298-9411
WEB SITE:	http://www.interlabsupply.com

E-MAIL:	pete@interlabsupply.com

The U.S. Environmental Protection Agency (EPA) supports the Environmental Technology Verification (ETV)
Program to facilitate the deployment of innovative or improved environmental technologies through performance
verification and dissemination of information. The goal of the ETV Program is to further environmental protection
by accelerating the acceptance and use of improved and cost-effective technologies. ETV seeks to achieve this goal
by providing high-quality, peer-reviewed data on technology performance to those involved in the design,
distribution, financing, permitting, purchase, and use of environmental technologies.

ETV works in partnership with recognized standards and testing organizations, with stakeholder groups
(consisting of buyers, vendor organizations, and permitters), and with individual technology developers. The
program evaluates the performance of innovative technologies by developing test plans that are responsive to the
needs of stakeholders, conducting field or laboratory tests (as appropriate), collecting and analyzing data, and pre-
paring peer-reviewed reports. All evaluations are conducted in accordance with rigorous quality assurance (QA)
protocols to ensure that data of known and adequate quality are generated and that the results are defensible.

The Advanced Monitoring Systems (AMS) Center, one of seven technology areas under ETV, is operated by
Batteiie in cooperation with EPA's National Exposure Research Laboratory. The AMS Center has recently
evaluated the performance of rapid toxicity testing systems used to detect toxicity in drinking water. This
verification statement provides a summary of the test results for the POLYTOX™ testing system.

VERIFICATION TEST DESCRIPTION

Rapid toxicity technologies use bacteria, enzymes, or small crustaceans that will produce light or use oxygen at a
steady rate in the absence of toxic contaminants. Toxic contaminants in drinking water are indicated by a change in
the color or intensity of light or by a decrease in the dissolved oxygen uptake rate (DOUR) in the presence of the


-------
contaminants. As part of this verification test, which took place between July 14 and August 22, 2003, various
contaminants were added to separate drinking water samples and analyzed by POLYTOX™. Response to
interfering compounds in clean drinking water also was evaluated. Dechlorinated drinking water samples from
Columbus, Ohio, (DDW) were fortified with contaminants at concentrations ranging from lethal levels to
concentrations several orders of magnitude less than the lethal dose and analyzed. Endpoint and precision, toxicity
threshold for each contaminant, false positive/negative responses, ease of use, and sample throughput were
evaluated.

Inhibition results (endpoints) from four replicates of each contaminant at each concentration level were evaluated
to assess the ability of the POLYTOX™ to detect toxicity at various concentrations of contaminants, as well as to
measure the precision of the POLYTOX™ results. The response of POLYTOX™ to compounds used during the
water treatment process (interfering compounds) was evaluated by analyzing separate aliquots of DDW fortified
with each potential interferent at approximately one-half of the concentration limit recommended by the EPA's
National Secondary Drinking Water Regulations guidance. For analysis of by-products of the chlorination process,
unspiked DDW was analyzed because Columbus, Ohio, uses chlorination as its disinfectant procedure. For the
analysis of by-products of the chloramination process, a separate drinking water sample from St. Petersburg,
Florida, which uses chloramination as its disinfection process, was obtained. The samples were analyzed after
residual chlorine was removed using sodium thiosulfate. Sample throughput was measured based on the number of
samples analyzed per hour. Ease of use and reliability were determined based on documented observations of the
operators and the verification test coordinator.

Quality control samples included method blank samples, which consisted of American Society for Testing and
Materials Type II deionized water; positive control samples fortified with cyanide; and negative control samples,
which consisted of the unspiked DDW.

QA oversight of verification testing was provided by Battelle and EPA. Battelle QA staff conducted a technical
systems audit, a performance evaluation audit, and a data quality audit of 10% of the test data. EPA QA staff also
performed a technical systems audit while testing was being conducted.

TECHNOLOGY DESCRIPTION

The following description of POLYTOX™ was provided by the vendor and was not subjected to verification in
this test.

POLYTOX™ uses the respiration of microorganisms to indicate the toxicity of a water or wastewater stream.

When activated in water, the mixture of bacterial cultures in POLYTOX™ begins to "breathe" like all other living
organisms. They breathe in oxygen and respire carbon dioxide. The inhibitory effect of toxicants in potable tap
water (or any water-based medium) to the bacterial cultures in POLYTOX™ is measured by evaluating the
culture's respiration rate in the presence of different concentrations of toxicants. The respiration rate is the oxygen
consumed by aerobic and facultative cultures (the dissolved oxygen update rate—DOUR) and is expressed as
milligrams (mg) of oxygen consumed per liter per minute.

The DOUR is determined by measuring the dissolved oxygen concentration at 19 and 21 minutes after adding the
POLYTOX™ microbial mixture to 300 milliliters (mL) of a drinking water sample. The DOUR of each drinking
water sample is compared to a baseline DOUR measured at the beginning of each day by adding POLYTOX™ to
a clean water matrix and measuring the oxygen concentrations in a manner similar to the test samples. For this
verification test, the vendor provided YSI 5000 and 5100 dissolved oxygen probes.

The toxicity of a contaminant was detectable if its inhibition was significantly greater than the negative control.
The average inhibition of the 50 negative controls analyzed using POLYTOX™ was 3 ± 15%; therefore, for any
result to be detected, the inhibition had to be greater than 18%.


-------
The POLYTOX™ test components include standard 300-mL biological oxygen demand (BOD) bottle(s) and a
dissolved oxygen probe (with stirrer) and meter. The probe must fit snugly into the neck of the BOD bottle,
eliminating all headspace. Also required, but not included in the test kit, are an aeration device, one- and two-liter
containers for aerating the DI water (control), and pH adjusting solutions. A thermometer and a stopwatch are also
provided. The dimensions of the POLYTOX™ test kit are 8 inches x 8 inches x 4 inches. With all necessary
components, the kit size is approximately 18 inches x 18 inches x 16 inches. The dissolved oxygen probe and
meter are 9-lA inches x 8-/2 inches x 6 inches. When a large number of tests are performed, data can be down-
loaded directly from the dissolved oxygen meter to a laptop or desktop computer for manipulation into a usable
form. The suggested price of the POLYTOX™ culture is $147 for 20 tests. The dissolved oxygen probe and meter
provided by the vendor for use during testing cost approximately $1,600 for the complete unit.

VERIFICATION OF PERFORMANCE

Endpoint and Precision/Toxicity Threshold: The table below presents POLYTOX™ percent inhibition data and
range of standard deviations for the contaminants and potential interferences that were tested. The toxicity
thresholds also are shown for each contaminant tested.

Parameter

Compound

Lethal
Dose (LD)
Cone.
(mg/L)

Average Inhibitions at
Concentrations Relative to the LD
Concentration (%)

Range of
Standard
Deviations (%)

Toxicity
Thresh.
(mg/L)

LD

LD/10

LD/100

LD/1,000

Contaminants in
DDW

Aldicarb

280

22

-3

-16

NA(b)

5-12

ND(C)

Colchicine

240

-13

-6

-9

-13

6-23

ND

Cyanide

250(a)

86

61

3

11

2-6

0.25

Dicrotophos

1,400

-5

8

18

16

4-31

ND

Lhallium
sulfate

2,400

34

20

3

2

4-14

2,400

Botulinum

toxin(d)

0.30

3

6

14

9

5-15

ND

Ricin(e)

15

44

26

25

16

7-21

15

Soman

0.18'°

1

6

5

-6

5-23

ND

VX

0.088'°

4

19

0

0

9-19

ND

Potential
interferences in
DDW

Interference

Cone.
(mg/L)

Average Inhibitions at a
Single Concentration (%)

Standard
Deviation (%)

Aluminum

0.36

-8

11

Copper

0.65

5

4

Iron

0.069

7

9

Manganese

0.26

6

5

Zinc

3.5

11

3

(a)	LD/10, LD/100, LD/1,000 concentrations for cyanide are 0.25, 0.0025, and 0.00025 mg/L respectively.

(b)	NA = Not applicable.

(c)	ND = Not detectable.

(d)	Lethal dose solution also contained 3 mg/L phosphate and 1 mg/L sodium chloride.

(e)	Lethal dose solution also contained 3 mg/L phosphate, 26 mg/L sodium chloride, and 2 mg/L sodium azide.

(f)	Due to the degradation of soman and VX in water, the stock solution confirmation analysis confirmed that the
concentration of the lethal dose of soman was 61% of the expected concentration of 0.30 mg/L and of VX was 44% of
the expected concentration of 0.20 mg/L.


-------
False Positive/Negative Responses: Average inhibitions for chloraminated and chlorinated water were 27% ± 4%
and 10% ± 15%, respectively. The chloraminated water produced false positive responses because the inhibition
was significantly greater than the negative control. However, both types of water had positive average inhibitions,
so a water matrix similar to the negative control should be used to compare background inhibition due to
disinfectant by-products and inhibition due to toxic contaminants. At the lethal concentration level, inhibitions
produced by aldicarb, colchicine, dicrotophos, botulinum toxin, soman, and VX were not significantly different
from the negative control or inhibition was generated by lower concentrations of the same contaminant, indicating
false negative responses.

Field Portability: The performance of POLYTOX™ in the field was similar to its performance in the laboratory
for the one contaminant (cyanide) that was tested in both locations. A flat, sturdy surface is needed for BOD
bottles and oxygen probes. Not including reference and background samples, 300 mL of waste were generated for
every sample. A carrying case was not provided. The overflow upon inserting the oxygen probe required a
secondary container.

Other Performance Factors: The pictorial manual was useful, and sample throughput was three samples per
hour. Each sample required 300 mL of water; 8 liters per day of waste were generated per oxygen probe. Oxygen
probe membranes were changed once every 40 to 50 samples. Although the operators for this test had scientific
backgrounds, based on the observations of the verification test coordinator, operators with little technical training
would probably be able to use POLYTOX™ to successfully analyze samples.

Original signed by Gabor J. Kovacs 11/13/03
Gabor J. Kovacs	Date

Vice President
Environmental Sector
Battelle

Original signed by Timothy E. Qppelt	12/1/03

Timothy E. Oppelt	Date

Director

National Homeland Security Research Center
U.S. Environmental Protection Agency

NOTICE: ETV verifications are based on an evaluation of technology performance under specific, predetermined
criteria and the appropriate quality assurance procedures. EPA and Battelle make no expressed or implied
warranties as to the performance of the technology and do not certify that a technology will always operate as
verified. The end user is solely responsible for complying with any and all applicable federal, state, and local
requirements. Mention of commercial product names does not imply endorsement.


-------