METHOD 524.2 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN WATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY Revision 4.0 August 1992 A. Alford-Stevens, J. W. Eichelberger, and W.L. Budde Method 524, Revision 1.0 (1983) R.W. Slater, Jr. Revision 2.0 (1986) J.W. Eichelberger, and W.L. Budde Revision 3.0 (1989) J.W. Eichelberger, J.W. Munch, and T.A.Bellar Revision 4.0 (1992) ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH AND DEVELOPMENT U.S. ENVIRONMENTAL PROTECTION AGENCY CINCINNATI, OHIO 45268 METHOD 524.2 524.2-1 ------- MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN WATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY SCOPE AND APPLICATION 1.1 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in surface water, ground water, and drinking water in any stage of treatment1,2. The method is applicable to a wide range of organic compounds, including the four trihalomethane disinfection by-products, that have sufficiently high volatility and low water solubility to be removed from water samples with purge and trap procedures. The following compounds can be determined by this method. Compound Chemical Abstract Service Registry Number Acetone* 67-64-1 Acrylonitrile* 107-13-1 Allyl chloride* 107-05-1 Benzene 71-43-2 Bromobenzene 108-86-1 Bromochloromethane 74-97-5 Bromodichloromethane 75-27-4 Bromoform 75-25-2 Bromomethane 74-83-9 2-Butanone* 78-93-3 n-Butylbenzene 104-51-8 sec-Butylbenzene 135-98-8 tert-Butylbenzene 98-06-6 Carbon disulfide* 75-15-0 Carbon tetrachloride 56-23-5 Chloroacetonitrile* 107-14-2 Chlorobenzene 108-90-7 1-Chlorobutane* 109-69-3 Chloroethane 75-00-3 Chloroform 67-66-3 Chloromethane 74-87-3 2-Chlorotoluene 95-49-8 4-Chlorotoluene 106-43-4 Dibromochloromethane 124-48-1 1,2-Dibromo-3-chloropropane 96-12-8 1,2-Dibromoethane 106-93-4 Dibromomethane 74-95-3 1,2-Dichlorobenzene 95-50-1 1,3-Dichlorobenzene 541-73-1 1,,4-Dichlorobenzene 106-46-7 524.2-2 ------- Compound Chemical Abstract Service Registry Number trans-1,4-Dichloro-2-butene* 110-57-6 Dichlorodifluoromethane 75-71-8 1,1 -Dichloroethane 75-34-3 1,2-Dichloroethane 107-06-2 1,1 -Dichloroethene 75-35-4 cis-1,2-Dichloroethene 156-59-4 trans-1,2-Dichloroethene 156-60-5 1,2-Dichloropropane 78-87-5 1,3-Dichloropropane 142-28-9 2,2-Dichloropropane 590-20-7 1,1 -Dichloropropene 563-58-6 1,1 -Dichloropropanone* 513-88-2 cis-1,3-Dichloropropene 10061-01-5 trans-1,3-Dichloropropene 10061-02-6 Diethyl ether* 60-29-7 Ethylbenzene 100-41-4 Ethyl methacrylate* 97-63-2 Hexachlorobutadiene 87-68-3 Hexachloroethane* 67-72-1 2-Hexanone* 591-78-6 Isopropylbenzene 98-82-8 4-Isopropyltoluene 99-87-6 Methacrylonitrile* 126-98-7 Methylacrylate* 96-33-3 Methylene chloride 75-09-2 Methyl iodide* 74-88-4 Methylmethacrylate* 80-62-6 4-Methyl-2-pentanone* 108-10-1 Methyl-t-butyl ether* 1634-04-4 Naphthalene 91-20-3 Nitrobenzene* 98-95-3 2-Nitropropane* 79-46-9 Pentachloroethane* 76-01-7 Propionitrile* 107-12-0 n-Propylbenzene 103-65-1 Styrene 100-42-5 1,1,1,2-T etrachloroethane 630-20-6 1,1,2,2-T etrachloroethane 79-34-5 T etrachlor oethene 127-18-4 T etrahydrofuran* 109-99-9 Toluene 108-88-3 1,2,3-Trichlorobenzene 87-61-6 1,2,4-Trichlorobenzene 120-82-1 1,1,1 -T richloroethane 71-55-6 1,1,2-T richloroethane 79-00-5 Trichloroethene 79-01-6 T richlorofluoromethane 75-69-4 524.2-3 ------- Compound Chemical Abstract Service Registry Number 1,2,3-Trichloropropane 96-18-4 1,2,4-Trimethylbenzene 95-63-6 1,3,5-Trimethylbenzene 108-67-8 Vinyl chloride 75-01-4 o-Xylene 95-47-6 m-Xylene 108-38-3 p-Xylene 106-42-3 *New Compound in Revision 4.0. 1.2 Method detection limits (MDLs)3 are compound, instrument and especially matrix dependent and vary from approximately 0.02-1.6 Hg/L. The applicable concentration range of this method is primarily column and matrix dependent, and is approximately 0.02-200 (Jg/L when a wide-bore thick-film capillary column is used. Narrow-bore thin-film columns may have a capacity which limits the range to about 0.02-20 Hg/L. Volatile water soluble, polar compounds which have relatively low purging efficiencies can be determined using this method. Such compounds may be more susceptible to matrix effects, and the quality of the data may be adversely influenced. 1.3 Analytes that are not separated chromatographically, but which have different mass spectra and noninterfering quantitation ions (Table 1), can be identified and measured in the same calibration mixture or water sample as long as their concentrations are somewhat similar (Section 11.6.2). Analytes that have very similar mass spectra cannot be individually identified and measured in the same calibration mixture or water sample unless they have different retention times (Section 11.6.3). Coeluting compounds with very similar mass spectra, typically many structural isomers, must be reported as an isomeric group or pair. Two of the three isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column, and if not, must be reported as isomeric pairs. The more water soluble compounds (>2% solubility) and compounds with boiling points above 200°C are purged from the water matrix with lower efficiencies. These analytes may be more susceptible to matrix effects. SUMMARY OF METHOD 2.1 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an inert gas through the aqueous sample. Purged sample components are trapped in a tube containing suitable sorbent materials. When purging is complete, the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (MS). The column is temperature programmed to facilitate the separation of the method analytes which are then detected with the MS. Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base. Reference 524.2-4 ------- spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples. The concentration of each identified component is measured by relating the MS response of the quantitation ion produced by that compound to the MS response of the quantitation ion produced by a compound that is used as an internal standard. Surrogate analytes, whose concentrations are known in every sample, are measured with the same internal standard calibration procedure. DEFINITIONS 3.1 Internal Standard (IS) - A pure analyte(s) added to a sample, extract, or standard solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same sample or solution. The internal standard must be an analyte that is not a sample component. 3.2 Surrogate Analyte (SA) — A pure analyte(s), which is extremely unlikely to be found in any sample, and which is added to a sample aliquot in known amount(s) before extraction or other processing and is measured with the same procedures used to measure other sample components. The purpose of the SA is to monitor method performance with each sample. 3.3 Laboratory Duplicates (LD1 and LD2) — Two aliquots of the same sample taken in the laboratory and analyzed separately with identical procedures. Analyses of LD1 and LD2 indicates precision associated with laboratory procedures, but not with sample collection, preservation, or storage procedures. 3.4 Field Duplicates (FD1 and FD2) — Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures. Analyses of FD1 and FD2 give a measure of the precision associated with sample collection, preservation and storage, as well as with laboratory procedures. 3.5 Laboratory Reagent Blank (LRB) — An aliquot of reagent water or other blank matrix that is treated exactly as a sample including exposure to all glassware, equipment, solvents, reagents, internal standards, and surrogates that are used with other samples. The LRB is used to determine if method analytes or other interferences are present in the laboratory environment, the reagents, or the apparatus. 3.6 Field Reagent Blank (FRB) — An aliquot of reagent water or other blank matrix that is placed in a sample container in the laboratory and treated as a sample in all respects, including shipment to the sampling site, exposure to sampling site conditions, storage, preservation, and all analytical procedures. The purpose of the FRB is to determine if method analytes or other interferences are present in the field environment. 3.7 Laboratory Performance Check Solution (LPC) — A solution of one or more compounds (analytes, surrogates, internal standard, or other test compounds) 524.2-5 ------- used to evaluate the performance of the instrument system with respect to a defined set of method criteria. 3.8 Laboratory Fortified Blank (LFB) — An aliquot of reagent water or other blank matrix to which known quantities of the method analytes are added in the laboratory. The LFB is analyzed exactly like a sample, and its purpose is to determine whether the methodology is in control, and whether the laboratory is capable of making accurate and precise measurements. 3.9 Laboratory Fortified Sample Matrix (LFM) — An aliquot of an environmental sample to which known quantities of the method analytes are added in the laboratory. The LFM is analyzed exactly like a sample, and its purpose is to determine whether the sample matrix contributes bias to the analytical results. The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values in the LFM corrected for background concentrations. 3.10 Stock Standard Solution (SSS) — A concentrated solution containing one or more method analytes prepared in the laboratory using assayed reference materials or purchased from a reputable commercial source. 3.11 Primary Dilution Standard Solution (PDS) — A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions. 3.12 Calibration Standard (CAL) — A solution prepared from the primary dilution standard solution or stock standard solutions and the internal standards and surrogate analytes. The CAL solutions are used to calibrate the instrument response with respect to analyte concentration. 3.13 Quality Control Sample (QCS) — A solution of method analytes of known concentrations which is used to fortify an aliquot of LRB or sample matrix. The QCS is obtained from a source external to the laboratory and different from the source of calibration standards. It is used to check laboratory performance with externally prepared test materials. INTERFERENCES 4.1 During analysis, major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap. The use of Teflon tubing, Teflon thread sealants, or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation. Analyses of laboratory reagent blanks provide information about the presence of contaminants. When potential interfering peaks are noted in laboratory reagent blanks, the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter. Subtracting blank values from sample results is not permitted. 524.2-6 ------- 4.2 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds. A preventive technique is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water. After analysis of a sample containing high concentrations of volatile organic compounds, one or more laboratory reagent blanks should be analyzed to check for cross-contamination. 4.3 Special precautions must be taken to determine methylene chloride. The analytical and sample storage area should be isolated from all atmospheric sources of methylene chloride, otherwise random background levels will result. Since methylene chloride will permeate Teflon tubing, all GC carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing. Laboratory worker's clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquid/liquid extraction procedures can contribute to sample contamination. 4.4 Traces of ketones, methylene chloride, and some other organic solvents can be present even in the highest purity methanol. This is another potential source of contamination, and should be assessed before standards are prepared in the methanol. 5.0 SAFETY 5.1 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined; each chemical should be treated as a potential health hazard, and exposure to these chemicals should be minimized. Each laboratory is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method. Additional references to laboratory safety are available4 6 for the information of the analyst. 5.2 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens: benzene, carbon tetrachloride, 1,4-dichlorobenzene, 1,2-dichlorethane, hexachlorobutadiene, 1,1,2,2-tetrachloroethane, 1,1,2-trichloroethane, chloroform, l,2-dibromoethane,tetrachloroethene, trichloroethene, and vinyl chloride. Pure standard materials and stock standard solutions of these compounds should be handled in a hood. A NIOSH/MESA approved toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds. 6.0 EQUIPMENT AND SUPPLIES 6.1 Sample Containers — 40-120 mL screw cap vials each equipped with a Teflon faced silicone septum. Prior to use, wash vials and septa with detergent and rinse with tap and distilled water. Allow the vials and septa to air dry at room temperature, place in a 105°C oven for one hour, then remove and allow to cool in an area known to be free of organics. 524.2-7 ------- Purge and Trap System — The purge and trap system consists of three separate pieces of equipment: purging device, trap, and desorber. Systems are commercially available from several sources that meet all of the following specifications. 6.2.1 The all glass purging device (Figure 1) should be designed to accept 25 mL samples with a water column at least 5 cm deep. A smaller (5 mL) purging device is recommended if the GC/MS system has adequate sensitivity to obtain the method detection limits required. Gaseous volumes above the sample must be kept to a minimum (<15 mL) to eliminate dead volume effects. A glass frit should be installed at the base of the sample chamber so the purge gas passes through the water column as finely divided bubbles with a diameter of <3 mm at the origin. Needle spargers may be used, however, the purge gas must be introduced at a point about 5 mm from the base of the water column. The use of a moisture control device is recommended to prohibit much of the trapped water vapor from entering the GC/MS and eventually causing instrumental problems. 6.2.2 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0.105 in. Starting from the inlet, the trap should contain 1.0 cm of methyl silicone coated packing and the following amounts of adsorbents: Va of 2,6-diphenylene oxide polymer, Va of silica gel, and Va of coconut charcoal. If it is not necessary to determine dichlorodifluoromethane, the charcoal can be eliminated and the polymer increased to fill % of the trap. Before initial use, the trap should be conditioned overnight at 180°C by backflushing with an inert gas flow of at least 20 mL/min. Vent the trap effluent to the room, not to the analytical column. Prior to daily use, the trap should be conditioned for 10 minutes at 180°C with backflushing. The trap may be vented to the analytical column during daily conditioning; however, the column must be run through the temperature program prior to analysis of samples. The use of alternative sorbents is acceptable, depending on the particular set of target analytes or other problems encountered, but the new trap packing must meet all quality control criteria described in Section 9.0. 6.2.3 The use of the methyl silicone coated packing is recommended, but not mandatory. The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols, and also of insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots. Alternatively, silanized glass wool may be used as a spacer at the trap inlet. 6.2.4 The desorber (Figure 2) must be capable of rapidly heating the trap to 180°C either prior to or at the beginning of the flow of desorption gas. The polymer section of the trap should not be heated higher than 200°C or the life expectancy of the trap will decrease. Trap failure is characterized by a pressure drop in excess of 3 lb/in2 across the trap 524.2-8 ------- during purging or by poor bromoform sensitivities. The desorber design illustrated in Figure 2 meets these criteria. Gas Chromatography/Mass Spectrometer/Data System (GC/MS/DS) 6.3.1 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation. If the column oven is to be cooled to 10°C or lower, a subambient oven controller will likely be required. If syringe injections of 4-bromofluorobenzene (BFB) will be used, a split/splitless injection port is required. 6.3.2 Capillary GC Columns — Any gas chromatography column that meets the performance specifications of this method may be used (Section 10.2.4.1). Separations of the calibration mixture must be equivalent or better than those described in this method. Four useful columns have been evaluated, and observed compound retention times for these columns are listed in Table 2. 6.3.2.1 Column 1 — 60 m x 0.75 mm ID VOCOL (Supelco, Inc.) glass wide-bore capillary with a 1.5 |im film thickness. Column 2 — 30 m x 0.53 mm ID DB-624 (J&W Scien-tific, Inc.) fused silica capillary with a 3 |im film thickness. Column 3 — 30 m x 0.32 mm ID DB-5 (J&W Scientific, Inc.) fused silica capillary with a 1 |im film thickness. Column 4 ~ 75 m x 0.53 mm id DB-624 (J&W Scien-tific, Inc.) fused silica capillary with a 3 |im film thickness. 6.3.3 Interfaces between the GC and MS — The interface used depends on the column selected and the gas flow rate. 6.3.3.1 The wide-bore Columns 1, 2, and 4 have the capacity to accept the standard gas flows from the trap during thermal desorption, and chromatography can begin with the onset of thermal desorption. Depending on the pumping capacity of the MS, an additional interface between the end of the column and the MS may be required. An open split interface7 or an all-glass jet separator is an acceptable interface. Any interface can be used if the performance specifications described in this method (Section 9.0 and 10.0) can be achieved. The end of the transfer line after the interface, or the end of the analytical column if no interface is used, should be placed within a few mm of the MS ion source. 6.3.3.2 When narrow-bore Column 3 is used, a cryogenic interface placed just in front of the column inlet is suggested. This interface condenses the desorbed sample components in a narrow band on an uncoated fused silica precolumn using liquid nitrogen cooling. 524.2-9 ------- When all analytes have been desorbed from the trap, the interface is rapidly heated to transfer them to the analytical column. The end of the analytical column should be placed within a few mm of the MS ion source. A potential problem with this interface is blockage of the interface by frozen water from the trap. This condition will result in a major loss in sensitivity and chromatographic resolution. 6.3.4 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV. The spectrometer must be capable of scanning from 35-260 amu with a complete scan cycle time (including scan overhead) of two seconds or less. (Scan cycle time = Total MS data acquisition time in seconds divided by number of scans in the chromatogram.) The spectrometer must produce a mass spectrum that meets all criteria in Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC. An average spectrum across the BFB GC peak may be used to test instrument performance. 6.3.5 An interfaced data system is required to acquire, store, reduce, and output mass spectral data. The computer software should have the capability of processing stored GC/MS data by recognizing a GC peak within any given retention time window, comparing the mass spectra from the GC peak with spectral data in a user-created data base, and generating a list of tentatively identified compounds with their retention times and scan numbers. The software must allow integration of the ion abundance of any specific ion between specified time or scan number limits. The software should also allow calculation of response factors as defined in Section 10.2.6 (or construction of a linear or second order regression calibration curve), calculation of response factor statistics (mean and standard deviation), and calculation of concentrations of analytes using either the calibration curve or the equation in Section 12.0. 6.4 Syringe and Syringe Valves 6.4.1 Two 5 mL or 25 mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used). 6.4.2 Three two-way syringe valves with Luer ends. 6.4.3 Micro syringes — 10, 100 |iL. 6.4.4 Syringes — 0.5, 1.0, and 5 mL, gas tight with shut-off valve. 6.5 Miscellaneous 6.5.1 Standard solution storage containers — 15 mL bottles with Teflon lined screw caps. REAGENTS AND STANDARDS 7.1 Trap Packing Materials 524.2-10 ------- 7.1.1 2,6-Diphenylene oxide polymer, 60/80 mesh, chromatographic grade (Tenax GC or equivalent). 7.1.2 Methyl silicone packing (optional) — OV-1 (3%) on Chromosorb W, 60/80 mesh, or equivalent. 7.1.3 Silica gel — 35/60 mesh, Davison, grade 15 or equivalent. 7.1.4 Coconut charcoal — Prepare from Barnebey Cheney, CA-580-26 lot #M-2649 by crushing through 26 mesh screen. 7.2 Reagents 7.2.1 Methanol — Demonstrated to be free of analytes. 7.2.2 Reagent water — Prepare reagent water by passing tap water through a filter bed containing about 0.5 kg of activated carbon, by using a water purification system, or by boiling distilled water for 15 minutes followed by a one hour purge with inert gas while the water temperature is held at 90°C. Store in clean, narrow-mouth bottles with Teflon lined septa and screw caps. 7.2.3 Hydrochloric acid (1 + 1) — Carefully add measured volume of conc. HC1 to equal volume of reagent water. 7.2.4 Vinyl chloride — Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several sources (for example, Matheson, Ideal Gas Products, and Scott Gases). 7.2.5 Ascorbic acid — ACS reagent grade, granular. 7.2.6 Sodium thiosulfate — ACS reagent grade, granular. 7.3 Stock Standard Solutions — These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures. One of these solutions is required for every analyte of concern, every surrogate, and the internal standard. A useful working concentration is about 1-5 mg/mL. 7.3.1 Place about 9.8 mL of methanol into a 10 mL ground-glass stoppered volumetric flask. Allow the flask to stand, unstoppered, for about 10 minutes or until all alcohol-wetted surfaces have dried and weigh to the nearest 0.1 mg. 7.3.2 If the analyte is a liquid at room temperature, use a 100 |iL syringe and immediately add two or more drops of reference standard to the flask. Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask. If the analyte is a gas at room temperature, fill a 5 mL valved gas-tight syringe with the standard to the 524.2-11 ------- 5.0 mL mark, lower the needle to 5 mm above the methanol meniscus, and slowly inject the standard into the neck area of the flask. The gas will rapidly dissolve in the methanol. 7.3.3 Reweigh, dilute to volume, stopper, then mix by inverting the flask several times. Calculate the concentration in |ig/|iL from the net gain in weight. When compound purity is certified at 96% or greater, the weight can be used without correction to calculate the concentration of the stock standard. 7.3.4 Store stock standard solutions in 15 mL bottles equipped with Teflon lined screw caps. Methanol solutions of acrylonitrile, methyl iodide, and methyl acrylate are stable for only one week at 4°C. Methanol solutions prepared from other liquid analytes are stable for at least four weeks when stored at 4°C. Methanol solutions prepared from gaseous analytes are not stable for more than one week when stored at <0°C; at room temperature, they must be discarded after one day. 7.4 Primary Dilution Standards — Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern in methanol or other suitable solvent. The primary dilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibration solutions that will bracket the working concentration range. Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation, especially just before preparing calibration solutions. Storage times described for stock standard solutions in Section 7.3.4 also apply to primary dilution standard solutions. 7.5 Fortification Solutions for Internal Standard and Surrogates 7.5.1 A solution containing the internal standard and the surrogate compounds is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution), and to fortify each sample. Prepare a fortification solution containing fluorobenzene (internal standard), l,2-dichlorobenzene-d4 (surrogate), and BFB (surrogate) in methanol at concentrations of 5 |ig/mL of each (any appropriate concentration is acceptable). A 5 |iL aliquot of this solution added to a 25 mL water sample volume gives concentrations of 1 Hg/L of each. A 5 |iL aliquot of this solution added to a 5 mL water sample volume gives a concentration of 5 Hg/L of each. Additional internal standards and surrogate analytes are optional. Additional surrogate compounds should be similar in physical and chemical characteristics to the analytes of concern. 7.6 Preparation of Laboratory Reagent Blank (LRB) — Fill a 25 mL (or 5 mL) syringe with reagent water and adjust to the mark (no air bubbles). Inject an appropriate volume of the fortification solution containing the internal standard and surrogates through the Luer Lok valve into the reagent water. Transfer the LRB to the purging device. See Section 11.1.2. 524.2-12 ------- 7.7 Preparation of Laboratory Fortified Blank — Prepare this exactly like a calibration standard (Section 7.8). This is a calibration standard that is treated as a sample. 7.8 Preparation of Calibration Standards 7.8.1 The number of calibration solutions (CALs) needed depends on the calibration range desired. A minimum of three CAL solutions is required to calibrate a range of a factor of 20 in concentration. For a factor of 50, use at least four standards, and for a factor of 100 at least five standards. One calibration standard should contain each analyte of concern at a concentration of 2-10 times the method detection limit (Tables 4, 5, and 7) for that compound. The other CAL standards should contain each analyte of concern at concentrations that define the range of the method. Every CAL solution contains the internal standard and the surrogate compounds at the same concentration (5 Hg/L suggested for a 5 mL sample; 1 Hg/L for a 25 mL sample). 7.8.2 To prepare a calibration standard, add an appropriate volume of a primary dilution standard containing all analytes of concern to an aliquot of acidified (pH 2) reagent water in a volumetric flask. Also add an appropriate volume of internal standard and surrogate compound solution from Section 7.5.1. Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask. Remove the needle as quickly as possible after injection. Mix by inverting the flask three times only. Discard the contents contained in the neck of the flask. Aqueous standards are not stable in a volumetric flask and should be discarded after one hour unless transferred to a sample bottle and sealed immediately. SAMPLE COLLECTION. PRESERVATION. AND STORAGE 8.1 Sample Collection, Dechlorination, and Preservation 8.1.1 Collect all samples in duplicate. If samples, such as finished drinking water or waste water, are suspected to contain residual chlorine, add about 25 mg of ascorbic acid per 40 mL of sample to the sample bottle before filling. If the residual chlorine is likely to be present >5 mg/L, a determination of the amount of the chlorine may be necessary. Diethyl-p- phenylenediamine (DPD) test kits are commercially available to determine residual chlorine in the field. Add an additional 25 mg of ascorbic acid per each 5 mg/L of residual chlorine. If compounds boiling below 25°C are not to be determined, sodium thiosulfate may be used to reduce the residual chlorine. Fill sample bottles to overflowing, but take care not to flush out the rapidly dissolving ascorbic acid. No air bubbles should pass through the sample as the bottle is filled, or be trapped in the sample when the bottle is sealed. Adjust the pH of the duplicate samples to <2 by carefully adding two drops of 1:1 HC1 for each 40 mL of sample. Seal the sample bottles, Teflon face down, and shake vigorously for one 524.2-13 ------- minute. Do not mix the ascorbic acid or sodium thiosulfate with the HC1 prior to sampling. 8.1.2 When sampling from a water tap, open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 minutes). Adjust the flow to about 500 mL/min and collect duplicate samples from the flowing stream. 8.1.3 When sampling from an open body of water, such as surface water, waste water, and possible leachate samples, partially fill a 1 qt wide-mouth bottle or 1 L beaker with sample from a representative area. Fill a 60 mL or a 120 mL sample vial with sample from the larger container, and adjust the pH of the sample to about 2 by adding 1 + 1 HC1 dropwise while stirring. Check the pH with narrow range (1.4-2.8) pH paper. Record the number of drops of acid necessary to adjust the pH to 2. To collect actual samples, refill the large container with fresh sample and pour sample into sample vials. Follow filling instructions in Section 8.1.1. Add the appropriate number of drops of 1 + 1 HC1 to each sample to adjust the pH to about 2. If samples are suspected to contain residual chlorine, add ascorbic acid or sodium thiosulfate according to Section 8.1.1. 8.1.4 The samples must be chilled to about 4°C when collected and maintained at that temperature until analysis. Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will arrive at the laboratory with a substantial amount of ice remaining in the cooler. 8.1.5 If a sample foams vigorously when HC1 is added, discard that sample. Collect a set of duplicate samples but do not acidify them. These samples must be flagged as "not acidified" and must be stored at 4°C or below. These samples must be analyzed within 24 hours of collection time. 8.2 Sample Storage 8.2.1 Store samples at <4°C until analysis. The sample storage area must be free of organic solvent vapors and direct or intense light. 8.2.2 Analyze all samples within 14 days of collection. Samples not analyzed within this period must be discarded and replaced. 8.3 Field Reagent Blanks (FRB) 8.3.1 Duplicate FRBs must be handled along with each sample set, which is composed of the samples collected from the same general sample site at approximately the same time. At the laboratory, fill field blank sample bottles with reagent water and sample preservatives, seal, and ship to the sampling site along with empty sample bottles and back to the laboratory with filled sample bottles. Wherever a set of samples is shipped and 524.2-14 ------- stored, it is accompanied by appropriate blanks. FRBs must remain hermetically sealed until analysis. 8.3.2 Use the same procedures used for samples to add ascorbic acid and HC1 to blanks (Section 8.1.1). The same batch of ascorbic acid and HC1 should be used for the field reagent blanks in the field. QUALITY CONTROL 9.1 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks, field reagent blanks, and laboratory fortified blanks. Each laboratory must maintain records to document the quality of the data generated. Additional quality control practices are recommended. 9.2 Initial demonstration of low system background. Before any samples are analyzed, it must be demonstrated that a laboratory reagent blank (LRB) is reasonably free of contamination that would prevent the determination of any analyte of concern. Sources of background contamination are glassware, purge gas, sorbents, and equipment. Background contamination must be reduced to an acceptable level before proceeding with the next section. In general, background from method analytes should be below the method detection limit. 9.3 Initial demonstration of laboratory accuracy and precision. Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concern at a concentration in the range of 0.2-5 Hg/L (see appropriate regulations and maximum contaminant levels for guidance on appropriate concentrations). 9.3.1 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water. If a quality control sample containing the method analytes is not available, a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used. Also add the appropriate amounts of internal standard and surrogate compounds. Analyze each replicate according to the procedures described in Section 11.0, and on a schedule that results in the analyses of all replicates over a period of several days. 9.3.2 Calculate the measured concentration of each analyte in each replicate, the mean concentration of each analyte in all replicates, and mean accuracy (as mean percentage of true value) for each analyte, and the precision (as relative standard deviation, RSD) of the measurements for each analyte. Calculate the MDL of each analyte using the equa-tion described in Section 13.23. 9.3.3 For each analyte, the mean accuracy, expressed as a percentage of the true value, should be 80-120% and the RSD should be <20%. Some analytes, particularly the early eluting gases and late eluting higher molecular weight compounds, are measured with less accuracy and precision than other analytes. The MDLs must be sufficient to detect analytes at the 524.2-15 ------- required levels according to the SDWA Regulations. If these criteria are not met for an analyte, take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed. 9.3.4 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time. Charting surrogate recoveries is an especially valuable activity because surrogates are present in every sample and the analytical results will form a significant record of data quality. 9.4 Monitor the integrated areas of the quantitation ions of the internal standards and surrogates (Table 1) in all samples, continuing calibration checks, and blanks. These should remain reasonably constant over time. An abrupt change may indicate a matrix effect or an instrument problem. If a cryogenic interface is utilized, it may indicate an inefficient transfer from the trap to the column. These samples must be reanalyzed or a laboratory fortified duplicate sample analyzed to test for matrix effect. A more gradual drift of more than 50% in any area is indicative of a loss in sensitivity, and the problem must be found and corrected. 9.5 Laboratory Reagent Blanks (LRB) — With each batch of samples processed as a group within a work shift, analyze a LRB to determine the background system contamination. A FRB (Section 9.7) may be used in place of a LRB. 9.6 With each batch of samples processed as a group within a work shift, analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in Section 9.3. If more than 20 samples are included in a batch, analyze one LFB for every 20 samples. Use the procedures described in Section 9.3.3 to evaluate the accuracy of the measurements, and to estimate whether the MDLs can be obtained. If acceptable accuracy and MDLs cannot be achieved, the problem must be located and corrected before further samples are analyzed. Add these results to the ongoing control charts to document data quality. 9.7 With each set of field samples a field reagent blank (FRB) should be analyzed. The results of these analyses will help define contamination resulting from field sampling and transportation activities. If the FRB shows unacceptable contamina- tion, a LRB must be measured to define the source of the impurities. 9.8 At least quarterly, replicate LFBs should be analyzed to determine the precision of the laboratory measurements. Add these results to the ongoing control charts to document data quality. 9.9 At least quarterly, analyze a quality control sample (QCS) from an external source. If measured analyte concentrations are not of acceptable accuracy, check the entire analytical procedure to locate and correct the problem source. 9.10 Sample matrix effects have not been observed when this method is used with distilled water, reagent water, drinking water, or ground water. Therefore, 524.2-16 ------- analysis of a laboratory fortified sample matrix (LFM) is not required unless the criteria in Section 9.4 are not met. If matrix effects are observed or suspected to be causing low recoveries, analyze a laboratory fortified matrix sample for that matrix. The sample results should be flagged and the LFM results should be reported with them. 9.11 Numerous other quality control measures are incorporated into other parts of this procedure, and serve to alert the analyst to potential problems. 10.0 CALIBRATION AND STANDARDIZATION 10.1 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and is required intermittently throughout sample analysis as dictated by results of continuing calibration checks. After initial calibration is successful, a continuing calibration check is required at the beginning of each eight hour period during which analyses are performed. Additional periodic calibration checks are good laboratory practice. 10.2 Initial Calibration 10.2.1 Calibrate the mass and abundance scales of the MS with calibration compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Section 10.2.2. 10.2.2 Introduce into the GC (either by purging a laboratory reagent blank or making a syringe injection) 25 ng or less of BFB and acquire mass spectra for m/z 35-260 at 70 eV (nominal). Use the purging procedure and/or GC conditions given in Section 11.0. If the spectrum does not meet all criteria in Table 3, the MS must be re tuned and adjusted to meet all criteria before proceeding with calibration. An average spectrum across the GC peak may be used to evaluate the performance of the system. 10.2.3 Purge a medium CAL solution, (e.g., 10-20 Hg/L) using the procedure given in Section 11.0. 10.2.4 Performance criteria for the medium calibration. Examine the stored GC/MS data with the data system software. Figures 3 and 4 shown acceptable total ion chromatograms. 10.2.4.1 GC performance — Good column performance will produce symmetrical peaks with minimum tailing for most com- pounds. If peaks are unusually broad, or if peaks are running together with little vallies between them, the wrong column has been selected or remedial action is probably necessary (Section 10.3.6). 10.2.4.2 MS sensitivity — The GC/MS/DS peak identification software should be able to recognize a GC peak in the appropriate retention time window for each of the com- pounds in calibration solution, and make correct tentative 524.2-17 ------- identifications. If fewer than 99% of the compounds are recognized, system maintenance is required. See Section 10.3.6. 10.2.5 If all performance criteria are met, purge an aliquot of each of the other CAL solutions using the same GC/MS conditions. 10.2.6 Calculate a response factor (RF) for each analyte and isomer pair for each CAL solution using the internal standard fluorobenzene. Table 1 contains suggested quantitation ions for all compounds. This calculation is supported in acceptable GC/MS data system software (Section 6.3.5), and many other software programs. RF is a unitless number, but units used to express quantities of analyte and internal standard must be equivalent. pr (A*> (QJ (Ais) (Qx) where: Ax = integrated abundance of the quantitation ion of the analyte. Ais = integrated abundance of the quantitation ion of the internal standard. Qx = quantity of analyte purged in nanograms or concentration units. Qis = quantity of internal standard purged in ng or concentration units. 10.2.6.1 For each analyte and surrogate, calculate the mean RF from analyses of CAL solutions. Calculate the standard deviation (SD) and the relative standard deviation (RSD) from each mean: RSD =100 (SD/M). If the RSD of any analyte or surrogate mean RF exceeds 20%, either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range, or take action to improve GC/MS performance Section 10.3.6). Surrogate compounds are present at the same concentration on every sample, calibration standard, and all types of blanks. 10.2.7 As an alternative to calculating mean response factors and applying the RSD test, use the GC/MS data system software or other available software to generate a linear or second order regression calibration curve. 10.3 Continuing Calibration Check — Verify the MS tune and initial calibration at the beginning of each eight-hour work shift during which analyses are performed using the following procedure. 10.3.1 Introduce into the GC (either by purging a laboratory reagent blank or making a syringe injection) 25 ng or less of BFB and acquire a mass spectrum that includes data for m/z 35-260. If the spectrum does not 524.2-18 ------- meet all criteria (Table 3), the MS must be retuned and adjusted to meet all criteria before proceeding with the continuing calibration check. 10.3.2 Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration. 10.3.3 Demonstrate acceptable performance for the criteria shown in Section 10.2.4. 10.3.4 Determine that the absolute areas of the quantitation ions of the internal standard and surrogates have not decreased by more than 30% from the areas measured in the most recent continuing calibration check, or by more than 50% from the areas measured during initial calibration. If these areas have decreased by more than these amounts, adjustments must be made to restore system sensitivity. These adjustments may require cleaning of the MS ion source, or other maintenance as indicated in Section 10.3.6, and recalibration. Control charts are useful aids in documenting system sensitivity changes. 10.3.5 Calculate the RF for each analyte of concern and surrogate compound from the data measured in the continuing calibration check. The RF for each analyte and surrogate must be within 30% of the mean value measured in the initial calibration. Alternatively, if a linear or second order regression is used, the concentration measured using the calibration curve must be within 30% of the true value of the concentration in the medium calibration solution. If these conditions do not exist, remedial action must be taken which may require recalibration. 10.3.6 Some possible remedial actions. Major maintenance such as cleaning an ion source, cleaning quadrupole rods, etc. require returning to the initial calibration step. 10.3.6.1 Check and adjust GC and/or MS operating conditions; check the MS resolution, and calibrate the mass scale. 10.3.6.2 Clean or replace the splitless injection liner; silanize a new injection liner. This applies only if the injection liner is an integral part of the system. 10.3.6.3 Flush the GC column with solvent according to manufacturer's instructions. 10.3.6.4 Break off a short portion (about one meter) of the column from the end near the injector; or replace GC column. This action will cause a slight change in retention times. Analyst may need to redefine retention windows. 10.3.6.5 Prepare fresh CAL solutions, and repeat the initial calibration step. 524.2-19 ------- 10.3.6.6 Clean the MS ion source and rods (if a quadrupole). 10.3.6.7 Replace any components that allow analytes to come into contact with hot metal surfaces. 10.3.6.8 Replace the MS electron multiplier, or any other faulty components. 10.3.6.9 Replace the trap, especially when only a few compounds fail the criteria in Section 10.3.5 while the majority are determined successfully. Also check for gas leaks in the purge and trap unit as well as the rest of the analytical system. 10.4 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps. 10.4.1 Fill the purging device with 25.0 mL (or 5 mL) of reagent water or aqueous calibration standard. 10.4.2 Start to purge the aqueous mixture. Inject a known volume (between 100 and 2000 pL) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe. Slowly inject the gaseous sample through a septum seal at the top of the purging device at 2000 |iL/min. If the injection of the standard is made through the aqueous sample inlet port, flush the dead volume with several mL of room air or carrier gas. Inject the gaseous standard before 5 min of the 11-minute purge time have elapsed. 10.4.3 Determine the aqueous equivalent concentration of vinyl chloride standard, in Hg/L, injected with the equation: S = 0.102 (C) (V) Aqueous equivalent concentration of vinyl chloride standard in iig/L. Concentration of gaseous standard in mg/L (v/v). Volume of standard injected in mL. 11.0 PROCEDURE where: S = C = V = 11.1 Sample Introduction and Purging 11.1.1 This method is designed for a 25 mL sample volume, but a smaller (5 mL) sample volume is recommended if the GC/MS system has adequate sensitivity to achieve the required method detection limits. Adjust the helium purge gas flow rate to 40 mL/min. Attach the trap inlet to the purging device and open the syringe valve on the purging device. 524.2-20 ------- 11.1.2 Remove the plungers from two 25 mL (or 5 mL depending on sample size) syringes and attach a closed syringe valve to each. Warm the sample to room temperature, open the sample bottle, and carefully pour the sample into one of the syringe barrels to just short of overflowing. Replace the syringe plunger, invert the syringe, and compress the sample. Open the syringe valve and vent any residual air while adjusting the sample volume to 25.0 mL (or 5 mL). To all samples, blanks, and calibration standards, add 5 |iL (or an appropriate volume) of the fortification solution containing the internal standard and the surrogates to the sample through the syringe valve. Close the valve. Fill the second syringe in an identical manner from the same sample bottle. Reserve this second syringe for a reanalysis if necessary. 11.1.3 Attach the sample syringe valve to the syringe valve on the purging device. Be sure that the trap is cooler than 25°C, then open the sample syringe valve and inject the sample into the purging chamber. Close both valves and initiate purging. Purge the sample for 11.0 minutes at ambient temperature. 11.1.4 Standards and samples must be analyzed in exactly the same manner. Room temperature changes in excess of 10°F may adversely affect the accuracy and precision of the method. 11.2 Sample Desorption 11.2.1 Non-cryogenic interface — After the 11-minute purge, place the purge and trap system in the desorb mode and preheat the trap to 180°C without a flow of desorption gas. Then simultaneously start the flow of desorption gas at a flow rate suitable for the column being used (optimum desorb flow rate is 15 mL/min) for about four minutes, begin the GC temperature program, and start data acquisition. 11.2.2 Cryogenic interface - After the 11-minute purge, place the purge and trap system in the desorb mode, make sure the cryogenic interface is a -150°C or lower, and rapidly heat the trap to 180°C while backflushing with an inert gas at 4 mL/min for about five minutes. At the end of the five minute desorption cycle, rapidly heat the cryogenic trap to 250°C, and simultaneously begin the temperature program of the gas chromatograph, and start data acquisition. 11.2.3 While the trapped components are being introduced into the gas chromatograph (or cryogenic interface), empty the purging device using the sample syringe and wash the chamber with two 25 mL flushes of reagent water. After the purging device has been emptied, leave syringe valve open to allow the purge gas to vent through the sample introduction needle. 11.3 Gas Chromatography/Mass Spectrometry — Acquire and store data over the nominal mass range 35-260 with a total cycle time (including scan overhead time) 524.2-21 ------- of two seconds or less. If water, methanol, or carbon dioxide cause a background problem, start at 47 or 48 m/z. If ketones are to be determined, data must be acquired starting at m/z 43. Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak. Suggested temperature programs are provided below. Alternative temperature programs can be used. 11.3.1 Single ramp linear temperature program for wide-bore Columns 1 and 2 with a jet separator. Adjust the helium carrier gas flow rate to within the capacity of the separator, or about 15 mL/min. The column temperature is reduced 10°C and held for five minutes from the beginning of desorption, then programmed to 160°C at 6°C/min, and held until all components have eluted. 11.3.2 Multi-ramp temperature program for wide-bore Column 2 with the open split interface. Adjust the helium carrier gas flow rate to about 4.6 mL/min. The column temperature is reduced to 10°C and held for six minutes from the beginning of desorption, then heated to 70°C at 10°/min, heated to 120°C at 5°/min, heated to 180°C at 8°/min, and held at 180°C until all compounds have eluted. 11.3.3 Single ramp linear temperature program for narrow-bore Column 3 with a cryogenic interface. Adjust the helium carrier gas flow rate to about 4 mL/min. The column temperature is reduced to 10°C and held for five minutes from the beginning of vaporization from the cryogenic trap, programmed at 6°/min for 10 minutes, then 15°/min for five minutes to 145°C, and held until all components have eluted. 11.3.4 Multi-ramp temperature program for wide-bore Column 4 with the open split interface. Adjust the helium carrier gas flow rate to about 7.0 mL/min. The column temperature is -10°C and held for six minutes from beginning of desorption, then heated to 100°C at 10°C/min, heated to 200°C at 5°C/min and held at 200°C for eight minutes or until all compounds of interest had eluted. 11.4 Trap Reconditioning — After desorbing the sample for four minutes, recondition the trap by returning the purge and trap system to the purge mode. Wait 15 seconds, then close the syringe valve on the purging device to begin gas flow through the trap. Maintain the trap temperature at 180°C. Maintain the moisture control module, if utilized, at 90°C to remove residual water. After approximately seven minutes, turn off the trap heater and open the syringe valve to stop the gas flow through the trap. When the trap is cool, the next sample can be analyzed. 11.5 Termination of Data Acquisition — When all the sample components have eluted from the GC, terminate MS data acquisition. Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time. If any ion abundance exceeds the system working range, dilute the sample aliquot in the second syringe with reagent water and analyze the diluted aliquot. 524.2-22 ------- 11.6 Identification of Analytes — Identify a sample component by comparison of its mass spectrum (after background subtraction) to a reference spectrum in the user-created data base. The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture. 11.6.1 In general, all ions that are present above 10% relative abundance in the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20%. For example, if an ion has a relative abundance of 30% in the standard spectrum, its abundance in the sample spectrum should be in the range of 10-50%. Some ions, particularly the molecular ion, are of special importance, and should be evaluated even if they are below 10% relative abundance. 11.6.2 Identification requires expert judgment when sample components are not resolved chromatographically and produce mass spectra containing ions contributed by more than one analyte. When GC peaks obviously represent more than one sample component (i.e., broadened peak with shoulder(s) or valley between two or more maxima), appropriate analyte spectra and background spectra can be selected by examining plots of characteristic ions for tentatively identified components. When analytes coelute (i.e., only one GC peak is apparent), the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound. Because purgeable organic compounds are relatively small molecules and produce comparatively simple mass spectra, this is not a significant problem for most method analytes. 11.6.3 Structural isomers that produce very similar mass spectra can be explicitly identified only if they have sufficiently different GC retention times. Acceptable resolution is achieved if the height of the valley between two peaks is less than 25% of the average height of the two peaks. Otherwise, structural isomers are identified as isomeric pairs. Two of the three isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary columns. If unresolved, these groups of isomers must be reported as isomeric pairs. 11.6.4 Methylene chloride, acetone, carbon disulfide, and other background components appear in variable quantities in laboratory and field reagent blanks, and generally cannot be accurately measured. Subtraction of the concentration in the blank from the concentration in the sample is not acceptable because the concentration of the background in the blank is highly variable. 12.0 DATA ANALYSIS AND CALCULATIONS 12.1 Complete chromatographic resolution is not necessary for accurate and precise measurements of analyte concentrations if unique ions with adequate intensities are available for quantitation. 524.2-23 ------- 12.1.1 Calculate analyte and surrogate concentrations. (Ax) (Qie) 1000 (A J RF V where: Cx = concentration of analyte or surrogate in Hg/L in the water sample. Ax = integrated abundance of the quantitation ion of the analyte in the sample. Ais = integrated abundance of the quantitation ion of the internal standard in the sample. Qis = total quantity (in micrograms) of internal standard added to the water sample. V = original water sample volume in mL. RF = mean response factor of analyte from the initial calibration. 12.1.2 Alternatively, use the GC/MS system software or other available proven software to compute the concentrations of the analytes and surrogates from the linear or second order regression curves. 12.1.3 Calculations should utilize all available digits of precision, but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty). Experience indicates that three significant figures may be used for concentrations above 99 Hg/L, two significant figures for concentrations between 1-99 Hg/L, and one significant figure for lower concentrations. 12.1.4 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations. 13.0 METHOD PERFORMANCE 13.1 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 Hg/L. Results were obtained using the four columns specified (Section 6.3.2.1) and the open split or jet separator (Section 6.3.3.1), or the cryogenic interface (Section 6.3.3.2). These data are shown in Tables 4-8. 13.2 With these data, method detection limits were calculated using the formula3: MDL = S t(n_u_alpha = o 99) where: t(n l l alpha = 0 99) = Student's t value for the 99% confidence level with n-1 degrees of freedom, n = Number of replicates. S = The standard deviation of the replicate analyses. 14.0 POLLUTION PREVENTION 524.2-24 ------- 14.1 No solvents are utilized in this method except the extremely small volumes of methanol needed to make calibration standards. The only other chemicals used in this method are the neat materials in preparing standards and sample preservatives. All are used in extremely small amounts and pose no threat to the environment. 15.0 WASTE MANAGEMENT 15.1 There are no waste management issues involved with this method. Due to the nature of this method, the discarded samples are chemically less contaminated than when they were collected. 16.0 REFERENCES 1. Alford-Stevens, A., Eichelberger, J.W., and Budde, W.L. "Purgeable Organic Compounds in Water by Gas Chromatography/Mass Spectrometry, Method 524." Environmental Monitoring and Support Laboratory, U.S. Environmental Protection Agency, Cincinnati, Ohio, February 1983. 2. Madding, C. "Volatile Organic Compounds in Water by Purge and Trap Capillary Column GC/MS," Proceedings of the Water Quality Technology Conference, American Water Works Association, Denver, CO, December 1984. 3. Glaser, J.A., Foerst, D.L., McKee, G.D., Quave, S.A., and Budde, W.L. "Trace Analyses for Wastewaters," Environ. Sci. Technol., JJ3, 1426, 1981. 4. "Carcinogens-Working with Carcinogens," Department of Health, Education, and Welfare, Public Health Service, Center for Disease Control, National Institute for Occupational Safety and Health, Publication No. 77-206, August 1977. 5. "OSHA Safety and Health Standards, General Industry," (29 CFR 1910), Occupational Safety and Health Administration, OSHA 2206, (Revised, January 1976). 6. "Safety in Academic Chemistry Laboratories," American Chemical Society Publication, Committee on Chemical Safety, 3rd Edition, 1979. 7. Arrendale, R.F., Severson, R.F., and Chortyk, O.T. "Open Split Interface for Capillary Gas Chromatography/Mass Spectrometry," Anal. Chem. 1984, 56, 1533. 8. Flesch, J.J. and Fair, P.S. "The Analysis of Cyanogen Chloride in Drinking Water," Proceedings of Water Quality Technology Conference, American Water Works Association, St. Louis, MO., November 14-16, 1988. 524.2-25 ------- 17.0 TABLES. DIAGRAMS. FLOWCHARTS. AND VALIDATION DATA TABLE 1. MOLECULAR WEIGHTS AND QUANTITATION IONS FOR METHOD ANALYTES Primary Secondary Quantitation Quantitation Compound MWa Ion Ions Internal standard Fluorobenzene 96 96 77 Surrogates 4-Bromofluorobenzene 174 95 174, 176 1,2-Dichlorobenzene-d4 150 152 115, 150 Target Analvtes Acetone 58 43 58 Acrylonitrile 53 52 53 Allyl chloride 76 76 49 Benzene 78 78 77 Bromobenzene 156 156 77, 158 Bromochloromethane 128 128 49, 130 Bromodichloromethane 162 83 85, 127 Bromoform 250 173 175, 252 Bromomethane 94 94 96 2-Butanone 72 43 57, 72 n-Butylbenzene 134 91 134 sec-Butylbenzene 134 105 134 tert-Butylbenzene 134 119 91 Carbon disulfide 76 76 - Carbon tetrachloride 152 117 119 Chloroacetonitrile 75 48 75 Chlorobenzene 112 112 77, 114 1-Chlorobutane 92 56 49 Chloroethane 64 64 66 Chloroform 118 83 85 Chloromethane 50 50 52 2-Chlorotoluene 126 91 126 4-Chlorotoluene 126 91 126 Dibromochloromethane 206 129 127 1,2-Dibromo-3-Chloropropane 234 75 155, 157 1,2-Dibromoethane 186 107 109, 188 Dibromomethane 172 93 95, 174 1,2-Dichlorobenzene 146 146 111, 148 1,3-Dichlorobenzene 146 146 111, 148 524.2-26 ------- TABLE 1. MOLECULAR WEIGHTS AND QUANTITATION IONS FOR METHOD ANALYTES Primary Secondary Quantitation Quantitation Compound MWa Ion Ions 1,4-Dichlorobenzene 146 146 111, 148 trans-1,4-Dichloro-2-butene 124 53 88, 75 Dichlorodifluoromethane 120 85 87 1,1 -Dichloroethane 98 63 65, 83 1,2-Dichloroethane 98 62 98 1,1 -Dichloroethene 96 96 61, 63 cis-1,2-Dichloroethene 96 96 61, 98 trans-1,2-Dichloroethene 96 96 61, 98 1,2-Dichloropropane 112 63 112 1,3-Dichloropropane 112 76 78 2,2-Dichloropropane 112 77 97 1,1 -Dichloropropene 110 75 110, 77 1,1 -Dichloropropanone 126 43 83 cis-1,3-dichloropropene 110 75 110 trans-1,3-dichloropropene 110 75 110 Diethyl ether 74 59 45, 73 Ethylbenzene 106 91 106 Ethyl methacrylate 114 69 99 Hexachlorobutadiene 258 225 260 Hexachloroethane 234 117 119, 201 2-Hexanone 100 43 58 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 134, 91 Methacrylonitrile 67 67 52 Methyl acrylate 86 55 85 Methylene chloride 84 84 86, 49 Methyl iodide 142 142 127 Methylmethacrylate 100 69 99 4-Methyl-2-pentanone 100 43 58, 85 Methyl-t-butyl ether 88 73 57 Naphthalene 128 128 - Nitrobenzene 123 51 77 2-Nitropropane 89 46 - Pentachloroethane 200 117 119, 167 Propionitrile 55 54 - n-Propylbenzene 120 91 120 Styrene 104 104 78 1,1,1,2-T etrachloroethane 166 131 133, 119 1,1,2,2-T etrachloroethane 166 83 131, 85 T etrachlor oethene 164 166 168, 129 T etrahydrofuran 72 71 72, 42 Toluene 92 92 91 524.2-27 ------- TABLE 1. MOLECULAR WEIGHTS AND QUANTITATION IONS FOR METHOD ANALYTES Primary Secondary Quantitation Quantitation Compound MWa Ion Ions 1,2,3-Trichlorobenzene 180 180 182 1,2,4-Trichlorobenzene 180 180 182 1,1,1 -T richloroethane 132 97 99, 61 1,1,2-T richloroethane 132 83 97, 85 Trichloroethene 130 95 130, 132 T richlorofluoromethane 136 101 103 1,2,3-Trichloropropane 146 75 77 1,2,4-Trimethylbenzene 120 105 120 1,3,5-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64 o-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91 aMonoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses. 524.2-28 ------- TABLE 2. CHROMATOGRAPHIC RETENTION TIMES FOR METHOD ANALYTES ON THREE COLUMNS WITH FOUR SETS OF CONDITIONS3 Compound Column lb Retention Time (min:sec) Column 2b Column 2C Column 3d Column 4e Internal standard Fluorobenzene 8:49 6:27 14:06 8:03 22:00 Surrogates 4-Bromofluorobenzene 18:38 15:43 23:38 31:21 1,2-Dichlorobenzene-d4 22:16 19:08 27:25 35:51 Target Analvtes Acetone 16:14 Acrylonitrile 17:49 Allyl chloride 16:58 Benzene 8:14 5:40 13:30 7:25 21:32 Bromobenzene 18:57 15:52 24:00 16:25 31:52 Bromochloromethane 6:44 4:23 12:22 5:38 20:20 Bromodichloromethane 10:35 8:29 15:48 9:20 23:36 Bromoform 17:56 14:53 22:46 15:42 30:32 Bromomethane 2:01 0:58 4:48 1:17 12:26 2-Butanone 19:41 n-Butylbenzene 22:13 19:29 27:32 17:57 35:41 sec-Butylbenzene 20:47 18:05 26:08 17:28 34:04 tert-Butylbenzene 20:17 17:34 25:36 17:19 33:26 Carbon Disulfide 16:30 Carbon Tetrachloride 7:37 5:16 13:10 7:25 21:11 Chloroacetonitrile 23:51 Chlorobenzene 15:46 13:01 20:40 14:20 28:26 1-Chlorobutane 21:00 Chloroethane 2:05 1:01 1:27 Chloroform 6:24 4:48 12:36 5:33 20:27 Chloromethane 1:38 0:44 3:24 0:58 9:11 2-Chlorotoluene 19:20 16:25 24:32 16:44 32:21 4-Chlorotoluene 19:30 16:43 24:46 16:49 32:38 Cyanogen chloride8 1:03 Dibromochloromethane 14:23 11:51 19:12 12:48 26:57 1,2-Dibromo-3-Chloropropan 24:32 21:05 18:02 38:20 G 1,2-Dibromoethane 14:44 11:50 19:24 13:36 27:19 Dibromomethane 10:39 7:56 15:26 9:05 23:22 1,2-Dichlorobenzene 22:31 19:10 27:26 17:47 35:55 1,3-Dichlorobenzene 21:13 18:08 26:22 17:28 34:31 1,4-Dichlorobenzene 21:33 18:23 26:36 17:38 34:45 524.2-29 ------- TABLE 2. CHROMATOGRAPHIC RETENTION TIMES FOR METHOD ANALYTES ON THREE COLUMNS WITH FOUR SETS OF CONDITIONS3 Retention Time (min:secr Compound Column lb Column 2b Column 2C Column 3d Column 4e t-1,4-Dichloro-2-butene 31:44 Dichlorodifluoromethane 1:33 0:42 3:08 0:53 7:16 1,1 -Dichloroethane 4:51 2:56 10:48 4:02 18:46 1,2-Dichloroethane 8:24 5:50 13:38 7:00 21:31 1,1 -Dichloroethene 2:53 1:34 7:50 2:20 16:01 cis-1,2-Dichloroethene 6:11 3:54 11:56 5:04 19:53 trans-1,2-Dichloroethene 3:59 2:22 9:54 3:32 17:54 1,2-Dichloropropane 10:05 7:40 15:12 8:56 23:08 1,3-Dichloropropane 14:02 11:19 18:42 12:29 26:23 2,2-Dichloropropane 6:01 3:48 11:52 5:19 19:54 1,1 -Dichloropropanone 24:52 1,1 -Dichloropropene 7:49 5:17 13:06 7:10 21:08 cis-1,3-dichloropropene 11.58 16:42 24:24 trans-1,3-dichloropropene 13.46 17:54 25:33 Diethyl ether 15:31 Ethylbenzene 15:59 13:23 21:00 14:44 28:37 Ethyl Methacrylate 25:35 Hexachlorobutadiene 26:59 23:41 32:04 19:14 42:03 Hexachloroethane 36:45 Hexanone 26:23 Isopropylbenzene 18:04 15:28 23:18 16:25 30:52 4-Isopropyltoluene 21:12 18:31 26:30 17:38 34:27 Methacrylonitrile 20:15 Methylacrylate 20:02 Methylene Chloride 3:36 2:04 9:16 2:40 17:18 Methyl Iodide 16:21 Methylmethacrylate 23:08 4-Methyl-2-pentanone 24:38 Methyl-t-butyl ether 17:56 Naphthalene 27:10 23:31 32:12 19:04 42:29 Nitrobenzene 39:02 2-Nitropropane 23:58 Pentachloroethane 33:33 Propionitrile 19:58 n-Propylbenzene 19:04 16:25 24:20 16:49 32:00 Styrene 17:19 14:36 22:24 15:47 29:57 1,1,1,2-T etrachloroethane 15:56 13:20 20:52 14:44 28:35 1,1,2,2-T etrachloroethane 18:43 16:21 24:04 15:47 31:35 T etrachlor oethene 13:44 11:09 18:36 13:12 26:27 T etrahydrofuran 20:26 Toluene 12:26 10:00 17:24 11:31 25:13 1,2,3-Trichlorobenzene 27:47 24:11 32:58 19:14 43:31 1,2,4-Trichlorobenzene 26:33 23:05 31:30 18:50 41:26 524.2-30 ------- TABLE 2. CHROMATOGRAPHIC RETENTION TIMES FOR METHOD ANALYTES ON THREE COLUMNS WITH FOUR SETS OF CONDITIONS3 Compound Column lb Retention Time (min:sec) Column 2b Column 2C Column 3d Column 4e 1,1,1 -Trichloroethane 7:16 4:50 12:50 6:46 20:51 1,1,2-T richloroethane 13:25 11:03 18:18 11:59 25:59 Trichloroethene 9:35 7:16 14:48 9:01 22:42 T richlorofluoromethane 2:16 1:11 6:12 1:46 14:18 1,2,3-Trichloropropane 19:01 16:14 24:08 16:16 31:47 1,2,4-Trimethylbenzene 20:20 17:42 31:30 17:19 33:33 1,3,5-Trimethylbenzene 19:28 16:54 24:50 16:59 32:26 Vinyl chloride 1:43 0:47 3:56 1:02 10:22 o-Xylene 17:07 14:31 22:16 15:47 29:56 m-Xylene 16:10 13:41 21:22 15:18 28:53 p-Xylene 16:07 13:41 21:18 15:18 28:53 aColumns 1-4 are those given in Section 6.3.2.1; retention times were measured from the beginning of thermal desorption from the trap (Columns 1-2, and 4) or from the beginning of thermal release from the cryogenic interface (Column 3). bGC conditions given in Section 11.3.1. CGC conditions given in Section 11.3.2. dGC conditions given in Section 11.3.3. eGC conditions given in Section 11.3.4. TABLE 3. ION ABUNDANCE CRITERIA FOR 4-BROMOFLUOROBENZENE (BFB) Mass (M/z) Relative Abundance Criteria 50 15-40% of mass 95 75 30-80% of mass 95 95 Base Peak, 100% Relative Abundance 96 5-9% of mass 95 173 <2% of mass 174 174 >50% of mass 95 175 5-9% of mass 174 176 >95% but <101% of mass 174 177 5-9% of mass 176 524.2-31 ------- TABLE 4. ACCURACY AND PRECISION DATA FROM 16-31 DETERMINATIONS OF THE METHOD ANALYTES IN REAGENT WATER USING WIDE-BORE CAPILLARY COLUMN la Compound True Cone. Range (Pg/L) Mean Accuracy (% of True Value) HeT Std. Dev. Method Det. Limitb (Pg/L) Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Dibromochloromethane l,2-Dibromo-3-chloropropane 1,2-Dibromoethane Dibromomethane 1.2-Dichlorobenzene 1.3-Dichlorobenzene 1.4-Dichlorobenzene Dichlorodifluoromethane 1,1 -Dichloroethane 1,2-Dichloroethane 1,1 -Dichloroethene cis-1,2 Dichloroethene trans-1,2-Dichloroethene 1.2-Dichloropropane 1.3-Dichloropropane 2,2-Dichloropropane 1,1 -Dichloropropene cis-1,2-Dichloropropene trans-1,2-Dichloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride 0.1-10 0.1-10 0.5-10 0.1-10 0.5-10 0.5-10 0.5-10 0.5-10 0.5-10 0.5-10 0.1-10 0.5-10 0.5-10 0.5-10 0.1-10 0.1-10 0.1-10 0.5-10 0.5-10 0.5-10 0.1-10 0.5-10 0.2-20 0.5-10 0.5-10 0.1-10 0.1-10 0.5-10 0.1-10 0.1-10 0.1-10 0.5-10 0.5-10 0.1-10 0.5-10 0.5-10 0.1-10 0.1-10 97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98 99 100 101 99 95 5.7 5.5 6.4 6.1 6.3 8.2 7.6 7.6 7.3 8.8 5.9 9.0 6.1 8.9 6.2 8.3 7.0 19.9 3.9 5.6 6.2 6.9 6.4 7.7 5.3 5.4 6.7 6.7 5.6 6.1 6.0 16.9 8.9 8.6 6.8 7.6 6.7 5.3 0.04 0.03 0.04 0.08 0.12 0.11 0.11 0.13 0.14 0.21 0.04 0.10 0.03 0.13 0.04 0.06 0.05 0.26 0.06 0.24 0.03 0.12 0.03 0.10 0.04 0.06 0.12 0.12 0.06 0.04 0.04 0.35 0.10 0.06 0.11 0.15 0.12 0.03 524.2-32 ------- TABLE 4. ACCURACY AND PRECISION DATA FROM 16-31 DETERMINATIONS OF THE METHOD ANALYTES IN REAGENT WATER USING WIDE-BORE CAPILLARY COLUMN la True Mean Rel. Method Cone. Accuracy Std. Det. Range (% of True Dev. Limitb Compound (Pg/L) Value) (%) (l-ig/L) Naphthalene 0.1-100 104 8.2 0.04 n-Propylbenzene 0.1-10 100 5.8 0.04 Styrene 0.1-100 102 7.2 0.04 1,1,1,2-T etrachloroethane 0.5-10 90 6.8 0.05 1,1,2,2-T etrachloroethane 0.1-10 91 6.3 0.04 T etrachlor oethene 0.5-10 89 6.8 0.14 Toluene 0.5-10 102 8.0 0.11 1,2,3-Trichlorobenzene 0.5-10 109 8.6 0.03 1,2,4-Trichlorobenzene 0.5-10 108 8.3 0.04 1,1,1 -T richloroethane 0.5-10 98 8.1 0.08 1,1,2-T richloroethane 0.5-10 104 7.3 0.10 Trichloroethene 0.5-10 90 7.3 0.19 T richlorofluoromethane 0.5-10 89 8.1 0.08 1,2,3-Trichloropropane 0.5-10 108 14.4 0.32 1,2,4-Trimethylbenzene 0.5-10 99 8.1 0.13 1,3,5-Trimethylbenzene 0.5-10 92 7.4 0.05 Vinyl chloride 0.5-10 98 6.7 0.17 o-Xylene 0.1-31 103 7.2 0.11 m-Xylene 0.1-10 97 6.5 0.05 p-Xylene 0.5-10 104 7.7 0.13 aData obtained by using Column 1 with a jet separator interface and a quadrupole mass spectrometer (Section 11.3.1) with analytes divided among three solutions. bReplicate samples at the lowest concentration listed in Column 2 of this table were analyzed. These results were used to calculate MDLs. 524.2-33 ------- TABLE 5. ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF METHOD ANALYTES IN REAGENT WATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW-BORE CAPILLARY COLUMN 3a Compound True Cone. (l-ig/L) Mean Accuracy (% of True Value) HeT Std. Dev. Method Dect. Limit (Pg/L) Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride5 Dibromochloromethane l,2-Dibromo-3-chloropropane 1,2-Dibromoethane Dibromomethane 1.2-Dichlorobenzene 1.3-Dichlorobenzene 1.4-Dichlorobenzene Dichlorodifluoromethane 1,1 -Dichloroethane 1,2-Dichloroethane 1,1 -Dichloroethene cis-1,2 Dichloroethene trans-1,2-Dichloroethene 1.2-Dichloropropane 1.3-Dichloropropane 2,2-Dichloropropane 1,1 -Dichloropropene cis-1,3-Dichloropropene trans-1,3-Dichloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene 0.1 0.5 0.5 0.1 0.1 0.1 0.5 0.5 0.5 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.5 0.5 99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98 99 100 98 87 6.2 7.4 5.8 4.6 5.4 7.1 6.0 7.1 2.5 6.8 5.8 5.8 3.2 4.7 4.6 7.0 10.6 5.6 10.0 5.6 6.9 3.5 6.0 5.7 8.8 6.2 6.3 9.0 3.7 7.2 6.0 5.8 4.9 7.4 5.2 6.7 6.4 13.0 0.03 0.11 0.07 0.03 0.20 0.06 0.03 0.12 0.33 0.08 0.03 0.02 0.02 0.05 0.05 0.05 0.30 0.07 0.05 0.02 0.03 0.05 0.05 0.04 0.11 0.03 0.02 0.05 0.06 0.03 0.02 0.04 0.05 0.02 0.03 0.04 0.10 0.26 524.2-34 ------- TABLE 5. ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF METHOD ANALYTES IN REAGENT WATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW-BORE CAPILLARY COLUMN 3a Mean Rel. Method True Accuracy Std. Dect. Cone. (% of True Dev. Limit Compound (Pg/L) Value) (%) (Pg/L) Methylene chloride 0.5 97 13.0 0.09 Naphthalene 0.1 98 7.2 0.04 n-Propylbenzene 0.1 99 6.6 0.06 Styrene 0.1 96 19.0 0.06 1,1,1,2-T etrachloroethane 0.1 100 4.7 0.04 1,1,2,2-T etrachloroethane 0.5 100 12.0 0.20 T etrachlor oethene 0.1 96 5.0 0.05 Toluene 0.1 100 5.9 0.08 1,2,3-Trichlorobenzene 0.1 98 8.9 0.04 1,2,4-Trichlorobenzene 0.1 91 16.0 0.20 1,1,1 -T richloroethane 0.1 100 4.0 0.04 1,1,2-Trichloroethane 0.1 98 4.9 0.03 Trichloroethene 0.1 96 2.0 0.02 T richlorofluoromethane 0.1 97 4.6 0.07 1,2,3-Trichloropropane 0.1 96 6.5 0.03 1,2,4-Trimethylbenzene 0.1 96 6.5 0.04 1,3,5-Trimethylbenzene 0.1 99 4.2 0.02 Vinyl chloride 0.1 96 0.2 0.04 o-Xylene 0.1 94 7.5 0.06 m-Xylene 0.1 94 4.6 0.03 p-Xylene 0.1 97 6.1 0.06 Data obtained by using Column 3 with a cryogenic interface and a quadrupoie mass spectrometer (Section 11.3.3). bReference 8. 524.2-35 ------- TABLE 6. ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF THE METHOD ANALYTES IN REAGENT WATER USING WIDE-BORE CAPILLARY COLUMN 2a Compound Mean Accuracy (% of True Value, 2 pg/L No.b Cone.) RSD Mean Accuracy (% of True Value, 2 pg/L Cone.) RSD Internal Standard Fluorobenzene Surrogates 4-Bromofluorobenzene 2 98 l,2-Dichlorobenzene-d4 3 97 Target Analvtes Benzene 37 97 Bromobenzene 38 102 Bromochloromethane 4 99 Bromodichloromethane 5 96 Bromoform 6 89 Bromomethane 7 55 n-Butylbenzene 39 89 sec-Butylbenzene 40 102 tert-Butylbenzene 41 101 Carbon tetrachloride 8 84 Chlorobenzene 42 104 Chloroethanec Chloroform 9 97 Chloromethane 10 110 2-Chlorotoluene 43 91 4-Chlorotoluene 44 89 Dibromochloromethane 11 95 l,2-Dibromo-3-chloropropanec l,2-Dibromoethanec Dibromomethane 13 99 1.2-Dichlorobenzene 45 93 1.3-Dichlorobenzene 46 100 1.4-Dichlorobenzene 47 98 Dichlorodifluoromethane 14 38 1.1-Dichloroethane 15 97 1.2-Dichloroethane 16 102 1,1-Dichloroethene 17 90 cis-l,2-Dichloroethene 18 100 trans-1,2-Dichloroethene 19 92 1.8 3.2 4.4 3.0 5.2 1.8 2.4 27 4.8 3.5 4.5 3.2 3.1 2.0 5.0 2.4 2.0 2.7 2.1 2.7 4.0 4.1 25 2.3 3.8 2.2 3.4 2.1 96 95 113 101 102 100 90 52 87 100 100 92 103 95 d 108 108 100 95 94 87 94 d 85 100 87 89 85 1.3 1.7 1.8 1.9 2.9 1.8 2.2 6.7 2.3 2.8 2.9 2.6 1.6 2.1 3.1 4.4 3.0 2.2 5.1 2.3 2.8 3.6 2.1 3.8 2.9 2.3 524.2-36 ------- TABLE 6. ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF THE METHOD ANALYTES IN REAGENT WATER USING WIDE-BORE CAPILLARY COLUMN 2a Mean Accuracy Mean Accuracy (% of True (% of True Value, 2 |ig/L RSD Value, 2 |ig/L RSD Compound No.b Cone.) (%) Cone.) (%) 1,2-Dichloropropane 20 102 2.2 103 2.9 1,3-Dichloropropane 21 92 3.7 93 3.2 2,2-Dichloropropanec 1,1 -Dichloropropenec cis-1,3-Dichloropropenec trans-1,3-Dichloropropene 25 96 1.7 99 2.1 Ethylbenzene 48 96 9.1 100 4.0 Hexachlorobutadiene 26 91 5.3 88 2.4 Isopropylbenzene 49 103 3.2 101 2.1 4-Isopropyltoluene 50 95 3.6 95 3.1 Methylene chloride 27 e e Naphthalene 51 93 7.6 78 8.3 n-Propylbenzene 52 102 4.9 97 2.1 Styrene 53 95 4.4 104 3.1 1,1,1,2-T etrachloroethane 28 99 2.7 95 3.8 1,1,2,2-T etrachloroethane 29 101 4.6 84 3.6 T etrachlor oethene 30 97 4.5 92 3.3 Toluene 54 105 2.8 126 1.7 1,2,3-Trichlorobenzene 55 90 5.7 78 2.9 1,2,4-Trichlorobenzene 56 92 5.2 83 5.9 1,1,1 -T richloroethane 31 94 3.9 94 2.5 1,1,2-T richloroethane 32 107 3.4 109 2.8 Trichloroethene 33 99 2.9 106 2.5 T richlorofluoromethane 34 81 4.6 48 13 1,2,3-Trichloropropane 35 97 3.9 91 2.8 1,2,4-Trimethylbenzene 57 93 3.1 106 2.2 1,3,5-Trimethylbenzene 58 88 2.4 97 3.2 Vinyl chloride 36 104 3.5 115 14 o-Xylene 59 97 1.8 98 1.7 m-Xylene 60 f f p-Xylene 61 98 2.3 103 1.4 aData obtained using column 2 with the open split interface and an ion trap mass spectrometer (Section 11.3.2) with all method analytes in the same reagent water solution. bDesignation in Figures 1 and 2. cNot measured; authentic standards were not available. dNot found at 0.2 ]ig/L. eNot measured; methylene chloride was in the laboratory reagent blank. fm-Xylene coelutes with and cannot be distinguished from its isomer p-Xylene, No 61. 524.2-37 ------- TABLE 7. ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF METHOD ANALYTES IN REAGENT WATER USING WIDE-BORE CAPILLARY COLUMN 4 Mean Rel. Method True Cone. Std. Dect. Cone. Detected Dev. Limit Compound (Pg/L) (l-ig/L) (%) (l-ig/L) Acetone 1.0 1.6 5.7 0.28 Acrylonitrile 1.0 0.81 8.7 0.22 Allyl chloride 1.0 0.90 4.7 0.13 2-Butanone 2.0 2.7 5.6 0.48 Carbon disulfide 0.20 0.19 15 0.093 Chloroacetonitrile 1.0 0.83 4.7 0.12 1-Chlorobutane 1.0 0.87 6.6 0.18 t-1,2-Dichloro-2-butene 1.0 1.3 8.7 0.36 1,1 -Dichloropropanone 5.0 4.2 7.7 1.0 Diethyl ether 1.0 0.92 9.5 0.28 Ethyl methacrylate 0.20 0.23 3.9 0.028 Hexachloroethane 0.20 0.18 10 0.057 2-Hexanone 1.0 1.1 12 0.39 Methacrylonitrile 1.0 0.92 4.2 0.12 Methylacrylate 1.0 1.2 12 0.45 Methyl iodide 0.20 0.19 3.1 0.019 Methylmethacrylate 1.0 1.0 13 0.43 4-Methyl-2-pentanone 0.40 0.56 9.7 0.17 Methyl-tert-butylether 0.40 0.52 5.6 0.090 Nitrobenzene 2.0 2.1 18 1.2 2-Nitropropane 1.0 0.83 6.2 0.16 Pentachloroethane 0.20 0.23 20 0.14 Propionitrile 1.0 0.87 5.3 0.14 T etrahydrofuran 5.0 3.9 13 1.6 524.2-38 ------- TABLE 8. ACCURACY AND PRECISION FROM FOUR DETERMINATIONS OF METHOD ANALYTES IN THREE WATER MATRICES FORTIFIED AT 20 pg/L Reagent Water Raw Water Tap Water (% of (% of (% of Mean Dev. True Mean Dev. True Mean Dev. True Compound (pg/L) (%) Value) (pg/L) (%) Value) (pg/L) (%) Value) Acetone 19 12 95 21 3.7 105 22 8.2 110 Acrylonitrile 20 4.7 100 22 3.4 110 21 1.3 105 Allyl chloride 20 5.1 100 20 2.8 100 19 3.5 95 2-Butanone 17 11 85 19 7.3 95 17 5.6 85 Carbon disulfide 19 6.4 95 18 2.5 90 18 3.0 90 Chloroacetonitrile 20 4.1 100 23 4.7 115 23 1.3 115 1-Chlorobutane 18 6.4 90 19 2.2 95 17 2.2 85 t-1,2-Dichloro-2-butene 19 4.1 95 22 2.9 110 21 0.90 105 1,1 -Dichloropropanone 20 5.6 100 22 6.4 110 21 7.7 105 Diethyl ether 18 6.7 90 22 3.4 110 22 2.6 110 Ethyl methacrylate 20 3.7 100 23 2.6 115 22 1.8 110 Hexachloroethane 20 6.1 100 21 2.5 105 21 2.0 105 2-Hexanone 19 6.3 95 21 3.8 105 21 4.0 105 Methacrylonitrile 20 3.4 100 23 2.9 115 22 2.0 110 Methylacrylate 20 3.7 100 22 3.1 110 21 2.1 105 Methyl iodide 20 4.4 100 19 3.8 95 19 3.0 95 Methylmethacrylate 20 3.7 100 23 3.3 115 23 2.7 115 4-Methyl-2-pentanone 19 8.7 95 21 5.5 105 22 7.2 110 Methyl-tert-butylether 19 3.5 95 22 2.5 110 22 3.6 110 Nitrobenzene 20 5.4 100 22 4.8 110 21 2.4 105 2-Nitropropane 20 6.1 100 23 5.1 115 22 3.2 110 Pentachloroethane 19 5.2 95 21 2.6 105 22 1.7 110 Propionitrile 20 4.5 100 23 3.9 115 23 2.4 115 T etrahydrofuran 20 2.8 100 24 3.2 120 21 2.9 105 524.2-39 ------- OPTIONAL »-eatxM. FOAM WAP v 0.0. L N-i«m o. o. v/L Jr *j Mm ir 0.0. •• II M KM. 0.0. BIT jp-t*«Af SfWVQI VAIVC SO «AU«H1N6C NEEDLE 0.0. lUMSt SETTlfli 1QML 0.0. t*-«UE7 MM.O.OJ GLASS mT WmWKMMTt 1/11 M. 0.0 yiTAWUSS isx molecular (SIEVE PURGE qam mm PURGE GAS iroi Icontrol FIGURE 1. PURGING DEVICE 524.2-40 ------- MCBN6 MOCEDUK CONSTRUCTION GLASS 1001 ACTIVATE), CHADC0AL7'7( ORAM 11 StUCAGS 1BIAI V ROV-1 GLASS 7 A/TOOT nft-SK?? KSSTANCt M"ITUWNUT ano renus mnura I0UD (DOUIiUTUQ J*/FO0U SSIfANCC iusiiuttis IOUD (SINGL! UTEW ini nil lHWOCOWLtf CONTROLS ^""SIISOJ BiCTHONIC" rerauTUKi CONTMX. \Rfil 9fMr% nniEiB TUHNQ2SCM LlOt IN. IX. LIS M. 0.0. stainless raa FISURE 2. TRAP PACKINGS AM) CONSTRUCTION TO INCLUDE DESORS CAPABILITY 524.2-41 ------- rXCWS 3. NORMALIZED TOTAL ION CURRENT CHIOMATOCIAM RON A VOLATILE COMPOUND CALIBRATION MIXTURE OONTA1NWC 23 «g (S Vf/L) OF MOST COMPOUNDS. THK COKPOIMD IDENTIFICATION NUMBERS All CIVEN IN TABLE 6. m TOT- '* ' */ 11 i"rTV> i Aj>rtf»tAwV| I 100 3:22 r 200 C42 M tl nl i jiflln1 ^ i i ¦ ¦/** IN M 2C42 3t:tt 13:: ------- ncuu 4. AWLinio nisT ucnr Knrores or a total iom cvudct crkohatockam now a volatile ookknmd calimatioh Mirruu OONTAIiraiC 23 n| (5 Pl/L) or EACH OOKPONENT. THE COMPOUND IDOfTIflCATION NUOEKS AXE CIVW 1M TABU 6. ------- |