METHOD 204D--VOLATILE ORGANIC COMPOUNDS EMISSIONS IN UNCAPTURED
STREAM FROM TEMPORARY TOTAL ENCLOSURE

1.	SCOPE AND APPLICATION

1.1	Applicability. This procedure is applicable for
determining the uncaptured volatile organic compounds (VOC)
emissions from a temporary total enclosure (TTE). It is intended
to be used as a segment in the development of liquid/gas or
gas/gas protocols for determining VOC capture efficiency (CE) for
surface coating and printing operations.

1.2	Principle. The amount of uncaptured VOC emissions (F)
from the TTE is calculated as the sum of the products of the VOC
content (CFj), the flow rate (QFj) from each uncaptured emissions
point, and the sampling time (0F) .

1.3	Sampling Requirements. A CE test shall consist of at
least three sampling runs. Each run shall cover at least one
complete production cycle, but shall be at least 3 hours long.
The sampling time for each run need not exceed

8 hours, even if the production cycle has not been completed.
Alternative sampling times may be used with the approval of the
Administrator.

2.	SUMMARY OF METHOD

A gas sample is extracted from the uncaptured exhaust duct
of a TTE through a heated sample line and, if necessary, a glass
fiber filter to a flame ionization analyzer (FIA).


-------
3.	SAFETY

Because this procedure is often applied in highly explosive
areas, caution and care should be exercised in choosing,
installing, and using the appropriate equipment.

4.	EQUIPMENT AND SUPPLIES

Mention of trade names or company products does not
constitute endorsement. All gas concentrations (percent, ppm)
are by volume, unless otherwise noted.

4.1 Gas VOC Concentration. A schematic of the measurement
system is shown in Figure 204D-1. The main components are as
follows:

4.1.1	Sample Probe. Stainless steel or equivalent. The
probe shall be heated to prevent VOC condensation.

4.1.2	Calibration Valve Assembly. Three-way valve assembly
at the outlet of the sample probe to direct the zero and
calibration gases to the analyzer. Other methods, such as
quick-connect lines, to route calibration gases to the outlet of
the sample probe are acceptable.

4.1.3	Sample Line. Stainless steel or Teflon tubing to
transport the sample gas to the analyzer. The sample line must
be heated to prevent condensation.

4.1.4	Sample Pump. A leak-free pump, to pull the sample
gas through the system at a flow rate sufficient to minimize the
response time of the measurement system. The components of the


-------
pump that contact the gas stream shall be constructed of
stainless steel or Teflon. The sample pump must be heated to
prevent condensation.

4.1.5	Sample Flow Rate Control. A sample flow rate control
valve and rotameter, or equivalent, to maintain a constant
sampling rate within 10 percent. The flow control valve and
rotameter must be heated to prevent condensation. A control
valve may also be located on the sample pump bypass loop to
assist in controlling the sample pressure and flow rate.

4.1.6	Sample Gas Manifold. Capable of diverting a portion
of the sample gas stream to the FIA, and the remainder to the
bypass discharge vent. The manifold components shall be
constructed of stainless steel or Teflon. If emissions are to be
measured at multiple locations, the measurement system shall be
designed to use separate sampling probes, lines, and pumps for
each measurement location and a common sample gas manifold and
FIA. The sample gas manifold and connecting lines to the FIA
must be heated to prevent condensation.

4.1.7	Organic Concentration Analyzer. An FIA with a span
value of 1.5 times the expected concentration as propane;
however, other span values may be used if it can be demonstrated
to the Administrator's satisfaction that they would provide more
accurate measurements. The system shall be capable of meeting or
exceeding the following specifications:

4.1.7.1 Zero Drift. Less than ±3.0 percent of the span


-------
value.

4.1.7.2	Calibration Drift. Less than ±3.0 percent of the
span value.

4.1.7.3	Calibration Error. Less than ±5.0 percent of the
calibration gas value.

4.1.7.4	Response Time. Less than 30 seconds.

4.1.8 Integrator/Data Acquisition System. An analog or
digital device or computerized data acquisition system used to
integrate the FIA response or compute the average response and
record measurement data. The minimum data sampling frequency for
computing average or integrated values is one measurement value
every 5 seconds. The device shall be capable of recording
average values at least once per minute.

4.2	Uncaptured Emissions Volumetric Flow Rate.

4.2.1	Method 2 or 2A Apparatus. For determining volumetric
flow rate.

4.2.2	Method 3 Apparatus and Reagents. For determining
molecular weight of the gas stream. An estimate of the molecular
weight of the gas stream may be used if approved by the
Administrator.

4.2.3	Method 4 Apparatus and Reagents. For determining
moisture content, if necessary.

4.3	Temporary Total Enclosure. The criteria for designing
an acceptable TTE are specified in Method 204.

5. REAGENTS AND STANDARDS


-------
5.1 Calibration and Other Gases. Gases used for
calibration, fuel, and combustion air (if required) are contained
in compressed gas cylinders. All calibration gases shall be
traceable to National Institute of Standards and Technology
standards and shall be certified by the manufacturer to
±1 percent of the tag value. Additionally, the manufacturer of
the cylinder should provide a recommended shelf life for each
calibration gas cylinder over which the concentration does not
change more than ±2 percent from the certified value. For
calibration gas values not generally available, dilution systems
calibrated using Method 205 may be used. Alternative methods for
preparing calibration gas mixtures may be used with the approval
of the Administrator.

5.1.1	Fuel. The FIA manufacturer's recommended fuel should
be used. A 40 percent H2/60 percent He or

40 percent H2/60 percent N2 gas mixture is recommended to avoid
an oxygen synergism effect that reportedly occurs when oxygen
concentration varies significantly from a mean value. Other
mixtures may be used provided the tester can demonstrate to the
Administrator that there is no oxygen synergism effect

5.1.2	Carrier Gas. High purity air with less than

1 ppm of organic material (as propane or carbon equivalent)
or less than 0.1 percent of the span value, whichever is greater.

5.1.3	FIA Linearity Calibration Gases. Low-, mid-, and
high-range gas mixture standards with nominal propane


-------
concentrations of 20-30, 45-55, and 70-80 percent of the span
value in air, respectively. Other calibration values and other
span values may be used if it can be shown to the Administrator's
satisfaction that equally accurate measurements would be
achieved.

5.2 Particulate Filter. An in-stack or an out-of-stack
glass fiber filter is recommended if exhaust gas particulate
loading is significant. An out-of-stack filter must be heated to
prevent any condensation unless it can be demonstrated that no
condensation occurs.

6. QUALITY CONTROL

6.1	Required instrument quality control parameters are
found in the following sections:

6.1.1	The FIA system must be calibrated as specified in
section 7.1.

6.1.2	The system drift check must be performed as specified
in section 7.2.

6.1.3	The system check must be conducted as specified in
section 7.3.

6.2	Audits.

6.2.1	Analysis Audit Procedure. Immediately before each
test, analyze an audit cylinder as described in section 7.2. The
analysis audit must agree with the audit cylinder concentration
within 10 percent.

6.2.2	Audit Samples and Audit Sample Availability. Audit


-------
samples will be supplied only to enforcement agencies for
compliance tests. The availability of audit samples may be
obtained by writing:

Source Test Audit Coordinator (STAC) (MD-77B)

Quality Assurance Division

Atmospheric Research and Exposure Assessment Laboratory

U.S. Environmental Protection Agency

Research Triangle Park, NC 27711

or by calling the STAC at (919) 541-7834. The request for
the audit sample must be made at least 30 days prior to the
scheduled compliance sample analysis.

6.2.3 Audit Results. Calculate the audit sample
concentration according to the calculation procedure described in
the audit instructions included with the audit sample. Fill in
the audit sample concentration and the analyst's name on the
audit response form included with the audit instructions. Send
one copy to the EPA Regional Office or the appropriate
enforcement agency, and a second copy to the STAC. The EPA
Regional Office or the appropriate enforcement agency will report
the results of the audit to the laboratory being audited.

Include this response with the results of the compliance samples
in relevant reports to the EPA Regional Office or the appropriate
enforcement agency.

7. CALIBRATION AND STANDARDIZATION

7.1 FIA Calibration and Linearity Check. Make necessary


-------
adjustments to the air and fuel supplies for the FIA and ignite
the burner. Allow the FIA to warm up for the period recommended
by the manufacturer. Inject a calibration gas into the
measurement system and adjust the back-pressure regulator to the
value required to achieve the flow rates specified by the
manufacturer. Inject the zero- and the high-range calibration
gases and adjust the analyzer calibration to provide the proper
responses. Inject the low- and mid-range gases and record the
responses of the measurement system. The calibration and
linearity of the system are acceptable if the responses for all
four gases are within 5 percent of the respective gas values. If
the performance of the system is not acceptable, repair or adjust
the system and repeat the linearity check. Conduct a calibration
and linearity check after assembling the analysis system and
after a major change is made to the system.

7.2 Systems Drift Checks. Select the calibration gas
concentration that most closely approximates that of the
uncaptured gas emissions concentration to conduct the drift
checks. Introduce the zero and calibration gases at the
calibration valve assembly and verify that the appropriate gas
flow rate and pressure are present at the FIA. Record
the measurement system responses to the zero and calibration
gases. The performance of the system is acceptable if the
difference between the drift check measurement and the value
obtained in section 7.1 is less than 3 percent of the span value.


-------
Alternatively, recalibrate the FIA as in section 7.1 and report
the results using both sets of calibration data (i.e., data
determined prior to the test period and data determined following
the test period). The data that results in the lowest CE value
shall be reported as the results for the test run. Conduct a
system drift check at the end of each run.

7.3 System Check. Inject the high-range calibration gas at
the inlet of the sampling probe and record the response. The
performance of the system is acceptable if the measurement system
response is within 5 percent of the value obtained in section 7.1
for the high-range calibration gas. Conduct a system check
before each test run.

8. PROCEDURE

8.1	Determination of Volumetric Flow Rate of Uncaptured
Emissions

8.1.1	Locate all points where uncaptured emissions are
exhausted from the TTE. Using Method 1, determine the sampling
points. Be sure to check each site for cyclonic or swirling
flow.

8.1.2	Measure the velocity at each sampling site at least
once every hour during each sampling run using Method 2 or 2A.

8.2	Determination of VOC Content of Uncaptured Emissions

8.2.1 Analysis Duration. Measure the VOC responses at each

uncaptured emission point during the entire test run or, if
applicable, while the process is operating. If there are


-------
multiple emission locations, design a sampling system to allow a
single FIA to be used to determine the VOC responses at all
sampling locations.

8.2.2 Gas VOC Concentration.

8.2.2.1	Assemble the sample train as shown in

Figure 204D-1. Calibrate the FIA and conduct a system check
according to the procedures in sections 7.1 and 7.3,
respectively.

8.2.2.2	Install the sample probe so that the probe is
centrally located in the stack, pipe, or duct, and is sealed
tightly at the stack port connection.

8.2.2.3	Inject zero gas at the calibration valve assembly.
Allow the measurement system response to reach zero. Measure the
system response time as the time required for the system to reach
the effluent concentration after the calibration valve has been
returned to the effluent sampling position.

8.2.2.4	Conduct a system check before, and a system drift
check after, each sampling run according to the procedures in
sections 7.2 and 7.3. If the drift check following a run
indicates unacceptable performance (see section 7.3), the run is
not valid. Alternatively, recalibrate the FIA as in section 7.1
and report the results using both sets of calibration data (i.e.,
data determined prior to the test period and data determined
following the test period). The data that results in the lowest


-------
CE value shall be reported as the results for the test run. The
tester may elect to perform system drift checks during the run
not to exceed one drift check per hour.

8.2.2.5	Verify that the sample lines, filter, and pump
temperatures are 120 ± 5°C.

8.2.2.6	Begin sampling at the start of the test period and
continue to sample during the entire run. Record the starting
and ending times and any required process information, as
appropriate. If multiple emission locations are sampled using a
single FIA, sample at each location for the same amount of time
(e.g., 2 min.) and continue to switch from one location to
another for the entire test run. Be sure that total sampling
time at each location is the same at the end of the test run.
Collect at least four separate measurements from each sample
point during each hour of testing. Disregard the response
measurements at each sampling location until 2 times the response
time of the measurement system has elapsed. Continue sampling
for at least 1 minute and record the concentration measurements.

8.2.3 Background Concentration.

8.2.3.1	Locate all natural draft openings (NDO's) of the
TTE. A sampling point shall be at the center of each NDO, unless
otherwise approved by the Administrator. If there are more than
six NDO's, choose six sampling points evenly spaced among the
NDO's.

8.2.3.2	Assemble the sample train as shown in


-------
Figure 204D-2. Calibrate the FIA and conduct a system check
according to the procedures in sections 7.1 and 7.3.

8.2.3.3	Position the probe at the sampling location.

8.2.3.4	Determine the response time, conduct the system
check, and sample according to the procedures described in
sections 8.2.2.3 through 8.2.2.6.

8.2.4 Alternative Procedure. The direct interface sampling
and analysis procedure described in section 7.2 of Method 18 may
be used to determine the gas VOC concentration. The system must
be designed to collect and analyze at least one sample every
10 minutes. If the alternative procedure is used to determine
the VOC concentration of the uncaptured emissions in a gas/gas
protocol, it must also be used to determine the VOC concentration
of the captured emissions. If a tester wishes to conduct a
liquid/gas protocol using a gas chromatograph, the tester must
use Method 204F for the liquid steam. A gas chromatograph is not
an acceptable alternative to the FIA in Method 204A.

9. DATA ANALYSIS AND CALCULATIONS

9.1 Nomenclature
A± = area of NDO i, ft2.

An = total area of all NDO's in the enclosure, ft2.

CB1 = corrected average VOC concentration of background
emissions at point i, ppm propane.

CB = average background concentration, ppm propane.

CDH = average measured concentration for the drift check
calibration gas, ppm propane.


-------
CD0 = average system drift check concentration for zero
concentration gas, ppm propane.

CFj = corrected average VOC concentration of uncaptured
emissions at point j, ppm propane.

CH = actual concentration of the drift check calibration gas,
ppm propane.

C± = uncorrected average background VOC concentration at point
i, ppm propane.

Cj = uncorrected average VOC concentration measured at point j,
ppm propane.

F = total VOC content of uncaptured emissions, kg.

Kx = 1.830 x 10~6 kg/ (m3-ppm) .

n = number of measurement points.

QFj = average effluent volumetric flow rate corrected to

standard conditions at uncaptured emissions point j,
m3/min.

0F = total duration of uncaptured emissions sampling run, min.

9.2 Calculations

9.2.1	Total Uncaptured VOC Emissions.

F = £ (CFj - Cb) Qfj 0/ K1	Eq. 204D-1

y=i

9.2.2	VOC Concentration of the Uncaptured Emissions at
Point j.

CH

CFj - (Cj - CDO) — ——	Eq. 204D-2

DH DO

9.2.3 Background VOC Concentration at Point i.


-------
Eq. 204D-3

9.2.4 Average Background Concentration.

n

£ CB, A

Eq. 204D-4

NOTE: If the concentration at each point is within 20 percent of
the average concentration of all points, use the arithmetic
average.

10.	METHOD PERFORMANCE.

The measurement uncertainties are estimated for each
uncaptured emission point as follows: QFj = ±5.5 percent and
CFj = ±5.0 percent. Based on these numbers, the probable
uncertainty for F is estimated at about ±7.4 percent.

11.	DIAGRAMS


-------
w SAMPLE
r BYPASS

FIA

^ EXHAUST

EXHAUST

DATA
ACQUISITION
SYSTEM

CHART
flECOHDER

irc(nnt measurement system.


-------
Figure 204D-2. Background measurement system.


-------